Sakura Tissue-Tek AutoTEC a120 Quick Reference Manual

Tissue Embedding
Tissue
-
Tek® AutoTEC®
a120
Tissue-Tek® AutoTEC® a120
The Tissue-Tek® AutoTEC® a120 Automated Embedding System is designed for the purpose of paraffin embedding human and animal tissue specimens.
The instrument, as part of the histopathology process, is intended to facilitate the in vitro examination of human and animal tissue for morphology changes by a pathologist.
The Tissue-Tek® AutoTEC® a120 can be programmed to initiate an automatic four hour system warm up so it is ready for operation when the workday begins.
Tissue-Tek® AutoTEC® a120
Run preparation
Prior to starting the first run:
1. Check that the 4 output doors are empty and
installed
2. Load a maximum of assembled 20 Tissue-Tek
®
Paraform® cassettes into each magazine:
a. Ensure that cassettes are properly
inserted in their frames, lids securely closed.
b. Insert cassettes parallel in magazines,
top sides facing up, beveled writing surface of the frames facing magazine lid.
c. Cassette retainer slides to one space
above topmost cassette.
d. Slide magazine cover on all the way.
3. LOG IN, touch LOG IN and enter your password
20 cassette magazine black
retainer
Do not exceed 20 cassettes
Tissue
-Tek® AutoTEC® a120
Tissue Embedding
Loading magazines
1. Press the LOAD MAGAZINE icon on the main
Monitor screen
2. When the message DOOR IS UNLOCKED is
displayed, the input door can be opened.
3. Up to 4 Magazines can be loaded, slide the
magazine to the bottom.
4. Securely close the input door, automatic
embedding will commence after the programmable warm-up time has completed.
Note: Cold magazines should not be loaded into the instrument, lids leveled to the posts of the input door.
Input door icon
Up to 4 magazines can be
loaded
Monitor screen
The Monitor screen provides the applicable information on Input Door, Output Doors and Basemold area during the embedding cycle.
Cassettes placed on the recovery area due to open lid or cassette assembly issues can be retrieved by pressing the INSTRUMENT ACCESS icon on the left bottom corner.
Monitor screen overview
Unloading
When the embedding of the blocks has been completed the blocks are placed in one of the Output door(s). The output door icon indicates the door can be removed of the instrument.
Output door icon
Up to 20 blocks are placed in
the output door
Troubleshooting
Note the Operating Manual Tissue-Tek® AutoTEC® a120, chapter 7.
Dropped cassettes: In case a cassette is dropped from the gripper a sensor will be triggered notifying the user and allow recovery of the cassette.
Error’s, Messages or attentions: If displayed on the screen follow press RESUME or shutdown and follow further instructions displayed on the screen.
Insufficient Paraffin: In case of insufficient molten paraffin in the paraffin reservoir, the instrument will stop pulling cassettes from the magazine in the input door and display 0 cassettes in the dashboard.
Cassette assembly: In case the Tissue-Tek® Paraform® cassette is not properly assembled or the lid is not closed completely, the instrument will palce the cassette in the recovery area and will display its presence in the Monitor Screen.
Maintenance
Note the Operating Manual Tissue-Tek® AutoTEC® a120, chapter 5 or the Application Bulletin Tissue-Tek® AutoTEC® a120 Cleaning instructions.