Remove the analyzer from the shipping box. The shipping box should contain:
•Analyzer (1)
•Enclosure plugs (5)
•Dome connectors (3)
•O-rings (6)
•Large lock nuts (6)
•Sealed dessicant pack (1) that will be placed in the installed analyzer
•Opened dessicant pack (1) that protected the instrument during shipping. Discard
after opening the analyzer.
Installation
WARNING!
Only properly trained and qualified personnel should install the RDO instrument described in
this manual. This instrument should be installed for use in nonhazardous locations only.
1.2Optional mounting kit
The mounting kit (PN R0087560) contains the following:
The nuts and screws supplied with the mounting tabs are suitable for use with the
mounting brackets.
Manual1
Installation
1.3Mounting options
1.3.1Analyzer dimensions
•Figure 1-1 calls out the analyzer dimensions with the lid closed.
Analyzer Dimensions with Lid ClosedFigure 1-1:
•Figure 1-2 shows the depth dimensions of the analyzer. When wall or panel
mounting, make sure to allow adequate space for opening the enclosure door or
making electrical connections.
Side View Dimensional DrawingsFigure 1-2:
2 Rosemount RDO
1.3.2Pipe mounting
Complete the following steps to mount the analyzer on a pipe.
1.Remove the four nuts and four screws from the mounting kit.
2.Open the enclosure and drop one nut into each drilled corner of the box (Figure 1-3).
Placing the Nut in the EnclosureFigure 1-3:
Installation
3.Use a screwdriver to push the nut down and set it in place (Figure 1-3).
Orient the two brackets on the back of the enclosure so that the flanges in the
4.
brackets point downward (Figure 1-4).
Manual3
Installation
Attaching the Mounting BracketsFigure 1-4:
5.Attach the mounting brackets to the enclosure using two screws for each bracket.
Make sure that the screws properly thread into the nuts you seated in steps 2 and 3.
6.Place two hose clamps over the mounting pipe and tighten them until they are
almost secure.
7.Place the flanges from the mounting bracket into each of the hose clamps
(Figure 1-3).
8.Tighten the hose clamps until the analyzer is secure on the pipe.
4 Rosemount RDO
Installation
Back and Side View of Pipe-Mounted AnalyzerFigure 1-5:
1.3.3Wall mounting using mounting tabs
The optional mounting kit contains a set of wall mounting hardware that includes four
screws, nuts, and tabs for mounting the analyzer to a wall or panel. Follow the instructions
included with the mounting tab hardware to attach the tabs (Figure 1-6).
Manual5
Installation
Back View of Analyzer with TabsFigure 1-6:
1.3.4Wall mounting with user-supplied screws and hardware
You can supply your own hardware and attach the analyzer to a wall.
Figure 1-3 indicates the placement of the nut. Use a screw that is best suited for your wall
material (Figure 1-7).
6 Rosemount RDO
Installation
Back View of AnalyzerFigure 1-7:
1.4Installing dome connectors and plugs
Complete the following steps to install the dome connector and plugs.
1.
Place the O-ring over the threaded end of the connector or plug.
2.Screw the connector or plug into the opening until it is snug. Then tighten an
additional 1/2 turn.
Do not overtighten.
3.Place the lock nut concave side down on the threads and, using a screwdriver blade
and mallet, tighten the lock nut until there is approximately one thread showing.
4.Thread cable through the dome connectors and secure by tightening the nut to
about 15 in.-lb of torque.
Confirm that cable does not slide or wiggle after the nut has been tightened. Leave
sufficient cable inside the enclosure to make secure connections without placing
stresss on the cable.
Manual7
Installation
Attached plug (left) and attached dome connector (right)Figure 1-8:
8 Rosemount RDO
2Wiring
WARNING!
Only properly trained and qualified personnel should install the RDO instrument described in
this manual. This instrument should be installed for use in nonhazardous locations only.
2.1Electrostatic discharge (ESD)
recommendations
•Before making wiring connections or touching circuit boards or other internal
components, discharge any static electricity from your body by touching a
grounded metal object.
•When making wiring connections, make sure to remain properly grounded by
wearing an ESD wriest strap or similar device.
Wiring
2.2Customer-supplied electrical equipment
You must supply the following:
•18-12 AWG wire for electrical connections using conduit or 100-240 Vac power
supply
•24-12 AWG cable for connecting relays and PLC devices
•A 100-240 Vac or 9-36 Vdc source with over current/disconnect protection for hard-
wired locations using conduit
•Approved suitable wiring for cord-connected plug and socket locations
2.3Ensuring good electrical connections
To ensure that all sensors and power sourced function properly, make sure that:
•Each individual wire is stripped and tinned to 1/4 in.
•Each wire is tightly screwed into the terminal strip.
•Each wire is touching the terminal strip. If the plastic wire jacket is clamped into the
terminal strip, connections will not be made.
•Clip or cap any unused wires.
Manual9
Wiring
2.4AC power and high voltage relay connections
The AC power board is located under the metal shield on the right side of the enclosure.
The terminal strip on the board accommodates the AC power connections as well as two
high voltage relay connections.
WARNING!
ELECTRIC SHOCK
Make sure that power to the instrument is disconnected before making any wiring
connections.
WARNING!
ELECTRIC SHOCK
Do not connect low voltage circuits (< 50 V) to the terminal connectors on the AC board!
See Figure 2-1.
10 Rosemount RDO
Wiring
AC Circuit Board and Terminal ConnectionsFigure 2-1:
Procedure
1.
Remove the front cover of the enclosure.
2.Remove the metal AC power board cover.
