The Model 396R
Sensor is specifically
designed for improved
life in harsh, dirty applications where a separate sample stream is
difficult to provide and
greater insertion depths
are required. It is designed for use with a 1-1/4 in. or
1-1/2 in. ball valve for hot tap installation.
The Model 396R is constructed of molded polypropylene housed in a titanium tube with EPDM seals to
provide maximum chemical resistance. The sensor
also features a titanium solution ground for advanced
sensor diagnostics when used with the Model 54 or
Model 3081 pH/ORP Analyzer/Transmitter. These
Rosemount Analytical 3
advanced sensor diagnostics provide preventative
maintenance by notifying the operator of the need for
replacement and cleaning of an aged or fouled sensor,
thus allowing continuous optimum performance.
The sensor also features a slotted tip for protection
from breakage while allowing the process to flow by the
glass electrode for accurate and reliable pH measurement.
The Model 396R is only available without an integral
preamplifier and has 15 ft or 9.5 in. of integral high quality 9 conductor cable. The preamplifier must be housed
in a remote location or in a junction box kit for attachment at the rear, cable end of the sensor (order separately) or must be integral to the Analyzer/Transmitter.
The Model 396R is compatible with all Rosemount
Analytical and other manufacturers’ instruments.
SPECIFICATIONS
MODEL 396 MODEL 396P MODEL 396R
Measurements and Ranges pH: 0-14 pH: 0-14 pH: 0-14
ORP: -1500 to 1500 mv ORP: -1500 to 1500 mv
Available pH ACCUGLASS GPHT hemi bulb or GPLR hemi bulb GPLR hemi bulb
Types GPHT flat bulb
Wetted Materials 316 SST, Polypropylene, Titanium, Polypropylene, Titanium, Polypropylene,
EPDM, glass EPDM, glass: EPDM, glass:
platinum (ORP only) platinum (ORP only)
Process Connection None, use 1 in. NPT process 1 in. MNPT front and rear None, use 1-in. process connector
connector, PN 23166-00 or facing threads or ball valve kit (1-1/2 in.
23166-01 (sold separately) or - 1/4 in.) (sold separately)
Temperature Range 0-100°C (32-212°F) 0-100°C (32-212°F) 0-100°C (32-212°F)
Pressure Range-
100-1136 kPa (0-150 psig) 100-1136 kPa (0-150 psig) 100-1136 kPa (0-150 psig)
Hemi bulb
Pressure Range-
100-790 kPa (0-100 psig) — —
Flat bulb
Maximum Pressure at Not Applicable Not Applicable Code 21: 542 kPa (64 psig)
Retraction or Insertion Code 25: 343 kPa (35 psig)
Minimum Conductivity 75 µS/cm, nominal 75 µS/cm, nominal 75 µS/cm, nominal
100 µS/cm 100 µS/cm 100 µS/cm
Integral Cable 15 ft coaxial with BNC Code 01: 25 ft 11 Standard 15 ft coaxial
conductor Code 60: 9.5 in. coaxial
Code 02: 15 ft coaxial with BNC
Code 61: 9.5 in. coaxial
Preamplifier Options Remote Remote or Integral Remote
Weight/Shipping Weight 0.45 kg/0.9 kg (1 lb/2 lb) 0.45 kg/0.9 kg (1 lb/2 lb) Sensor: Code 21:
0.9 kg/1.40 kg (2.0 lb/3.0 lb)
Code 25: 1.40 kg/1.80 kg
(3.0 lb/4.0 lb)
Ball Valve PN 23240-00:
2.25 kg/3.20 kg (5.0 lb/7.0 lb)
Junction box; 1.40 kg/1.80 kg
(3 lb/4 lb)