•Foruse in Zones1 (explosive gases)and Zone 21 (explosive dusts)
• ATEX:II2G ExiaIICT4GbII2DExiaIIICT80°CDb
•Allowsthe connection ofone analogue SartoriusplatformsCombics-Ex,IFXS4,IUXS4,oroneIS-Xprecisionplatforms,SignumExscale orload cellswith suitable specificationsto be connected in Zone 1,21,20
• VerifiableclassIII,IIII
•Veryeasyto read display
Sartorius Combics 2 Ex Indicator
for Use in Areas at Risk to Explosion
Model CAIXS2
98648-019-74
Contents
Contents
Notes on Using these Instructions ..................................... 3
Warnings and Safety Precautions ............................................. 4
Declaration of Conformity .................................................. 119
Declaration of Conformity ................................................... 120
Ex Safety Notes ........................................................................ 122
Menu Structure ..................................................................124
External Data Interface 66015-741-50 ........................... 140
Verification of Intrinsic Safety ....................................... 145
FM Certificate ................................................................... 150
General Password ............................................................... 177
2 Operating Instructions Combics CAIXS2
Notes on Using These Instructions
t Please read these instructions carefully and completely before using,
maintaining or repairing the device.
t Observe the safety instructions.
t These instructions are part of the product. Keep it in a safe and easily accessible
location.
t If the instructions should be lost or misplaced, please contact Sartorius
for a replacement or download the latest version from our website:
www.sartorius.com
Symbols and Signs
The following symbols are used in this manual:
3
Warning symbol for various types of dangers.
Notes on Using These Instructions
This symbol indicates useful information and tips.
h
e, 1, This and similar symbols mean that the respective key should be pressed.
T T ..., This means that this key must be pressed more than once.
t Indicates a required action
y Describes the result of an action
1. If a procedure has multiple steps...
2. ... the steps are numbered consecutively
– Indicates an item in a list
Menu Descriptions
In some cases, text descriptions are used to describe menu settings and in other
cases only the number structure of the menu is used for faster orientation for
experienced users (e.g. “Menu item 1.9« contains the parameter settings for
calibration/adjustment). The Setup menu is shown on the display when “codes”
is selected as the language.
The Combics CAIXS2 indicator complies with the European Council Directives
as well as international regulations and standards for electrical equipment,
electromagnetic compatibility, and the stipulated safety requirements. Improper use
or handling, however, can result in damage and/or injury.
t Read these operating instructions carefully before use.
This will prevent damage to the equipment.
Please observe safety instructions 65015-750-16 in the
section.
Please also bear the following points in mind:
safety information
Make absolutely sure to unplug the indicator from the power before you connect or
3
3
3
3
3
disconnect any electronic peripheral devices to or from the interface port.
If you use electrical equipment in installations and under ambient conditions
requiring higher safety standards, you must comply with the provisions as specified
in the applicable regulations for installation in your country.
The operator shall be responsible for any modifications to the equipment and for
any connections of cables or equipment not supplied by Sartorius and must check
and, if necessary, correct these modifications and connections.
Warning when using RS-232 cables, purchased from other manufacturers:
These often have incorrect pin assignments for use with Sartorius equipment.
Connect only Sartorius accessories and options, since these are optimally designed
for use with your device. Therefore, do not use any proprietary solutions. The
operator shall be solely responsible for installation and testing of any modifications
to Sartorius equipment, including connection of cables or equipment not supplied
by Sartorius. Information on operational quality (in line with norms pertaining to
immunity) is available on request.
Clean your equipment only as directed in the cleaning instructions (see “Care and
Maintenance").
The display value can be affected by extreme electromagnetic influences. Once the
disturbance has ceased, the instrument can be used again in accordance with its
intended purpose.
Information on operational quality is available upon request from Sartorius (in line
with norms pertaining to immunity).
t If you have any problems with your device, contact your local Sartorius office,
dealer or service center.
IP Protection
– The model meets the requirements of protection class IP69K.
4 Operating Instructions Combics CAIXS2
IP Rating
–
The IP65/IP69k protection rating is ensured only if the rubber gasket is
installed and all connections are fastened securely (including the caps on
unused sockets). Weighing platforms must be installed and tested by a certified
technician.
Equipment Description
Equipment Description
CAIXS2 The CAIXS2 Ex-Indicator offers the following features:
– robust and durable, thanks to its stainless steel housing
– easy to clean and disinfect
– easy to operate, thanks to the following features:
– large backlit display elements (14 segments)
– large keys with positive click action
– operation independent of the platform location
– range of interfaces for flexible use
– password protection to prevent unauthorized alteration of operating parameters.
CAIXS2 Offers the following practical functions:
– easy calibration via a separate key
– automatic tare for loading
– alibi memory installed optionally
– automatic printout for loading
– configurable print-out
– FlexPrint.
