Sartorius Combics 1, Combics 2, CISL2U, CAPP.U-***-LU, CAPS.U-***-LU Service Manual

...
WCI5005-e03104
Service Manual
Sartorius Combics 1 | Combics 2
Models CIS1U | CIS2U | CISL1U | CISL2U for Combics Indicators
and
Models CAPP.U-...-LU | CAPS.U-...-LU for Combics Weighing Platforms
C
F
REF
O
K
2
Contents
Overview
04 Service Concept 04 General Information 05 When to Use Which Procedure 05 Combics Indicators 07 Overview of the Models 07 Auxiliary Service Tools and Equipment 08 Accompanying Literature
09 Operating Concept
09 Keys below the Display 10 Configuring the Combics 10 Purpose, Features
12 Function of the Keys when Configuring the Menu 12 Printing the Parameter Settings 13 Description of the Equipment 14 Function of the Service Keys and Switches
14 A. Menu Access Switch 14 B. SBI/BPI Key 15 Activating the BPI Mode 15 C. Boot Key 16 Activating the xBPI Protocol
17 Activating the Service Mode
17 Activating the Service Mode
19 Working in the Service Mode
19 Additional Menus in Service Mode 22 Menu Item 1-18:
Entering Calibration/Linearization Weights
22 1-18-1: External User-defined Calibration
Weight (Service Mode not Required) 23 1-18-2 1. Entering 1st Linearization Weight 24 Calibration/Adjustment Routines in Service Mode (General
Information) 25 Valid areas for Use 26 Zone “D” 27 Menu Line 1-9: Calibration/Adjustment Functions 28 1-9-1 External Calibration/Adjustment with Default
Weights
28 1-9-3 External Calibration/Adjustment with User-
defined Weights
30 1-9-5 Internal Linearization (Only for WP1 on COM1 or
UniCOM Ports) 30 1-9-7 External Linearization with User-defined Weights 33 1-9-8 Setting the Preload 34 1-9-9 Clearing the Preload
36 Adjustment without Weights
36 Menu Item 1-19:
Entering the Specifications of the Load Cell(s)
39 Menu Item 1-20: Entering the Geographical Latitude and
Altitude, or the Acceleration of Earth Gravity
40 Entering the Place of Adjustment with Subsequent
Adjustment (Example)
40 Menu Item 1-13: Zero at On
46 Configuring the A/D Converter of the Combics Indicator
(Service Mode)
46 Menu Item 11: A/D Converter Settings (Configuration:
Standard, Trade)
3
47 A/D Converter Configuration (General Information) 47 Notes on A/D Converter Configuration 48 Descriptions of the Individual Menu Items 49 Selecting the Weighing Range Structure
(Menu Item 11-3)
52 Menu Item 9-1: A/D Converter Configuration (Example:
Trade Configuration)
57 Menu Item 9-1: A/D Converter Configuration with Load
Cell(s) Connected (Example: Standard Configuration)
58 Loading the Standard Configuration
(if the Trade configuration is currently loaded)
62 Menu Item 1-19: Entering the Specifications of the Load
Cell(s) 64 Entering the Date of Service 65 Entering the Serial Number of the Combics Indicator 67 Entering the Model Designation of the Combics Indicator 68 Defining Transaction Numbers for Data Records (Weight
Values) which were stored in the Alibi Memory
69 Repairing the Combics Indicator
69 Replacing the Front Panel 70 Blank Display 70 Replacing the Power Cord 71 Replacing the Power PCB 72 Replacing the DC/DC Converter 73 Replacing the A/D Converter 73 Quick-test of the A/D Converter 74 Replacing the Digital PCB 75 Checking the IP67 Protection 75 Torque Values for Cable Glands and Vent Valve
76 Adjusting the Weighing Platform
76 Adjusting the Off-center Load 76 Procedure (Example Illustrated on the Left) 77 Diagram: Determining the Adjustment Resistance 78 Off-center Load Adjustment: 0-ohm Resistors (Example) 79 Setting the Overload Stops 79 1. Weighing Platforms with 1 Load Cell: Steel 80 2. Weighing Platforms with 1 Load Cell: Stainless Steel 81 3. Weighing Platforms with 4 Load Cells: Steel 82 4. Weighing Platforms with 4 Load Cells: Stainless Steel
83 Repairing the Weighing Platforms
83 Replacing the Connecting Cable 84 Replacing Load Cells 85 Torque Values 86 Pin Assignments in the Junction Box 87 Replacing the Load Cell Foot
87 Color Codes of the Wiring for Weighing Platforms, Models
CAPS../CAPS.. 88 Error Codes 90 The Combics Weighing Platforms 91 Type Designation
91 Complete Combics scale 92 Complete Combics stainless steel scale
93 Service Specifications
93 CAPP/CAPS1U-.....-LU
94 CAPP/CAPS4U-.....-LU
4
Overview
Service Concept
Prerequisites for performing maintenance and repair work on Sartorius Combics scales requires considerable experience with both indicators and weighing platforms. In case of defects, repairs are performed on site. Generally, the equipment is not replaced.
General Information
– Do not connect or disconnect cables to or from the equipment; always disconnect the
power cable from the wall socket (mains supply) first!
– To ensure safety, an isolating transformer must be installed between the indicator and
the power supply before performing work that entails opening the Combics indicator housing.
On Combics scales, check and adjust as necessary:
(The procedure is the same as for all scale and weighing platform models.)
– Repeatability (standard deviation) – Off-center load – Span – Linearity
5
When to Use Which Procedure
Combics Indicators
Activate the service mode to perform all adjustments See page 17
Important: All adjustments required after repair work (such as
replacement of the A/D converter or the load cells) are performed using the terminal. The Sartocas service software is not required for these adjustments. Using this software, however, can make some adjustments unnecessary (see „Use Sartocas software“ below).
Activate the BPI mode to load a modified data record in the controller
of the A/D converter See page 15
Activate the xBPI protocal to perform calibration and adjustment, if necessary See page 16
using Sartocas
Use Sartocas software - to reactivate the SBI protocol by selecting the „Close“
command on the Combics terminal (or by restoring factory defaults (menu line 9-1)
- to program a new A/D converter PCB with data read from the old PCB
The necessary adjustments and customer specific menu settings are not necessary
6
Use PPLoader software to update the application software in the terminal
Drift, display Inf 02 when starting an adjustment routine to check the A/D converter (quick-test) See page 73
or with the strain-gauge simulator, check load cell(s) with the load cell tester
Display blank (dark) Disconnect the equipment from the power supply and
disconnect all cables and wires from the subassemblies, connect equipment to power again, connect subassemblies in turn and measure voltages See pages 70-72
No communication with connected equipment Check connections and data transfer parameters;
start internal test program; if necessary, replace
7
Overview of the Models
The hardware configuration consists of either a Combics 1 or 2 indicator and Sartorius weighing platform(s), or a Combics 1 or 2 indicator and weighing platform(s) (strain-gauge load cells) from another manufacturer.
Important: Mechanical and electrical service or repair work on the Combics scale requires
considerable experience, and for this reason should be performed only by Sartorius technicians trained at the factory. Any attempt to perform repair work can result in damage to the equipment.
Auxiliary Service Tools and Equipment
In addition to standard tools, you will need the following special tools to work on the Combics scale:
Qty. Designation Order No.
1 Laptop 1 Sartocas service software, version 1.46 or later 6740-33
or 1 Psion Server version 4.10 or later 1 PPLoader software 1 Isolating transformer 1 Flintec LCT-01 load cell tester 1 Socket wrench, double hexagon, 10-32 mm,
“ square driver 1 Adapter for Allen screws, SW 5 - 17, ” square driver
8
Qty. Designation Order No.
1 Torque wrench, 10-120 Nm 14x18, stainless steel 1 Torque wrench, 60-320 Nm 14x18, stainless steel 1 Set of open-ended wrenches with sockets 1 Set of sockets, up to 30 mm (hexagonal, stainless steel) 2 Transport belt (to move load plates or weighing
platforms; available from specialist suppliers)
2 Ring lugs (stainless steel; for moving the load plates or
weighing platforms; available from specialist suppliers) 1 Set open-ended wrenches (spanners) (up to 24 mm, stainless steel) 1 Threaded rod, M 16x120, for lifting the weighing platform
(available from specialist supplier) 1 Testing device for checking IP66/67 protection (not yet defined) 1 Set of screwdrivers (slotted) 1 Set of screwdrivers (Phillips head ) 1 Set of Allen wrenches 1 Digital voltmeter (Beckman) 6738-62 1 Strain gauge simulator 6740-74 1 RS-232/485 converter 6740-68
Accompanying Literature
Operating instructions for “Combics 1 | 2 Indicators“ WCI6004-e03081 98648-012-27 Operating instructions for “Painted or Stainless Steel Weighing Platforms“ WCA6004-e03081 98648-012-26
9
Operating Concept
Keys below the Display
e On/off
(On) Turns the Combics on and off. (Standby) Combics switched off displays
“off“.
n Toggle the display between (Scale #) the weighing platforms
(Combics 2 only) With two weighing platforms connected, this key toggles the display between the two readouts.
( During weighing: (Zero) Zero the scale
In the operating menu:
Closes active submenu and returns to next higher menu level and menu level (unless the first menu level is already shown) During alphanumeric input
in the operating menu:
– Press briefly:
Activate character to the left of the currently active character (with the first character active: Exit the input mode without saving changes)
– Press and hold (> 2 sec ):
Exit the input mode without saving changes
) During weighing: (Tare) – Press briefly:
Tare the scale
– Press and hold (> 2 sec ):
Calibration/adjustment
In the operating menu:
– Press briefly:
Display the next lower menu
level or Select and store a menu item
– Press and hold (> 2 sec ):
Exit the operating menu During alphanumeric input in the operating menu: – Press briefly:
Activate character/s to the
right of the currently active
character ®
(after the last character:
store input) – Press and hold (> 2 sec ):
Store current input and display
the menu item
k During weighing: (Function)Toggle (depends on operating
menu settings) between: – first and second weight unit or – gross and net or – normal and 10-fold increased
display resolution In the operating menu:
Show the next item on the
same menu level (scroll
through values in series)
Combics_1Folie_u.eps
Combics 1
10
During alphanumeric input in the operating menu: – Cursor in first position,
no characters entered yet: Delete entire string and enter “0“
Change the displayed
character; scroll forward (sequence: 0 through 9, decimal point, minus sign, Z through A, pace)
p During weighing: (Print) Press briefly:
Print
Press and hold (> 2 sec ):
Print GMP footer In the operating menu: Print the menu settings starting from the current position; or print Info data During alphanumeric input in the operating menu: – Cursor in first position, no
characters entered yet:
Delete entire string and
enter a space
Change the displayed
character; scroll backwards ÿ (sequence: Space, A through Z, minus sign, decimal point, 9 through 0)
Note:
The sequence in which keys are pressed in the following descriptions (“Key sequence:“) are given as examples only. Other sequences are possible.