3.Thread the power cable and relay wires (if needed) through the dome fittings
nearest the power board.
4.Using the green screw, connect the green earth wire of the power cable to the metal
frame in the bottom of the enclosure.
See Figure 2-2. A green wire connecting the metal frame to the terminal strip should
already be in place.
Manual11
Wiring
Figure 2-2:
Connect the Ground Wire to the Metal Frame Using the Green
Screw
5.Connect the live and neutral power wires to the L and N terminals.
See Figure 2-1.
TerminalDescriptionNorth America colorEuro color
LLiveBlackBrown
NNeutralWhiteBlue
Protective earthGreenGreen and yellow
6.Connect relay wires (if required) to the terminal strip.
See Figure 2-1.
TerminalDescription
COMCommon
NC1 or NC2Normally closed
NO1 or NO2Normally open
7.Replace the metal AC power board cover.
Replace the front cover of the enclosure.
8.
Be sure the dessicant pack (PN R0087630) is present.
12 Rosemount RDO
2.5Analog and digital outputs and low voltage
relay connections
Complete the following steps to connect the analog and digital outputs and low voltage
relay.
WARNING!
ELECTRIC SHOCK
Make sure that the power to the instrument is disconnected before making any wiring
connections.
WARNING!
ELECTRIC SHOCK
The maximum voltage that can be applied across the loop terminals is 36 V.
WARNING!
Wiring
ELECTRIC SHOCK
Do not connect high voltage circuits (> 50 V) to the terminal connections on the I/O board!
Procedure
1.Remove the front cover of the enclosure.
2.Connect the digital output (Modbus/RS-485) to the PLC terminals.
See Figure 2-3. There is no connection to the GND and POWER terminals. However,
the analyzer supplies 24 Vdc to the GND and POWER terminals, which can be used
to power external accessories.
TerminalDescription
GNDSignal ground
485-RS-485 negative
485+RS-485 positive
POWER24 Vdc
Manual13
Wiring
Input/output boardFigure 2-3:
3.Wire the analog current loops to the 4-20 mA terminals.
Each analog output loop must be powered separately. The maximum voltage at the
terminal must not exceed 36 Vdc. The minimum voltage is 9 Vdc.
TerminalDescription
LOOP2-To negative end of device
LOOP2+To positive end of device
LOOP1-To negative end of device
14 Rosemount RDO
TerminalDescription
LOOP1+To positive end of device
GREEN EARTHCable shield
4.Connect the shield to the green earth ground screw.
See Figure 2-4. Do not ground the device at both ends of the cable.
Earth Ground Screws for Analog Outputs and SensorsFigure 2-4:
Wiring
5.Make the low voltage relay connections.
Important
Do not apply more than 50 V across the low voltage relays.
TerminalDescription
GNDSignal ground
COM1 or COM2Common
NC1 or NC2Normally closed
NO1 or NO2Normally open
6.Replace the front cover of the enclosure.
Be sure the dessicant pack (PN R0087630) is present.
Manual15
Wiring
2.6RDO probe wiring
Complete the following steps to wire the RDO probe.
WARNING!
ELECTRIC SHOCK
Make sure that power to the instrument is disconnected before making any wiring
connections.
Procedure
1.Remove the front cover of the enclosure.
2.Wire the RDO probe to the PROBE A or PROBE B terminals on the input/output card.
TerminalDescriptionWire color
GNDSignal groundBlack
485A-RS485 negativeGreen
485+RS485 positiveBlue
POWERInput DC/Output ACRed
GREEN EARTHShieldShield
3.Connect shield wire to the green earth ground screw.
See Figure 2-4.
4.Cut off or cap the unused brown and white wires.
5.Replace the front cover of the enclosure.
Be sure that the dessicant pack (PN R0087630) is present.
2.7Tighten dome connectors
The dome connectors installed in Section 1.4 must be tightened securely after the sensor
and electrical cables have been routed through them. The dome connectors are not meant
to be weight bearing. Be sure that:
•The cable does not slide or wiggle after the nut has been tightened.
•Sufficient cable has been threaded through the connector so that secure wiring
connections remain secure.
•There is no weight or force from water pulling on the cable.
16 Rosemount RDO
Placing the RDO probe in service
3Placing the RDO probe in service
3.1Unpacking
Remove the probe and sensing cap from the box. The sensing cap is shipped in a small
plastic cylinder inserted in a hole in the internal cardboard packaging.
3.2Assembling the probe
Complete the following steps to assemble the RDO probe.
1.Unscrew the nose cone from the probe and remove the red protective dust cap.
Save the dust cap for later use. Make sure the O-ring grooves are dry and the O-rings
are not rolled or pinched.
2.Remove the RDO sensing cap from its shipping/storage sleeve.
The expected operating lifetime of the cap is two years after the first reading has
been taken. Install the cap by the date printed on the packaging (for PN
R00804230X only).
Note
Keep the cap in its sealed packaging until you are ready to install it. Install promptly. Avoid
allowing moisture, including humidity, inside the cap.
3.Align the arrow on the cap with the index mark on the probe and firmly press the cap
onto the probe, without twisting, until it seals.
Manual17
Placing the RDO probe in service
4.Reattach the nose cone.
Note
The nose cone must be in place whenever the probe is submerged in water.
5.Wire the probe to the analyzer.
See Section 2.6.
Note
If you are replacing an existing RDO probe with a new probe, the analyzer may not recognize
the new probe. If this is the case, the main display will show the probe not talking icon (letter A
or B in reverse video). To establish communication, follow the procedure in Section 5.3.
