Salter BPW-9100 Instructions And Guarantee

Page 1
BPW-9100
BLOOD PRESSURE MONITOR
Instructions and Guarantee
E
BLOOD PRESSURE MONITOR
Instructions and Guarantee
E
Page 2
GB
WEEE EXPLANATION
This marking indicates that this product should not be disposed with other household wastes throughout the EU. To prevent possible harm to the environment or human health from uncontrolled waste disposal, recycle it responsibly to promote the sustainable reuse of material resources. To return your used device, please use the return and collection systems or contact the retailer where the product
was purchased. They can take this product for environmental safe recycling.
BATTERY DIRECTIVE
This symbol indicates that batteries must not be disposed of in the domestic waste as they contain substances which can be damaging to the environment and health. Please dispose of batteries in designated collection points.
GUARANTEE
Salter will repair or replace the product, or any part of this product, free of charge if within 2 years of the date of purchase, it can be shown to have failed through defective workmanship or materials. This guarantee covers working parts that aect the function of the blood pressure monitor. It does not cover cosmetic deterioration caused by fair wear and tear or damage caused by accident or misuse. Opening or taking apart the blood pressure monitor or its components will void the guarantee. Claims under guarantee must be supported by proof of purchase and be returned carriage paid to Salter (or local Salter appointed agent if outside the UK). Care should be taken in packing the blood pressure monitor so that it is not damaged while in transit. This undertaking is in addition to a consumer’s statutory rights and does not aect those rights in any way. For UK Sales and Service contact HoMedics Group Ltd, HoMedics House, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP, UK. Helpline Tel No: (01732) 360783. e-mail: support@homedics.co.uk. For Ireland, please contact Petra Brand Masters, Unit J4 Maynooth Business Campus, Maynooth, Co. Kildare, Ireland. Tel +00 353 (0) 1 6510660. e-mail sales@petrabrandmasters.ie.
www.salterhousewares.com/servicecentres
NORMAL BLOOD PRESSURE FLUCTUATION
All physical activity, excitement, stress, eating, drinking, smoking, body posture and many other activities or factors (including taking a blood pressure measurement) will influence blood pressure value. Because of this, it is mostly unusual to obtain identical multiple blood pressure readings. Blood pressure fluctuates continually ----- day and night. The highest value usually appears in the daytime and lowest one usually at midnight. Typically, the value begins to increase at around 3:00AM, and reaches to highest level in the daytime while most people are awake and active. Considering the above information, it is recommended that you measure your blood pressure at approximately the same time each day. Too frequent measurements may cause injury due to blood flow interference, please always relax a minimum of 1 to 1.5 minutes between measurements to allow the blood circulation in your arm to recover. It is rare that you obtain identical blood pressure readings each time.
Page 3
GB
INTENDED USE
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cu is wrapped around the wrist.
CONTRAINDICATION
It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer.
CONTENTS AND DISPLAY INDICATORS
LCD display
Blood Pressure Level
Classification Indicator
Cu
Memory Indicator
Date/Time Display
Systolic Pressure
MEM button
START button
Diastolic Pressure/ Pulse Rate Display (Alternating)
Low Battery Indicator
Blood pressure level
Classification Indicator
Irregular heartbeat symbol
Ready to inflate indicator
Page 4
GB
PRODUCT DESCRIPTION
Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can be measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate.The most recent 60 measurements can be stored in the memory with date and time stamp.The Electronic Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance), IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests), IEC 80601-2-30:2009+Cor.2010/EN 80601-2-30:2010(Medical electrical equipment –Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers),EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems). ANSI/AAMI SP-10:2002+A1:2003+A2:2006
SPECIFICATIONS
Product name: Blood Pressure Monitor
Model BPW-9100
Classification
Internally powered, Type BF applied part,IPX0,No AP or APG,Continuous operation
Machine size Approx. 85mm x 64.5mm x 28mm (3 11/32”x 2 17/32”x 1 3/32”)
Accuracy
Pressure: 5ºC-40ºc within±0.4kpa(3mmHg) pulse value:±5%
Cu circumference 14cm - 19.5cm(5 1/2” - 7 11/16”)
Weight Approx. 110g (3 7/8 oz.) (exclude batteries)
Measuring method Oscillometric method, automatic inflation and measurement
Memory volume 60 times with time and date stamp
Power source
Batteries: 2 × 1.5V Size AAA
Measurement range
Cu pressure: 0-300mmHg Systolic: 60-260mmHg Diastolic: 40-199mmHg Pulse rate: 40-180 beats/minute
Accuracy
Pressure: ±3mmHg Pulse rate: ±5%
Environmental temperature for operation 5ºC~40ºC (41ºF~104ºF)
Environmental humidity for operation ≤90%RH
Environmental temperature for storage and transport -20ºC~55ºC (-4ºC~131ºC)
Environmental humidity for storage and transport ≤90%RH
Environmental pressure 80kPa-105kPa
Battery life Approx 270 times All components belonging to the pressure measuring
system, including accessories
Pump, Valve, LCD, Cu, Sensor
Note: These specifications are subject to change without notice.
