Sartorius CPA224S, CPA124S, CPA64, CPA64-WDS, GCA1603P Operating Instructions Manual

...
Operating Instructions
Sartorius CP|Gem
plus
Series
CPA, GCA and GPA Models Electronic Micro-, Analytical and Precision Balances and Precious Metal Scales
98648-015-70
Contents
Safety Precautions 3
Getting Started 5
Installation 6
Operation 15
Overview of Display and Operating Elements 15 Basic Weighing Function 16 Below-Balance Weighing 18 Calibration and Adjustment 20
Configuration 25 Printing the Parameter Settings 25 Setting the Parameters (Menu Codes) 26 Parameter Settings (Overview) 27 Setting IDs, Time, Date and Display Brightness 32
Application Programs 35 Net-total Formulation 36 Counting 40 Reference Balance/Scale for Counting 40 Weighing in Percent 44 Animal Weighing/Averaging 47 Toggling between Weight Units 50
Generating a Printout 51 ISO/GLP-compliant Printout/Record 53
Interface Port 55 Data Input Format 59 Pin Assignment Charts 62 Cabling Diagram 63
Troubleshooting Guide 64
Preparing CPA2P.. Models for Transport 65
Care and Maintenance 66
Instructions for Recycling 68
Overview 69
Specifications 69 Accessories (Options) 81 Declarations of Conformity 84 EC Type-approval Certificate 87 Plates and Markings 88
2
Safety Precautions
Safety Instructions
§ Please read these operating instructions
carefully before using your balance to prevent damage to the equipment.
!Do not use this equipment in hazardous
areas/locations.
!The balance housing may be opened
only by Sartorius service technicians who have been trained at the factory.
!Make sure you disconnect the balance
from power before connecting or disconnecting peripheral devices to or from the balance.
!If you operate the equipment under
ambient conditions that require higher safety standards, you must comply with the installation regulations applicable in your country.
When cleaning your balance, make sure that no liquid enters the balance housing; use only a slightly moistened cloth to clean the balance.
Installation
!Make sure the voltage rating printed
on the AC adapter is identical to your local line voltage.
– Proceed with extreme caution when
using pre-wired RS-232 connecting cables, as the pin assignments may not be compatible with Sartorius equipment. Check all pin assignments against the cabling diagrams and dis­connect any lines that do not match.
!If there is visible damage to the equip-
ment or power cord, disconnect the equipment from power and lock it in a secure place to ensure that it cannot be used for the time being.
– Connect only Sartorius accessories
and options, as these are optimally designed for use with your balance. The operator shall be responsible for any modifications to Sartorius equipment and for any connection of cables or equipment not supplied by Sartorius and must check and, if necessary, correct these modifications and connections. On request, Sartorius will be happy to provide information on operating specifications (in accordance with the Standards for defined immunity to interference).
$ Do not open the balance. If the seal
is broken, this will result in forfeiture of all claims under the manufacturer's warranty.
$ If you have any problems with your
balance, please contact your local Sartorius office, dealer or service center.
3
Equipment Supplied
The equipment supplied includes the components listed below:
CPA2P, CPA2P-F – Balance with display and control unit – Kit of standard accessories – AC adapter – Dust cover – Filter pan and lid (model CPA2P-F only) – Spacer (model CPA2P-F only)
The kit of standard accessories contains
the following:
– Weighing pan
– Interior draft shield
– Hanger for below-balance weighing
– 1 brush
– 1 pair of forceps
– 1 piece of lint-free cloth
CPA Balances with a Readability of < 0.1 mg; CPA…-DS, GCA Scales
Balance/scale with display and
control unit – Electronics box (model CPA225D only) – Draft shield with base plate
(not available for the CPA64-WDS) – AC adapter – Weighing pan – Shield disk – Centering plate (only for CPA…DS) – In-use dust cover – Gem tray (GCA scales only)
CPA Balances with a Readability of 1 mg
(except for the CPA…-DS models) – Balance with display and control unit – Draft shield with cover – AC adapter – Weighing pan – Pan support – Base plate – In-use dust cover
CPA-Balances with a Readability of
0.01 g/0.1 g, GPA Scales
Balance/scale with display and
control unit – AC adapter – Weighing pan – Gem tray (GPA scales only) – In-use dust cover
CPA34001S, CPA34001P, CPA16001S,
CPA12001S, CPA34000 – Balance with display and control unit – AC adapter – Weighing pan – In-use dust cover
4
Getting Started
Storage and Shipping Conditions
– Do not expose the balance to
extreme temperatures, moisture, shocks, blows or vibration.
