Sam 271 Directions For Use Manual

Page 1
Directions
for Use
Model sam-271
Page 2
sam
Product Support 1-888-202-9831
Page 3
Contents
1. Introduction .............................................................................................................................................. 4
2. Indications for Use ................................................................................................................................ 4
3. Safety .......................................................................................................................................................... 5
3.1. Contraindications ................................................................................................................................... 5
3.2. Warnings .................................................................................................................................................... 5
3.3. Precautions ............................................................................................................................................... 6
5. sam Components ............................................................................................................................... 9
6. Operator Interface ............................................................................................................................... 10
7. Accessories ............................................................................................................................................ 10
8. LED Display .............................................................................................................................................. 11
9. Initial Setup Instructions ................................................................................................................... 13
10. Treatment Options ............................................................................................................................... 13
11. Application Instructions .................................................................................................................... 16
12. Device Power Down and Removal ............................................................................................... 22
13. Modes of Operation ........................................................................................................................... 23
14. Cleaning and Maintenance ............................................................................................................. 24
15. Storage and Operating Conditions ............................................................................................. 24
16. Disposal of Waste Products ........................................................................................................... 25
17. Appendix ................................................................................................................................................. 26
A. Symbols ........................................................................................................................................ 26
B. Specifications ............................................................................................................................. 27
C. Notice of Compliance, Calibration and Operational Period .................................. 28
D. Technical Information ............................................................................................................. 29
G. Troubleshooting ....................................................................................................................... 34
3ZetrOZ Systems Product Support 1-800-202-9831
Page 4
1. Introduction
Thank you for choosing sam! This manual contains general instructions for operation, application, precautions, and maintenance. In order to obtain maximum life and efficiency from the sam Device and to assist in the proper operation of the device, please read and understand this manual thoroughly. This device is only to be used as directed in this manual.
sam was developed as a next generation wearable ultrasound therapy system which
combines miniaturization technology into a small and portable ultrasound therapy system. It is designed to work with the human body and maximize the safe and effective delivery of long-duration therapeutic ultrasound. Simple to administer and operate on a broad range of body types, sam allows the delivery of ultrasound treatment for up to four hours. It operates at a preset frequency and allows the use of up to two applicators simultaneously. Applicators are applied and secured to the surface of the body using convenient sam Ultrasound Coupling Patches.
The specifications put forth in this manual were in effect at the time of publication.
1.1. GENERAL SAFETY
Thoroughly read and understand the precautionary and operating instructions before attempting to operate the sam Ultrasonic Diathermy Device. Know the limitations and hazards associated with using any ultrasound device. Observe the precautionary and operational information on the device. Periodically review the operation procedures and safety precautions outlined in this manual.
1.2. USA: PRESCRIPTION USE ONLY
Caution: Federal law restricts this device to sale by or on the order of a practitioner
licensed by the law of the state in which he/she practices to use or order the use of the device.
This device complies with 21 C.F.R. § 1050.10
2. Indications for Use
The sam Ultrasonic Diathermy Device is intended to apply ultrasonic energy to generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures and the local increase of circulation.
The sam Ultrasonic Diathermy Device is a prescription use device. sam should only be administered and monitored by a licensed healthcare practitioner.
4 Product Support 1-888-202-9831
Page 5
3. Safety
3.1. CONTRAINDICATIONS
Contraindications for the use of ultrasound include:
• Over an area of the body where a malignancy is known to be present
• Over the eyes
• Over or near growth centers until bone growth is complete
• Over the reproductive organs
• Over the pregnant uterus
• Over a healing bone fracture
• On the thoracic area if the patient is using a cardiac pacemaker
• Over an active implanted medical device such as an implanted deep brain stimulation device
• On the brain, spinal cord, or large subcutaneous peripheral nerves
• Ischemic tissues in individuals with vascular disease where the blood supply would be unable to follow the increase in metabolic demand and tissue necrosis might result
3.2. WARNINGS
• If the treatment is reported as painful or too hot at any point during treatment, turn off device and remove the device from the skin.
• Instruct the patient to inform the practitioner if the patient feels any pain or burning during treatment.
• Instruct the patient how to turn off the sam Device and remove the sam Applicator if the patient feels any pain or burning during treatment.
• ALWAYS administer treatment using a new sam Ultrasound Coupling Patch. Use one sam Ultrasound Coupling Patch per applicator. Use of the sam Ultrasound Applicator without a new sam Ultrasound Coupling Patch MAY RESULT IN BURN and/or REPEATED SHUTOFF of the sam Applicator.
• The sam Device should be kept out of the reach of children.
DO NOT apply the sam Applicator with alternative coupling media as a replacement for the sam Ultrasound Coupling Patch. Use of alternative coupling media in lieu of the sam Ultrasound Coupling Patches may reduce the effectiveness of treatment, lead to automatic shutoff of the applicator, or cause a burn.
