Rosenberger PIM User Manual

PIM Rack Analyzer
User Manual
Table of Contents
1 General Information and Safety Instructions .......................................................................................... 3
2 Package Contents ...................................................................................................................................... 8
3 PIM Rack Analyzer Product Description .................................................................................................. 9
3.1 Overview .................................................................................................................................................... 9
3.2 Filter ......................................................................................................................................................... 12
4 Setting Up the Device ............................................................................................................................... 13
4.1 Switch Matrix Installation and Cabling of First Filter Unit ........................................................................ 14
4.2 Cabling of Additional Filter....................................................................................................................... 20
4.3 Software Setup ........................................................................................................................................ 24
5 Measurement Settings ............................................................................................................................. 26
5.1 Main Menu ............................................................................................................................................... 26
5.2 Manual Mode ........................................................................................................................................... 27
5.3 2 Tone ...................................................................................................................................................... 28
5.4 Sweep ...................................................................................................................................................... 29
5.5 VSWR \ RL .............................................................................................................................................. 30
5.6 DTF .......................................................................................................................................................... 31
5.7 Isolation ................................................................................................................................................... 32
5.8 Power Sweep .......................................................................................................................................... 33
5.9 Spectrum Analyzer .................................................................................................................................. 34
6 Software Operation .................................................................................................................................. 34
6.1 Common Settings .................................................................................................................................... 35
6.2 Filter / Band Selection ............................................................................................................................ 36
6.3 Measurement Screen .............................................................................................................................. 37
6.3.1 History................................................................................................................................................... 38
6.4 Report Generation ................................................................................................................................... 39
6.4.1 Creating a Full Report .......................................................................................................................... 39
6.4.2 Creating a Single-Page Report ............................................................................................................ 42
6.4.3 Exporting Measurement Data as CSV ................................................................................................. 42
6.4.4 Saving a Screenshot ............................................................................................................................ 42
6.5 Zeroing PIM ............................................................................................................................................. 43
6.5.1 Zeroing Return Loss ............................................................................................................................. 44
6.5.2 Setting the velocity factor of the signal path (+Application Note DTF) ................................................. 45
6.6 Device Info ............................................................................................................................................... 46
6.7 Defining Own Buttons for Section Names ............................................................................................... 47
6.8 Updating the PIM Analyzer Software ...................................................................................................... 48
7 Remote Control ......................................................................................................................................... 49
8 Maintenance of the Device ...................................................................................................................... 49
8.1 Handling................................................................................................................................................... 49
8.2 Cleaning................................................................................................................................................... 49
8.3 Calibration & Repair ................................................................................................................................ 49
9 Measurement Examples........................................................................................................................... 50
9.1 RF Measurements ................................................................................................................................... 50
9.1.1 PIM Acceptance Test ........................................................................................................................... 50
9.1.2 Troubleshooting a PIM Problem ........................................................................................................... 50
10 Troubleshooting ..................................................................................................................................... 52
10.1 Reporting Software Problems to Rosenberger ..................................................................................... 52
10.2 How to Avoid Common Problems ......................................................................................................... 52
11 Technical Specifications........................................................................................................................ 53
11.1 Base Unit ............................................................................................................................................... 53
11.1.1 Main Features ..................................................................................................................................... 53
11.1.2 Product Description ............................................................................................................................ 53
11.1.3 General ............................................................................................................................................... 54
11.1.4 RF PIM Analyzer (Base Unit) -150W Option ...................................................................................... 54
11.1.5 CPRI PIM Analyzer (SW Option) ........................................................................................................ 54
11.1.6 RF Isolation Measurement ................................................................................................................. 55
11.1.7 RF VSWR / Return Loss..................................................................................................................... 55
11.1.8 RF Spectrum Analyzer ....................................................................................................................... 55
11.1.9 Environmental ..................................................................................................................................... 55
11.1.10 User Interfaces ................................................................................................................................. 56
11.2 Filter Units ............................................................................................................................................. 56
11.2.1 Main Features ..................................................................................................................................... 56
11.2.2 General ............................................................................................................................................... 56
11.2.3 Environmental ..................................................................................................................................... 56
11.2.4 Compatibility / Supported Base Unit ................................................................................................... 57
11.2.5 IM-R-FI-07/B12-14-R (LTE700 L+U) .................................................................................................. 57
11.2.6 IM-R-FI-07/B28-R (APT700) .............................................................................................................. 57
11.2.7 IM-R-FI-08/B20-R (DigDiv800) ........................................................................................................... 57
11.2.8 IM-R-FI-08/B5-R (AMPS850) ............................................................................................................. 58
11.2.9 IM-R-FI-09/B8-R (EGSM900) ............................................................................................................. 58
11.2.10 IM-R-FI-14/B11+21-R (LTE1400) ..................................................................................................... 58
11.2.11 IM-R-FI-18/B3-R (DCS1800) ............................................................................................................ 59
11.2.12 IM-R-FI-19/B2+4-R (PCS/AWS1900) ............................................................................................... 59
11.2.13 IM-R-FI-21/B1-R (UMTS2100) ......................................................................................................... 59
11.2.14 IM-R-FI-23/B30-R (WCS) (opt. on request) ..................................................................................... 59
11.2.15 IM-R-FI-26/B7-R (LTE II 2600) ......................................................................................................... 60
11.3 Part Number Designation ...................................................................................................................... 60
11.4 Order Number Example......................................................................................................................... 61
12 CE Declaration of Conformity ............................................................................................................... 62
13 Accessories ............................................................................................................................................ 63
13.1 Filter Units ............................................................................................................................................. 63
13.2 Other Accessories ................................................................................................................................. 64
14 Support and Sales Locations ................................................................................................................ 65
14.1 Europe, Middle East, Africa ................................................................................................................... 65
14.2 Americas ................................................................................................................................................ 65
14.3 Brazil ...................................................................................................................................................... 65
14.4 Asia Pacific ............................................................................................................................................ 66
14.5 India ....................................................................................................................................................... 66
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 4 of 66
1 General Information and Safety Instructions
Operation which deviates from the intended purpose of the product, failure to ob­serve this documentation, the use of insufficiently qualified personnel as well as unauthorized modifications will exclude the manufacturer from any liability for re­sulting damage and will render any warranty void.
People with cardiac simulators must not be exposed to the magnetic field of the batteries, the battery unit, the power unit or the charging cable.
Rosenberger makes every effort to keep the safety standards of our products up to date to offer our custom­ers the highest possible degree of safety. Our products and the required accessories are designed, built and tested in accordance with the safety standards that apply in each case. Compliance with these standards is monitored by our quality assurance system. The product described here has been designed, built and tested in accordance with the attached CE Certificate of Conformity and has left the manufacturer’s plant in a condi­tion fully complying with safety standards. To maintain this condition and to ensure safe operation, you must observe all instructions and warnings provided in this manual. If you have any questions regarding these safety instructions, please contact Rosenberger.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is designed for production & laboratory use and must not be used in any way that may cause personal injury or property damage. You are responsible if the product is used for any intention other than its designated purpose or in disregard of the manufacturer's instructions. The manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its product documentation and within its performance limits (see data sheet, documentation, the following safety instructions). Using the product requires technical skills and a basic knowledge of the English language. It is therefore essential that only skilled and specialized staff or thoroughly trained personnel with the required skills are allowed to use the product. If personal safety gear is required in order to use the Rosenberger PIM Rack Analyzer, this will be indicated at the appropriate place in the product documentation. Keep the basic safety instructions and the product documentation in a safe place and pass them on to any subsequent users.
Observing the safety instructions will help prevent personal injury or damage of any kind caused by danger­ous situations. Therefore, carefully read through and adhere to the following safety instructions before and when using the product. It is also absolutely essential to observe the additional safety instructions on per­sonal safety, for example, that appear in relevant parts of the product documentation.