Configuring the Combics
Purpose
You can configure the Combics to meet individual requirements by entering user data and setting selected parameters in the operating menu. The operating menu is a combination of text levels and numeric codes.
Features
The operating menu parameters are divided into the following categories
1. Text-menu level
Application programs »APPL« Key assignments k »fn-key« Device parameters »setup« Device information »InFo« Language for calibration and
adjustment printouts »LAnG«
Combics 2
Combics_2Folie_u.eps
11
You can display, enter or change the
following parameters:
In addition to basic weighing functions, you can use the Combics 2 to calculate and
display weighing data as follows: – Counting – Neutral measurement – Averaging (animal weighing) – Checkweighing – Classification – Weighing in percent – Net-total formulation – Totalizing
Function: Key assignment for k
»fn-key«
Gross/net toggling
»gro net«
Toggle between the 1st, 2nd and
3nd weight unit
(are eycled in the display)
»3. unit«
10-fold higher resolution
»Res 10«
Device parameters »setup«
Device configuration, i.e.,
to meet individual requirements by selecting predefined menu parameters in the operating menu. The device parameters are combined in the following
groups : – Weighing platform »wp1« Interfaces (COM1 and optional
UniCOM) – Digital control line »ctrl in« – Bar code scanner »barcode«
Printer protocols »prtport«
Extra Functions: Utilities
»utilit«. – Timer »time« Date »date« Set user password »code«
Display device information
»info« (serial number, etc.)
Language for calibration and
adjustment records »lang«.
Select the language.
The rest of the menu
is language-independent.
12
(Function)
(O) (Tare)
(Print)
Scroll upwards (0, 1, 2, 3, etc.):
Press the k key
Scroll downwards ¬ (A, B, C, etc.):
Press the p key – Next position: Press the ) key – Previous position: Press the (
key – Exit input mode without saving
changes; with cursor in first
position: Press the ( key – Exit the input mode without
saving changes:
Press and hold the ( key
(> 2 sec) – To confirm your input: Press the
) key with the cursor 1 space
beyond the last position; or press
and hold (> 2 sec) – Confirm currently active
character and move 1 position
Functions of the Keys when Configuring the Menu
Open the operating menu:
Turn the Combics off and on
again by pressing e; while all segments are displayed, press the ) key briefly.
Navigating in the operating
menu:
(On) (Tare)
(Function)
Scroll :
Press the k key – To the left ¬: Press ( – Change and store setting:
Press the ) key – Exit the operating menu:
Press and hold the ) key (> 2
sec)
During alphanumeric input in
the operating menu:
to the right: Press the ) key
Move cursor 1 position to the left
Press the ( key
Printing the Parameter Settings
Generate a printout of the settings on
the current menu level:
Press the p key
> Printout (example)
The maximum width of this
printout is 20 characters.
--------------------
MENU SETUP WP1
-------------------­ 1
1.1
1.1.2
1.2.1 ...
1.18
1.18.1 CAL.
10.000 kg
etc.
13
Description of the Equipment
The Combics scales from Sartorius are primarily used in industry. Because there are a number of hardware configurations to choose from, Combics models are constructed according to customer requirements.
The Combics scale consists of:
– A choice of weighing platforms with capacities from 10 lbs to 10000 lbs (in stainless steel or steel) and indicators
The Combics 1 / 2 indicators consist of:
– Front panel with keypad and
digital display (Combics 2 has additional keys and LEDs)
– Combics indicators
are equipped with either cable glands or D-Sub 25 connectors.
Note: Additional electronic subassemblies are optional.
Digital PCB with application processor, memory, RS-232 interfaces
Analog/digital converter
Power PCB
DC/DC converter (Combics2 has a clock chip)
combics_15_klemmen.eps
Combics 2
14
Function of the Service Keys and Switches
A. Menu Access Switch
This switch must be open to work with the service software or configure the A/ D converter (e. g., “Trade” configuration). If the menu access switch is closed, the error message “ACCESS SWITCH LOCKED” is displayed in the service program and the indicator shows “L” (underload) or “H” (overload). In this case, the A/D converter cannot be configured. If adjustment is necessary (span, linearity), the “± 2%” window is active.
B. SBI/BPI Key
To use the Sartocas program (v 1.46 or later) or Psion server (v 4.10 or later). the Combics indicator must be in BPI (binary processor interface) mode (for example, to load a modified data record from the A/D converter into its controller). This setting is valid for the COM1 port.
Note: To perform adjustment
only with the service software, the COM1 interface is set to the xBPI (eXtended Binary Processor Interface) protocol and the laptop or Psion server connected to that port.
verriegelungs_schalter.eps
closed open
A
sbi_bpi_taster.jpg
B
15
Activating the BPI Mode
– Turn on the Combics. – In weighing mode, press and hold the SBI/BPI key for approx.
5 seconds (the weight readout goes blank). – Release the switch. – Turn the Combics off and then on again. – The Combics is now in BPI mode.
Important: Peripheral devices that communicate over the SBI (Sartorius Balance Interface)
protocol cannot be operated while the Combics is in BPI mode. If you press the p key, only InF 30 is displayed.
To return the indicator to the SBI mode, either use the “Close” function in the Sartocas service software or a Psion server (and then turn the Combics off and back on again), or select the Menu Reset option (9 - 1 - 1) in the Combics operating menu.
C. Boot Key
Activate this switch only if an attempt to program the application memory with the PPLoader program fails (programming routine stops responding).
Procedure for making the PCB bootable: – Disconnect the Combics from the main power. – Press and hold the boot switch (approx. 3 sec.) while reconnecting the Combics
to power.
– Release the switch. After a brief period, the PCB is bootable again
(load application software again with the PPLoader program).
boot_taster.jpg
C
16
Activating the xBPI Protocol
Activate the xBPI protocol to perform adjustment with the service software Sartocas software or Psion server without activating the BPI mode through the SBI/BPI key.
Note: This menu item is accessible in the standard menu.
–Via Setup select the Device parameters menu item (for the COM1 port).
– Open the Device parameters menu and scroll down to the Interface
submenu.
– Open the Interface submenu and activate the xBPI menu item.
Note: Under xBPI a submenu is opened for assigning addresses, but this function is only
available when using xBPI with RS-485.
Display_xbpi.eps
17
Activating the Service Mode
Activating the Service Mode
– Turn off the Combics (key: e). – Turn on the Combics (key: e) and briefly press the ) key while all
segments are displayed.
The display shows APPL.
– Press the k key to scroll to the Setup item.
– Press the ) key; WP-1 is displayed.
– Press the k key to scroll to the CodE item.
setup.eps
display.eps
appl.eps
display_wp1.eps
code.eps
18
– Press the ) key; the display shows only a flashing cursor.
Press k to select “2” and then press ) to confirm; the cursor jumps to the
second position; now you can enter the next number. Repeat this procedure with the appropriate numbers to enter the service password (202122).
Note: If you find that one of the numbers is wrong, press ( to go back to that digit
and change it; then press the ) key repeatedly to return to the sixth digit (the 2 flashes).
– After entering the last digit, press ) (< 2 sec.) and ( ( (in that order)
to activate the service mode. Setup appears in the display. A “S” in the upper right-hand corner of the display indicates that the service mode is active.
To cancel the password input process before it is finished, press ( repeatedly until CodE is displayed.
Note: The service mode can be deactivated only by turning the off the Combics indicator.
Important: When the service mode is active (password 202122), you can only view the
customer password. The customer password can be deleted only by entering the general password (40414243) or the customers password.
d_vore12.eps
202122.eps
code_s.eps
19
Working in the Service Mode
Note: When performing maintenance on an existing system, the service mode must be
activated to perform linearization, to set or clear the preload, and to enter the date of service.
Additional Menus in Service Mode
In the Text Menu (“Setup”) Under the “Date” (DATE) and “Password” (CodE) menu items:
– Date of service S-DATE – Memory number MEM-NO – Indicator serial number SER-NO – Model designation ModEL
In the numeric menu: The Setup menu for WP1 (WP-1) and the COM1 (COM1) and
UniCOM (UNICOM) ports is extended by the following items for configuring weighing platforms WP1 (WP-1) and WP2 (WP-2):
1-9 Calibration/adjustment functions:
1-9-1: Ext. calibration/adjustment with default weight
(service mode not required)
1-9-3: Ext. calibration/adjustment with user-defined weight
(service mode not required) 1-9-5: Internal linearization (only for WP2 on COM1 and UniCOM ports) 1-9-7: External linearization with user-defined weights 1-9-8: Set preload 1-9-9: Clear preload 1-9-10: Key blocked (service mode not required)
20
Note: After necessary adjustments, install the suitable
adjustment routine for the customer.
1-18 Enter the calibration and linearization weights
1-18-1: Enter adjustment weight 1-18-2: Enter linearization weight 1 1-18-3: Enter linearization weight 2 1-18-4: Enter linearization weight 3 1-18-5: Enter linearization weight 4
1-19 Calibration without weights (by entering the
specifications of the load cell(s)):
1-19-1: Nominal load 1-19-2: Resolution 1-19-3: Sensitivity in mV/V for cell 1 (or mean derived
from all load cells) 1-19-4: Sensitivity in mV/V for cell 2 1-19-5: Sensitivity in mV/V for cell 3 1-19-6: Sensitivity in mV/V for cell 4
When multiple cells are connected, either the
individual values are entered under items
1-19-3 to 1-19-6, or the mean derived from all
load cells is entered under 1-19-3. 1-19-7: Store values entered for 1-19
1-20 Place of adjustment (geographical latitude
and altitude or, alternatively, the acceleration of gravity at the place of installation):
1-20-1: Latitude in degrees 1-20-2: Altitude in meters above sea level 1-20-3: Acceleration of gravity 1-20-4: Store values entered for 1-20
8-12 Geographical data is displayed before every
span adjustment
9-1 Restore factory settings / Reset menu
9-1-3: Load “Standard” configuration 9-1-4: Load “Trade” configuration 9-1-10: Load default setting and delete
all Parameters
11 A/D Converter Settings:
11-1 Accuracy classes: 11-1-4: Accuracy class l 11-2 1. Weight unit (copy from menu subset 1-7): 11-2-1: long ton 11-2-2: Grams 11-2-3: Kilograms ... 11-2-21: Tons
21
11-3 Weighing range:
11-3-1: Single-range scale 11-3-2: Multiple-range scale 11-3-3: Multi-interval scale
11-4 Metrological data for single-range scale:
11-4-1: Scale interval d 11-4-2: Verification scale interval e 11-4-3: Minimum load 11-4-4: Maximum load
11-5 Metrological data for multi-interval scale:
11-5-1: Scale interval d 11-5-2: Verification scale interval e 11-5-3: Minimum load 11-5-4: Range 1 11-5-5: Range 2 11-5-6: Range 3 11-5-7: Maximum load
11-6 Metrological data for multiple-range scale:
11-6-1: Scale interval d 11-6-2: Verification scale interval e 11-6-3: Minimum load 11-6-4: Range 1 11-6-5: Range 2 11-6-6: Range 3 11-6-7: Maximum load
11-7 Available weight units:
11-7-1: User-defined unit 11-7-2: Grams 11-7-3: Kilograms ... 11-7-5: lb ... 11-7-21: Tons
11-8 Calibration/adjustment unit: 11-8-1: User-defined unit 11-8-2: Grams 11-8-3: Kilograms ... 11-8-5: lb ... 11-8-21: Tons
11-10 Store A/D converter configuration:
11-10-1: Store 11-10-2: Do not store
22
display_wps1_serv.eps
display_serv_01.eps
menu_1181.eps
user1Ib.eps
user2Ib.eps
menu_1181.eps
Menu Item 1-18: Entering Calibration/Linearization Weights
1-18-1: External User-defined Calibration Weight (Service Mode not Required)
Activate the service mode (see page 16).Select weighing platform “WP1”.
