RS PRO RSDPB5000 Series, RSDPB4000 Series, RSDPB5150, RSDPB4080 Quick Start Manual

Quick Start
RSDPB5000/RSDPB4000
2
Please read this manual carefully before use
Safety precautions
1.
Be cautious of an electric shock
2.
Pay attention to the maximum input voltage
Please do not use in humid or in flammable and explosive environment
3
Model
Maximum Input Voltage
Bandwidth
Attenuation Rate
RSDPB5150
1500V
70MHz
50X/500X
RSDPB5000 Series High Voltage Differential Probe
RSDPB5000 Series Summary
Overview
RSDPB5000 series high voltage differential probes are designed for the measurement of high voltage differential signal, to meet the demand for floating measurement. The bandwidth can be as high as 100MHz, meeting the demand for majority of measurement systems.
There are a variety of ranges to choose from, and their differential voltage measurement range can meet with the demand for majority of tested circuits. Users may go into test mode to adjust the offset voltage, and also to adjust automatically to prevent probes being disordered after years’ of use. The electronic touch buttons give them a longer working life.
The function of 5MHz bandwidth limit selection, whose frequency bandwidth fits the FETs switching frequency measurement in most switching power supplies, and they can filter out higher frequency noise and interference. With the sound & light alarming function, this can also be closed manually, with a USB power supply connector to make it easier and more flexible for use. The probes are equipped with standard BNC input connectors, they can be used with any manufacturer oscilloscope ( oscilloscope input impedance should set to 1MΩ. when 50Ω is selected, the attenuation multiply attenuates double.) to test waveform of the tested circuits, there is an automatic save function, to prevent users re-operating in case of power supply drops. The probes have good capability of common mode noise suppression and can be widely used in the research and development, debugging or overhauling work for switching power supply, frequency converter, electronic ballast, frequency conversion electronic appliances and other electric power equipments.
Application
Floating voltage measurement
Frequency converter
Switching power supplies designs
Welding, plating power supplies
Induction heating, electromagnetic oven
Motor driven design
Electronic ballast design
CRT displayer design
Inverter, UPS power supplies
Frequency conversion electrical appliances
Power conversion and related design
Electrical engineering experiment
Low voltage appliances experiment
Power electronics and electric drive experiment
4
Input Port
Maximum differential Voltage 1500V Danger! High Voltage!
Commom-mode Voltage Security Level
Overrange Indicator
Attenuation Button
Attenuation Indicator Bandwidth Limit Button Bandwidth Limit Indicator
Audible Overrange Button Audible Overrange ON/OFF Indicator
Terminal Load
Power Interface: USB B Output Interface: BNC
Product and Accessories Description
Probe body description
Take RSDPB5150 as an example, voltage, range and bandwidth are varied with different
products.
Detailed Description:
1.
output will reverse when is reversely connected. Use together with standard red black input c a b l e s .
2.
500X indicate the maximum test voltage is 1500V.50X indicates the maximum test voltage is 150V. Oscilloscope attenuation factor should be set accordingly based on the probe attenuation selection.
Input connector: Standard red and black socket. Red is positive, black negative,
ATTENUATION: Different attenuation indicates different ranges, such as RSDPB5150:
5
3. BANDWIDTH: The series products have bandwidth selection function. The default is full bandwidth(FULL) of the product. When testing low frequency signal, you can choose 5MHz bandwidth limit to prevent being interfered by high frequency signal.
4. AUDIBLE OVERRANGE: When the test range exceeds probe range, audible and visual alarm will start. The function is to control buzzer alarm on or off, ON is to open audible alarm and OFF closes the alarm.
5. Output connector: Standard BNC input connectors, can be connected to any Manufacturer oscilloscope, oscilloscope input impedance should set to 1MΩ.Ifset to 50Ω, the output attenuation is a half of the practical value.
6. Power interface: Standard USB type B interface, supply power with standard USB adapter, can be supplied by the oscilloscope, it can also be supplied by portable power source, convenient for outdoor test.