CAIXS2 Simplifies and speeds up your daily work with:
– integrated programs for applications (some can be combined):
– counting
– neutral measurement
– averaging/animal weighing
– weighing in percent
– checkweighing
– classification
– totalizing
– net-total formulation
– automatic initialization when the scale is switched on
– option to be controlled via an external computer using various protocols
– possibility of inputting tare values via the number block
– connection option for a second balance
– external battery
– product data memory.
Intended Usage
It is robust electrical equipment and suitable for use in daily quality control in industry for the
tasks previously specified. The Combics 2 CAIXS2 Ex-Indicator is designed for use with suitable
scales or weighing platforms that correspond to the described technical specifications. To do
this, the CAIXS2 and accessories must be used within the parameters of the specifications (see
Appendix).
The CAIXS2 Ex-Indicator meets the requirements set in EC Directive 94/9/EG for electrical
equipment in category II 2 GD and as such is suitable for use in potentially explosive Zone 1
and Zone 21 atmospheres.
Any other use beyond this is considered improper.
Operating Instructions Combics CAIXS2 5
General View of the Equipment
•Networking ofup to 8 indicators
•Foruse in Zones1 (explosive gases)and Zone 21 (explosive dusts)
• ATEX:II2G ExiaIICT4GbII2DExiaIIICT80°CDb
•Allowsthe connection ofone analogue SartoriusplatformsCombics-Ex,IFXS4,IUXS4,oroneIS-Xprecisionplatforms,SignumExscale orload cellswith suitable specificationsto be connected in Zone 1,21,20
• VerifiableclassIII,IIII
•Veryeasyto read display
2
1
5
CAIXS2 2
Display (for details, see “Operating Design" chapter)
4
3
1
On/Standby switch
2
General function keys: Zero, Tare, Switch function, Adjustment/
3
Calibration, Print/Data output
(see “Operating Design")
10 digit keypad for entering values
4
Additional function keys (see “Operating Design")
5
8
12
9
13
10
11
Rear View
WP weighing platform connection
8
Connection options for COM1
9
Adapter cable with plug for EX power supply
10
11 Equipotential bonding conductor
Input for menu access switch (standard or legal-for-trade mode)
12
Vent valve
13
6 Operating Instructions Combics CAIXS2
Installation
When a CAIXS2 indicator is ordered with special equipment, the desired options
come pre-loaded from the factory.
Storage and Shipping Conditions
Installation
Excessive vibrations may compromise the safety of the equipment.
3
– Do not expose the equipment to unnecessarily extreme temperatures, moisture,
shocks, blows or vibration.
– Permissible storage temperature: -20 to +60°C.
Installation Location
Avoid adverse influences at the place of installation:
– extreme temperatures (operating temperature: -10°C to +40°C)
– aggressive chemical vapors
– extreme moisture (according to IP protection rating).
Unpacking the Equipment
t After unpacking the device, check it for any visible damage as a result of rough
handling during shipment.
y If you detect any damage, proceed as directed under “Safety Inspection" in the
chapter entitled “Care and Maintenance."
tSave the original packaging for any future transport.
Unplug all connected cables before packing the equipment.
Checking Package Contents
– Indicator
– Operating instructions
– Options (special accessories) as listed on the bill of delivery
Acclimatizing the Device
Condensation can form on the surfaces of a cold device when it is brought into a
substantially warmer area.
t Allow the device to acclimatize for about 2 hours at room temperature, leaving
it unplugged from the supply voltage.
Operating Instructions Combics CAIXS2 7
Getting Started
Getting Started
1.) Connect weighing platform to the indicator.
2.) CAIXS2 Indicators have an intrinsically safe data interface which can be
connected to a computer (or any other peripheral device) using a barrier
(e.g. YDI05-Z).
3.) Connect the AC adapter.
4.) Configure the analog/digital converter (ADC).
5.) Carry out an alignment: Adjustment.
Connecting Weighing Platforms
You can connect any intrinsically safe, analog Sartorius platform to your CAIXS2
Indicator. Refer also to the Verification of Intrinsic Safety, the EC Type Examination
Certificate for the CAIXS2 and the load cell or analog weighing platform to be
connected.
t Open the Combics indicator:
The load cell should be connected by a certified technician who has received
3
3
specialized training from Sartorius. Any installation work that does not conform
to the instructions in this manual results in forfeiture of all claims under the
manufacturer’s warranty.
Make sure the CAIXS2 is disconnected from the power before starting any
connection work.
t Set up the weighing platform (see operating instructions for the weighing
platform).
t Place the cable from the weighing platform next to the indicator.
Loosen the ten cap nuts on the front panel. Remove the front panel.
t Remove the cover plate by removing the 4 screws.
Connection terminals are now visible.
3
8 Operating Instructions Combics CAIXS2
Installing Connection and Interface Cables
The cable gland (IP69K protection) is pre-mounted on the indicator.
Please use extreme caution when performing any work on the equipment that
affects this cable gland.
You must use a torque wrench. The torque for this cable gland is 5 Nm.