3.3Installing the probe
The RDO dissolved oxygen probe can be submerged in basins or ponds.
The nose cone and the thermistor (the small metal disc on the side of the sensor about
1/6 in. [3 mm] above the nose cone) must be completely submerged. Use the 1 1/4 in.
FNPT fitting at the rear of the probe to attach it to a pipe. Be sure the connection is
watertight and the upper end of the pipe is closed to keep out water.
CAUTION!
EQUIPMENT DAMAGE
Do not allow the back end of the sensor to get wet.
18 Rosemount RDO
4Display and Operation
4.1Display and keypad
Figure 4-1 shows the keypad and display window.
Keypad and displayFigure 4-1:
Display and Operation
The Up/Left arrow key has three functions: it moves the cursor to the left; it moves the
cursor up a list; it increases the value of a digit. The Down/Right arrow key has three
functions: it moves the cursor to the right; it moves the cursor down a list; it reduces the
value of a digit. The left and right function keys perform the operation shown in the screen
immediately above the key.
The main display has two lines. The measurement from probe A appears in the top line,
and the measurement from probe B appears in the bottom line. If no probe B is connected,
a row labeled C appears. References to probe C refer to the analyzer, which is called the
ConTROLL Pro. See Figure 4-2.
Manual19
Display and Operation
Two-line displayFigure 4-2:
Lines labeled A and B refer to probes A and B. C refers to measurements made by the analyzer. See also
Section 4.2.
Display Icons and Their DescriptionsTable 4-1:
IconDescription
Error with parameter
Calibration stabilizing
Calibration stable
View
View locked
View unlocked
Factory calibration for RDO cap has expired.
User calibration has expired, based on the Calibrate Interval.
Probe A not talking
Probe B not talking
Relay 1 active
Relay 2 active
20 Rosemount RDO
Display and Operation
4.2Using the view function and customizing the
display
The view function allows you to view parameters measured by each probe and the
analyzer. It also allows you to customize the display.
1.Select the View option.
An arrow points to the first line in the display.
2.Select View again to move the arrow to the second line in the display.
3.With the arrow pointing at any line, use the Up and Down arrow keys to cycle
through the parameters being measured by the probe and analyzer.
4.To customize the display, move the arrow to the desired line and press the Up or
Down arrow key until the desired measurement appears. Then select View to move
to the next line.
Manual21
Display and Operation
22 Rosemount RDO
5Programming the analyzer
5.1Initial configuration
Enter the following information upon first use or any time after you have restored the
default settings.
1.Choose English, Spanish, or French. Use the arrow keys to move up or down to the
appropriate language. Select Enter.
2.Use the arrow keys to adjust the contrast level. Select Enter.
3.Enter the time zone.
a. Refer to Table 5-1. Choose the appropriate time zone and find the corresponding
UTC value (Universal Time, Coordinated, formerly GMT).
Selected Regions and Their Time in UTCTable 5-1:
RegionUTC value (± numeric value)
Australia Northern Territory+9.5
Australia Lord Howe Island+10.5 (Daylight Savings Time +11)
Australia New South Wales+10 (Daylight Savings Time +11)
Australia Queensland+10
Australia Victoria+10 (Daylight Savings Time +11)
Australia Australian Cap. Terr.+10 (Daylight Savings Time +11)
Australia South+9.5 (Daylight Savings Time +10.5)
Australia Tasmania+10 (Daylight Savings Time +11)
Australia Western+8
Canada Central-6 (Daylight Savings Time -5)
Canada Eastern-5 (Daylight Savings Time -4)
Canada Mountain-7 (Daylight Savings Time -6)
Canada Yukon & Pacific-8 (Daylight Savings Time -7)
Canada Atlantic-4 (Daylight Savings Time -3)
Canada Newfoundland-3.5 (Daylight Savings Time -2.5)
England0 hours (Daylight Savings Time (+1)
USA Puerto Rico-4
USA Central-6 (Daylight Savings Time -5)
USA Eastern-5 (Daylight Savings Time -4)
USA Mountain-7 (Daylight Savings Time -6)
USA Arizona-7
USA Indiana East-5
Programming the analyzer
Manual23
Programming the analyzer
4.Enter the date using the Up/Left arrow key. Select Enter.
5.Enter the time using the Up/Left arrow key. Select Enter.
Selected Regions and Their Time in UTC (continued)Table 5-1:
RegionUTC value (± numeric value)
USA Pacific-8 (Daylight Savings Time -7)
USA Alaska-9 (Daylight Savings Time -8)
USA Aleutian-10
USA Hawaii-10
b. Use the Up/Left arrow key to enter a positive or negative value.
c. Press the Down/Right arrow key to move to the appropriate numeric position.
d. Use the Up/Left arrow key to enter the correct numeric value. Repeat for
additional digits as needed.
e. Select Enter.
The Add Probe A screen appears. The name of the probe wired in the position also
appears.
6.Select Enter to configure this probe.
The Add Probe B screen appears. The name of the probe wired in this position also
appears. If there is no probe B, the display shows Probe not connected.
7.Select Enter to configure this probe if it exists or to proceed to the next screen if it
does not.
The main display appears.
For additional information on setting up the probes, see Section 5.3.3 and Section 5.3.4.
24 Rosemount RDO
5.2Menu tree
The menu tree for the analyzer is shown in Figure 5-1.
