Rosemount Analytical T1056 Operating Manual

Clarity II
Turbidimeter
Turbidity Measurement System
Instruction Manual
PN 51T1056/rev.D August 2014
Emerson Process Management
http://www.rosemountanalytical.com
© Rosemount Analytical Inc. 2014
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your instrument purchase from Rosemount Analytical, Inc. is one of the finest available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their con tinued operation to the design specifications, per sonnel should read this manual thoroughly before proceeding with installation, commissioning, opera tion, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property dam age; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 18008548257 or 9497578500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protec tive covers must be in place unless qualified per sonnel are performing maintenance.
Equipment protected throughout by double insulation.
• Installation and servicing of this product may expose personel to dangerous voltages.
• Main power wired to separate power source must be disconnected before servicing.
• Do not operate or energize instrument with case open!
• Signal wiring connected in this box must be rated at least 240 V.
• Nonmetallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and jumper wires.
• Unused cable conduit entries must be securely sealed by nonflammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
• Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA70) and/or any other applicable national or local codes.
• Operate only with front panel fastened and in place.
Proper use and configuration is the responsibility of the user.
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
This product is not intended for use in the light industrial, residential or commercial environments per the instru ment’s certification to EN500812.
WARNING
RISK OF ELECTRICAL SHOCK
CAUTION
CAUTION
QUICK START GUIDE
FOR CLARITY II TURBIDIMETER
1. Refer to Section 2.0 for installation instructions.
2. The sensor cable is prewired to a plug that inserts into a receiving socket in the analyzer. The cable also passes through a strain relief fitting. To install the cable…
a. Remove the wrenching nut from the strain relief fitting.
b. Insert the plug through the hole in the bottom of the enclosure nearest the sensor socket. Seat the fitting in the
hole.
c. Slide the wrenching nut over the plug and screw it onto the fitting.
d. Loosen the cable nut so the cable slides easily.
e. Insert the plug into the appropriate receptacle on the circuit board.
f. Adjust the cable slack in the enclosure and tighten the cable nut. For wall/pipe mounting, be sure to leave suffi
cient cable in the enclosure to avoid stress on the cable and connections.
g. Plug the cable into the back of the sensor.
h. Place the sensor in either the measuring chamber or the calibration cup. The sensor must be in a dark place
when power is first applied to the analyzer.
3. Make power, alarm, and output connections as shown in section 3.0 wiring.
4. Once connections are secured and verified, apply power to the analyzer.
5. When the analyzer is powered up for the first time Quick Start screens appear. Follow the Quick Start Guide to enable live readings.
a. A blinking field shows the position of the cursor.
b. Use the t or u key to move the cursor left or right. Use the p or q key to increase or decrease the value of a
digit. Use the p or q key to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the language selection screen.
IMPORTANT NOTE:
When using EPA/incandescent sensors (PN 801080002EPA):
n DO NOT power up the instrument without the sensor connected n DO NOT disconnect and reconnect a sensor while an analyzer is powered
If this is inconvenient or cannot be avoided:
1. Cycle power to the instrument after connecting the sensor or..
2. Perform a Slope Calibration or Standard Calibration routine after connecting the sensor. Following these guide
lines will extend the life of the incandescent lamp and avoid premature warnings and faults due to reduced
lamp life.
CONTINUED ON THE FOLLOWING PAGE
QUICK START GUIDE
Figure A. QUICK START GUIDE, Model 1056
QUICK REFERENCE GUIDE
Figure B. MODEL 1056 MENU TREE
About This Document
This manual contains instructions for installation and operation of the Clarity II Model T1056 Turbidimeter.
The following list provides notes concerning all revisions of this document.
Rev. Level
Date Notes
A 12/07 This is the initiallaunch version.