Operating conditions and operating positions
The product must only be operated under the operating conditions and in the positions specified by the man­ufacturer. The product’s ventilation must not be obstructed. If the manufacturer's specifications are not ob­served, this can result in electric shock, fire and/or serious personal injury or death and product damage. Applicable local or national safety regulations and rules for the prevention of accidents must be observed during all work.
Never switch output power on (in manual mode or remote mode) without a load or terminated OUT connected to the test port.
Unless otherwise specified, the following requirements apply to this product: IP protection 2X, pollution severity 2, overvoltage category 2, max. operating altitude 2000 m above sea level, max. transport altitude 4500 m above sea level. A tolerance of +-10 % shall apply to the nominal voltage and +- 5 % to the nominal frequency.
Do not place the product on surfaces, vehicles, cabinets or tables that, for reasons of weight or stability, are unsuitable for this purpose. Always follow the manufacturer’s installation instructions when installing the product and fastening it to objects or structures (e.g. walls and shelves). Installation that is not carried out as described in the product documentation could result in personal injury or death.
Do not cover the heat sink or ventilation openings.
Do not place the product on heat-generating devices such as radiators or fan heaters. The ambient
temperature must not exceed the maximum temperature specified in the product documentation or in the data sheet. Product overheating can cause electric shock, fire and/or serious personal injury or death.
Electrical safety
If the information on electrical safety is not observed at all or to the extent necessary, electric shock, fire and/or serious personal injury or death may occur.
Prior to switching on the product, always ensure that the nominal voltage setting on the product matches the nominal voltage of the AC supply network. If the equipment is used at different voltages, the power fuse of the product may have to be changed accordingly.
In the case of products in safety class I with movable power cord and connector, operation is only permitted using sockets with an ground contact and protective ground connection.
Intentionally breaking the protective ground connection either in the feed line or in the product itself is not permitted. Doing so may result in an electric shock from the product. If extension cords or connector strips are implemented, they must be checked on a regular basis to ensure that they are safe to use.
The plug on the connecting cable serves as the disconnecting device for disconnecting the product from the AC supply network. In such cases, always ensure that the power plug is easily reachable and accessible at all times (corresponding to the length of the connecting cable, approx. 2.5 m). Functional or electronic switches are not suitable for disconnecting the product from the AC supply network. If products without power switches are integrated into racks or systems, a disconnecting device must be provided at the system level.
Never use the product if the power cable is damaged. Check the power cable on a regular basis to ensure that it is in proper working order. By taking appropriate safety measures and laying the power cable carefully, you can ensure that the cable will not be damaged and that no one can be hurt by, for example, tripping over the cable or suffering an electric shock.
The product must only be operated in TN/TT networks fused with max. 16 A (230 VAC) / 20 A (110 VAC)
Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the socket. Otherwise, sparks may occur and result in fire and/or injuries.
Do not overload any sockets, extension cords or connector strips; doing so can result in fire or electric shocks.
For measurements in circuits with voltages Vrms > 30 V, suitable measures (e.g. appropriate measuring equipment, fusing, current limiting, electrical separation, insulation) should be taken to avoid any hazards.
Ensure that connections with information technology equipment, e.g. PCs or other industrial computers, comply with the IEC60950-1 / EN60950-1 or IEC61010-1 / EN 61010-1 standards that apply in each case.
Unless expressly permitted, never remove the cover or any part of the housing while the product is in operation. Doing so will expose circuits and components and can lead to injuries, fire or damage to the product.
If a product is to be permanently installed, the connection between the PG terminal on site and the product's PG conductor must be made first before any other connection is made. The product must only be installed and connected by a licensed electrician.
For permanently installed equipment without built-in fuses, circuit breakers or similar protective devices, the supply circuit must be fused in such a way that anyone who has access to the product, as well as the product itself, is adequately protected from injury or damage.
Use suitable overvoltage protection to ensure that no overvoltage (such as that caused by a bolt of lightning) can reach the product. Otherwise, the person operating the product will be exposed to the danger of an electric shock.
Any object that is not designed to be placed in the openings of the housing must not be used for this purpose. Doing so can cause short circuits inside the product and/or electric shocks, fire or injuries.
Unless specified otherwise, products are not liquid-proof (see also section "Operating conditions and operating positions", item 1.) Therefore, the equipment must be protected against penetration by liquids. If the necessary precautions are not taken, the user may suffer an electric shock or the product itself may be damaged, which can also lead to personal injury.
Never use the product under conditions in which condensation has formed or can form in or on the product, e.g. if the product has been moved from a cold to a warm environment. Penetration by water increases the risk of electric shock.
Operation
Operating the products requires special training and intense concentration. Make sure that persons who use the products are physically, mentally and emotionally fit enough to do so; otherwise, injuries or material damage may occur. It is the responsibility of the employer/operator to select suitable personnel for operating the products.
Operation of the PIM Rack Analyzer can produce electromagnetic radiation. Ensure that the radiation levels do not exceed the limits stipulated by national regulations. Persons with pacemakers and pregnant women are especially at risk.
Before you move or transport the product, read and observe the section titled "Transport".
Before you start processing the product mechanically and/or thermally, or before you take it apart, be sure to read and pay special attention to the section titled "Waste disposal", item 1.
Should a fire occur, the product may release hazardous substances (gases, fluids, etc.) that can cause health problems. Therefore, suitable measures must be taken, e.g. protective masks and protective clothing must be worn.
Repair and service
The product may only be opened by authorized, specially trained personnel. Before any work is performed on the product or before the product is opened, it must be disconnected from the AC supply network. Otherwise, personnel will be exposed to the risk of an electric shock.
Adjustments, replacement of parts, maintenance and repairs may only be performed by electrical experts authorized by Rosenberger. Only original parts can be used for replacing safety parts (e.g. power switches, power transformers, fuses). A safety test must always be performed after safety parts have been replaced (visual inspection, PG conductor test, insulation resistance measurement, leakage current measurement, functional test). This helps to ensure the continued safety of the product.
Transport
The product may be very heavy. Therefore, it must be handled with care. In some cases, the user may require a suitable means of lifting or moving the product (e.g. with a second person) to avoid back or other physical injuries.
The user is responsible for securely fastening the products to or on the transport or lifting equipment. Observe the safety regulations issued by the manufacturer of the transport or lifting equipment. Non­compliance may result in personal injury or material damage.
If you use the product in a vehicle, it is the sole responsibility of the driver to drive the vehicle safely and properly. The manufacturer assumes no responsibility for accidents or collisions. Never use the product in a moving vehicle if doing so could distract the driver of the vehicle. Adequately secure the product in the vehicle to prevent injuries or other damage in the event of an accident.
Waste disposal
If products or their components are mechanically and/or thermally processed in a manner that goes beyond their intended use, hazardous substances (heavy-metal dust such as lead, beryllium, nickel) may be released. For this reason, the product may only be disassembled by specially trained personnel. Improper disassembly may be hazardous to your health. National waste disposal regulations must be observed.
If handling the product releases hazardous substances or fuels that must be disposed of in a special way, e.g. coolants or engine oils that must be replenished regularly, the safety instructions issued by the manufacturer of the hazardous substances or fuels and the applicable regional waste disposal regulations must be observed. Also observe the relevant safety instructions in the product documentation. The improper disposal of hazardous substances or fuels can cause health problems and lead to environmental damage.
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 8 of 66
Cleaning
Prior to cleaning the product, disconnect it completely from the power supply. Use a soft, non-linting cloth to clean the product. Never use chemical cleaning agents such as alcohol (except in the case of the test port), acetone or diluents for cellulose lacquers.
Use pressurized air or alcohol-soaked cotton swabs to clean the test port.
2 Package Contents
Please keep the original boxes and all packaging materials and use them in the same way as received when sending back the unit, e.g. for calibration, service. This way the risk of damage during transport can be minimized.
Before doing anything else, please check that all parts are contained in your shipment. If anything is missing, contact Rosenberger. See the information below for a description of the contents of the main unit packaging. The filter unit comes in a separate box.
