The contents of this manual correspond to firmware version 1.40 and later.
1
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent
personal injury or instrument damage.
● Do not open the instrument casing.
● Read and observe the "Basic Safety Instructions" delivered as
printed brochure with the instrument.
●Read and observe the safety instructions in the following sections.
Note that the data sheet may specify additional operating conditions.
●Keep the "Basic Safety Instructions" and the product documentation
in a safe place and pass them on to the subsequent users.
Riesgo de lesiones y daños en el instrumento
El instrumento se debe usar de manera adecuada para prevenir
descargas eléctricas, incendios, lesiones o daños materiales.
● No abrir la carcasa del instrumento.
● Lea y cumpla las "Instrucciones de seguridad elementales"
suministradas con el instrumento como folleto impreso.
●Lea y cumpla las instrucciones de seguridad incluidas en las
siguientes secciones. Se debe tener en cuenta que las
especificaciones técnicas pueden contener condiciones adicionales
para su uso.
●Guarde bien las instrucciones de seguridad elementales, así como
la documentación del producto, y entréguelas a usuarios
posteriores.
Safety Instructions
Instrucciones de seguridad
Sicherheitshinweise
Consignes de sécurité
1171.1307.42 - 05
2
Gefahr von Verletzungen und Schäden am Gerät
Betreiben Sie das Gerät immer ordnungsgemäß, um elektrischen
Schlag, Brand, Verletzungen von Personen oder Geräteschäden zu
verhindern.
● Öffnen Sie das Gerätegehäuse nicht.
● Lesen und beachten Sie die "Grundlegenden Sicherheitshinweise",
die als gedruckte Broschüre dem Gerät beiliegen.
●Lesen und beachten Sie die Sicherheitshinweise in den folgenden
Abschnitten; möglicherweise enthält das Datenblatt weitere
Hinweise zu speziellen Betriebsbedingungen.
●Bewahren Sie die "Grundlegenden Sicherheitshinweise" und die
Produktdokumentation gut auf und geben Sie diese an weitere
Benutzer des Produkts weiter.
Risque de blessures et d'endommagement de l'appareil
L'appareil doit être utilisé conformément aux prescriptions afin d'éviter
les électrocutions, incendies, dommages corporels et matériels.
● N'ouvrez pas le boîtier de l'appareil.
● Lisez et respectez les "consignes de sécurité fondamentales"
fournies avec l’appareil sous forme de brochure imprimée.
● Lisez et respectez les instructions de sécurité dans les sections
suivantes. Il ne faut pas oublier que la fiche technique peut indiquer
des conditions d’exploitation supplémentaires.
● Gardez les consignes de sécurité fondamentales et la
documentation produit dans un lieu sûr et transmettez ces
documents aux autres utilisateurs.
1171.1307.42 - 05
Customer Support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support
Centers. A team of highly qualified engineers provides telephone support and will work with you to find a
solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz
equipment.
Up-to-date information and upgrades
To keep your instrument up-to-date and to be informed about new application notes related to your
instrument, please send an e-mail to the Customer Support Center stating your instrument and your wish.
We will take care that you will get the right information.
This section provides an overview of the R&S FPC user documentation. You can find it
on the product page at:
www.rohde-schwarz.com/manual/fpc
Getting started manual
Introduces the R&S FPC and describes how to set up and start working with the product. A printed version is included in the delivery.
User manual
The user manual contains the description of all instrument modes and functions. It also
provides an introduction to remote control, a complete description of the remote control
commands with programming examples, and information on maintenance, instrument
interfaces and error messages.
In addition to the R&S FPC user manual, there is a separate user manual for the
R&S InstrumentView software package. This manual contains a description of all features of the R&S InstrumentView software package.
The online version (html format) of the user manual provides the complete contents for
immediate display on the internet.
The user manual is also integrated into the firmware (.chm format). You can export the
file to a memory stick ("Setup" > "User Preferences" > "Export Documentation". After
the export, you can connect the memory stick to a PC and read the .chm file.
Service manual
Describes the performance test for checking the rated specifications, module replacement and repair, firmware update, troubleshooting and fault elimination, and contains
mechanical drawings and spare part lists.
The service manual is available for registered users on the global Rohde & Schwarz
information system (GLORIS, https://gloris.rohde-schwarz.com).
Basic safety instructions
Contains safety instructions, operating conditions and further important information.