2
3
Getting Started
Preparing the Cable
t Strip approx. 14 cm from the end of the cable.
t Shorten the shielding to approx. 2 cm and pull back over the insulation.
t Strip approximately 5 mm of the insulation from the wires of the connecting
cable and affix ferrules to the wire ends.
Attaching the Cable Entry
Please use extreme caution when performing any work on the equipment that
1
4
5
t Insert the cable through the cable gland until the shielding (2) comes into
affects this cable gland.
You must use a torque wrench.
The torque for this cable gland is 5 Nm.
t Remove the protective caps from the bore hole on the indicator.
t Insert the included cable gland through the bore hole and secure from the
inside using the locknut (1).
contact with the clamps (3). Tighten the screw-down nut (4) until the gasket
(5) inserted between the screw-down nut and cable forms a small beaded rim.
t Check the shielding and clamps.
t Securely connect the wires of the connecting cable in accordance with the
terminal assignments.
Connecting the Cable to the Analog/Digital Converter (ADC)
t Insert all cable wires through the ferrite case, wind them around the ferrite case
and then reinsert back through the ferrite case.
t Screw the wires tightly into the clamps.
Pin Assignmentt Refer to the data sheet or operating instructions of the weighing platform for
details on the assignment of wire colors/signals. Ensure any lines that are not
assigned are insulated correctly.
Connect weighing platform to ADU, Option A15
1 EXC+ blue
2 Sense+ green
3 OUT+ white
4 OUT- red
5 Sense- gray
6 EXC- black
t When connecting a load receptor that uses 4-conductor technology (the cable
of the weighing platform to be connected only has 4 lines), connect clamp
pairs 1 and 2 (EXC+ and SENSE+), and 5 and 6 (SENSE- and EXC-) with a wire
jumper.
Operating Instructions Combics CAIXS2 9
Getting Started
1
2
Instead of an analog/digital converter (ADC), you could also install a data interface
to connect an intrinsically safe digital weighing platform or balance (e.g. an
Connect intrinsically safe data cables to COM 1 (RS232, RS485 or RS422 and
intrinsically safe control signals). For pin assignments on the data interface board,
see 66015-741-50.
External voltage supply for AP1
At AP1 connected IS platform requires an external
Voltage supply, see document 66015-741-50 Page 2/9.
10 Operating Instructions Combics CAIXS2
Pin Allocations on the CAIXS2 Data Outputs (COM1)
O
D
K
A
M
N
J
G
B
C
E
F
H
L
Option A21 A22 A23
RS232 + RS422 RS485 + Pin*) Pin**)
Digital I/Os Digital I/Os
CTS GND GND A 1
RxD GND TxD-RxD_P J 2
TxD TxD_N TxD_RxD_N K 3
DTR TxD_P --- N 4
GND DRT_P GND C 5
GND RxD_N GND M 6
GND DTR_N GND B 7
UNI_IN --- UNI_IN O 8
SET --- SET D 9
PAR CTS_N PAR E 10
MIN CTS_P MIN F 11
MAJ RxD_P MAJ G 12
Getting Started
1 2 3 4 5 6 7 8 9 10 11 12
* 14-pin plug on adapter cable:
** 12-pin terminal block on the data adapter board:
Plug the 12-pin connection cable into the corresponding type of data output
(see data sheet External Data Interface).
External voltage supply for COM1
At COM1 connected IS platform requires an external
Voltage supply, see document 66015-741-50 Page 3/9.
Operating Instructions Combics CAIXS2 11
Getting Started
t Insert the cover plate and tighten the screws.
t Re-attach the front panel and tighten the ten cap nuts with a torque of 1 Nm.
Connecting the Device to AC Power
3
Power is supplied via an external power supply device, which is provided with the
equipment.
The power connection must be made in accordance with the regulations applicable
in your country.
Make sure that the voltage rating printed on the manufacturer's ID label is identical
to that of your local mains voltage. If the voltage specified on the label or the plug
design of the AC adapter do not match the rating or standard you use, please
contact your Sartorius office or dealer.
t Check the voltage rating and plug design.
t The device must be plugged into a properly installed wall outlet.
Protection Class 1 Device
t The device must be plugged into a properly installed wall outlet that has
a protective grounding conductor (PE).
Safety Precautions
If you use an electrical outlet that does not have a protective grounding conductor,
3
ensure that an equivalent protective conductor is installed by a certified electrician
(as specified in the applicable regulations for installation in your country).
The protective effect must not be negated by using an extension cord without
a protective grounding conductor.
Before using for the first time, any superstructure parts must be completely
installed.
Avoid connecting the equipment to lines that have a heavy electrical load,
e.g. compressors, large machinery, etc.
Warm-Up Time
To deliver exact results, the device must warm up for at least 30 minutes after
connection to AC power. Only after this time will the device have reached the
required operating temperature.
