Infrared Thermometer
Remote Temperature Measurement
(
Non-Contact )
1327
INSTRUCTION MANUAL

TABLE OF CONTENTS
TITLE PAGE
INTRODUCTION....................................................................... 1
Ⅰ. SAFETY INFORMATION.....................................................
1
Ⅱ. FEATURES................................................................. ........ .
3
Ⅲ. SPECIFICATIONS...............................................................
3
Ⅳ. GENERAL LAYOUT AND OPERATING CONTROLS........
5
Ⅴ. LCD DISPLAY DESCRIPTION............................................
7
Ⅵ. EMISSIVITY ADJUSTMENT ...............................................
8
Ⅶ. TEMPERATURE MEASUREMENT...................................
11
Ⅷ. SETTING MODE................................................................
13
Ⅸ.BATTERY REPLACEMENT...............................................
15

1
INTRODUCTION
The RS 1327 hand-held Infrared Thermometer is easy to use and
operate. The thermometer has a laser to assist in sighting the sensor
for making measurements. The Liquid Crystal Display has a backlight
function which may be turned on to assist measurements made in a
dark place. Other features include reading auto-hold, auto power-off,
maximum and minimum readings, all of which can be selected for
maximum operating convenience.
The instrument does not rely on physical contact with the object
whose temperature is to be taken, but measures the radiated
infrared energy from the object. This allows temperature
measurements to be made of items such as rotating machinery,
bus-bars carrying electric current, or inaccessible objects which
would be difficult or hazardous to touch.
Ⅰ. SAFETY INFORMATION
1. Read the following safety information carefully before attempting
to operate or service the meter.
Laser Safety Warning
LASER RADIATION DO
EN 60825-1:1994/A11:1996
If the MEAS button is pressed while the symbol is shown on the
LCD, the meter will radiate the lase r. To prevent injury , do not look or
stare into the beam. Do not point the l aser directly a t eyes or
indirectly off reflective surfaces.
˙ If the target object has a smooth surfa ce, the laser may be reflected
and create a hazard.
˙ Do not radiate the laser at flammable gasses or other possible

2
sources of ignition.
2. Precautions:
a). Do not operate the thermometer in the presence of strong
electrical or magnetic f i elds, as this may prevent c orrect
operation.
b). Use and store the thermometer away from direct sunlight
or other strong source of light, high temperatures, high
humidity or dust.
c). If the thermometer has been kept in area of high or low
temperature, wait until it has reached a stable temperature
before use.
d). Condensation may form on the focal lens if the
thermometer is moved quickly from a cold to a hot
environment. Before taking measurements, wait for the
condensation to dissi pate.
e). Do not touch the foc al lens.
3. Environmental conditions:
a). Altitude up to 2000 meters.

3
b). Relatively humidity: 80% max.
c). Operating ambient t emperature: 0 to 50
o
C
4. Maintenance & Clearing:
a). Repairs or servicing should only be carried out by
competent persons using the correct parts, test equipment
and procedures. If repair or servicing is required, contact
the supplier or RS Components. The address is given at
the end of these instructions.
b). Periodically clean the case with a dry cloth. Do not use
abrasives or solvents.
5. Safety symbols:
Complies with the EMC Directive 89/336/EEC
Ⅱ. FEATURES
¾
o
C/ oF Selectable.
¾ Backlit LCD display.
¾ Sighting laser.
¾ Auto-hold function.
¾ Maximum/minimum reading recorder function.
¾ Auto power-off.
Ⅲ. SPECIFICATIONS
3-1 General Information:
Display: Backlit LCD display.
Auto power-off time: Approx.15 sec.
Data memory capacity: 50 set. (Direct readin g on
LCD display)

4
Over-range indication: “OL” or “-OL”.
Low battery indication: The
+
symbol is displayed on
LCD when the battery voltage drops
below the operating voltage.
Power supply: Single 9V battery 006P 9V,
IEC6F22, or NEDA1604.
Battery life: Approx. 100 hours (with las er
pointer and backlight off)
Operating temperature : 0
o
C to 50 oC (32 oF to 122 oF),
Operating humidity: Below 80%RH.
Storage temperature –10
o
C to 60 oC (14 oF to 140 oF),
Humidity: Below 70%RH.
Dimensions: 172(L)×118( W)×46(H) mm
6.8(L)×4.6(W)×1.8(H) inches.
Weight: Approx. 220g with battery.
Accessories: Instruction manual, battery.
3-2 Electrical specifications:
Measuring range: -20
o
C to 500 oC (-4 oF to 932 oF)
Resolution: 0.1
o
C, 0.2 oF
Accuracy: ± 2% of reading or ± 2
o
C or 4 oF
(whichever is greater).
Temperature
coefficient: 0.1 times the applicable accuracy
specification per
o
C from 0 oC to
18
o
C and 28 oC to 50 oC (32 oF to