Unpacking the Balance
§ After unpacking the equipment,
please check it immediately for any external damage
$ If damage is evident, refer to the
instructions under “Safety Inspection" in the chapter entitled “Care and Maintenance."
$ Save the box and all parts of the
packaging for any future transport. Disconnect all cables before packing the balance for shipping!
Installation
Choose a location that is not subject
to the following negative influences: – Heat (heater or direct sunlight) – Drafts from open windows and doors – Extreme vibrations during weighing – Excessive moisture
Conditioning the Balance
Moisture in the air can condense on
the surfaces of a cold balance whenever
it is brought into a substantially warmer
place. If you transfer the balance to
a warmer area, make sure to condition it
for about 2 hours at room temperature,
leaving it unplugged from AC power.
Seal on Balances Verified for Use
in Legal Metrology in the EU*:
EU legislation requires that a control
seal be affixed to verified balances of
accuracy class K. The control seal
consists of a sticker with the “Sartorius"
logo. If the seal is broken, the verifi-
cation becomes null and void and the
balance must be re-verified.
* Including the Signatories of the Agree-
ment on the European Economic Area
5
Installation
Model CPA2P
§ Remove the adhesive tape from the chamber doors
§ Place the components listed below inside the
chamber in the order given: – Interior draft shield – Weighing pan
Model CPA2P-F
1) Weighing filters of up to 125 mm d:
§ Lift the chamber lid gently and turn it to the
left or right
§ Place the components listed below inside
the chamber in the order given: – Interior draft shield – Filter pan – Cover the sample (filter) with the lid
or
2) Weighing with the standard weighing pan
(20 mm d):
§ Lift the chamber lid gently and turn it to the
left or right
§ Place the components listed below inside
the chamber in the order given: – Interior draft shield – Spacer ring – Weighing pan
Balances with an Analytical Draft Shield
!Check the sliding lock device on the back of the
draft shield; make sure it is in the “open" position
(to the right).
§ Position the draft shield carefully on the balance
6
§ Secure the draft shield by pressing lightly on
the draft shield base and moving the sliding lock device to the left
§ Place components inside the chamber in the
following order: – Base plate – Shield ring – Pan support – Weighing pan – Gem tray (GCA scales only)
§ Place components inside the chamber in the
following order: – Base plate – Centering plate – Shield ring – Pan support – Weighing pan
7
Round weighing pan shape
Triangular weighing pan shape
Connecting Model CPA26P, CPA225D (-0CE) to the Electronics Box
– Plug the male connector on the cable into the
female connector on the electronics box
!Do not exchange the balance or electronics box with
a component of a different balance!
Balances with a 3-Sided Draft Shield
§ Place draft shield on the balance with the cover
opening in front on the right
§ Turn the draft shield clockwise until it is firmly
in position
§ Place components inside the chamber in the
following order:
– Base plate
– Weighing pan receptor
– Weighing pan
$ To access the weighing chamber from the side,
remove side panels as desired
8
1
2
9
Balances/Scales with a Rectangular Weighing Pan
and a Weighing Capacity up to 10 kg
§ Place the components listed below on the
balance/scale in the order given:
– Weighing pan
– Gem tray (GPA scales only)
Balances with a Rectangular Weighing Pan and
a Weighing Capacity over 10 kg
§ Place the weighing pan on the balance
10
Connecting the Balance to AC Power/ Safety Precautions
$ Use only original Sartorius equipment.