The printed document is included in the delivery.
Data sheet and brochure
The data sheet contains the technical specifications of the R&S FPC. It also lists the
options and their order numbers as well as optional accessories.
The brochure provides an overview of the R&S FPC and shows its specific characteristics.
7User Manual 1178.4130.02 ─ 07
R&S®FPC
Documentation Overview
Release notes and open source acknowledgment
The release notes list new features, improvements and known issues of the current
firmware version, and describe the firmware installation.
The open source acknowledgment document provides verbatim license texts of the
used open source software.
www.rohde-schwarz.com/manual/fpc
The open source acknowledgement is also integrated into the firmware (.chm format).
You can export the file to a memory stick ("Setup" > "User Preferences" > "Export Documentation". After the export, you can connect the memory stick to a PC and read
the .chm file.
Application notes, application cards, white papers, etc.
These documents contain information about possible applications and background
information on various topics, see www.rohde-schwarz.com/appnotes.
Calibration certificates
The calibration certificates of your device are available online. Visit the R&S FPC product page and select the item to download the calibration certificate. You will be forwarded to a Gloris page.
https://gloris.rohde-schwarz.com/calcert
Enter the device ID of your R&S FPC and download the certificate. You can find the
device ID either in the "Setup" menu or on the label on the rear panel.
8User Manual 1178.4130.02 ─ 07
R&S®FPC
Regulatory Information
2Regulatory Information
You can access the regulatory information in the firmware of the R&S FPC.
1. Press the "Setup" key.
The R&S FPC opens the "Instrument Setup" menu.
2. Select the "Regulatory Information" menu item in the "WiFi" category with the
"Enter" key.
The R&S FPC shows the regulations it complies with.
9User Manual 1178.4130.02 ─ 07
R&S®FPC
Preparing for Use
3Preparing for Use
The R&S FPC is designated for use in industrial, administrative and laboratory environments. Use the R&S FPC only for its designated purpose. Observe the safety and
usage instructions documented in the operating manual, as well as the operating conditions and performance limits stated in the data sheet.
Make sure to consider the following information before using the R&S FPC for the first
time.
Risk of injury due to disregarding safety information
Observe the information on appropriate operating conditions provided in the data sheet
to prevent personal injury or damage to the instrument. Read and observe the basic
safety instructions provided with the instrument, in addition to the safety instructions in
the following sections. In particular:
●
Do not open the instrument casing.
Risk of instrument damage due to inappropriate operating conditions
Specific operating conditions are required to ensure accurate measurements and to
avoid damage to the instrument. Observe the information on appropriate operating
conditions provided in the basic safety instructions and the instrument's data sheet.
Instrument damage caused by electrostatic discharge
Electrostatic discharge (ESD) can damage the electronic components of the instrument
and the device under test (DUT). Electrostatic discharge is most likely to occur when
you connect or disconnect a DUT or test fixture to the instrument's test ports. To prevent electrostatic discharge, use a wrist strap and cord and connect yourself to the
ground, or use a conductive floor mat and heel strap combination.
10User Manual 1178.4130.02 ─ 07
R&S®FPC
Preparing for Use
Unpacking and Checking the Instrument
Risk of instrument damage during operation
An unsuitable operating site or test setup can damage the instrument and connected
devices. Ensure the following operating conditions before you switch on the instrument:
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are correctly connected and are not overloaded.
EMI impact on measurement results
Electromagnetic interference (EMI) may affect the measurement results.
To suppress generated electromagnetic interference (EMI):
●
Use suitable shielded cables of high quality. For example, use double-shielded RF
and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
3.1Unpacking and Checking the Instrument
Unpack the R&SFPC carefully and check the contents of the package.
●
Check if all items listed on the delivery note, including the getting started manual,
are included in the delivery.
●
Check the R&S FPC for any damage.
If the contents are damaged, immediately contact the carrier who delivered the
package.
●
Keep the box and packing material.
11User Manual 1178.4130.02 ─ 07
R&S®FPC
Preparing for Use
Placing or Mounting the Instrument
Risk of instrument damage during transportation and shipment
Insufficient protection against mechanical and electrostatic effects during transportation
and shipment can damage the instrument.
●
Always make sure that sufficient mechanical and electrostatic protection is provided.
●
When shipping an instrument, use the original packaging. If it is not available, allow
for sufficient padding to prevent the instrument from moving around inside the box.
