Rosemount Hx338+ IQ/OQ Instruction Sheet
1Rosemount Hx338+ IQ/OQ
Instruction Sheet
1.1Purpose
The purpose of this document is to describe the procedure for an Installation Qualification
(IQ) and Operation Qualification (OQ) for Rosemount Hx338+ pH sensors.
The IQ/OQ may be performed at a customer site by following these instructions.
1.2Applicable documents
To carry out the IQ/OQ procedures, the following documents must be available:
•Quick Start Guide for the Rosemount Hx338+ sensor
•Instruction Manual for the pH transmitter used to conduct the IQ/OQ
•IQ/OQ Form (Section 1.5)
1.3
Installation Qualification (IQ)
1.Visual inspection.
Check to see if there is any obvious physical damage to the sensor.
No damage = passed, damage = failed.
Note any visible defects.
2.Verification of documentation.
a. Is the final test inspection (Declaration of Quality) for the sensor available?
Attach a copy to the completed IQ report.
b. Is the Quick Start Guide for the Rosemount Hx338+ sensor available? Attach a
copy to the IQ report.
3.Test connection of sensor to transmitter.
With the sensor connected to a pH transmitter according to the instruction manual
for each, perform the following:
a. Remove the protective cap on the sensor and rinse in deionized water.
b. Immerse the sensor in pH 4 buffer and swirl.
c. Monitor the mV value for at least three minutes until stable.
d. mV reading must be positive and over 100 mV to pass.
Record value and note pass or fail.
Rosemount Hx338+ IQ/OQ Instructions Sheet 1
Page 6
Rosemount Hx338+ IQ/OQ Instruction Sheet
4.If the above steps are completed without any problem, note that the sensor has
passed IQ.
5.Sign and date on the IQ section of the IQ/OQ Form (Section 1.5) .
1.4Operation Qualification (OQ)
This test verifies the slope and offset of the pH sensor are within specification.
1.Perform a two-point calibration as directed by the transmitter instruction manual.
2.Use buffers of pH 7 and pH 4.
3.Make sure to rinse the sensor carefully when moving the sensor from one buffer to
another.
4.Make sure that the mV reading has stabilized in one buffer solution before moving
on to the second buffer solution.
5.Once the calibration is completed, verify the sensor slope and offset which should
be available from the calibration menu of the transmitter.
a. Offset must be between -20 mV and +20 mV to pass.
b. Slope must be between 50.3 mV/pH unit and 65.0 mV/pH unit to pass (or 85% to
110% on some transmitters).
6.If the above steps are completed without any problem, note that the sensor has
passed OQ.
7.Sign and date on the OQ section of the IQ/OQ Form (Section 1.5).
2Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 7
1.5IQ/OQ form
Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 3
Page 8
Rosemount Hx338+ IQ/OQ Instruction Sheet
4Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 9
Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 5
Page 10
Rosemount Hx338+ IQ/OQ Instruction Sheet
6Rosemount Hx338+ IQ/OQ Instructions Sheet
Page 11
Rosemount Hx338+ IQ/OQ Instruction Sheet
Rosemount Hx338+ IQ/OQ Instructions Sheet 7
Page 12
GLOBAL HEADQUARTERS
Emerson Automation Solutions
6021 Innovation Blvd
Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888
F +1 952 949 7001
liquid.csc@emerson.com
www.Emerson.com/
RosemountLiquidAnalysis
NORTH AMERICA
Emerson Automation Solutions
8200 Market Blvd
Chanhassen, MN 55317
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a
mark of one of the Emerson Process Management family of companies. All other marks
are the property of their respective owners. The contents of this publication are presented
for information purposes only, and, while effort has been made to ensure their accuracy,
they are not to be construed as warranties or guarantees, express or implied, regarding
the products or services described herein or their use or applicability. All sales are
governed by our terms and conditions, which are available on request. We reserve the
right to modify or improve the designs or specifications of our products at any time
without notice.