Salutron HardyTest D700 User Manual

Portable Hardness Tester
H
H
arr
a
d
d
y
T
e
y
T
e
stt
s
D
D
7
7
0
0
0
0
®
®
User´s Manual HardyTest D700®
3
Content
1 OVERVIEW......................................................................................................................................... 5
1.1 ADVANTAGES .................................................................................................................................. 5
1.2 MAIN APPLICATION &TESTING RANGE ........................................................................................... 5
1.2.1 Main Application ................................................................................................................... 5
1.2.2 Testing Range ...................................................................................................................... 5
1.3 CONFIGURATION .............................................................................................................................. 6
1.4 WORKING CONDITIONS ................................................................................................................... 6
1.5 SAFETY INSTRUCTIONS .................................................................................................................... 6
2 STRUCTURE FEATURE &TESTING PRINCIPLE ...................................................................... 7
2.1 STRUCTURE FEAT UR E ...................................................................................................................... 7
2.1.1 The Hardness Tester Appearance .................................................................................... 7
2.1.2 Parts of the Main body ........................................................................................................ 7
2.1.3 D Type Impact Device ......................................................................................................... 8
2.1.4 Different Types of Impact Device....................................................................................... 8
2.2 LEEB HARDNESS TESTING PRINCIPLE .............................................................................................. 8
3 TECHNICAL SPECIFICATIONS ..................................................................................................... 9
4 PREPARATION & TESTING.......................................................................................................... 10
4.1 PREPARATION & INSPECTION BEFORE TESTING ............................................................................. 10
4.1.1 Preparation of Sample Surface ....................................................................................... 10
4.1.2 System Setting ................................................................................................................... 10
4.1.3 Presetting Testing condition ............................................................................................. 10
4.2 TESTING PROGRAM ....................................................................................................................... 11
4.2.1 Start-Up ............................................................................................................................... 11
4.2.2 Loading ............................................................................................................................... 11
4.2.3 Localization ......................................................................................................................... 11
4.2.4 Testing ................................................................................................................................. 11
4.2.5 Read measured value ....................................................................................................... 12
4.2.6 Power Off ............................................................................................................................ 12
5 ADVICE ............................................................................................................................................. 12
6 OPERATION IN DETAILS .............................................................................................................. 13
6.1 POWER ON .................................................................................................................................... 13
6.2 Power Off ............................................................................................................................... 13
6.3 Testing .................................................................................................................................... 13
6.3.1 Instruction of the Main Display Interface ........................................................................ 13
6.3.2 Testing Operation at the Main Display Interface ........................................................... 14
6.3.3 Key Operation at the Main Display Interface ................................................................. 14
6.4 MENU STRUCTURE ........................................................................................................................ 14
6.5 TEST SET ....................................................................................................................................... 15
6.5.1 Impact Direction Setting.................................................................................................... 15
User´s Manual HardyTest D700®
4
6.5.2 Average Times Setting ...................................................................................................... 16
6.5.3 Material Setting .................................................................................................................. 16
6.5.4 Hardness Scale Setting .................................................................................................... 17
6.5.5 Tolerance Limit Setting ..................................................................................................... 17
6.5.6 Hardness/бb Setting .......................................................................................................... 17
6.6 PRINT FUNCTION ........................................................................................................................... 18
6.6.1 Print Current ....................................................................................................................... 18
6.6.2 Print Memory ...................................................................................................................... 18
6.6.3 Print All Memory ................................................................................................................. 18
6.6.4 Paper Feeding .................................................................................................................... 18
6.7 MEM ORY MANAGER ...................................................................................................................... 18
6.7.1 View from No.1 Group/View from Ending Group .......................................................... 19
6.7.2 View from Selected No. Group ........................................................................................ 19
6.7.3 Data Transfer ...................................................................................................................... 19
6.7.4 Delete by Group No. ......................................................................................................... 19
6.7.5 Delete All Data ................................................................................................................... 20
6.7.6 Deletion Confirmation ....................................................................................................... 20
6.8 BROWSING MEM ORY DATA GROUPS .............................................................................................. 20
6.9 SYSTEM SET .................................................................................................................................. 21
6.9.1 LCD Brightness Set ........................................................................................................... 22
6.9.2 Time Date Set .................................................................................................................... 22
6.10 SOFTWAR E INFORMATION ............................................................................................................ 22
6.11 SYSTEM CALIBRATION ................................................................................................................. 22
6.12 EL BACKGROUND LIGHT ............................................................................................................. 23
6.13 AUTO POWER OFF ....................................................................................................................... 23
6.14 PAPER LOADING .......................................................................................................................... 23
6.15 BATTERY CHARGE ....................................................................................................................... 24
6.16 BATTERY REPLACEMENT ............................................................................................................. 24
6.17 CONNECTION OF DATA TRANSMISSION CABLE ............................................................................ 24
7 FAULT ANALYSIS & EVA C UATION ............................................................................................ 24
8 SERVICING & MAINTENANCE ................................................................................................... 25
8.1 IMPAC T DEVICE SERVIC ING ........................................................................................................... 25
8.2 NOR MAL MAINTENANCE PROGRAM .............................................................................................. 25
9 CALIBRATION TIME ..................................................................................................................... 25
10 NOTICE OF TRANSPORT AND STORAGE CONDITIONS ................................................... 25
APPENDIX ........................................................................................................................................... 26
TABLE 1 HARDNESS VALUES MATERIALS ............................................................................................ 26
TABLE 2 MAT E R IA L STRENGTH ........................................................................................................... 27
TABLE 3 TECHNICAL DETAILS IM PA C T DEVICES .................................................................................. 28
TABLE 4 SUPPORTING RINGS ............................................................................................................... 29
TABLE 5 SPECIFICATIONS .................................................................................................................... 30
User´s Manual HardyTest D700®
5
1 Overview
1.1 Advantages
Wide measuring range. Based on the principle of Leeb hardness testing theory. It
can measure the Leeb hardness of all metallic materials.