5ZetrOZ Systems Product Support 1-800-202-9831
Page 6
DO NOT administer treatment if the applicator is not connected to a sam Ultrasound Coupling Patch.
• Applicators and sam Ultrasound Coupling Patches are not sterile. DO NOT apply this device over an open wound or inflamed skin.
DO NOT use the sam Ultrasound Coupling Patch if the sam Ultrasound Media is dried out. Indications of a dried out Patch include: the cup is not full of gel, there is dry residue or film in the cup, or there is any cut, break, or opening in the Patch or seals.
DO NOT apply directly over a bone that is near the skin surface.
3.3. PRECAUTIONS
Precaution should be taken when using the device:
• Over an area of the spinal cord following a laminectomy, i.e. when major covering tissues have been removed
• On patients with hemorrhagic diatheses
• Over areas where metal prosthesis or other metallic implants are embedded in tissue which may form a reflective surface to the ultrasound energy causing unintended irradiation of tissue and excessive heating
• Over an acute infection or sepsis
• On patients with peripheral artery disease
• Over a deep vein thrombosis
• Over an anesthetized area or in conjunction with a condition that causes impairment of sensation, such as caused by chemotherapy
• When using the sam Ultrasound Coupling Patch, ensure the top and bottom seals have been completely removed before attaching the applicator to the skin
3.4. INFLAMMABLE GASES AND ANESTHETICS
• Warning: Explosion hazard if used in the presence of flammable anesthetics, open flame, or oxygen-rich environment
6 Product Support 1-888-202-9831
Page 7
3.5. ELECTRONICS AND BATTERY
WARNINGS
• This device is rated IPX0; therefore, it is Not Waterproof. DO NOT apply a direct stream of any liquid onto the device, submerge the device, or allow any liquid to pool on the surface of the device. DO NOT use if device has been submerged in water.
• This device contains a rechargeable lithium-ion battery. DO NOT disassemble,
DO NOT heat above 100°C, DO NOT incinerate or expose to water and DO NOT ingest.
• The use of accessories, transducers and/or cables other than those specified, with the exception of those sold by the manufacturer as replacement parts for internal components, may result in increased emissions or decreased electrical immunity of the equipment or system.
• DO NOT open or modify any component of the sam Device. Hazards such as shock, burn or inappropriate functionality can result from unauthorized modification of the sam Device.
• Use of the sam Device around electromagnetic interference may negatively affect the output performance and safety of the device. Do not use the device if any abnormal functionality occurs.
PRECAUTIONS
• Only recharge the sam Device using the sam Electrical Charger. Use with any other recharging device may result in damage to the system and void all warranties.
• When not in use, power ‘OFF’ the device to protect the functionality of the components.
• Avoid dropping the applicator or power controller and avoid scratching the lens of the applicator. Rough handling may reduce the device’s acoustic output power, thereby reducing the effectiveness of therapy.
• The power controller and applicators should be routinely checked for cracks and other damage before each use to determine that the device functions normally.
DO NOT place the device in a location where the power charging cord could be a trip hazard.
DO NOT use sharp objects such as a pencil point or ball point pen to operate the buttons on the control panel as damage may result.
• Prevent potential electromagnetic or other interference. DO NOT open the sam Device or connect the device or components of the device to any non-sam part. Keep the device clean and make sure no exposed non-insulated wires are visible. If damage is present, do not administer treatment.
7ZetrOZ Systems Product Support 1-800-202-9831
Page 8
4. Features of the sam Device
4.1. PRESET TREATMENT
The sam Device is preconfigured to provide continuous ultrasonic output at a preset frequency and intensity which cannot be modified by the user. The user can set the treatment duration to be 1, 2, 3, or 4 hours.
4.2. sam ULTRASOUND APPLICATORS
The sam Ultrasound Applicators serve as the ultrasound transducers of the
sam Device. The applicators offer low-profile design with light emitting diode (LED) on/
off notification. The ergonomic plastic housing and smooth contours provide enhanced comfort.
4.3. sam SENSING
sam is designed to work with the human body and maximize the safe and effective
delivery of long-duration therapeutic ultrasound. Each sam Applicator is equipped with closed-loop continuous temperature monitoring which maintains treatment site temperatures below 44°C during normal operation. See section 13.4 on page 24 for more information.
4.4. sam ULTRASOUND COUPLING Patches
The sam Device utilizes ultrasound coupling Patches which are manufactured with ultrasound coupling media sealed inside. The ultrasound coupling Patches ARE REQUIRED to secure the sam Applicators to the body.
4.5. BATTERY OPERATION
Powered by a rechargeable lithium-ion battery, the sam Device can provide 4 hours of therapy on a single battery charge.
