RTI Black Piranha, GE Senographe Pristina Quick Manual

Quick Guide
9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 1(4) v.2017.3D
This Quick Guide refers to use of the Black Piranha on the GE Senographe Pristina.
1. Before starting
a. The Black Piranha
1
must have a calibration for mammography performed at RTI
after May 1, 2016.
b. The Ocean software version should be 2016.12.27.243 or higher. c. The GE Quick Check templates, version 1.1D, for Ocean must be installed. d. Make sure to have the Piranha Pristina holder, GE part # 5758284, available (image
to right).
1
Piranha ordered under GE Part # 5758293 is Black, and must have a calibration for
mammography performed at RTI after May 1, 2016. Red Piranhas ordered under GE Part # 5401610, or 5458262 must have been upgraded to Black Piranha with calibration after May 1, 2016. If the Piranha has a calibration date from May 1, 2016 to March 9, 2017, it also must have been updated for correct CF calibration using the IQST phantom (see internal GE service notes).
2. Measure using GE Quick Check Templates
Setup
e. Place the Black Piranha in the holder as shown in image (right).
The Black Piranha will only fit in one way.
f. Slide the holder into the GE dosimeter support. g. The Black Piranha can be left in this positon for all measurements described below. h. Follow the job card instructions when it comes to image detector protection, IQST
phantom set-up, compression paddle set-up, etc…
Measure
When starting Ocean the measure guide as shown in image (right) should appear. If not, the option “Auto-start Quick Check” is deactivated. This function can be activated under Options.
i. Select GE Mammography… j. In next screen select GE Mammography[Senographe Pristina]
k. In next screen (image to right) select the proper application: Conversion
factor, HVL, or kVp.
Quick Guide
9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 2(4) v.2017.3D
l. When starting one of the test a set-up instruction (hint)
will pop-up, see image.
m. Press the right arrow key to get to the measure screen.
Notice:
Depending on configuration of Ocean, a Position Check may be suggested (yellow bar besides the green “ready” box). If so, just
abort the Position Check by clicking anywhere in the yellow area.
This will be the case for every new target/filter selection.
With the good geometry of the Pristina unit, and use of the Piranha Pristina holder, GE part # 5758284, the Position check does not add confidence to the measurement.
Conversion Factor – CAL 009
The image to the right shows the template for the conversion factor measurement. Follow the GE job card instructions for the measurement.
Notice:
The template is built for mR units. If mGy units shall be used, just right click on the unit and select mGy.
When ready for next test, just press the New Quick Check button. You shall now be back to the selection screen as described under (i) above.
HVL – CAL A010
The template is built for the sequence to measure small and large focus in the same sequence, but can be used also for one at the time. The template can be used for one shot HVL as well as for traditional measurements by adding layers of aluminum sheets.
Quick Guide
9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 3(4) v.2017.3D
The image to the right shows the template for the HVL measurement on Small Focus. The template supports one shot HVL as well as traditional method by adding layers of Aluminum. Follow the job card instructions for the measurement. No settings in Ocean should have to be made.
Notice: The Position check may be suggested (yellow bar besides the green
“ready” box). Is so, abort by clicking in the yellow field as described above.
When ready for next test, just press the New Quick Check button. You shall now be back to the selection screen as described under (i) above, or exit Ocean.
Tube Voltage – ELE 010
The template is built for measurement without compression paddle. The compression paddle may be used. Then you need to go Meter Settings¸ and change the setting to Use compression paddle.
Follow the GE job card instructions for the measurement.
Notice:
The Position check may be suggested. If so, abort by clicking in the anywhere in the yellow area as described above.
When ready for next test, just press the New Quick Check button. You shall now be back to the selection screen as described under (i) above, or exit Ocean.
3. Measure using GE Mammo Quick Check
Setup
n. Place the Black Piranha in the holder as shown in image (right) with the left pin
inside the tripod mounting hole. The Black Piranha will only fit in one way.
o. Slide the holder into the GE dosimeter support. p. The Black Piranha can be left in this positon for all measurements described below. q. Follow the job card instructions when it comes to image detector protection, IQST
phantom set-up, compression paddle set-up, etc…
Measure
When starting Ocean the measure guide as shown in image (right) should appear. If not, the option “Auto-start Quick Check” is deactivated. This function can be activated under Options.
r. Select Mammo quick check. s. A set-up instruction (hint) will pop-up. t. Press the right arrow key in the lower left corner to get to the measure screen.
Quick Guide
9604554-01 Quick Guide Black Piranha on Pristina 2017.3D.docx 4(4) v.2017.3D
u. Select calibration, by clicking the model/beam
quality. See image to the right. As default compression paddle is in use. If compression paddle is not in use, go to Meter settings and deselect.
Notice:
The Position check may be suggested (yellow bar
besides the green “ready” box). If so, abort by clicking in anywhere in the yellow area as described above.
Notice:
If doing traditional HVL measurements by adding layers of Al sheets, the thickness of the filter must be entered in the field for Added filtration under Meter Settings.
Notice:
When measuring the CF factor using the IQST phantom the displayed kV may not be within 34 ±1 kV. This does not affect the accuracy of the dose reading which is the primary target of this measurement.
4. Measure using standard Ocean configuration
In case the GE Quick Check Templates are not configured with the Ocean in use, the procedure is slightly different from what is describe under heading 3 above. After entering Quick Check (see image below), the measure guide will look a little different.
Follow the guide sequence and select Mammography - GE Healthcare - Pristina - Rh/Ag IQST Pristina - Compression Paddle not used.
The image to the right shows the template for the conversion factor measurement. Follow the GE job card instructions for the measurement.
Notice: The Position check may be suggested (yellow bar
besides the green “ready” box). If so, abort by clicking in
anywhere in the yellow area as described above.
Notice:
Depending on Ocean configuration the unit will be mR or mGy. Unit can be changed - Just right click on the unit and select unit.
Notice:
When measuring the CF factor using the IQST phantom the displayed kV may not be within 34 ±1 kV. This does not affect the accuracy of the dose reading which is the primary target of this measurement.
When ready for next test, just press the New Quick Check button. You shall now be back to the selection guide described above.
For deeper understanding and guidance, please see User’s Manuals for Piranha and Ocean.
*** END ***
Loading...