Large screen (128×64 dot matrix LCD), showing all functions and parameters.
Test at any angle, even upside down.
Direct display of hardness scales HRB, HRC, HRA, HV, HB, HS, HL.
Seven impact devices are available for special application. Automatically identify
the type of impact devices.
Large capacity memory could store 500 groups (Relative to average times 32~1 )
information including single measured value, mean value, testing date, impact
direction, impact times, material and hardness scale etc.
Upper and lower limit can be preset. It will alarm automatically when the result
value exceeding the limit.
Battery information indicates the rest capacity of the battery and the charge status.
User calibration function.
Software to connect with PC via USB port.
With EL background light.
Thermal printer integrated, convenient for in field printing.
NI-MH rechargeable battery as the power source. Charge circuit integrated insid
t
he instrument. Continuous working period of no less than 150 hours (EL off
inting).
no pr
Auto power off to save energy. Outline dimensions212mm×80mm×32mm
1.2 Main Application &Testing Range
1.2.1 Main Application
Die cavity of molds
Bearings and other parts
Failure analysis of pressure vessel, steam generator and other equipment
Heavy work piece
The installed machinery and permanently assembled parts
Testing surface of a small hollow space
Material identification in the warehouse of metallic materials
Rapid testing in large range and multi-measuring areas for large-scale work piece
e
and
1.2.2 Testing Range
Testing range refer to Table 1 and Table 2 in the Appendix.
6
1.3 Configuration
No.
Item
Quantity
Remarks
Standard
1
Main Unit
1 2
D type impact device
1
With cable
3
Standard test block
1 4
Cleaning brush (I)
1 5
Small support ring
1 6
Battery Charger
1
9V 500mA
7
Paper for printing
1 8
Manual
1 9
Instrument case
1
Optional
11
Cleaning brush (II)
1
For use with G type impact device
12
Other type of impact
Refer to Table 3 and
appendix.
13
DataPro software
1 14
Communication cable
1 15 16
Table 1-1
Configuration
Configuration
User´s Manual HardyTest D700®
devices and support rings
1.4 Working Conditions
Working temperature -10°C~+50°C;
Storage temperature:-30°C~+60°C;
Relative humidity: ≤ 90;
The surrounding environment should avoid of vibration, strong magnetic field,
corrosive medium and heavy dust.
1.5 Safety Instructions
The instrument can only work with the specially designed battery pack and power
adapter (charger) supplied by us. Working with others may result in damage of the
instrument, battery leakage, fire or even explosion.
Do not cast the battery pack into fire and do not short circuit, disassemble or heat
the battery pack, otherwise battery leakage, fire or even explosion may occur.
Do not open the cover of the paper compartment or come into contact with the
heating head of the printer by hand or any part of your body to avoid burns due to
high temperature when the printer is printing.
Table 4 in the
User´s Manual HardyTest D700®
7
DEL
SAVE
ESC
MENU
ENTER
MAT
DIR
FEED
PRT
1
2
HD
SaluTron
SCALE
DEL
SAVE
ESC
MENU
ENTER
MAT
DIR
FEED
PRT
11
10
9
8
7
6
5
4
3
2
1
OFF---ON
HL0108070811
MODEL: HardyTest D700
BATTERY
: 6V NI-MH
Portable Hardness Tester
CAUTION:
1 Do not disassemble
3 Use only supplied AC-DC adapter for charging
CHARGER: 9V DC
2 Do not dispose of in fire or water
SaluTron
SaluTron
2 Structure Feature &Testing Principle
2.1 Structure Feature
2.1.1 The Hardness Tester Appearance
1. Main unit
2. Impact device
2.1.2 Parts of the Main body
1 Socket of impact device 2 Socket of USB 3 LOGO 4 Paper compartment cover 5
LCD display 6 Keypad 7 Battery switch 8 Power jack 9 Serial Number 10 Product
label 11 Battery compartment cover
User´s Manual HardyTest D700®
8
2.1.3 D Type Impact Device
1 2 3 4 5 6 7
1 Release button 2 Loading tube 3 Guide tube 4 Coil unit
5 Connection cable 6 Impact body 7 Support ring
2.1.4 Different Types of Impact Device
D DC DL C D+15 E G
2.2 Leeb Hardness Testing Principle
The basic principle is: use an impact body of certain weight impacts against the testing
surface under certain test force, then measure the impacting velocity and the rebounding
velocity of the impact body respectively when the spherically test tip is located 1mm above the
testing surface.
