Sanyo EMF1010M Schematic

File No :
SERVICE MANUAL Microwave Oven EM-F1010MW
EM-F1010MB
Model No : Product Code No :
EM-F1010MW 437 373 72
EM-F1010MB 437 373 73
(U.S.A.)
CAUTION
WARNING TO SERVICE TECHNICIANS
PRECAUTIONS TO BE OBSERVED BEFORE AND DURING
SERVICING TO AVOID POSSIBLE EXPOSURE TO
EXCESSIVE MICROWAVE ENERGY
(a) Do not operate or allow the oven to be operated with the door open.
(b) Make the following safety checks on all ovens to be serviced before activating the magnetron or other
(c) Before turning on microwave power for any service test or inspection within the microwave generating
compartments, check the magnetron, wave guide or transmission line, and cavity for proper alignment, integrity, and connections.
(d) Any defective or misadjusted components in the interlock, monitor, door seal, and microwave generation
and transmission systems shall be repaired, replaced, or adjusted by procedures described in this manual before the oven is released to the owner.
(e) A microwave leakage check to verify compliance with the Federal performance standard should be
performed on each oven prior to release to the owner.
REFERENCE NO. SM-860329
CAUTION
For microwave energy emission
In the area emitting the highest reading, switch the meter to SLOW RESPONSE, and take a reading for minimum of three (3) seconds. We recommended the pattern outline shown below when the door surface is surveyed.
On every service calls, check for microwave energy emission, must be made according to the following manner.
Measurement of energy emission
Measurement must be made with the microwave oven operating at its maximum output and containing a load of 275±15 milliliters of tap water initially at 20o±5o celsius (68 ±9oF) placed within the cavity at the center.
NOTE : The water container must be a 600 milliliter beaker
and made of an electrically none conductive material such as glass or plastic. The cook tray must be in place when measuring emission.
A properly operating door and seal assembly will normally register emission on greater than 4mW/cm measurement uncertainty with the cooking shelf or tray in place.
2
to allow for
All repairs must be performed in such a manner that microwave energy emission are minimal.
Follow the instructions supplied with a detector being used and performed an R.F. emission test around the door front and edges and all edges and vent of the outer case. The cabinet (wrapper) must be in place and the oven fully assembled.
When performing emission survey, with the meter on FAST RESPONSE the movement of a detector probe shall not exceed one (1) inch per second.
NOTE : Periodically check to be sure that the probe tip is
not worn or dirty.
The following U.S. standard applies to microwave ovens : 21 CFR 1030.10, Performance Standard for Microwave Ovens.
It requires that the power density of the microwave radiation emitted by a microwave oven shall not exceed five (5) milliwatts per square centimeter at any point 5 centimeter (about 2 inches) or more from the external surface of the oven.
All microwave ovens exceeding the emission level of 4mW/cm microwave ovens and the manufacturer immediately and the owner should be told not to use the microwave oven until it has been repaired completely.
If a microwave oven is found to operate with the door open, report to Dept. of Service, the manufacturer and CDRH* immediately. Also tell the owner not to use the oven. *CDRH : Center for Device and Radiological Health.
The Interlock Monitor Switch acts as the final safety switch protecting the customer from microwave radiation. If the Interlock Monitor Switch operates to blow the Fuse with interlocks failed, you must replace all Interlock Switches Primary and Secondary Interlock Switches and the Monitor Switch with new ones because the contacts of those Interlock Switches may be melted and welded together.
2
must be reported to Dept. of Service for
- TABLE OF CONTENTS -
Adjustment Procedures ......................................... 1
Specifications ....................................................... 2
Power Output Measurement .................................. 2
Precautions and Repair Service Tips ........................ 2
1.ADJUSTMENT PROCEDURES
TO AVOID POSSIBLE EXPOSURE TO MICROWAVE ENERGY LEAKAGE, THE FOLLOWING
ADJUSTMENT OF THE INTERLOCK SWITCHES SHOULD BE MADE ONLY BY AUTHORIZED SERVICE PERSONNEL.
The SANYO service center should have the designated detector to measure the microwave energy leakage after the repair or adjustment.
NOTE : Detector to be used at the service center is
NARDA 8100, 8200 or the equivalent.
PRIMARY INTERLOCK SWITCH, INTERLOCK MONITOR SWITCH AND DOOR SENSING SWITCH ADJUSTMENT
(Figure 1)
(1) Loosen 2 screws securing the Lever Stopper.
(2) Adjust the Lever Stopper position so that it is pushed
forward and pull backward until there is about zero gap between the Latch Lever and the switch body on the Primary Interlock Switch and at the same time there is about zero gap between the Door Latch and the switch body on the Door Sensing Switch when the Door Latch is securely locked.
(3) Tighten the Lever Stopper screws securely.
(4) Make sure the Interlock Monitor Switch closes after
the Primary Interlock Switch opens when the door is opened very slowly, according to “CHECKOUT PROCEDURE FOR SWITCHES” on page 6.
(5) Make sure the Interlock Monitor Switch opens before
the Primary Interlock Switch closes when the door is closed very slowly, according to “CHECKOUT PROCEDURE FOR SWITCHES” on page 6.
(6) Make sure the microwave energy leakage should be
no greater than 4mW/cm2 to allow for measurement uncertainty when measured with a detector.