7. Factory Setting: The default factory setting is high attenuation ratio, FULL bandwidth, audible alarm is on. The product has an automatic memory; this will automatically save the state before power off.
Accessories Description
Alligator Clips (CK-261 red black 1pair) Alligator Clips (CK-262 red black 1 pair)
Pincer Clips (CK-281 red black 1 pair) Hook Clips (CK-284 red black1 p a i r)
Input differential lead (CK-28 one pair)
6
Model
RSDPB5150
Alligator Clips (CK-261)
CATIII 1000V CATIV 600V
Alligator Clips (CK-262)
-----
Pincer Clips (CK-281 )
CATIII 1000V Hook Clips (CK-284 )
CATIII 1000V
High Voltage Input
differential lead (CK-28)
10A CATIII 1000V
Coaxial Output Line (CK-310)
Double-ended BNC connector coaxial line 1m Standard accessories
Coaxial Output Line (CK-320)
Double-ended BNC connector coaxial line 2m non-standard, purchased individually
USB lineCK-315
AM-BM, 1.5 m
Power adapterCK-605
USB 5V/1A
Coaxial Output Line (CK-310)
Coaxial Output Line (CK-320)
USB line (CK-315 AM-BM, 1.5m)
Power adapter (CK-605) USB 5V/1A
Product standard accessories description
7
Model
RSDPB5150
BW(-3dB)
DC-70MHz
Rise time
≤5ns
Accuracy
±2%
Attenuation Rate
50X/500X 100X/1000X
Max Differential Test Voltage (DC + Peak AC)
50X:150V 100X:700V
500X:1500V
Max input common mode voltage( voltage-to-earth
Vrms)
600V CATIII 1000V CATII
Input Impedance Single-ended to ground
5MΩ
Two inputs
10MΩ
Input Capacitance Single-ended to ground
4pF
Two inputs
2pF
CMRR DC
80dB 100kHz
60dB
1MHz
50dB
NoiseVrms
50X: 50mV 500X: 50mV 100X: 200mV 1000X: 1.2V
Propagation delay (standard 1 m output lead)
18ns±1ns
Bandwidth limit(5MHz)
-3dB@5MHz
Differential overvoltage detection level
50X:≥150V
Overload indicator (red light)
YES
Overload Alarm
Yes(Can shut up manually)
Automatic Save
YES
Offset Setting function
Yes (Set in test mode
Terminate Load
1MΩ
Power Supply
USB 5V/1A adapter
Electric Specification
8
Model
RSDPB5150
Differential Input lead
CK-28
Approx 28 cm Output Lead
CK-310
Approx 1 m
CK-320
Approx 2 m Alligator Clips CK-261
Approx 85*40*17 mm Alligator Clips CK-261
Approx 106*43*16 mm Pincer Clips CK-281
Approx 152*50*13 mm Hook Clips CK-284
Approx 121*23*23 mm Probe body dimensions
195*65*28 mm Probe body weight
Approx 188 g
Model
RSDPB5150
Operating Temperature
050
Non-operating
-3070
Operating Humidity
≤85%RH
Non-operating Humidity
≤90%RH
Operating Altitude
3000m
Non-operating Altitude
12000m
Mechanical Specification
Environment Characteristics
9
Operating the Probe Safely
1. You should estimate the tested voltage amplitude before testing, please do not use if
exceeds the voltage range, because it is likely the probe is damaged.
2. Connect the input lead and output lead to the probe, and then connect the probe to the
oscilloscope or other instruments.
3. Connect the power adapter to voltage probe, the power indicator light turns on green.
Please select proper range based on the tested voltage. When the tested voltage exceeds range, the overload indicator light is on with alarming sound, which can be manually turned off.
4. Please set proper attenuation rate for the oscilloscope or other instruments according to
the probe range. And adjust the oscilloscope sensitivity based on the tested voltage.
5. Connect the probe clips based on needs, start after connecting to the circuits to be tested.
When testing, the probe body should be kept away from high voltage pulse circuits to reduce interference to the probe.