The calculation formula is as follows: HL= VB / VA
HL—— Leeb hardness value, VB—— Rebounding velocity of the impact bodyand VA——
Impacting velocity of the impact body
User´s Manual HardyTest D700®
9
Type of impact
device
Hardness value of Leeb
standard hardness block
Error of displayed
value
Measuring range:HLD(170~960)HLD
Communication interface:USB1.1
3 Technical Specifications
Error and repeatability of displayed value see Table 3-1.
Table 3-1
No.
1 D
2 DC
3 DL
4 D+15
5 G
6 E
7 C
760±30HLD 530±40HLD
760±30HLDC 530±40HLDC
878±30HLDL 736±40HLDL
766±30HLD+15 544±40HLD+15
590±40HLG 500±40HLG
725±30HLE 508±40HLE
822±30HLC 590±40HLC
Repeatability
±6 HLD
±10 HLD
±6 HLDC
±10 HLDC
±12 HLDL 12 HLDL
±12 HLD+15 12 HLD+15
±12 HLG 12 HLG
±12 HLE 12 HLE
±12 HLC 12 HLC
6 HLD
10 HLD
6 HLD
10 HLD
Measuring direction360° Hardness Scale:HL, HB, HRB, HRC, HRA, HV, HS Displaydot matrix LCD, 128×64 dots Data memory500 groups max.relative to impact times 32~1) Printing paper: width is (57.5 ± 0.5mm, diameter is 30mm Battery pack: 6V NI-MH Battery charger: 9V/500mA Continuous working periodabout 150 hours(With backlight off, no printing
User´s Manual HardyTest D700®
10
4 Preparation & Testing
4.1 Preparation & Inspection before Testing
4.1.1 Preparation of Sample Surface
Preparation for sample surface should conform to the relative requirement in the Appendix
Table 3.
In the preparation processing for sample surface, the hardness effect of being
heated or cold processing on the surface of sample should be avoided.
Too big roughness of the being measured surface could cause measure error. So,
the surface of the sample to be measured must appear metallic luster, smoothing
and polish, without oil stain.
Support of test sample. Support is no necessary for heavy sample.
Medium-weight parts must be set on the smoothing and stable plane. The sample
must set absolutely equability and without any wobble.
Curved surface: The best testing surface of sample is flat. When the curvature
radius R of the surface to be tested is smaller than 30mm (D, DC, D+15,C, E and
DL type of impact device) and smaller than 50mm (G type of impact device), the
small support ring or the shaped support rings should be chosen.
The sample should have enough thickness, minimum thickness of sample should
conform to Table 3.
For the sample with hardened layer on surface, the depth of hardened layer
should conform to Table 3.
Coupling. Light-weight sample must be firmly coupled with a heavy base plate.
Both coupled surface must be flat and smooth, and there is no redundant coupling
agent existing. The impact direction must be vertical to the coupled surface. When
the sample is a big plate, long rod or bending piece, it can be deformed and
become unstable, even though its weight and thickness is big enough, and
accordingly, the test value may not be accurate. So the sample should be
reinforced or supported at its back.
Magnetism of the sample itself should be avoided.
4.1.2 System Setting
See 6.9 for details.
4.1.3 Presetting Testing condition
See 6.5 for details.
User´s Manual HardyTest D700®
11
Use a calibrated hardness tester, test the standard test block
4.2 Testing Program
Verification of the tester is by using standard test block. The error and repeatability of
displayed value should be within the regulation of Appendix table 2.
Note
downward vertically for 5 times, the arithmetical average value compare with the value of standard test block. If this value exceeds the standard value, could use the function of software calibration to adjusting.
4.2.1 Start-Up
Insert the plug of the impact device into the socket of impact device on the tester.
Press [
4.2.2 Loading
Pushing the loading-tube downwards until contact is felt. Then allow it to slowly return to the
starting position or using other method locking the impact body.
4.2.3 Localization
Press the impact device supporting ring on the surface of the sample firmly, the impact
direction should be vertical to the testing surface.
4.2.4 Testing
Press the release button on the upside of the impact device to test. The sample
Each measure area of the sample usually need 5 times of testing operation. The
The distance between any two impact points or from the center of any impact
If want accurate conversion from the Leeb hardness value to other hardness value,
]keynow power is connected. The instrument is in testing condition.
and the impact device as well as the operator are all required to be stable now.
The action direction should pass the axis of the impact device.
result data dispersion should not more than mean value±15HL.
point to the edge of testing sample should conform to the regulation of Table 4-1.
contrastive test is needed to get conversion relations for the special material. Use
inspection qualified Leeb hardness tester and corresponding hardness tester to
test at the same sample respectively. For each hardness value, each measure
homogeneously 5 points of Leeb hardness value in the surrounding of more than
User´s Manual HardyTest D700®
12
three indentations which need conversion hardness, using Leeb hardness
arithmetic average value and corresponding hardness average value as
correlative value respectively, make individual hardness contrastive curve.
Contrastive curve at least should include three group of correlative data.
Table 4-1
Type of Impact
Device
D, DC 3 5
DL 3 5
D+15 3 5
G 4 8
E 3 5
C 2 4
4.2.5 Read measured value
4.2.6 Power Off
Distance of center of the two
Distance of center of the
indentations
Not less than (mm) Not less than (mm)
indentation to sample edge
Press [
5 Advice
Replacing the impact device must be done during Power off. Otherwise the main
body can not identify the type of the impact device, and it can damage the circuit
board of the main body.