B 2/09 Update Specifications
C 12/09 Update DNV logo, company name
D 08/14 Change turbidity specifications, add CSA Nonincendive field
wiring installation drawings
MODEL CLARITY II TABLE OF CONTENTS
MODEL CLARITY II TURBIDIMETER
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS .............................................................................. 1
1.1 Features and Applications...................................................................................................... 1
1.2 Specifications ......................................................................................................................... 2
2.0 INSTALLATION...................................................................................................................... 5
2.1 Unpacking and Inspection ...................................................................................................... 5
2.2 Installation ............................................................................................................................. 5
2.3 Installation — Debubbler Assembly........................................................................................ 8
2.4 Installation — Sensor ............................................................................................................. 10
2.5 Sample Point .......................................................................................................................... 10
3.0 WIRING .................................................................................................................................. 13
3.1 General................................................................................................................................... 13
3.2 Preparing Conduit Openings ................................................................................................. 13
3.3 Preparing Sensor Cable ......................................................................................................... 14
3.4 Power, Output, and Sensor Connections ............................................................................... 14
4.0 DISPLAY AND OPERATION ................................................................................................. 19
4.1 User Interface......................................................................................................................... 19
4.2 Instrument Keypad ................................................................................................................. 19
4.3 Main Display........................................................................................................................... 20
4.4 Menu System ......................................................................................................................... 21
5.0 PROGRAMMING THE ANALYZER....................................................................................... 23
5.1 General................................................................................................................................... 23
5.2 Changing StartUp Settings..................................................................................................... 23
5.3 Configuring and Ranging the Current Outputs ....................................................................... 23
5.4 Setting a Security Code.......................................................................................................... 25
5.5 Security Access ...................................................................................................................... 26
5.6 Using Hold.............................................................................................................................. 26
5.7 Resetting Factory Default Settings......................................................................................... 27
5.8 Programming Alarm Relays.................................................................................................... 28
6.0 PROGRAMMING TURBIDITY ............................................................................................... 31
6.1 Programming Measurements  Introduction ........................................................................... 31
6.2 Turbidity Measurement Programming .................................................................................... 32
6.3 Choosing Turbidity or Total Suspended solids ....................................................................... 35
6.4 Entering a Turbidity to TSS Conversion Equation.................................................................. 38
7.0 CALIBRATION ....................................................................................................................... 41
7.1 Calibration Introduction .......................................................................................................... 41
7.2 Turbidity Calibration................................................................................................................ 41
8.0 MAINTENANCE .................................................................................................................... 45
8.1 Model 1056............................................................................................................................. 45
8.2 Sensor .................................................................................................................................... 46
8.3 Debubbler and Measuring Chamber ...................................................................................... 47
8.4 List of Replacement Parts ...................................................................................................... 48
9.0 TROUBLESHOOTING........................................................................................................... 49
9.1 Overview ................................................................................................................................ 49
9.2 Troubleshooting Using Fault Codes ....................................................................................... 45
9.3 Troubleshooting Calibration Problems ................................................................................... 47
9.4 Troubleshooting Other Problems............................................................................................ 52
10.0 RETURN OF MATERIAL ...................................................................................................... 55
10.1 General................................................................................................................................... 56
10.2 Warranty Repair ..................................................................................................................... 57
10.3 NONWarranty Repair ............................................................................................................ 57
i
MODEL CLARITY II TABLE OF CONTENTS
