Sakura Tissue-Tek Glas Quick Reference Manual

Coverslipping
Tissue-Tek® Glas™ Coverslipper
Quick Reference Guide
Tissue-Tek® Glas™ Coverslipper
Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL
Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | sakura@sakura.eu | www.sakura.eu
The Tissue-Tek® GlasTM Coverslipper is a system that dispenses mounting medium onto stained histopathological and cytopathological samples on glass slides as well as coverslipping these samples using cover glasses.
Loading door
Adjustable
Feet level
Coverglass
holder
Available in 40,
50, 55 and 60
mm
Display Panel
Debris tray
Mounting medium
bottle
Receiving racks
Tissue-Tek® Glas™ Coverslipper
Coverslipping
Quick Reference Guide
Tissue-Tek® Glas™ Coverslipper
page 2 | 6 QRG | Glas Coverslipper | 2011 | V-2
Preparation before cover slipping
Hardware
Placing cover glasses
1. Put cover glass holder in place.
2. Make sure the tab of the cover glass holder fits
in the holder (check on the CG size).
Placing coverglasses
Placing mounting medium
When setting the mounting medium bottle for the first time, you must fill the discharging channel with mounting medium.
1. Place the waste bottle below the dispensing
nozzle.
2. Fill the priming bottle with mounting medium.
3. Connect the tip of the introduction tube with
the priming bottle.
4. Squeeze the priming bottle to feed the
mounting media until it reaches the dispensing nozzle.
5. Remove the priming bottle from the
introduction tube.
6. Set the mounting medium bottle.
7. Tilt the mounting medium tube towards the
front and lift the tube.
8. In order to prevent air bubbles, insert the tube
gently into the bottle.
9. Perform an [initial priming] page 2.7 OM.
10. Press [EMER. STOP] to enter the set up
menu.
11. Highlight [1. Prime Dispenser].
12. Press [Verify].
13. Press [Start] and again [Start] to prime.
Changing mounting medium brands Since mounting mediums vary in viscosity and formulation, it will be necessary to follow the same priming procedure when changing to a different brand of mounting medium.
Follow the detailed instructions written in the Operating Manual, “priming the instrument” beginning on page 2.7.
Placing mounting medium Priming bottle connected
to introduction tube
Cover glass holder
Holder hole
Cover glasses
Introduction tube