Test Mode (Offset Setting)
Users may enter the test mode to adjust offset based on needs. The probes may be disordered
after years’ of use. The adjustment method is as follows if not at zero:
Please press these two keys, and make short circuits for input terminals.
Turn power on to start, entering test mode, and the overload indicator light is on,
release the two keys.
Go into the high attenuation fact or offset adjustments under the state, press the
offset increase. Press offset decrease.
After the adjustment, press key to switch to low attenuation ratio offset adjustment,
press key offset increases, press , offset dec re as es.
Then press key to exit the test mode, offset adjustment is completed and the
overload indicator light is off, entering into normal operation mode.
Safety Notices
1. Please try to wind up the input leads when testing, this is better for eliminating noise and
to improve the ability of high frequency response. Please view below for the winding method:
10
Equipments
Minimum Requirements
Usages
Oscilloscope
Bandwidth≥100MHZ.Accuracy1.5%, e.g.: SO/DSO4000
Displays probe output
Test bandwidth.
Standard signal
Amplitude:accuracy≤0.75%.rise time≤3ns
AC accuracy.
Generator or calibrator
e.g.: FLUKE/WAVETEK 9100
Common mode rejection
Digital millimeter
Accuracy of not less than 6 and a half
e.g.KEITHLEY 2000
Test the DC accuracy
Insulation pincer clips
Supplied in the accessories
Testing clips BNC adapter 1
BNC-male-to-female-dual show as Figure 1
Test adapter
BNC adapter 2
BNC-male-to-dual binding post show as Figure 2
Test adapter
BNC adapter 3
BNC-female-to-dual binding post show as Figure3
Test adapter
Load terminal
BNC-male-to 50Ωload show as Figure 4
Signal source load
2. It is better not to extend input lead when testing; otherwise it may introduce
more noise. If extra extension lead is necessary, please ensure the extension leads are the same length, and the input frequency is under 10MHz, errors may exist if exceeds 10MHzoutput.
HF transient response with (left) and without (right) extension leads
Performance Verification
The below operation is for performance verification of the electric specification, requirement
for test equipment is shown below:
Figure 1 BNC-male-
to
-female-dual
Figure 2 BNC-female-
to
-dual binding post
Figure 3 BNC-male-
to
-dual binding post
Figure 4 BNC-male-
to
50 Ω load
11
Model
Attenuation Rate
Signal source output voltage
Probe expected output voltage
RSDPB5150
50X
5V
100mV±2mV 500X
50V
100mV±2mV
Model
Attenuation Rate
Signal source voltage, frequency setting
Expected probe Rise time
RSDPB5150
50X
20Vpp70MHz
≤5ns 500X
20Vpp70MHz
≤5ns
Setup
1 Connect power adapter to voltage probe, which turns on green light, to ensure accuracy, test
the probe index after 20minutes.
2 Uncover the red black plastic cover of the BNC-male-to-dual binding post.
DC Accuracy
1) Connect the probe output to the BNC-female-to-dual binding post; plug the two input
terminals of the digital multi-meter into the binding posthole.
2) Connect the probe input to insulation pincer clips, and then connect the calibrator output
and the generator closed, connect the red clip to the positive pole, black clip to negative pole.
3) Set the probe attenuation factor in the first gear.
4) Follow the chart below to set output values for the signal source.
5) Enable the signal output, observe and record the output voltage for the attenuation.
6) Close the signal source output.
7) Switch the probe attenuation factor to the second gear.
8) Repeat step 4~6, and calculate whether is within the accuracy ranges.
Rise Time
1) Configure the fast rise output of the generator for a 50 Ω load. Attach a 50 Ω terminator
to the generator fast-rise output and attach the modified BNC adapter to the terminator.
Attach the differential probe input leads (without attachment accessories) by sliding the banana
plug of the leads onto the binding posts metal sleeves on the modified BNC adapter.
2) Connect the probe output to the oscilloscope, set attenuation factor in the first gear.