8 Product Support 1-888-202-9831
Page 9
UB-12
AT- 1 2
5. sam Components
PC271 UB14
4 HRS
K
C
E
H
C
Y
R
E
T
T
A
B
3
2
1
DONE
CT271
YB271
YB-12
if included
AT271
AT-1 2
figure 1: sam Model 271 Components
sam MODEL 271 COMPONENTS
AT271: Ultrasound Applicators
UB14: Ultrasound Coupling Patches
OM271: User Manual
PC271: Power Controller
CT271: Electrical Charger
YB271: Y-branch Adapter
AT271
AT-1 2
9ZetrOZ Systems Product Support 1-800-202-9831
Page 10
6. Operator Interface
Battery
Indicator LED
Power Button
K
C
E
H
C
Y
R
E
T
T
A
B
Treatment Timer LED
Front
Charge Port
Indicator LED
4 HRS
3
2
1
DONE
figure 2: sam Interface Components
Side
Bottom
Toggle Button
Indicator LED
Applicator
Connection
7. Accessories
The sam Device may be used with any of the following accessories:
sam Belt Clip
sam Armband
sam Carrying Case
* Additional sam Ultrasound Coupling Patches may be obtained by contacting
the manufacturer
10 Product Support 1-888-202-9831
Page 11
8. LED Display
The sam Device contains Light Emitting Diode (LED) displays that indicate the functions of the device (see Figure 2).
8.1. POWER CONTROLLER INDICATOR LED
The Indicator LED provides power, charging and error checking information.
Color of Indicator LED Meaning
Opaque (absence of light) Power Controller is OFF. No power is being generated Blue Power Controller is ON. Note – Ultrasonic
energy is generated only when the Power Controller is connected to the sam Applicator(s)
Green Power Controller battery is fully charged Amber Power Controller battery is charging Red Too many applicators are attached. A
maximum of two applicators may be attached to a single power controller at any time
Table 1. Power Controller LED Color Definitions
11ZetrOZ Systems Product Support 1-800-202-9831
Page 12
8.2. CONTROL PANEL
The front face of the sam Power Controller contains a control panel. The LEDs on the control panel are divided into a field for Battery Indicator and a field for Treatment Timer. The operator is able to make selections by pressing the Power Button and the Toggle
Button.
Color of Battery Indicator LED* Meaning
Red Only Low Battery, Re-Charge required Blinking Red Re-Charge required Amber Power Controller is partially charged Green Power Controller fully charged *These lights remain on for 4 seconds and subsequently turn off
Table 2. Battery Indicator LED Color Definitions
Treatment Time Indicator LED Meaning
Blue Treatment time remaining. Treatment
timer is set in one hour increments.
Amber Treatment complete. This single “done”
light will activate only when a treatment cycle has completed and will remain illuminated for 8 hours, then turn off
Table 3. Treatment Time Indicator LED Color Definitions
8.3. APPLICATOR INDICATOR LED
LED Color Meaning
Opaque (absence of light) Applicator is not receiving power Blue Applicator is receiving power from the
Power Controller. Ultrasonic energy is being generated at all times this light is illuminated
Red (accompanied by vibration) sam Sensing Mode. No Ultrasonic energy
is being generated from this applicator while this light is illuminated
Table 4. Applicator LED Color Definitions
12 Product Support 1-888-202-9831
Page 13
Bottom of sam Power Controller
Micro USB
Bottom of sam Power Controller
Micro USB
Wall Charger
Outlet
9. Initial Setup Instructions
Remove the sam Device components (Figure
1) from the packaging and inspect for any damage
that may have occurred during shipment. Charge the device for up to 6 hours or until the Power Controller Indicator LED is Green. The sam Device arrives only partially charged due to generally accepted shipping practices.
9.1. CHARGING THE POWER
CONTROLLER
A. Peel back the dust cover on the bottom
of the sam Power Controller and plug the micro USB end of the sam Electrical Charger into the charging port.
B. Plug the electrical charger into a 120230
VAC wall outlet. The sam Power Controller indicator LED will be amber. When the device is fully charged the top right indicator LED will change from amber to green.
figure 3A
10. Treatment Options
10.1. TREATMENT DURATION
The sam Ultrasonic Diathermy Device provides ultrasound therapy at a preset frequency and intensity. The user can program the treatment duration to be 1, 2, 3, or 4 hours. The maximum treatment duration setting on the sam Device is 4 hours. Treatment duration should be set to the minimum increment required for effective therapy. Due to individual differences in skin type and tolerance, it is recommended to begin with a 1 hour treatment. Increase treatment duration in subsequent applications only as tolerated. Maximum usage time is 4 hours per day per treatment area.
The sam Device may be used while charging. Charging the device during treatment has no impact on device frequency or intensity output settings. If using sam while charging, position the device in such a way that the charger cable is easily removed from the power Controller if needed.
figure 3B
13ZetrOZ Systems Product Support 1-800-202-9831
Page 14
10.2. TREATMENT LOCATIONS
Knee
Shoulder
Figures 4 and 5 depict examples of device placement on two treatment locations. Figure 4A and 4B illustrates using the sam Device with one applicator on the shoulder and knee. Figure 5A and 5B illustrates using the sam Device with two applicators on the shoulder and knee. When determining treatment location, consider that the maximum diathermic effect will occur directly under the applicator face; therefore, placing the applicator near or directly over the target area while following all warning and cautionary instructions is advised.
