Sartorius Cubis Series User Manual

User Manual
Sartorius Cubis Series
Precision and Analytical Balances MSU Models
Contents
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Overview of Equipment. . . . . . . . . . . . . . . . . . . . . . . . . 7
Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Equipment Supplied. . . . . . . . . . . . . . . . . . . . . . . . . . 10
Moving the Balance. . . . . . . . . . . . . . . . . . . . . . . . . . 15
Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . 16
Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Anti-theft Locking Device . . . . . . . . . . . . . . . . . . . . . 19
Below-Cell Weighing . . . . . . . . . . . . . . . . . . . . . . . . . 26
Using the Balance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Turning the Balance On/Off . . . . . . . . . . . . . . . . . . . . 29
Operating Design: Q-Guide . . . . . . . . . . . . . . . . . . . . 29
Using the Display and Control Unit. . . . . . . . . . . . . . . 30
Setting the Language. . . . . . . . . . . . . . . . . . . . . . . . . 31
Quick Guide: First Weighing. . . . . . . . . . . . . . . . . . . . 31
Display in Operating Mode . . . . . . . . . . . . . . . . . . . . 32
Display in Setup Mode . . . . . . . . . . . . . . . . . . . . . . . . 33
Activating/Switching Users. . . . . . . . . . . . . . . . . . . . . 34
System Settings (Menu) . . . . . . . . . . . . . . . . . . . . . . . . 35
Level the Balance (Q-Level) . . . . . . . . . . . . . . . . . . . . 36
Calibration/Adjustment Data . . . . . . . . . . . . . . . . . . . 37
Configure Timer Controlled Actions . . . . . . . . . . . . . . 38
Device Information . . . . . . . . . . . . . . . . . . . . . . . . . . 40
View Alibi Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Device Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Task Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Using Applications with the Factory Settings . . . . . . . 54
Creating New Tasks (Configuration) . . . . . . . . . . . . . . 54
Configuring Printouts . . . . . . . . . . . . . . . . . . . . . . . . 57
Combining Applications into One Task . . . . . . . . . . . . 62
Executing Tasks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Weighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Mass Unit Conversion . . . . . . . . . . . . . . . . . . . . . . . . . . 65
SQmin Minimum Weighing . . . . . . . . . . . . . . . . . . . . . . 68
Individual Identifiers. . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Density Determination . . . . . . . . . . . . . . . . . . . . . . . . . 73
Statistics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Formulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Weighing in Percent . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Timer-controlled Functions. . . . . . . . . . . . . . . . . . . . . . 97
Totalizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
2 Cubis MSU User Manual
DKD Measurement Uncertainty. . . . . . . . . . . . . . . . . . 101
Second Tare Memory (Preset Tare) . . . . . . . . . . . . . . . 103
Parts Counting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Checkweighing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Importing/Exporting Data . . . . . . . . . . . . . . . . . . . . . . 113
Calibration and Adjustment . . . . . . . . . . . . . . . . . . . . 116
Calibration/Adjustment Using Internal
Calibration Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Calibration/Adjustment Using External
Calibration Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
User Management . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
User Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Creating User Profiles. . . . . . . . . . . . . . . . . . . . . . . . . . 119
Activating Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
Editing User Profiles. . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
USB Port (PC). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
PS2 Interfaces for Barcode Scanner or Keyboard . . . . . . 127
Interfaces (RS-232) 25-pin and 9-pin . . . . . . . . . . . . . . 128
Configuring Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . 131
Bluetooth®Interface (COM C, Optional) . . . . . . . . . . . . 134
Network Interface (Ethernet) . . . . . . . . . . . . . . . . . . . . 136
Data Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Data Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Updating the Software . . . . . . . . . . . . . . . . . . . . . . . . 148
Error and Status Messages . . . . . . . . . . . . . . . . . . . . . 150
GPL License. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 152
Shipping the Balance . . . . . . . . . . . . . . . . . . . . . . . . . 155
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Balance Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
Declarations of Conformity. . . . . . . . . . . . . . . . . . . . . 177
EC Type-Approval Certificate . . . . . . . . . . . . . . . . . . . 180
Plates and Markings . . . . . . . . . . . . . . . . . . . . . . . . . . 181
Contents
Cubis MSU User Manual 3
Notes on Using this Manual
t Please read this entire manual carefully and completely before using the device. t Read the safety precautions carefully. t This manual is part of the product. Keep it in a safe and easily accessible
location.
t If the manual should be lost or misplaced, please contact Sartorius for
a replacement or download the latest manual from our website: www.sartorius.com
Symbols and Signs
The following symbols are used in this manual:
2
Warning symbol for various types of dangers.