Pack the instrument in antistatic wrap to protect it from electrostatic charging.
●
Secure the instrument to prevent any movement and other mechanical effects during transportation.
Packing material
Retain the original packing material. If the instrument needs to be transported or shipped later, you can use the material to protect the control elements and connectors.
3.2Placing or Mounting the Instrument
The R&S FPC is designed for use under laboratory conditions, either on a bench top or
in a rack.
Benchtop operation
If the R&S FPC is operated on a bench top, the surface should be flat. The instrument
can be used in horizontal position, standing on its feet, or with the support feet on the
bottom extended.
Mounting the R&S FPC in a rack
The R&S FPC can be installed in the 19" rack mount kit R&S ZZA-FPC1 (order no.
1328.7080.02). The installation instructions are part of the adapter kit.
12User Manual 1178.4130.02 ─ 07
R&S®FPC
Preparing for Use
Turning the R&S
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted.
This can cause damage or injury.
●
Fold the feet completely in or out to ensure stability of the instrument. Never shift
the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 200 N.
F
max
FPC on and off
3.3Connecting AC Power
The AC power connector on the rear panel of the R&S FPC allows you to connect it to
the primary power supply.
Included in the delivery of the R&S FPC are several common power plug types.
1. Select the cable with the plug type you need and firmly connect it to the R&S FPC.
2. Connect the AC plug to the power outlet to supply the R&S FPC with power.
The R&S FPC is assembled in line with the specifications for safety class
EN61010. Therefore, it may only be connected to an outlet that has a ground contact.
The AC power supply has the following characteristics.
●Line voltage: 100 V AC to 240 V AC
●Line frequency: 50 Hz to 60 Hz; 400 Hz
●Current: 0.6 A to 0.4 A
3.4Turning the R&S FPC on and off
After you have established a connection to the power supply, you can turn on the
R&S FPC.
13User Manual 1178.4130.02 ─ 07
R&S®FPC
Preparing for Use
Turning the R&S FPC on and off
Turning on the R&S FPC
► Turn on the main AC power switch on the rear panel of the R&S FPC (position "I").
The instrument is now supplied with AC power.
●"Power" key is highlighted orange: R&S FPC is in standby mode (main AC
power switch is in position "I").
●"Power" key is highlighted green: R&S FPC is running and ready for operation.
Turning off the R&S FPC
► Turn off the main AC power switch on the rear panel of the R&S FPC (position
"O").
The instrument is no longer supplied with AC power.
Changing the AC supply fuse
Only fuses of the type 2A T IEC60127-2/V should be used.
1. Disconnect the power cable.
2. Open the flap covering the voltage selector using a small screwdriver (or similar).
3. Remove the cylinder labeled with the nominal voltages. Remove the fuse and
install the new one. Reinsert the cylinder so that the value visible through the hole
in the cover flap is the same nominal voltage as before.
4. Close the flap.
14User Manual 1178.4130.02 ─ 07
R&S®FPC
Instrument Tour
Front Panel
4Instrument Tour
The R&S FPC has various connectors on the front and rear panel.
4.1Front Panel
1632574
Figure 4-1: Front panel of the R&S FPC
1 = USB ports (type A)
2 = Headphone jack
3 = Softkeys
4 = Signal source output
5 = Function keys and alphanumeric keypad
6 = RF input
7 = Power switch
Instrument damage caused by cleaning agents
Cleaning agents contain substances such as solvents (thinners, acetone, etc.), acids,
bases, or other substances. Solvents can damage the front panel labeling, plastic
parts, or screens, for example.
Never use cleaning agents to clean the outside of the instrument. Use a soft, dry, lintfree dust cloth instead.
Power switch
The power switch turns the R&S FPC on and off when it is supplied with power.
For more information, see Chapter 3.4, "Turning the R&S FPC on and off", on page 13.
15User Manual 1178.4130.02 ─ 07
R&S®FPC
Instrument Tour
Front Panel
RF input
The RF input with an impedance of 50 Ω allows you to connect a DUT to the
R&S FPC. Typically, you connect the DUT with a cable and an appropriate connector
(for example a male N connector).
The frequency range of the RF input is specified in the datasheet.
The attenuation range is between 0 dB and 40 dB.
Risk of instrument damage
Make sure not to overload the RF input and keep within the maximum allowed signal
levels. Refer to the datasheet for the maximum allowed signal levels.
