T56 Product Data Sheet
LIQ_PDS_T56 August 2014
T56
Clarity®II Turbidimeter
COMPLETE SYSTEM includes single or dual input
analyzer, sensor(s), and debubbler assembly
CHOOSE U.S. EPA METHOD 180.1 or ISO
METHOD 7027 compliant sensors
RESOLUTION 0.001 NTU
EASY TO USE 56 ANALYZER with four analog
outputs and four programmable alarm relays
INTUITIVE, USER-FRIENDLY MENU in nine
languages makes setup and calibration easy
HART
Data and Event Logger – stores thirty days of
®
DIGITAL COMMUNICATIONS – standard
data for output on USB 2.0 memory stick
Features and Applications
The Clarity II turbidimeter is intended for the determination of
turbidity in water. Low stray light, high stability, efficient bubble
rejection, and a display resolution of 0.001 NTU make Clarity II
ideal for monitoring the turbidity of filtered drinking water. Clarity
II is also suitable for applications other than drinking water treatment including monitoring of filtered water discharges, condensate returns, and clarifiers.
Both US EPA 180.1 and ISO 7027 compliant sensors are available.
US EPA 180.1 sensors use a visible light source. ISO 7027 sensors
use a near infrared LED. For regulatory monitoring in the United
States, US EPA 180.1 sensors must be used. Regulatory agencies in
other countries august have different requirements.
The Clarity II turbidimeter consists of an analyzer, with either one
or two signal input boards, sensor(s), and a debubbler/measuring
chamber and cable for each sensor. The cable plugs into the sensor
and the analyzer, making setup fast and easy. Sensors can be
located as far as 50 ft (15.2 m) away from the analyzer.
The Clarity II turbidimeter incorporates the easy-to-use 56 intelligent analyzer. Menu screens and prompts are so intuitive on the
large LCD display that a manual is practically not needed. 4-20mA
analog outputs are fully scalable. Every unit includes four alarm
relays with interval timer functions. Alarms are fully programmable for high/low logic and dead band. To simplify start-up and
programming, the analyzer automatically recognizes installed
turbidity signal boards and detects whether an EPA 180.1 or ISO
7027 sensor is being used.
Clarity II is available in an optional configuration in which the
analyzer, sensor(s), and debubbling flow cell(s) are mounted on a
single back plate. The sensor cables are pre-wired to the analyzer,
so setup is exceptionally fast and easy. All the user does is mount
the unit on a wall, bring in power and sample, and provide a drain.
To order this option, consult the factory.
T56 August 2014
Specifications - Analyzer
Enclosure: Polycarbonate. Type 4X/CSA 4X IP66
Current Output Accuracy: ±0.05 mA @25°C
Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132 mm)
onduit openings: Accepts 1/2” or PG 13.5 conduit fittings.
C
Display: Large 3.75 x 2.2 in. (95.3 x 55.9 mm) high resolution
color LCD for large process variables and user-definable display
of diagnostic parameters. Back-Lighting is user adjustable. Main
display can be customized to meet user requirements.
Measurement character height: 0.5 in. (13mm)
Security Code: 3-digit code prevents accidental or unauthorized
changes in instrument settings and calibration.
Languages: English, French, German, Italian, Spanish,
Portuguese, Chinese, Russian and Polish
Units: Turbidity (NTU, FTU, or FNU); total suspended solids
(mg/L, ppm, or no units)
Display resolution-turbidity: 4 digits; decimal point moves
from x.xxx to xxx.x
Display resolution-TSS: 4 digits; decimal point moves from
x.xxx to xxxx
Calibration methods: User-prepared standard, commercially
prepared standard, or grab sample. For total suspended solids
user must provide a linear calibration equation.
Ambient Temperature and Humidity: 0 to 55°C, (32 to 131°F);
RH 5 to 95% (non-condensing)
Altitude: For use up to 2000 meters.
Storage Temperature: -20 to 60°C, (-4 to 140°F)
Real time clock back-up: 24 hours
Power: 85 to 265 VAC, 47.5 to 65.0 Hz. 20W min. input power.
Equipment protected by double insulation.