12 Operating Instructions Combics CAIXS2
Operating Design
•Networking ofup to 8 indicators
•Foruse in Zones1 (explosive gases)and Zone 21 (explosive dusts)
• ATEX:II2G ExiaIICT4GbII2DExiaIIICT80°CDb
•Allowsthe connection ofone analogue SartoriusplatformsCombics-Ex,IFXS4,IUXS4,oroneIS-Xprecisionplatforms,SignumExscale orload cellswith suitable specificationsto be connected in Zone 1,21,20
• VerifiableclassIII,IIII
•Veryeasyto read display
Operating Design
Display and Keypad
17 16
18 15
19 14
1 13
2
3 10
4
5 8
6 7
12
11
9
1 Display
2 On/Standby key
3 Keys with no function
4 Zero key
5 Tare key
6 Function key unit conversion
7 Start calibration or adjustment
8 Print key (data output)
9 Toggle unit between normal and 10-fold
higher resolution
10 View gross value (net value plus tare)
View net value (gross value minus tare)
11 Product data memory
12 ID key for entering the operator ID
13 Numeric keypad
14 Toggle between application program and
application- specific information
15 Display of applications and manual tare
values
16 Toggle key
(function depends on application)
17 OK key
(function depends on application)
18 Reference value key
(function depends on application)
19 Clear function key
(function depends on active application)
Operating Instructions Combics CAIXS2 13
Operating Design
•Networking ofup to 8 indicators
•Foruse in Zones1 (explosive gases)and Zone 21 (explosive dusts)
• ATEX:II2G ExiaIICT4GbII2DExiaIIICT80°CDb
•Allowsthe connection ofone analogue SartoriusplatformsCombics-Ex,IFXS4,IUXS4,oroneIS-Xprecisionplatforms,SignumExscale orload cellswith suitable specificationsto be connected in Zone 1,21,20
• VerifiableclassIII,IIII
•Veryeasyto read display
Key Functions
e On/Standby switch
When in Standby mode, STANDBY is displayed.
( Zeroing key
– Press key for less than 2 seconds: Zero
– Press key longer than 2 seconds: Display the adjustment/configuration
counter
) Tare key
– Saves the numeric input as the tare weight
– Press key longer than 2 seconds: Save the preset tare weight
kFunction key: Depending on the configuration in the Setup menu, switches
between the
– first and second weighing unit
– results display and SQmin display.
JISO test: Start calibration or adjustmentpPrint key
– Press key for less than 2 seconds: Print
– Press key longer than 2 seconds: Print GMP footer
nToggle key: When a second platform is connected (COM1), this key toggles the
display between the two readouts.
The following four keys are used for operating the individual applications. Their
exact function is described in the respective section for the application.
cDelete key: Deletes initialization values or totalizing memory. During numeric entry
the last character entered is deleted.
rReference value key: Changes the set reference value.
OOK key: Applies values or starts an application program.
wToggle key: Toggles between display modes within an application program.
14 Operating Instructions Combics CAIXS2
Operating Design
1, 2, 3 ... ., 0Number block: Used to enter numeric values
DApplication toggle key: Toggles between available applicationsdID key: Used to enter operator IDs
RSave key: Used to save values to the product data memory or load to the
KResolution toggle key: Toggles unit between normal and 10-fold higher resolutionLGross/Net value key: Toggles between the gross or net value
I Info key: Used to display application parameters and manual tare values (Info after
pressing a follow-up key, e.g. ))
t To apply the value, press the corresponding function key (e.g. key ) to save
the entry as a manual tare value.
t To delete the last character entered, press the c key.
application
Saving Settings in Weighing Mode
All application parameters saved (e.g. reference values) remain in memory and are
available when:
– the device has been switched off and then on again
– you return to the originally selected application from a second one (e.g. when
you switch from Averaging back to Counting. all parameters saved for Counting
are available).
Applying the Tare Weight
t Place the tare object on the weighing platform.
t Press the T key.
y The value is applied as the tare value.
Input Through the Digital Control Port
You can connect an external hard drive or foot switch to the control port (universal
interface). You can assign one of the following functions to the control port in the
SETUP / CTRL IO / INPUT / PARAMET / EXT.KEYB menu:
– p key
– p key (hold)
–
) key
– J key
– k key
– n key
– O key
– ( key
– e key
– c key
– I key
– D key
– K key
– L key
Operating Instructions Combics CAIXS2 15
Operating Design
13
4
5
6
7
8
9
10
11
12131415
16
17
18
19
20
Appl. 1Appl. 2Appl. 3
2
The Display
Display in Weighing Mode
There are two display modes:
– display for weighing (weighing values and calculated values)
– display in “Menu mode" (device settings).
1 Bar graph showing 10% intervals
– shows the percentage of the weighing platform’s capacity that is “used
up" by the load on the scale (0% = lower limit, 100% = upper limit)
or
– shows the measured value in relation to a target value (with the
“Checkweighing" or “Classification" applications).