A DC input voltage of 50 V must never be exceeded.
Signal source output
Available on the R&S FPC1500.
The "Signal Source" output allows you to generate a signal that can be fed into the
DUT or other external accessories like frequency dividers or amplifiers. You can connect the DUT or accessories with a cable that has a male N connector.
Risk of instrument damage
The maximum supported reverse power is 23 dBm. Make sure that this value is not
exceeded to avoid damage to the R&S FPC.
Headphone jack
The female headphone jack allows you to connect headphones (or external speakers)
with a miniature jack plug.
You can control the output voltage with the volume control integrated into the firmware.
Refer to the user manual for details.
If you connect headphones or external speakers, the R&S FPC automatically turns off
the internal speaker.
Risk of hearing damage
Before putting on the headphones, make sure that the volume setting is not too high to
protect your hearing.
16User Manual 1178.4130.02 ─ 07
R&S®FPC
Instrument Tour
Rear Panel
USB ports (type A)
The two USB 2.0 ports on the front panel (type A) allow you to connect devices like
memory sticks.
Function keys and alphanumeric keypad
The function keys provide access to the measurement settings and functions. The
alphanumeric keypad allows you to enter alphanumeric data if necessary.
Refer to the user manual for a comprehensive description of the function keys.
Softkeys
The softkeys allow you to access measurement settings and functions.
Softkeys are dynamic. A different list of softkeys is displayed depending on the
selected function key. A list of softkeys for a certain function key is also called a menu.
Softkeys can either perform a specific function or open a dialog box.
Refer to the user manual for a comprehensive description of the function keys.
4.2Rear Panel
Figure 4-2: Rear panel of the R&S
1 = Trigger input / external reference
2 = LAN
3 = USB port (type B)
4 = Power supply
FPC
1432
17User Manual 1178.4130.02 ─ 07
R&S®FPC
Instrument Tour
Rear Panel
Power supply
The AC power supply and main power switch are located in a unit on the rear panel of
the instrument.
The main power switch has the following states.
●
Position "1": The instrument is supplied with power.
●
Position "0": The instrument is disconnected from the power supply.
Trigger input / external reference
This female BNC connector allows you to connect an external trigger signal or an
external reference signal.
When you are using the connector as a trigger input, you can trigger measurements
with an external trigger. For more information about triggered measurements, refer to
the user manual.
Alternatively, you can use the connector to connect a 10 MHz reference signal to synchronize the frequency with the external reference. Note that the reference signal must
be stronger than 0 dBm.
LAN
The LAN interface allows you to connect the R&S FPC to a local network for remote
control, printouts or data transfer. The assignment of the RJ-45 connector supports
twisted-pair category 5 UTP/STP cables in a star configuration (UTP stands for
unshielded twisted pair, and STP for shielded twisted pair).
USB port (type B)
The USB port (type B) allows you to connect the R&S FPC to a computer and establish
a remote control connection.
18User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Manual Operation
5Operation
You can operate the R&S FPC manually with the keys available on its front panel, or
remotely from another computer.
The front panel of the R&S FPC provides various types of buttons (or keys) on its front
panel that allow you operate it manually.
Using function keys
Function keys are the group of keys with a light grey background.
Figure 5-1: Function keys
Pressing one of the function keys opens a list of parameters on the display that contains settings that thematically belong together. This list of parameters is also referred
to as a menu.
For example, pressing the "Freq" key opens the "Frequency" menu that contains
parameters to configure the frequency characteristics of the measurement.
The contents of the menu depend on the measurement application that you are using.
For example, the "Frequency" menu in the spectrum application looks different than the
"Frequency" menu of the analog demodulation application.
Exceptions are the "Preset" key and the camera key. These do not open a menu, but
initiate an action (performing a preset and taking a screenshot, respectively).
You can modify a specific parameter (menu item) with one of the softkeys.
Using softkeys
Softkeys are the group of keys to the right of the display.
Pressing one of the softkeys allows you to modify the properties of the parameter that
is displayed next to it. The parameter assigned to each softkey is variable and depends
on the function key that you pressed last.
For example, when you press the "Freq" function key in the spectrum application, the
first item in the menu is the "Center Frequency". When you press the softkey next to
19User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Manual Operation
that menu item, the R&S FPC opens an input field that allows you to change the center
frequency.
Depending on the parameter, pressing a softkey has one of the following effects.