LIST OF FIGURES
Number Title Page
21 Panel Mount Dimensions ......................................................................................... 6
22 Pipe and Wall Mount Dimensions............................................................................. 7
23 Debubbler and Flow Chamber ................................................................................. 9
24 Sensor .................................................................................................... 10
25 Sampling for Turbidity............................................................................................... 10
26 CSA Nonincendive Field Wiring Installation for Turbidity, page 1............................ 11
27 CSA Nonincendive Field Wiring Installation for Turbidity, page 2............................ 12
31 24VDC Power Supply............................................................................................... 14
32 Switching AC Power Supply..................................................................................... 14
33 Current Output Wiring .............................................................................................. 15
34 Alarm Relay Wiring................................................................................................... 15
35 Turbidity Signal Board .............................................................................................. 16
36 Power Wiring for Model 1056 85265 VAC Power Supply ....................................... 17
37 Power Wiring for Model 1056 24VDC Power Supply ............................................... 17
38 Output Wiring for Model 1056 Main PCB ................................................................. 18
41 Formatting the Main Display..................................................................................... 22
52 Configuring and Ranging the Current Outputs ........................................................ 24
53 Setting a Security Code............................................................................................ 25
54 Using Hold................................................................................................................ 26
55 Resetting Factory Default Settings........................................................................... 27
61 Converting Turbidity to TSS ..................................................................................... 33
64 Turbidity Sensor — General..................................................................................... 35
65 Turbidity Sensor — EPA 108.1................................................................................. 35
66 Turbidity Sensor — ISO 7027................................................................................... 36
62 Converting Turbidity to TSS ..................................................................................... 38
62 Lowest Turbidity (TSS) ............................................................................................. 38
67 Configure Turbidity Measurement ............................................................................ 40
78 Calibrate Turbidity .................................................................................................... 44
83 Replacing the Lamp/LED Board............................................................................... 46
84 Molded Buffer Assembly........................................................................................... 48
LIST OF TABLES
Number Title Page
21 Approximate Debubbler Pressure as a Function of Flow ........................................ 8
611 Turbidity Measurement Programming ...................................................................... 32
712 Turbidity Calibration Routines ..................................................................................... 41
81 Replacement Parts for Model 1506.......................................................................... 45
APPENDIX
APPENDIX .............................................................................................................. 56
ii
1
MODEL CLARITY II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION 1.0.
DESCRIPTION AND SPECIFICATIONS
• COMPLETE SYSTEM includes single or dual input analyzer, sensor(s), and debubbler
assembly
• CHOOSE U.S. EPA METHOD 180.1 or ISO METHOD 7027 compliant sensors
• RESOLUTION 0.001 NTU
• FULL FEATURED ANALYZER with fully scalable analog outputs and fully programmable
alarms with interval timers
• INTUITIVE, USERFRIENDLY MENU in seven languages makes setup and calibration
easy
Clarity II is a trademark of Emerson Process Management.
1.1 FEATURES AND APPLICATIONS
The Clarity II turbidimeter is intended for the determi nation of turbidity in water. Low stray light, high stability, efficient bubble rejection, and a display resolution of
0.001 NTU make Clarity II ideal for monitoring the turbidity of filtered drinking water. The Clarity II turbidimeter can be used in applications other than drinking water treatment. Examples are monitoring wastewater discharges, condensate returns, and clarifiers.
Both USEPA 180.1 and ISO 7027compliant sensors are available. USEPA 180.1 sensors use a visible light source. ISO 7027 sensors use a near infrared LED. For regulatory monitoring in the United States, USEPA
180.1 sensors must be used. Regulatory agencies in other countries may have different requirements.
The Clarity II turbidimeter consists of an analyzer, which accepts either one or two sensors, the sensors themselves, and a debubbler/measuring chamber and cable for each sensor. The cable plugs into the sensor and the analyzer, making setup fast and easy. Sensors can be located as far as 50 ft (15.2 m) away from the analyzer.
The Clarity II turbidimeter incorporates the popular and easy to use Model 1056 analyzer. Menu flows and prompts are so intuitive that a manual is practically not needed. Analog outputs are fully scalable. Alarms are fully programmable for high/low logic and dead band. To simplify programming, the analyzer automati cally detects whether an EPA 180.1 or ISO 7027 sensor is being used.
Clarity II is available in an optional configuration in which the analyzer, sensor(s), and debubbling flow cell(s) are mounted on a single back plate. The sensor cables are prewired to the analyzer, so setup is exceptionally fast and easy. All the user does is mount the unit on a wall, bring in power and sample, and pro vide a drain. To order this option, consult the factory.
2
MODEL CLARITY II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SPECIFICATIONS  General
Enclosure: Polycarbonate. Type 4X/CSA 4 (IP65). Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10
x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x
5.45 in.)
Conduit Openings: Accepts 1/2” or PG13.5 conduit
fittings
Display: Monochromatic graphic liquid crystal display.
128 x 96 pixel display resolution. Backlit. Active display area: 58 x 78mm (2.3 x 3.0 in.).