You could not save the current test value if the test times are less than the
presetting times value.
Only type D and type DC of impact device have the function of strength measure
option. You can not modify the [Set hardness orбb ]setting when using other
types of impact device. The [Set hardness orбb ]setting would be set to
[Hardness ] automatically after replacing the impact device whether the setting
is [Hardness ] or not before.
Not all materials could convert to all hardness style value. The hardness style is
reset to HL automatically after changing the material. So select material first
before changing the hardness style.
]key to turn off the instrument.
User´s Manual HardyTest D700®
13
Times count
Probe Type:D
6 Operation in Details
6.1 Power On
Press [ ] to power on the system. The screen shows as below:
The system would automatically detect the type of the impact device during power up, and
would display this information on the screen. Users should pay attention to the probe type
displayed on the screen. After pausing for several second, the screen will exit and enter the
main display interface.
Note: If the instrument is in power off condition, it will turns on automatically after the
charge power is connected.
6.2 Power Off
HARDNESS TESTER
D700
Press key [
Note: If the charge power is connected, the instrument will turns on automatically
after pressing the power off key.
6.3 Testing
Below is the main display interface:
Measured value
Material
6.3.1 Instruction of the Main Display Interface
Battery information
charging status.
Impact direction
Average value indicator
] could power off the system in any conditions.
Hardness scale Average value indicator
Display the information of the rest capacity of the battery, and the
The present impact direction.
It appears to show the mean value of the samples when reaching
Battery information
Impact direction
14
the presetting impact times.
--------------
--------------
YES
Hardness scale
Hardness method of the present measured value.
User´s Manual HardyTest D700®
Measured value
or display the present mean value (with average value indicator prompting).
conversion or measure range.
Material
Impact times count
6.3.2 Testing Operation at the Main Display Interface
Testing operation could be carried out under this interface. After each impact operation, it can
display the current measured value, impact times count plus one, the buzzer would alert a
long howl if the measured value is not within the tolerance limit. When reaching the presetting
impact times, the buzzer will alert a long howl. After 2 seconds, the buzzer will alert a short
howl, and display the mean value.
6.3.3 Key Operation at the Main Display Interface
Press key [SAVE]to store present group of measured value into memory. This operation
Press key [DEL]to delete the latest single measured value. After pressing this key, the
is only valid after displaying the mean value.
screen will displays as below:
Display present single time measured value(without mean value indicator),
means lower than conversion or measure range.
The present presetting material.
Times that have been impacted.
means over
Press key [] or [] to move the
Confirm delete
Press key [] or [] could display single measured value.
Press key [ ]could switch on of off the background light of LCD.
Press key [MENU ]could enter the system presetting menu.
Press key [DIR ]to set the impact direction.
Press key [CNT ]to change the impact times in one group. The impact times
count item will be highlighted when first pressing the key [CNT ], and the
impact times count value will plus one with each pressing. The value will roll back
to 1 when it reaches 32.
Press key [SCALE ]to change the hardness scale.
Press key [MAT ]to change the material set. Presetting hardness scale recovers to
HL automatically after material presetting changed.
6.4 Menu Structure
Both presetting system parameters and the additional function could come true by menu
operation. At the main display interface, press key [MENU] into the main menu.
NO
cursor to [YES ]or [NO ].
Press key [ENTER ] to confirm
operation. Press key [ESC ]to cancel
delete operation.
User´s Manual HardyTest D700®
15
Impact Direc.
Average
Material
Print Current
Print Memory
Print All Mem
View From No.1
View From End
View From No.
Transfer
Delete by No.
Delete All
Auto Save:
Auto Print:
Auto Delete:
Auto Trans.:
Software Info
Print Set Memory Manager System Set
Test Set
Impact Direction
---------------
Impact Direc. Average Material Hardness Scale Tolerance Limit Hard/
Test Set
Print Set
The Main Display Interface
6.5 Test Set
At the main display interface, press key [MENU ] to enter the main menu.
Memory Manager
System Set
Press key [ENTER ] to enter Test Set
Menu.
The symbol at the left side of underside
menu indicates that the menu has not
ended. Press key [
glance downward. The symbol ↑ at the left
side of the upside menu indicates that the
]could continuously
6.5.1 Impact Direction Setting
бb:Hard
menu has not ended. Press [
continuously glance upward.
Press key [] or [] to move the cursor
to the line you want to set, and press key
[ENTER ]to confirm it.
Note: 1. When [Hard/бb] is switched to
бb, the hardness scale could not be
selected. The cursor will skip over
[Hardness Scale] while moves the
cursor.
2. Only D type of impact device has the
function of бb measure. So the cursor
could not move to [Hard/бb ]while
use other impact device.
Press key [] or [] to move the cursor
to the impact direction that you will preset.
Press key [ENTER] to confirm it.
] could
User´s Manual HardyTest D700®
16
-------------
0
Copper Wrought Copper Wrought Steel
Copper-Zinc
Press key [ESC] to cancel it.
6.5.2 Average Times Setting
Press [] or [] to move the cursor.
Mean Times
3
6.5.3 Material Setting
When [Hard/бb ]is preset to hardness, it will display the following material: Steel and Cast
Steel, Cold Work Tool Steel, Stainless Steel, Gray Cast Iron, Nodular Cast Iron, Cast
Aluminum Alloys, Copper-Zinc Alloys, Copper-Aluminum Alloys, Wrought Copper and
Wrought Steel.