The AC adapter meets the requirements of IP20 in accordance with EN 60529.
$ For AC adapters with higher protection ratings
or for an external rechargeable battery pack, please see “Accessories."
CPA26P, CPA225D:
G Insert the right-angle plug from the AC adapter into
the jack on the electronics box.
All other balances with a weighing capacity up to 10 kg:
G Insert the right-angle plug from the AC adapter
into the jack on the balance.
G Connect the AC adapter to an electrical outlet
(mains supply)
AC Adapter with Country-specific Power Cord
Some models come with separate country-specific power cords for the AC adapter. In Europe, use only original Sartorius AC adapter part no. 6971983.
§ Connect the angle plug to the balance/scale
§ Select the power cord for your area and connect it to
the AC adapter
§ Plug the power cord into the wall outlet (mains)
Connect the power cord to the AC adapter (on balances with weighing capacities up to 10 kg)
$ Use an original Sartorius AC adapter with a wide input
voltage range (100 to 240 V~), order no. 6971966, and replaceable power cord: 6900900 (Europe) 6900901 (US/CDN) 6971945 (UK) 6900905 (AUS) 6900902 (ZA)
11
Balances with a Weighing Capacity over 10 kg: G Insert the right-angle plug into the jack and
tighten the screw.
G Connect the AC adapter to an electrical outlet
(mains supply)
Safety Precautions
Plug-in AC Adapter:
The balance/scale is intended to be supplied by a
listed direct plug-in power unit marked “Class 2."
Universal AC Adapter 6971966:
The AC adapter rated to Class 1 can be plugged
into any wall outlet without additional safety
precautions.
The ground terminal is connected to the balance
housing, which can be additionally grounded
for operation. The data interface is also electrically
connected to the balance housing (ground).
Note:
This equipment has been tested and found to
comply with the limits pursuant to part 15 of FCC
Rules. These limits are designed to provide reason-
able protection against harmful interference. This
equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in
accordance with these instructions, may cause
harmful interference to radio communications.
For information on the specific limits and class
of this equipment, please refer to the Declaration
of Conformity. Depending on the particular class,
you are either required or requested to correct the
interference.
If you have a Class A digital device, you need to comply with the FCC statements as follows: “Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.”
If you have a Class B digital device, please read and follow the FCC information given below:
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna. – Increase the separation between the equipment
and receiver.
– Connect the equipment into an outlet on
a circuit different from that to which the receiver is connected.
– Consult the dealer or an experienced radio/TV
technician for help.
Before you operate this equipment, check which FCC class (Class A or Class B) it has according to the Declaration of Conformity included. Be sure to observe the information of this Declaration.
12
Connecting Electronic Peripheral Devices
§ Make sure to unplug the balance from AC power
before you connect or disconnect a peripheral device
(printer or PC) to or from the interface port.
Warmup Time
To deliver exact results, the balance must warm up
as listed below after initial connection to AC power
or after a relatively long power outage. – Model CPA2P…, CPA26P…: at least 4 hours – All other precision and analytical models:
at least 30 minutes
Only after this time will the balance have reached
the required operating temperature.
Using Verified Balances in Legal Metrology:
$ Allow the equipment to warm up for at least
24 hours after initial connection to AC power.
Antitheft Locking Device on Balances with
a Weighing Capacity of up to 10 kg
G To secure the balance at the place of installation,
fasten a chain or a lock to the lug located on the
rear panel of the balance (order no.: LC1).
13
14
Leveling the Balance
Purpose:
– To compensate for unevenness at the place
of installation
Leveling Balances with a Weighing Capacity of up to 10 kg
Only the 2 front feet are adjusted to level the balance.
§ Retract the two rear feet (only on models with
a rectangular weighing pan).
§ Turn the 2 front feet as shown in the diagram
until the air bubble is centered within the circle of the level indicator.
> In most cases this will require several adjustment
steps.
§ For weighing heavy samples: Extend the 2 rear
feet until they touch the surface on which the balance rests (only on models with a rectangular weighing pan).