Ambient Temperature and Humidity: 0 to 55°C
(32 to 131°F). Turbidity only: 0 to 50°C (32 to 12 2°F), RH 5 to 95% (noncondensing)
Storage Temperature Effect: 20 to 60ºC (4 to 140°F)
Power: Code 02: 20 to 30 VDC. 15 W.
Code 03: 85 to 265 VAC, 47.5 to 65.0 Hz, switching.
15 W.
Note: Code 02 and 03 power supplies include 4 pro
grammable relays
Equipment protected by double insulation
Alarms relays*: Four alarm relays for process meas urement(s) or temperature. Any relay can be config ured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed with interval timer settings.
Relays: Form C, SPDT, epoxy sealed
Inductive load: 1/8 HP motor (max.), 40 VAC
*Relays only available with 02 power supply (20  30 VDC) or 03 switching power supply (85  265 VAC)
Inputs: One or two isolated sensor inputs Outputs: Two 420 mA or 020 mA isolated current out
puts. Fully scalable. Max Load: 550 Ohm. Output 1 has superimposed HART signal (configurations 10560X2X3XHT only)
Current Output Accuracy: ±0.05 mA @ 25 ºC Terminal Connections Rating: Power connector
(3leads): 2412 AWG wire size. Signal board ter minal blocks: 2616 AWG wire size. Current output connectors (2leads): 2416 AWG wire size. Alarm relay terminal blocks: 2412 AWG wire size (02 24 VDC power supply and 03 85265VAC power supply)
RFI/EMI: EN61326 LVD: EN610101
Hazardous Location Approvals 
Options for CSA: 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, AN, and HT.
Class I, Division 2, Groups A, B, C, & D Class Il, Division 2, Groups E, F, & G
Class Ill T4A Tamb= 50
°C
Evaluated to the ANSI/UL Standards. The ‘C’ and ‘US’ indi cators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively
CAUTION
RISK OF ELECTRICAL SHOCK
Maximum Relay Current
Resistive
28 VDC 5.0 A 115 VAC 5.0 A 230 VAC 5.0 A
POLLUTION DEGREE 2: Normally only nonconductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
WARNING
Exposure to some chemicals may degrade the sealing properties used in the following devices: Zettler Relays (K1K4) PN AZ81CH12DSEA
WARNING
3
SPECIFICATIONS — SENSOR
Method: EPA 180.1 or ISO 7027 (using 860 nm LED
source). Must be specified when ordering.
Incandescent lamp life: two years LED life: five years Wetted materials: Delrin
1
, glass, EPDM
Accuracy after calibration at 20.0 NTU:
0  1 NTU: ±2% of reading or ±0.015 NTU, whichever is greater.
0  20 NTU: ±2% of reading Note: Turbidity values of 2200 NTU can be meas
ured, but frequent cleaning may be required to main tain turbidity measurements.
Cable: 20 ft (6.1 m) or 50 ft (15.2 m). Maximum 50 ft
(15.2 m). Connector is IP65.
Maximum Pressure: 30 psig (308 kPa abs) Temperature: 40  95°F (5  35°C) Sensor body rating: IP65 when cable is connected
SPECIFICATIONS — DEBUBBLER AND FLOW CHAMBER
Dimensions: 18.1 in. x 4.1 in. diam. (460 mm x 104 mm
diam.) (approx.)
Wetted materials: ABS, EPDM, Delrin
1
, polypropylene,
nylon
Inlet: compression fitting accepts 1/4 in. OD tubing; fit
ting can be removed to provide 1/4 in. FNPT
Drain: barbed fitting accepts 3/8 in. ID tubing; fitting can
be removed to provide 1/4 in. FNPT. Must drain to atmosphere.
Sample temperature: 40  95°F (5  35°C) Minimum inlet pressure : 3.5 psig (125 kPa abs). 3.5
psig will provide about 250 mL/min sample flow.
Maximum inlet pressure: 30 psig (308 kPa abs). Do
not block drain tube.
Recommended sample flow: 250  750 mL/min Response Time: The table shows the time in minutes to
percent of final value following a step change in turbid ity.
1
Delrin is a registered trademark of DuPont Performance
Elastomers.