Packaging
- Base unit
- Power cable (country-specific)
- DTF zeroing adapter
- Torque wrench
- Adjustable flat wrench
- USB stick
Unpacking the box
When opening the box, remove the accessory box and foam at the top first. With the help of a second per­son,lift the unit out of the box using the foldable handles at the side and on the front panel. The center of gravity is at the side handles – do not carry the unit by the front panel handles!
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 9 of 66
3 PIM Rack Analyzer Product Description
Dear customer, thank you for purchasing the Rosenberger PIM Rack Analyzer. The PIM Rack Analyzer al­lows you to perform PIM and VSWER_Distance_to_Fault (DTF) measurements. The plug-and-play filter units and the broadband base units from 700-2200 MHz and 2100-2700 MHz ensure maximum flexibility.
3.1 Overview
Do not cover cooling fan or ventilation inlets, otherwise the device may overheat.
Front
Air dust filter
Ventilation inlet
Type/Model
Handles
USB
Touch­display
On/Off pushbutton
Back
Power socket with main switch
10 MHz reference input
Safety port
USB
LAN 1 Gbit/s
USB remote con­trol
RF receiver inputs PIM
FWD
REFL
TX1
TX2
Amplifier outputs
3.2 Filter
Filter units are available with preassembled 7-16 (4.3-10 on request).
Front
Back
4
PIM
TX1
TX2
REFL
Handles Type/Model
Ventilation inlet air dust filter
Test port RF in- & outputs
FWD
Bus in
Bus out
Termination switch
Ventilation outlet
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 13 of 66
4 Setting Up the Device
Always ensure that the device is stable and that the environmental conditions – listed under Chapter 1 – Section: Operating conditions and operating positions – are acceptable before starting to use the PIM Rack Analyzer.
Please disconnect the unit from the power supply before reconfiguring the RF ca­bling of the switch matrix.
Required tools (not included in the scope of delivery)
Small flat-head screwdriver (for switching the termination switch)
8 mm + 18 mm flat wrench (for tightening the SMA and N plugs)
Recommended: 8 mm (0.9 Nm) + 18 mm (1.1 Nm) torque wrenches (for tightening the SMA and N plugs)
Small flat-head screwdriver
18 mm (1.1 Nm)
Flat wrench
8 mm
18 mm
Torque wrench
8 mm (0.9 Nm)
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 14 of 66
4.1 Switch Matrix Installation and Cabling of First Filter Unit
1. Mount the switch matrix on the back of the base unit (IM-R-BU-xxx).
Hand-tighten the knurled screws.
Switch matrix
2. 11-way switch matrix only: Mount the cables W125, W126, W127 on PIM RX (X to X),
TX1 (X to X) and TX2 (X to X).
3. Mount cable W122 between REFL and REFL-Com, W121 between FWD and FWD-Com and
W120 between RX and RX-C (make sure that the longer end of cable W120 is mounted on the RX connector).
W125 PIM RX (X to X)
W120 (RX – RX-C)
W121 (FWD – FWD-Com)
W122 (REFL – REFL-Com)
W126
TX1 (X to X)
W127 TX2 (X to X)
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 16 of 66
4. Mount cables W123 and W124 between the amplifier outputs and relay common (TX1 and TX1-
C, Tx2 and TX2-C).
W123
TX1 – TX1-C
W124
TX2 – TX2-C
5. Mount cable LY8-C039-xxx for each path (xxx corresponds to the length; two different lengths
depending on the distance between base unit and filter)
PIM RX (on switch matrix) to RF out PIM (on filter unit) TX1 (on MPX) to RF in TX1 (on filter unit) TX2 (on MPX) to RF in TX2 (on filter unit)
When mounting, please make sure that the position on the switch matrix is the same for all paths. E.g. all cables for one filter must be in position 1.
LY8-C039-xxx
PIM RX 1 (on MPX) to RF out PIM
LY8-C039-xxx TX1 1 (on MPX) to RF in TX1
LY8-C039-xxx TX2 1 (on MPX) to RF in TX2
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 18 of 66
6. Take two L71-A0036-xxx cables (xxx corresponds to the length; two different lengths depending
on the distance between base unit and filter) and mount them from FWD to FWD and from REFL to REFL. Please make sure that the position on the switch matrix is the same for all paths. E.g.
both cables must be in position 1.
Please insert the SMP plug of the L71-A0036-xxx cable carefully and straight
L71-A0036-xxx
FWD - FWD 1
L71-A0036-xxx
REFL - REFL 1
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 19 of 66
7. Take the LD5-101-0550-E-C cable and plug it into “Bus out” (green) on the switch matrix and
“Bus in” (blue) on the filter.
If you are not connecting another filter, change the position of the termination switch to ON. If you are connecting further filters, the position of the termination switch must be set to OFF.
LD5-101-0550-E-C
Bus in
LD5-101-0550-E-C
Bus out
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 20 of 66
LY8-C039-xxx
TX2 2 to RF in TX2
4.2 Cabling of Additional Filter
When mounting, please make sure that the position on the switch matrix is the same for all paths. E.g. all cables for this filter must be in position 2.
LY8-C039-xxx
PIM RX 2 – RF out PIM
LY8-C039-xxx
TX1 2 to RF in TX1
1. Mount cable LY8-C039-xxx for each path. PIM RX (on MPX) to RF out PIM (on filter unit) TX1 (on MPX) to RF in TX1 (on filter unit)
TX2 (on MPX) to RF in TX2 (on filter unit)
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 21 of 66
L71-A0036-xxx FWD- FWD 2
L71-A0036-xxx
REFL - REFL 2
2. Take two L71-A0036-xxx cables and mount them from FWD to FWD and from REFL to REFL. Please make sure that the position on the switch matrix is the same for all paths. E.g. both cables must be in position 2.
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 22 of 66
LD5-101-0350-E-C Bus in
LD5-101-0350-E-C Bus out
3. Take the LD5-101-0350-E-C cable and plug it into “Bus out” on the previous filter and “Bus in” on the current filter.
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 23 of 66
4. If you are not connecting another filter, change the position of the termination switch to ON and change the switch on the previous filter to OFF. After checking all connections for correctness and torque, connect the unit to the power supply again.
4.3 Software Setup
Open the main menu and “Filter Connections”
Select the connected filter unit for each switch matrix position
Select the cables used for each path (see cable connection dialog)
4. Repeat for every filter unit
5. Apply settings
6. Select filter as described in the “Software Operation” chapter (See page 36)
1. Either select one of the predefined and standard-accessory cables or
2. Select a custom­defined *.s2p file or use a single attenuation value
3. Optional: Copy the same selection for the remaining paths of this filter.
5 Measurement Settings
5.1 Main Menu
The main menu can be accessed from every screen with the button in the upper left corner. It allows you to switch between all the different measurement types and settings.
Switch to specific measurement
Settings for specific measurement
Toggle fullscreen (go to Windows)
Toggle in menu
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 27 of 66
5.2 Manual Mode
Manual controlling of test carrier’s power and frequency while observing re­ceiver frequency and signal strength in real time. Used to characterize DUT be­havior and find and try out appropriate settings for regular RF measurements.
Group Parameter Description
TX1
Frequency f
1
Frequency setting for CW carrier 1
Power f
1
Power setting for CW carrier 1
On | Off f
2
Enable or disable carrier 2
TX2
Frequency f
2
Frequency setting for CW carrier 2
Power f
2
Power setting for CW carrier 2
On | Off f
2
Enable or disable carrier 2
RX1
Path RX
1
Choose between PIM or FWD monitor path
Frequency RX
1
Set frequency where RX1is listening
IM order + Use IM Frequency
Choose between IM2| IM3| IM5| IM7… depending on filter and set RX
1
to a frequency derived from f1, f2and IM order settings
RX2
Path RX
2
Choose between ISOLation or REFL monitor path
Frequency RX
2
Set frequency where RX2is listening
Power Detector Type
AVG | Peak Choose between average and peak mode for receiver
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 28 of 66
5.3 2 Tone
Measuring PIM under dynamic stress (e.g. tap with a tool to simulate tower movements during wind). Two fixed-frequency CW carriers stimulate PIM at the DUT while the PIM level is recorded over time and displayed.