These symbols are explained in more detail in Section “Safety Instructions.”
h
This symbol indicates useful information and tips.
This symbol indicates notes on use of the balance in legal metrology within the scope of validity of Council Directive No. 90/384/EEC, replaced by 2009/23/EC (models MS...–CE...).
H This symbol means that the USER key should be pressed.
D This symbol means that the TASK key should be pressed.
When individual functions are displayed, press the associated softkey.
t Indicates a required action
y Describes the result of an action
1. If a procedure has multiple steps...
2. ... the steps are numbered consecutively.
Indicates an item in a list
h
Technical advice/hotline:
Phone (in Germany): +49.551.308.4440 Fax: +49.551.308.4449
Information for the User
4 Cubis MSU User Manual
Safety Instructions
This device complies with European Council Directives as well as international regulations and standards for electrical equipment, electromagnetic compatibility, and the stipulated safety requirements. Improper use or handling can, however, result in damage and/or injury. This device should only be operated by trained personnel. Operators must read these installation instructions, particularly the safety informa­tion, and must be familiar with the operation of the equipment. The operator is required to supplement these safety precautions as appropriate. The operating personnel must be provided with the appropriate training. The requirements per­taining to applicable installation regulations must be followed when using electrical equipment in systems and environmental conditions with increased safety require­ments. Relevant laws, standards, regulations, guidelines, and environmental protection laws valid in your country must be observed. Always keep the equipment and balance freely accessible. Any installation work or balance usage that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer’s warranty.
Danger of explosion!
1
Do not use this equipment in hazardous areas in which explosive materials are present.
Danger of electrical shock!
2
Make sure that the voltage rating printed on the AC adaptor is identical to your local line voltage. The wall outlet must have a protective grounding conductor. The operating instructions included with the AC adaptor must be followed.
2
Do not switch on the equipment if the AC adaptor or power cord are damaged. If the balance, the AC adaptor or any of the cables are damaged, disconnect the equipment from the power supply and secure it so that it cannot be turned on.
2
Chemicals (e.g., gases or liquids) that can corrode and damage the inside or outside of the balance, AC adaptor, power cable, or peripherals must be kept away from the equipment.
2
Do not operate the balance unless the housing and the display are undamaged and the housing is closed so that inadvertent contact with parts inside the equipment is not possible. Make sure that no liquids penetrate the balance.
2
Installation note: The operator shall be solely responsible for any modifications to the equipment and for connecting any cables or equipment not supplied by Sartorius. Information on operational quality is available upon request from Sartorius. You should only use peripherals and options available from Sartorius.
Note on cable quality: CE conformity is only guaranteed with well shielded cables. The shielding braid of the cable must be made of metal with at least 80% coverage and also covered with metal foil as much as possible. The shielding on both sides on the full metal or metal connector casing, as well as on the casing inlet, must be high-frequency compliant and have a flat design.
Safety Instructions
Cubis MSU User Manual 5
Safety Instructions
Power to the device can only be switched off by pulling the plug or disconnecting the mains connection line, or by unscrewing the four-pin plug at the DC supply cable for the balance.
3
Equipment may be damaged by objects with points or sharp edges (e. g., ballpoint pens).
Use only your fingertips to operate the buttons and softkeys.
Liquid penetration can damage the electrical components. Use only a slightly damp cloth for cleaning. See Section “Care and Maintenance” for
instructions on cleaning the device. Do not compromise the IP protection of the balance. Do not open the balance housing. If the seal is broken, this will result in forfeiture
of all claims under the manufacturer’s warranty. The device should only be opened by specialized personnel trained by Sartorius.
Disconnect the balance from the wall outlet prior to moving the device. Avoid condensation damage to the balance. When the balance is turned on, con-
densation build-up is not possible. If the device is moved from a cool environment to a warm environment, you must allow the device to acclimatize for approx. two hours before connecting to the power supply.
Good ventilation is required for devices with a built-in fan. Heat build-up should be generally avoided. Maintain a minimum distance of approx. 10 cm around the device.