Leveling Balances with a Weighing Capacity of over 10 kg
§ Adjust the leveling feet until the air bubble
is centered within the circle on the level indicator.
15
Operation
Overview of Display and Operating Elements
Position Designation
1 Weight units 2 Tare key 3 Symbol: “GLP printing mode
active" 4 Symbol: “Printing mode active" 5 Display: Data in memory for
net-total formulation program 6 Data output: Press this key
to output readout values to the
built-in data interface. 7 Function key:
Start application program 8 Symbols for active application 9 Start calibration/adjustment
routine
Position Designation
10 Delete (Clear Function)
This key is generally used to cancel functions. – Quit application program – Cancel calibration/
adjustment routine 11 On/off 12 Display: Calibration/adjustment
function
13 Display: Animal weighing with
automatic start
14 Symbols for stand-by mode
or zero range
15 Weight value displayed in selected
weight unit
15 1
2 2
14 3
4 5 6
13 12 11
10
9
7 8
Purpose
The basic weighing function can be used alone or in combination with an application program (counting, weighing in percent, etc.).
Features
– Taring the balance – Assigning IDs to weights (as needed) – Printing weights
Using Verified Balances as Legal Measuring Instruments in the EU*:
The type-approval certificate for veri­fication applies only to non-automatic weighing instruments. For automatic operation with or without auxiliary measuring devices, you must comply with the regulations applicable to the place of installation.
G Before using the balance as a legal
measuring instrument, calibrate and adjust it at the place of use using the built-in motorized calibration weight; for details, see “Calibration and Adjustment" in the next chapter.
$ The temperature range (°C) indicated
on the verification label may not be exceeded during operation.
Example: BD BL 200 +10°C to +30°C 0°C to +40°C isoCAL
K
* Including the Signatories of the Agree-
ment on the European Economic Area
Working with CPA2P, CPA26P… Models:
Working with the microbalance requires a steady hand and a smooth, uninterrupted technique.
Use forceps or other suitable utensil to place the sample on the weighing pan.
Perform a number of test measurements before you begin weighing, to allow the temperature inside the weighing chamber to adjust to the ambient temperature outside the chamber. Otherwise, if the chamber door was closed for a longer period of time prior to beginning weighing, the sudden change in tem­perature inside the chamber when you open the door might affect the weight readout. This is why a series of test measurements is recommended; the repeated opening and closing of the weighing chamber door, at the same rate of speed as will be used during the actual weighing sequence, will both compensate this difference in tempera­ture to some extent and help you develop a smooth working rhythm.
Place the sample gently on the weighing pan. The weight readout should stabilize within 15 to 20 seconds. The degree of precision attained increases in propor­tion as the weighing operations become more consistent.
16
Basic Weighing Function
17
Preparation
A circle in the upper right-hand corner of the display indicates that the balance was disconnected from power. This symbol is shown, for example, the first time the balance is put into operation, or after a power outage.
G Switch on the balance: Press (E) > All symbols on the display light up briefly.
> The balance performs a display test.
$ Tare the balance, if necessary: Press (TARE)
When you turn on the balance, the b symbol is displayed until you press a key. If the b symbol is displayed during operation, this indicates that the processor is performing a function and cannot receive further commands at the moment.
Additional Functions
G Switching off the balance: Press (E)
A circle in the lower left-hand corner of the display indicates that the balance has been switched off and is in stand-by mode.
Filter Weighing with Model CPA2P-F
The CPA2P-F filter microbalance comes with a filter pan that has a utilizable diameter of 125 mm as standard equipment. Place the filter on the pan and close the lid.
18
Below-Balance Weighing
A port for a below-balance weighing hanger is located on the bottom of the balance (for models with a weighing capacity of 12 kg or more, see chapter on “Accessories.”
$ Below-balance weighing is not permitted in
legal metrology.
G Open cover plate on the bottom of the balance.
G Using the built-in hanger 1: Attach the sample
(e.g., using a suspension wire) to the hanger.
$ If necessary, install a shield for protection against
drafts.