Select power detector type “peak” to measure short, intermittent signals. Switch to “average” to get the lowest possible noise floor.
Group Parameter Description
Stimulus
Frequency f
1
Frequency setting for CW carrier 1
Frequency f
2
Frequency setting for CW carrier 2
Power f
1
Power setting for CW carrier 1
Power f
2
Power setting for CW carrier 2 – can be coupled to Power f
1
Measurement Setup
IM order Choose between IM
2
| IM3| IM5| IM7… measurements
Time Set time duration of 2 Tone measurement
Endless Mode Run 2 Tone measurement until stop is pressed
Display Unit Choose between dBm (absolute) and dBc (relative to carriers)
Alarm Setup Error Limit Set error limit level
Power Detector Type
AVG | Peak Choose between average and peak mode for receiver
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 29 of 66
5.4 Sweep
Measuring PIM with swept carrier frequencies. Allows detection of multiple PIM sources in the signal path. An unsteady graph would indicate that more than one PIM source is present because all reflected PIM waves would interfere posi­tively and negatively over frequency.
Group Parameter Description
Upsweep
Frequency f
1
Frequency from-to setting of swept carrier 1
Frequency f
2
Frequency setting for fixed carrier 2
Step size Frequency step size for swept carrier
Power f
1
Power setting for carrier 1
Downsweep
Frequency f
2
Frequency from-to setting of swept carrier 2
Frequency f
1
Frequency setting for fixed carrier 1
Step size Frequency step size for swept carrier
Power f
2
Power setting for carrier 2 (must be equal if unit dBc is used)
Measurement Setup
IM order Choose between IM
2
| IM3| IM5| IM7… measurements
No. of Cycles Set wait time between multiple sweeps
Endless Mode Run sweep measurement until stop is pressed
Display Unit Switches scale between dBm (absolute) and dBc (relative to carriers)
Alarm Setup Error Limit Set error limit level
Power Detector Type
AVG | Peak Choose between average and peak mode for receiver
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 30 of 66
5.5 VSWR \ RL
Measures if components in the signal path are impedance-matched or not. High RL or VSWR values mean that signals are reflected back instead of pass­ing the mismatched component.
Group Parameter Description
Stimulus
Frequency Frequency from-to setting of swept carrier
Step Size Frequency step size for swept carrier
Power Power setting for carrier
Measurement Setup
Display Unit Choose between dB (return loss) and
VSWR factor (1:1.0 – 1:∞)
Error Limit Set error limit level
No. of Cycles Set number of sweep cycles per measurement
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 31 of 66
5.6 DTF
Measuring the distance to the PIM fault and return loss fault position. The graph plots the PIM and RL values over distance, so even multiple sources can be detected and the level measured. The velocity factor has to be known for accu­rate display scaling.
In order to perform DTF measurements, the filter needs to be zeroed (PIM and return loss) first. The zeroing only needs to be done once after a new filter is inserted for the first time (if the filter is changed afterwards the equipment will use the stored value that was created for this filter by means of zeroing) and it can only be done on the device, not on the tablet.
Group Parameter Description
Measurement Setup
Measurement Mode
Choose between distance to PIM only | distance to RL only | both
Velocity Factor Sets the cable type or user-defined velocity factor for proper scaling
of the distance measurement (x-axis)
Distance to PIM
Power f
1
Power setting for carrier 1
Power f
2
Power setting for carrier 2 (must be equal if unit dBc is used)
Equal Power Couples Power f2to Power f1so that only one has to be adjusted
Error Limit Sets error limit level
Display Unit Switches scale between dBm (absolute) and dBc (relative to carriers)
Distance to RL
Power Choose between IM
2
| IM3| IM5| IM7… measurements
Error Limit Set error limit level
Freq. / Channel Select channel where previous measurement showed worst point
Display Unit
Select scale between dB (return loss) and VSWR factor (1:1.0 – 1:∞)
Setting New Zero Point
On / Off Enables the zeroing buttons to allow reconfiguration of zero distance
(e.g. set an offset to end of jumper)
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 32 of 66
5.7 Isolation
Measuring the isolation between separate RF paths (e.g. decoupling of two an­tennas or separate polarizations). For this purpose, a signal is transmitted from the PIM measuring port (7-16 or 4.3-10) and the leakage is received at the N­connector.
Group Parameter Description
Stimulus
Frequency Frequency from-to setting of swept carrier
Step Size Frequency step size for swept carrier
Power Power setting for carrier
Measurement Setup
Error Limit Set error limit level
No. of Cycles Set number of sweep cycles per measurement
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 33 of 66
5.8 Power Sweep
Measuring PIM as a function of power to the DUT. Because PIM is an extremely nonlinear effect with many causes, it is hard to predict the power level at which PIM will start to rise. PIM sources or measurement headroom can be identified in this manner.
Group Parameter Description
Sweep Setup
Frequency f
1
Frequency setting for CW carrier 1
Frequency f
2
Frequency setting for CW carrier 2
Frequency f
IM
Shows resulting IM frequency from f1f2setting
Power Power sweep from-to setting of both carriers
Step size Power increase per step for power sweep
Measurement Setup
IM order Choose between IM
2
| IM3| IM5| IM7… measurements
No. of Cycles Set number of sweep cycles per measurement
Display Unit Switches scale between dBm (absolute) and dBc (relative to carriers)
Alarm Setup Error Limit Set error limit level
Power Detector Type
AVG | Peak Choose between average and peak mode for receiver
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 34 of 66
5.9 Spectrum Analyzer
Measuring the uplink spectrum prior to PIM measurements is useful to detect if interferences (e.g. mobile devices transmitting unintentionally) are disturbing the PIM measurement. The RF power density on the graph shows if there are static signals which must be ignored on PIM measurements.
Group Parameter Description
Measurement Setup
Frequency Frequency from-to setting of monitoring range (in RX range)
Reference Level Level of reference line (Line 9)
Scale Power scale between lines
Resolution BW Resolution bandwidth (frequency resolution and noise floor)
Video BW Video bandwidth (smoothens dynamic signals)
Display Setup
Detector Choose between peak | average | min
Persistence Switch between persistence mode # of cycles | Off
Alarm Setup Error Limit Set error limit level
6 Software Operation
The PIM Rack Analyzer runs on Microsoft Windows 7. If you install additional soft­ware or change system settings, Rosenberger can no longer guarantee flawless operation. Please back up your data (e.g. reports) on a regular basis.
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 35 of 66
6.1 Common Settings
If output offset is active, output power is increased by that value. Thus the configu­rable output power for each measurement may be reduced depending on the maxi­mum power of the amplifiers and the currently selected filter.
Tap to cycle through available languages.
The color theme “Day” offers better readability in bright environments, e.g. with bright light shining on the screen
Import a CSV file containing your custom jumper cable velocities. See application
note for details
Various special functions or features can be activated or deactivated here.
It is advisable to keep “Check RL before PIM measurements” activated – it protects the unit from excessive VSWR and device damage caused by hot plugging.
Specify how long measurement data should be kept in the history. Too many stored measure­ments may slow down synchronization with remote software.
Select a predefined jumper cable or set an offset in dB. If enabled, output power and receive level are increased by that value to compensate cable loss.
WiFi Configuration: Activate internal HotSpot for remote control or join a WiFi
network. This feature is not available in the rack system
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 36 of 66
6.2 Filter / Band Selection
Filter units with multi-band support: Additional band is selected here. Standard filter unit: Band is automatically selected.
Selected filter unit
Selec­ted band
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 37 of 66
6.3 Measurement Screen
The following picture shows the measurement screen for “VSWR \ RL”. The explanations of the screen elements apply to all other measurement types too.