Hazards at the Place of Installation and During Operation
Avoid generating static electricity and establish equipotential bonding. Grounding (10 kilohms) is used when connecting the balance to the power supply. Disconnecting the ground conductor is not permitted.
3
Exercise caution to avoid glass breakage. There is a danger of lacerations or cuts caused by sharp edges on glass or metal (contact springs), particularly when removing or replacing the glass draft shield panels, during cleaning, if a glass sample container breaks, or when modifying the balance hardware configuration.
3
Make sure all equipment operators have received proper instructions when working with hazardous or toxic materials. Failure to follow correct handling procedures can result in breakage of parts or spillage of liquids or other substances. Make sure the necessary protective clothing or equipment is worn, such as protective gloves, clothing, and eyewear. All safety regulations applicable in the laboratory must be observed.
3
Exercise caution to avoid pinching or crushing fingers when – Operating the electronic draft shield – Removing or installing the panels – Adjusting the display
Do not expose the device to aggressive chemical vapors or unnecessarily extreme temperatures, moisture, shocks, or vibration.
The device should be setup in areas away from equipment or devices that have strong magnetic properties. Extreme electromagnetic fields from electrical lines should also be avoided.
The setup location should be kept clean and free of dust. Always keep the balance freely accessible.
Intended Use
Cubis models are high-resolution precision and analytical balances. They were specially developed for exact determination of material mass in liquid, paste, powder, or solid form.
Appropriate containers must be used for each type of sample material. Cubis models are designed specially for use in research, educational, and day-to-day laboratory tasks in science, technology, and industry.
They are designed to be used exclusively indoors.
Cubis models can be operated as standalone, connected to a PC, or on a network.
6 Cubis MSU User Manual
Safety Instructions
Overview of Equipment
Pos. Designation
1 Upper draft shield panel/handle 2 Rear panel 3 Weighing pan 4 Pan support (not for MSx225.../MSx125... models) 5 Shield plate 6 Pan retainer 7 Right sliding panel/handle 8 SD card slot (for MSU, MSA models)
9 Leveling foot 10 Display 11 Display and control unit 12 Draft shield/shield disk (only for models
with a readability of 1 mg and 10 mg)
13 Below-balance weighing port
(on the bottom of the balance) 14 Leveling foot 15 Level indicator 16 Power socket 17 Slot for attaching an anti-theft device 18 USB socket for a PC connection 19 Hook for below-cell weighing 20 Slot for optional interfaces, e.g., 9-pin data output
and PS2 (as shown) or Bluetooth 21 Communication port (PERIPHERALS) for accessories 22 Menu access switch 23 Left sliding panel/handle 24 Ethernet port
(on the bottom of the display and control unit) 25 Semi-microbalances: weigh cell socket 26 Semi-microbalances: electronics module
(for MSx225.../MSx125... models)
Cubis MSU User Manual 7
10
11
23
3
4
5
5
6
6
13
13
11
11
14
14
6
12
12
4
3
3
15 16 17
18 19
20
21
22
4
25
26
15
17
20
21
22
1
2
3
4
5
6
7
9
8
24
Overview of Equipment
Getting Started
Unpacking the Balance
t Open the box.
t Use both hands to lift the balance, with the packaging,
out of the cardboard box.
t Place the packaging with the balance on the floor.
t Remove the top part of the packaging.
8 Cubis MSU User Manual
Getting Started
Balance with Draft Shield
t Remove the packages (containing draft shield panels, weighing pan, pan sup-
port, AC adaptor, etc.) from the lower packaging and place them to one side.
t Use both hands to lift the balance out of the packaging.
3
Exercise caution to avoid glass breakage. Never lift the equipment by the draft shield panels to remove it from the packaging.
t Place the balance on an even surface.
Balance without Draft Shield
t Use both hands to remove the model without a draft shield from
the packaging.
t Remove the AC adaptor and the balance parts from the packaging.
Cubis MSU User Manual 9
Getting Started
Equipment Supplied
t After unpacking the device and all parts, check it
immediately for any visible damage as a result of rough
handling during shipment.
If you detect any damage, proceed as directed in Section
“Care and Maintenance” under “Safety Inspection”.
t Save the box and all parts of the packaging for
any future transport.
Unplug all connected cables before packing
the equipment.