●
It opens an input field that allows you to modify a parameter by entering a specific
value. When an input field is open, the corresponding menu item turns orange.
Example: Pressing the "Freq" key and selecting the "Center Frequency" menu item
opens an input field.
Figure 5-2: Softkey turns orange when an input field is open
Pressing the softkey again, closes the input field. If you have changed the value in
the input field, the new value is adopted.
Figure 5-3: Softkey turns grey when the input field is closed
●
It opens a submenu which contains another list of parameters that thematically
belong together.
A softkey that opens a submenu shows an arrow (">") in its label.
Example: Pressing the "Freq" key opens the "Frequency" menu. The second softkey in the list ("Center Freq Step Size") opens a submenu.
Figure 5-4: Softkey with an arrow opens a submenu
●
It initiates some kind of action.
Example: Pressing the "Span" key and selecting the "Full Span" menu item
restores the full frequency span.
●
It changes the state of something (like a toggle button).
The corresponding menu item turns light blue when its state is "on".
Example: Pressing the "BW" key and selecting the "RBW: Auto" menu item turns
automatic selection of the resolution bandwidth on and off.
Figure 5-5: Softkey in the "on" (left) and "off" (right) state
●
It opens a dialog box that contains additional settings.
Example: Pressing the "Setup" key and selecting the "Instrument Setup" menu item
opens a dialog box to configure general instrument properties like date and time.
Using input keys
Input keys are the group of keys with a dark grey background.
20User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Manual Operation
Figure 5-6: Input keys
The input keys only have an effect when any kind of input field is active on the display.
There are several types of input keys.
●
Alphanumeric keys
The alphanumeric keys allow you to enter numeric values or characters: the numbers from 0 to 9, the alphabet, a minus and plus sign and a dot. Pressing the number 0 repeatedly allows you to enter an empty space and an underscore.
Press the corresponding key to enter a numeric value. In case of numeric values,
each key covers just the number that is printed on it. You can enter negative values
with the minus sign key and enter values that contain decimal places with the dot
key.
Example: Press the "Freq" key and select the "Center Frequency" menu item.
Enter a number with the alphanumeric keys.
If you need to enter a character (for example for file names), the key assignment
changes. Each key covers one number and more than one character with the first
choice being a lower case character. If you need to enter a character, press the key
several times until the character you require is selected.
Example: Press the "Freq" key and select the "User Preference" menu item. In the
dialog box, select the "Site Name" menu item and enter some characters (for
"Home", for example, press 6x"4", 3x"6", 1x"6" and 2x"9").
An empty space or special characters like _ ! @ # $ % ^ & * = ? are available by
pressing the "0" key several times.
●
Unit keys
The unit keys confirm entries of values that have a unit. The units assigned to each
key are indicated on the key itself or next to the key.
Example: Press the "Freq" key and select the "Center Frequency" menu item. To
define a center frequency of 100 MHz, enter "100" and confirm with the [MHz] key.
●
Escape key
The "Esc" key cancels entries and closes the input field. The firmware does not
adopt the entry you have made.
●
Backspace key
The "Backspace" key ( ) moves the cursor one position backwards and deletes
the character / number that was in that place.
Tip: If an input field is visible, the backspace key toggles between the last two
entries. Using this feature, you can, for example, quickly toggle between two center
frequencies.
●
"Enter" key ("✓")
The "✓" key confirms an entry you have made.
21User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Screen Layout
If you confirm a value that has a unit with the "✓" key, the R&S FPC always uses
the highest possible unit (for example GHz or s).
In addition, the group of input keys also contains cursor keys and the rotary knob.
Using the cursor keys
The cursor keys do several things, depending on the situation when you use them.
●
The cursor keys navigate through dialog boxes.
●
The up and down keys increase or decrease any kind of numeric value if an input
field is active.
The cursor keys change numeric values with a fix step size.
●
The up and down keys move markers around.
The step size is fix.
●
The left and right keys move the cursor in an input field in the corresponding direction.
Using the rotary knob
The rotary knob does several things, depending on the situation when you use it.
●
The rotary knob works like a cursor key in dialog boxes. In that case, you can navigate to one of the items with the rotary knob. If the dialog box covers more than
one screen page, it also scrolls through the dialog box.
Turning it to the right corresponds to a downward movement. Moving it to the left to
an upward movement.
●
The rotary knob increases or decreases any kind of numeric value if an input field
is active.
