Rosemount Analytical RBIs, RB Specifications

Page 1
RBI pH Sensors Product Data Sheet
LIQ-PDS-RBI-pH-Sensors September 2015
Threaded In-Line, Submerged or Ball-valve Insertion pH Sensor
 Maximum sensor life due to proprietary
multi-junction which hinders reference poisoning.
 Chemically resistant KYNAR
 Rugged, quick change, quick clean.
 Optional installation methods allows
standardization of entire plant with a minimum of sensor part numbers.
 Assembled in a reusable Titanium sheath,
Hi-Pressure body (RB-547).
RBI-546
Submersion/In-line pH Sensor
RBI-547
Ball Valve Insertion pH Sensor
Installation
RBI-546
Threaded in-line: ¾'' MPT threaded nose for installation into
process, sample line. Submerged: ¾'' MNPT threaded top for connection to ¾'' FNPT coupling extension pipe.
RBI-547
Insertion: 1'' MPT process tting or, 1¼'' and 1½'' ball valve assemblies. Insertion depth is user selectable.
Submerged: 1'' MNPT process tting, into 1'' FNPT coupling or ¾'' submersion extension pipe.
Sensor & Transmitter types supported: pH: Compatible with all Rosemount Analytical industrial
transmitters and analyzers.
Wetted Materials:
Kynar, Titanium (RBI-547 only), Porous TEFLON, Wood and Glass, choice of: Viton
Maximum Pressure & Temperature Ratings: Max Temperature: 266 °F/130 °C at 40 psig/276 KPa
Max Pressure: 150 psig/1035 KPa at 158 °F/70 °C
Maximum Insertion/Retraction Pressure:
(option 546) 65 psig/448 KPa @ 158 °F/70 °C (option 547) 40 psig/276 KPa @ 266 °F/130 °C
Measurement Range:
pH: 0–14 pH
®
, EPDM®, or Kalrez®.
Page 2
RBI pH Sensors September 2015
Sensor Dimensions, Submersion/In-Line RBI-546
Typical Mounting Details - RBI-547 Retraction Version
Note: Sensor must be mounted at an angle between 10° and 90° above the horizontal. Pipe tees and weldalets provided by customer.
2 EmersonProcess.com/LiquidAnalysis
Page 3
September 2015
Ball Valve Kit (PN 23240-00) used with RBI-547 retractable sensor A process connector (PN 23166-00 or -01) must be used
RBI pH Sensors
to connect the sensor to Ball Valve Kit 23240-00.
Process connector can be purchased separately.
A B C
MM/IN MM/IN MM/IN
482/19.0 290–11.4 521–20.5
Dimensional Drawings of RBI-547 with and without 1–½ in. Ball Valve Kit PN 23240-00
For the ball valve installation shown, the ball valve kit (PN 23240-00) and 1 in. x 1 in. process connector (PN 23166-00 or 23166-01) must be purchased separately. Note: The 1¼'' ball valve assembly (PN 23765-00) is not shown, but is compatible with RBI-547. Note: Add ve (5) inches to length of sensor if mounting a sensor-head junction box onto the sensor.
EmersonProcess.com/LiquidAnalysis 3
Page 4
RBI pH Sensors September 2015
RBI-547 Sensor Cartridge
RBI-[ ]-547_ _ _ PT100 TC
4 EmersonProcess.com/LiquidAnalysis
Page 5
September 2015 RBI pH Sensors
RBI-547 Sheath
EmersonProcess.com/LiquidAnalysis 5
Page 6
RBI pH Sensors September 2015
RBI-547 Assembly (Cartridge and Stretch)
6 EmersonProcess.com/LiquidAnalysis
Page 7
September 2015 RBI pH Sensors
Ordering Information - RBI
The RBI pH Sensor is built with a Kynar® body and is designed in two congurations. The 546 sensor is housed in a fully encapsulated
body for use in pipe tees or submersed in a tank. The 547 is the ball valve retractable version; the sensor body can be replaced and
the titanium sensor sheath assembly (sold separately) can be reused. Both sensor types are built with a high performance J glass,
notched tip with a Teon junction, and a PT100 RTD. The integral cable is prepared with tinned leads for easy wiring.
Model Description Example: RBI-V-546-R-DT-C-S-05
RBI pH Sensor with Multi-Junction
O-ring Material
V Viton
E EPDM
K FFKM (peruoro-elasteme), generic Kalrez
Body
546 ¾'' MNPT Inline/Submersion
547 Replacement cartridge valve insertion (tube ordered separately)
Measuring Electrode
R Ruggedized, Hemi-glass (0–14 pH), 15 to 130 °C
FG Ruggedized, Flat-glass (0–14 pH), 15 to 130 °C, Not available with option -546
PX Platinum ORP, Flat Solid Billet (0 to ± 1500 m V), 0 to 130 °C
Tip Conguration
DT Dual notch teflon junction
FT Flush no tip protection, Not available with option 546, -R and -PX
Temperature Compensation
C PT100 RTD (3-wire), Not available with option -PX
N None, Only available with option -PX
Body Options
S Standard body
Insertion Depth
05 0.5'' for option 546 only
N None, standard replacement sensor cartridge for option 547 only
Cable Conguration
15 15' cable
N 30' cable
Cable Conguration
E Reference on separate wire
Lead Terminations
TT All tinned leads
B2 BNC (with Molex for temp comp), Not available with option -FG, -PX, and -FT
EmersonProcess.com/LiquidAnalysis 7
Page 8
RBI pH Sensors Product Data Sheet
LIQ-PDS-RBI-pH-Sensors September 2015
Retraction Accessories
Part # Description
RB5104-0058E 8'' Titanium,Kynar, EPDM, 547 retractable insertion tube
RB5104-0078E 8'' Hastelloy-C, Kynar, EPDM, 547 retractable insertion tube
RB5104-0120V 20'' Titanium, Kynar, Viton, 547 retractable insertion tube
RB5104-0120E 20'' Titanium, Kynar, EPDM, 547 retractable insertion tube
RB5104-0320E 20'' Hastelloy-C, Kynar, EPDM, 547 retractable insertion tube
RB5104-0136E 36'' Titanium, Kynar, EPDM, 547 retractable insertion tube
RB5104-0336E 36'' Hastelloy-C, Kynar, EPDM, 547 retractable insertion tube
EmersonProcess.com/LiquidAnalysis
YouTube.com/user/RosemountAnalytical
Twitter.com/Rosemount_News
Emerson Process Management
Rosemount Analytical 2400 Barranca Parkway
Irvine, CA 92606
USA Toll Free + 800 854 8257 T + 949 757 8500 F + 949 474 7250
Liquid.CSC@Emerson.com
Analyticexpert.com
Facebook.com/Rosemount
©2015 Emerson Process Management. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount Analytical
is a mark of one of the Emerson Process Management family of companies. All other marks are
the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees,
express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specications of our products at any
time without notice.
Loading...