The model with the analytical draft shield comes with the following:
1) Upper draft shield panel
2) Right draft shield panel
3) Left draft shield panel
4) Weighing pan
5) Pan support (not for MSx225.../MSx125... models)
6) Shield plate
Not pictured: – AC adaptor – Operating instructions (this manual) – USB cable
The model with the milligram draft shield comes with the following:
1) Upper draft shield panel
2) Right draft shield panel
3) Left draft shield panel
4) Weighing pan
5) Pan support
6) Shield plate
Not pictured: – AC adaptor – Operating instructions (this manual) – USB cable
The model without a draft shield comes with the following:
1) Weighing pan
2) Pan support
3) Shield disk (only for models with a readability of 10 mg)
Not pictured: – AC adaptor – Operating instructions (this manual) – USB cable
10 Cubis MSU User Manual
Getting Started
1
3
4
2
5
6
1
2
3
4
5
6
1
2
3
Assembly
Installing the Analytical Draft Shield; Balance Assembly (Draft Shield Identification: DA, DI, and DU)
t Fit parts onto the balance as shown in the picture.
1. Shield plate
2. Pan support (not for MSx225.../MSx125... models)
3. Weighing pan
t Slide the upper draft shield panel into the guide rails from the rear.
t Slide the side draft shield panels into the guide rails from the rear.
Make sure the panels are fully inserted into the upper and lower guide rails.
t Slide the panels in until they snap into place.
1. Right draft shield panel
2. Left draft shield panel
y This completes the balance assembly.
Cubis MSU User Manual 11
Getting Started
2
1
3
Installing the Draft Shield on Milligram Balances, Balance Assembly (Draft Shield Identification: DE)
t Fit parts onto the balance as shown in the picture.
1. Shield plate
2. Pan support
3. Weighing pan
Assembly of Milligram Balances without Glass Draft Shield (DR Option): Balance Assembly
t Fit all components listed below onto the balance in the order given:
1. Pan support (see also next section)
2. Shield plate
3. Weighing pan
4. Draft shield frame
t First insert the shield plate.
t Make sure the pin on the pan support is facing toward the front, and slide
it under the clip on the pan retainer.
1. Insert the pan support.
t After inserting the pan support, press it down against the shield plate
so that it lies flat.
2. Press down on pan support.
3
Exercise caution to avoid pinching or crushing fingers.
t Place the weighing pan on the pan support.
3. Weighing pan
12 Cubis MSU User Manual
Getting Started
2
1
3
1
2
3
4
2
1
3
t Slide the upper draft shield panel into the guide rails from the rear while
pressing the locking tab.
t Slide the side draft shield panel into the guide rails from the rear while
pressing the locking tab.
1. Right draft shield panel
2. Left draft shield panel
y This completes the balance assembly.
Assembly for Models without a Draft Shield
t Fit parts onto the balance as shown in the picture.
1. Pan support
2. Shield plate/draft shield
3. Weighing pan
Cubis MSU User Manual 13
Getting Started
2
1
2
1
2
1
3
t Place the pan support diagonally and press down lightly.
1. Pan support
t Carefully turn the pan support clockwise until the two buttons
engage. The pan support is now attached.
t 2. Insert the shield plate/draft shield.
t 3. Place the weighing pan on the pan support.
y This completes the balance assembly.
Models MSx225..., MSx125... (Semi-microbalances): Connecting the Weigh Cell to the Electronics Module
t Connect the weigh cell and electronics module using the connection cable.
14 Cubis MSU User Manual
Getting Started
2
3
1
Moving the Balance
Moving the Balance over Short Distances
t Always use both hands to carry the balance.
Lift it up evenly either from the front and back
or
t from the sides.
3
Exercise caution to avoid glass breakage. Never lift and carry the balance by its draft shield.
Storage and Shipping Conditions
Permitted storage and shipping temperature: –10 to +60°C – Unpacked devices can lose their precision if subject to extreme vibrations. – Excessive vibrations may compromise the safety of the equipment. Save the packaging for any future storage or shipment of the balance. Only the original packaging provides optimum protection for the equipment. – Follow all warnings and safety precautions. – Follow the instructions in Section “Shipping the Balance.”
Cubis MSU User Manual 15
Getting Started
Installation Instructions
Adjustment to the ambient conditions: – Set up the device on a stable, even surface that is not exposed to vibrations
(e.g. weighing stone). – Ensure the device is located in a place where excessive heat cannot build up. – Maintain free access to the device at all times.