Page 5
GB
NOTICE
1. Read all of the information in the operation guide and any other literature in the box before operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cu should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same wrist for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation in your arm to recover. Prolonged over-inflation (cu pressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes) of the bladder may cause ecchymoma of your wrist.
7. Consult your physician if you have any doubt about below cases:
1) The application of the cu over a wound or inflammation diseases;
2) The application of the cu on any limb where intravascular access or therapy, or an arterio-venous (A-V) shunt, is
present;
3) The application of the cu on the arm on the side of a mastectomy;
4) Simultaneously used with other monitoring medical equipments on the same limb;
5) Need to check the blood circulation of the user.
8. This Electronic Sphygmomanometers is designed for adults and should never be used on infants or young children. Consult your physician or other health care professionals before use on older children.
9. Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained by a trained observer using the cu/stethoscope auscultation method, within the limits prescribed by the American National Standard Institute, Electronic or automated sphygmomanometers.
11. Information regarding potential electromagnetic or other interference between the blood pressure monitor and other
devices together with advice regarding avoidance of such interference please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION.
12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood pressure
measurement, a signal of will be displayed. Under this condition, the Electronic Sphygmomanometers can keep function, but the results may not be accurate, it’s suggested that you consult with your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coecient of variation (CV) of pulse period >25%.
2) The dierence of adjacent pulse period≥0.14s, and the number of such pulse takes more than 53 percentage of the
total number of pulse.
13. Please do not use the cu other than supplied by the manufacturer, otherwise it may bring biocompatible hazard and might result in measurement error.
14. The monitor might not meet its performance specifications or cause safety hazard if stored or used outside the specified temperature and humidity ranges in specifications.
15. Please do not share the cu with other infective person to avoid cross-infection.
16. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:
— Reorient or relocate the receiving antenna. — Increase the separation between the equipment and receiver. — Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected. — Consult the dealer or an experienced radio/TV technician for help
Page 6
GB
17. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
18. This blood pressure monitor is verified by auscultatory method. It is recommended that you check annex B of ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for details of verification method if you need.
SETUP AND OPERATING PROCEDURES
Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cu is wrapped around the wrist.
1. BATTERY LOADING
a. Open battery cover at the back of the monitor. b. Load two “AAA” size batteries. Please pay attention to polarity. c. Close the battery cover. When LCD shows battery symbol , replace all batteries with new ones. Rechargeable batteries are not suitable for this monitor. Remove the batteries if the monitor will not be used for a month or more to avoid relevant damage of battery leakage.
Avoid the battery fluid to get in your eyes. If it should get in your eyes, immediately rinse with plenty of clean water and
contact a physician.
The monitor, the batteries and the cu, must be disposed of according to local regulations at the end of their usage.
2. CLOCK AND DATE ADJUSTMENT
a. Once you install the battery or turn o the monitor, it will enter Clock Mode, and LCD will display time and date by turns. See picture 2&2-1.
Picture 2 Picture 2-1 Picture 2-2
b. While the monitor is in Clock Mode, pressing both the “START” and “MEM” button simultaneously, a beep is heard and the
month will blink at first. See picture 2-2. Press the button “START” repeatedly, the day, hour and minute will blink in turn. While the number is blinking, press the button “MEM” to increase the number. Keep on pressing the button “MEM”, the number will increase fast.
c. You can turn o the monitor by pressing “START” button when the minute is blinking, then the time and date is
confirmed.
d. The monitor will turn o automatically after 1 minute of no operation, with the time and date unchanged. e. Once you change the batteries, you should readjust the time and date.
Page 7
GB
3. CONNECTING THE CUFF TO THE MONITOR
The cu is attached to the monitor when it is packaged. Should the cu become unattached, align the two plugs and four brackets of the cu with the plug sockets and bracket sockets of the monitor and press the cu to the monitor until the plugs and brackets are securely attached.
4. APPLYING THE CUFF
a. Place the cu around a bare wrist 1-2cm above the wrist joint on the palm side of the wrist. b. While seated, place the arm with the cued wrist in front of your body on a desk or table with the palm up.