Minimum for checkweighing
Maximum for checkweighing
Target value for checkweighing
2
S
Symbol for active print job
3R8 Displays the active range on multiple-range scales
4 Indicates active weighing platform; flashes to prompt calibration/
adjustment
51 2 Selected weighing platform 1 or 26B/G NET Net/Gross value on the main display (with tare in memory or preset tare)7k Identifies the value on the main display as calculated (value not valid in
legal metrology)
8Battery charge status
9pGMP-compliant printing in progress
10Weight unit of the value displayed
11Numeric display; e.g. showing the reference value
12 Numeric display; e.g. showing the reference value
– Interface initialized (profibus/Ethernet)
– Flashes during data transfer (RS-232/485)
13MemSymbol for product data memory
14 In legal metrology, on equipment for which e is not equal to d, the digit
bordered for identification is not taken into account
15AUTO/OPT
– AUTO: Depending on the weight value, a reaction is triggered in the
– OPT: Automatic optimization takes place for the Counting application
application
16 Measured value line: weight value or calculated value
16 Operating Instructions Combics CAIXS2
Operating Design
13
4
5
67
89
10
2
17 Symbols for applications: An active application is identified by a line above and
below the symbol (
).
Application 1: A “Counting"/ “Neutral Measurement"
B “Weighing in percent"
V “Averaging" (animal weighing)
Application 2: H “Checkweighing"
W “Classification"
Y “Checkweighing toward zero"
Manually batching toward “zero"
Application 3: L “Totalizing"
M “Net total formulation"
18
The zero-setting symbol is displayed after the active scale or weighing
U
platform has been zeroed (verified models only)
19+ –Plus or minus sign of the value displayed
20 l Busy symbol indicates that an internal process is in progress
Menu Operating Concept
Switching to the Menu
et Turn on the device.
If it is already on: turn off and then on again.
y The menu will open. The top most level is always displayed (APPL, ); for menu
Navigating the Menu
t During the display test, briefly press the ) key.
)
structure see page 136.
You can navigate the menu using the keys with the white arrows under them.
( Back to the superordinate menu level
k Access the next menu item on the same level
This continues to page through on the same level
) Press less than 2 seconds: Select the menu item and save
Press longer than 2 seconds: Exit the menu and switch to weighing mode
p Print the menu settings starting from the current position, or print Info data
Operating Instructions Combics CAIXS2 17
Operating Design
Entering Numbers and Letters (without a Number Block)
( – Press the key for less than 2 seconds: Activate character to the left of the
currently active character (when first character is active: exit the input
mode without saving changes)
– Press the key for longer than 2 seconds: Exit the input mode without
saving changes
( – Press the key for less than 2 seconds:
Confirm currently active character and move 1 position to the right (after
the last character: save input)
– Press key for longer than 2 seconds: Save current input and display the
menu item
k – Cursor in first position, no characters entered yet: Delete character(s) and
p – Cursor in first position, no characters entered yet: Delete character(s) and
enter a space
–
Enter number values (date and time, etc.) using the 10-key numeric keypad.
Menu Display
Both illustrations depict all of the main display elements and symbols that can be
shown in Menu mode.
1 Selected menu item (e.g. printer for setting the connected printer)
2 Menu history (note at highest menu level in the Setup menu)
1
4567
3 Note that other submenus are available
2
3
Display with the “codes" language setting
4 First level in the Setup menu
5 Second level in the Setup menu
6 Third level in the Setup menu
7 Currently active setting
2
3
Saving Menu Settings
Change the displayed character; scroll backward (sequence:
space, A _ Z, minus sign, decimal point, 9 _ 0)
18 Operating Instructions Combics CAIXS2
The parameters selected in the menu remain saved when you switch to weighing
mode or turn off the device. You can block access to the Setup menu by requiring
a password to prevent unauthorized or accidental setting changes.
Configurations
Basic settings are made in the Menu mode by selecting the desired parameters.
These are divided into the following groups (first menu level); for menu structure
see page 136:
– Application parameter APPLIC.
– Function key fn-key
– Device parameter SETUP
– Device-specific information Info
– User language LANGUAG.
When used in legal metrology, not all parameters can be accessed. Only those
parameters that can be selected are displayed. Factory-set parameters are identified
by an “*" in the list starting on page 137.
Printing Parameter Settings
t Access the Menu mode (see page 35).
t Press the p key.
The scope of the printout depends on the position in the setup. It may take several
seconds.
Operating Design
Language Settings
Example: Select the language “German." The factory setting for language is
“English."
Menu: appl / LANG .
et Turn on the device.
y The first item in the main menu is shown: APPL.
k k ... t Press the k key until the LANGuag. menu item is displayed for the language
t While all segments are lit, press the ( key briefly.
setting.
uag.
)t Press the ) key to access the language setting sub-menu.
y The currently set language is displayed.
k k ... t Press the k key until GERMAN is displayed.
)t Press the ) key to save the selection.