Avoid adverse influences at the place of installation: – Heat (heater or direct sunlight) – Direct drafts from open windows, air conditioning systems, and doors – Extreme vibrations during weighing – Excessive moisture
Acclimatization
Condensation from humidity can form on the surfaces of a cold device when it is brought into a substantially warmer area. To avoid the effects of condensation, condition the balance for about two hours at room temperature, leaving it unplugged from the power supply. Follow all warnings and safety precautions from the respective section of this manual.
Setting the Angle for the Display and Control Unit
The angle of the display and control unit can be adjusted by the user in order to ensure optimal readability of the weight values on the display at all times.
t The display and control unit can be tilted as desired.
Power Connection
t Check the voltage rating and plug design.
Make sure that the voltage rating printed on the manufacturer’s ID label
is identical to that of your local line voltage. – If the stated supply voltage or the plug design of the power cord does
not comply with your country’s standard, please inform the nearest Sartorius
representative or your dealer. – The power connection must be made in accordance with the regulations appli-
cable in your country. – In order to connect the equipment to the power supply (protection class 1),
use a correctly installed wall outlet with a protective grounding conductor (PE)
and a fuse of maximum 16 A. – If a main feeder from the ceiling is required or if installation of a CEE plug
is necessary, these must be carried out by a qualified electrician. – Only use original Sartorius power cables.
16 Cubis MSU User Manual
Getting Started
°C
Analytical and precision balances:
1. Plug the DC power supply cable into the balance and tighten
the threaded fitting.
Semi-microbalances (MSx1225..., MSx125... models):
1. Plug the DC supply cable into the electronics module
and tighten the threaded fitting.
2. Plug the other end of the power cord into the AC adaptor.
3. Insert the AC adaptor plug into the wall outlet.
y The balance is now ready to use.
Safety Precautions
The output line of the adaptor has a terminal (GND) connected to the metal hous­ing of the balance. The data port is also galvanically linked to the balance housing (GND).
Cubis MSU User Manual 17
Getting Started
2
1
Leveling the Balance, Setting the Level Indicator
h
Leveling the balance compensates for slant or unevenness at the place of installation. The balance must be perfectly horizontal to ensure con­sistent, reproducible weighing results. All models are equipped with an electronic tilt angle detection feature. If the balance is not level, all balance models issue an alert on the display. Push-button automatic leveling is available on models with motorized leveling feet. For models with manual leveling feet, please follow the instructions on the display.
Leveling the Balance Manually
t The balance is leveled using both front leveling feet. t Screw in both back leveling feet (only for models with back leveling feet). t Turn the two front feet as shown in the illustration until the air bubble is cen-
tered within the circle of the level indicator.
y Normally, several leveling steps are required.
t Turn both back leveling feet until they touch the setup surface
(only for models with back leveling feet).
Warm-up Time
y To return precise results, the balance must warm up for at least 30 minutes
after initial connection to the power supply.
Only after this time will the device have reached the required operating temper-
ature.
When a verified balance used in legal metrology (legal-for-trade applica­tions) is connected to the power, it must warm up for at least two hours before operation.
t Turn on the devices via the A key. The following appears on the display:
y The operator can now follow the brief instructions to configure the balance
settings before beginning weighing operations.
18 Cubis MSU User Manual
Getting Started
R
L
L
L
R
R
Equipment Downtime
t Press the A key to switch the equipment to standby mode when it will not
be in use for a short period of time.
t For longer periods of time, the device should be disconnected from the power
supply.
Connecting Electronic Devices (Peripherals)
t Make absolutely sure that the device is unplugged from the power supply
before connecting/disconnecting any peripheral device (printer, scanner, PC) to or from the data port.
3
A device connected to the power supply should never be opened.
Anti-theft Locking Device (Accessory)
t An optional anti-theft locking device can be installed to the back
of the balance.
Cubis MSU User Manual 19
Getting Started
Modifying the Balance
Setting Up the Display and Control Unit at the Place of Use
The display and control unit can be removed from all models and set-up wherever desired to enable the operator to customize the work space.
Removing the Retainer with the Display and Control Unit
t Remove all items (such as weights) from the draft shield.
t Carefully remove all parts as shown in the illustration.