Press the ) key to open the numeric menu.
Select menu item 1-18-1 (external user-defined weight) by pressing k )
repeatedly.
Activate input mode for the user-defined weight (press ) briefly).
The first digit flashes.
Enter the user-defined weight (in this example: 10,000 lb) (key sequence:
) k ) p ) ) ) k )). The last number entered flashes.
Press the ) key again to store the user-defined weight; the menu jumps back
to 1-18-1.
To exit the service mode, press and hold the ) key for longer than 2 seconds.
23
1-18-2: 1. Entering Linearization Weight
– Scroll to menu item 1-18-2 (press k repeatedly).
Note: The service mode must be active.
– Activate the input mode for the first linearization weight (press the ) key
briefly). The first number flashes.
– 1. Enter the linearization weight (in this example: 2,500 lb) (key sequence:
3 x k, ) p ) 6 x k, ) k ) k). The last number entered flashes.
– Press and hold the ) key (> 2 sec.) to store the value for the first
linearization weight. The menu jumps back to item 1-18-2.
– Follow the above steps for menu items 1-18-3, 1-18-4, and 1-18-5 to enter
values for the second, third and fourth linearization weights as well. Enter “0” instead of a weight value for unused linearization points.
– At the end of input, the display must show menu item 1-18-5.
– To exit the service mode, press and hold the ) key for longer than 2 seconds.
menu_1182.eps
user1Ib.eps
userl2Ib.eps
menu_1182.eps
menu_1185.eps
24
Calibration/Adjustment Routines in Service Mode (General Information)
There are three ways to adjust the span: – Using external weights – By entering the specifications of the load cell(s) (sensitivity in mV/V) – By entering geographical data (latitude and altitude of the place of use) or by
entering the acceleration of gravity (at the place of use)
Span adjustment with external weights always takes precedence.
What does this mean when performing service work?
If a weighing system is used in a location other than that for which it was originally intended (i.e., the geographical data at the new place of use does not match the entered at Sartorius AG in Goettingen, Germany), the scale span can be adjusted by using external weights or by entering the specifications of the load cell(s). In this case, the geographical data entered for the original place of use is no longer valid. The same applies for span adjustment performed after any repair work on the weighing system. Once you store the geographical data or acceleration of gravity at the place of use and subsequently perform span adjustment using weights, the scale is correctly programmed and adjusted. You can now enter the geographical data of the new place of use to adapt the scale for use at that location.
25
If the acceleration of gravity has been entered, this data takes precedence over any geographical data (latitude and altitude) previously entered.
Valid Locations for Use
Once the exact geographical data or the acceleration of gravity has been entered and subsequent span adjustment performed, a scale with a resolution of 3000 digits can be used within a tolerance zone of ±100 km from the latitude entered and ±200 m from the altitude. The same applies when the acceleration of gravity is entered.
The following values are entered when the Combics scale is adjusted in Goettingen, Germany: – Latitude: 51.53 degrees (= 51° 32’) – Altitude: 151 m – or the corresponding acceleration of gravity, 9.811590 m/s-2
The greater the precision of the geographical data entered, the greater the precision achieved with the weighing instrument; the tolerance range, however, is restricted accordingly (see above).
26
Zone “D”
For a complete scale used in the Federal Republic of Germany, the following data can be entered for scales with a resolution of 3000 digits at the place of span adjustment:
– Latitude: 51.00 degrees – Altitude: 513 m – or the corresponding acceleration of gravity, 9.810 m/s
-2
The area of use valid for this scale is designated “Zone D” (valid only in Germany).
27
Menu Item 1-9: Calibration/Adjustment Functions
Note: When using a verified weighing platform, open the cover plate on the left side
of the back of the indicator and move the menu access switch to the right (“accessible”; see page 14).
– Activate the service mode (see page 17).
1-9-1: External Calibration/Adjustment with Default Weights
– Select weighing platform »WP1«. – Press the ) key to open the numeric menu.
Select menu item 1-9-1 (external calibration/adjustment with default weights)
(key sequence: ) several times k )) and confirm () key briefly). When this menu setting is active, you can exit the Service mode by turning the Combics indicator off and then on again.
display_wps1_serv.eps
display_serv_01.eps
menu_191.eps
28
display_cal.eps
5lb.eps
display_serv_01lb.eps
5.0lb.eps
display_wps1_serv.eps
display_serv_01.eps
– After taring or zeroing the scale, press and hold the ) key until Cal is
displayed. If menu item 8-12-2 is active, the geographical data is displayed before the calibration weight is prompted. Confirm each of the values shown by pressing the ) key to continue with the calibration routine.
– After approx. 2 seconds the calibration weight required is shown on the display.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last is displayed (calibration).
– Press the ( key to stop the calibration/adjustment routine.
– Press the ) key to start the adjustment.
1-9-3 External Calibration/Adjustment with User-defined Weights
– Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
29
– Press ), then, several times, k ) to select menu item 1-9-3, “External
calibration/adjustment with user-defined weights” and confirm by pressing )
briefly. – Press ( to jump to the first columns of the numeric menu. – Press ) to store the selected menu item. – Turn the Combics indicator off and then on again to exit the Service mode.
– After taring or zeroing the scale, press and hold the ) key until Cal is
displayed.
If menu item 8-12-2 is active, the geographical data is displayed before the
calibration weight is prompted. Confirm each of the values shown by pressing
the ) key to continue with the calibration routine. – After approx. 2 seconds the calibration weight (user-defined weight) required
is shown on the display. Select menu item “1-18-1” to enter the user-defined
weight.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last span adjustment is displayed
(calibration). – Press the ( key to stop the calibration/adjustment routine.
– Press the ) key to start the adjustment.
display_cal.eps
10Ib.eps
display_serv_01Ib.eps
10Ibw.eps
menu_193.eps
30
display_serv_01.eps
menu_197.eps
1-9-5 Internal Linearization (Only for WP1 on COM1 or UniCOM Ports)
1-9-7 External Linearization with User-defined Weights
– Activate the service mode (see page 17). – Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Select menu item 1-9-7 (External linearization with user-defined weight) (key
sequence: ), then, several times, k )) and confirm by pressing )
briefly. – Press and hold the ) key to store the selected menu items and return the
Combics indicator to the weighing mode.
When this menu setting is already active, you can exit the Service mode by
turning the Combics indicator off and then on again. – After taring or zeroing the scale, press and hold the ) key until Lin is
displayed.
Display_WP1_Serv.eps
display_lin.eps
31
– After approx. 2 seconds the first linearization weight (user-defined weight)
is shown on the display. Select menu items 1-18-2 through 1-18-5 to enter the
linearization weights.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration). – Press the ( key to stop the calibration/adjustment routine.
– Press the ) key to store the value for the first linearization weight;
the second linearization weight is prompted.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
– Press the ) key to store the value for the second linearization weight;
the third linearization weight is prompted.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
2.5Ib.eps
display_serv_02Ib.eps
display_5Ib.eps
display_serv_03lb.eps
7.5Ib.eps
display_serv_04lb.eps
32
– Press the ) key to store the value for the third linearization weight;
the fourth linearization weight is prompted.
– Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last adjustment is displayed (calibration).
– Press the ) key to store the value for the fourth linearization weight.
The scale now prompts a requires the zero point (remove all weights from the
load plate).
– Once the zero point is stored, the Combics automatically returns to the
weighing mode.
10Ib.eps
display_serv_06Ib.eps
0kg.eps
display_serv_03lb.eps
33
1-9-8 Setting the Preload
Note: When using a verified weighing platform, open the cover plate on the left side
of the back of the indicator and move the menu access switch to the right (“accessible”; see page 14).
– Activate the service mode (see page 17).
– Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Select menu item 1-9-8 (Set preload) (key sequence: ), then, several times,
k )) and confirm by pressing ) briefly. – Press and hold the ) key to store the selected menu items and return the
Combics indicator to the weighing mode. – Press )) to tare the Combics indicator, or press ( to zero it.
If you tare the indicator, “NET” might be displayed.
– Place the preload on the scale.
display_wp1_serv.eps
display_serv_01.eps
menu_198.eps
0lb_o.eps
vorl1.eps
34
– Press ) until SEt VOR is displayed.
– Release the ) key; after a brief pause, the Combics indicator automatically
returns to the weighing mode.
1-9-9 Clearing the Preload
Note: When using a verified weighing platform, open the cover plate on the left side
of the back of the indicator and move the menu access switch to the right (“accessible”; see page 14).
– Activate the service mode (see page 17). – Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Select menu item 1-9-9 (Clear preload) (key sequence: ), then, several times,
k )) and confirm by pressing ) briefly. – Press and hold the ) key to store the selected menu items and return the
Combics indicator to the weighing mode.
vorl2.eps
0lb_o.eps
display_serv_01.eps
display_wps1_serv.eps
vorl3.eps
35
– Remove the preload weight from the scale. A weight value with a preceding
“minus” sign is displayed (for example, -0.335 lb).
– Press ) until CLr VOR is displayed.
– Release the ) key; the preload is cleared and, after a brief pause, the Combics
indicator automatically returns to the weighing mode.
Display_-0335lb.eps
vorl5.eps
0lb_o.eps
36
Adjustment without Weights
Menu Item 1-19: Entering the Specifications of the Load Cell(s)
– Activate the service mode (see page 17). – Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Press the ) key, then, several times, k repeatedly and then ) again to
open menu item 1-19, “Calibration without weights”; then select item 1-19-1,
“Nominal capacity”.
– Press ) again to activate menu item 1-19-1. – Enter the nominal capacity of the load cell(s) (e.g., 30.00 lb).
The nominal capacity has to match the maximum load set under menu item
11-4-4.
Note: If a weighing platform has more than one load cell, multiply the nominal capacity
accordingly. Example: The weighing platform consists of 4 load cells, each of them has a capacity of 30 kg. In this case, the nominal capacity (= maximum load) equals 4 x 30 kg, or 120 kg.
display_wps1_serv.eps
display_serv_01.eps
display_serv_01191.eps
display_mxl30Ib.eps
37
– To store the setting, press and hold the ) key (> 2 sec).
– Select menu item1-19-2 (key sequence: k, )).
– Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
– To store the setting, press and hold the ) key (> 2 sec).
– Select menu item1-19-3 (key sequence: k, )).