SPECIFICATIONS — MISCELLANEOUS
Weight/shipping weight:
Sensor: 1 lb/2 lb (0.5 kg/1.0 kg)
Analyzer: 2 lb/3 lb (1.0 kg/1.5 kg)
Debubbler: 3 lb/4 lb (1.5 kg/2.0 kg)
(rounded to the nearest lb or 0.5 kg
MODEL CLARITY II SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
4
This page left blank intentionally.
SECTION 2.0.
INSTALLATION
MODEL CLARITY II SECTION 2.0
INSTALLATION
5
2.1 UNPACKING AND INSPECTION
2.2 INSTALLATION
Type of Mounting Figure
Panel 21
Wall and Pipe 22
2.1 UNPACKING AND INSPECTION
The Clarity II Turbidimeter is a complete system for the determination of turbidity in drinking water. The system consists of the analyzer, sensor(s), cable(s), and flow chamber/debubbler(s). Consult the table to verify that you have received the parts for the option you ordered.
(1)
The analyzer model number is printed on a label attached to the side of the instrument.
2.2 INSTALLATION
2.2.1 General Information
1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. Refer to the table below.
Item
Model/part number
Single Input Turbidity Analyzer 1056032738AN
Dual Input Turbidity Analyzer 1056032737AN
Single Input Turbidity Analyzer with HART 1056032738HT
Dual Input Turbidity Analyzer with HART 1056032737HT
SensorEPA standard 801080002EPA
SensorISO standard 801080003ISO
Cable3 ft (0.9 m) 2413800
Cable20 ft (6.1 m) 2409700
Cable50 ft (15.2 m) 2409800
Calibration cup 2410100
Molded chamber/debubbler 2417000
WARNING
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA70) and/or any other applicable national or local codes.
WARNING
RISK OF ELECTRICAL SHOCK
6
Bottom View
Front View
Side View
FIGURE 21 PANEL MOUNTING DIMENSIONS
Note: Panel mounting seal integrity (4/4X) for outdoor applications is the responsibility of the end user.
MILLIMETER
INCH
154.9
6.1
154.9
6.1
126.4
5.0
101.6
4.00
17.13
1.1
126.4
5.0
)
(
76.2
3.0
41.4
1.6
152.73
6.0
7
FIGURE 22 PIPE AND WALL MOUNTING DIMENSIONS
(Mounting bracket PN:2382000)
The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
Bottom View
Front View
Side View
Side View
Wall / Surface Mount
Pipe Mount
MILLIMETER
INCH
154.9
6.1
102
4.0
187
7.4
154.9
6.1
232
9.1
33.5
1.3
130
5.1
165
6.5
232
9.1
130
5.1
33.5
1.3
165
6.5
108.9
4.3
45.21
1.8
80.01
3.2
71.37
2.8
8
2.3 INSTALLATION — DEBUBBLER ASSEMBLY
See Figure 23 for installation.
Connect the sample line to the inlet fitting. The fitting accepts 1/4inch OD tubing. See Section 2.6 for recom mended installation of the sample port.
Attach a piece of 3/8 inch ID soft tubing to the drain fitting. The debubbler must drain to atmosphere.
NOTE
During operation, the debubbler is under pressure. A 0.040 inch (1 mm) orifice in the outlet pro vides the pressure. Back pressure helps prevent outgassing, which can lead to bubbles accu mulating on the sensor face resulting in erroneous readings. DO NOT EXCEED 30 psig (308 kPa abs) inlet pressure.
The amount of pressure in the debubbler can be estimated from the flow rate. See Table 21.
To control and monitor sample flow, a valved rotameter with fittings is available (PN 2410300). Attach the rotameter to the debubbler outlet. The rotameter can also be used to increase back pressure on the debubbler if additional pressure is needed to prevent outgassing.
