12/09/09 Ver sion No. 00
Instruction Manual
RS-1360A
Digital Humidity / Temperature Meter
FREN
IT
DE
ES
HUMIDITY = 10 ~ 95 %RH
- 4 F ~ 140 F
TEMP = - 20 C ~ 60 C
°°°
°

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01
TABLE OF CONTENTS
Title Page
1. SAFETY INFORMATION .......................................1
2. SPECIFICATIONS..................................................2
3. CONTROLS AND FUNCTIONS.............................4
4. LCD DISPLAY DESCRIPTION ..............................6
5. MEASUREMENT....................................................7
6. BATTERY REPLACEMENT.................................13

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 1
1. SAFETY INFORMATION
Read the following safety information carefully before
attempting to operate or service the meter.
Use the meter only as specified in this manual;
otherwise, the protection provided by the meter may be
impaired.
Environment conditions :
Altitude up to 2000 meters
Relative humidity 95% max.
Operating Temperature 0 to 60℃
Indoor use only
Maintenance & Clearing :
Repairs or servicing that is not covered in this manual
should only be performed by qualified personnel.
Periodically wipe the case with a dry cloth.
Do not use abrasives or solvents on this instrument.
Safety symbols:
Conforms to EU directives

Digital Humidity / Temperature Meter / EN
2 08/15/16 Version No. 01
2. SPECIFICATIONS
Measurement Range :
Humidity : 10% to 95% RH
Temperature : -20 to + 60℃, -4 to +140℉
Resolution : 0.1% RH, 0.1℃, 0.1℉
Accuracy : Humidity : ±3%RH ( at 30 to 95% RH ).
±5%RH ( at 10 to 30% RH ).
Temperature : ±0.8℃, ±1.5℉
Sensor Type : Humidity : precision capacitance sensor.
Temperature : NTC
Response Time : 45% RH → 95% RH ≦3 min
95% RH → 45% RH ≦5 min
Sampling Rate : 3 times per second.
Manual Memory Capacity : 99 sets
Auto Power off : Approx. 30 minutes
Performance : Operating : 0 to 60℃,below 95% RH
Storage : -10 to 60℃,below 70% RH
Power Supply : Single 9V Battery 006P 9V or IEC 6F22,
or NEDA 1604.
Battery Life : Approx. 100 hours

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 3
Size : 270x68x25mm
Humidity & Temp. sensor cap diameter: 20φ
Weight : Approx. 210g (including battery)
Accessories : Instruction manual, Battery, Carrying case.

Digital Humidity / Temperature Meter / EN
4 08/15/16 Version No. 01
3. CONTROLS AND FUNCTIONS
HUMIDITY = 10 ~ 95 %RH
- 4 F ~ 140 F
TEMP = - 20 C ~ 60 C
°°°
°
OPEN
Humidity & Temp. sensor cap
Display
Power and data hold switch

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 5
℃ / ℉ switch
Main display (Temperature / %RH / Dew Point) on
LCD switch.
Manual recording key
Manual recorded data reading key
Battery cover

Digital Humidity / Temperature Meter / EN
6 08/15/16 Version No. 01
4. LCD DISPLAY DESCRIPTION
1. — :Negative sign
7. Measuring value (2nd display)
2.
+
: Low-battery
8. %RH
3. M : Data recording
9. ℃, ℉
4. R : Reading mode
10. DEW : Dew point
temperature
5. No. of manual recorded data
11. Measuring value
(main display)
6.
: Auto power off
12. HOLD : Data hold
13. Memory FULL

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 7
5. MEASUREMENT
5-1 Temperature, humidity and dew point temperature
measurement
1. Slide power switch to “ON” position.
2. Slide temperature units switch to desired ℃ or ℉
position.
3. Slide TEMP, %RH and DEW switch to desired position.
(see below)
Temperature :
%RH :

Digital Humidity / Temperature Meter / EN
8 08/15/16 Version No. 01
DEW :
4. Slide power switch to “HOLD” position. Shows the
“ HOLD ” symbol on the LCD. This is data holding
mode.
5. “ ” symbol indicates that when there’s no operation
on any switches / keys for 30 minutes from powering on,
the meter will auto power off.
5-2 Manual Data Memory and Reading Mode
1. To Memorize the reading
Press “MEMORY” key
momentarily.

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 9
Shows the “
M
” symbol
on the LCD about 0.5
second.
The number of memorized
records increases by 1 to
the maximum of readings
of 99.
When 99 sets, the
“
M
FULL” symbol will
show on the LCD about
0.5 second. No further
readings can be saved.

Digital Humidity / Temperature Meter / EN
10 08/15/16 Version No. 01
2. To Recall and Read Manual Memorized Reading
to read manual recorded data.
The “
R
” symbol will show on
the LCD to indicate you are in
manual read mode.
The LCD shows the data from
the 34th record.
Press “READ” key again to
read previous recorded data.
Press “MEMORY” momentarily
to decrease the displayed
record by 10, continuing to
press the “MEMORY” button
momentarily will cycle the
readings continually.

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 11
To exit the reading mode
Keep Pressing “READ” key
about 2 seconds.
3. To Clear the manual memorized data
Turn off the meter then keep
pressing “MEMORY” key and
turn on the meter until “
M
CLr ” shows on the LCD to
clear all memory data.
5-3 To Cancel auto power off mode
Turn off the meter then keep
pressing “READ” key and turn
on the meter until “ OFF”
symbol shows on the LCD,
then the “
” symbol will
disappear, exit the auto power
off mode.

Digital Humidity / Temperature Meter / EN
12 08/15/16 Version No. 01
5-4 User calibration procedure
Use a standard calibrated humidity meter or humidity
calibration capsules to verify the reading
1. Press and hold down “READ” and “MEMORY” key, then
power until the LCD show the “CAL” symbol.
2. Press “READ” or “MEM” keys until the LCD value is
same as the standard humidity meter.
3. Set power switch to “HOLD” position to store the
calibrated value.
4. It is recommended that the unit is fully
calibrated at least once a year.

Digital Humidity / Temperature Meter / EN
08/15/16 Version No. 01 13
6. BATTERY REPLACEMENT
1. If battery power is not sufficient the LCD will display
+
.
Replacement with one new battery, type 9V, is required.
2. Press battery cover and push in the direction of the arrow
to open.
3. Disconnect battery from instrument and replace with a
standard 9-Volt battery and replace battery cover.
4. Dispose of the used batteries in accordance with local
regulations.