Menu TreeFigure 5-1:
CalibrateHold outputsSettingsDiagnostics
Saturation
One point
Two point
Concentration
Default
Programming the analyzer
Menu
Probes
(A or B)
Options
Barometer
Salinity
Cal Interval
Parameters
Concentration*
Temperature*
Saturation*
Pressure*
Replace probe
Delete probe
Probe C
(analyzer)
Barometer*
Temperature*
Outputs
Current loops
Enable/disable
Parameter
4-20 mA values
Trim
Relays
Enable/disable
High/low
Parameter
Setpoint
Deadband
Test
Date and
time
Note
*For each parameter:
•Enable or disable
•Choose units
•Choose resolution
•Set sentinel
5.3Configuring the RDO probes
Display
Lock/unlock
Contrast
Language
Power save
Polling interval
Communications
Device address
RS485 parameters
Security code
Default
Configure the RDO probes one at a time.
There are four submenus available for each probe: Options, Parameters, Replace Probe,
and Delete Probe.
Manual25
Programming the analyzer
5.3.1Options (Barometer, Salinity, and Calibration Interval)
Three settings can be made under Options.
•Barometer: The analyzer has a built-in pressure sensor to measure barometric
pressure, which is needed for percent saturation (air) calibration. Emerson strongly
recommends the use of automatic barometric pressure.
•Salinity: Salinity is an important setting if dissolved oxygen is being measured in
brackish water and percent saturation (air) calibration is being used to calibrate the
probe.
•Calibration Interval: The analyzer can be programmed to remind you when a calibration
is due.
1.
Select Menu > Settings > Probes. Select RDO Probe A or RDO Probe B, if it is present,
and then select Options.
2.Scroll to Barometer and then select it. The choices are Fixed and Automatic. If you
choose Fixed, enter the barometric pressure here before you start percent saturation
(air) calibration. If you choose Automatic, the analyzer automatically measures the
barometric pressure and uses it during percent saturation (air) calibration.
3.Scroll to Salinity and select it. Choose Fixed. Automatic is not available. Enter the salinity
of the process water in PSU (practical salinity units in parts per thousand).
4.Scroll to Calibration Interval. Select a calibration interval. The analyzer will display a
message when the next calibration is due.
5.3.2Parameters (Concentration, Temperature, Saturation,
and Partial Pressure
Each RDO probe can measure concentration, temperature, percent saturation, and partial
pressure. There are four settings to make for each parameter.
•Enable or disable the measurement. If the measurement is enabled, the probe
makes the measurement, and the analyzer displays the result. If the measurement is
disabled, the probe does not make the measurement.
•Select units in which measurement results will be displayed.
•Set the display resolution.
•Set a sentinel. The sentinel is the value that will be displayed when the probe returns
an error condition for the measurement.
Procedure
1.
Select Menu > Settings > Probes.
2.Select RDO Probe A or RDO Probe B, if it is present, and then select Parameters.
3.Select Concentration.
a. Press the Select key to toggle between Enable and Disable.
If the box is checked, Concentration has been selected.
b. Scroll to Units and select it. Choose between mg/L and ug/L.
26 Rosemount RDO
Programming the analyzer
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
4.Select Temperature.
a. Press Select to toggle between Enable and Disable.
If the box is checked, Temperature has been selected.
b. Scroll to Units and select it. Choose between °C and °F.
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
5.Select Saturation.
a. Press Select to toggle between Enable and Disable.
If the box is checked, Percent Saturation has been selected.
b. Scroll to Units.
There is no selection to make. the units are % saturation.
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
6.Select Pressure.
a. Press Select to toggle between Enable and Disable.
If the box is checked, Partial Pressure has been selected.
b. Scroll to Units.
There is no selection to make. The units are torr.
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
5.3.3Replacing a probe
If you replace the RDO probe with another RDO probe, use Replace Probe to transfer the
existing setting to the new probe.
Manual27
Programming the analyzer
1.Disconnect the existing RDO probe and replace it with the new probe.
See Section 3.3.
2.Select Menu > Settings > Probes. Select RDO Probe A or RDO Probe B, if it is present,
and then select Replace Probe.
The analyzer automatically configures the new probe.
3.Press and hold Back to return to the main display.
5.3.4Deleting a probe
If you plan to permanently remove a probe, follow the steps below.
1.
Select Menu > Settings > Probes.
2.Highlight the probe you wish to delete and press Select.
3.Scroll to Delete Probe.
4.Press Select and then Enter to confirm.
The probe is deleted from the configuration.
5.3.5Adding a probe
To add a probe, follow the steps below.
1.
Select Menu > Settings > Probes.
2.Select Probe A or Probe B and then Add Probe.
The Add Probe screen appears with the name of the sensor that is wired in this
position. If there is no Probe B, the display says No probe connected.
3.Select Enter to add the probe.
5.4Configuring the analyzer
The analyzer (called the ConTROLL PRO) measures two parameters, barometric pressure
and temperature. There are four settings to make for each parameter.
•Enable or disable the measurement. If the measurement is enabled, the analyzer
makes the measurement and displays the result. If the measurement is disabled, the
analyzer does not make the measurement.
•Select units in which measurement results will be displayed.
•Set the display resolution.
•Set a sentinel. The sentinel is the value that will be displayed when the analyzer
returns an error condition for the parameter.
Procedure
1.
Select Menu > Settings > Probes. Select ConTROLL PRO.
28 Rosemount RDO
Programming the analyzer
2.Select Barometer.
a. Press Select to toggle between Enable and Disable.
If the box is checked, barometric pressure has been selected.
b. Scroll to Units and select it. Choose the desired units for barometric pressure.
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
3.Select Temperature.
a. Press Select to toggle between Enable and Disable.
