Ross Companion Operating Manual

No. 54748–Purchase
No. 00084–Lease/Rent
Operating Manual
For Enteral Use Only Not for IV Use
CONTENTS
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Pediatric Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Special Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Illustrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Control Dial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Visual Displays and Audio Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Pump/Charger Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Disinfectants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Ambulatory Transporter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
USER QUALIFICATIONS
The Ross Companion
®
Enteral Pump is for use at the direction of or under the super­vision of licensed physicians, and by licensed or certified health care professionals who are trained in the use of the pump and the administration of enteral fluids.
© 2001, 2004 Abbott Laboratories
2
PUMP CHARACTERISTICS
®
The Ross Companion
Enteral Nutrition Pump is a volumetric infusion pump that uses a specially designed administration set, including a cassette with bellows, to deliver measured amounts of enteral formula. The pump operates on AC or battery power. The fluid delivery system and safety alarms function when the pump is used by an ambulatory patient or is mounted on a pole. The volume-monitoring device does not depend on drop counting, so pump adapts well to patient activities.
The pump is controlled by a microprocessor (computer), providing accurate delivery rates, easy-to-read displays, and simple controls. A charger unit with a power cord is provided with each Companion Pump. The charger may be clamped to a feeding stand or pole or positioned on a flat surface.
Additionally, the Companion Pump offers these features:
1. Accuracy to ±10% at rates of 5 to 300 mL/hr
2. 8-hour battery operation at 150 mL/hr, when fully charged
3. Alarms:
OCCLUSION
EMPTY FEEDING CONTAINER/MISSING OR IMPROPERLY LOADED
CASSETTE
SELECT RUN
LOW BATTERY
RESET RATE
4. Fluid flow and fluid monitoring independent of pump position (no drop counting)
5. Adjustable alarm volume for low or high setting
6. Flow rate selection of 5 to 300 mL/hr in 1-mL/hr increments
7. VOLUME FED accumulation display
8. User-friendly operating controls
9. Small and lightweight
10. Simple setup
11. Backlit visual display for easier viewing in a darkened room
12. Self-test capability
INDICATIONS FOR USE
The Ross Companion Enteral Nutrition Pump can be used for adult and pediatric patients provided the patients can tolerate a feeding range within the pump’s operational specifications. Those specifications are:
•The flow rate range is 5-300 mL/hr in 1-mL/hr increments
•The flow rate accuracy is ±10%
•Pumps against 24 psi nominal back pressure before no FLO alarm.
3
This pump is designed to deliver only a liquid enteral feeding product (standard liquid product, infant formula or reconstituted powder product that has been thoroughly mixed into solution).
If the specifications are not appropriate for a given patient, the
®
Companion
Pump should not be used.
PEDIATRIC USE
Do not use this pump for feeding newborn infants or infants in a neonatal intensive care unit (NICU). Enteral pumps have the potential to bolus-feed small
amounts, which is an important consideration in feeding volume-sensitive patients. The Ross Companion Enteral Nutrition Pump should normally be used only for children 12 months of age or older, and only if the rate of feeding is 25 mL/hr or greater. In these patients, do not hang more volume of nutritional product than can be tolerated as a bolus.
PRECAUTIONS
This pump is not for IV (intravenous) use.
Confirm proper placement and function of patient’s enteral feeding tube (nasogastric, jejunostomy, gastrostomy, etc) and verify the following before initiating feeding:
1. Only
a Ross Companion®Pump set is being used.
2. Cassette is properly seated in pump.
3. When connected to AC power, the pump is fully seated in charger.
4. Flow rate is set at the prescribed mL/hr.
SPECIAL FEATURES
Small Size: The Companion Pump is small and lightweight. The pump weighs 1.5 lbs; the charger weighs 2.5 lbs. With the Companion Ambulatory Transporter, a patient can easily carry the pump and up to 500 mL of enteral formula during routine activities.
Position Independent: The Companion Pump fluid delivery and alarm systems function when pump is in any position that might be customary for ambulatory use.
4
ROSS COMPANION®ENTERAL NUTRITION PUMP
Illustrations
Touch Pads
Select the numerical value of flow rate from 5 to 300 mL/hr in 1-mL increments.
Pump
Display Light
Press to temporarily illuminate display panel while on battery power.
Display Panel
Shows flow rate, cumulative
VOLUME FED, RUN indicator,
and visual expla­nation of all intermittent audio alarms.
Control Dial
Select pump functions.
HI/LO Alarm Volume
Select volume of audio alarm.
Cassette Cavity
Using shape for orientation, insert cassette into pump until seated.
Cassette Release Latch
Press down to release cassette for removal.
Pump Set
Cassette With Bellows
To Container
5
Adapter
Adapter Cover
Charger
Carrying Handle
Pump/Charger Connector
Pump Release Latch
Pole Clamp Knob
Pole Clamp
Power Cord
6
Self-Test Procedure: Each time pump is turned on, the microprocessor initiates a self-test procedure. Audio alarm, visual displays, and readout (8888) turn on for 5 seconds. If the self-test fails, F1 may appear. If this happens, DO NOT USE THE
PUMP; SERVICING IS REQUIRED.
Flashing Displays: Three flashing visual displays indicate the need for immediate
attention:
EMPTY
OCCL
LOW BATTERY
CONTROL DIAL SETTINGS
POWER OFF Stops all pump functions. Battery charges in all control
dial settings when pump is plugged into AC power.
SET RATE Allows flow rate to be set (5 to 300 mL/hr in 1-mL
increments) by pressing the touch pads (arrow buttons). No pumping action occurs in this mode.
RUN Only setting that activates pumping action. Visual display
shows flashing RUN and rate of flow in mL/hr.
HOLD Stops pumping action and silences audio alarm (if one
has occurred) without erasing previous commands or memory functions. HOLD is used when correcting an alarm condition, or whenever feeding is temporarily interrupted (such as when connecting a new feeding container).
VOLUME FED Stops pumping action and displays cumulative volume fed
in mL, since this value was last cleared. (NOTE: If 9999 mL is fed, numerical value turns to 0 and a new accumulation begins.) To clear VOLUME FED, turn knob to VOLUME CLEAR and wait for 4 beeps to sound.
VOLUME CLEAR Stops pumping action and automatically clears to zero any
accumulated numerical value of volume fed to that point.
NOTE: Pump activates flow only in RUN setting. Pump maintains memory in all settings except POWER OFF and VOLUME CLEAR.
7
VISUAL DISPLAYS AND AUDIO ALARMS
VISUAL DISPLAYS
CONTROL VISUAL DISPLAY AUDIO ALARM DIAL SETTING
POWER OFF None None
(Charge)
SET RATE Previous flow rate or None
(No pumping action) 000 mL/hr
RUN Numerical rate in None
mL/hr and RUN
HOLD Existing flow rate or None (No pumping action) 000 mL/hr (Any previous
Visual alarm display audio alarm remains visible (if one ceases) occurred)
VOLUME FED Numerical value of None (No pumping action) accumulated volume fed
in mL or 0000
VOLUME CLEAR Numerical value of None (No pumping action) accumulated volume fed
is automatically cleared to zero (0000 mL)
8
Loading...
+ 16 hidden pages