1. Weighing pan
2. Pan support (not for MSx225.../MSx125... models)
3. Shield plate/draft shield
t Keep all parts in a safe place.
t Remove the panels (right and left).
Models with the analytical draft shield:
1. Press on the locking tab.
2. Remove the panel.
Models with the milligram draft shield:
1. Press on the locking tab.
2. Remove the panel.
t Keep all parts in a safe place.
20 Cubis MSU User Manual
Getting Started
2
1
3
2
1
2
1
t Remove the upper draft shield panel.
Models with the analytical draft shield (left figure):
1. Press on the locking tab.
2. Remove the panel.
Models with the milligram draft shield (right figure):
1. Press on the locking tab.
2. Remove the panel.
t Keep all parts in a safe place.
t Turn over the balance and place on a soft surface.
3
Exercise caution to avoid breaking the glass on models with a draft shield.
t Use a 2.5 mm Allen key to remove the two screws from the display and control
unit retainer bracket.
t Remove the display and re-insert both screws back into their holes.
t Lengthen the cable and position the display and control unit as desired.
t Turn the balance over and place it on an even surface.
t Carefully fit all parts onto the balance.
1. Shield plate/draft shield
2. Pan support (not for MSx225.../MSx125... models)
3. Weighing pan
Cubis MSU User Manual 21
Getting Started
2
1
2
1
2
1
3
t Replace the upper and side shield panel.
1. Upper draft shield panel
2. Right draft shield panel
3. Left draft shield panel
t Level the balance.
h
Leveling the balance compensates for slant or unevenness at the place of installation. The balance must be perfectly horizontal to ensure consistent, reproducible weighing results. All models are equipped with an electronic tilt angle detection feature. If the balance is not level, the balance issues an alert (MSA and MSU models, see brief instructions). The display shows a graphical representa­tion of a level indicator and text prompts that guide the user through the leveling process.
t The balance is leveled using both front leveling feet.
t Turn both back leveling feet (only for models with back leveling feet).
t Turn the two front feet as shown in the illustration until the air bubble
is centered within the circle of the level indicator.
y Normally, several leveling steps are required.
t Turn both back leveling feet until they touch the setup surface
(only for models with back leveling feet).
Removing the Display and Control Unit for Models without a Draft Shield
t Carefully remove all parts as shown in the illustration.
1. Weighing pan
2. Shield plate/draft shield
3. Pan support
t Keep all parts in a safe place.
t Turn over the balance and place on a soft surface.
22 Cubis MSU User Manual
Getting Started
3
2
1
2
1
3
t Remove the two retaining screws.
t Remove the display and re-insert both screws back into their holes.
t Carefully pull the cable connected between the display and control unit from
the retainer.
t Determine the required cable length.
t Return the balance to an upright position and fit the parts onto the balance.
1. Attach the pan support.
2. Shield disk (only for models with a readability of 10 mg)
3. Weighing pan
t Level the balance.
Cubis MSU User Manual 23
Getting Started
2
1
3
Semi-microbalances: Attaching the Display and Control Unit to the Electronics Module (MSx225.../MSx125... models)
The display and control unit can also be attached to the electronics module if required for operation.
t Turn over the balance and place on a soft surface.
Remove the connection cable from the cable channel:
t Remove two screws from beneath the weigh cell and detach the plate.
t Remove the connection cable plug.
t Then reattach the plate to the slot.
t Remove the display and control unit from the weigh cell:
Remove two retaining screws.
t Remove the display and control unit.
24 Cubis MSU User Manual
Getting Started
tAttach the display and control unit to the electronics module:
Reattach the two retaining screws.
Open the slot for the connection cable on the electronics module:
t Remove two screws from beneath the electronics module and detach the plate.
Connect the display and control unit to the electronics module:
t Plug in the connection cable.
t Then reattach the plate to the slot.
t Slide the cable that sticks out into the cable channel.
Cubis MSU User Manual 25
Getting Started
Below-Cell Weighing
A port for a below-cell weighing hanger is located on the bottom of the balance.
1. Remove the hook for below-cell weighing from the clip on the bottom
of the balance.
2. Push the cover of the below-cell weighing port to one side.
3. Carefully screw on the below-cell weighing hook.
3
Do not screw on the below-cell weighing hook too tightly. This may damage the threading or the balance itself.
h
Install a draft protection shield.
t Attach the sample (e.g., using a suspension wire) to the hanger.
