Sartorius ENTRIS64I-1SUS, ENTRIS8201-1SUS, ENTRIS8201I-1SUS, ENTRIS623I-1SUS, ENTRIS64-1SUS Data Sheet

...
Entris
®
– Monolithic weigh cell for long-term high accuracy – Built-in applications and function keys for convenient
operation
– Backlit, high-contrast display for user-friendly
guidance

Product Information

Get the reliability and quality of a Sartorius balance, but only pay for the features you really need.
®
Entris
has been specially designed to deliver effective and reliable weighing results in your daily work. The easy and clearly structured user interface, the logical key assignment and excellent readability are ideal features for error-free operation.
Laboratory Balances

Technical Specifications

Specifications Int. weight circuit
calibration
All models with the designation ENTRIS...i-1xx are equipped with an internal weight
Mains connection, voltage, frequency
via desktop power supply, 100-240 V AC, ±10%, 50 – 60 Hz
Power consumption maximum 16 VA; average 8 VA
(including power supply)
Operating time with
approx. 35 h external battery YRB11Z (display backlighting on)
Standard Equipment Levelling Glass level indicator with air
bubble for centering Calibration External calibration
1)
Selectable weight units
Grams, Kilograms, Carats, Pounds,
Ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwan taels, Grains,
Pennyweights, Milligrams, Parts per
pound, China taels, Mommes, Austrian
carats, Tola, Baht, Mesghal, Tons,
Pounds: ounces, Newton Interface RS232C 25-pin Display White backlit, high-contrast
display with 15 mm digits Standard built-in
applications
Weighing, Density, Percentage,
Counting, Animal Weighing,
Conversion Languages English, French, German, Italian,
Polish, Russian, Spanish Anti-theft lock Lockdown capability for cable
or chain Underfloor weighing Integrated
Ambient Conditions The specifications apply when the following ambient conditions
are in place: Environment for indoor use only Ambient temperature* +10° C … +30° C (+50° F ... +86° F) Operational capacity guaranteed between
+5° C … +40° C (+41° F … +104° F) Storage and shipping –10 ° C ... +60 ° C (+14° F ... +140° F) Elevation up to 3000 m above sea level Relative humidity 15% to 80% for temperatures up to
31 °C; non-condensing, decreasing
linearly to 50% relative humidity at
40 °C and 20% at 50 °C Safety of electrical
equipment
in accordance with
EN 61010-1/IEC 61010-1
Safety requirements for electrical
equipment for mea surement,
control, and laboratory use – Part 1:
General requirements Electromagnetic
compatibility
in accordance with
EN 61326-1/IEC 61326-1
Electrical equipment for measure-
ment, control, and laboratory
use – EMC requirements - Part 1:
General requirements Defined immunity to
Suitable for use in industrial areas interference
Interference emission Class B (suitable for use in residential
areas and areas that are connected
to a low voltage network that also
supplies residential buildings). The
device can therefore be used in both
areas.
1)
Depending on country-specific model
Design 1 Design 2 Design 3 Design 4
Analytical Balances Entris
®
Model ENTRIS224i-1x
1)
ENTRIS124i-1x
1)
ENTRIS64i-1x
1)
Design 1 1 1 Weighing capacity g 220 120 60 Readability mg 0.1 0.1 0.1 Repeatability (standard deviation) mg 0.1 0.1 0.1 Linearity deviation mg 0.2 0.2 0.2
Sensitivity drift between +10 °C and +30 °C ± ppm/K 3 3 3 Typical stabilization time s 2.5 2.5 2.5 Weighing pan size mm
d 90 d 90 d 90 Weighing chamber height* mm 230 230 230 Net weight, approx. kg 4.8 4.8 4.8 Calibration Internal Internal Internal
Precision Balances Entris
Model ENTRIS
®
623i-1x
1)
ENTRIS 423i-1x
1)
ENTRIS 323i-1x
1)
ENTRIS 153i-1x
1)
ENTRIS 6202i-1x
ENTRIS
1)
4202i-1x
ENTRIS
1)
3202i-1x
ENTRIS
1)
2202i-1x
Design 2 2 2 2 4 4 4 4 Weighing capacity g 620 420 320 150 6,200 4,200 3,200 2,200 Readability mg 1 1 1 1 10 10 10 10 Repeatability (standard deviation) mg 1 1 1 1 10 10 10 10 Linearity deviation mg 2 2 2 2 30 30 30 30 Sensitivity drift between
± ppm/K 3 3 3 3 4 4 4 4
+10 °C and +30 °C Typical stabilization time s 1.0 1.0 1.1 1.3 1.5 1.5 1.5 1.5 Weighing pan size mm
d 115 d 115 d 115 d 115
180 × 180 180 × 180 180 × 180 180 × 180 Weighing chamber height* mm 50 50 50 50 – Net weight, approx. kg 3.6 3.6 3.6 3.0 3.5 3.5 3.5 3.5 Calibration Internal Internal Internal Internal Internal Internal Internal Internal
1)
Model ENTRIS822i-1x
1)
ENTRIS8201i-1x
1)
ENTRIS5201i-1x
1)
ENTRIS2201i-1x
Design 3 4 4 4 Weighing capacity g 820 8,200 5,200 2,200 Readability mg 10 100 100 100 Repeatability (standard deviation) mg 10 100 100 100 Linearity deviation mg 30 300 300 300 Sensitivity drift between
± ppm/K 4 8 8 8
+10 °C and +30 °C Typical stabilization time s 1.5 1.5 1.5 1.5 Weighing pan size mm
d 150
180 × 180 180 × 180 180 × 180 Net weight, approx. kg 2.6 3.5 3.5 3.5 Calibration Internal Internal Internal Internal
* Upper edge of the weighing pan to the lower edge of the upper draft shield panel
1)
Possible terms for country-specific models:
x = S: Standard balances without country-specific additions x = SUS: Standard balances with country-specific additions for USA