Press [
Press key [ENTER ]to confirm it. Press
key [ESC ]to cancel it.
] or [] to change the number.
Press key [] or [] to move the cursor
to the material you want to preset.
-Alumin
When [Hard/ бb ]is preset to бb, it will display the following material: Mild Steel,
High-Carbon Steel, Cr Ste el, Cr-V Steel, Cr-Ni Steel, Cr-Mo Steel, Cr-Ni-Mo Steel, Cr-Mn-Si
Steel, Super Strength Steel and Stainless Steel.
Press key [ENTER ]to confirm it.
Press key [ESC ]to cancel it.
Note 1. Presetting hardness scale
recovers to HL automatically after
material presetting is changed.
2. Please select material first, then
select hardness scale.
User´s Manual HardyTest D700®
17
Hard of Material
HV HB HRC
HS HRB HRA
HL
Tolerance Limit
Mild Steel High Cr Steel
Material Hardness Scale Tolerance Limit Hard/бb
Cr
-C Steel
-V Steel
6.5.4 Hardness Scale Setting
---------------
6.5.5 Tolerance Limit Setting
--------------­Min Max
0200 0890
6.5.6 Hardness/бb Setting
: Hard
Press key [] or [] to move the cursor
to the material to want to preset.
Press key [ENTER ]to confirm it.
Press key [ESC ]to cancel it.
Press key [] or [] to move the cursor
to the hardness scale you want to preset.
Press [ENTER ] to confirm setting.
Press [ESC ] to cancel setting.
Note
1. Here only displays the valid
hardness scale for the present selected
impact device and material. It would not
display the hardness scale which is not
valid.
2. Please select material first, then
select hardness scale.
3. Presetting hardness scale recovers
to HL automatically after presetting
material is changed.
Press [] or [] to move the cursor.
Press [
Press [ENTER ] to confirm setting.
Press [ESC ] to cancel setting.
Note:1. If the setting value exceeds the
measure range, the instrument will
remind you to reset.
2. If the bottom limit is larger than the
upper limit, they will exchange
automatically.
Press [ENTER ] to switch between
Hard and бb .
Note: Only D and DC type of impact
] or [] to change the number.
device has the function of бb
measure. So hard is the only
selection if the impact device is
not D or DC type.
User´s Manual HardyTest D700®
18
System Set
Memory Manager
Test Set Print Set
6.6 Print Function
At the main display interface, press [MENU ] to enter the main menu. [] or [] to
move the cursor to print menu and press [ENTER ]to enter the print menu.
Note:
Printing function is unavailable while charging.
Printing can be stopped by pressing the [ESC ]key.
Do not open the cover of the paper compartment during printing. Otherwise
the instrument may not print normally.
Over high ambient humidity (above 85% of relative humidity) or over low
ambient humidity (below 20% of relative humidity) may reduce the print
quality.
Printing with paper that has been stored for over long period of time or of
poor quality may reduce the print quality or even damage the printer.
6.6.1 Print Current
Print out the data report just finished testing. If the instrument hasn’t been switched off,
and hasn’t changed any testing condition during continuous printing process, it will only print
out single measured value and average value when printing again.
6.6.2 Print Memory
Print out the selected group of measured value stored inside the instrument.
6.6.3 Print All Memory
Print out all the measured value stored inside the instrument.
6.6.4 Paper Feeding
When the printer is powered on and ready for printing, press [FEED ] key then the
instrument will start manual paper feeding. Press and keep holding the [FEED ] key to
start paper feeding, while releasing the key to stop paper feeding.
Note: Manual paper feeding is unavailable while charging.
6.7 Memory Manager
At the main display interface, press key [MENU ] enter the main menu.
Press key [] or [] to move the cursor
to [Memory Manager ].Press key
[ENTER ]into [Memory
Manager ]menu.
If there is no data in the memory, displays:
<No Data!>. Then return.
19
View From No.1
View From End View From No. Transfer Delete By No. Delete All
6.7.1 View from No.1 Group/View from Ending Group
[ View from No.1 ] Start display values in the memory from the first group.
[View from End ] Start display values in the memory from the ending group.
6.7.2 View from Selected No. Group
Select Group
From 001 to 010
--------------
001
6.7.3 Data Transfer
[Transfer ]export the values stored in the memory as text format to PC through USB port.
This function is not available now.
6.7.4 Delete by Group No.
Select Group
From 001 to 010
--------------
001
Note: 1. If the preset group number exceeds the actual range, then deletes the actual
groups among them.
2. Do not shut down the instrument while deleting data. It could lead to unpredicted
consequence if shutting down while deleting.
User´s Manual HardyTest D700®
Press key [] or [] to move the cursor
to the function wanted, then press key
[ENTER ]to confirm.
Press [] or [] to move the cursor.
Press [
Press key [ENTER ] to start displaying
memory data from the selected beginning
group. Press key [ESC ]to cancel
current operation.
[ Delete by No. ]displays selecting the
range of deleting groups.
Press [] or [] to move the cursor.
Press [
Press key [ENTER ]to delete the
selected groups. Press key [ESC ]to
cancel operation.
] or [] to change the number.