If the cu is correctly placed, you can read the LCD display.
c. The cu must be neither too tight nor too loose.
Note:
1. Please refer to the cu circumference range in “SPECIFICATIONS” to make sure that the appropriate cu is used.
2. Measuring on same wrist each time.
3. Do not move your arm, body, or the monitor during measurement.
4. Stay quiet, calm for 5 minutes before blood pressure measurement.
5. Please keep the cu clean. Clean the cu by wet soft cloth and mild detergent if the cu becomes dirty. Do not remove the cu from the monitor.Clean the cu after the usage of every 200 times is recommended.
5. BODY POSTURE DURING MEASUREMENT
Sitting Comfortably Measurement
a. Be seated with your feet flat on the floor, and don’t cross your legs. b. Place palm upside in front of you on a flat surface such as a desk or table. c. The middle of the cu should be at the level of the right atrium of the heart.
6. TAKING YOUR BLOOD PRESSURE READING
a. After applying the cu and your body is in a comfortable position, press the “START” button. A beep is heard and all
display characters are shown for self-test. See picture 6. Please contact the service center if segment is missing.
Picture 6 Picture 6-1 Picture 6-2 Picture 6-3
b. Then the monitor starts to seek zero pressure. See picture 6-1. c. The monitor inflates the cu until sucient pressure has built up for a measurement. Then the monitor slowly releases
air from the cu and carries out the measurement. Finally the blood pressure and pulse rate will be calculated and displayed on the LCD screen separately. Irregular heartbeat symbol (if any) will blink. See picture 6-2&6-3. The result will be automatically stored in the memory bank.
d. After measurement, the monitor will turn o automatically after 1 minute of no operation. Alternatively, you can press
the “START” button to turn o the monitor manually.
e. During measurement, you can press the “START” button to turn o the monitor manually. Note: Please consult a health care professional for interpretation of pressure measurements.
Page 8
GB
7. DISPLAYING STORED RESULTS
a. After measurement, you can review the results in the memory bank by pressing the “MEM” button. Alternatively, you can
press “MEM” button in Clock Mode to display the stored results. If it no result stored, LCD will show dashes as picture 7, while press the button “MEM” or “START”, machine will turn o. If there are results in the memory bank, the LCD will display the amount of the results in the memory bank. See picture 7-1.
Picture 7 Picture 7-1 Picture 7-2
Picture 7-3 Picture 7-4 Picture 7-5
b. And then, the most recent result will be displayed with date and time stamp. See picture 7-2. Followed by, the blood
pressure and pulse rate will be shown separately. Irregular heartbeat symbol (if any) will blink. See picture 7-3&7-4. Press “MEM” button again to review the next result. See picture 7-5. In this way, repeatedly pressing the MEM button displays the respective results measured previously.
c. When displaying the stored results, the monitor will turn o automatically after 1 minute of no operation. You can also
press the button “START” to turn o the monitor manually.
8. DELETING MEASUREMENTS FROM THE MEMORY
When any result is displaying, keeping on pressing button “MEM” for three seconds, all results in the current memory bank will be deleted after three “beep”. LCD will show picture 8, Press the button “MEM” or “START”, the monitor will turn o.
Picture 8
Page 9
GB
9. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS
The following guidelines for assessing high blood pressure (without regard to age or gender) have been established by the World Health Organization (WHO). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) need to be taken into consideration. Consult with your physician for accurate assessment, and never change your treatment by yourself.
Classification of blood pressure for adults
10. TECHNICAL ALARM DESCRIPTION
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if the determined blood pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICATIONS. In this case, you should consult a physician or check if your operation violated the instructions. The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8. The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically after about 8 seconds.
11. TROUBLESHOOTING (1)
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display shows abnormal result
The cu position was not correct or it was not properly tightened
Apply the cu correctly and try again
Body posture was not correct during testing
Review the “BODY POSTURE DURING MEASUREMENT” sections of the instructions and re-test.
Speaking, arm or body movement, angry, excited or nervous during testing
Re-test when calm and without speaking or moving during the test
Irregular heartbeat (arrhythmia)
It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer
Page 10
GB
10
11. TROUBLESHOOTING (2)
PROBLEM POSSIBLE CAUSE SOLUTION
LCD shows low battery symbol
LCD shows low battery symbol Change the batteries
LCD shows “Er 0” Pressure system is unstable before
measurement
Don’t move and try again.