Settings: Adjust settings for current measurement Start (Stop): Start / Stop measurement Marker: Set / Adjust markers History: Show history of measurements Report: Create measurement report Comment: Add a comment to the measurement performed (appears in history and report) Exit: Exit PIM Analyzer application
Each measurement trace is kept for fourteen days (default) with a measurement comment and date/time. The time span can be changed in the Common Settings screen.
Selected
filter / band
Settings for
current measurement
Measurement type
Control buttons
6.3.1 History
Check entries to compare several measurement traces or reload them to add them to a report.
The history is synchronized with the tablet during connection – old measurements can be viewed and reports can be created offline using the tablet only.
6.4 Report Generation
6.4.1 Creating a Full Report
A report is not only a collection of measurements – it contains a complete set of documentation to provide evidence of the PIM performance of a “device under test”. It includes test parameters, comments, pass/fail check and title sheet with company logo, operator name, equipment serial and so on. By default, similar measurements relating to one section are grouped into a graph on one page to save paper and increase comparability; this can be switched off in the menu. The following graphics show examples of how reports can be structured:
The “Report” button can be found on every measurement screen. After a measurement has been performed, you can tap on it to display a sidebar menu.
A report is divided into sections which hold the individual measurements. To add a measurement to your report you will have to create a section first. If a section already exists you can select it here.
Add this measurement to the selected section of your report.
Show all the saved meas­urement graphs and create a final report with them. This function can also be accessed in the main menu via the “Report” button.
In the “Manage Report” screen all your saved measurements can be seen and arranged for the report.
Additionally you can set common attributes of your report.
Create PDF: The configured report is saved as a pdf file in the default folder automatically. You can copy the file to a USB stick or different location by hitting “Share” in the popup dialog.
New Report: Start a new blank report file, set the overall report title and create sections (e.g. Sector East, 2nd Floor, … or Production Lot1, Production Lot2, …). Measurements which were added are listed with Fre­quency Band, Measurement Type, Pass/Fail and Date in the selected target section.
Report History: Reload report configurations from the past (e.g. if you noticed a typo in a comment in the report yesterday, or failed to add a trace and want to overwrite the old report).
Report Settings: Configure operator’s name, logo and the default report output path. If “merge same meas­urements per section” is checked, all measurements with similar settings are combined in one graph with a legend showing the comments. If unchecked each measurement produces one page of paper.
6.4.2 Creating a Single-Page Report
6.4.3 Exporting Measurement Data as CSV
The raw data of each measurement trace can be exported into a comma-separated file (CSV) to enable cus­tomized postprocessing and data archiving. The CSV file is simply generated from the graph that is currently shown using the Report and Export as CSV buttons. The file name is automatically created from the meas­urement type and the current date + time. The file is saved to the folder selected in Reports -> Manage -> Settings.
The CSV file can be imported into Microsoft Excel with a semicolon separator and decimal dot. Important (especially in Germany): When files are opened directly with Excel, it uses the global operating system lan­guage setting for the decimal separator – therefore Excel will display a value of 1.2 as February 1st. Please change your regional Windows settings to a decimal dot instead of a comma. The file itself contains correct data – it is only shown differently in Excel.
We reserve the right to modify the data structure without notice.
6.4.4 Saving a Screenshot
Exports a picture of the current graph.
Export current measurement values into a CSV text file. Storage path can be set in “Report Settings” via “Man­age Report”
Export current measure­ment into a single-page PDF. Storage path can be set in “Report Settings” via “Manage Report”
6.5 Zeroing PIM
Before the first PIM DTF measurement, the unit must be zeroed. In the settings screen for the DTF meas­urement, choose “PIM Location” under “Measurement Mode”, check the “Enable” checkbox under “Setting New Zero Point” and apply the settings.
After that, connect the DTF zeroing adapter and a low PIM load at the desired zero point. To start zeroing PIM, click on “Zero PIM” at the bottom of the DTF measurement screen.
After successful zeroing, “PASS” is displayed and the zero point is then set. To avoid mistakes in subse­quent measurements, please uncheck the “Enable” checkbox in the DTF settings screen. The “Zero PIM” button on the measurement screen then disappears. The zeroing can also be used to set an offset.
6.5.1 Zeroing Return Loss
Return loss is zeroed without a load / adapter at the end of the cable. “Return Loss Location” must be checked in the DTF settings screen. Also make sure that “Output Power RL” is set to 23 dBm.
After successful zeroing, “PASS” is displayed and the zero point is then set. To avoid mistakes in subse­quent measurements, please uncheck the “Enable” checkbox in the DTF settings screen. The “Zero RL” button on the measurement screen then disappears.
6.5.2 Setting the velocity factor of the signal path (+Application Note DTF)
In order to increase the accuracy of distance measurements, the velocity factor of the signal path can be set up in sections. Enter the cable length and select a predefined cable from the table (or a custom-defined fac­tor) and hit “Add new cable” to create a list of subsequent cable sections from the zero point to the “end” of possible PIM occurrences.
The given example includes a 3 m jumper (1/2” R) and a 30 m feeder (7/8” R) + 1.5 m jumper (1/2” R). Eve­rything beyond the last entry is treated as “Air” with 100% because the antenna is almost unknown anyway.
6.6 Device Info
The “Device Info” screen shows useful information about your PIM Rack Analyzer device, e.g. manufacturing and calibration dates of the filters and base unit.
6.7 Defining Own Buttons for Section Names
When generating reports, measurement results are put into a file structure in the following hierarchy: Report name => Section => Measurement Type => Comment
Whereas measurement type information is provided automatically, the report name and comments are filled in manually by the operator, as this is individual measurement-specific data. However, in many applications, the designations for sectors are the same. In order to avoid having to create the same section names indi­vidually with every new report, it is possible to pre-define your own buttons for section names as default but­tons in the report menu. In order to use this functionality, SW version 2.10 or later is required.
The following example shows how to define your own buttons for section names step by step.
1. Open CSV file With SW version 2.10 (or later), the required .csv file will be installed automatically.
In order to process the data, open the .csv file from the following directory:
C:\Software\PimAnalyzer\userconfig Datei section_text_blocks.csv
2. In order to start defining your own buttons for section names, uncomment the last two lines by removing the hash key:
4. Replace the words in the last two lines with your desired section name. Make sure the field names are separated by a semicolon:
5. Close the .csv file and save the entered data.
6. Open the PIM Analyzer application and go to the report section. Now you can start to add measurement data to the new sectors by selecting “create new” in the sector section and selecting one of the pre-defined buttons.
6.8 Updating the PIM Analyzer Software
To benefit from the latest improvements and to ensure reliable and secure operation of the PIM Rack Ana­lyzer it is important to keep the software up to date. The following steps describe how to install software up­dates for the PIM Analyzer software:
1. Go to http://www.rosenberger.com/pia to download the latest software version and register for up­date notifications.
2. Close the PIM Analyzer application.
3. Run the Setup file PimAlphaSetup-x.x.exe on the PIM Analyzer’s Windows system, typically from a USB stick.
4. Start the PIM Analyzer application, and check the software version under Menu -> Device Info.
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 49 of 66
7 Remote Control
Remote control will be implemented soon in an upcoming release. Documentation will be provided as soon as the update is released. We apologize and ask for your understanding.
8 Maintenance of the Device
8.1 Handling
Only use the PIM Rack Analyzer in line with the intended purpose and comply with this documentation to prevent damage.
8.2 Cleaning
Prior to cleaning the product, disconnect it completely from the power supply (e.g. AC supply network or battery). Use a soft, non-linting cloth to clean the product. Never use chemical cleaning agents such as alcohol (except in the case of the test port), acetone or diluents for cellulose lacquers. Use pressurized air or alcohol-soaked cotton swabs to clean the test port. Ventilation outlets can be cleaned with pressurized air.
