RS ICM A1 Instruction Manual

Instruction Manual ICM A1 AC/DC Clamp Meter
EN
ICM A1
1
Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
Safety Information
- Use the instrument only as specified in this manual or the protection by the instrument might be impaired.
- Always use proper terminals, switch position, and range for measurements.
- Verify the instrument’s operation by measure ing a known voltage. If in doubt, have the instrument serviced.
- Do not apply more than the rated voltage, as marked on the instrument.
- To avoid false readings that can lead to electric shock and injury, replace battery as soon as low battery indicator.
- Do not use the instrument around explosive gas or vapor.
- To reduce the risk of fire or electric shock do not expose the instrument to rain or moisture.
- Individual protective equipment should be used if HAZARDOUS LIVE parts in the installation where measurement is to be carried out could be ACCESSIBLE.
- Do not allow fingers to protrude beyond the Tactile Barrier when fitting or removing the instrument from around a Hazardous Live conductor, as this may cause a shock.
2
Risk of electric shock
See instruction card
DC measurement
AC measurement
Equipment protected by double or reinforced insulation
<
Battery
Earth
Conforms to EU directives
E
Application around and removal from hazardous live conductors is permitted
Do not dispose this electrical or electronic product in domestic household waste.
Caution
Do not expose the instrument to extremes in temperature or high humidity.
Symbols as marked on the instrument and instruction manual
3
Power on / off
Auto Power Off
4
Auto Power Off Disable
Hold down and press
ACA & DCA
Barrier for Hand Guard
earth for the jaw
with respect to
The barrier on the JAW
safe access of the hand­held part, do not hold
in normal use.
over the barrier when
is indicating the limit of
CAT.600V
5
Auto ACA&DCA Detection / ACA&DCA
Due to the high sensitivity of the instrument, perform DCA Zero in the same direction as measurement to avoid interference by an external magnetic field.
6
Data Hold
The display will flash continuously if the measured signal is larger by 50counts than the display reading. However, it cannot detect across the AC and DC Current.
7
The instrument calculates the RMS value for 100ms as detecting a more than 5A current.
DCA ZERO
Inrush Current
8
Low Pass Filter
The cut -off frequency of the low pass filter is about 160Hz with attenuation characteristic of approx-24db/octave.
Backlight on/off
9
Voltsense
The number of dashes displaying on the LCD indicates the electric field intensity. If no indication, voltage could still be present.
10
Battery Replacement
Maintenance
Do not attempt to repair this instrument. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or solvents.
11
Specifications
1-1 General Specifications
Display Count : 6000 count Overrange Display : “OL” or “-OL” Conversion Rate : 2 times / second Dimensions (W x H x D) : 60 x 147 x 31.5 mm Weight : 140 g Power requirement :
LR44 Size Button Battery 1.5V x 2
Battery Life : 20 hours. Maximum Conductor Size : 22 diameter LVD : EN61010-1, EN61010-2-030,
EN61010-2-032
EMC : EN61326-1
Installation category : CAT.. 600V.
CAT. Application field
The circuits is not connected to mains.
The circuits is directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
12
1-2 Environmental Conditions
Indoor Use.
Maximum operating altitude :
2000m (6562 ft)
Operating temperature :
0°C ~ 30°C, ≦80%RH 30°C ~ 40°C, 75%RH 40°C ~ 50°C, 45%RH
Storage temperature :
-20 to +60°C, 0 to 80% RH (no batteries).
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C
Pollution Degree : 2 Shock vibration :
MIL-PRF-28800F for A class 2 Instrument
Drop Protection :
4 Feet Drop to hardwood on concrete Floor
13
1-3 Electrical Specifications
Accuracy is given as ±(% of reading + counts of least significant digit) at 23°C ± 5°C,with relative humidity Less than 80% R.H. ACV and ACA specifications are ac coupled, true R.M.S. The crest factor may be up to 3.0 as 4000 counts. For non-sinusoidal waveforms, Additional Accuracy by Crest Factor (C.F.) : Add 3.0% for C.F. 1.0 ~ 2.0. Add 5.0% for C.F. 2.0 ~ 2.5. Add 7.0% for C.F. 2.5 ~ 3.0. Position Error of Clamp: ±1.5% of LCD reading.
DC Current
Range Resolution Accuracy
60.00A
(1)
0.01A ±(1.5% +10D)
(2)
300.0A 0.1A ±(1.5% + 5D)
All specifications are valid from 5% to 100% of each range. (1) There is less than 0.3A variation as measuring in different directions. (2) Add 10D to accuracy in Auto AC & DC Sense Mode.
14
AC Current
Range Resolution
Accuracy
(50 ~ 100Hz)
Accuracy
(100 ~ 400Hz)
60.00A 0.01A ±(1.5% + 5D) ±(2.5% +5D)
300.0A 0.1A
All specifications are valid from 5% to 100% of each range. Frequency Response : 50 ~ 400Hz (Sine Wave)
Low-pass Filter
Range Resolution
Accuracy
(50Hz/60Hz)
60.00A 0.01A ±(3.5%+5D)
300.0A 0.1A ±(3.5%+5D)
All specifications are valid from 5% to 100% of each range.
Cut-off Frequency (-3dB) : Approx. 160Hz Attenuation Characteristic :
Approx. -24dB / Oct
15
Inrush Current
Range Resolution
300.0A 0.1A
Integration Time : 100ms Trigger Current : 5.0A
VoltSense
Voltage Range : 80V ~ 600V
(At the tip of clamp)
16
Limited Warranty
This meter is warranted to the original purchaser against defects in material and workmanship for 3 years from the date of purchase. During this warranty period, RS Components will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction. This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorized repair, alteration, contamination, or abnormal conditions of operation or handling. Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. RS Components shall not be liable for loss of use of the instrument or Other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you. For full terms and conditions, refer to the RS website.
Africa RS Components SA
P.O. Box 12182, Vorna Valley, 1686 20 Indianapolis Street, Kyalami Business Park, Kyalami, Midrand, South Africa
www.rs-components.com
Asia RS Components Pte Ltd.
31 Tech Park Crescent Singapore 638040
www.rs-components.com
China RS Components Ltd.
Suite 23 A-C , East Sea Business Centre Phase 2 , No. 618 Yan'an Eastern Road Shanghai, 200001, China
www.rs-components.com
Europe RS Components Ltd.
PO Box 99, Corby, Northants. NN17 9RS, United Kingdom
www.rs-components.com
Japan RS Components Ltd.
West Tower (12th Floor), Yokohama Business Park, 134 Godocho, Hodogaya, Yokohama, Kanagawa 240-0005, Japan
www.rs-components.com
U.S.A Allied Electronics
7151 Jack Newell Blvd. S. Fort Worth, Texas 76118 , U.S.A.
www.alliedelec.com
South America RS Componentes Limitada
Av. Pdte. Eduardo Frei M. 6001-71 Centro Empresas El Cortijo Conchali, Santiago, Chile
www.rs-components.com
Loading...