The TUpH helical reference pathway resists reference poisoning.
Ions diffuse through the reference pathways and a charge is
passed to the reference element. The reference element must
be protected from contamination by poisoning ions such as
sulfide, mercury, cyanide, and ammonia that will cause reference
drift and offsets. The TUpH sensor's long internal reference
pathway hinders the contaminants' migration to the reference
element thereby providing an increased sensor life.
All TUpH sensor models have been specifically designed for
improved performance in harsh, dirty, or abrasive applications
where large quantities of suspended solids are present.
The Model TF396 offers a watertight sensor-to-cable connector
that prevents cable twisting and eliminates the need for rewiring
when replacing the sensor. The sensor cable includes an integral preamplifier cap with mating connector and is compatible
with most Rosemount Analytical analyzers and transmitters.
SPECIFICATIONS
Measurement Range: 2-12 pH
Measuring element: ISFET
Repeatability: ±0.05 pH
Response time to pH change: 99% in 30 seconds step change
Wetted Materials: Code -01: Polypropylene, EPDM, Ryton, Silicon.
Code -02: 316SST, Titanium, Polypropylene, EPDM, Ryton, Silicon
Process Connections: Code -01: 1 in. MNPT front and rear facing threads
Code -02: 2 in. Tri-Clamp connection
Temperature Range: 0 to 100°C (32 to 212°F) @ 50 psig
Pressure Range: -10 to 100 psig (32 to 790 kPa [abs]) at 50°C
Minimum Process Conductivity: 100 µS/cm
Cable Lengths: 20 ft, 50 ft (6 m, 18 m)
Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)