Entris
®
Benefits
– Monolithic weigh cell for long-term high accuracy
– Built-in applications and function keys for convenient
operation
– Backlit, high-contrast display for user-friendly
guidance
Product Information
Get the reliability and quality of a Sartorius balance, but only pay for the features you really need.
®
Entris
has been specially designed to deliver effective and reliable weighing results in your daily work.
The easy and clearly structured user interface, the logical key assignment and excellent readability are
ideal features for error-free operation.
Laboratory Balances
Technical Specifications
Specifications
Int. weight circuit
calibration
All models with the designation
ENTRIS...i-1xx are equipped with
an internal weight
Mains connection,
voltage, frequency
via desktop power supply,
100-240 V AC, ±10%, 50 – 60 Hz
Power consumption maximum 16 VA; average 8 VA
(including power supply)
Operating time with
approx. 35 h
external battery YRB11Z
(display backlighting on)
Standard Equipment
Levelling Glass level indicator with air
bubble for centering
Calibration External calibration
1)
Selectable weight units
Grams, Kilograms, Carats, Pounds,
Ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwan taels, Grains,
Pennyweights, Milligrams, Parts per
pound, China taels, Mommes, Austrian
carats, Tola, Baht, Mesghal, Tons,
Pounds: ounces, Newton
Interface RS232C 25-pin
Display White backlit, high-contrast
display with 15 mm digits
Standard built-in
applications
Weighing, Density, Percentage,
Counting, Animal Weighing,
Conversion
Languages English, French, German, Italian,
Polish, Russian, Spanish
Anti-theft lock Lockdown capability for cable
or chain
Underfloor weighing Integrated
Ambient Conditions
The specifications apply when the following ambient conditions
are in place:
Environment for indoor use only
Ambient temperature* +10° C … +30° C (+50° F ... +86° F)
Operational capacity guaranteed between
+5° C … +40° C (+41° F … +104° F)
Storage and shipping –10 ° C ... +60 ° C (+14° F ... +140° F)
Elevation up to 3000 m above sea level
Relative humidity 15% to 80% for temperatures up to
31 °C; non-condensing, decreasing
linearly to 50% relative humidity at
40 °C and 20% at 50 °C
Safety of electrical
equipment
in accordance with
EN 61010-1/IEC 61010-1
Safety requirements for electrical
equipment for mea surement,
control, and laboratory use – Part 1:
General requirements
Electromagnetic
compatibility
in accordance with
EN 61326-1/IEC 61326-1
Electrical equipment for measure-
ment, control, and laboratory
use – EMC requirements - Part 1:
General requirements
Defined immunity to
Suitable for use in industrial areas
interference
Interference emission Class B (suitable for use in residential
areas and areas that are connected
to a low voltage network that also
supplies residential buildings). The
device can therefore be used in both
areas.
1)
Depending on country-specific model
Design 1 Design 2 Design 3 Design 4
Analytical Balances Entris
®
Model ENTRIS224i-1x
1)
ENTRIS124i-1x
1)
ENTRIS64i-1x
1)
Design 1 1 1
Weighing capacity g 220 120 60
Readability mg 0.1 0.1 0.1
Repeatability (standard deviation) mg 0.1 0.1 0.1
Linearity deviation mg 0.2 0.2 0.2
Sensitivity drift between +10 °C and +30 °C ± ppm/K 3 3 3
Typical stabilization time s 2.5 2.5 2.5
Weighing pan size mm
d 90 d 90 d 90
Weighing chamber height* mm 230 230 230
Net weight, approx. kg 4.8 4.8 4.8
Calibration Internal Internal Internal
Precision Balances Entris
Model ENTRIS
®
623i-1x
1)
ENTRIS
423i-1x
1)
ENTRIS
323i-1x
1)
ENTRIS
153i-1x
1)
ENTRIS
6202i-1x
ENTRIS
1)
4202i-1x
ENTRIS
1)
3202i-1x
ENTRIS
1)
2202i-1x
Design 2 2 2 2 4 4 4 4
Weighing capacity g 620 420 320 150 6,200 4,200 3,200 2,200
Readability mg 1 1 1 1 10 10 10 10
Repeatability (standard deviation) mg 1 1 1 1 10 10 10 10
Linearity deviation mg 2 2 2 2 30 30 30 30
Sensitivity drift between
± ppm/K 3 3 3 3 4 4 4 4
+10 °C and +30 °C
Typical stabilization time s 1.0 1.0 1.1 1.3 1.5 1.5 1.5 1.5
Weighing pan size mm
d 115 d 115 d 115 d 115
180 × 180 180 × 180 180 × 180 180 × 180
Weighing chamber height* mm 50 50 50 50 – – – –
Net weight, approx. kg 3.6 3.6 3.6 3.0 3.5 3.5 3.5 3.5
Calibration Internal Internal Internal Internal Internal Internal Internal Internal
1)
Model ENTRIS822i-1x
1)
ENTRIS8201i-1x
1)
ENTRIS5201i-1x
1)
ENTRIS2201i-1x
Design 3 4 4 4
Weighing capacity g 820 8,200 5,200 2,200
Readability mg 10 100 100 100
Repeatability (standard deviation) mg 10 100 100 100
Linearity deviation mg 30 300 300 300
Sensitivity drift between
± ppm/K 4 8 8 8
+10 °C and +30 °C
Typical stabilization time s 1.5 1.5 1.5 1.5
Weighing pan size mm
d 150
180 × 180 180 × 180 180 × 180
Net weight, approx. kg 2.6 3.5 3.5 3.5
Calibration Internal Internal Internal Internal
* Upper edge of the weighing pan to the lower edge of the upper draft shield panel
1)
Possible terms for country-specific models:
x = S: Standard balances without country-specific additions
x = SUS: Standard balances with country-specific additions for USA
Laboratory Balances 3
1)