Enter the sensitivity of the load cell in mV/V (usually listed in the load cell
specification sheets). If a weighing platform consists of more than one load cell,
enter here the sensitivity of load cell 1.
Example: The sensitivity of the load cell is 1.9440 mV/V.
– To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2) through 1-19-6 (for load cell 4) and confirm each value (k ) keys). If the platform has less than four load cells, enter “0.000” under the unused menu items.
display_schrittw_001Ib.eps
display_wze19440.eps
38
– Press k and ) to select menu item 1-19-7,
Store parameters set in menu item 1-19.
– Press ) ( ( to store the load cell specifications entered; the menu
automatically jumps back to the first menu page.
– Exit the operating menu and return to the weighing mode.
display_serv_01197.eps
display_serv_01.eps
39
Menu Item 1-20: Entering the Geographical Latitude and Altitude, or the Acceleration of Gravity
When you activate menu item 1-20-1, the value 99999.99 is displayed; under 1-20-2, is 9999999 displayed too. Otherwise, if values have already been entered for geographical latitude and altitude, then 0.000000 is displayed under menu item 1-20-3. During calibration, the Combics show will whether and which gravimetric data (geographical latitude and altitude, or acceleration of gravity) is used, if the menu item 8-12-2 is activated.
If latitude and altitude are used, the display shows “Altitude” for 2 seconds when the calibration procedure is started (CAL), followed by the configured elevation above sea level in meters. Press ) to confirm this information. Next, the display shows “LATITUDE” for 2 seconds, followed by the value set for the geographical latitude (in degrees). Press ) to confirm this information. The calibration weight is now prompted. If the acceleration of gravity is given rather than the geographical latitude and elevation, the display shows “GRAVITY” for 2 seconds (after “CAL”), followed by the value entered for the local acceleration of gravity. Press ) to confirm this information.
However, if the calibration weight is prompted as soon as the “CAL” display goes out, this means whether the correction of the calibration factor has been deactivated nor the menu item 8-12-2 is activated; in other words, neither the geographical latitude and altitude nor the acceleration of gravity have been entered under menu line 1-20.
40
Entering the Place of Adjustment with Subsequent Adjustment (Example)
– Activate the service mode (see page 17). – Press ) to select weighing platform »WP1«.
– Press the ) key to open the numeric menu. The first menu page is displayed.
– Open menu line 1-20, “Place of adjustment”, by pressing k and ).
Menu item 1-20-1 (latitude) is displayed.
If the geographical latitude and altitude of the place of installation are known, enter this data under menu items 1-20-1 (latitude) and 1-20-2 (altitude). If you wish to use the acceleration of gravity at the place of installation for the basis of adjustment rather than these two parameters, enter this value under menu item 1-20-3.
Important: A value entered for acceleration of gravity takes precedence over the
combination of geographical latitude and altitude.
display_serv_01201.eps
display_wp1_serv.eps
display_serv_01.eps
41
Example: Entering geographical latitude and elevation:
Select menu item 1-20-1 (latitude) (press )).
In the example shown here, the most recently entered parameter was the value for the acceleration of gravity (menu item 1-20-3). For this reason, the input fields for menu items 1-20-1 and 1-20-2 are filled with 999…
In the example shown here, the geographical latitude of Goettingen, Germany, was the value entered most recently. This value is the reference value, for which no correction of the span adjustment was made with respect to the place of installation, assuming that “151” (elevation in meters above sea level) has been entered under menu item 1-20-2.
Enter the geographical latitude of the place of installation (or the reference value 51.53) as a positive decimal number (convert angular minutes to decimal places). Press ) to confirm your input.
– Select menu item 1-20-2 (“Elevation at the place of installation”) (key
sequence: ) k )).
The input field is filled with 999…, because the last value entered in menu line
1-20 was the acceleration of gravity (menu item 1-20-3).
display_999.eps
display_5153.eps
display_999.eps
42
– Enter a value for the local elevation at the place of installation in meters above
sea level (reference value: 151) (key sequence: ) k k).
Note: A negative value may be entered for local elevation; for example,
if the scale is installed in a mine.
– Press ) to confirm your input. – Activate menu item 1-20-4 to store new data
(key sequence: ) k k).
Rather than the geographical latitude and local elevation of the place of installation, you can enter a value for the acceleration of gravity at the place of installation (menu item 1-20-3).
Example: Entering the acceleration of gravity place of installation.
– Select menu item 1-20-3 (acceleration of gravity) (key sequence: k k )).
Example: In the example shown here, the geographical latitude and local
altitude (menu items 1-20-1 and 1-20-2) were the last parameters entered in menu line 1-20. Thus the input field for acceleration of gravity is filled with 000…
In the example shown here, the most recently entered parameter
was the value for the acceleration of gravity at Goettingen, Germany (menu item 1-20-3). This value is the reference value, which means no correction of the span adjustment for the place of installation is required.
display_000.eps
display_981.eps
display_151.eps
43
– Enter the acceleration of gravity at the place of installation (in ms-2 / or the
reference value: 9.811590). Press ) to confirm your input.
– Activate menu item 1-20-4 to store new data (key sequence: ) k).
The display shows “busy”, indicating that the parameters entered are being stored.
– Restart the scale: Turn the indicator off and then on again (key sequence: e
e, or press and hold the ) key (> 2 sec).
After the self-test of the display, the scale is ready to operate.
Note: Once the precise geographical data or acceleration of gravity has been entered, the
tolerance range is restricted (see page 23).
After data has been entered as described above, the span must
be adjusted.
Example Settings (changes in the factory settings required for this example): Menu item 1-9-1: External calibration/adjustment with default weight; Menu item 1-10-1: Calibration/adjustment in one operation; Menu item 1-16-1: External calibration/adjustment/linearization
accessible.
display_busy.eps
display.eps
44
The parameters for altitude at the place of installation and geographical latitude have been entered (menu items 1-20-1 and 1-20-2). The display will show “ALTITUD” for 2 seconds, if the menu item 8-12-2 is activated.
The altitude at the place of installation is displayed in meters above sea level. In the example shown here, the altitude of Goettingen, Germany (reference value) is shown.
– Press the ) key to confirm the displayed value.
The display shows “LATITUD” for 2 seconds.
The geographical latitude of the place of installation is shown in degrees north or degrees south. In the example shown here, the altitude of Goettingen, Germany (reference value) is shown.
– Press the ) key to confirm the displayed value.
– Place the prompted calibration weight on the scale (see page 25 for details on
calibration and adjustment).
or
display_latitud.eps
display_lat_5153.eps
10Ib.eps
display_altitud.eps
display_alt_151.eps
45
Instead of “ALTITUD” and “LATITUD”, “GRAVITY” is displayed for 2 seconds if the acceleration of gravity (menu item 1-20-3) was entered rather than the elevation and geographical latitude (menu items 1-20-1 and 1-20-2). The value entered for the acceleration of gravity at the place of installation is displayed (in m/s-2).
In the example shown here, the acceleration of gravity at Goettingen, Germany (reference value) is shown.
This data is shown when menu item 8-12-2 is active.
– Press the ) key to confirm the displayed value.
– Place the prompted calibration weight on the scale (see page 25 for details on
calibration and adjustment).
display_gravity.eps
display_grav_981.eps
10Ib.eps
46
Configuring the A/D Converter in the Combics Indicator (Service Mode)
The A/D converter must be configured in the following cases:
– To adapt the existing configuration to customer requirements (for example,
to change the nominal capacity or the resolution). – To adapt the Combics for use with any commercially available strain-gauge
load cell or analog Sartorius CAPP or CAPS weighing platform, when installing
a new system. – After replacing the electronic A/D converter during repair work, also on
combics scales, when the Sartocas software is not used.
Menu Item 11: A/D Converter Settings (Configuration: Standard, Trade)
Note: Before the A/D converter can be configured, you need to load either the
“Standard” configuration or the “Trade” configuration, depending on customer requirements.
– 9-1-3: Load “Standard” configuration – 9-1-4: Load “Trade” configuration
47
Menu Item 11: A/D Converter Configuration (General Information)
Note: Access is restricted by a special password.
Features With the menu access switch open, you can
configure most of the parameters affecting the following specifications:
– Toggling between Standard and Trade configurations
(for use in legal metrology)
– Verification scale interval e – Scale interval d – Minimum load – Maximum load – Maximum load for a given range – Verification scale interval e for a given range
– User-definable weight units
These parameters are not reset when you restore the factory defaults in the operating menu. Parameters not listed above are not affected by your choice of Standard or Trade configuration; the same restrictions apply as for Sartorius weighing instruments which cannot be configured.
Note on A/D Converter Configuration
A/D converter configuration is performed using the numeric menu (weighing platform 1).
Important: The A/D converter can be configured only in
the service mode.
The maximum load for each weighing range must be entered in an appropriate weight unit. Select the desired weight unit for operation under menu line 11-7, “Weight units”. The maximum load for each range must be a value that can be displayed in any available weight unit. Weight units that are accessible (or blocked) for use as Weight unit 1 are also available (or blocked) for use as Weight unit 2. The weight unit used for the configuration cannot be blocked.
Checking and Configuring the Equipment for Use in Legal Metrology:
48
Following A/D converter configuration, write the metrological data on a tag for all weighing ranges. Afterwards, affix the enclosed protective acetate overlay to the tag.
After completing calibration and adjustment, move the menu access switch back to the “closed” position and then turn the Combics indicator off and back on again to activate the new configuration.
Open menu line 1-7 to make sure only the permitted weight units are accessible.
Perform necessary routines for scales used in legal metrology.
Descriptions of the Individual Menu Items
Menu Item 9-1-3 / 9-1-4: Selecting and loading configuration data
Before selecting the menu item for A/D converter configuration, check whether the weighing platform is used in the standard operating mode (Standard configuration) or in legal metrology (Trade configuration):
– Standard Configuration
(menu item 9-1-3) or
– Trade Configuration
(menu item 9-1-4)
Menu Line 11-1 (Accuracy Class) This menu line is not shown when the Standard configuration is active.
When the Trade configuration is active (for verifiable or verified weighing platforms), only menu item 11-1-4 is shown. Select the accuracy class appropriate for the weighing platform connected. Which class is appropriate depends
on the load cell(s) installed.
In legal metrology, only classes l (menu item 11-1-4) can be selected. Press k to select the accuracy classes and ) to confirm. A circle (o) after the menu code indicates the active setting.
49
Selecting the Weighing Range Structure (Menu Item 11-3)
The capacity of the weighing platform can be divided into multiple ranges. These settings are configured under menu item 11-3:
Single-range Scale (11-3-1) The entire weighing range is divided into scale intervals on the basis of the lowest interval d and the maximum load. In this case, the readability over the entire weighing range is always the lowest scale interval d.
Multiple-range Scale (11-3-3) A multiple-range scale has two weighing ranges. When the maximum capacity of the lower range is exceeded, the scale switches to the higher range (lower resolution) and remains in that range. The scale can be returned to the lower weighing range (higher resolution) only by unloading the scale and then pressing the ( key.
