Rosemount Analytical Hx338+ Operating Manual

Page 1
Rosemount Hx338+ IQ/OQ Instructions Sheet
LIQ-IQ/OQ-Instruction Sheet, Rev BA
October 2017
Rosemount™ Hx338+
Steam Sterilizable and Autoclavable Pre-Pressurized pH Sensor
Page 2
Page 3

Contents

Contents
1.1 Purpose ..........................................................................................................................................1
1.2 Applicable documents ....................................................................................................................1
1.3 Installation Qualification (IQ) ..........................................................................................................1
1.4 Operation Qualification (OQ) ......................................................................................................... 2
1.5 IQ/OQ form .................................................................................................................................... 3
Rosemount Hx338+ IQ/OQ Instructions Sheet i
Page 4
Contents
ii Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 5

Rosemount Hx338+ IQ/OQ Instruction Sheet

1 Rosemount Hx338+ IQ/OQ
Instruction Sheet

1.1 Purpose

The purpose of this document is to describe the procedure for an Installation Qualification (IQ) and Operation Qualification (OQ) for Rosemount Hx338+ pH sensors.
The IQ/OQ may be performed at a customer site by following these instructions.

1.2 Applicable documents

To carry out the IQ/OQ procedures, the following documents must be available:
Quick Start Guide for the Rosemount Hx338+ sensor
Instruction Manual for the pH transmitter used to conduct the IQ/OQ
IQ/OQ Form (Section 1.5)
1.3

Installation Qualification (IQ)

1. Visual inspection.
Check to see if there is any obvious physical damage to the sensor.
No damage = passed, damage = failed.
Note any visible defects.
2. Verification of documentation.
a. Is the final test inspection (Declaration of Quality) for the sensor available?
Attach a copy to the completed IQ report.
b. Is the Quick Start Guide for the Rosemount Hx338+ sensor available? Attach a
copy to the IQ report.
3. Test connection of sensor to transmitter.
With the sensor connected to a pH transmitter according to the instruction manual for each, perform the following:
a. Remove the protective cap on the sensor and rinse in deionized water.
b. Immerse the sensor in pH 4 buffer and swirl.
c. Monitor the mV value for at least three minutes until stable.
d. mV reading must be positive and over 100 mV to pass.
Record value and note pass or fail.
Rosemount Hx338+ IQ/OQ Instructions Sheet 1
Page 6
Rosemount Hx338+ IQ/OQ Instruction Sheet
4. If the above steps are completed without any problem, note that the sensor has passed IQ.
5. Sign and date on the IQ section of the IQ/OQ Form (Section 1.5) .

1.4 Operation Qualification (OQ)

This test verifies the slope and offset of the pH sensor are within specification.
1. Perform a two-point calibration as directed by the transmitter instruction manual.
2. Use buffers of pH 7 and pH 4.
3. Make sure to rinse the sensor carefully when moving the sensor from one buffer to another.
4. Make sure that the mV reading has stabilized in one buffer solution before moving on to the second buffer solution.
5. Once the calibration is completed, verify the sensor slope and offset which should be available from the calibration menu of the transmitter.
a. Offset must be between -20 mV and +20 mV to pass.
b. Slope must be between 50.3 mV/pH unit and 65.0 mV/pH unit to pass (or 85% to
110% on some transmitters).
6. If the above steps are completed without any problem, note that the sensor has passed OQ.
7. Sign and date on the OQ section of the IQ/OQ Form (Section 1.5).
2 Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 7

1.5 IQ/OQ form

Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 3
Page 8
Rosemount Hx338+ IQ/OQ Instruction Sheet
4 Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 9
Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 5
Page 10
Rosemount Hx338+ IQ/OQ Instruction Sheet
6 Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 11
Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 7
Page 12
GLOBAL HEADQUARTERS Emerson Automation Solutions 6021 Innovation Blvd Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 F +1 952 949 7001
liquid.csc@emerson.com www.Emerson.com/ RosemountLiquidAnalysis NORTH AMERICA Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317
Toll Free +1 800 999 9307
F +1 952 949 7001
liquid.csc@emerson.com www.Emerson.com/ RosemountLiquidAnalysis
LIQ-IQ/OQ-Instruction Sheet
Rev BA
2017
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland
T + 41 (0) 41 768 6111
F + 41 (0) 41 768 6300
liquid.csc@emerson.com www.Emerson.com/ RosemountLiquidAnalysis
Linkedin.com/company/Emerson-Automation-Solutions
twitter.com/rosemount_news
Facebook.com/Rosemount
youtube.com/RosemountMeasurement
google.com/+RosemountMeasurement
AnalyticExpert.com
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
T +971 4 811 8100 F +971 4 886 5465
liquid.csc@emerson.com www.Emerson.com/ RosemountLiquidAnalysis
©
2017 Rosemount. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and, while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore
T +65 777 8211 F +65 777 0947
liquid.csc@emerson.com www.Emerson.com/ RosemountLiquidAnalysis
Loading...