8.3 Calibration & Repair
Note that any repairs or calibration of the device not performed by Rosenberger can have a negative impact on the flawless operation of the device and will render any warranty void.
Device is deleted / reset to factory settings during calibration / service. Please save all required data beforehand.
Always use the original packaging material for shipping.
To ensure maximum measurement precision, we recommend a calibration interval of 12 months. If you wish to send back a unit for calibration or repair, please contact us prior to shipment to ensure a smooth transaction. Refer to the last chapter of this manual for global support addresses.
Calibration FAQs
- Calibration: Calibration of the unit based on the values defined in the initial factory-provided calibration. Check for latest firmware update.
- Calibration frequency: To ensure maximum measurement precision we recommend a calibration interval of 12 months.
- Cycle time: Standard cycle time is 10 working days after receipt of unit.
- Service options: We are happy to provide information about additional service options
PIM Rack Analyzer Manual
© Rosenberger 2018 - www.rosenberger.com/pia
Page 50 of 66
9 Measurement Examples
The following instructions are only recommendations. The actual requirements of PIM testing and verification can vary depending on specifications.
9.1 RF Measurements
General checklist for determining PIM:
1. Tighten RF connectors using appropriate torque
2. Clean connector interfaces with alcohol, Q-tip and compressed air
3. Replace components (e.g. feeder connector)
9.1.1 PIM Acceptance Test
1. Select appropriate filter unit according to your frequency requirements
2. Connect DUT to the test port of the selected filter unit (LED green)
3. Terminate DUT with a low PIM load (60Z150-001, -012, -020)
3. Perform a 2 Tone measurement while tapping on your DUT to simulate mechanical stress (e.g. wind)
4. Perform a sweep measurement over the widest possible frequency range
5. If PIM is not within specified limits, refer to 9.1.2
6. If PIM is within specified limits, create a test report and save it
9.1.2 Troubleshooting a PIM Problem
1. Select appropriate filter unit according to your frequency requirements
2. Connect DUT to the test port of the selected filter unit (LED green)
3. Terminate DUT with a low PIM load (60Z150-001, -012, -020)
4. Use the spectrum analyzer to evaluate interference signals from external sources (e.g. mobile phones or other operator creating strong PIM signals) and eliminate them if present
5. Perform a sweep measurement to evaluate whether the PIM depends on frequency
i. If no PIM is present, perform a 2 Tone measurement with a tap test
ii. If PIM is present, perform a DTF measurement to measure the distance to the
static PIM in meters. If the value is not within cabling length, an environmental influence is causing the problem (e.g. absorbers, other material in test chamber)
9. Fix the PIM problem you have located
10. Perform an acceptance test as described in 9.1.1.
10 Troubleshooting
10.1 Reporting Software Problems to Rosenberger
If you experience any problems relating to the use of the PIM Analyzer software, in order to make sure we can analyze the problem quickly and provide an effective solution, please follow these steps to provide relat­ed debug data to your Rosenberger sales/service partner:
If the issue is reproducible without using the tablet:
1. Make sure that the latest software is installed
2. Restart the PIM Analyzer software a. If the problem is that the PIM Analyzer software is not starting, execute
C:\Software\PimAnalyzer\PiaDebugData.exe instead
3. Reproduce the issue
4. Go to Device Info -> Debug Data and generate a debug data zip file.
5. Send the file to pia_service@rosenberger.com along with the exact steps describing what has been
done with the device
10.2 How to Avoid Common Problems
Problem
Probable causes
Possible remedy
Device will not turn on
- Device is not connected to power grid
- Main switch is turned off
- Malfunction within the power grid
Check all cabling connections for power supply. Check if the main switch is turned to ‘on’ and ensure your power grid is working properly.
Software is no longer re­sponding to any user ac­tions.
- Microsoft Windows operating system is not working properly anymore
Press and hold the power button until the device powers off. You can now turn it on by pressing the button again.
Unstable PIM
- Test port loose, worn out or dirty
- Interference from external device (e.g. mobile phone)
Change the test port connector. Use the spectrum analyzer function to ob­serve possible interferences.
11 Technical Specifications
11.1 Base Unit
11.1.1 Main Features
Broadband RX & TX base model 698-2200 MHz with outstanding PIM performance (typ. <-130 dBm)
Continuous wave signal (no pulse), conformity with IEC 62037 – 1, full power to PIM source
Up to 11 filters connected to one base unit
(optional 6-way / 11-way switch matrix)
No production downtime when setup is rearranged
Intuitive software operation
Automated report generation
Antenna isolation measurements
DTF measurement
o PIM vs. distance (< 0.3 m accuracy) o VSWR vs. distance
Measurement modes
o Passive intermodulation:
PIM vs. frequency ▪ PIM vs. time
o VSWR/return loss with high RF power o RF spectrum analyzer o Isolation measurement
Made for 24/7 production use, temperature-controlled fan system, easy-to-replace dust filter
CPRI PIM tests (HW option on request, option for later SW release)
11.1.2 Product Description
The versatile rack-type Passive Intermodulation Analyzer (PIA) system provides an easy way to precisely determine the intermodulation characteristics and other RF parameters in a wide frequency range for
Antennas
Transmission lines
Connectors
Jumpers
Filters and combiners
Splitters
This PIA is designed to measure the reversed/transmitted intermodulation products in production lines. The base unit (BU) is equipped with a Windows computer, RF baseband hardware and two broadband power amplifiers according to the frequency range. It can be operated manually from the built-in touchscreen or via a network connection either via remote desktop or a remote command interface for integration in production tools. The test system is completed by attaching one or more (up to eleven) filter inserts (FI) using a 1:1 cable or a switch matrix (MPX) which can be installed easily by the user. Band switching is carried out automatically when the measurement band is changed.
The user-friendly graphical touch interface enables easy report generation for every measurement, while guided measurements (planned in future release) force staff to follow predefined test patterns.
The test setup complies with the test methods suggested by proposal paper IEC 62037 (IEC SC 46D.WG6).
11.1.3 General
Display
9” touch screen, readable in sunlight
Dimensions
6 HU 550 mm (266 × 483 × 650 H×W×D)
Weight
43 kg
RF ports
- F TX out
- RX, isolation & VSWR
- Opt. switch matrix
N-type (if no switch matrix is used) SMA-type SMA (TX & RX) and SMP-type (VSWR)
User interface ports
2×USB, LAN
AC power supply
100-250 VAC
Frequency stability
± 2.5 ppm
11.1.4 RF PIM Analyzer (Base Unit) -150W Option
IM order
3rd, 5th, 7th, 9th, 11th, 13th, 15th, 17th
Output power At test port of most 3 dB
coupled filters*
26 – 52 dBm 23 – 46 dBm
Residual PIM
< -128 dBm (> 171 dBc @ 2x +43 dBm) <- 131 dBm (> 174 dBc @2x +43 dBm) typ.