Scale Interval d
The lowest scale interval d indicates the resolution of the weighing instrument. The scale interval d can be entered only in increments of 1, 2, 5, 10, 20, 50, etc. When working on a verifiable or verified scale in accuracy class l, you are not prompted to enter the lowest scale interval, as this is equal to the verification scale interval e.
Verification Scale Interval e
The verification scale interval e indicates the resolution of the weighing instrument in legal metrology. When accuracy class l is active, this is equal to the scale interval d. Thus with classes l, the lowest scale interval d is not prompted.
50
Range 1, Range 2, Range 3
Here you can enter the limits for each of the weighing ranges. When a limit is exceeded, the accuracy changes. The following applies when entering range limits: Range 1 < range 2 < range 3 < maximum capacity. Thus the weighing capacity can be divided into 4 ranges. The resolution changes in intervals of 1, 2, 5, 10, 20, 50, etc. The lowest resolution is equal to the specified lowest scale interval d. Set unused ranges to 0.
Maximum Capacity
The maximum capacity is the maximum load that may be placed on the weighing instrument. If a heavier load is placed on the platform, the display shows H. The scale intervals are derived from the maximum load and the lowest scale interval d. In legal metrology, the number of intervals must not exceed 6000 e, or 3000 e per range in a multi-interval scale. In standard operation, as opposed to legal metrology, you can define a “SuperRange” weighing instrument with more intervals. For example, a maximum load of 60,000 lb with a lowest scale interval d of 0.001 lb 60,000 scale intervals. These parameters, however, may be influenced by physical restrictions.
Available Weight Units (Menu Item 11-7)
With this function, you can make particular weight units inaccessible during weighing. Available units are indicated by a circle (o) on the display (more than one can be selected).
Calibration/Adjustment Unit (Menu Item 11-8)
This setting defines the weight unit with which calibration must be performed. The calibration unit remains the same, even when a different weight unit is used during normal weighing operation.
Calibration without Weights (Menu Item 1-19)
The values entered for nominal capacity (in lb; menu item 1-19-1; specification of the strain-gauge system), resolution in kg; (menu item 1-19-2) and sensitivity in mV/ V menu item 1-19-3) are converted to internal quantities. Once the A/D converter configuration data has been stored, these parameters can no longer be read. For weighing platforms with multiple load cells, enter the sensitivities of the other load cells under menu items 1­19-4 (cell 2) through 1-19-6 (cell 4). Select menu item 1-19-7 to store the data entered.
51
Important: Calibration without weights cannot be
performed on weighing instruments used in legal metrology.
Storing Configuration Data (Menu Item 11-10)
Select menu item 11-10-1 to store the A/D converter configuration data.
Once these parameters have been configured, the A/D converter in conjunction with the load cell is defined as a weighing instrument. The A/D converter, in conjunction with the weighing platform, can now be used like any standard weighing platform.
In addition, the weight unit must be defined and the weighing platform adjusted (calibration/adjustment and linearization must be performed). For a detailed description of these procedures, see the chapter entitled “Calibration/Adjustment in Service Mode”.
Important: After configuring the A/D converter
and adjusting the weighing instrument (calibration/adjustment and linearization), return the menu access switch to the “closed” position (i.e. for use in legal metrology). Turn the Combics indicator off and then on again (key sequence: e e to activate the new configuration.
Once the A/D converter configuration has been locked (menu access switch closed), the indicator can no longer be used to influence weighing results. The scope of functions available in the weighing instrument is defined by the A/D converter. Scale functions which can be activated are, for example: weight readout, taring, adjustment, reading tare value, saving/deleting tare values.
52
Menu Item 9-1: A/D Converter Configuration (Example: Trade Configuration)
– Activate the service mode (see page 17). – Press k repeatedly to select weighing platform »WP1«.
– Press the ) key to open the numeric menu. The first menu page is displayed.
– Open menu item 9-1 (Factory setting/Menu reset)
(key sequence: k k )).
Loading the Trade Configuration
– Select menu item 9-1-4 (Trade configuration) and confirm
(key sequence: ) k k)).
The selected configuration is loaded. »Busy« is displayed briefly.
Afterwards, the program returns to the display of menu item 9-1-2
(Restore factory defaults: »Off«; factory setting).
display_wp1_serv.eps
display_serv_01.eps
display_serv_91.eps
display_serv_914.eps
display_busy.eps
display_serv_912.eps
53
– Select menu page 11 (key sequence: ( ( k).
Note: If the Trade configuration is loaded, menu line 11-1 (for selecting
accuracy class) is displayed at this point.
Otherwise (Standard configuration), that menu line is skipped and 11-2
(weight unit 1) is displayed.
In this case, press k to open menu line 11-3.
– Press ) to confirm your selection.
– To use the weighing instrument in legal metrology (i.e., with the Trade
configuration loaded) press ) to enter the accuracy class:
– Select menu item 11-1-4 for accuracy class l.
– Press ) to confirm.
The configured menu item (in the picture here,
11-1-4 for accuracy class l) is marked by a circle (o).
display_serv_11.eps
display_serv_111.eps
display_serv_1114.eps
display_serv_1114w.eps
54
– Press the key sequence ( k k to open menu item 11-3 (Selecting the
weighing range structure).
– Press ) to confirm your selection.
In the example shown here, menu item 11-3-1 (Single-range scale) has been
selected.
– Press ( to return to menu item 11-3.
Note: Enter the scale interval, minimum load, range limits (multiple-range or multi-
interval scales only) and maximum capacity: The menu line corresponding to the selected weighing range structure is displayed; i.e.
– for single-range scale (menu item 11-3-1): in relation to menu item 11-4; – for multiple-range scale (menu item 11-3-2): in relation to menu item 11-5; – for multi-interval scale (menu item 11-3-3): in relation to menu item 11-6.
In each case, the menu lines for the other two weighing range structures are not shown.
In this example, 11-3-1 (Single-range scale) has been selected. Thus menu line 11-4 is opened automatically. Menu lines 11-5 and 11-6 are not available.
display_serv_113.eps
display_serv_1131w.eps
55
– The input field for the verification scale interval e is to open.
– If the Trade configuration is active, select menu item 11-4-2 (Verification scale
interval e) (key sequence: ( k k k )).
– Press the ) key again to open this item; 0.001 kg is displayed.
– Enter the value for e (for example, e = 0.1 lb) by pressing the key sequence )
) k ) ) p ) and store (press and hold ) > 2 sec).
When menu item 11-4-2 is opened again, the display shows 0.1 lb.
– Select menu item 11-4-3 (Minimum load) (key sequence: ( k).
– Press the ) key again to open this item; then enter the minimum load (key
sequence: k k k ) p ) k).
Example for accuracy class K: Minimum load = 20 x e = 2.0 lb – To store the new setting, press and hold the ) key (> 2 sec).
display_serv_114.eps
display_serv_1142.eps
eiche1lb.eps
eiche2lb.eps
display_schrittw_01Ib.eps
display_serv_1143.eps
display_mnl20Ib.eps
56
– Press k to select menu item 11-4-4 (maximum load).
– Press ) again to open this menu item; then enter the maximum capacity (for
example 300.0 kg) (key sequence: k k k k ) k ) k )
p ) k). Enter “300” (it is not necessary to enter the decimal point and
the last “0”). – To store the new setting, press and hold the ) key (> 2 sec).
Note: If “Multiple-range scale” or “Multi-interval scale” was selected during A/D
converter configuration rather than “Single-range scale”, then the range limits must be entered under menu line 11-5 (multiple-range) or 11-6 (multi-interval).
Each of these menu lines has 3 items to choose from, for setting range limits 1, 2 and 3.
– Store the configuration data: Select menu line 11-10 (key sequence: ( (
k k k )).
The default setting is 11-10-2, “Do not store configuration”.
– Press k to select menu item 11-10-1 (Store configuration).
– To store the new settings, press and hold the ) key (> 2 sec).
Display_Serv_1144.eps
Display_Mxl30.eps
Display_Serv_11102.eps
Display_Serv_11101.eps
57
Busy” is displayed briefly.
Afterwards, the Combics is re-started. The scale is in the normal weighing
mode.
Then adjust the weighing instrument and, if the Trade configuration is active, return the menu access switch from the “accessible” to the “closed” position (see page 12).
Menu Item 9-1: A/D Converter Configuration with Load Cell(s) Connected
(Example: Standard Configuration
– Activate the service mode (see page 17). – Press ) repeatedly to select weighing platform »WP1«.
– Press the ) key to open the numeric menu. The first menu page is displayed.
– Open menu line 9-1 (Factory settings/Reset menu).
Display_busy.eps
Display.eps
Display_WP1_Serv.eps
Display_Serv_01.eps
58
Loading the Standard Configuration (if the Trade configuration is currently loaded)
– Open menu item 9-1 (Standard configuration) and confirm (key sequence: k
k )).
– Select menu code 9-1-3 (key sequence: ) k) and press ) to store.
The selected configuration is loaded. “Busy” is displayed briefly.
Afterwards, the program returns to the display of menu item 9-1-2 (Restore
defaults: »Off«; factory setting).
– Open menu page 11 (key sequence: ( ( k).
Configuration: Enter the desired weighing range structure (single-range scale,
multiple-range scale, multi-interval scale), scale interval and maximum capacity
(with range limits, if necessary). – Press ) k to select the weighing range structure (menu line 11-3).
Display_Serv_912.eps
Display_Serv_11.eps
Display_Serv_113.eps
display_serv_91.eps
Display_Serv_913.eps
Display_busy.eps
59
– Press ) to confirm your selection.
In the example shown here, menu item 11-3-1 (Single-range scale) has been
selected.
If desired, press k to change the weighing range structure:
11-3-2: Multiple-range scale; 11-2-3: Multi-interval scale. – Press ( to return to menu line 11-3.
Note: Entering the scale interval, range limits (multiple-range or multi-interval scales
only) and maximum capacity: The menu line corresponding to the selected weighing range structure is displayed, i.e., – for single-range scale (menu item 11-3-1): in relation to menu item 11-4; – for multiple-range scale (menu item 11-3-2): in relation to menu item 11-5; – for multi-interval scale (menu item 11-3-3): in relation to menu item 11-6.
In each case, the menu lines for the other two weighing range structures are not shown. – When you activate menu code 11-3-1 (Single-range scale) (by pressing the key),
menu line 11-4 is opened automatically. In this case, menu lines 11-5 and 11-6
are not shown.
– Open menu item 11-4-1 to enter the scale interval d.
– Press ) to select the menu item.
Display_Serv_1131W.eps
Display_Serv_1141.eps
Display_Serv_114.eps
60
The scale interval d is shown, and a flashing digit indicates the cursor position.
You can change the digit that the cursor is on.
Move the cursor as needed to change other digits; in this manner, you can
change the number to the desired value.
– Defining the scale interval d for the weighing instrument:
If the currently active value (displayed; in this example, 0.001 lb) matches the
desired value, press and hold ) (> 2 sec) to confirm it.
– Otherwise, overwrite the value as desired (in this example, by entering 0.01 lb;
key sequence: k ) p ) k ) k k); then press and hold the
) (> 2 sec) to confirm.
