Rossmax RA500 Instruction Manual

www.qutie-rossmax.com
4. Measuring the ear temperature: Use the index nger to trig­ger. Press “Start” button, a long beep will be sounded when measurement is done.
5. Power Off: Device will automatically shut off if left idle for more than 1 minute to extend buttery life. Or press “On/ Off/Memory” button for at least 3 seconds to turn the device off.
37.5°C(99.5°F), the display will ashing and 4 short beeps will be sounded. The measuring temperature is higher than
42.2°C(108.0°F), the display will show “Hi“and 4 short beep will be sounded.
a. It is recommended that you measure 3 times with the
same ear. If the 3 measurements are different, select the
highest temperature.
b. To avoid the risk of cross contamination, please clean
the probe according to “Clean and Storage“ section after each use.
Ear Battery
°C/°F unit
Temperature display
1. To achieve accurate readings, make sure the ear canal is clean.
2. Press the “On/Off/Memory” button. The thermometer is ready for use after the ear icon is ashing and two beeps are sounded.
3. Gently pull the ear back to straighten the ear canal and snugly position the probe into the ear canal, aiming towards the membrane of the eardrum to obtain an accurate reading.
There are total 9 set memories for body measurement. If the reading of the thermometers is within the normal temperature range of 34 °C to 42.2 °C (93.2 °F to 108°F ), when the meas­urement is done, the measurement data is saved into memory.
• Press “On/Off/Memory” button again to see the tempera­ture stored.
The probe is the most delicate part of the ther­mometer. Use with care when cleaning the lens to avoid damage.
1. Please use the cotton swab with the Alcohol to
clean the lens (on the inside of the probe).
2. Allow the probe to fully dry for at least 5 minutes.
a. Please check the device if it falls and damages. If you
can’t make sure of it, please send the complete device to the nearest retailer for recalibration.
b. Keep the unit dry and away from any liquid and direct
sunlight. The probe should not be submerging into liquids.
• It means measurable.
• When battery icon appears, please replace the battery.
How to use the RA500 Infrared Ear Thermometer
LCD dislpay
Memory function
Cleaning and storage
Please read the instruction manual carefully before using the product and be sure to keep this manual.
1. Turn off the device.
2. Hold down the “Start” button, then press and hold the“ON/ OFF/Memory” button until the LCD shows °C.
3. Press the “ON/OFF/Memory” change beep sound from on to off, then release the “ON/OFF/Memory” button.
4. After 6 seconds without changing, the beep On/Off setting will be completed, and the unit will be entered into the meas­urement mode.
If you continuously obtain different measurement results, it is recommended that you use Scanning Mode to measure.
Measuring the ear temperature: Use the index nger to trig­ger, press and hold the “Start” button for 3 seconds to take measurement. During the measuring process, gently pivoting the probe side-to-side to view the entire ear canal or in order to nd the highest temperature. Release the “Start” button when hearing a long beep sound and the measurement is completed.
Enabling or disabling the beep sound
Scanning Mode
The Rossmax Infrared Ear thermometer has been carefully de­veloped for accurate, safe and fast temperature measurements
in the ear.
It is a non-invasive ear thermometer using an infrared detector to detect body temperature from the auditory canal for adult and child.
• The quality of the infrared ear thermometer has been veried and conforms to the provisions of the EC council directive 93/42/ EEC(Medical Device Directive) Annex I essential requirements and applied harmonized standards. EN 12470-5: 2003 Clinical ther­mometers-Part 5: Performance of infrared ear thermometers (with maximum device.)
• This thermometer converts the ear temperature to display its “oral equivalent.” (according to the result of the clinical evaluation)
RA500 Infrared Ear Thermometer
1. This thermometer is supplied with one lithium cell (CR2032).
2. Turn off the thermometer before replace the battery.
Change the battery
Error message Problem Solution
bi-bi-bi
The ambient temperature is not within the range be­tween 10°C~40°C (50°F~104°F)
Allow the thermometer to rest in a room for at least 30 minutes at room temperature: 10°C to 40°C (50°F to 104°F)
bi-bi-bi
The system is not functioning properly.