Operating Instructions Combics CAIXS2 19
Operating Design
y The small circle indicates that the setting has been saved.
°
)Press the ) key longer than 2 seconds to exit the menu.
( Use the ( key to exit the menu level to make additional settings if required.
or
Setting Up Password Protection
et Turn on the device.
y The first item in the main menu is shown: APPLIC.
k k ... t Press the k key until the SETUP menu item is displayed.
t While all segments are lit, press the ( key briefly.
) t Press the ) key to open the Setup sub-menu.
y The first parameter in the Setup sub-menu is displayed: WP-1.
k k ... t Press the k key until BEN.-CODE is displayed.
) t Press the ) key to open the menu item.
y The position for the first character to be entered flashes.
p kt Use the p and k keys to select the desired character.
p starts the character selection with A alphabetically and
k starts the character selection with 0 and counts upward.
Alternatively, enter a number value using the 10-key numeric keypad directly.
20 Operating Instructions Combics CAIXS2
Operating Design
) t To apply a character, press the ) key.
t Enter all additional characters of the password as described above.
t Press and hold the ) key to save the password.
( Use the ( key to exit the menu level to make additional settings if required.
or
)Press the ) key longer than 2 seconds to exit the menu.
Changing or Deleting Passwords
t In the setup sub-menu, open the ben.code menu item as described above.
y The old password must be entered to change or delete a password.
t To change a password, overwrite the old password.
t To delete a password, enter spaces and press the ) key.
Operating Instructions Combics CAIXS2 21
Configuring Weighing Platforms
Configuring Weighing Platforms
Service mode
Purpose The Service mode enables access to additional menu items in the Setup menu
(setup) which are not displayed when the Service mode is not active. The most
important calibration and adjustment work for the indicator and for the connected
weighing platform can be carried out in the Service menu, e.g. ADC configuration.
When the Service mode is active, an “S" is shown in the top right-hand corner of
the display. To deactivate the Service mode, restart the indicator (turn the indicator
off and back on again).
In Service mode, the Setup menu is expanded with the following parameters after
entering the user password:
– S-DATE for entering the next service date
– SER.No. for entering the device serial number
– MODEL with the model description
– S-SQMIN
– ALIB.MEM for deleting thealibi memory.
The Setup menu for WP1 and WP2can be extended to include the following setting
options to confugure the weighing platforms:
Param1
lIN.ext. external linearization with default weights 1.9.6
lIN.e.usr external linearization with user-defined weights (entered under 1.18) 1.9.7
SET.PREL. set the preload (not for use in legal metrology) 1.9.8
DEL.PREL delete the preload (not for use in legal metrology) 1.9.9
HND.XT.G/CAL.ADJ enter the adjustment and linearization weights 1.18CAL.ADJ.1 enter adjustment weight 1.18.1
LIN. WT.1 enter lin. weight 1 1.18.2
LIN. WT.2 enter lin. weight 2 1.18.3
LIN. WT.3 enter lin. weight 3 1.18.4
LIN. WT.4 enter lin. weight 4 1.18.5
ADJ.W/O.W adjust without weights (enter the characteristic data of the load cells) 1.19 NOM.LOAD. nominal load 1.19.1
RESOLUT resolution (only visible for older ADCs)* 1.19.2
SENSIT.1 sensitivity in mV/V for cell 1 (or average value for all cells) 1.19.3
SENSIT.2 sensitivity in mV/V for cell 2 1.19.4
SENSIT.3 sensitivity in mV/V for cell 3 1.19.5
SENSIT.4 sensitivity in mV/V for cell 4 1.19.6
ZER.POIN zero point or offset of system in mV/V 1.19.7
SAVE. save values for 1.19 1.19.8
GEOG.DAT adjustment location (geograph. data; or alternatively the gravitational acceleration
at the place of installation) 1.20
LATITUD latitude in degrees 1 20.1ALTITUD elevation in meters above sea level 1 20.2GRAVITY. gravitational acceleration 1 20.3SAVE. save values for 1.20 1 20.4
ADC settings (see Analog/Digital Converter) 11
Apply the serial number of the IS weighing platform 12.1
Apply the serial number 12.1.1
Inactive (standard WP) 12.1.2
22 Operating Instructions Combics CAIXS2
Activating the Service Mode
e ... ) t Switch to the Menu mode.
k k ... t Access the Setup menu.
) t Select Setup.
If a password is requested at this point, enter the service password
(see Appendix) and continue with “Saving the service password."
Configuring Weighing Platforms
k k ... t Access menu item U-CODE.
) t Select ben.Code.
t Enter the service password (see Appendix).
) t Apply the service password.
y The Service mode is active: an “S" appears in the top right-hand corner of the
display.
( ( t Return to Setup in the Service mode.
Operating Instructions Combics CAIXS2 23
Configuring Weighing Platforms
Analog/Digital Converter (ADC)
Purpose Adjust the parameters of the analog/digital converter to the connected load cell or
weighing platform. After ADC configuration, the ADC in connection with the load
sensor is defined as a scale.