G Remove the brass screw
G Suspend the below-balance weighing hook
supplied from the hanger
G If necessary, install a shield for protection against
drafts
1
Microbalance CPA2P..:
Analytical and precision balances:
19
Example
Simple Weighing
Step Key (or instruction) Display/Data Output
1. Switch on the balance. (E)
+ 0.0 g
Self-test is performed, followed by automatic initial tare function.
2. Place container on the balance
+ 11.5 g
(in this example, 11.5 g).
3. Tare the balance. (TARE)
+ 0.0 g
4. Place sample in container
+ 132.0 g
on balance (in this example, 132 g).
6. Print weight. (A)
N + 132.0 g
20
Calibration and Adjustment
Purpose
Calibration is the determination of the difference between the weight readout and the true weight (mass) of a sample. Calibration does not entail making any changes within the balance.
Adjustment is the correction of any difference between the measured value displayed and the true weight (mass) of the sample, or the reduction of the difference to an allowable level within the maximum permissible error limits.
Using Verified Balances as Legal Measuring Instruments in the EU*:
Before using your balance as a legal measuring instrument, you must perform “internal calibration/adjustment" at the place of installation after the warmup period.
Features
Calibration/adjustment can be performed
only when: – there is no load on the balance, – the balance is tared, and – the internal signal is stable.
If these conditions are not met,
an error message is displayed (Err 02).
Adjustment can be performed – automatically following calibration
(1 10 1) or – manually, at operator discretion,
after calibration (1 10 2)
* Including the Signatories of the Agree-
ment on the European Economic Area
– The weight displayed for the sample
on the balance must not differ from the nominal weight by more than 2%.
– You can use any of the following
weight units to calibrate/adjust the balance: g, kg, lb (1 11 1 to 3, factory setting: 1 11 1)
You can block calibration/adjustment
of the balance as follows: – Select menu code1 9 7, or – Close the menu access switch on the
back of the balance
– You can have calibration/adjustment
start automatically when a specified
time or temperature limit is reached
(isoCAL function; 1 15 3).
– You can have calibration and
adjustment results documented as
an ISO/GLP-compliant printout;
see the chapter on ISO/GLP-compliant
Printout/Record.
External Calibration/Adjustment
in Verified Balances of Accuracy
Class K
– When the balance is used in legal
metrology, external calibration/adjust-
ment is blocked by a seal over the menu
access switch.
21
Internal Calibration/Adjustment
The menu code setting 1 9 3 must be selected in the Setup menu.
Inside the balance housing is a motorized calibration weight which is applied and removed automatically for internal calibration.
G Activate calibration: Press (CAL)
> The built-in calibration weight is
applied automatically.
> The balance is calibrated.
> If “Calibrate, then auto adjust in one
operation" is selected in the Setup menu, the balance is now adjusted automatically.
> The internal calibration weight is
removed.
! Important note for calibration/adjust-
ment of the CPA2P-F model: position the cover before calibration/adjust­ment
Calibration and Adjustment Sequence
In the Setup menu, you can configure whether:
– Calibration is always followed auto-
matically by adjustment (1 10 1; factory setting), or
– You have the choice of ending the
sequence or starting adjustment after calibration (1 10 2)
If no difference is determined between nominal and actual weights, you can end the calibration/adjustment routine following calibration. Two keys are active at this point:
– (CAL) = start calibration/adjustment
– (CF) = end the sequence
22
isoCAL*: Automatic Calibration and Adjustment
The menu code setting 1 15 3 must be selected in the Setup menu.
– Temperature range with isoCAL:
0°C to +40°C The “AUTOCAL" display automatically begins flashing if the ambient temperature has changed in relation to the tempera­ture at the time of the last calibration/ adjustment, or after a defined time inter­val has elapsed. The balance is telling you that it wants to self-calibrate and adjust.