] or [] to change the number.
User´s Manual HardyTest D700®
20
No.001 12/03 652HL
No.001 12/03 514HL
No.001 12/03/02
6.7.5 Delete All Data
[Delete All ]will delete all the data in the memory.
6.7.6 Deletion Confirmation
Press [] or [] to move the cursor to
---------------
[YES ] and press key [ENTER ]to
Confirm delete
---------------
YES NO
6.8 Browsing Memory Data Groups
No.002 12/03 587HL No.003 12/03 820HL No.004 12/03 693HL No.005 12/03 783HL No.006 12/03 782HL No.007 12/03 579HL No.008 12/03 687HL
No.002 12/03 785HL No.003 12/03 516HL No.004 12/03 789HL No.005 12/03 570HL No.006 12/03 852HL No.007 12/03 523HL No.008 12/03 796HL
confirm deleting operation.
Press [] or [] to move the cursor to
[NO ]and press key [ENTER ] to
cancel deleting operation.
Press key [ESC ]could cancel deleting
operation, no matter where the cursor is.
Press [] or [] to see previous or next
page.
Press key [ESC ]to exit browsing.
Press key [ENTER ], then press []
or [
] to move the cursor to the line which
you want to see details. Press
[ENTER ]to see details of that group.
Average= 514HL D 05 times Steel
Press [] or [] to browse details
including average value, test set and each
single value.
Press [ESC ] to return to previous display.
21
Test Set Print Set Memory Manager
Auto Save:Off Auto Print Auto Delete Auto Trans. Key Sound Warn. Sound LCD Brightness Time Date Set
511 513 516
514 515
6.9 System Set
At the main display interface, press key [MENU ] enter the main menu.
System Set
When [Auto Save ]is set to <On>, could store the data of current group automatically after
measuring and displaying average value.
When [Auto Print ]is set to <On>, could print the data of current group automatically after
measuring and displaying average value.
When [Auto Delete ]is set to <On>, according to 3σ rule, could cancel gross error
automatically after having measured presetting average times or pressing end in advance. If
there is data canceled, it needs supplemental measure to reach presetting times.
When [Auto Trans. ]is set to <On>could export the value of present group through
communication port after measuring and displaying average value. It’s not available now.
When [Key Sound ]is set to <On>, the buzzer would make a short hoot while press key
each time.
When [Warn. Sound ]is set to <On>, if the measured value exceeds the tolerance limit,
reached the presetting average times or deleting data, the buzzer would make a long hoot.
User´s Manual HardyTest D700®
Press key [] or [] to move the cursor
to [System Set ]Menu.
Press key [ENTER ]to enter [System
:Off
:Off :Off
:On
:On
Set ]menu.
Press key [] or [] to move the cursor
to the item wanted.
Press key [ENTER ]to modify the setting
directly or into corresponding screen.
Press key [ESC ]to exit.
[Auto Save ] [Auto Delete ] [Auto
Trans ] [Key Sound ] [Warn.
Sound ] could be switched on or off.
22
6.9.1 LCD Brightness Set
LCD brightness
Time Date Set
Print Set Memory Manager System Set
Software Info
D700
Version:1.00 Code:HL20000000 SN:HL2000000000
User´s Manual HardyTest D700®
Press key [] to enhance the brightness.
--------------­Bright:Press[ Dark:Press[ ]
6.9.2 Time Date Set
-------------­ 10/05/2008
6.10 Software Information
At the main display interface, press key [MENU ] enter the main menu.
6.11 System Calibration
The tester and impact device must be calibrated using hardness block before use as the first
time, or having not been used for a long time, or having reset the system.
Press key [
software calibration screen shows as below.
Press key [
Press key [ENTER ]to confirm the
modifying. Press key [ESC ]to cancel
] to weaken the brightness.
]
the modifying.
Present time and date is displayed
as“M/D/Y H/M”. Press the figure [
or [
] keys to modify the present figure.
Move the cursor by pressing the [
[
] keys. Press key [ENTER ]to confirm
modifying. Press key [ESC ]to cancel
11:02
], meanwhile pressing [ENTER ]to power on the system. Then the
modifying and exit.
Press figure [] or [] to move the
cursor to
Press key [ENTER ]into [Software
Info ]screen.
This screen displays the information about
the main body and the firmware. The
version, the Code and the SN would
change with the firmware.
[Software Info ].
]
] or
User´s Manual HardyTest D700®
23
Hold Both ends of the paper
compartment cover with moderate
, put the
attention to the paper direction. If the paper is misplaced, the instrument will
Pull a trip of paper out of the
Calibration
Calibration
780
--------------
0/5 times
--------------
Set the impact direction as [
Measure 5 points on the standard
hardness block.
It would display the average value after
measuring 5 times.
Press key figure [] or [] to input the
nominal value.
].
Average=780
Press key [ENTER ]to confirm.
Nominal=
6.12 EL Background Light
With the background light, it is convenient to work in the dark condition. Press key [
switch on or switch off the background light at any moment as you like after power on.
6.13 Auto Power Off
The instrument has the function of powering off automatically to save power.  The system would power down automatically if there’s neither measuring nor any key
Press key [ESC ]to cancel this
operation.
Range of adjustment: ±15HL.
]to
operation within 5 minutes. Except key [
screen and stop the operation of power off at the moment.