LCD shows “Er 1” Fail to detect systolic pressure
LCD shows “Er 1” Fail to detect diastolic pressure
LCD shows “Er 1” Pneumatic system blocked or cu is
too tight during inflation
Apply the cu correctly and try again
LCD shows “Er 4” Pneumatic system leakage or cu is
too loose during inflation
LCD shows “Er 5” Cu pressure above 300mmHg
Measure again after five minutes. If the monitor is still abnormal, please contact the local distributor or the factory.
LCD shows “Er 6” More than 3 minutes with cu
pressure above 15 mmHg
LCD shows “Er 7” EEPROM accessing error
LCD shows “Er 8” Device parameter checking error
LCD shows “Er A” Pressure sensor parameter error
No response when you press button or load battery.
Incorrect operation or strong electromagnetic interference
Take out batteries for five minutes, and then reinstall all batteries.
NOTICE
01. Do not drop this monitor or subject it to strong impact.
02. Avoid high temperature and solarization. Do not immerse the monitor in water as this will result in damage to the
monitor.
03. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
04. Do not attempt to disassemble this monitor.
05. If you do not use the monitor for a long time, please remove the batteries.
06. It is recommended the performance should be checked every 2 years or after repair. Please contact the service center.
07. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after moistened with water, diluted disinfectant
alcohol, or diluted detergent.
08. No component can be maintained by user in the monitor. The circuit diagrams, component part lists, descriptions,
calibration instructions, or other information which will assist the user’s appropriately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied.
09. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three
years, and the cu integrity is maintained after 1,000 open–close cycles of the closure.
10. It is recommended the cu should be disinfected 2 times every week if needed (For example, in hospital or in clinique).
Wipe the inner side (the side contacts skin) of the cu by a soft cloth squeezed after moistened with Ethyl alcohol
(75-90%), then dry the cu by airing.
Page 11
GB
11
CONTACT INFORMATION
For more information about our products, please visit www.salterhousewares.com For UK Sales and Service contact HoMedics Group Ltd, HoMedics House, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP, UK. Helpline Tel No: (01732) 360783. For Ireland, please contact Petra Brand Masters, Unit J4 Maynooth Business Campus, Maynooth, Co. Kildare, Ireland. Tel +00 353 (0) 1 6510660. e-mail sales@petrabrandmasters.ie.
www.salterhousewares.com/servicecentres
EXPLANATION OF SYMBOLS ON UNIT
Symbol for” THE OPERATION GUIDE MUST BE READ”(The sign background colour: blue.The sign graphical symbol:
white)
Symbol for “WARNING”
Symbol for “TYPE BF APPLIED PARTS” (The cu is type BF applied part)
Symbol for “ENVIRONMENT PROTECTION – Waste electrical products should not be disposed of with household
waste. Please recycle where facilities exist. Check with your local Authority or retailer for recycling advice”.
Symbol for “MANUFACTURER”
0120
Symbol for “COMPILES WITH MDD93/42/EEC REQUIREMENTS”
Symbol for “DATE OF MANUFACTURE”
EC REP
Symbol for “EUROPEAN REPRESENTATION”
SN Symbol for “SERIAL NUMBER”
Symbol for “KEEP DRY”
Page 12
GB
12
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The [BPW-9100] is intended for use in the electromagnetic environment specified below. The customer or the user of the [BPW-9100] should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The [BPW-9100] uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11
Class B
The [BPW-9100] is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions IEC 61000-3-3
Not applicable
Table 2 For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The [BPW-9100] is intended for use in the electromagnetic environment specified below. The customer or the user of the [BPW-9100] should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3A/m 3A/m Power frequency magnetic
fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
Page 13
GB
13
Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The [BPW-9100] is intended for use in the electromagnetic environment specified below. The customer or the user of the [BPW-9100] should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the BPW-9100, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
P 80 MHz to 800 MHz
d = 2.3
P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the [BPW-9100] is used exceeds the applicable RF compliance level above, the [BPW-9100] should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the [BPW-9100].
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Page 14
14
GB
Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the [BPW-9100]
The [BPW-9100] is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the [BPW-9100] can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the [BPW-9100] as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2
P
80 MHz to 800 MHz
d = 1.2
P
800 MHz to 2.5 GHz
d = 2.3
P
0.01 0.12 0.12
0.23
0.1 0.38 0.38
0.73
1 1.2 1.2
2.3
10
3.8 3.8
7.3
100
12 12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and
reflection from structures, objects and people.
Page 15
Page 16
HoMedics Group Ltd
HoMedics House, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP, UK. www.salterhousewares.co.uk IB-BPW-9100-GB-0314-01
E
Register your product today at:
www.homedicsgroup.com/register
0120