If the box is checked, temperature has been selected.
b. Scroll to Units and select it. Choose the desired units for temperature.
c. Scroll to Resolution. Press Select repeatedly to change the resolution.
The setting is stored as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the arrow keys to change the setting.
5.5Configuring outputs
Outputs refer to the analog current loops and the alarm relays.
Important
Configure the probes before configuring the outputs.
5.5.1Current loops
There are six settings to make:
•Enable/Disable: Choose to enable or disable the output loop.
•Parameter: Assign a probe and parameter to the output.
•4 mA Value: Enter the value to correspond with 4 mA.
•20 mA Value: Enter the value to correspond with 20 mA.
•4 mA Trim: Adjust the 4 mA value of the analyzer to achieve a loop current of 4 mA.
•20 mA Trim: Adjust the 20 mA value of the analyzer to achieve a loop current of
20 mA.
Procedure
1.
Select Menu > Settings > Outputs > Current Loops.
2.Choose current loop 1 or current loop 2.
3.Scroll to Enable/Disable and select it. Press Select to toggle between Enable and Disable.
Manual29
Programming the analyzer
If the box is checked, the current loop has been enabled.
4.Scroll to Parameter and select it. Scroll through the list of probes and measurements.
A is probe A, B is probe B, and C is the analyzer.
5.Scroll to 4 mA value and select it. Use the arrow keys to change the setting.
6.Scroll to 20 mA value and select it. Use the arrow keys to change the setting.
7.Scroll to 4 mA Trim. Use the arrow keys to adjust the loop current to 4 mA.
8.Scroll to 20 mA Trim. Use the arrow keys to adjust the loop current to 20 mA.
5.5.2Alarm relays
There are six settings to make.
•Enable/Disable: Choose to enable or disable the relay.
•Mode: Choose High Alarm or Low Alarm.
•Parameter: Assign a probe and a parameter to the relay.
•Setpoint: See Figure 5-2.
•Deadband: See Figure 5-2.
•Test: Manually energize or de-energize a relay to test it.
30 Rosemount RDO
Programming the analyzer
High and Low Alarm Relay ActionFigure 5-2:
Procedure
Select Menu > Settings > Outputs > Relays.
1.
2.Choose Relay 1 or Relay 2.
For each relay, there is a high voltage and low voltage relay. Configuring the relay
configures both the high voltage and low voltage relay.
3.Scroll to Enable/Disable and select it. Press Select to toggle between Enable and
Disable.
If the box is checked, the relay has been enabled.
4.Scroll to Mode and select it. Choose High alarm or Low alarm.
5.Scroll to Parameter and select it. Scroll through the list of probes and measurements.
A is probe A, B is probe B, and C is the analyzer.
6.Scroll to Setpoint and select it. Use the arrow keys to change the setting.
7.Scroll to Deadband and select it. Use the arrow keys to change the setting.
Manual31
Programming the analyzer
8.Scroll to Test and select it. Use the arrow keys to change from yes to no.
5.6Changing the date and time
Complete the following steps to change the date and time on the RDO analyzer.
1.
Select Menu > Settings > Date & Time.
Three menu items appear:
• Date shown as yyyy-mm-dd
• Time shown as hh:mm:ss (24 hour clock)
• UTC shown as ±xx:xx (see Table 5-1)
2.Scroll to the menu item and select it.
3.Use the arrow keys to change the setting.
4.Select Enter.
5.7Changing the appearance of the screens
5.7.1Locking the display
Locking the display prevents the operator from scrolling through the measurement
parameters using the View function.
1.
Select Menu > Settings > Display.
2.Choose Lock View.
3.Use Select to toggle between the lock view and the unlocked view.
The lock symbol at the top of the screen shows whether the screen is locked or
unlocked.
5.7.2Changing the display contrast
Complete the following steps to brighten or darken the screen.
1.
Select Menu > Settings > Contrast.
2.Use the arrow keys to brighten or darken the screen.
3.Press Enter.
5.7.3Changing the language
Complete the following steps to change the language.
1.Select Menu > Settings > Language.
2.Choose English, Spanish, or French.
32 Rosemount RDO
3.Press Enter.
5.8Changing power settings
To conserve power the analyzer can be set to turn off the display after idling for a certain
length of time.
1.Select Menu > Settings > Power Save.
2.Select External.
Do not choose Battery.
3.Select the appropriate interval - Off (always on), 15 sec, 30 sec, 1 min, 5 min, or 15 min.
5.9Setting the polling (update) interval
The polling interval specifies how often the analyzer checks the probe and updates the
configured outputs when the analyzer is in power saving mode.
See Section 5.8. The default polling rate is every 15 minutes. The fastest polling rate is
every minute; the slowest is every 24 hours.
Programming the analyzer
Important
If the analyzer is not in power saving mode, the display, relays, and outputs update once every 5 sec.
Procedure
1.
Select Menu > Settings > Polling Interval.
2.Select the desired interval.
3.Press Enter.
5.10RS485 communications
5.10.1Setting the device address for the analyzer
The device address is a number between 1 and 245. The device address of Probe A is the
analyzer address plus 1. The device address of Probe B is the analyzer address plus 2.
1.Select Menu > Settings > Communications > Address.
2.Use the arrow keys to enter a number between 1 and 245.
3.Press Enter.
5.10.2Setting RS485 parameters for the analyzer
Complete the following steps to set the RS485 parameters for the analyzer.
Manual33
Programming the analyzer
1.Select Menu > Settings > Communication > RS485.