PIM vs. distance
- Accuracy / resolution
- Range
< 0.3 m , all bands
Depends on number of PIM sources and accuracy of cable velocity factor
Down to -120 dBm PIM, 0 – 150 m
Frequency range (seamless)
698 … 2700 MHz
Filter units
Switchable between connected filter inserts via software
11.1.5 CPRI PIM Analyzer (SW Option)
Fiber interface
CPRI up to Rate 7
SFPs built into base unit (various types on request)
IM order
3rd, 5th,7th
Carrier types
LTE5, LTE10, LTE20, (LTE15 on request)
PIM range
-130 dBm noise floor (depends on RRH NF & bandwidth)
* Max. adjustable power is reduced automatically depending on hardware setup. Some special filters/bands may have higher attenuation and possible output power is reduced – see datasheet for details
11.1.6 RF Isolation Measurement
Frequency
Downlink frequency band of filter unit
RF output
+23 – 46 dBm
RF input
+27 dBm max. operating
+30 dBm max. no damage
+50 VDC max. no damage
Isolation
- Accuracy
- Resolution
0 – 60 dB
1.5 dB
0.1 dB
11.1.7 RF VSWR / Return Loss
Frequency
Downlink frequency band of filter unit
VSWR
1.10 – 20.00
Return loss
1.00 – 25.00 dB
Distance to VSWR fault
- Accuracy / resolution
- Range
0.2 m (typ.) Depends on number of mismatch sources and accuracy of cable velocity factor
0 – 150 m
11.1.8 RF Spectrum Analyzer
Frequency
Uplink frequency band of filter unit
Resolution bandwidth
120 Hz to 20 MHz RBW
Noise floor
-135 dBm DANL at 1 kHz
Amplitude accuracy
±1.0 dB typ, ±1.5 dB max
RF input
-40 dBm max. operating
+10 dBm max. no damage
11.1.9 Environmental
Operating temperature range
-10°C to +40°C
Storage temperature range
-20°C to +80°C
Relative humidity
5% to 95% RH non-condensing
Mechanical shock
1G
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 55 of 66
11.1.10 User Interfaces
Keyboard/USB memory 2 × USB A connector (one each on front and back panel)
Remote control 1 × LAN, 1 × Micro-USB B connector
Reference 1 × BNC 10 MHz
CPRI 2 × SFP built in on request, LC duplex on front panel
Isolation Port 1: 7-16 Port 2: SMA-type
Supply 1 × DC magnetic connector
11.2 Filter Units
11.2.1 Main Features
Outstanding PIM performance
Industrial 24/7 ready
o Forced air cooled depending on usage o Easy-to-replace air filters
Plug-and-play system reconfiguration
o Automatic detection in base unit
o Factory calibrated
Wideband hybrid coupler design
Space-saving 3 HU height
Field-replaceable test port
o 7-16 DIN and 4.3-10 DIN
11.2.2 General
Dimensions (w/o connectors) 482.6 × 550 × 132.5 mm (W×L×H)
19”, 3 HU, depth 550 mm
Weight 10-16 kg (depends on frequency band)
Internal ports 3 × N, 2 × SMA, 2 × HSD
RF port Field-replaceable test port 7-16 DIN, 4.3-10 DIN (opt)
11.2.3 Environmental
Operating temperature range -10°C to +40°C
Storage temperature range -20°C to +80°C
Relative humidity 5% to 95% RH non-condensing
Mechanical shock 1G
11.2.4 Compatibility / Supported Base Unit
The filter unit must correspond to the frequency range of the base units.
Base unit 0722
Filter units 07, 08, …, 21
Base unit 2127
Filter units 21, …, 26
11.2.5 IM-R-FI-07/B12-14-R (LTE700 L+U)
ETSI band
B12 - 14, B17
Transmit path
- Range
728 – 764 MHz
Receive path
- Range B12+17
- Range B13+14
698 – 716 MHz 776 – 798 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
15.5 kg
11.2.6 IM-R-FI-07/B28-R (APT700)
ETSI band
B28
Transmit path
- Range
758 – 803 MHz
Receive path
- Range
703 – 748 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
16 kg
11.2.7 IM-R-FI-08/B20-R (DigDiv800)
ETSI band
B20
Transmit path
- Range
791 – 821 MHz
Receive path
- Range
832 – 862 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
13.5 kg
11.2.8 IM-R-FI-08/B5-R (AMPS850)
ETSI band
B5
Transmit path
- Range
869 – 894 MHz
Receive path
- Range
824 – 851 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
13 kg
11.2.9 IM-R-FI-09/B8-R (EGSM900)
ETSI band
B8
Transmit path
- Range
925 – 960 MHz
Receive path
- Range
880 – 915 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
13 kg
11.2.10 IM-R-FI-14/B11+21-R (LTE1400)
ETSI band
B11 + 21
Transmit path
- Range
1475.9 – 1510.9 MHz
Receive path
- Range
1427.9 – 1462.9 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
kg
11.2.11 IM-R-FI-18/B3-R (DCS1800)
ETSI band
B3
Transmit path
- Range
1805 – 1880 MHz
Receive path
- Range
1710 – 1785 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
11.5 kg
11.2.12 IM-R-FI-19/B2+4-R (PCS/AWS1900)
ETSI band
B2 + 4
Transmit path
- Range TX1
- Range TX2 (B2 PCS)
- Range TX2 (B4 AWS)
1930 – 1950 MHz 1970 – 1995 MHz 1970 – 2155 MHz
Receive path
- Range (B2 PCS)
- Range (B4 AWS)
1850 – 1910 MHz 1710 – 1755 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
13.5 kg
11.2.13 IM-R-FI-21/B1-R (UMTS2100)
ETSI band
B1
Transmit path
- Range
2110 – 2170 MHz
Receive path
- Range
1920 – 2060 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
12 kg
11.2.14 IM-R-FI-23/B30-R (WCS) (opt. on request)
ETSI band
B30
Transmit path
- Range
2345 MHz – 2360 MHz
Receive path
- Range
2305 MHz – 2335 MHz
Residual PIM
-128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
kg
11.2.15 IM-R-FI-26/B7-R (LTE II 2600)
ETSI band
B7
Transmit path
- Range
2620 – 2695 MHz
Receive path
- Range
2445 – 2590 MHz
Residual PIM
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
Weight
12.5 kg
11.3 Part Number Designation
IM-R-FI-xxxx-yy
Base unit
xxxx: 0722: 2127:
yy: 150W:
700-2200 MHz broadband amplifier & receiver 2100-2200 MHz broadband amplifier & receiver
150 W output power
M-R-MPX-xxxx
Base unit
xxxx: 6way:
11way:
6-way switch matrix (connect up to 6 filters)
11-way switch matrix (connect up to 11 filters)
IM-R-FI-xxxx-y
Filter unit 7-16 test port
xxxx: 07/B12-14:
LTE700LU (ETSI band 12 to 14)
07/B28:
APT700 (ETSI band 28)
08/B20:
DigDiv (ETSI band 20)
08/B5:
AMPS (CDMA 800) (ETSI band 5)
09/B8:
EGSM (ETSI band 8)
14/B11+21:
LTE1400 (ETSI band 11 & 21)
18/B3:
DCS (ETSI band 3)
19/B2+4:
PCS + AWS (ETSI band 2 & 4)
21/B1:
UMTS (ETSI band 1)
23/B30:
WCS (ETSI band 30)
26/B7:
W UMTS II / LTE II (ETSI band 7)
y: R:
Reflected PIM measurements (1 port)
T:
Transmitted and dual-port measurements (2 ports) (on request)
IM-R-HWO-xxxx
Hardware option
xxxx: ExtCtrl
External control interface to enable/disable amplifiers and receive RF-ON warning
IM-R-SWO-xxxx
Software option
xxxx:
Available on request
IM-R-ACSRY-xxxx
Accessory
xxxx: Fil-BU:
9 cs air filter mats for base unit
52 cs air filter mats for filter unit
11.4 Order Number Example
IM-R-BU-0722-150W
Base unit with broadband amplifier 700-2700 MHz, receiv­er, fiber CPRI unit
IM-R-MPX-6way
Single battery pack
IM-R-FI-07/B28-R
Filter unit for APT700 (ETSI band 28)
IM-R-FI-19/B2+4-R
Filter unit for EGSM900 (ETSI band 2+4)
IM-R-ACSRY-Fil-BU
Filter mats for base unit
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 61 of 66
12 CE Declaration of Conformity
Place and date:
Fridolfing, August 21st, 2007, update
Manufacturer:
Rosenberger Hochfrequenztechnik GmbH & Co. KG
Address:
Hauptstraße 1, 83413 Fridolfing, Germany
declare under our sole responsibility that the product
Intermodulation test set
IM-07S, IM-08S, IM-09S, IM-18S, IM-19S, IM-21S, IM-26S, IM-35S
IM-07P, IM-08P, IM-85P, IM-09P, IM-18P, IM-19P, IM-21P, IM-26P, IM-35P
IM-07P-BB, IM-08P-BB, IM-85P-BB, IM-09P-BB, IM-18P-BB, IM-19P-BB, IM-21P-BB,
IM-0710-BB, IM-1822-BB, IM-2526-BB, IM-3435-BB, IM-0722-BB
IM-A-BU-0727, IM-B-BU-0727, IM-R-BU-0722-150W, IM-R-BU-2127-150W
to which this declaration relates is in conformity with the following EC directives:
73/23/EEC Low Voltage Directive 89/336/EEC Electromagnetic Compatibility Directive
and complies with the following standards or normative documents:
Safety
- EN 61010-1-1
EMC
- EN 55022
- EN 50082-2
- EN55011: Group 1 Class A
- EN 61000-4-2
- EN 61000-4-3
- EN 61000-4-4
- EN 61000-4-6
Christian Entsfellner
Product Manager
R&D PIA Manager
13 Accessories
IM-R-MPX-6way or -11way
Switch matrix with support for a maximum of 6 or 11 filters
IM-R-ACSRY-Fil-BU or -FI Set of dust air filter mats for base/filter unit
IM-R-HWO-ExtCtrl External control / signaling interface
13.1 Filter Units
Rosenberger no.
