Rossmax X9 User Manual

Model: X9
Blood Pressure Monitor
www.rossmax.com
Introduction
Blood pressure measurements determined with X9 are equivalent to those obtained by a trained observer using cu/stethoscope auscultation method, within the limits prescribed by the American National Standard, Electronic or Automated Sphygmoma­nometers. This unit is to be used by adult consumers in home, physicians’ oces, hospi­tals, clinics and other medical facilities. The patient is an intended operator. Do not use this device on infants or neonates. X9 is protected against manufacturing defects by an established International Warranty Program. For warranty information, you can contact the manufacturer, Rossmax International Ltd.
Attention: Consult the accompanying documents. Please read this manual care-
fully before use. For specic information on your own blood pressure, contact your physician. Please be sure to keep this manual.
PARR(Pulse Arrhythmia) Technology
Pulse Arrhythmia (PARR) technology specically detects the existence of pulse arrhyth­mia, including atrial brillation (AFib), Atrial and / or Ventricular Premature Contractions (PC), Tachycardia (TACH), and Bradycardia (BRAD). Pulse Arrhythmia may be related to cardiac disorders, needs medical attention and thus early diagnosis is of paramount importance. The PARR technology detects arrhythmia during regular blood pressure checks without any additional user skills, user interaction and measurement prolonga­tion. Beside the blood pressure diagnosis a specic pulse arrhythmia diagnosis is pro­vided with PARR. Note: The PARR detection of AFib, PC, TACH and BRAD is provided with a clinically
proven high detection probability [1]. However, the sensitivity and specicity is limited, thus most, but not all pulse arrhythmia will be detected and displayed. In certain patients with uncommon clinical conditions the PARR technology may not be able to detect pulse arrhythmia. This partly comes from the fact that some arrhythmia can only be found with an ECG diagnosis, but not with a pulse diagno­sis. Thus PARR is not meant to replace any medical ECG diagnosis by your doctor. PARR provides an early detection of certain pulse arrhythmia, which inevitably need to be presented to your doctor in charge.
Remark: [1] Clinical Investigation of PARR - A new Oscillometric Pulse Arrhythmia Type
Discriminating Detection Technology
Atrial Fibrillation Detection (AFib)
The upper chambers of the heart (the atria) do not contract, but quiver and thus blood is driven irregularly and with lower eciency into the ventricles. Subsequently irregular heartbeats occurs, which mostly are associated with a fast, yet highly instable heart rate.
This condition is associated with a higher risk for the formation of cardiac blood clots. Amongst others, they may elevate the risk of brain strokes. Beside this atrial brillation may contribute to the severity of a chronic or acute heart failure condition and may be associated with other heart-related complications. Age dependent, about 10 %- 20 % percent of patients who suer from an ischemic stroke also suer from atrial brillation. Atrial brillation most often initially occurs with temporary periods of arrhythmia and may progress to a permanent state of this disorder in the course of time. No matter, whether you intent to safeguard yourself from an undetected AFib state, or you meas­ure during an ongoing period of active atrial brillation, or you measure in between pe­riods of AFib, the PARR technology can be applied at any of these conditions. This unit is able to detect Atrial brillation (AFib). The ARR and AFib icons ( right after the measurement if Atrial Fibrillation was detected. Note: It is strongly recommended, that you consult your physician, if either the AFib icon
occurs newly for several times, or, if your AFib is known to your doctor, but the incidence of AFib readings changes over time. Your doctor will then be able to provide all required medical test and possible therapeutic procedures.
Note: The presence of a cardiac pacemaker may impair the AFib detection by PARR.
) are displayed
Premature Contraction Detection (PC)
Extra abnormal heartbeats generated in irregular excitation sites of your heart, either in the atria (PAC), the ventricle (PVC) or the cardiac conduction nodes (PNC). These extra beats may disrupt your regular rhythm, they may come in early or cause a signicant pauses regarding your perceivable pulse. This is called palpitations, which can be felt in your chest. They may occur as isolated, single events, as a series of irregular pulses or can be distributed all over your pulse beats. If they are not related to mental stress, or acute demanding physical load, they may be a marker for a multitude of cardiac disorders. Some of these disorders go along with an elevated risk prole for ischemic events, either in the heart (e,g, coronary heart disease) or outside the heart, e.g. an elevated risk for a stroke. Some PCs may indicate on valvular or myocardial disorders and become very important if a myocarditis (infection of the heart muscle) is suspected. This unit is able to detect premature contractions. The ARR and PC icons ( ) are displayed right after the measurement if premature contractions have been detected. Note: It is strongly recommended, that you consult your physician, if either the PC icon
occurs newly for several times, or, if your PC is known to your doctor, but the inci­dence of PC readings changes over time. Your doctor will then be able to provide all required medical test and possible therapeutic procedures.
