Sartorius Element Series,GD502,ELT103,ELT602,ELT402,ELT202,ELT6001,ELT4001,ELT2001,ELT1001,ELT601,ELT6000 Operating Instructions Manual

Operating Instructions
Sartorius Element Series
Electronic Precision Balances, Electronic Gem Scale
98648-013-44
Contents Warnings and
Safety Precautions
3 Getting Started
7 Operation 7 Basic Weighing Function 7 Description of the Keys
8 Application Programs 8 Toggling between Weight Units
9 Counting 10 Weighing in Percent 11 “Hold” Display 12 Totalizing 13 Specific Gravity
14 Calibration/Span Adjustment
15 Configuration (Setup Menu)
16 Error Codes
17 Overview 17 Specifications 18 Accessories (Options) 19 C Marking
Safety Information
§ To prevent damage to the equipment,
please read these operating instructions carefully before using your balance/scale.
! Do not use this equipment in
hazardous areas.
! Make sure the voltage rating printed
on the power supply is identical to your local line voltage.
! Use only commercially available non-
rechargeable or rechargeable battery: 8+ AA, Mignon (not included). The rechargeable batteries are not charged when you use the AC adapter. Please use the charger to recharge the batteries outside the balance/scale.
– The balance/scale is energized at all
times unless you disconnect the AC adapter and, if connected, the battery.
– Protect the AC adapter from contact
with liquid
! Exposure to excessive electromagnetic
interference can cause the readout value to change. Once the disturbance has ceased, the instrument can be used again in accordance with its intended purpose.
Installation
– It is recommended to connect Sartorius
accessories and options, as these are optimally designed for use with your balance.
– Do not open the balance/scale housing
as this may void the manufacturer’s warranty.
2
Getting Started
Equipment Supplied
– Balance/scale – Weighing pan – Plug-in AC adapter – Level indicator and adjustable feet
Additional equipment with models ELT602, ELT402, ELT202, ELT1001, ELT601:
– Calibration weight
Additional equipment with model ELT103, GD502: – Calibration weight – Round glass draft shield – Only with GD502: Gem tray
Installation
Choose a location that is not subject to the
following negative influences: – Heat (heater or direct sunlight) – Drafts from open windows and doors – Extreme vibrations during weighing – Excessive moisture
Setting Up the Balance/Scale
§ Place the components on the balance/scale in the
following order: – Weighing pan – Round glass draft shield on model ELT103, GD502
3
4
! Use only original Sartorius AC adapters:
– for Asia: 400081.1 – for the U.S.: 400314.1
Connecting the Balance/Scale to AC Power
§ Insert plug into the jack (located on back of the
balance/scale)
§ Plug the AC adapter into an electrical outlet
! LISTED power supply 11 V–21 V compliant with
NEC Class 2 output.
Below-Balance/Scale Weighing
A port for a below-balance/scale weighing hanger is located on the bottom of the balance/scale.
§ Open the cover plate on the bottom of the
balance/scale.
§ Attach the sample (e.g., using a suspension wire)
to the hanger.
$ Install a shield for protection against drafts if
necessary.
Using a Non-Rechargeable/Rechargeable Battery
(not for model ELT103, GD502)
$ A non-rechargeable or rechargeable battery is not
included with the equipment supplied.
! Use only a commercially available non-rechargeable
(8+ AA/Mignon) or rechargeable battery.
! When using a rechargeable battery, be sure to
recharge it outside the balance/scale using a sepa-
rate charger.
Lay the balance/scale on its side
Lift the compartment cover
Insert the non-rechargeable (8+ AA/Mignon)
or rechargeable batteries into the compartment.
$
Make sure to connect the positive and negative poles
correctly.
Close the battery compartment: press down on
the cover until it clicks into place.
! All used batteries are classified as waste that requires
special handling (not “household” waste). Dispose
of rechargeable batteries in accordance with the
applicable special waste disposal regulations.
5
Leveling the Balance/Scale
§ Always level the balance/scale again any time after it
has been moved to a different location. Example: moving bubble from R to L
§ Turn the feet as shown in the diagram until the
air bubble is centered within the circle of the level indicator.
> In most cases this will require several adjustment
steps.
Removing Weights for Calibration/Span Adjustment
(only for models ELT103, ELT602, ELT402, ELT202, ELT1001, ELT601, GD502)
§ Grasp the tab to turn and pull out the weight
compartment.
$ Follow instructions on page 14 for calibration/span
adjustment.
6