4 When weighing has been completed, unscrew the hook and return it to the clip.
t Close the cover of the below-cell weighing port.
The below-cell weighing port may not be opened or used on balances used for legal metrology.
26 Cubis MSU User Manual
Getting Started
1
2
3
4
Using the Cable Opening of the Draft Shield
Models with an analytical draft shield have an opening for passing a cable (for example, for a temperature sensor) through to the interior of the weighing chamber.
1. Lift the locking tab on the rear panel of the analytical draft shield.
2. Lift the panel out of the draft shield.
t Turn the panel clockwise 180° so that the opening is at the bottom.
t Install the desired sensor.
t Insert the panel into the guide rail.
t Lift the locking tab and gently press the panel into position.
t Press the locking tab down and close it.
t You can now begin weighing.
Cubis MSU User Manual 27
Getting Started
Removing the Side Draft Shield Panels for Milligram Balances
The side draft shield panels on milligram balances can be removed and stored at the back of the analytical draft shield. This can be useful, for example, when perform­ing continuous weighing or filling operations.
h
To work with an open draft shield, the ambient conditions must be extremely stable.
t Remove the panel:
1. Press on the locking tab.
2. Remove the panel.
t Place the panel in the storage slot at the back of the balance.
28 Cubis MSU User Manual
Getting Started
2
1
Cubis MSU User Manual 29
Getting Started
Using the Balance
Turning the Balance On/Off
t Make sure the balance has been installed and put into operation in accordance
with the installation instructions.
A t Press the on/off key A on the control unit.
y The start screen appears briefly on the display, and then the user interface
appears.
h
On subsequent starts, the most recently active user profile and task are opened (if at least one user profile has been set up).
t If prompted, level the balance (see Section “Level the Balance” for details).
A t To switch the balance to standby mode or turn it off, press the A key.
t Close the draft shield (if your model has one).
Operating Design: Q-Guide
Cubis precision and analytical balances are controlled by application software with interactive operator guidance. Once you open a menu or select an application, brief instructions appear on the display to guide you through the menu or application step-by-step. At each step along the way, the display shows only those options that are relevant to your process; this helps prevent unnecessary “detours” and enables you to reach your goal more quickly. This type of user guidance is mostly intuitive. Although operation of the balance is also mostly intuitive, this manual includes a section with very detailed, step-by­step instructions where you can learn about all of the available options (see “Creating User Profiles” in Section “User Management”).
Basic Operational Structure
The most basic functions, weighing and taring, can be carried out as soon as the balance is switched on. The application software is divided into three general areas in which individual settings can be configured: In Task Management (TASK), task profiles with specific applications, weighing settings, and printout settings can be configured.
User Management (USER) lets you set up user profiles with specific configuration settings and user rights. You can also set up password protection to meet your security requirements. Your balance can also be used without User Management.
Simple Weighing
D TASK
Task Management (from page 53)
Selecting and starting applications
Configured
at factory
Configured
by user
System Settings (from page 35)
Configuring applications Configuring tasks
Creating user profiles
H USER
User Management (from page 119)
Selecting user profiles
System Settings (MENU) contains all basic settings and parameters which used to operate the balance, affect the entire operation of the balance, and can also be used for creating new tasks.
Wizard
You can choose between two display types in several menus: The overview lists all parameters with their set options. You can select each
individual option to change its setting. When you activate the Wizard in the same menu, the program guides you step by
step: The individual parameters with their options are displayed in succession.
Using the Display and Control Unit
Control Keys (Softkeys):
D
TASK key: opens Task Management, in which applications can be selected and tasks defined. While working in this menu, you can also use the TASK key to cancel at any time and return to task overview.
H
USER key: opens User Management, in which user profiles can be selected and user settings defined. While working in this menu, you can also use the USER key to cancel at any time and return to user overview.
J
Left and right TARE keys: used to tare the balance
P
PRINT key: used to print the currently displayed measurement results or a user-specific print-out
A
On/off key: used to turn on/off and switch to standby
-
Softkeys: used to activate the function displayed directly above them
qwe
Cursor softkeys: used for navigation on the display
O
Operating elements for electronic draft shield (optional)
Using the Balance
30 Cubis MSU User Manual
Loading...
+ 154 hidden pages