Circuit Diagram ............................................. 3
Test Procedures and Troubleshooting............... 4 ~ 7
Disassembly Instructions ................................ 8 ~ 10
Exploded View and Parts List .......................... 11 ~ 14
NOTE
Screws
: If the Interlock Monitor Circuit operates and
at the same time the fuse blows with the door opened, be sure to replace the Door Sensing Switch and the electric circuit related on the Door Sensing Switch, which act as Secondary Interlock Switch.
Door Latch
Lever Stopper
Door Sensing Switch
Interlock Monitor Switch
(All service adjustments must be made for minimum microwave energy leakage readings.)
- 1 -
Figure 1
Latch Lever
Primary Interlock Switch
2. SPECIFICATIONS
Rated Power Consumption .. 920W.
Microwave Output .............. 600W.
Frequency .......................... 2,450MHz±50MHz.
Power Supply ..................... 120V±12V, 60Hz.
Rated Current ..................... 8 Amp.
Safety Devices ................... Thermal Fuse open at 332°F
(167°C) for Cavity. Thermal Protector open at 252°F (122°C) for Magnetron. Fuse (Cartridge Type 15A) Primary Interlock Switch, Door Sensing Switch and
Interlock Monitor Switch.
Timer ................................ 60Mins.
Overall Dimensions ............... 18
Oven Cavity Size ................ 113/4”(W)x121/4”(D)x71/2”(H)
Turn Table Diameter ............ 107/8”
Effective Capacity of
Oven Cavity ....................... 0.7 Cubic Feet.
Net Weight ........................ Approx. 23 Lbs.
1
/16”(W)x1 33/4”(D)x913/16”(H)
3. POWER OUTPUT MEASUREMENT
NOTE
The Power Output specification, 600W. on this model is measured with IEC measurement. The Power Output is measured with two (2) liters water is equivalent to 600W. in measurement with IEC, when measured with the following Power Output.
(1) Fill two beakers (glass or plastic) with each one liter
of tap water (about 20°C) and measure the water temperature. (Use a thermometer with a 1/10 degree gauge).
(2) Place the beakers side by side in the center of the
glass tray.
(3) (a) Set the Timer at any position more than 2 minutes.
(b) Heat the water correctly for 2 minutes.
(4) Take the beakers out, immediately stir the water and
measure the water temperature respectively.
(5) Calculate the temperature rise of water in each beaker.
Then calculate the average value of two temperature rises. Output power can be calculated by the equation :
Power Output (W) = 70 x ∆t Where t is an average temperature rise in degrees
Centigrade.
(6) Power Output shall be in the following range :
Average Temperature Rise Power Output
Minimum : 13.9°F (7.7°C) 540W. Maximum : 16.9°F (9.4°C) 660W.
(7) Power Output is affected by the line voltage under
load. For correct Power Output measurement, the line voltage under load must be 120±1 volts.
4. PRECAUTIONS AND REPAIR SERVICE TIPS
PRELIMINARY
A. SINCE NEARLY 2,000 VOLTS EXISTS IN SOME
CIRCUITS OF THIS MICROWAVE OVEN, REPAIRS SHOULD BE CARRIED OUT WITH GREAT CARE.
B. TO AVOID POSSIBLE EXPOSURE TO MICROWAVE
ENERGY LEAKAGE, THE FOLLOWING PRECAUTIONS MUST BE TAKEN BEFORE SERVICING.
(1) Before the power is applied :
(a) Open and close door several times to make sure the
Primary Interlock Switch, the Interlock Monitor Switch and the Door Sensing Switch operate properly. (Listen for the clicking sound from the switches). Make sure the Interlock Monitor Switch closes after the Primary Interlock Switch is opens when the door is opened. (See pages 1 and 6).
(b) Make sure the perforated screen and the Choke
Dielectric of the door are correctly mounted.
(2) After the power is applied :
(a) Open and close the door to see if the interlock
mechanism operates properly.
(b) Check microwave energy leakage with a leakage
detector and confirm the energy leakage should be no greater than 4mW/cm2 to allow for measurement uncertainty.
(3) Do not operate the unit until it is completely repaired,
if any of the following conditions exists :
(a) Door does not close firmly against the cavity front. (b) The hinge is broken. (c) The Choke Dielectric or the door seal is damaged. (d) The door is bent or warped, or there is any other
visible damage to the oven that may cause microwave energy leakage.
NOTE : Always keep the seal clean.
(e) Make sure that there are no defective parts in the
interlock mechanism.
(f) Make sure there are no defective parts in the
microwave generating and transmission assembly. (especially waveguide).
(4) The following items should be checked after the unit
is repaired :
(a) The Interlock Monitor Switch is connected correctly
and firmly.
(b) The magnetron gasket on the Magnetron is properly
positioned.
(c) Waveguide and Oven Cavity are intact (no leakage
of microwave energy).
(d) The door can be properly closed and the safety
switches work properly.
(e) The oven must be stopped when the door is opened
or the time is up.
The oven must not be operated with any of the above components removed or bypassed.
- 2 -
5. CIRCUIT DIAGRAM
Figure 2
CONDITION OF OVEN
TIMER : OFF DOOR : OPEN
PRIMARY INTERLOCK SWITCH :
LOWER DOOR INTERLOCK SWITCH
SECONDARY INTERLOCK SWITCH:
UPPER DOOR INTERLOCK SWITCH
WIRING COLOUR
BK : BLACK BL : BLUE W : WHITE Y : YELLOW R : RED
NOTE : The parts marked with have special characteristics of safety,
if replace the marked parts, use only SANYO specified parts.
- 3 -
Loading...
+ 11 hidden pages