While the voltage of the battery is too low, the screen will show < Battery Empty!>, then
power off automatically.
When the instrument is being charged, the Auto Power Off will not function.
6.14 Paper Loading
compartment with fingers and open the
strength.
According to the illustration
paper into the paper compartment with
fail to print.
compartment.
Make sure that the paper is well in place
and close the paper compartment cover.
], press any key could stop the twinkle of LCD
User´s Manual HardyTest D700®
24
Fault Appearance
Fault Analysis
Handling method
Battery failure
Replace the battery with a
new pack
The battery is switched off
Switch on the battery
No measured value
Impact device cable failure
Replace the cable
Battery exhaustion
Charge the battery
The battery is switched off
Switch on the battery
6.15 Battery Charge
The instrument uses a NI-MH battery pack as its power source. When the battery pack
almost runs out, the battery symbol on the display will glint soon as possible. Try to drain your battery pack as fully as possible before it is charged for
longest battery service.
The battery switch should be at ‘ON’ condition before charging.
Plug the power adapter into the mains supply power socket and then plug the
charger connector into the power jack of the instrument. If the instrument is in
power off condition, it will turns on automatically after the charger plug is inserted
into the power jack. The battery symbol will alternately shows between
when charging. The more of the dark part indicates the more close to full
capacity.
When the battery is fully charged, the battery symbol on the display will glint
.
Please use the configured AC-DC adapter to charge the battery pack.
Warning: When the battery pack is being charged, printing or paper feeding is
unavailable.
. It needs charging as
and
6.16 Battery Replacement
When the battery pack fails to be charged, the user should replace the batteries following the
program below:
Power down the instrument.
Take off the battery compartment cover and take out the battery pack.
Insert the connection plug of the new battery pack into the socket on the circuit
board.
Reset the battery cover.
Turn on the instrument to check.
Warning: Please pay much attention to the polarity of the battery during battery
replacement.
6.17 Connection of Data Transmission Cable
Insert one connection plug of the transmission cable into the USB socket on the right side of
main body, and insert the another plug into the USB port on the back of computer box. Refer
to the manual of the DataPro software for detailed information.
7 Fault Analysis & Evacuation
Charge failure
Failure power on
User´s Manual HardyTest D700®
25
8 Servicing & Maintenance
8.1 Impact Device Servicing
After the impact device has been used for 1000--2000 times, please use the nylon brush
provided to clean the guide tube and impact body. When cleaning the guide tube, unscrew the
support ring first, then take out the impact body, spiral the nylon brush in counter-clock
direction into the bottom of guide tube and take it out for 5 times, and then install the impact
body and support ring again.
Release the impact body after use.
Any lubricant is absolutely prohibited inside the impact device.
8.2 Normal Maintenance Program
When using standard Rockwell hardness block to testing, if all the error is bigger than 2 HRC,
it may be the invalidation of impacted ball top caused by abrasion. Changing the spherical
test tip or impact object should be considered.
When the hardness tester appears some other abnormal phenomena, please do not
dismantle or adjust any fixedly assembled parts. Fill in and present the warranty card to us.
The warranty service can be carried on.
9 Calibration Time
Calibration is needed every 1 year.
10 Notice of Transport and Storage Conditions
Keep it away from vibration, strong magnetic field, corrosive medium, dumpiness
and dust. Storage in ordinary temperature.
With original packing, transport is allowed on the third grade highway.
User´s Manual HardyTest D700®
26
scale
Impact device
alloys
/tin alloys)
alloys
APPENDIX
Table 1 Hardness Values Materials
Hard-
Material
ness
HRC 20-68.5 19.3-67.9 20.0-69.5 22.4-70.7 20.6-68.2
HRB 38.4-99.6 47.7-99.9 37.0-99.9
D/DC D+15 C G E DL
Steel and
cast steel
Cold work
tool steel
Stainless
steel
Grey cast
iron
Nodular cast
iron
HRA 59.1-85.8 61.7-88.0
HB 127-651 80-638 80-683 90-646 83-663 81-646
HV 83-976 80-937 80-996 84-1042 80-950
HS 32.2-99.5
HRC 20.4-67.1 19.8-68.2 20.7-68.2 22.6-70.2
HV 80-898 80-935 100-941 82-1009
HRB 46.5-101.7
HB 85-655
HV 85-802
HRC
HB 93-334 92-326
HV
HRC
HB 131-387 127-364
HV