Once the ADC configuration has been locked, the indicator can no longer be used
3
to influence weighing results. The scope of functions available in the weighing
instrument is defined by the A/D converter. Weighing functions that can be
activated include reading weight values, taring, adjustment, reading the tare
value and saving/deleting the tare entry.
Setup information – ADC configuration is only possible when the menu access switch is open. Close
the menu access switch after ADC configuration, as otherwise there will not be
any display of the conditions “overload" ( H ) and “underload" ( L ).
– Before ADC configuration, you must first set whether or not the weighing
platform will be used as a standard or verifiable weighing platform under menu
item 9.1.
– When the Service mode is activated, the ADC configuration takes place in the
Setup menu under WP-1 for the first weighing platform and under COm1 /
WP-2 for the second weighing platform.
If you return to the highest level of the Setup menu without saving the
3
configuration parameters beforehand (menu item 11.10) any settings that have
been made will be deleted.
– The settings are made in the corresponding Setup menu under menu item 11.
– Enter the maximum capacities in a suitable weight unit, without any decimal
places (decimal places will be truncated by the rounding function).
– Entries made in the ADC configuration will not be affected by a menu reset
(returning the setup parameters to their factory settings).
Standard or verifiable configuration In ADC configuration, you must first select whether the weighing platform should
be configured as a standard or verifiable (for use in legal metrology) weighing
platform.
– Standard configuration STANDRD (9.1.3)
– Verifiable configuration VERIF. (9.1.4)
Configuration unit1.WT.UNIT Menu item 1.7
The weight unit used in the ADC configuration must have previously been selected
here.
Range selectionRANGE Menu item 11.3
Depending on the setting under this menu item, the Menu items 11.5, 11.6 and
11.7 will either be displayed or will not be displayed for further configurations.
– Single-range scale (11.3.1)
The entire weighing capacity is divided into decimal numbers dependent on the
smallest scale interval d and the maximum weight. The readability corresponds
to the scale interval d.
– Multiple-range scale (11.3.2)
A multiple-range scale has two or three weighing ranges. When the range limit
for the lower weighing range is exceeded, the scale switches into the next
highest weighing range (lower resolution). The scale only switches back to the
lower weighing range (higher resolution) when the weighing platform has been
completely unloaded after pressing the key (.
– Multi-interval scale (11.3.3)
The function “Multi-interval scale" divides the weighing capacity into a
maximum of three ranges with differing readability. The corresponding change
takes place automatically at the defined range limits. Once the scale has been
tared, the highest possible resolution is available even if the weighing platform
is loaded.
Scale interval d Scale interval d indicates the resolution of the weighing instrument.
The scale interval can only be entered in increments of 1, 2, 5, 10, 20, etc.
When “Verifiable configuration” is used, this menu item is not displayed. When
using verifiable or verified weighing platforms (classes l and m), scale interval d is
the same as verification scale interval e.
Verification scale interval e Verification scale interval e indicates the resolution of the weighing instrument in
legal metrology.
The scale interval can only be entered in increments of 1, 2, 5, 10, 20, etc.
When “Standard configuration" is used, this menu item is not displayed.
Maximum load (max. load) The maximum load is the maximum amount of weight that may be placed on the
weighing platform. When heavier weights are used the weighing instrument
displays overload “H".
The scale intervals of the weighing instrument are calculated using the maximum
load and the scale interval d (e.g. max. capacity = 15.000 kg, smallest scale interval
d = 0.005 kg yields 3000 scale intervals).
In legal metrology the total number of intervals must be no more than 3000 e, and
when using multi-interval scales there must not be more than 3000 e intervals per
range.
In standard operation, as opposed to legal metrology, you can define a “Super
Range” weighing instrument of over 3000 intervals. These parameters, however,
may be influenced by physical restrictions.
26 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Minimum load (min. load) When “Standard configuration” is used, this menu item is not displayed.
The minimum load of the connected weighing platform is entered under this menu
item. The minimum load for scales of class l is 20 e and 10 e for class m. Attention: The function of the minimum load setting is to warn operators that
below this limit, the summation of tolerances might lead to significant measurement errors. In Germany, for example, initial weights below the minimum load are
not allowed.
Available weight unitsWT.UNIT Menu item 11.7
This menu item is used to select the weighing units that have been cleared for use
in weighing. All units marked with a circle (o) have been cleared for use, multiple
selection is possible.
Calibration/Adjustment unitCal.Unit Menu item 11.8
This menu item is used to select the weighing unit that must be used for a
calibration/adjustment. The selected unit is then valid as a calibration/adjustment
unit even when a different unit is used during normal weighing operation.
Save parametersSAVE Menu item 11.10
The ADC configuration data is saved by selecting Menu item 11.10.1.
Operating Instructions Combics CAIXS2 27
Configuring Weighing Platforms
Configuring the A/D Converter (ADC)
The weighing platform must already be connected.