There are six settings to make:
• Mode (choose RTU or ASCII)
• Baud Rate (choose 9600, 19200, 38400, or 57600)
• Data Bits (choose 8)
• Parity (choose Even, Odd, or None)
• Stop Bits (choose 1 or 2)
• Defaults (choose yes to restore RS485 default settings)
2.
Scroll to the menu item and select it.
3.Use the arrow keys to change the setting.
4.Press Enter.
5.11Setting security codes
You may set up the analyzer with passwords for an administrator and a user. The user can
access the Calibrate, Hold, Options, and Diagnostics menus. The administrator can access
all functions, most notably the Settings menu. If there is no administrator code set, the
user becomes the default administrator and can access all functions. To remove pass codeprotected access, enter 0000 for all codes.
1.
Select Menu > Settings > Security Code.
2.Select Administrator and enter a 4-digit code for the Administrator using the arrow
keys. Press Enter.
3.Select User. Set a 4-digit code for the User using the arrow keys. Press Enter.
5.12Defaults
Selecting defaults returns all user settings and calibrations to factory default values.
The display also returns to the initial configuration screens described in Section 5.1.
5.13Placing the output and relays on hold
Occasionally, you may wish to temporarily pause the outputs.
1.
Select Menu, then Hold Outputs.
All measurement updates stop, the current values and outputs are held, and the
phrase Hold Outputs appears at the top of the main display.
2.To take the analyzer out of hold, select Resume.
34 Rosemount RDO
5.14Diagnostics
The following diagnostic information is available for the RDO probe:
•Serial number
•Firmware version number
•Date of last user calibration
•Last calibration slope and offset
•Date next calibration is due
•Expiration date for sensing cap (one year from the date the sensing cap took its first
measurement, for PN R0084230 only)
•For replacement cap, PN R0084230X, there is no expiration date (expected life is
two years)
•Sensing cap serial number
The following diagnostic information is available for analyzer Probe C, also called
ConTROLL PRO:
•Serial number
•Firmware version number
•Hardware version number
•Power source
Programming the analyzer
Procedure
1.Select Menu, then Diagnostics.
2.Select the desired probe (A or B). For analyzer diagnostics, select Probe C.
3.Use the arrow keys to scroll through the diagnostics.
4.To read the last calibration slope and offset, scroll to Last calibration and select it.
The slope should be between 0.80 and 1.20, and the offset should be between -0.2
and +0.2.
Manual35
Programming the analyzer
36 Rosemount RDO
6Calibration
6.1Calibration options
There are three ways to calibrate the RDO probe.
•Saturation: Both a two-point (100% and 0% saturation) and one-point (100%
saturation) calibration are available. 100% saturation refers to water completely
saturated with atmospheric oxygen. 0% refers to water containing no dissolved
oxygen.
The solubility of atmospheric oxygen in water depends on the barometric pressure,
temperature, and humidity. If these are known, the concentration of oxygen in airsaturated water can be readily calculated, making it a useful calibration standard. In
practice, however, air-saturated water is almost never used. Air is used instead. Air
works because, according to Henry's Law, the concentration of oxygen in airsaturated water (expressed in mg/L) is in equilibrium with the partial pressure of
oxygen in the air used to saturate the water. Therefore, whether the probe is in air or
in air-saturated water, the fluorescence quenching will be exactly the same.
Calibration
During calibration at 100% saturation, the micro-processor uses the temperature
measured by the probe and the barometric pressure measured by the analyzer to
calculate the equilibrium solubility of oxygen in water. Because the calibration
assumes the air is completely saturated with water, the calibration must be done
with the probe in air having 100% humidity.
Calibrating at 0% saturation requires water containing no dissolved oxygen. A good
0% saturation standard is water containing about 5% sodium sulfite (Na2SO3). Add
about one teaspoonful of sodium sulfite crystals to a cup of water.
•Concentration: If the probe is installed in a waste-water aeration basin, it is often
inconvenient to remove it for saturation calibration. In this case, concentration
calibration, in which the probe is calibrated against a referee instrument, is more
suitable.
•Default: Default restores the default calibration.
Although a new probe can be used as received from the factory, Emerson recommends
that you do a two-point saturation calibration when the probe is first placed in service.
Manual37
Calibration
6.2One-point calibration (100% saturation)
Complete the following steps to do a one-point calibration.
1.Remove storage cap on top of calibration chamber and replace it with the
calibration cap (cap with the vent hole).
Calibration and storage capsFigure 6-1:
2.Fill the calibration chamber to the lower fill line with approximately 10 mL water.
The water temperature should be as close to the sample temperature as possible.
Fill the calibration chamber to the lower line with water.Figure 6-2:
3.Remove the sensor from the process liquid. If the sensor is fouled or dirty, clean it by
following the procedure in Section 7.2.2. Gently dry the probe and sensor with a soft
cloth, making sure there is no water on the body of the sensor or on the sensing foil.
Important
The sensing foil must remain dry during 100% saturation calibration.
38 Rosemount RDO
Calibration
4.Place the probe in the calibration chamber keeping the sensing foil about 1 in. (25
mm) above the surface of the water.
5.Allow at least five minutes for the temperature to stabilize before starting the
calibration.
Keeping the sensor in the shade will help reduce drift caused by the sun's heat. Do
not leave the calibration chamber for more than 30 minutes, lest condensation form
on the surface of the foil, leading to false low readings after calibration. If
condensation does occur, remove the probe and dry the foil. Return the sensor to
the calibration chamber and continue.
6.Once readings are stable, start the calibration. Select Menu > Calibrate > RDO PRO >
Saturation > One-point cal.