Unload the battery, wait for 1 minute and re-power it. If the mes­sage reappears, contact the retailer for service.
Device can not be powered on to the ready stage.
Change with a new battery. If the message reappears, contact the retailer for service.
bi-bi-bi-bi
Temperature taken is higher than
42.2°C (108.0°F)
Ensure no break or dirt of the probe cover and take a new temperature
measurement.
bi-bi-bi-bi
Temperature taken is lower than 34°C (93.2°F)
Make sure the probe cover is clean and take a new temperature
measurement.
Troubleshooting
Instruction Manual
Model: RA500
EN
3. Open the battery cover: Hold the device and ip the battery out with a small screwdriver.
4. Insert the new battery and press the right side of the battery down until you hear a click.
5. Place back the battery cover.
* The positive (+) side up and the negative (-) side pointed down.
To change the Unit from °C to °F:
1. Turn off the device.
2. Hold down the “Start” button, then press and hold the“ON/OFF/Memory” button until the LCD shows °C.
3. Press the “Start” to change the °C to °F, then release the “Start” button.
4. During the setting adjustment, each old or new tem­perature scale setting will be ashing for 6 seconds. After 6 seconds without changing, the temperature scale adjustment will be completed, and the unit will be entered into measurement mode.
Switching between temperature scale Fahrenheit and Celsius
LCD display
Battery cover
ON/OFF/Memory button
Probe
Start button
This instrument is covered by a 5 year guarantee from the date of purchase, batteries and accessories are not included. The guarantee is valid only on presentation of the guarantee card completed by the dealer conrming date of purchase or the receipt. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact Rossmax Service.
compare them. Take the temperature periodically to nd out
the normal ear temperature reading , then use that reading as the basis for comparison with any measurement taken when you suspect a fever.
2. This thermometer is designed for ear temperature meas­urement. Do not use this thermometer for other body site
measurement.
3. Keep the unit dry and away from where it might be exposed to moisture, liquids, direct sunlight, high temperature, high humidity, or excessive dust.
4. This device is not shock-protected. Do not drop, heavily pres­sure the unit or fall from high place.
5. Do not bend the device.
6. Do not disassemble or make modications the device.
7. Please do not dispose of the product in the household waste at the end of its useful life. Disposal can take place at your local retailer or at appropriate collection points provided in your country.
8. Do not boil the probe.
9. Do not use the device if it operates abnormally or error mas­sage is showed.
10. Do not use thinner or benzene to clean the device.
11. Wipe the device clean before storing.
12. When taking the device from storage below or above 10~
40°C(50~104°F), place it in 10~ 40°C(50~104°F) temperature range for at least 30 minutes before use.
13. Remove the battery if this device is not used for a period of time.
14. If this device is used according to the operation instruction, pe­riodic re-calibration is not required. If you still have questions, please send the complete device to dealers.
15. Please note that this is a home healthcare product only, and it is not intended to serve as a substitute for the advice of a physician or medical professional.
16. Do not use this device for diagnosis or treatment of any health problem on disease. Measurement results are for ref­erence only. Contact your physician if you have or suspect any medical problems. Do not change your medications without the advice of your physician or healthcare profes­sional.
17. This device may not meet its performance specication if stored or used outside temperature and humidity ranges specied in specications.
18. Battery should neither be charged nor placed into the ex­treme environment, or it may explode.
19. The thermometer contains small parts (ex: battery, etc.) that can be swallowed by children. Therefore never leave the thermometer unattended to children.
1. Besides the oral temperature, the body temperature meas-
ured by ear may be different from measurements taken by rectal or axillary. Therefore, there is no any meaning to
Warranty
IN0RA5000000000XX
OBM_IB_RA500Q_EN_ver1304
Product Information
Date of purchase:
Store where purchased:
Price Paid (excl. Tax):
Purchase for:
EMC guidance and manufacturer’s declaration
Specication
Caution
Temperature measurement range
34.0~42.2°C(93.2~108.0°F)
Accuracy ±0.2°C(0.4°F) during
35.0~42.2°C(95~108.0°F) ±0.3°C(0.5°F) for other range
Operating
temperature range
10 ~ 40°C (50~104°F), RH≤95%
Storage and transportation tem-
perature
Temperature:-25~55°C(-13~131°F), RH ≤95%
Memory 9 set Display resolution 0.1 Battery 3V, lithium CR2032 x 1 Weight (with battery) 55g Size 12.0cm(L) x 3.5cm(w) x 2.8cm(H) Auto shut down 60 sec. Battery life 3000 consecutive measurements or 1
year with 1-2 measurements per day including stand-by mode.