Note: These figures are examples and are not
intended to be the suggested or only allowable applicator configurations for those body locations.
Warning: Do Not apply directly over a bone that is
near the skin surface.
figure 4A: Single Shoulder Application
10.3. ONE OR TWO APPLICATORS
The sam Device may be used with one or two applicators simultaneously. The decision to use one or two applicators is dependent on the size and anatomical area of treatment. Using two applicators allows a larger area of tissue to receive ultrasound treatment. For example, large anatomical regions, such as the shoulder could benefit from two applicators, whereas smaller treatment areas such as the forearm may only require one.
10.3.1. SINGLEAPPLICATOR MODE
Connect the power controller wire to a single applicator by inserting the power controller wire jack into the matching cavity at the base of the applicator. The port for the power controller wire is described as ‘Applicator connection’ on the Power Controller drawing shown in Figure 2, page 10.
figure 4B: Single knee Application
figure 5A: Dual Shoulder Application
figure 5B: Dual Knee Application
14 Product Support 1-888-202-9831
Page 15
10.3.2. DUALAPPLICATOR MODE
The Y-adapter (optional) may be used to power two Applicators simultaneously. As shown in Figure 8, connect the Y-adapter directly to the Power Controller wire and then connect each Applicator to a wire jack at the other end of the Y-adapter.
When using two sam Applicators and two sam Ultrasound Coupling Patches simultaneously, the sam Ultrasound Coupling Patches should be positioned so that their footprints do not overlap.
Caution: DO NOT overlap sam Ultrasound Coupling Patches (Figure 6).
Caution: DO NOT use the Y-adapter when using only one applicator for treatment.
Caution: Do NOT attach more than one Y-adapter to the sam Device.
Note: If two applicators will be used, two sam Ultrasound Coupling Patches (one for
each applicator) and the Y-adapter must be used for the treatment.
figure 6: NO Patch Overlap
15ZetrOZ Systems Product Support 1-800-202-9831
Page 16
11. Application Instructions
11.1. CHECK DEVICE CHARGE
A. Check to make sure the Power Controller
holds enough charge to provide the desired duration of treatment. Press the Toggle Button on the sam Power Controller to view the battery indicator lights (Figure 7). The lights will illuminate to show how much battery life is remaining in the device. If the device is not fully charged it may not be able to deliver treatment for 4 hours. The device will only permit the user to set a treatment duration for which the battery charge is capable of fulfilling.
i. To check the treatment duration for
which the battery charge is capable of fulfilling, simply continue to press the Toggle Button sequentially to view all timer settings allowed by the current battery charge. See section 9 for charging instructions.
figure 7: Check Device Charger
Toggle Button
16 Product Support 1-888-202-9831
Page 17
11.2. CONNECT APPLICATOR TO POWER CONTROLLER
4
HRS
3
2
1
DONE
B
A
T
T
E
R
Y
C
H
E
C
K
45
Rotate Clockwise
45
* If using 2 applicators: first connect the Y-adapter directly to the power controller
wire jack and then connect each applicator to a wire jack at the other end of the Y-adapter (Figure 8*).
A. Insert the power controller wire into the matching cavity at the base of the
applicator at a 45 degree angle (Figure 8A).
B. Gently twist the wire jack clockwise until the applicator edge is flush with the
wire jack (Figure 8B).
figure 8*: Connecting Y-Adapter
figure 8B: Gently twist to Lock
figure 8A: Insert power controller wire at 450 angle
17ZetrOZ Systems Product Support 1-800-202-9831
Page 18
11.3. ATTACH APPLICATOR TO PATCH
A. Remove the circular seal from the top of
the Patch to reveal the coupling media within the gel cup (Figure 9A).
B. While supporting the edges of the cup
with your fingers, align the applicator, face down, above the cup and coupling media (Figure 9B).
C. While continuing to support the edges
of the cup, firmly press the applicator down onto the center of the gel cup until a clicking noise is heard or clicking sensation is felt (Figure 9C).
If using dual applicators, repeat these steps with the second applicator and the second Patch.
Note: When the applicator is pressed down onto
the gel cup, keep your fingers on the back of the patch, keeping them on the edges of the cup and NOT at the center of the Gel pack.
Note: Do not be concerned if coupling media
flows out the edges of the gel cup. Wipe any excess coupling media away with a tissue or towel
figure 9A: Removing Foil Seal
figure 9B: Attach Applicator
Note: Ensure the applicator is fully attached to the
Patch.