The calibration begins immediately. The Calibration Beginning screen shown in
Figure 6-3 appears.
Figure 6-3:
Note the Calibration Beginning icon next to the mg/L reading. As the reading
stabilizes, the bar in the icon shrinks. When the word Nominal appears, the
stabilization is almost complete, and the calibration can be selected at this point by
selecting Nominal. However, for best results, wait until Stable appears.
7.
Select Stable to accept the calibration. To end and return to the previous
calibration, select Cancel.
The figure shows the screens that appear as the calibration
progresses.
6.3Two-point calibration (100% and 0%
saturation)
Complete the following steps to perform a two-point calibration on your RDO analyzer.
1.
Set up the calibration chamber and probe as described in Section 6.2, steps 1
through 5.
Manual39
Calibration
Important
Do the 100% calibration step first.
2.Once readings are stable, start the calibration. Select Menu > Calibrate > RDO PRO >
Saturation > Two-point cal.
The calibration begins immediately. The Calibration Beginning screen shown in
Figure 6-3 appears, and the other two screens appear as the reading stabilizes. Wait
until the word Stable appears.
3.Select Stable to accept the calibraiton and proceed to the 0% saturation step. To
end and return to the previous calibration, select Cancel.
The 0% saturation step starts automatically.
4.Remove the probe and fill the calibration chamber to the upper fill line with
approximately 60 mL of fresh sodium sulfite solution (see Figure 6-4).
Figure 6-4:
Fill the Calibration Chamber to the Upper line with Fresh Sodium
Sulfite Solution.
A teaspoonful of sodium sulfite in a cup of water is adequate.
Place the probe in the sodium sulfite solution. Be sure the thermistor thermowell
5.
(the small metallic disc about 1/8 inch [3 mm] above the nose cone) is completely
submerged. Leave at least 1/2 inch (12 mm) between the surface of the foil and the
bottom of the chamber. Do not allow the sensing foil to rest on the bottom of the
calibration chamber. Allow at least five minutes for readings to stabilize.
The Calibration Beginning screen shown in Figure 6-5 appears.
40 Rosemount RDO
Calibration
Figure 6-5:
Note the Calibration Beginning icon next to the mg/L reading. As the reading
stabilizes, the bar in the icon shrinks. When the word Nominal appears, the
stabilization is almost complete. However, for best results, wait until Stable
appears.
6.
Select Stable to accept the calibration. To end and return to the previous
calibration, select Cancel.
The calibration report screen appears. The slope will be between 0.8 and 1.2, and
the offset will be between -0.2 and +0.2.
The Figure Shows the Screens that Appear as the Calibration
Progresses.
7.Press OK.
8.Remove the probe from the calibration chamber. Rinse it and return it to the process
liquid.
6.4Concentration calibration
Complete the following steps to do a concentration calibration of your RDO analyzer.
1.
Calibrate the referree instrument following the manufacturer's instructions.
2.Immerse the referree sensor in the process liquid as close to the RDO probe as
possible.
3.Allow adequate time for the referree sensor to come to equilibrium with the process
liquid.
4.Once readings are stable, start the calibration. Select Menu > Calibrate > RDO >
Concentration.
5.Enter the value from the referree instrument as the Standard Value shown on
the display. Press Enter.
Manual41
Calibration
42 Rosemount RDO
7Maintenance
7.1Analyzer
7.1.1Cleaning the analyzer enclosure
Clean the analyzer case and front panel by wiping with a clean, soft cloth dampened with
water only.
CAUTION!
EQUIPMENT DAMAGE
Do not use organic solvents.
7.1.2Replacing the dessicant
Maintenance
Periodically inspect the dessicant bag in the analyzer enclosure. Replace the dessicant (PN
R0087630) when the indicator beads turn from blue to pink.
7.1.3Replacing the clock battery
The clock battery is located at the bottom of the back of the enclosure cover. To replace
the battery:
1.Disconnect AC power to the instrument.
2.Remove the enclosure cover.
3.Remove the clock battery and replace it with a 3 V MnO2-Li battery, CR2032 or
equivalent.
4.Dispose of spent battery properly.
5.Replace the cover.
6.Restore AC power.
Manual43
Maintenance
7.2RDO probe
7.2.1Cleaning the sensing cap
Complete the following steps to clean the sensing cap.
Important
Leave the sensing cap and cone attached to the probe body. Do not remove the sensing cap to clean
it.
Procedure
1.Rinse the cap with clean water from a squirt bottle.
2.If biofouling is present, gently wipe the cap with a soft-bristeled brush or soft cloth.
If oil or grease is present, wash with a gentle detergent.
3.If extensive fouling or mineral build-up is present, soak the cap of the sensor - do not
remove the sensing cap - in vinegar for 15 minutes; then soak in deionized water for
15 minutes.
CAUTION!
EQUIPMENT DAMAGE
Do not use organic solvents.
4.After cleaning the sensor, check the calibration in the air. Perform a one or two-
point saturation calibration if necessary.
7.2.2Cleaning the probe body
Complete the following steps to clean the RDO probe body.
Leave the sensing cap and nose cone attached to the probe body while cleaning it. Do not
remove the sensing cap.
Procedure
1.Gently scrub the probe body with a soft-bristled brush of a nylon dish scrubber. Use
a mild detergent to remove oil or grease.
2.Soak in vinegar followed by deionized water to remove mineral deposits or extensive
fouling.
See step 4 in Section 7.2.1.
7.2.3Cleaning the optical window
Clean the optical window only when the sensing cap is replaced. See Section 7.2.1.