After you store the data, the program jumps back to menu item 11-4-1.
– Press k to select menu line 11-4-4 (maximum load).
– Press ) to open the menu item.
Display_Serv_1141.eps
Display_Serv_1144.eps
display_schrittw_001Ib.eps
display_schrittw_0001Ib.eps
standard1Ib.eps
61
– Enter the desired maximum capacity (in this example, 30.0 kg; key sequence:
k k k k ) k ) p ) k) and then press and hold )
(> 2 sec) to store the value. Enter “30” (it is not necessary to enter the decimal
point and the last “0”).
After you store the data, the program jumps back to menu item 11-4-4.
Note: If “Multiple-range scale” or “Multi-interval scale” was selected during A/D
converter configuration rather than “Single-range scale”, then the range limits must be entered under menu line 11-5 (multiple-range) or 11-6 (multi-interval). Each of these menu lines has 3 items to choose (range limits 1, 2 and 3).
– Store the configuration data: Select menu line 11-10 (key sequence: ( k
k k
– Press ) to open menu item 11-10-2 (“Do not store configuration”)
(active setting).
– Press k to select menu item 11-10-1 (Store configuration).
Display_Serv_1144.eps
display_serv_1110.eps
Display_Serv_11102.eps
Display_Serv_11101.eps
standard1Ib.eps
62
Press ) to store the A/D converter configuration data. “Busy” is displayed briefly; afterwards, the Combics is re-started. After the self-test of the display, the scale is ready to operate.
Menu Item 1-19: Entering the Specifications of the Load Cell(s)
– Activate the service mode (see page 17). – Select weighing platform »WP1«.
– Press the ) key to open the numeric menu.
– Press the ) key, then seveal times, k ) to open menu line 1-19,
“Calibration without weights”; then select item 1-19-1, “Nominal capacity”.
– Press ) again to activate menu item 1-19-1.
– Enter the nominal capacity of the load cell(s) (e.g., 30.00 kg).
The nominal capacity has to match the maximum load set under menu item 11-4-4.
Display_Serv_01191.eps
display_mxl30Ib.eps
Display_busy.eps
Display_WP1_Serv.eps
Display_Serv_01.eps
63
Note: If a weighing platform has more than one load cell, multiply the nominal capacity
accordingly. Example: The weighing platform consists of 4 load cells, each of which has a capacity of 30 kg. In this case, the nominal capacity (= maximum load) equals 4 x 30 kg, or 120 kg.
– To store the setting, press and hold the ) key (> 2 sec). – Select menu item1-19-2 (key sequence: k, )).
– Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
– To store the setting, press and hold the ) key (> 2 sec). – Open menu item 1-19-3 (key sequence: k )).
Enter the sensitivity of the load cell in mV/V (usually listed in the load cell
specification sheets). If a weighing platform consists of more than one load
cell, enter the sensitivity of load cell 1.
Example: The sensitivity of the load cell is 1.9440 mV/V.
– To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2) through 1-19-6 (for load cell 4) and confirm each value (k and ) keys). If the platform has less than four load cells, enter “0.000” under the unused menu items.
display_schrittw_001Ib.eps
Display_WZE19440.eps
64
– Press ) and then ) to select menu item 1-19-7,
(Store values for menu item 1-19).
– Press ) ( ( to store the load cell specifications entered; the menu
automatically jumps back to the first menu page.
– Press e twice; the indicator is now in weighing mode. – Close the menu access switch (see page 14). – Turn the Combics indicator off and then on again to activate the new
configuration.
Entering the Date of Service
After performing maintenance or repair, you can enter the date of the next scheduled maintenance visit.
– Activate the service mode (see page 17).
– Press k repeatedly to scroll through the menu until the display
shows S-dAtE.
s-date.eps
Display_Serv_01197.eps
Display_Serv_01.eps
code_s.eps
65
– Press ) activate the input mode (the cursor flashes in the first position).
– Press the ) k p keys as needed to enter the next scheduled
maintenance date (for example, 25 Apr 03).
– Press and hold the ) key (with the cursor to the right of the last character)
to store the date; the display returns automatically to S-dAtE.
– Press e twice to return the Combics indicator to the weighing mode.
Entering the Serial Number of the Combics Indicator
After replacing the digital PCB, the serial number of the Combics scale or Combics indicator must be written in memory.
– Activate the service mode (see page 17).
cursor.eps
s-date.eps
0lb_o.eps
s-date1.eps
code_s.eps
66
– Press k k k to scroll through the menu until the display shows Ser-
no.
– Press ) activate the input mode.
– Press the ) k p keys as needed to enter the serial number ( for
example, 13800012) of the existing weighing system.
Note: After you enter the last digit, the first digit is no longer displayed.
– Press and hold the ) key to store the serial number; the display returns
automatically to Ser-no.
– Press e twice to return the Combics indicator to the weighing mode.
0lb_o.eps
cursor.eps
ser-no1.eps
ser-no.eps
ser-no.eps
67
Entering the Model Designation of the Combics Indicator
After replacing the digital PCB, the model designation of the scale or Combics indicator must be written in memory.
– Activate the service mode (see page 17).
– Press k k k k to scroll through the menu until the display shows
ModEL.
– Press ) activate the input mode (the cursor flashes in the first position).
– Press the ) k keys as needed to enter the model designation (for
example, CISL_1U of the weighing system.
– Press and hold the ) key (with the cursor to the right of the last character)
to store the model designation; the display returns automatically to ModEL.
– Press e twice to return the Combics indicator to the weighing mode.
cursor.eps
Display_CISLIU_.eps
model.eps
model.eps
code_s.eps
0lb_o.eps
68
Defining Transaction Numbers for Data Records (Weight Values) which were stored in the Alibi Memory
– Activate the service mode (see page 17).
– Press k k to scroll through the menu until the display shows MEM-no.
– Press ) activate the input mode (the cursor flashes in the first position).
– Press the ) k keys as needed to enter the allocation number in the Alibi
memory.
– Press and hold the ) key (with the cursor to the right of the last character)
to store the allocation number; the display returns automatically to MEM-no.
– Press e twice to return the Combics indicator to the weighing mode.
code_s.eps
memory.eps
cursor.eps
memory1.eps
memory.eps
0lb_o.eps
69
Repairing the Combics Indicator
Important: An isolating transformer must be installed between the indicator and mains
before performing work that entails opening the Combics indicator housing.
To open the Combics indicator, remove the four nuts as shown in the illustration on the left.
Note: After completing maintenance or repair work, check the seal between the
front panel and the housing body for damage and replace if necessary.
If the Combics indicator in question has an IP67 protection rating,
a special test procedure is used to check the IP67 protection after the housing has been closed.
Replacing the Front Panel
In the case of a defective keypad overlay, keys, LEDs, or display, the entire front panel must be replaced.
combics_10.eps
70
Blank Display
If the display is blank (dark), disconnect the equipment from mains, open the housing and disconnect all cables and wires from the subassemblies. Then connect the equipment to power again through an isolating transformer and connect all subassemblies again, measuring the supply voltage in each case.
– The voltage at the power supply output (connector A; see page 71) is
15V ± 0.3 V direct current. – The output voltages of the DC/DC converter are listed on page 72.
Replacing the Power Cable
– Disconnect the cable from the power source.
– After replacing the power cable, use a torque wrench to tighten the cable gland
to 3 Nm.
pg_netz.jpg
71
Replacing the Power PCB
– Disconnect the cable from the power source.
– Pry the protective cap from the power PCB.
To do this, insert the end of a slotted-head screwdriver into each of the
four openings on the protective cap and carefully push the retainer clips,
located further down, to one side.
– Remove the protective cap and disconnect the two wires (blue/brown).
– Unplug connector A, remove the 2 screws and replace the power supply.
After replaceing the PCB, make sure to return the protective cap to its
original position.
netzt_kappe.jpg
netzteil.jpg
A
72
Replacing the DC/DC Converter
Note: Use a digital voltmeter (DC setting) to measure voltages against ground (earth).
No particular points of measurement are defined for this procedure; for this reason, use a thin test pin and exercise extreme caution to avoid short-circuiting the equipment.
The various supply voltages delivered by the DC/DC converter can be measured at the following points. M 1: 13V ± 1.3V Supply voltage for a connected printer. M 2: 15V ± 0.3V Input voltage for the DC/DC converter. M 3: –29V ± 1.5V Supply voltage for the front panel electronics. M 4: –8.5V ± 0.2V Supply voltage for the A/D converter M 5: 8.5V ± 0.2V Supply voltage for the A/D converter M 6: 5V ± 0.13V Supply voltage for the digital PCB
M 1
M 3 M 4M 2 M 5 M 6
dc_dc.gif
73
Replacing the A/D Converter
– Before replacing the A/D converter, try to read out the data from the A/D
converter’s controller using the service software; if this is successful, you can
load the adjustment data in the controller of the new A/D converter once it is
installed. In many cases, this precludes the need to adjust the weighing system.
– Because the model designation and serial number are not stored in the A/D
converter’s controller, you do not need to enter this data again after replacing
the A/D converter. – When connecting the weighing platform (load cell(s)) to the A/D converter,
it is important to install a ferrite ring over the signal lines leading from the
platform, to ensure electromagnetic compatibility.
Quick-test of the A/D Converter
Pin Assignment Chart
No. Signal designation Meaning Voltage
1 BR_POS Bridge supply voltage (+) 4.5V ± 0.18V
2 SENSE_POS Sense (+)
Bridge supply voltage 3 OUT_POS Measuring voltage positive 4 OUT_NEG Measuring voltage negative 5 SENSE_NEG Sense (-)
Bridge supply voltage 6 BR_NEG Bridge supply voltage (-) –4.5V ± 0.23V
adu_001.jpg
Combics_18an.eps
74
1. Disconnect the platform measuring voltage lines from the terminal strip (pins 3 and 4) in the A/D converter. Then short circuit these two lines and connect them to pin 7 (GND) of the terminal connector. If the A/D converter is intact, the weight readout should be stable. Disconnect the cable from the platform.
2. Connect the strain-gauge simulator and turn the switch to simulate different input signals. The weight readout should be stable.
Replacing the Digital PCB
Note: If communcation problems occur during communication with connected devices
(such as a printer), check the cable connections and the data transfer parameters first; if the error source is not found, start the internal test program.
– Turn off the indicator. – Press and hold the k key and turn the indicator on again. The test program
starts.
In case of defect, replace the digital PCB.
Some Combics indicators are equipped with D-Sub 25 female connectors, and some with cable glands; make sure you order the correct PCB for your indicator (see the illustration on the left).
– After replacing the digital PCB, use the PPLoader program to load the
application software, if necessary.
– Afterwards, enter the model designation and the Serial number of the Combics
scale (see pages 65 and 67).
haupt_pg.jpg
or
haupt_25.jpg
75
Checking the IP67 Protection
The test procedure and the equipment used have yet to be specified.