18 Product Support 1-888-202-9831
figure 9C: Press Applicator Down
Page 19
11.4. APPLY ULTRASOUND
Knee
COUPLING PATCH TO TREATMENT LOCATION
A. Hold the applicator so the bottom of the
Patch faces up. Remove the circular foil seal, revealing the gel.
B. Remove the paper liner, revealing the
Patch adhesive (Figure 10A).
i. Add more sam Coupling Media
as needed.
C. Turn the Patch over and adhere the Patch
to the desired treatment location (Figure 10B).
i. The sam Coupling Media must be in
direct contact with the skin.
If using dual applicators, repeat these steps to adhere the second Patch and applicator to the second treatment site.
Note: When applying Patches, both sides of
the ultrasound coupling media should be uncovered. One side should be in contact with the skin. The other side should be in contact with the face of the applicator.
figure 10A: Removing Paper Liner
figure 10B: Adhere Patch to Site
19ZetrOZ Systems Product Support 1-800-202-9831
Page 20
11.5. TURN THE DEVICE ‘ON’
4
HRS
3
2
1
DONE
Power Button
A. Press and hold the Power Button on the
sam Power Controller for at least 1 second
to turn the device ‘ON’ (Figure 11A).
i. As confirmation that the device is ‘ON’,
the LEDs on the sam Power Controller will illuminate blue (treatment timer and indicator LEDs) and the indicator LED on the applicator will illuminate blue.
ii. The battery indicator LEDs will remain
‘ON’ for about 4 seconds after turning ‘ON’ the device.
iii. The treatment timer LEDs and blue
indicator LEDs (power controller and applicator) will remain ‘ON’ for the duration of treatment.
B. Press the Toggle Button to select 1, 2, 3 or
4 hour treatment duration (as allowed by current battery charge) (Figure 11B).
i. If the Toggle Button is pressed up to
the 4 hour setting and a decreased treatment time is desired, pressing it again will return the treatment timer to the 1 hour setting.
figure 11B: Press Power Button
Toggle Button
figure 11B: Select Duration
20 Product Support 1-888-202-9831
Page 21
11.6. TREATMENT DELIVERY
Treatment will be delivered for the duration set by the user. The treatment timer LEDs will count down in hour increments as treatment progresses. Upon completion, the device will automatically shut off and the ‘treatment complete’ LED will illuminate for approximately 8 hours before turning off. See section 12 for how to manually power ‘OFF’ the device.
Warning: ALWAYS instruct the patient to alert the practitioner in the event of discomfort
during treatment.
Warning: As an additional safety measure, ALWAYS instruct the patient how to turn off
the sam Device and remove the sam Applicator in the event of discomfort or emergency.
Warning: ALWAYS keep the sam Power Controller within easy reach so that power may
be ceased at any time.
21ZetrOZ Systems Product Support 1-800-202-9831
Page 22
12. Device Power Down and Removal
MODEL
AT-12
Pull Tab
12.1. POWER OFF THE POWER CONTROLLER
Press and hold the Power Button on the sam Power Controller for at least 1 second to turn the device ‘OFF’ (See Figure 11A for device diagram).
12.2. REMOVE DEVICE FROM SKIN
A. Remove the Patch(s) from the skin.
B. Remove the applicator from the sam Ultrasound Coupling Patch by pulling the
tab on the side of the gel cup. This unlocks the cup from the applicator (See Figure 12).
C. Throw away the sam Patches.
Warning: NEVER use a sam Ultrasound Coupling Patch for more than one use.
Misuse of sam Ultrasound Coupling Patches or use for more than the intended treatment duration MAY RESULT IN BURN OR REPEATED SHUTOFF of the applicator.
12.3. CLEAN THE APPLICATORS
Clean any residual ultrasound coupling media off of the applicator and off of the skin. See section 14 for ‘Cleaning and Maintenance’ details.
12.4. RECHARGE THE DEVICE
Connect the sam Power Controller to the sam Electrical Charger for recharging. Prior to the next use, allow up to 6 hours to recharge the battery to 100% charge. See section 9 for charging instructions.
figure 12: Removing Applicator
22 Product Support 1-888-202-9831
Page 23
13. Modes of Operation
13.1. ON MODE
To turn the device ‘ON’ press and hold the Power Button for at least 1 second. Press the Toggle Button on the right side of the sam Power Controller to set the treatment duration. The treatment duration may be set for 1, 2, 3 or 4 hours. The battery indicator LED and treatment timer LED will be displayed on the control panel (see section 8: LED Display). The applicator indicator LED will be displayed on each attached applicator (See Figure 2, page 10). Each connected applicator will emit ultrasound energy immediately upon the sam Power Controller entering the ‘ON’ mode of operation. A powered sam Applicator will always emit ultrasonic energy unless it enters the sam Sensing Mode or the treatment duration concludes.
