| 15
Activating the BPI Mode
The BPI switch (1) is used to prepare for working the service software and is located
on the rear panel.
On models CP622, CP4202S, CP3202S, CP3202P, CP2202S, GP5202, GP3202, CP8201,
CP6201, CP4201, CP2201, GP8201, CP423S, CP323S, CP323P, CP153, GC2502,
CP225D, CP324S, CP224S, CP124S, CP64, GC1603, GC803S, GC803P, the middle
opening is intended for the BPI switch (1) (see fi gure left).
With the CP225D it´s in an external electronics box
On models CP34001S, CP34001P, CP34000, CP16001S, CP12001S, on opening is
intended for the BPI switch (1) (see fi gure left).
Note:
To use the CAS programs (Version 1.44 and later) or the SARTORIUS MC1 Server
(version 4.9 and later and later), the balance/scale must be set to the BPI mode (=
Binary Processor Interface Mode) (e.g. for linearization /span adjustment or when
replacing the PCB).
- Turn on the balance/scale using the e key and wait for the segment test on the
balance/scale to run, i.e. until (e.g. CP423S) „ 0.000 g „ is displayed.
- Remove the capped plug from the rear panel of the balance, located to the left
of the DC jack (screw).
- The BPI switch is now accessible through the opening. Press the BPI switch once,
e.g. with a ball-point pen and hold it in:
- The weight readout in the display disappears (8-er-check it is executed).
- After about 3 seconds, the interface will be in the BPI mode.
- Release the BPI switch, the balance/scale now returns automatically to the
normal weight readout mode, but is still in the BPI Mode.
(Only SARTOCAS is functional).
- Insert the plug to close the opening on the back panel of the balance/scale.
- You can now use the balance/scale together with the SARTOCAS program
(version 1.44 and later) for PCs or the Psion server (version 4.9 and later) in
the BPI Mode.
Caution! After working in the BPI Mode, make sure to set the write-protect (with the
SARTOCAS program for SARTORIUS MC1 Server (Version 4.9 and later)), so that
the balance/scale returns to the standard data record output mode (SBI mode =
Sartorius Balance Interface).
If the balance/scale is in the BPI Mode (= Binary Processor Interface Mode) when you
press p key, the error code ERR 30 will be displayed!
You cannot run the standard peripheral devices until you set the balance/scale back
to the SBI mode!
Note: If necessary, it is possible to switch from SBI to BPI by activating the menu setting
»9 - 1« (menu reset).
Anz_00-F.eps
Anz_88-F.eps
Anz_OFF.eps
Aut_4106.ipg / CP_bpi_t.eps
Aut_4106.ipg / CP_bpi_t_sch.eps
1
1
Function Service BPI Switch
Anz_err30-F.eps