Torque Values for Cable Glands and Vent Valve
Cable gland: power cord 3 Nm Cable gland: optional connections 3 Nm (brass: 5 Nm) Cable gland (brass): platform connection 5 Nm Vent valve 1.5 Nm
76
Adjusting the Weighing Platform
Adjusting the Off-center Load
If the weighing platform has only one load cell, the off-center load is not adjusted in case of error; rather, the load cell must be replaced. The off-center load is adjusted only in weighing platforms that have four load cells.
Procedure (Example Illustrated on the Left)
– Place the test weight (1000 lb) on the weighing platform and write down the
weight values measured.
– The load cell with the lowest weight value (9972 lb) is the reference load cell.
This load cell does not require adjustment.
– Calculate the difference between the weight value registered by this load cell
and those registered by the other load cells.
– Plot the point of intersection of the calculated off-center load difference
(x-axis) and the test weight (see page 77).
– The adjustment resistance (total) and the best resistor combination can be read
from the y-axis (see the diagram on page 77).
In the example given here, load cell 3 has an off-center loading error of 8 lb,
which yields an adjustment resistance of 1.5 ohm + 0.82 ohm.
– Depending on the resistor combinations of each individual load cell, the
required resistors are activated by disconnecting the corresponding “0-ohm resistor” (see page 78).
Note: The resistance values detected are based on 350-ohm load cells.
For 700-ohm load cells, double the resistance values.
Eckl_2_usa.eps
Load cell 1
1260 lb
Load cell 2
1255 lb
Load cell 3
1254 lb
reference Load cell 4
1246 lb
(Example: reference Load cell +80 lb)
Test weight
1250 lb
77
Diagram: Determining the Adjustment Resistance
Note: If a 125 lb test weight is used (corresponding to the maximum capacity of the weighing platform) rather than a 12.5 klb test weight, and the off-center load error is 0.8 lb, for example, then the same adjustment resistance is set as when a 12.5 klb test weight is used.
Diagram_usa.eps
Test load in klb
0,2123 45
2102030 4050
20 100 200 300 400 500
50lb 100lb 125lb 150lb 200lb 250lb 300lb
0,5klb 1klb 1,25klb 1,5klb 2klb 2,5klb 3klb
5klb 10klb 12,5klb 15klb 20klb 25klb 30klb
Corner load error in lb
Corner load balancing for 350 ohm load cells
(For 700-ohm load cells the computedvalue must be doubed)
Example: Test load: 12,5 klb Corner load error: 80 lb
0,39+0,82+1,5+3
+1,5+3
1,5+3
0,39+0,82+3
0,82+3
0,39+3
3
0,39+0,82+1,5
0,82+1,5
0,39+1,5
1,5
0,39+0,82
0,82
0,39
.
0,39
Ω Ω Ω
Ω Ω Ω
Ω Ω Ω
.
.
.
.
.
.
.
.
.
.
.
.
Balancing resistence in ohm
Resistance
Individual
cholce
78
Off-center Load Adjustment: 0-ohm Resistors (Example)
In the example given on page 75, the two 0-ohm resistors (marked by a white dot) must be removed. This activates the adjustment resistors (0.82 and 1.5 ohms). Follow the same procedure to determine the resistance values for load cells 1 and 2, and remove the 0-ohm resistors if necessary.
The allocation of load cells in the weighing platform to contact on the adjustment PCB can be determined by checking the wiring in the weighing platform.
Note: On stainless steel models, the IP67 protection must be checked after closing the
junction box.
eckl_4.jpg
0-ohm-resistors and the corresponding ad­justment resistors of load cell no 4
0-ohm-resistors and the corresponding ad­justment resistors of load cell no 3
0-ohm-resistors and the corresponding ad­justment resistors of load cell no 2
0-ohm-resistors and the corresponding ad­justment resistors of load cell no 1
79
Setting the Overload Stops
1. Weighing Platforms with One Load Cell: Steel
Loosen the overload stops until the weight value displayed for the load on the platform begins to decrease (stops touch the load plate support). Then retract (tighten) the stops by 4 turn.
Platform nominal Test weight for Test weight for capacity in Ib. middle stops in Ib. corner stops in Ib.
10 12 12 25 30 30
50 60 60 100 120 120 200 250 250 250 300 300 500 600 600
1000 1200 1200
80
2. Weighing Platforms with One Load Cell: Stainless Steel
Extend (loosen) the overload stops until the weight value displayed for the load on the platform begins to decrease (i.e., until the stops touch the load plate support).
Platform nominal Test weight for Test weight for capacity in Ib. middle stops in Ib. corner stops in Ib.
10 12 12
25 30 30
50 60 60 100 120 120 200 250 250
81
3. Weighing Platforms with Four Load Cells: Steel
Extend (loosen) the overload stops until the weight value displayed for the load on the platform begins to decrease (i.e., until the stops touch the load plate support).
Platform nominal Test weight for Test weight for capacity in Ib. middle stops in Ib. corner stops in Ib.
1000 n/a 600 2500 n/a 1200 5000 n/a 2500
10000 n/a 4000
Alternatively, you can use a feeler gauge to adjust the corner stops. In this case, weights are not required.
Platform nominal Feeler Gauge capacity in Ib. in inch.
1000 0.015 2500 0.015 5000 0.020
10000 0.015
82
4. Weighing Platforms with Four Load Cells: Stainless Steel
Extend (loosen) the overload stops until the weight value displayed for the load on the platform begins to decrease (i.e., until the stops touch the load plate support).
Platform nominal Test weight for Test weight for capacity in Ib. middle stops in Ib. corner stops in Ib.
50 n/a 50 100 n/a 100 200 n/a 100 250 n/a 120 500 n/a 200
1000 n/a 500 2500 n/a 1200 5000 n/a 2500
10000 n/a 5000
Alternatively, you can use a feeler gauge to adjust the corner stops. In this case, weights are not required.
Platform nominal Feeler Gauge capacity in Lb. in inch.
50 0.12 100 0.12 200 0.12 250 0.15 500 0.12
1000 0.15 2500 0.15 5000 0.20
10000 0.15
83
Repairing the Weighing Platforms
Replacing the Connecting Cable
If the connecting cable (from junction box to Combics indicator) needs to be replaced on a weighing platform with four load cells, or if the load cell needs to be replaced on a platform with only one load cell , open the Combics indicator and disconnect the cable from the terminal strip of the A/D converter. Unscrew the cable gland; the cable or load cell can now be replaced.
Remove the isolation and connect the new cable: – Expose approx. 6 cm (2.4 in.) of the wires (3) in the cable. – Remove the isolation from approx. 1 cm (0.5 in.) of the wires and affix ferrules
to the wire ends. – Thread the cable through the cable gland. – The shielding (1) must have contact with the clamps (2). Connection to ground
via the shield. – Make sure you install the required ferrite ring (see page 73).
– After replacing the cable or load cell, use a torque wrench to tighten the cable
gland to 5 Nm.
Note: On stainless steel models, the IP67 protection must be checked after
closing the junction box.
pg9_04.tif
pg9_01.tif
fct01_23.eps
84
Replacing Load Cells
If an off-center load error is detected (platforms with one load cell), or if the output signal from the load cell is too high or too low (InF 02 is displayed during span adjustment), replace the load cell.
Because the mechanical construction of the platforms is basically uncomplicated, no detailed description of the disassembly procedure is included here.
Important:
– When replacing load cells in a weighing platform with four load cells, lift the
corner using an M16x120 threaded rod to access the cells.
The threaded rod can be obtained from a specialized supplier.
– When reassembling the weighing platform, make sure to replace any plates that
were under the platform frame.
– When replacing a load cell designated “GWT Type 011275/500 C3”, the three
bushings (two for affixing the load cell; one for the load cell base) must be
removed from the defective load cell and installed in the new load cell. These
bushings are located in bore holes in the load cell.
lastz_1.jpg
eckl_5.jpg
85
– When installing the new load cell, make sure any arrow (see the illustration on
the left) always points in the movement of the load cell. – Tighten the load cells with the required torque.
Torque Values
– After replacing load cells, you need to check the overload stops and adjust
them if necessary (see page 77-80).
lastz_2.jpg
lastz_3.jpg
Lad cell designation Torque Lad cell designation Torque
GW T Type 011231/550 C3 130Nm GW T Type 011296/50 C3 10Nm GW T Type 011232/1100 C3 130Nm GW T Type 011297/100 C3 10Nm GW T Type 011233/1760 C3 130Nm GW T Type 011298/200 C3 20Nm GW T Type 011241/7,5 C3 10N m GW T Type 011299/7,5 C3 10Nm GW T Type 011242/15 C3 10Nm GW T Type 011300/15 C3 10Nm GW T Type 011243/30 C3 10Nm GW T Type 011301/30 C3 10Nm GW T Type 011244/50 C3 10Nm GW T Type 011302/50 C3 10Nm GW T Type 011245/100 C3 14Nm GW T Type 011303/100 C3 10Nm GW T Type 011246/50 C3 14Nm GW T Type 011304/200 C3 20Nm GW T Type 011247/100 C3 14Nm GW T Type 011305/50 C3 35Nm GW T Type 011248/200 C3 14Nm GW T Type 011306/100 C3 35Nm GW T Type 011249/100 C3 35Nm GW T Type 011307/200 C3 35Nm GW T Type 011250/200 C3 35Nm GW T Type 011308/500 C3 35Nm GW T Type 011251/50 C3 35Nm GW T Type 011309/550 C3 130Nm GW T Type 011252/150 C3 35Nm GW T Type 011310/1100 C3 130Nm GW T Type 011253/250 C3 35Nm GW T Type 011311/1760 C3 130Nm GW T Type 011272/50 C3 35Nm GW T Type 011461/ 20kg 35Nm GW T Type 011273/100 C3 35Nm GW T Type 011462/ 500lb 130Nm GW T Type 011274/200 C3 35Nm GW T Type 011463/ 1250lb 130Nm GW T Type 011275/500 C3 35Nm GW T Type 011464/ 2500lb 130Nm GW T Type 011290/500 C3 35Nm GW T Type 011465 / 4000lb 130Nm GW T Type 011293/7,5 C3 10N m GW T Type 011466/ 5000lb 130Nm GW T Type 011294/15 C3 10Nm GW T Type 011467/ 40kg 10Nm GW T Type 011295/30 C3 10Nm GW T Type 011468/ 635kg 42Nm
Lastz_Drehm.xls
86
Pin Assignments in the Junction Box
Color Codes: Load Cell Cables and Connecting Cable
(Weighing Platform to the Junction Box) (Junction Box to A/D Converter)
Load cell 1-4 Output
Ye (yellow) = Ye (yellow) = Shield = Ye (yellow) Bl (blue) = Gn (green) = (1) BR_POS (Bridge supply voltage +) = Bl (blue) Gn (green) = ---- = (2) SENSE_POS (Sense +) = Gn (green) Blk (black) = Blk (black) = (6) (BR_NEG (Bridge supply voltage -) = Blk (black) or (brown) Gr (gray) = ---- = (5) SENSE_NEG (Sense -) = Gr (gray) Rd (red) = Rd (red) = (4) OUT_NEG (Measuring voltage negative) = Rd (red) Wh (white) = Wh (white) = (3) OUT_POS (Measuring voltage positive) = Wh (white)
(Wiring Bridge )
Note: The color coding of the connecting cable (junction box to A/D converter) might be
different when connecting a non-Sartorius platform to the Combics indicator.
kabelan1_e.gif
adjustment resistor
87
Replacing the Load Cells Foot
– To replace the load cells foot, move the retainer spring to the left hand and right
hand and forth with a screwdriver to push it out.
