MODEL 56 SPECIFICATIONS
Case: Polycarbonate.
Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132mm)
Conduit openings: Accepts (6) PG13.5 or 1/2 in. con-
duit fittings
Display: Large 3.75 x 2.2 in. (95.3 x 55.9mm) high res-
olution color LCD displays large process variables and
user-definable display of diagnostic parameters.
Calibration, programming and information screens display
clear, easy-to-read characters. The color display is
back-lit and backlighting intensity is user adjustable.
Measurement character height: (.5") 13mm. Main display
can be customized to meet user requirements.
Ambient temperature and humidity: -10 to 60°C, (14
to 140°F) RH 5 to 95% (non-condensing). For Turbidity
only: 0 to 55°C (32 to 131°F). RH 5 to 95% (non-condensing).
Note: Some degradation in display response or performance may occur below -5°C (23°F) and above
55°C (131°F). Above 60°C, the following components
will progressively and automatically shut down: display,
USB communications port, current outputs, alarm
relays, main circuit board.
WARNING: Always remove USB memory device if
ambient temp exceeds 60°C. Do not access USB
port if combustible atmosphere is present.
Storage temperature: -20 to 60°C, (-4 to 140°F)
Power: Code -02: 20 to 30 VDC. 20 W
Code –03: 85 to 264 VAC, 47.5 to 65.0 Hz,
20 W
Real time clock back-up: 24 hours.
Ordinary Locations:
Options for UL: 02, 03, 20, 21, 22, 24, 25,
26, 27, 30, 31, 32, 34, 35, 36, 37, 38, and HT.
Pollution Degree 2: Normally only non-conductive
pollution occurs. Occasionally, however, a temporary
conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
RFI/EMI: – EN-61326
LVD: – EN-61010-1
Input: One or two isolated sensor inputs. Measurement
choices of pH/ORP, resistivity/conductivity/TDS, % concentration, ratio conductivity, total and free chlorine,
monochloramine, dissolved oxygen, dissolved ozone,
turbidity, pulse flow, temperature and raw 4-20mA input.
For contacting conductivity measurements, temperature
element must be a Pt1000 RTD. For other measurements (except ORP, flow and turbidity), use either a
PT100 RTD, PT1000 RTD, or 22k NTC (D.O. only).
Outputs: Four actively powered 4-20 mA or 0-20 mA
isolated current outputs. Fully scalable. Max Load: 550
Ohms. Outputs can be programmed for PID control.
Output dampening can be enabled with time constants
from 0 to 999 seconds. HART digital communications
which is superimposed on output 1 is standard on all
units (option code –HT).
Alarms: Four alarm relays for process measurement(s) or temperature. Any relay can be programmed
for any measurement, timer, TPC or fault alarm operation, instead of a process alarm. When selected, a fault
alarm will activate the relay when a sensor or analyzer
fault occurs. Each relay can be configured independently. Alarm logic (high or low activation or USP*) and
deadband are user-programmable.
*USP alarm can be programmed to activate when the conductivity is
within a user-selectable percentage of the limit. conductivity/resistivity
measurement only)
Relays: Form C, SPDT, epoxy sealed
Inductive load: 1/8 HP motor (max.), 115/240 VAC
Terminal Connections Rating:
Power connector (-02 24VDC power supply and -03
85-264VAC power supply): 24-12 AWG wire size.
Signal board terminal blocks: 26-16 AWG wire size.
Current output connectors: 26-16 AWG wire size.
Alarm relay terminal blocks: 24-12 AWG wire size.
Weight/Shipping Weight: (rounded up to nearest lb or
nearest 0.5 kg): 3 lbs/4 lbs (1.5 kg/2.0 kg)
Maximum Relay Current
Power Resistive
28 VDC 5.0 A 5.0 A
115 VAC 5.0 A 5.0 A
230 VAC 5.0 A 5.0 A