3) Refer to the below stable to set standard signal generator.
4) Enable signal source output and record the rise time.
5) Close signal source output.
6) Switch the probe attenuation factor to the second gear.
7) Repeat step 3-5, and calculate whether is in the range.
12
Model
Attenuation Rate
Probe expected output voltage
RSDPB5150
50X
≤1mV
DC Common Mode Rejection Ration (CMRR)
1. Set RDPB5000 series probes at low attenuation ration, respectively (10X, 50X,100X).
2. Set 500V DC voltage for signal source, now the voltage output shutoff.
3. Connect the two probe inputs to 500Vvoltage.
4. Connect the probe output to BNC-female- to- dual binding post, and plug into the two
inputs of the digital millimeter.
5. Enable signal source output, respectively record voltage output values; check with the
following chart to calculate whether is within the ranges.
6. Close the calibrator after completion of the test.
Note: High voltage500 V is used during the testing, please pay attention to personal safety, to reduce voltage fluctuation, be sure to make the calibrator output 500 V high voltages after the completion of all connections.
13
Model
Maximum Input Voltage
Bandwidth
Attenuation Rate
RSDPB4080
800V
50MHz
10X/100X
RSDPB4080 High Voltage Differential Probe
RSDPB4080 Summary
Overview
The RSPDB4080 differential probe provides a safety means for measuring differential
voltage to all models of oscilloscopes. It can convert the high differential voltage (≤800Vpeak) into a low voltage (≤8V) and display this on the oscilloscope. Its bandwidth is up to 50MHz, which is ideal for large power testing, development and maintenance.
The RDPB4080 is designed to operate with the 1MΩ impedance oscilloscopes. When
combined with the 50Ω load, the attenuation will be 2 times as much.
RDPB4080 is recommend for use with our own manufactured PL-10 to expand the measuring with the electricity meter to observe more accurate measurement. The accuracy of oscilloscope is 1% and the electricity meter is about10%.
Electric Specification
(1) Bandwidth:DC-50MHz
(2) Attenuation: X100,X10
(3) Accuracy:+/-1%
(4) Input voltage range (DC+AC PEAK TOPEAK)
≤+/-80Vforx10, (about 30V RMS or DC)
≤+/-800Vforx100, (about 290V RMS or DC)
(5) Permitted max input voltage:
Max differential voltage: 800V (DC+AC PEAK TO PEAK)
Max voltage between each input terminal and ground: 800V RMS
6Input Impedance: Differential: 54MΩ/1.2pF
Between terminal and ground: 27MΩ /2.3pF
7Output voltage:≤+/-8V 8Output impedance:50Ω
14
Reference
Use
Storage
Temperature
+20+30
0+50
-30+70
Relative Humidity
≤70%RH
10%85%RH
10%90%RH
Attenuation Rate
X100
X10
Input Voltage(DC+AC Peak)
±800V
±80V
Operating environmental conditions
(1) Dimensions and weight69x26x165mm.500g (2) Electrical safety to IEC 1010-1 Dual insulation Installation category Degree of Pollution 2 Related voltage or max line-earth6500V RMS CEEN50081-1 and 50082-1
Operating procedure
Connect the probe to the oscilloscope with the BP-250 BNC TO BNC cable. Adjust the vertical zero adjustment of the oscilloscope if necessary. Select the attenuation rate and the vertical deviation of the oscilloscope in accordance with the conversion table below.
Note: The power must be on
Attention: The real vertical deviation is equal to the attenuation rate multiplied by the
range of vertical deviation selected on the oscilloscope. It will be double in the case of a 50Ω
load.
15
Care and Maintenance
1. Keep the probe clean and dry.
2. Please wipe with soft dry cloth to clean. Chemical must not be used to clean.
3. Please put the probe in the package provided, and put it in cool, clean and dry places.
4. Please put the probe in the package provided to prevent shock.
5. Do not forcefully pull the input and output lead to prevent bending, twisted and folding.
16
Loading...