13.2. OFF MODE
When the device is ‘OFF’, the LEDs on the applicator will not be illuminated, the LEDs on the sam Power Controller will not be illuminated, and no ultrasound energy will emanate from the device. Always turn the device ‘OFF’ when not in use or if pain or uncomfortable heating is indicated by the patient.
13.3. SAM SENSING MODE
In the event that the treatment site underneath the applicator reaches the temperature threshold, the applicator will pause ultrasound output and vibrate once with a red LED notification to signify that the device has begun a cooling or rest cycle. sam Applicator will automatically resume treatment with a blue LED notification after the site has cooled.
Warning: Should the patient ever report discomfort or a painful sensation from the site
under the sam Applicator, the applicator should be removed immediately.
Caution: At no time during treatment should the applicator be covered by thick
insulating material such as a coat, blanket or sports wrap. This may cause
sam to disable and remain disabled throughout the therapy session.
13.4. END OF TREATMENT MODE
At the end of treatment, an amber-colored “Treatment Complete” LED will illuminate on the sam Power Controller and all other displays (battery indicator, treatment timer, indicator LED, and applicator LEDs) will be inactive. All ultrasound emissions from the applicator will cease. The “Treatment Complete” LED will remain illuminated for approximately 8 hours after completion of therapy, at which point it will time out and turn off. To set the sam Device for a new treatment, assess whether the sam Device requires recharging.
23ZetrOZ Systems Product Support 1-800-202-9831
Page 24
14. Cleaning and Maintenance
The exterior of the sam Power Controller and the applicator surfaces may be cleaned with a soft cloth, tissue, or towel and one of the following cleaning agents: mild detergent and water or disinfecting medical wipes.
Caution: Properly clean the applicator between treatments.
Caution: The device is not waterproof. Do not apply a direct stream of liquid onto the
device, submerge the device, or allow any liquid to pool on the surface of the device.
Caution: DO NOT USE: Phenolic-based disinfectants, quaternary ammonium, chlorine-
based disinfectants, solvent-based cleaners, or abrasive materials. Doing so may damage the plastic housing and void the warranty.
24 Product Support 1-888-202-9831
Page 25
15. Storage & Operating Conditions
15.1. STORAGE
Store the sam Device in the following conditions:
Temperature: 557°C;
Humidity: 1080%;
Atmospheric pressure range: 7001060 hPa
Store the sam Ultrasound Coupling Patches in the following conditions:
Temperature: 530°C;
Humidity: 1080%;
Atmospheric pressure range: 7001060 hPa
15.2. OPERATION
Only operate the sam Device in the following conditions:
Temperature: 144°C;
Humidity: 1080%;
Atmospheric pressure range: 7001060 hPa
Caution: Do not keep the device in extreme hot or cold temperatures (above 50°C or
below 0°C). Do not leave the sam Device in a hot or freezing car. Do not leave the device in direct sunlight for extended periods. UV light may damage or discolor the device.
16. Disposal of Waste Products
sam Ultrasound Coupling Patches are one time use and may be disposed of in regular
sanitation trash. No special disposal procedures are necessary.
Old, damaged or expired sam Power Controllers and applicators should be recycled or returned to the manufacturer for proper disposal.
25ZetrOZ Systems Product Support 1-800-202-9831
Page 26
17. Appendix
Li-ion
LOT
A. SYMBOLS
Consult User Manual/Instructions for Use
Manufacturer/Date of Manufacture
Lithium-ion battery inside
Class BF Applied Part
Do not use if package is damaged
Do not reuse
Non-ionizing radiation
Separate collection for electrical and electronic equipment.
be disposed of in unsorted municipal waste.