44 Rosemount RDO
7.2.4Replacing the sensing cap
The replacement sensing cap kit (PN R0084230X and PN R084230) contains a sensing cap,
two O-rings, O-ring lubricant, and a lens wipe.
Note
Keep the cap in its sealed packaging until you are ready to install it. Install promptly. Avoid allowing
moisture, including humidity, inside the cap.
Procedure
1.Remove the sensor from the process liquid. Rinse with water and dry the probe body
and nose cone.
2.Unscrew the nose cone.
3.Pull the sensing cap straight off the probe body.
Do not twist.
4.Remove and discard the existing O-rings.
5.Remove any moisture in the O-ring grooves.
Maintenance
Be careful not to touch the lens.
6.Use your finger to apply a thin layer of lubricant around the O-ring grooves.
CAUTION!
EQUIPMENT DAMAGE
Be careful not to get grease on the lens or on the sensor pins.
7.Slide the two O-rings into grooves. Check that the O-rings are not twisted or
pinched. Apply a thin layer of lubricant over the O-rings and grooves.
CAUTION!
EQUIPMENT DAMAGE
Be careful not to get grease on the lens or on the sensor pins.
8.Clean the lens with the wipe provided in the kit and allow it to dry thoroughly.
Inspect the lens for scratches or dirt.
9.Remove the new sensing cap from its sealed packaging. Align the arrow on the cap
with the index mark on the probe and firmly press the cap onto the probe, without
twisting, until it seals. Replace the nose cone.
10.Perform a one or two-point saturation calibration.
Manual45
Maintenance
7.2.5Storing the probe
Store the probe in the calibration chamber using the storage cap (cap without notch).
Place a few drops of water in the chamber before inserting the probe.
7.3Replacement parts
Part numberDescription
R0094030RDO analyzer, AC power, no data logging
R0086460XRDO Pro-X sensor with 32 ft (10 m) of integral cable
R0082490XRDO Pro-X sensor with twist lock connector
R0087560RDO analyzer pipe and wall mounting kit
R00CBL10Twist lock connector cable, 32 ft (10 m)
R00CBL20Twist lock connector cable, 64 ft (20 m)
R00CBL30Twist lock connector cable, 96 ft (30 m)
R0084230RDO sensor cap replacement kit
R0080810RDO O-ring replacement kit
R0080820RDO replacement nose cone kit
R0088890RDO replacement calibration cup
R00087630RDO replacement dessicant bag for analyzer
R0084230XRDO Pro-X sensor cap replacement kit
46 Rosemount RDO
8Specifications
8.1RDO probe
Specifications
Wetted materialsDelrin
DimensionsLength: 8.0 in. (203 mm); Diameter: 1.9 in. (47 mm)
RatingIP-67 with cap off; IP-68 with cap installed
Process connection1-1/4 in. FNPT
Integral cable length32 ft (10 m)
Maximum cable length (quick disconnect cable only)4000 ft (1219 m)
PressureUp to 314 psig (2060 kPa abs)
Temperature0 to 50 °C (32 to 122 °F)
Range0 to 20 ppm (mg/L) or 0 to 200% saturation
Accuracy±0.1 ppm between 0 and 8 ppm, ±0.2 ppm between 8 and
Resolution0.01 ppm (mg/L)
Digital outputModbus/RS485
Response time30 sec to 90% of final value; 37 sec to 95% of final value (at
Operating life of sensing cap2 years from first reading
Safety directive73/23/EEC
EU directives2004/108/EC for electro-magnetic compatibility (EMC)
ImmunityEN 61000-6-2, electromagnetic compatibility (EMC) part
(1)
, ABS, Viton
poly(methyl, methacrylate) blend
20 ppm
25 °C [77 °F])
and 72/23/EEC for Safety
(2)
, titanium, polycarbonate/
6-2
EmissionsClass A requirements of CISPR 11: 2004
(1) Delrin is a registered trademark of DuPoint DeNemours, LLC.
(2)
Viton is a registered trademark of DuPoint Dow Elastomers, LLC.
8.2RDO analyzer
EnclosurePolycarbonate, rated NEMA 4X, IP67
Dimensions (W x H x D)6.3 x 6.3 x 3.6 in. (16 x 16 x 9.0 cm)
DisplayLiquid crystal; character height: 0.4 in. (6 mm)
MountingSuitable for pipe or wall mounting
Manual47
Specifications
Conduit openingsSix; PG 13.5 (1/2 in.); three gland fittings and five plugs
ship with the analyzer
Ambient temperature and humidity-20 to 70 °C (-4 to 158 °F); 95% relative humidity (non-con-
densing)
Power100 to 240 Vac, approximately 0.15 A, 50 - 60 Hz
Analog outputsTwo fully scalable 4-20 mA outputs, each loop-powered
(9-36 Vdc)
Digital outputModbus/RS485
RelaysTwo low voltage relays, <50 Vac or Vdc, maximum current
2 A (resistive or inductive) two high voltage relays, 264 Vac
max, maximum current 5 A (resistive or inductive)
Barometric pressure range8.86 to 29.53 in Hg (300 to 1000 mbar)
Barometric presssure accuracy±0.09 in Hg (±3 mbar)
EU directive2004/108/EC for electromagnetic compatibility (EMC)P
ImmunityEN61000-6-2, electromagnetic compatibility (EMC) part
6-2
EmissionsEN61000-6-4 electromagnetic compatibility; includes IEC/
The Emerson logo is a trademark and service mark of Emerson
Electric Co. Rosemount is a mark of one of the Emerson family of
companies. All other marks are the property of their respective
owners.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.