Opening the Menu Access Switch
The menu access switch is located on the back of the indicator, behind the cover.
t Remove the cap.
t Slide the switch to the right (= “open" position).
Menu Access Switch/Calibration Switch Cover
yADC-CON appears briefly on the display, followed briefly by S-CODE.
y The cursor flashes on the display.
) t Confirm your entry using the ) key.
y The device is in Service mode. This can be recognized by the small S in the top
) t Confirm your selection using the ) key.
t Select the Configuration mode using the k key: STANDRD or VERIF.
et Switch off and restart the device.
t While all segments are lit, press the ( key briefly.
t Enter the service password (see Appendix).
right of the display.
t Select the weighing platform to be configured, using the k key to switch to
WP-2 if required.
t Carry out ADC configuration (see menu tree).
t Once you have completed the configuration, save the data using the SAVE
menu item.
y The indicator will restart automatically.
28 Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
The A/D converter can now be treated like a standard weighing platform in
connection with the load sensor.
t Close the menu access switch (left position) and reattach the cap.
Installation location in Germany An exception to this is the setting for “Germany (Zone D)": If during external
Once ADC configuration has been completed, an adjustment of the weighing
h
platform (calibration/adjustment and linearization) must be carried out
(see “Calibration/Adjustment without Weights").
adjustment of weighing equipment within Germany the geographical data
– geographical latitude: 51.00 degrees N
– elevation: 513 m
is entered, the weighing equipment can be used throughout Germany.
Gravitational acceleration for “Germany (Zone D)" is 9.810 m/s
On delivery the geographical data for “Germany (Zone D)” is entered in the output
device.
It is recommended to use the geographical data settings for “Germany (Zone D)"
when adjusting and delivering the weighing equipment within Germany. Entering
exact geographical data will lead to a higher level of accuracy but will also restrict
the tolerance zone.
2
.
Setup information – It is only possible to enter geographical data when the menu access switch is
open.
– When the Service mode is active, geographical data can be entered in the
Setup menu under “WP-1" for the first weighing platform and under COm1 /
WP-2 for the second weighing platform. The settings are made in the
corresponding Setup menu under menu item 1.20.
– Either the “geographical latitude in degrees" (latitude menu item 1.20.1)
and “elevation in m above sea level" (elevation menu item 1.20.2) or the
value for gravitational acceleration (GRAVITY menu item 1.20.3).
Gravitational acceleration takes precedence over the geographical latitude and
elevation of the location: If it has been entered, input fields for latitude and
elevation show the values 99999.99 and 9999999 respectively. If only elevation
and latitude have been entered, 0000000 is displayed for gravitational
acceleration.
If you return to the highest level of the Setup menu without saving the
3
configuration parameters beforehand (menu item 1.20.4) any settings that
have been made will be deleted.
Operating Instructions Combics CAIXS2 29
Configuring Weighing Platforms
Proceduret Open menu access switch.
If the device is part of a verified weighing facility, this will only be possible if
the verification seal is broken. The weighing equipment must then be verified
again.
t Activate the Service mode.
t Select the weighing platform.
t Enter the geographical data for the place of adjustment under menu items
1.20.1 to 1.20.3 and save them under menu item 1.20.4. The data can be
obtained from the relevant land registry or Ordnance Survey.
t Carry out external calibration.
t After the calibration, enter the geographical data for the place of installation
under menu items 1.20.1 to 1.20.3 and save them under menu item 1.20.4.
t Close the menu access switch.
y The weighing equipment can now be operated at the place of installation, and
within the abovementioned tolerance zone.
Note: The set geographical values are displayed during the adjustment procedure if the
display of the data has been activated in the Setup menu under UTILIT menu item
8.12.2 (factory setting: 8.12.1, display deactivated).
When the display is activated the adjustment procedure is as follows:
y If the elevation and geographical latitude are used, after the start of the CAL
adjustment procedure the word ALTITUD will appear briefly followed by the
set elevation (in meters above sea level).
t Confirm the display using the ) key (cancel using the () key.
y Then the word “LATITUD" will be displayed briefly followed by the set
geographical latitude in degrees.
t Confirm the display using the ) key (cancel using the () key.
y You are then asked to place the calibration weight on the weighing platform.
If gravitational acceleration has been entered instead of elevation and
geographical latitude, the word GRAVITY t will appear briefly, followed by
the value set for gravitational acceleration.
t Confirm the display using the ) key (cancel using the () key.
Menu structure for entering the geographical data
GEOG.DAT adjustment location (geograph. data; or alternatively the gravitational
acceleration at the place of installation) 1.20
LATITUD latitude in degrees 1.20.1ALTITUD elevation in meters above sea level 1.20.2GRAVITY. gravitational acceleration 1.20.3SAVE. save values for 1. 20 1.20.4
30 Operating Instructions Combics CAIXS2
Loading...
+ 150 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.