Sakura Cyto-Tek 2500 Quick Reference Manual

Page 1
Cytology
Cyto-Tek® 2500 Cytocentrifuge
Quick Reference Guide
Cyto-Tek® 2500 Cytocentrifuge
Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL
Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | sakura@sakura.eu | www.sakura.eu
Cyto-Tek® 2500 Cytocentrifuge
Cyto-Tek® 2500 Cytocentrifuge
1
2
3
Page 2
Cyto-Tek® 2500 Cytocentrifuge
Cytology
Quick Reference Guide
Cyto-Tek® 2500 Cytocentrifuge
page 2 | 4 QRG | Cytocentrifuge | V-1
Opening the Cyto-Tek® 2500 Cytocentrifuge
Guard bowl including rotor
Detachable lid
Unlock guard bowl
Remove guard bowl
6. Speed Display:
Displays a program speed or actual speed.
7. Time display:
Displays a programmed time or process time left.
8. Lid Lock indicator:
A LED lights up when the lid is locked at the start of operation and goes out when the lid is unlocked after the motor has stopped.
9. Power indicator:
A LED lights up while the power is supplied to the instrument.
Control Panel
Page 3
Cytology
Cyto-Tek® 2500 Cytocentrifuge
Quick Reference Guide
Cyto-Tek® 2500 Cytocentrifuge
QRG | Cytocentrifuge | V-1 page 3 | 4
Guard Bowl loading always in pairs
Incorrectly set specimen chamber
Starting the operation
Page 4
Cyto-Tek® 2500 Cytocentrifuge
Cytology
Quick Reference Guide
Cyto-Tek® 2500 Cytocentrifuge
page 4 | 4 QRG | Cytocentrifuge | V-1
Maintenance
More information in operating manual section 4: Care of the instrument
Daily
Control panel cleaning
1. Wipe the control panel with a clean cloth
moistened with neutral detergent.
Interior and exterior surfaces
1. Clean with a dry clean cloth or a cloth
moistened with ethylalcohol or isopropyl­alcohol 70%-80%.
Caution: Do not use volatile fluids like xylene, thinner, benzene Caution: Do not spray a disinfectant directly on or over the instrument.
Monthly
Guard bowl and rotor cleaning
1. Disassemble the guard bowl unit.
2. Steralize the guard bowl body and
components in an autoclave for > 20 minutes at 121°C
When no autoclave is available:
1. Disassemble the guard bowl unit.
2. Soak guard bowl body and components in
an ethylalcohol or isopropylalcohol 70%­80%.
3. or sodium hypochlorite 0.5% solution for at
20 minutes.
4. Wash with water and soap.
5. Rinse with clear water.
6. Dry thoroughly (when using an oven keep
temperature below 120°C).
Yearly
1. Replace the lid and top gasket.
2. Preventative maintenance by a Sakura
representative or a qualified service engineer.
Control panel
Guard bowl & rotor and lid gaskets
Error handling
When a power outage occurs while the instrument is in operation, the instrument lid will remain locked. When the motor stopped the lid can be opened mechanically.
1. Remove a small cap located on the left side
of the instrument.
2. Insert and push an unlock bar into the hole.
Remove a small cap
Insert an unlock bar