. Caution, consult accompanying documents
Diverging beam
Continuous wave (CW)
Use only with
W
Watts, ultrasonic power
MHz
Frequency in megahertz
ERA
Effective radiating area
BNR
Beam non uniformity ratio
Must not
Use by date
Temperature limit
Humidity limitation
SN
Serial number
Catalogue/Reorder Number
Lot Number
Atmospheric Pressure limitation
VO
Output Voltage
P
Acoustic Power
Ie
Average Intensity
f
awf
Frequency
26 Product Support 1-888-202-9831
Page 27
B. SPECIFICATIONS
The sam Device does not contain microprocessor or software to control function
Maximum Acoustic 0.65 W +20% per transducer
Power Output 1.3 W +20% for 2 transducers
Maximum Intensity 0.132 W/cm2 +20%
Frequency 3 MHz +20%
Duty Cycle 100% - continuous wave
Beam Form Wide Beam – 5 degree diverging lens
Individual Transducer Dimension 5 cm2 emitting surface area (circular)
BNR <5:1
2
ERA 6 cm
Maximum Treatment Duration 4 hours
Other Electrical Ratings for the sam Device Components
Electrical Charger (Model SINGOF10U050200)
Input Voltage 100240 V, 5060 Hz
Input Current 0.120.3 A
Output Voltage 4.755.25 V DC
Output Current 2.0 A
Max Output Power 10.5 W
+
20 %
Power Controller
Output Voltage 3.7 V DC +10%
Max Output Amperes 700 mA
Battery Protection Max Current: 2 Amps
Max Voltage: 4.2 V
Min Voltage: 3.1 V
Cable Voltage Rating 300 V
Cable 20°C Resistance: 94/km
27ZetrOZ Systems Product Support 1-800-202-9831
Page 28
Applicator
Input Voltage: 3.7 V DC +10%
Ultrasound Frequency: 3 MHz +20%
Ultrasonic Output Power: 1 Transducer: 0.65 W +20%
2 Transducers Together: 1.3 W +20%
Max Input Current: 400 mA
YAdapter
Voltage Rating 300 V
20°C Resistance 94/km
C. NOTICE OF COMPLIANCE, CALIBRATION AND
OPERATIONAL PERIOD
The sam Device meets performance standards under 21 C.F.R. § 1050.10 ­PERFORMANCE STANDARDS FOR SONIC, INFRASONIC, AND ULTRASONIC RADIATION­EMITTING PRODUCTS. The sam Device meets IEC 606011, 3 ed. (2012); IEC 60601­12, 3 ed. (2007); Medical Electrical Equipment, General Requirements for Safety; Electromagnetic Compatibility. The sam Device is calibrated to provide 3 MHz +20% ultrasound diathermy at 0.65 W +20% per ultrasound applicator.
Operational Period: The sam Power Controller is intended to maintain calibration for up to 300 charging cycles and each ultrasound applicator is intended to withstand 1500 hours of run time, after which the system should be recycled or returned to the manufacturer for proper disposal.
28 Product Support 1-888-202-9831
Page 29
D. TECHNICAL INFORMATION
figure A. Ultrasound Field Scan Across Applicator Face
figure B. Ultrasound Field Scan Away From the Applicator Face
29ZetrOZ Systems Product Support 1-800-202-9831
Page 30
E. ELECTRICAL IMMUNITY AND EMISSIONS
The following components of the sam Ultrasonic Diathermy Device are
compliant with the requirements of IEC 6060112 ed 3.0 (200703):
sam Model 271
Cable Length
AT271: Ultrasound Applicators
PC271: Power Controller 48 inches +1.2
CT271: Electrical Charger 70.8 inches +3.8
YB271: Y-branch Adapter 8 inches +.6 x 2
Guidance and Manufacturer’s Declaration – Emissions
Emissions Test Compliance Electromagnetic Environment – Guidance
RF Emissions Group 1 The sam uses RF energy only for its internal
CISPR 11 function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions Class B The sam is suitable for use in all establishments,
CISPR 11 including domestic, and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonics Class A
IEC 6100032
Flicker
IEC 6100033
30 Product Support 1-888-202-9831
Page 31
Guidance and Manufacturer’s Declaration – Immunity All
ME Equipment and ME Systems
The sam is intended for use in the electromagnetic environment
specified below. The customer or user of the sam should ensure
that it is used in such an environment.
Immunity Test IEC 60601 Compliance Electromagnetic
Test Level Level Environment - Guidance
ESD
IEC 6100042 +8kV Air
EFT 2kV Mains
IEC 61000044 +1kV IOs
or hospital environment.
Surge
IEC 6100045 +2kV Common Common NA that of a typical commercial or
(no ground) hospital environment.
Voltage >95% Dip for >95% Dip for Mains power quality should be
Dips/Dropout 0.5 Cycle 0.5 Cycle that of a typical commercial or
IEC 61000411 60% Dip for 60% Dip hospital environment. If the user
5 Cycles 5 Cycles of the sam requires continued
30% Dip for 30% Dip for operation during power mains
25 Cycles 25 Cycles interruptions, it is recommended
>95% Dip for >95% Dip for that the sam be powered from
5 Seconds 5 Seconds an uninterruptible power supply
or battery.
Power 3A/m 3A/m Power frequency magnetic fields
Frequency should be that of a typical
5060Hz commercial or hospital
Magnetic Field environment.
IEC 6100048
+
6kV Contact +6kV Contact Floors should be wood, concrete
+
8kV Air or ceramic tile. If floors are
synthetic, the r/h should be at
least 30%.
+
2kV Mains Mains power quality should be
+
1kV IOs that of a typical commercial
+
1kV Differential +1kV Differential Mains power quality should be
31ZetrOZ Systems Product Support 1-800-202-9831
Page 32
Guidance and Manufacturer’s Declaration – Immunity
The sam is intended for use in the electromagnetic environment
specified below. The customer or user of the sam should ensure
that it is used in such an environment.