Turning it to the right corresponds to an increase, turning it to the left to a decrease
of a numeric value.
The rotary knob changes numeric values with a fix step size.
●
The rotary knob moves markers around in the diagram area.
To do so, you have to select a marker first. The step size is fix.
●
Pressing the rotary knob has the same effect as pressing the "✓" key and confirms
an entry or selection.
5.2Screen Layout
The following image shows the screen layout in the spectrum application. It shows all
elements that are the same in all applications of the R&S FPC. If an application or
measurement has a completely different layout or elements, the differences are noted
in the description of those.
22User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Screen Layout
123
4
5
6
7
8
9
101112131410
Figure 5-7: Screen layout in the spectrum application
1 = Title bar
2 = Input field
3 = Softkeys
4 = Measurement settings
5 = Trace information
6 = Marker information
7 = Limit line information
8 = Diagram area
9 = X-axis information
10 = Sweep count
11 = Trace
12 = Delta marker
13 = Limit line
14 = Marker
15 = Star icon
Title bar
The title bar shows the application that you are currently using as well as the date and
time.
Input field
An input field opens when you select a feature that requires you to enter a value. The
header of the input field shows the name of the softkey menu you have currently
selected. If no input field is open, the R&S FPC just shows the name of the softkey
menu you have currently selected.
Softkey menu
Softkeys provide access to measurement settings and functions. They are dynamic: a
different list of softkeys is displayed depending on the selected function key. A list of
softkeys for a certain function key is also called a menu.
23User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Screen Layout
Softkeys can either perform a specific function or open an input field or dialog box.
Measurement settings bar
The measurement settings bar contains information about the current measurement
settings. A blue dot in front of a setting indicates that automatic selection for that setting has been turned off.
Trace information
The trace information indicates the current trace configuration for each active trace
(trace color, trace number, trace mode and detector).
Table 5-1: Overview of trace modes
AbbreviationDetector
ClrwClear / write trace mode
MaxMax hold trace mode
MinMin hold trace mode
AvgAverage trace mode
ViewView trace mode
No labelBlank trace mode (trace is off)
Table 5-2: Overview of detectors
AbbreviationDetector
APAuto peak detector
PkMax peak detector
MiMin peak detector
RmRMS detector
SaSample detector
QPQuasipeak detector (receiver only)
Marker information
The marker information indicates the position of each active marker. If a marker function is active, it also shows the result for that marker function.
Limit line information
The limit line information indicates the characteristics of any active limit line (type of
limit line, name of limit line etc.), including the result of a limit check (pass or fail).
Diagram area
The diagram area contains the graphical representation of the measurement results.
24User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Remote Operation
The diagram area can contain additional information.
●
The sweep count indicates the process of trace averaging. The first number is the
current measurement, the second number the total number of measurements included in the trace averaging.
●
The trace indicates the characteristics of the measured spectrum.
●
The position of markers or delta markers is indicated by a straight horizontal line
(markers) or a dotted line (delta markers). In addition, the marker position is also
indicated by an icon at the bottom of the diagram area.
●
The position and shape of a limit line is indicated by a red vertical line.
●
The star icon (
current measurement settings. The star will appear, for example, when you change
settings after a single measurement is done.
X-axis information
The x-axis information indicates the current scale and range of the x-axis.
) indicates that the displayed results no longer correspond to the
5.3Remote Operation
There are various tools that allow you to operate the R&S FPC remotely from another
computer (for example PC or tablet).
To use the R&S FPC this way, you have to establish a connection between both devices via the LAN, Wi-Fi or USB interfaces of the R&S FPC.
Remote desktop (R&S InstrumentView)
R&S InstrumentView is a software that contains various useful tools, including a
remote desktop application. You can download the software, including a user manual,
from the Rohde & Schwarz website.
1. Install the software on a desktop PC or notebook.
2. Connect the R&S FPC to the computer with a LAN cable or a USB cable.
3. Start the software on the computer (use the desktop icon, or search for "InstrumentView" in the start menu).
4. Establish a connection to the R&S FPC in the "Connection Manager".
a) Select the "LAN" tab to configure the LAN properties.
b)
Select the pencil icon ( ).
c) Enter the IP address of the R&S FPC you want to connect to.
d) Confirm with the "Connect" button.
You can connect to as many instruments as you like (including other supported
instruments like R&S FSH or R&S FPH). To do so, simply "Add Another
Device" and define the network properties for them.
