INTRODUCTION
1-1 Unpacking and Inspection
Upon removing your new current transducer from it's packing , you should have the following items:
1. Current Transducer with coiled cable out put plugs.
2. Instruction manual.
3. Carrying Case.
1-2 Front View
Refer to Figure 1 and the following numbered steps to familiarize yourself with the transducer.
1. Transformer Jaws — Designed to pick up the a.c. current following through the conductor.
2. Hand Guard Designed to protect user for safety.
3. Trigger Press the lever to open the trans­ former jaws.
Page 5
4
RED
BLACK
Figure 1
Page 6
5
Meter Safety
Terms marked on Equipment.
ATTENTION Refer to Manual. DOUBLE INSULATION — Protection Class II. DANGER Risk of electric shock.
Symbols in this Manual
This symbol indicates where cautionary or
other information is found in the manual.
SPECIFICATIONS
Current Range : 0.1A to 400A a.c. RMS.
Output Voltage :1mV a.c. per 1Amp a.c.
Working Voltage : 600V CAT. II per IEC 1010-1
Maximum Altitude : 2000m
Operating Temperature : 0°C to 45°C <75%R.H.
Storage Temperature : -20°C ~ 60°C
Type of Sensing : Induction coil sensing for AC
Page 7
6
Temperature Coefficient :
0.2x(Spec Acc'y)/ °C, < 18°C or > 28°C
Maximum Output Impedance : 75Ω. Maximum Jaw Opening : 30mm. Maximum Conductor Size : 29mm diameter. Size : 72mm (W) x 148mm (L) x 36mm(D)
W/O cable.
Weight : 250grams. Indoor use. Maximum Altitude : 2000 meter. Accessories : Manual , Carrying Case.
Precautions and Preparations for measurement
1. Do not apply the voltage to the output plugs.
2. Do not use or store this instrument in a high temperature or high humidity environment and do not store the unit in direct sunlight.
3. Do not measure current before the unit is not combined with DMM.
Page 8
7
4. If the instrument is used near noise generating equipment , be aware that output voltage may become unstable or get more errors.
THIS INSTRUMENT MUST NOT BE USED
ON UNINSULATED CONDUCTORS AT A VOLTAGE GREATER THAN 250V ac/dc.
Electrical Specification
Accuracy is ± (%reading + number of Ampere) at
23°C ± 5 °C, less than 75% R.H.
Range Measure Output Accuracy
400A
3Amp 3mV
±(1.9% + 0.5A)
50 ~ 60Hz
30Amp 30mV
350Amp 350mV
400Amp 400mV
±(3.2% + 1A)
50 ~ 60Hz
Page 9
8
OPERATION
AC Current Measurement
1. Set the DMM at ACV Function and proper range.
2. Connect the plug of transducer with DMM firmly.
3. Press the trigger to open transformer jaws and clamp one conductor only, making sure that the jaw is firmly closed around the conductor and reading the result from the digital display of DMM.
MAINTENANCE
To keep the instrument clean , wipe the case
with a damp cloth and detergent , do not use
abrasives or solvents.
Any adjustment , maintenance and repair
shall be conducted by a service personnel.
Page 10
Africa
RS Components SA P.O. Box 12182, Vorna Valley, 1686 20 Indianapolis Street, Kyalami Business Park, Kyalami, Midrand, South Africa
www.rs -components.com
Asia
RS Components Pte Ltd. 31 Tech Park Crescent Singapore 638040
www.rs -components.com
China
RS Components Ltd. Suite 23 A-C , East Sea Business Centre Phase 2 , No. 618 Yan'an Eastern Road Shanghai, 200001, China
www.rs -components.com
Europe
RS Components Ltd. PO Box 99, Corby, Northants. NN17 9RS, United Kingdom
www.rs -components.com
Japan
RS Components Ltd. West Tower (12th Floor), Yokohama Business Park, 134 Godocho, Hodogaya, Yokohama, Kanagawa 240 Japan
www.rs -components.com
U.S.A
Allied Electronics 7151 Jack Newell Blvd. S. Fort Worth, Texas 76118 , U.S.A.
www.alliedelec.com
South America
RS Componentes Limitada Av. Pdte. Eduardo Frei M. 6001, Centro Empresas El Cortijo Conchali, Santiago, Chile
www.rs -components.com
Loading...