Safety classication Type BF equipment Clinical repeatability 0.09°C (< 1 year old)
0.07°C (1~5 years old)
0.07°C (> 5 years old)
*Dispose of device and batteries according to local regulations
Low battery: Battery icon is ashing, still can be measurable.
Suggest to replace the battery.
Dead battery: Bat­tery icon always enabled, no more
measurements are
possible.
Replace the battery
20. The ear thermometer must be inserted correctly to receive correct temperature measurement.
21. Be sure to consult a doctor if you feel that your health is in poor condition.
22. Do not judge your health only on the presence or absence of a fever.
Guidance and manufacturer’s declaration-electromagnetic immunity The RA500 is intended for use in the electromagnetic environment specied below. The customer or the user of the RA500 should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance
level
Electromagnetic environ-
ment-guidance Electrostatic discharge (ESD) IEC 61000-4-2
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30% Electrical fast tran­sient/burst IEC 61000-
4-4
± 2kV for power sup­ply lines ± 1kV for input/out­put lines
Not applicable
Not applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Surge IEC 61000-4-5 ± 1kV line(s) to line(s)
± 2kV line(s) to earth
Not applicable Not applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment. Voltage Dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
<5% UT(>95% dip in UT) for 0,5 cycle 40% UT(60% dip in UT) for 5 cycles 70% UT(30% dip in UT) for 25 cycles <5% UT(>95% dip in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the RA500
series requires continued
operation during power
mains interruptions, it is rec-
ommended that the RA500
series be powered from an
uninterruptible power supply
or a battery. Power frequency (50/60 Hz) magnetic eld IEC 61000-4-8
3 A/m 3 A/m The RA500 power frequency
magnetic elds should be
at levels characteristic of a
typical location in a typical
commercial or hospital envi-
ronment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distance between portable and mobile RF communications equip­ment and the RA500 The RA500 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the RA500 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the RA500 as recommended below, according to the maximum output power of the communica­tions equipment. Rated maximum output power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz /
d=1,2√P
80 MHz to 800 MHz
/ d=1,2√P
800 MHz to 2,5 GHz
/ d=2,3√P
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1,2 2,3
10 N/A 3,8 7,3
100 N/A 12 23 For transmitters rated at a maximum output power not listed above, the recommend­ed separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reection from structures, objects and people.
Guidance and manufacturer’s declaration-electromagnetic emissions The RA500 is intended for use in the electromagnetic environment specied below. The customer or the user of the RA500 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1 The RA500 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The RA500 is suitable for use in all establish-
ments, including domestic establishments and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Not applicable
Voltage uctuations/icker emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity The RA500 is intended for use in the electromagnetic environment specied below. The customer or the user of the RA500 should assure that it is used in such an environment.
Immunity
test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidan
ce
Conducted RF IEC 61000-
4-6
Radiated RF IEC 61000­4-3
3 Vrms 150 KHz to 80 MHz
3 V/m 80MHz to 2,5 GHz
Not applica­ble
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the RA500 series, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d = 1,2 √P d = 1,2 √P 80MHz to 800 MHz d = 2,3 √P 800MHz to 2,5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the rec­ommended separation distance in metres (m). Field strengths from xed RF transmitters, as determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol: NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec­tromagnetic environment due to xed RF transmitters, an electromagnetic site sur­vey should be considered. If the measured eld strength in the location in which the RA500 series is used exceeds the applicable RF compliance level above, the RA500 series should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the RA500 series.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Rossmax International Ltd.
12F., No.189, Kang Chien Rd., Taipei 114, Taiwan
Rossmax Swiss GmbH
Tramstrasse 16, CH-9442 Berneck, Switzerland
www.qutie-rossmax.com
Loading...