25User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Remote Operation
Alternatively, you can establish a connection via USB in the "USB" tab of the "Connection Manager".
5. Select "Instrument" > "Remote Display".
The software shows a live simulation of the R&S FPC, including all keys available
on the front panel.
Everything you do in the remote desktop application is also done on the connected
R&S FPC.
For a comprehensive description of the R&S InstrumentView software, refer to its
user manual.
Tip: Lab display. The lab display allows you to access several instruments at the
same time via the remote desktop. For more information, refer to the documentation of the R&S InstrumentView software.
Remote desktop (R&S MobileView)
R&S MobileView is an app that allows you to remote control the R&S FPC from a tablet or smartphone. It basically works like the remote desktop application built into the
R&S InstrumentView software.
You can download the app from the Google Play Store (Android based devices) or the
AppStore (iOS devices). A search for "Rohde Schwarz MobileView" should find the
app.
1. Start the R&S FPC.
2. Configure your tablet or phone as a hotspot for an ad-hoc connection (usually in
the phone settings).
3. Connect the R&S FPC to the mobile device.
a) Press the "Setup" key to open the instrument setup menu.
b) Select "WiFi": "On" to turn on the Wi-Fi functionality.
c) Select the "Connect" menu item to start a search for Wi-Fi networks in the
area.
The R&S FPC starts a search for all Wi-Fi networks it can find (mobile devices,
routers etc.).
d) Select the network you want to connect to.
e) Enter the password of the network with the alphanumeric keys. The network
password should be displayed in the hotspot settings of the mobile device.
The R&S FPC establishes the connection.
4. Start the R&S MobileView app on your mobile device.
The app shows a list of all R&S FPC devices that are connected to the hotspot.
5. Select the R&S FPC you want to connect to.
The app shows a live simulation of the R&S FPC, including all keys available on
the front panel.
Everything you do in the app is also done on the connected R&S FPC.
26User Manual 1178.4130.02 ─ 07
R&S®FPC
Operation
Remote Operation
Note: If you have a device that does not support ad-hoc connections, you have to
connect both devices to an existing wireless infrastructure (for example through a
router). When you connect both the mobile device and the R&S FPC to the infrastructure, you can use the app just like with an ad-hoc connection.
Remote operation (SCPI commands)
Remote operation with SCPI commands is a method to automate measurements by
writing scripts or programs that contain all instructions required for the measurement.
When you start the program, the measurement is done automatically from start to finish. Manual configuration of a measurement with function keys, softkeys or apps is no
longer necessary.
Remote operation is thus a useful tool for measurements that you perform repeatedly
and for which you know the configuration and for which the configuration is always the
same. Testing devices during production is an example for such measurement tasks.
Each device needs to be tested under the same conditions.
You can operate an instrument remotely using SCPI commands (Standard Commands
for Programmable Instruments). Sending a SCPI command initiates one or more
actions on the instrument, for example change the state of a parameter or start a measurement. Usually, each parameter available in the firmware has a corresponding SCPI
command.
Example: The SCPI command to change the center frequency is
[SENSe:]FREQuency:CENTer.
When you send a succession of these commands, you can configure and perform
complete measurements.
There are several tools that allow you to connect to an instrument, send an individual
(or more) command(s) and compile automated test sequences.
One of these tools is delivered with the R&S VISA library that is available for download
on the Rohde & Schwarz website (search term: "r&s visa"). The VISA library itself provides input and output functions that are required to communicate with the instrument.
Another tool with more complex scripting functions is R&S Forum, also available for
download on the Rohde & Schwarz website (search term: "forum"). It allows you to run
and edit example script sequences and to write your own script files, as well as on-thefly remote control of instruments. Script files can range from simple command sequences (Winbatch syntax) to complex programs using the programming language Python.
User documentation and background information is available for both tools. If you want
to learn more about their features, refer to the corresponding user manuals and application notes.
The following basic introduction to remote control is based on the tool that is delivered
with the R&S VISA.
1. Install R&S VISA on a desktop PC or notebook.
2. Connect the R&S FPC to the computer with a LAN cable or a USB cable.
3. Start the RsVisaTester software (search for "tester" in the start menu). Use the 32bit or 64-bit version, depending on your operating system.
27User Manual 1178.4130.02 ─ 07
Loading...
+ 389 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.