Inputs: Single or dual input, EPA 180.1 or ISO 7027 sensors
Outputs: Four 4-20 mA or 0-20 mA isolated current outputs.
Fully scalable. Maximum load: 550 ohm. Output 1 has superimposed HART signal
Output Dampening: 0-999 seconds
Alarms: Four process alarm relays for turbidity or temperature.
Relays can also be programmed for timer, TPC, or fault alarm
operation instead of as a process alarm. Each relay can be
configured independently. Alarm logic (high or low activation)
and deadband are user-programmable.
Terminal Connections Rating:
Power connector (3-leads); 24-12 AWG wire size.
Signal board terminal blocks; 26-16 AWG wire size.
Current output connectors (2-leads); 24-16 wire size.
Alarm relay terminal blocks: 24-12 AWG wire size.
RFI/EMI: EN-61326
LVD: EN-61010-1
Hazardous Location Approvals – Analyzer
CSA approvals:
Class I, Division 2, Groups A, B, C, & D
Class Il, Division 2, Groups E, F, & G
Class Ill T4A Tamb= 50°C
Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indicators
adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in
Canada and the U.S. respectively.
FM approvals:
Class I, Division 2, Groups A, B, C, & D
Class Il & lll, Division 2, Groups E, F, & G
T4 Tamp = -10 deg C to 60 deg C
Relays: Form C, single pole, double throw, epoxy sealed.
Maximum Relay Current
Power Resistive
28 VDC 5.0 A 5.0 A
115 VAC 5.0 A 5.0 A
230 VAC 5.0 A 5.0 A
Field wiring terminals: Removable terminal blocks for power,
analog outputs, and sensors
2 RosemountAnalytical.com
August 2014 T56
Specifications - Sensor
Method: EPA 180.1 or ISO 7027 (using 860 nm LED source).
Must be specified when ordering.
Incandescent lamp life: one year (EPA 180.1)
LED life: three years (ISO 7027)
®
Wetted materials: Delrin
Accuracy after calibration at 20.0 NTU:
0 -1 NTU: ±2% of reading or ±0.015 NTU, whichever is greater.
0 - 20 NTU: ±2% of reading
NOTE: turbidity values of 2-200 NTU can be measured but frequent
cleaning may be required to maintain turbidity measurements.
Cable: 20 ft (6.1 m) or 50 ft (15.2 m). Maximum 50 ft (15.2 m).
Connector is IP65.
Maximum Pressure: 30 psig (308 kPa abs)
Temperature: 40 - 95°F (5 - 35°C)
Sensor body rating: IP65 when cable is connected
, glass, EPDM
Specifications - General
Weight/shipping weight:
Sensor: 1 lb/2 lb (0.5 kg/1.0 kg)
Analyzer: 2 lb/3 lb (1.0 kg/1.5 kg)
Debubbler: 3 lb/4 lb (1.5 kg/2.0 kg)
(rounded to the nearest lb or 0.5 kg)
Specifications subject to change without notice.
Specifications - Debubbler
and Flow Chamber
Dimensions: 18.1 in. x 4.1 in. diam. (460 mm x 104 mm diam.)
(approx.)
Wetted materials: ABS, EPDM, Delrin
Inlet: Compression fitting accepts 1/4 in. OD tubing; fitting can
be removed to provide 1/4 in. FNPT
Drain: Barbed fitting accepts 3/8 in. ID tubing; fitting can be
removed to provide 1/4 in. FNPT. Must drain to atmosphere.
Sample temperature: 40 - 95°F (5 - 35°C)
Minimum inlet pressure: 3.5 psig (125 kPa abs). 3.5 psig will
provide about 250 mL/min sample flow.
Maximum inlet pressure: 30 psig (308 kPa abs). Do not block
drain tube.
Recommended sample flow: 250 - 750 mL/min
1
Delrin is a registered trademark of DuPont Performance Elastomers
1
, polypropylene, nylon
RosemountAnalytical.com 3
T56 August 2014
ILLIMETER
M
INCH
Front View
Bottom View
Side View
Panel mount dimensions. The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
4 RosemountAnalytical.com