INSTRUCTION MANUAL
IDM67 DIGITAL MULTIMETER
EN IT DE FR JP

3
INTRODUCTION
1-1 Unpacking and Inspection
Upon removing your new Digital Multimeter from its packing, you should have the following items:
1. Digital Multimeter.
2. Test lead set (one black, one red).
3. Instruction Manual.
4. Holster.
1-2 Meter Safety
Terms as Marked on Equipment
ATTENTION — Refer to Manual.
DOUBLE INSULATION — Protection Class II.
DANGER — Risk of electric shock.

4
Symbols in this Manual
This symbol indicates where cautionary or other information is found in the manual.
Fuse
< Battery
1-3 Front Panel
Refer to Figure 1 and the following numbered steps to familiarise yourself with the meter's front panel controls and
connectors.
1. Digital Display — The digital display has a 3200 counts LCD readout with 65 segments analog bar graph, auto
polarity, decimal point, <, =, DC, AC, ;, I , and Unit annunciators.
2. Rotary Switch — Selects the desired function and range.
3. COM Input Terminal — Ground input connector.
4. VΩ Input Terminal — Positive input connector for Volts, Ohms and Diode measurements.

5
5. µA mA Input Terminal — Positive input connector for current measurements. (up to 300mA).
6. A Input Terminal — Positive input connector for current measurements. (up to 10A)
7. Function Switch (Blue) — Press the switch to measure AC Current or DC Current in the current mode, or to
measure continuity or diode check in ;/9 mode.
8. Reset Switch - The meter can be turned back on by pushing "RESET" key switch.
9. Hold Switch - This switch is used to hold measured values for all functions. When pressed the "=" annunciator
is displayed. Conversions are made but the display is not updated.
10. Range Switch (Manual Range) – The "Range" switch is pressed to select manual ranging and to change
ranges. When the "Range" switch is pressed once, the "RANGE" annunciator on the LCD disappears. Press the
"Range" switch to select the appropriate range to be used. Press the "Range" switch and hold for 2 seconds to
return to Auto-ranging.

7
SPECIFICATIONS
2-1 General Specifications
Display : The Liquid Crystal Display (LCD) has a maximum reading of 3200, and 65 segments bar graph.
Polarity Indication : Automatic, positive implied, negative indicated.
Overrange Indication : "OL" or "-OL".
Low Battery Indication : "<" is displayed when the battery voltage drops below operating voltage.
Sampling : 2 times/sec for digit readout. 12 times/sec for analog bar graph readout.
Auto Power Off : Approximately 10 minutes unless the measured value changes within this time.
Temperature Coefficient : 0.15 x (Specified accuracy) / °C, < 18°C or > 28°C.
Power Requirements : Alkaline 1.5V x 2.
Battery Life : Alkaline 900 hours.
Dimensions (HxWxD) : 66mm x 155mm x 34mm, without holster.
85mm x 165mm x 40mm, with holster.
Weight (including battery) : 240 gms, without holster.
360 gms, with holster.
Supplied Accessories: Test leads, battery (installed), operator's manual and holster.

8
2-2 Environmental Conditions
For indoor use.
Maximum Altitude : 2000m.
Installation Category : IEC 1010, 600V CAT II, 300V CAT III.
Pollution Degree : 2
Operating Ambient : 0°C to 50°C, 0 to 80% R.H.
Storage Temperature : -20°C to 60°C, 0 to 80% R.H. with battery removed from meter.