TABLE 21. Approximate debubbler pressure
as a function of flow (0.040 inch outlet orifice)
gph psig
21
43
68
814
10 21
11 26
12 31
——
mL/min kPa abs
100 110
200 120
300 140
400 160
500 190
600 240
700 280
800 340
MODEL CLARITY II SECTION 2.0
INSTALLATION
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
WARNING
9
FIGURE 23. Debubbler and Flow Chamber
INCH
MILLIMETER
MODEL CLARITY II SECTION 2.0
INSTALLATION
10
2.4 INSTALLATION — SENSOR
Unscrew the nut on the side of the debubbler. Insert the sensor in the mouth of the measuring chamber. Be sure the pin on the debubbler lines up with the hole in the sensor. Replace the nut. Remove the protective cap from the sensor and screw the cable onto the receptacle. The plug and receptacle are keyed for proper alignment.
The sensor is rated to IP65 when properly connected to the cable. To prevent possible water damage to the connector contacts, be sure the cable receptacle and the connector on the back of the sensor are dry when connecting or disconnecting the cable.
2.5 SAMPLE POINT
Locate the sample tap to minimize pickup of sediment or air. See Figure 25. If possible, install a sampling port that extends one or two inches (25  50 mm) into the pipe. Use ¼ inch OD rigid plastic tubing. Avoid soft plastic tubing if possible. To reduce sample lag time, install the debubbler and flow chamber as close to the sample tap as possible.
FIGURE 25. Sampling for Turbidity
FIGURE 24. Sensor
INCH
MILLIMETER
DWG. NO. REV.
40T105501 A
MODEL CLARITY II SECTION 2.0
INSTALLATION
11
Figure 26
REV
D
0000
SEE
ECO
BJ
C
REV
REV
REV
REV
REV
REV
CSA
THIS DOCUMENT I S
CERTIFIED BY
TURBIDITY
CLARITY II
SENSOR #2
(OPTIONAL)
TURBIDITY
CLARITY II
SENSOR #1
6
6
5
ANALYTICAL
ROSEMOUNT
W/O AGENCY APPROVAL
REVISIONS NOT PERMITTED
D3
NON INCENDIVE FIELD
Emerson
PROCESS MAN AGEMENT
TITLE
DATE
12/8/04
D
SHEET 1 OF 1
1056-27/37
1400314
NONE
DWG NO
WIRING INSTALLATION (CSA)
SIZE
SCALE: WEI GHT:
C
1/5/05
1/5/05
5
WARNING
!
PROCESS MEDIA ONLY
FOR USE WITH NON-FLAMMABLE
SENSOR CABLE
IS SHIELDED
SENSOR CABLE
IS SHIELDED
AREA
UNCLASSIFIED
B. JOHNSON
J. FLOCK
J. FLOCK
APPR OVALS
THIS FILE CREATED US ING
DRAWN
ANGLE S ± 1/2 °.
DIMENSIONS ARE IN INCHES
NOMINAL SURFACE FINISH: 125
MACHINE FILLET RADII .020 MAX
REMOVE BURRS & SHARP EDGES
SOLID EDGE
ENG APVD
CHECKED
.XX ± .03 .XXX ± .010
FINISH
MATERIAL
1056-27/37
ALARM
WIRING (VAC)
(OPTIONAL)
HAZARDOUS AREA
CLASS II, III, DIV 2, GPS E-G
CLASS I, DIV. 2, GPS A-D, 0° - 50 ° C
D1
3
ANALOG OUTPUT
METAL CONDUIT
D2
POWER SUPPLY
METAL CONDUIT
METAL CONDUIT
AREA
UNCLASSIFIED
5 MAX CABLE LENGTH IS 50 FEET.
4 DURING INSTALLATION, LEAVE MAXIMUM AMOUNT OF JACKET INSULATION POSSIBLE ON N.I. FIELD
3 GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
2. SEAL REQUIRED AT EACH CONDUIT ENTRANCE.
WIRING WITHIN INSTRUMENT ENCLOSURE. AFTER TERMINATION, WRAP N.I. FIELD WIRING WITHIN
ENCLOSURE WITH M YLAR TAPE, TO ENSURE ADEQUATE DOUBLE INSULATION REMAINS.
1. INSTALLATION MUST CONFORM TO THE CEC. NOTES: UNLESS OTHERWISE SPECIFIED
12
Figure 27
Loading...
+ 46 hidden pages