Immunity Test IEC 60601 Compliance Electromagnetic
Test Level Level Environment - Guidance
Portable and mobile communications
equipment should be separated from
the sam by no less than the distances
calculated/listed below:
Conducted RF 3 Vrms (V1) = 3 Vrms D = (3.5V1)(Sqrt P)
IEC 6100046 150 kHz to 150 kHz to 80 MHz
80 MHz
Radiated RF 3V/m D = (3.5Eq)(Sqrt P)
IEC 6100043 80 MHz to (E1) = 3 V/m 80 to 800 MHz
2.5 GHz
D=7E1)(Sqet P)
800 MHz to 2.5 GHz
Where P is the max power in watts and D
is the recommended separation distance
in meters.
Field strength from fixed transmitters,
as determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
32 Product Support 1-888-202-9831
Page 33
Recommended Separation Distances between portable and
mobile RF Communications equipment and the sam ME Equipment and
ME Systems that are NOT Life-supporting
The sam is intended for use in the electromagnetic environment in which radiated
disturbances are controlled. The customer or user of the sam can help prevent
electromagnetic interference by maintaining a minimum distance between portable
and mobile RF Communications Equipment and the sam as recommended below,
according to the maximum output power of the communications equipment.
Max Output Separation (m) Separation (m) Separation (m)
Power (Watts) 150kHz to 80MHz 80 to 800MHz 800MHz to 2.5GHz
D=3.5V1)(Sqrt P) D=3.5E1)(Sqrt P) D=3.5E1)(Sqrt P)
0.01 0.11667 0.11667 0.23333
0.1 0.36894 0.36894 0.73785
1.0 1.1667 1.1667 2.3333
10.0 3.6894 3.6894 7.3785
100.0 11.667 11.667 23.333
F. WARRANTY
ZetrOZ Systems offers a 1-year manufacturer’s warranty for the sam Device. If the
sam Device fails due to defects in material or workmanship, ZetrOZ Systems, at its
discretion, will:
1. REPAIR the sam Device OR
2. REPLACE the sam Device with another sam Device
THIS LIMITED WARRANTY AND ANY IMPLIED WARRANTIES THAT MAY EXIST UNDER STATE LAW APPLY ONLY TO THE ORIGINAL PURCHASER OF THE sam DEVICE AND ARE NONTRANSFERABLE.
Extent of Limited Warranty
This limited warranty does not cover damages due to external causes, including, without limitation, accident, usage not in accordance with product instructions, misuse, neglect, alteration or repair.
33ZetrOZ Systems Product Support 1-800-202-9831
Page 34
G. TROUBLESHOOTING
Question or Problem Solution
1. How to determine if As confirmation that the device is delivering ultrasound,
the device is delivering the LEDs on the Power Controller will illuminate blue treatment (treatment timer and indicator LEDs) and the indicator LED
on the applicator will illuminate blue. See section 8 on LED Display.
2. The indicator LED on Ensure that the applicator is fully connected to the power
the Applicator does not controller wire jack and/or the Y-adapter wire jack. If illuminate when the recently in sam Sensing Mode, ensure that the applicator device is turned ON has had time to reach normal operating temperature and
the indicator LED on the applicator has turned from red back to blue. If the applicator LED still fails to illuminate blue, contact the manufacturer with serial and model number details.
3. How to observe the The Battery Indicator display can only be viewed when
Battery level during the device is powered ON or OFF by pressing the Power treatment Button, or when the Toggle Button is pressed. (See section 1 1.1) The lights remain illuminated for approximately 4
seconds after the device is turned ON or OFF. The Battery Indicator LEDs do not remain illuminated during treatment. The treatment duration is indicated by the blue treatment timer LEDs (See section 8 on LED display).
34 Product Support 1-888-202-9831
Page 35
G. TROUBLESHOOTING
Question or Problem Solution
4. The Applicator is Turn off the device. Remove the sam Ultrasound Coupling
not staying secured Patch from the applicator. Replace with new sam during treatment Ultrasound Coupling Patch. See section 11.3 for gel cup
application instructions.
5. The Applicators are The device has entered ‘sam Sensing Mode’ (See section
vibrating and red 13.4) Each sam Applicator is equipped with closed-loop
continuous temperature monitoring which maintains treatment site temperatures below 44°C during normal operation. In the event that the treatment site underneath the applicator reaches the temperature threshold, the applicator will pause ultrasound output and vibrate once with a red LED notification that the device has begun a cooling or rest cycle. sam will automatically resume treatment with a blue LED notification after the site has cooled. Caution: At no time during treatment should the applicator be covered by thick insulating material such as a coat, blanket or sports wrap. This may cause sam to disable and remain disabled throughout the therapy session.
35ZetrOZ Systems Product Support 1-800-202-9831
Page 36
sam
0086
ZetrOZ Systems, LLC
56 Quarry Road
Trumbull, CT 0661 1
USA
MDSS GmbH
Schiffgraben 41
30175 Hannover, Germany
Toll Free Tel 18882029831
sam User Manual (OM271) — LL452600 Rev. A
Published 04042017
Loading...