This adjustment prompt is activated when:
– The change in temperature or the elapsed
time interval is greater than that shown in the table below
– The load on the pan has not been
changed within the last 2 minutes
– The balance has not been operated within
the last 2 minutes
– The weight on the pan is no more than 2%
of the maximum capacity of the balance
*= CPA64-WDS models:
factory setting isoCAL off (Code 1 15 1)
When these requirements are met,
the following symbols are displayed: – C in the measured value line – AUTOCAL flashes in the symbol display
In the Setup menu, you can configure
the balance to display the adjustment
prompt only, without performing
calibration/adjustment automatically
(menu code 1 15 2).
! Important note for calibration/adjust-
ment of the CPA2P-F model: position
the cover before calibration/adjustment.
This is necessary to enable automatic
calibration (isoCAL) to take place.
isoCAL Deactivated on Verified
Balances:
The permitted operating temperature
range for balances used in applications
subject to legal metrology (legal for
trade) is restricted as follows: – Balances of accuracy class k:
+15°C to 25°C (+59 to +77°F) – Balances of accuracy class K:
+10°C to +30°C (+50 to 86°F)
Fully automatic adjustment is initiated under the following conditions:
Model When the After a time
temperature interval of changes by
CPA2P, CPA2P-F, CPA26P, CPA225D, CPA324S, CPA224S, CPA124S, CPA64, CPA1003S, CPA1003P, GCA1603, GCA803 1.5 Kelvin 4 hours
CPA623S, CPA423S, CPA6202S, CPA5202S-DS, CPA6202P, CPA4202S, CPA523S-PCE, GPA5202, GPA3202 2 Kelvin 6 hours
CPA323S, CPA2202S…, CPA34001P, CPA34001S, CPA223S, CPA3202S, GCA2502 4 Kelvin 12 hours
CPA5201, CPA2201-0CE, CPA10001, CPA16001S CPA12001S, CPA34000, CPA8201 4 Kelvin 24 hours
These values are also set in the corresponding verified balances (CPA…-0CE/-PCE models with the -0CE designation).
23
Internal Calibration
Step Key (or instruction) Display
1. Zero the balance. (TARE)
0.0 g
CPA2P-F model: Either position the standard pan (8 g) or the filter pan (5 g) along with the cover (3 g) beforehand.
2. Start calibration. (CAL)
C
The internal weight is CAL applied automatically.
3. The balance is calibrated
0.2 g
(displayed only if menu CAL ± G code 1 10 2 is set).
4. If the “Calibrate, then auto
Adjust*
adjust" setting is selected CAL (1 10 1), the balance is now adjusted automatically.
5. The calibration sequence
CC
is completed. CAL
6. The internal weight is removed.
0.0 g
* = Adjust displayed only if menu item 1 10 2 is selected.
24
External Calibration
Settings: Calibration/adjustment mode: External calibration/adjustment (menu code 1 9 1)
The weight required for calibration/adjustment is defined in the factory settings (see “Specifications").
Step Key (or instruction) Display
1. Zero the balance. (TARE)
0.0 g
2. Start calibration. (CAL)
+ 5000.0 g
CAL G
3. Apply the prompted calibration weight (in this example, 5000 g).
4. The balance is calibrated
0.2 g
(displayed only if menu CAL ± G code 1 10 2 is set).
5. If the “Calibrate, then auto
Adjust
*
adjust" setting is selected CAL (1 10 1), the balance is now adjusted automatically.
6. The calibration/adjustment
CC
sequence is completed. CAL
7. After calibration/adjustment, the
+ 5000.0 g
weight is displayed with weight unit.
8. Remove the calibration weight.
0.0 g
* Adjust shown only if menu code 1 10 2 is set.
Important note: Afterwards, do not perform internal calibration/adjustment again.
25
Configuration
Purpose
To adapt the balance to individual requirements by choosing from parameters options in the Setup menu.
Features
To open the Setup menu, switch the balance off and then on again by pressing (E). While all segments are lit, press (TARE) briefly.
Scroll upward ↑: Press (CAL) Scroll to the right : Press (A) Confirm input: Press (TARE) Save settings and exit menu: Press and hold (TARE) (> 2 sec.)