2 3
Tachycardia Detection (TACH)
A fast heart rate of higher than 100 beats per minute (bpm) in adults. Unless being caused by physical or mental stress, a tachycardia may be an indicator for both cardiac (e.g. Coronary heart disease, valvular disorder), or extra-cardiac disorders (e.g. hyperthy­roidism, fever, hypoxemia), as well as medication and stimulant substance side eects (e.g. caeine). The unit is able to detect Tachycardia (TACH). The ARR and TACH icons (
) are displayed right after the measurement if tachycardia has been detected.
Note: It is strongly recommended, that you consult your physician, if either the TACH
icon occurs newly for several times, or, if your TACH is known to your doctor, but the incidence of TACH readings changes over time. Your doctor will then be able to provide all required medical test and possible therapeutic procedures.
Bradycardia Detection (BRAD)
A slow heart rate of less than 55 beats per minute (bpm) in adults. Unless not genetically determined or subsequent to a high cardiac endurance training adaptation, bradycar­dia may be related to multitude of cardiac disorders (e.g. valvular heart disease, heart failure) or extra-cardiac disorders (e.g. hypothyroidism, electrolyte imbalance) or medi­cations (e.g. beta-receptor blocker). This unit is able to detect Bradycardia (BRAD). The ARR and BRAD icons ( ) are displayed right after the measurement if bradycar­dia was detected. Note: It is strongly recommended, that you consult your physician, if either the BRAD
icon occurs newly for several times, or, if your BRAD is known to your doctor, but the incidence of BRAD readings changes over time. Your doctor will then be able to provide all required medical test and possible therapeutic procedures.
Pulse Arrhythmia Detection (ARR)
Once the occurrence of pulse arrhythmia has been detected in the course of your blood pressure measurement, the icon ARR is displayed. In the case, that the found pulse ar­rhythmia can be specied by the PARR technology, the ARR icon is accompanied by the specically detected type of arrhythmia, e.g. PC, AFib, TACH or BRAD. Once the kind of found pulse arrhythmia cannot be safely determined by PARR, the device is displaying ARR without any additional pulse arrhythmia type icon. Note: It is strongly recommended, that you consult your physician, if either the ARR icon
occurs newly for several times, or, if your ARR is known to your doctor, but the in­cidence of ARR readings changes over time. This is independent whether the ARR icon is specied by another pulse arrhythmia icon or not. Your doctor will then be able to provide all required medical test and possible therapeutic procedures.
The PARR technology is able to detect and display combined pulse arrhythmia ndings.
Display Results
- Normal nding ARR Pulse Arrhythmia without type-specic detection
ARR PC
ARR AFib Pulse Arrhythmia-Atrial brillation detection ARR TACH Tachycardia detection ARR BRAD Bradycardia detection
ARR PC BRAD
ARR PC TACH
ARR AFib TACH
ARR AFib PC
ARR AFib PC TACH
Pulse Arrhythmia-Premature ventricular, atrial or nodal beat detection
Combined Pulse Arrhythmia: Premature beats & Bradycardia detection
Combined Pulse Arrhythmia: Premature beats & Tachycardia detection
Combined Pulse Arrhythmia: Atrial brillation & Tachycardia detection
Combined Pulse Arrhythmia: Atrial brillation & Premature beats detection
Combined Pulse Arrhythmia: Detection of Atrial Fibrillation, Premature Beats and Tachycardia.
Real Fuzzy Measuring Technology
This unit uses the oscillometric method to detect your blood pressure. Before the cu starts inating, the device will establish a baseline cu pressure equivalent to the air pressure. This unit will automatically determine the appropriate ination level based on pressure oscillations, followed by cu deation. During the deation, the device will detect the amplitude and slope of the pressure os­cillations and thereby determine your actual the systolic blood pressure, diastolic blood pressure, and pulse rate.
Preliminary Remarks
This Blood Pressure Monitor complies with the European regulations and bears the CE mark “CE 0120”. The quality of the device has been veried and conforms to the provi­sions of the EC council directive 93/42/EEC (Medical Device Directive), Annex I essential requirements and applied harmonized standards. EN 1060-1: 1995/A2: 2009 Non-invasive sphygmomanometers - Part 1 - General require­ments
4 5
Loading...
+ 7 hidden pages