Instrument Declassification
R&S®ZPH
6Instrument Security Procedures 1178.9390.02 ─ 01
The sanitizing procedure is part of the instruments maintenance system which can be
accessed by pressing the front panel buttons [PRESET] and softkey [5] during poweron.
After activating the sanitizing procedure, the following steps occur:
●
A full sector erase command as per manufacturer data sheet is applied to each
sector of the instrument settings and user data section. This explicitly includes sectors which might be declared as defect.
●
Every addressable location of the instrument settings and user data section is
overwritten by a single character.
●
Again, a full sector erase command as per manufacturer data sheet is applied to
each sector of the instrument settings and user data section, including defect sectors.
The "Sanitize internal memory" procedure meets the memory sanitization requirements
specified in the "Clearing and Sanitization Matrix" in Section 14.1.16 of the ISFO "Manual for the Certification and Accreditation of Classified Systems under the NISPOM".
5 Instrument Declassification
Firmware greater or equal 1.30 is required for the instrument declassification.
Before you can remove the R&S ZPH Cable and Antenna Analyzer from a secured
area (for example to perform service or calibration), all classified user data needs to be
removed. You can declassify the R&S ZPH as follows:
1. Turn off the R&S ZPH. This will sanitize the volatile memory.
2. To sanitize the internal Flash memory, perform the following steps:
a) Make sure, that no USB mass memory device is connected.
b) Press the front panel buttons [PRESET] and softkey [5] and hold them while
switching on the instrument again.
After a few seconds, the sanitizing procedure starts.
Sanitizing is indicated by the message "Secure Formatting Flash, please wait!" on
the instrument's screen. The sanitizing procedure takes approximately 8 minutes.
Afterwards, the instrument reboots. Since permanent adjustment values are not located in instrument settings and user data section of the flash, the validity of the
R&S ZPH Cable and Antenna Analyzer’s calibration is maintained throughout the sanitization.
Following these steps removes all user data from the R&S ZPH Cable and Antenna
Analyzer. The instrument can now leave the secured area.