RBI pH Sensors Product Data Sheet
LIQ-PDS-RBI-pH-Sensors September 2015
Threaded In-Line, Submerged or
Ball-valve Insertion pH Sensor
Maximum sensor life due to proprietary
multi-junction which hinders reference
poisoning.
Chemically resistant KYNAR
®
Rugged, quick change, quick clean.
Optional installation methods allows
standardization of entire plant with a minimum
of sensor part numbers.
Assembled in a reusable Titanium sheath,
Hi-Pressure body (RB-547).
RBI-546
Submersion/In-line pH Sensor
RBI-547
Ball Valve Insertion pH Sensor
Installation
RBI-546
Threaded in-line: ¾'' MPT threaded nose for installation into
process, sample line.
Submerged: ¾'' MNPT threaded top for connection to ¾'' FNPT
coupling extension pipe.
RBI-547
Insertion: 1'' MPT process tting or, 1¼'' and 1½'' ball valve
assemblies. Insertion depth is user selectable.
Submerged: 1'' MNPT process tting, into 1'' FNPT coupling
or ¾'' submersion extension pipe.
Sensor & Transmitter types supported:
pH: Compatible with all Rosemount Analytical industrial
transmitters and analyzers.
Wetted Materials:
Kynar, Titanium (RBI-547 only), Porous TEFLON,
Wood and Glass, choice of: Viton
Maximum Pressure & Temperature Ratings:
Max Temperature: 266 °F/130 °C at 40 psig/276 KPa
Max Pressure: 150 psig/1035 KPa at 158 °F/70 °C
Maximum Insertion/Retraction Pressure:
(option 546) 65 psig/448 KPa @ 158 °F/70 °C
(option 547) 40 psig/276 KPa @ 266 °F/130 °C
Measurement Range:
pH: 0–14 pH
®
, EPDM®, or Kalrez®.
RBI pH Sensors September 2015
Sensor Dimensions, Submersion/In-Line RBI-546
Typical Mounting Details - RBI-547 Retraction Version
Note: Sensor must be mounted at an angle between 10° and 90° above the horizontal. Pipe tees and weldalets provided by customer.
2 EmersonProcess.com/LiquidAnalysis
September 2015
Ball Valve Kit (PN 23240-00) used with RBI-547 retractable sensor A process connector (PN 23166-00 or -01) must be used
RBI pH Sensors
to connect the sensor to Ball Valve Kit 23240-00.
Process connector can be purchased separately.
A B C
MM/IN MM/IN MM/IN
482/19.0 290–11.4 521–20.5
Dimensional Drawings of RBI-547 with and without 1–½ in. Ball Valve Kit PN 23240-00
For the ball valve installation shown, the ball valve kit (PN 23240-00) and 1 in. x 1 in. process connector (PN 23166-00 or 23166-01)
must be purchased separately. Note: The 1¼'' ball valve assembly (PN 23765-00) is not shown, but is compatible with RBI-547.
Note: Add ve (5) inches to length of sensor if mounting a sensor-head junction box onto the sensor.
EmersonProcess.com/LiquidAnalysis 3