Frequency band
ETSI band
Transmit path
range
Receive path
range
Residual PIM
IM-R-FI-06/B71-R (opt. on request)
71
617 – 652 MHz
663 – 698 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-07/B12-14-R
(LTE700 L+U)
12 ­14, 17
728 – 764 MHz
B12+17 698 – 716 MHz B13+14 776 – 798 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-07/B28-R
APT 700
28
758 – 803 MHz
703 – 748 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-08/B20-R
DigDiv 800
20
791 – 821 MHz
832 – 862 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-08/B5-R
AMPS850
5
869 – 894 MHz
824 – 851 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-09/B8-R
EGSM 900
8
925 – 960 MHz
880 – 915 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-14/B11+21-R
LTE 1400
11, 21
1475.9 – 1510.9 MHz
1427.9 – 1462.9 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-18/B3-
DCS 1800
3
1805 – 1880 MHz
1710 – 1785 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-19/B2+4-R
PCS/AWS1900
2, 4
TX1 1930 – 1950 MHz TX2 (B2 PCS) 1970 – 1995 MHz TX2 (B4 AWS) 1970 – 2155 MHz
(B2 PCS) 1850 – 1910 MHz (B4 AWS) 1710 – 1755 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-21/B1-R
UMTS2100
1
2110 – 2170 MHz
1920 – 2060 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-23/B30-R (opt. on request)
WCS
30
2345 MHz – 2360 MHz
2305 MHz – 2335 MHz
-128 dBm (< -171 dBc), referred to 2 × +43 dBm
IM-R-FI-26/B7-R
LTE II 2600
7
2620 – 2695 MHz
2445 – 2590 MHz
< -128 dBm (< -171 dBc), referred to 2 × +43 dBm
13.2 Other Accessories
Part number
Description
Useful acces-
sories
IM-R-HWO-ExtCtrl
External control/signaling interface
IM-R-ACSRY-Fil-BU
10 air filter mats for base unit (420x375x15 mm, polyolefin, filter class G2, flammability F1)
IM-R-ACSRY-Fil-FI
50 air filter mats for filter unit (130x45x15 mm, polyolefin, filter class G3, flammability F1)
Spare parts
60S101-KIMN1
7/16 test port saver for use on filter
60S164-K00N1
4.3/10 test port saver for use on filter
Low PIM adapters
60S101-KIMN1
7/16 male to 7/16 female adapter
60S101-SIMN1
7/16 male to 7/16 male adapter
60K101-KIMN1
7/16 female to 7/16 female adapter
60S153-KIMN1
7/16 male to N female adapter
53S160-KIMN1
7/16 female to N male adapter
60S164-K00N1
7/16 male to 4.3/10 female adapter
60S164-S00N1
7/16 male to 4.3/10 male adapter
60K164-S00N1
7/16 female to 4.3/10 male adapter
PIM stand-
ards
60S110-KxxN1
-110 dBm standard adapt-
er (band-specific) included with every filter unit (7/16 DIN type)
64S110-KxxN1
-110 dBm standard adapt-
er (band-specific) included with every filter unit (4.3/10 DIN type)
xx: 07 LTE700; 08 DigDiv|AMPS; 09 EGSM; 18
DCS; 19 PCS/AWS;
21 UMTS; 26 LTE2600; other frequencies on re-
quest
Low PIM ter-
minations
60Z150-001
Low PIM termination (19” rack type, 3 HU)
60Z150-012
Low PIM termination (benchtop type)
60Z150-020
Low PIM termination (portable, with male & female ports)
Tools
60W000-002
32 mm torque wrench
53W010-000
18 mm torque wrench
99W057-000
Adjustable flat wrench
Corrugated
Test cables
LC02-186-4000
Test cable 7/16 male / 7/16 male 4.0 m
LC02-186-1500
Test cable 7/16 male / 7/16 male 1.5 m
LC02-188-4000
Test cable 7/16 male / N male 4.0 m
LC02-188-1500
Test cable 7/16 male / N male 1.5 m
SLJ12SP-60M64M-2.0m-00
Test cable 7/16 male / 4.3/10 male 2.0 m
SLJ12SP-64M64M-2.0m-00
Test cable 4.3/10 male / 4.3/10 male 2.0 m
Super flex
test cables
IM-Cable-716m-716m-3000
Test cable 7/16 male / 7/16 male 3.0 m
IM-Cable-716m-4310m-3000
Test cable 7/16 male / 4.3/10 male 3.0 m
IM-Cable-4310m-4310m-3000
Test cable 4.3/10 male / 4.3/10 male 3.0 m
14 Support and Sales Locations
14.1 Europe, Middle East, Africa
Rosenberger Hochfrequenztechnik GmbH & Co. KG
Hauptstraße 1 83413 Fridolfing, Germany
Phone +49 8684 18-0
Fax +49 8684 18-1499
info@rosenberger.de
www.rosenberger.com
14.2 Americas
Rosenberger Site Solutions, LLC
P.O. Box 8817, Lake Charles, LA 70606, USA
Phone +1 337 598 5250
Fax +1 337 598 5290
rlss@rlss.us www.rlss.us
14.3 Brazil
Rosenberger Domex Telecom
Cabletech Avenue, 601
Guaramirin CEP 12295-230
Cacapava – Sao Paulo, Brazil
Phone +55 12 3221 8500
Fax +55 12 3221 8543
vendas@rosenbergerdomex.com.br www.rosenberger.com
14.4 Asia Pacific
Rosenberger Asia Pacific Electronic Co., Ltd.
No. 3, Anxiang Road, Block B
Tianzhu Airport Industrial Zone
Beijing, 101300, PR China
Phone +86 10 80 48 1995
Fax +86 10 80 48 2438
info@rosenbergerap.com
www.rosenbergerap.com
14.5 India
Rosenberger Electronic Co. (India) Pvt Limited
Plot No. N3B3, Phase-IV Verna Industrial Estate
IND - 403722 Goa
info@rosenbergerap.com
www.rosenbergerap.com
PIM Rack Analyzer Manual
© Rosenberger 2018 -
www.rosenberger.com/pia
Page 66 of 66
Hochfrequenztechnik GmbH & Co. KG Hauptstraße 1
| 83413 Fridolfing
P.O. Box 1260
| 84526 Tittmoning
Germany Phone +49 8684 18
-0
info@rosenber
ger.com
www.rosenberger.com Certified
to ISO/TS 16949 · DIN EN 9100 · ISO 9001 · ISO 14001
Order
no. 394581 IM-R-ACSRY-OM-en
8/2018
Rosenberger
®
is a registered trademark of Rosenberger Hochfrequenztechnik GmbH & Co. KG.
All rights reserved.
Loading...