G Printing the Parameter Settings
– At the 3
rd
menu level (lowest level;
see also the next page): Press and hold (A) (> 2 sec.).
> Printout (Example)
Menu711
– At the 2nd menu level:
Press and hold (A) (> 2 sec.).
> Printout (Example)
Menu711 Menu721 Menu731
– All current menu settings are printed
when the 1
st
menu level (highest level)
is displayed: Press and hold (A) (> 2 sec.).
26
Setting the Parameters (Menu Codes)
Example: Adapting the balance to “very unstable" ambient conditions (menu code 1 1 4).
Step Key (or instruction) Display
1. Switch off the balance. (E)
2. Switch the balance on; (E)
while all segments are displayed: (TARE) briefly
1
$ Scroll upward within a (CAL) 2
menu level; after the last repeatedly ... menu code, the first
9
code is displayed again. 1
3. Select menu level 2 (A) 1 1 (scroll to the right).
4. Select menu level 3 (A)
1 1 2
o
(scroll to the right).
5. Menu level 3: (CAL) Scroll until the desired repeatedly
1 1 4
number is shown.
6. Confirm change; (TARE)
1 1 4
o
“o” on display indicates active setting.
$ Return to higher menu level (A)
1
(from the third level).
$ Set other codes as desired. (A), (CAL)
7. Save changes and exit Press and hold the menu (TARE) (> 2 sec.)
or
$ Exit menu without saving
changes. (E)
27
Parameter Settings (Overview)
o Factory setting User-defined setting
Setup 1 Weighing 1 1 Adapt filter 111 Very stable conditions
112 o Stable conditions 113 Unstable conditions 114 Very unstable conditions
1 2 Application filter 121 o Final readout
122 Filling mode
1 3 Stability range 131 4 digit
The stability symbol is 132 1 digit displayed when the 133 1 digit value is stable within 134 o 2 digits this number of digits. 135 4 digits
136 8 digits*
1 5 Tare function* 151 Without stability
152 o After stability
1 6 Auto zero 161oOn
162 Off
1 7 Weight unit 1 171 Grams (display: o)*
172 o Grams (display: g) 173 Kilograms
1
)
174 Carats 175 Pounds* 176 Ounces* 177 Troy ounces* 178 Hong Kong taels* 179 Singapore taels* 1710 Taiwanese taels* 1711 Grains* 1712 Pennyweights* 1713 Milligrams
2
)
1714 Parts per pound* 1715 Chinese taels* 1716 Mommes* 1717 Austrian carats* 1718 Tola* 1719 Baht* 1720 Mesghal*
1 8 see next page
2 through 9 see following pages
* = Not available in balances verified for use in legal metrology
1
) = Not available for models with a readability < 0.1 mg
2
) = Not available for verified balances of accuracy class K
Menu level 1
Menu level 2
Menu level 3
Factory setting
Menu item
Menu 1 Weighing 1 1 through 1 7 see previous page
1 8 Display accuracy 1 * 181 o All digits
185 Reduced by 1 digit*
1 9 (CAL) key function 191 External cal./adj.
1
)
193 o Internal cal./adj. 196 CPA225D model:
Internal linearization
197 (CAL) key blocked
1 10 Calibration/ 1101 o Calibrate, then auto adjust
adjustment sequence in one operation
1102 Calibrate, then manual
adjust
1 11 Weight unit for 1111 o Grams
calibration weight * 1112 Kilograms
1113 Pounds
1 15 isoCAL function 1151 Off
2
)
1152 Only adjustment prompt 1153oOn
2 Application 2 1 Program selection 211 o Weighing
programs 212 Toggle weight units
214 Counting 215 Weighing in percent
3
)
216 Net-total formulation 217 Animal weighing
3
)
3 through 9 see next page
* = Setting cannot be changed on verified balances
1
) = Not available for verified balances of accuracy class K
2
) = Factory setting for CPA64-WDS
3
) = Not available for CPA…-PCE models
28
Menu level 1
Menu level 2
Menu level 3
Factory setting
Menu item
Loading...
+ 64 hidden pages