33.3-99.3 31.8-102.1 35.8-102.6 30.6-96.8
Cast
aluminum
Brass
(copper-zinc
alloys)
BRONZE
(copper-alu
minum
Wrought
copper
HB 19-164 23-210 32-168
HRB 23.8-84.6 22.7-85.0 23.8-85.5
HB 40-173
HRB 13.5-95.3
HB 60-290
HB 45-315
User´s Manual HardyTest D700®
27
Table 2 Material Strength
No. Material HLD Strength σ
1 Mild steel 350522 374780
2 High-Carbon steel 500710 7371670
3 Cr steel 500730 7071829
4 Cr-V steel 500750 7041980
5 Cr-Ni steel 500750 7632007
6 Cr-Mo steel 500738 7211875
7 Cr-Ni-Mo steel 540738 8441933
8 Cr-Mn-Si steel 500750 7551993
9 Super strength steel 630800 11802652
10 Stainless steel 500710 7031676
(MPa)
b
User´s Manual HardyTest D700®
28
Type of impact device
D, DC, DL
D+15
C G E
Test tip hardness:
1600HV
carbide
1600HV
carbide
1600HV
carbide
1600HV
carbide
5000HV
diamond
Max. hardness of sample:
940HV
940HV
1000HV
650HB
1200HV
Min. weight of sample:
Need coupling tightly
0,05-2kg
0,05-2kg
0,02-0,5kg
0,5-5kg
0,05-2kg
Min. thickness of sample
surface hardening:
0.8mm
≥0.8 mm
0.2mm
1.2mm
0.8mm
Hardness
Indentation
indentation:
24μm
24μm
12μm
53μm
24μm
Hardness
Indentation
indentation:
17μm
17μm
8μm
41μm
17μm
Hardness
Indentation
indentation:
10μm
10μm
7μm
--
10μm
D: normal test
Table 3 Technical Details Impact Devices
Impacting energy: Mass of impact body:
Dia. Test tip: Material of test tip:
Impact device diameter: Impact device length: Impact device weight:
Mean roughness value of sample surface Ra:
Measure directly Need support firmly
Coupling tightly: Min. layer thickness for
Size of tip indentation
11mJ
5.5g/7.2g
3mm Tungsten
20mm 86(147)/ 75mm 50g
1.6μm 1.6μm 0.4μm 6.3μm 1.6μm
> 5kg 2-5kg
5mm
11mJ
7.8g
3mm Tungsten
20mm 162mm 80g
> 5kg 2-5kg
5mm
2.7mJ
3.0g
3mm Tungsten
20mm 141mm 75g
> 1,5kg 0,5-1,5kg
1mm
90mJ
20.0g
5mm Tungsten
30mm 254mm 250g
> 15kg 5-15kg
10mm
11mJ
5.5g
3mm synthetic
20mm 155mm 80g
> 5kg 2-5kg
5mm
300HV
600HV
800HV
Available type of impact device:
diameter: Depth of
diameter: Depth of
diameter:
Depth of
0.54mm
0.54mm
0.35mm
DC: for hole or hollow cylindrical; DL: for slender narrow groove or hole
0.54mm
0.54mm
0.35mm
D+15: fot
groove or reentrant surface
0.38mm
0.32mm
0.35mm
C: for small, light or thin parts and surface of hardened layer
1.03mm
0.90mm
--
G: for large, thick, heavy and rough surface steel
0.54mm
0.54mm
0.35mm
E: for super high hardness material
User´s Manual HardyTest D700®
29
surface R10~R15
surface R25~R50
R12.5~R17
SR11~SR13
SR16.5~SR30
Table 4 Supporting Rings
No. Type Sketch of non-conventional
Supporting ring
1 Z10-15
2 Z14.5-30
3 Z25-50
4 HZ11-13
5 HZ12.5-17
6 HZ16.5-30
7
8 K14.5-30
K10-15
Remarks
For testing cylindrical outside
For testing cylindrical outside
surface R14.5R30
For testing cylindrical outside
For testing cylindrical inside surface
R11R13
For testing cylindrical inside surface
For testing cylindrical inside surface
R16.5R30
For testing spherical outside surface
SR10SR15
For testing spherical outside surface
SR14.5SR30
9 H K 11 -13
10 HK12.5-17
11 HK16.5-30
12
UN
For testing spherical inside surface
For testing spherical inside surface
SR12.5SR17
For testing spherical inside surface
For testing cylindrical outside
surface,radius adjustable R10
30
Table 5 Specifications
Specifications
HardyTest D700®
Specifications are subject to change without prior notice.
Hardness units:
HL (Leeb), HB (Brinell), HRB (Rockwell B),
ers), HS
(Shore D)
Measuring range:
170 – 960 HLD
Measuring direction:
360°
Standard impact device:
D
Memory size:
Max. 500 groups (relative to 1 - 32 impact
Statistics: Number of measurements, date, average
value, impact device, material, measured value, hardness unit
Setting of limits:
Acoustic signal when exceeding preset
Weight of sample:
> 5 kg solid material; 2-5 kg on stable surface; < 2 kg with coupling paste on
Memory function:
Manually oder automatically
Data transfer:
USB-cable and software
Menu languages:
English and german
Display:
128 × 64 Dot-Matrix-LCD
Backlight:
Adjustable brightness
Battery capacity:
150 hours (without backlight and printing)
Power supply: Battery-pack (6V NI-MH) with integrated charging cable
Battery charging cable:
9 V / 500 mA
Print-out paper (length x width):
57,5 (± 0,5 mm) x 30 mm Storage temperature:
-30°C to + 60°C
Relative humidity:
≤ 90%
Working temperature:
-10°C to + 50°C
Weight:
340 g (with batteries)
Size (length x width x height):
212 x 80 x 32 mm
User´s Manual HardyTest D700®
HRC (Rockwell C), HV (Vick
times)
min. und max. limits
stable surface
charging circuit and
SaluTron Messtechnik GmbH
Dr.-Gottfried-Cremer-Allee 30/7 · D-50226 Frechen
Tel. +49 (0) 2234 9999960 · Fax. +49 (0) 2234 9999962
Email: info@salutron.de · www.salutron.de
Loading...