Color Codes of the Wiring for Weighing Platforms, Models CAPP../CAPS...
Platform size Pin Assignment Chart for the Indicator (see Page 71)
(1)
or Pin Assignment Chart for the Junction Box (see Page 84)
(2)
in inches No.: 1 2 3 4 5 6
12.6 x 9.5
(1)
Blue Green White Red Gray Black
15.8 x 11.8
(1)
Blue Green White Red Gray Black
19.7 x 15.8
(1)
Blue Green White Red Gray Black
24 x 18
(1 or 2)
Blue Green White Red Gray Black
24 x 24
(1 or 2)
Blue Green White Red Gray Black
30 x 24
(1 or 2)
Blue Green White Red Gray Black
30 x 30
(1 or 2)
Blue Green White Red Gray Black
Platform size Pin Assignment Chart for the Junction Box (see Page 84) in inches No.: 1 2 3 4 5 6
(Wiring Bridge )
36 x 36 Green --- White Red --- Black 48 x 48 Green --- White Red --- Black 60 x 48 Green --- White Red --- Black 60 x 60 Green --- White Red --- Black
lastz_2.jpg
88
Error Codes
Error codes are shown on the main display. Err codes are shown continuously; Inf messages are shown for 2 seconds, after which the program returns automatically to the weighing mode.
Display Cause Solution Err 101 - 104 Key is stuck Release key or
Key pressed at power on Contact your local Sartorius Service Center
Err 320 Program memory defective Contact your local Sartorius Service Center Err 335 Verified weighing platform not compatible Connect a compatible weighing platform
with the connected terminal
Err 340 Operating parameter memory (EEPROM) defective Turn the scale off and then on again
If Err 340 is still displayed, contact your local Sartorius Service Center
Err 341 Data lost from RAM; battery needs to be recharged Leave the scale power on for at least 10 hrs. Err 343 Data lost from the memory module Contact your local Sartorius Service Center
for transaction numbers in external alibi memory
Inf 01 Data output not compatible with output format Change the menu settings Inf 02 Calibration/adjustment condition Calibrate only when zero is displayed.
not met; for example, not tared or Press ) to tare there is a load on the weighing pan Unload the scale
Inf 03 Adjustment could not Allow the scale to warm up again and then
be d within a certain time repeat the adjustment process
Inf 06 Built-in calibration weight defective Contact your local Sartorius Service Center Inf 07 Function not allowed in scales Contact your local Sartorius Service Center
verified for use in legal metrology for details on changing settings
Inf 08 The load on the scale is too heavy to zero the readout Check whether “Tare/zero at power on” is set (1.12)
89
Display Cause Solution Inf 09 Taring is not possible when the gross weight Zero the scale
is a minus value
Inf 10 Tare key is blocked when there The data stored in the 2nd tare memory
is data in the tare memory (Combics 2 only) must be deleted (clear
the memory) before taring
Inf 22 Error in storing reference value, load is too light Put a heavier weight on the scale Inf 23 Error in initializing an application Contact your local Sartorius Service Center Inf 29 Minimum load not reached Change menu, or perform “Close” function check
the interface and cable of the connected devicen Contact your local Sartorius Service Center
Inf 30, 31 Indicator is in the xBPI-mode Define a lower value for the minimum load
(in the Application settings, menu item 3.6)
Inf 71 Cannot store the current weight value None
(e.g., if control limits are too low or too high)
Inf 72 Cannot store the current weight value None
(e.g., the transaction counter has reached its limit)
Inf 73 Data not found or unreadable Contact your local Sartorius Service Center Inf 74 Function is blocked (e.g., menu is locked) None Inf 98 No weighing platform connected Connect weighing platform Inf 99 No weighing platform connected Connect weighing platform NO WP No weighing platform connected Connect weighing platform
90
The Combics Weighing Platforms
Painted version:
- Weighing ranges from 10 lbs to 10,000 lbs
- 11 different dimensions
- 11 weighing ranges
- 37 models
- up to 5,000e NTEP; up to 25,000d
- Version: painted steel
- IP65 protection rating
- Drive-on ramps, built-in frame
- Customized dimensions possible
Stainless steel version:
- Weighing ranges from 10 lbs to 10,000 lbs
- 11 different dimensions
- 11 weighing ranges
- 37 models
- up to 5,000e NTEP; up to 25,000d
- AISI 304 (1.4301), AISI 316Ti (1.4571) stainless steel, electropolished, with tread plate structure
- IP67 / IP68 protection rating
- Weighing cell in stainless steel
- Drive-on ramps, built-in frame, customized dimensions
91
Type Designation
Example for the order number of a Combics weighing platform:
CAPP4U-1000KK-LU
weighing platform.jpg
92
Example for the order number of a stainless steel Combics weighing platform:
CAPS4U-500HH-LU
weighing platform_2.jpg
93
Service Specification CAPP/CAPS1U-.....-LU
Example for the order number of a Combics full range scale: CAPP1U-1000HH-LU (Type Designation see page 91-92)
Applications Material | Number of Weighing Platform size Resolution
Design load cells Capacity (Ibs) inches | order code
CAP P 1 U - 1000 HH - LU
just_combics_USA.xls
Repro ducibilit y
Off-ce nter lo ad
Eccen tricity
Span Line arity TCS iso CAL
Model
Weigh ing
capaci ty
Readability
Test
weight
Per mis sib le
tolerance
s
Test
weight
Per mis sib le
tolerance
(±)
Class
Adjus tm.
weight
Test
weight
Per mis sib le
tolerance
(±)
Tarew eight Test we ight
Pe rm iss ib le
tolerance
(±)
ppm
/K
K
CAPP1U-10..-LU CAPS1U-10..-LU
10 lbs 0,0002 lbs 5 lbs 0,002 lbs 5 lbs 0,004 lbs M1 1 lbs 3 lb s 0,002 lbs
--­(1)
lbs 0,7/1,5/2,2/3 lbs 0,002 lbs 10 n
CAPP1U-25..-LU CAPS1U-25..-LU
25 lbs 0,0005 lbs 12,5 lbs 0,005 lbs 12,5 lbs 0,01 lbs M1 2 lbs 6 lb s 0,004 lbs
--­(2)
lbs 1,5/3/4/6 lbs 0,005 lbs 10 n
CAPP1U-50..-LU CAPS1U-50..-LU
50 lbs 0,001 lbs 25 lbs 0,01 lbs 25 lbs 0,02 lbs M1 5 lbs 15 lbs 0,01 lbs
--­(5)
lbs 3/7/11/15 lbs 0,01 lbs 10 n
CAPP1U-100..-LU CAPS1U-100..-LU
100 lbs 0,002 lbs 50 lbs 0,02 lbs 50 lbs 0,04 lbs M1 10 lb s 30 lb s 0,02 lbs
---
(10)
lbs 7/15/22/30 lbs 0,02 lbs 10 n
CAPP1U-200..-LU CAPS1U-200..-LU
200 lbs 0,005 lbs 100 lbs 0,05 lbs 100 lbs 0,05 lbs M1 20 lbs 60 lb s 0,04 lbs
---
(20)
lbs 15/30/45/60 lbs 0,05 lbs 10 n
CAPP1U-250..-LU CAPS1U-250..-LU
250 lbs 0,005 lbs 125 lbs 0,05 lbs 125 lbs 0,1 lbs M1 50 lb s 150 lbs 0,1 lbs
---
(50)
lbs 30/70/110/150 lbs 0,05 lbs 10 n
CAPP1U-500..-LU CAPS1U-500..-LU
500 lbs 0,01 lbs 250 lbs 0,1 lbs 250 lbs 0,2 lbs M1 100 lbs 300 lbs 0,2 lbs
---
(100)
lbs 70/150/220/300 lbs 0,1 lb s 10 n
CAPP1U-1000..-LU CAPS1U-1000..-LU
1000 lbs 0,02 lbs 500 lbs 0,2 lbs 500 lbs 0,4 lbs M1 500 lbs 1500 lbs 1 lb s
---
(500)
lbs
300/700/
1100/1500
lbs 0 ,2 lbs 10 n
94
Service Specification CAPP/CAPS4U-.....-LU
Example for the order number of a Combics full range scale: CAPS4U-5000KK-RR (Type Designation see page 91-92)
Applications Material | Number of Weighing Platform size Resolution
Design load cells Capacity (Ibs) inches | order code
CAP S 4 U - 5000 RR - LU
just_combics.xls
Repro ducibilit y
Off-ce nter lo ad
Eccen tricity
Span Line arity TCS iso CAL
Model
Weigh ing
capaci ty
Readability
Test
weight
Per mis sib le
tolerance
s
Test
weight
Per mis sib le
tolerance
(±)
Class
Adjus tm.
weight
Test
weight
Per mis sib le
tolerance
(±)
Tarew eight Test we ight
Pe rm iss ib le
tolerance
(±)
ppm
/K
K
CAPP4U-1000..-LU CAPS4U-1000..-LU
1000 lbs 0,02 lbs 500 lbs 0,2 lbs 500 lbs 0,2 lbs M1 500 lbs 1500 lbs 1 lbs
---
(500)
lbs
300/700/
1100/1500
lbs 0 ,2 lbs 10 n
CAPP4U-2500..-LU CAPS4U-2500..-LU
2500 lbs 0,05 lbs 1250 lbs 0,5 lbs 1250 lbs 0,5 lbs M1 2500 lb s 3 lbs 0,002 lbs
--­(1)
lbs 0,7/1,5/2,2/3 lbs 0,5 lbs 10 n
CAPP4U-5000..-LU CAPS4U-5000..-LU
5000 lbs 0,1 lbs 2500 lbs 1 lbs 2500 lbs 1 lbs M1 5000 lbs 3 lb s 0,002 lbs
--­(1)
lbs 0,7/1,5/2,2/3 lbs 1 lbs 10 n
CAPP4U-10000..-LU CAPS4U-10000..-LU
10000 lbs 0,2 lbs 5000 lbs 2 lbs 5000 lbs 2 lbs M1 10000 lbs 3 lb s 0,002 lb s
--­(1)
lbs 0,7/1,5/2,2/3 lbs 2 lbs 10 n
95
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Hotline +49.551.308-4440 Fax +49.551.308-4449 www.sartorius.com e-mail: Int.Service@Sartorius.com
Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG.
The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to the technology, features, specifications, and design of the equipment without notice.
Status: October 2003, Sartorius AG, Goettingen, Germany
Printed in Germany on paper that has been bleached without any use of chlorine · W1A000 · I.K. Publication No.: WCI5005-e03104
Loading...