Rosemount Analytical 56 Operating Manual

Page 1
Instruction Manual
LIQ-MAN-56
Rev. D
April 2017
Rosemount
56
Advanced Dual-Input Analyzer
Page 2
Page 3
Essential Instructions
Read this page before proceeding
ave been designed, and tested to meet many national and international standards. Experience indicates that its perform-
h ance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instru­ment. To ensure their continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards august be impaired.
• Failure to follow the proper instructions august cause any one of the following situations to occur: Loss of life; per­sonal injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual cov­ers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and proce­dures can affect the performance and cause unsafe operation of your process.
• All instrument enclosures must be closed and protective covers must be in place unless qualified personnel are per­forming maintenance.
WARNING
RIS
K
OF
EL
EC
TRIC
Equipme
I
ns
tallation and
M
ain powe
Do
no
t o
Si
gnal w
N
on
-
metallic
ings and jumper wires.
Unus
ed ca
pli
ance with
Ty
pe 4X or IP6
Electric a
p
plicable
•Operate
Safe
t
y
and
s
ou
r
ce
.
Proper
nt
pr
se
r
w
ir
ed to
pe
r
at
e or
iring connect
c
able
ble condu
pers
o
nal saf
6
condu
al installation must
nati
o
nal or
o
n
l
y
with
p
erformance
u
s
e
and
co
n
AL
S
HO
o
t
ect
ed th
rvicing of th
se
p
ar
at
e
ne
rgi
z
e
ins
ed in th
s
tr
ain
r
e
it entrie
s must be securely sealed by non-flammable closures to provide enclosure integrity in com
ety and enviro
it plugs t
local
f
ro
n
t
pan re
qui
f
igu
rati
o
C
K
r
o
ugho
ut by
do
uble ins
ulation.
is
pr
oduct a
ugus
t ex
pose
e
powe
r
so
ur
ce
m
us
t
be disconnec
trume
nt w
ith
c
a
se o
pe
n!
is
box
m
us
t be r
at
ed at lea
liefs
do
no
t
pr
o
vi
de
gr
o
unding
nmental protection requirements. Unused conduit openings must be sealed with
o maint
ain the ingress protection rating (Type 4X)
be in ac codes.
el
n
c
ordance with the National Electrical Code (ANSI/NFPA-70) and/or any other
f
asten
e
d and in place.
re
that this instrument be connected and properly grounded through a three-wire power
is
th
e
responsibility of the user.
pe
s
t 2
rsonne
t
40
V
betwee
l
to dange
ed
befor
for
Eur
n
e
se
o
pean m
conduit
r
o
us
rvicing.
connec
volt
ains
ages
o
pe
tions
.
r
ation.
!
U
se
gr
o
unding
ty
pe
bus
h
-
-
Page 4
CAUTION
This product generates, uses, and can radiate radio frequency energy and thus can cause radio com­munication interference. Improper installation, or operation, august increase such interference. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of
lass A computing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to pro-
C vide reasonable protection against such interference. Operation of this equipment in a residential area august cause interference, in which case the user at his own expense, will be required to take whatever measures august be required to correct the interference.
CAUTION
This product is not intended for use in the light industrial, residential or commercial environments per the instrument’s certification to EN61326-1:2006.
Page 5
Section i: Quick Start Guide
1. Refer to Section 2.0 for mechanical installation instructions.
. Wire sensor(s) to the signal boards. See Section 3.0 for wiring instructions. Refer to the
2
sensor instruction sheet for additional details. Make current output, alarm relay and power connections
3. Once connections are secured and verified, apply power to the analyzer.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol appears, consult the manual for appropriate action.
4. When the analyzer is powered up for the first time, Time/Date and Quick Start screens appear. Quick Start operating tips are as follows:
a. Window screens will appear. The field with the focus will appear with dark blue back-
lighting. The field with focus can be edited by press ENTER/MENU.
b. The Time and Date screen to set the real-time clock will appear. Accept the displayed
time by pressing ENTER on Time and date OK or press the down key to Change the
time and date.
c. The first Quick Start screen appears. Choose the desired language by pressing
ENTER/MENU to edit the active field and scrolling to the language of choice. Press ENTER/MENU and press the down arrow to highlight NEXT.
d. The Navigation Rules for operating the keypad will be displayed.
e. Choose the measurement for Sensor 1 (and Sensor 2) and proceed to the remaining
Quick Start steps.
f. Keypad operation guidelines will appear to guide the user how operate the user inter-
face.
g. NOTE: To edit a field with backlit focus, press ENTER/MENU. To scroll up or down, use
the keys to above or below the ENTER key. To move the cursor left or right, use the keys to the left or right of the ENTER key. To edit a numeric value including decimal points, use the alphanumeric keypad then press ENTER.
h. NOTE: Press ENTER to store a setting or value. Press EXIT to leave without storing
changes. Pressing EXIT during Quick Start returns the display to the initial start-up screen (select language). To proceed to the next Quick Start step, use the right key or the down key to highlight NEXT. Press ENTER.
5. After the last step, the main display appears. The current outputs are assigned to default
values before probes are wired to the analyzer. After the last step, the main display ap­pears. The outputs are assigned to default values.
6. To change output, and all settings, press ENTER/MENU from the live screen. Using the down and right arrow keys, select one of the following menus and navigate the screen of choice.
7. To return the analyzer to the default settings, choose Reset under the Menu selection screen.
Page 6
About This Document
This manual contains instructions for installation and operation of the 56 Advanced Dual-Input Analyzer. The following list provides notes concerning all revisions of this document.
Rev. Level Date Notes
A 08/11 This is the initial release of the product manual. The manual has been
reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.
B 11/12 Add new feature - configuration transfer via USB. Add new section for
C 11/13 Add sec. 9 - Advanced Relay Functions. Add procedure for Software Update
D 04/17 Updated the Address and Emerson Logo.
SAFETY MESSAGES
Procedures and instructions in this section august require special precautions to ensure the safety of the personnel performing the operations. Information that raises potential safety issues is indicated by a warning symbol ( ). This symbol identifies a potential hazard.
existing features - PID control and TPC relay activation, Non-Incendive Field Wiring drawings.
and Configuration Transfer via USB (for units manufactured October 2012 and later). Add procedure for Data Logger and Event Logger Download to USB. Add new sections for HART and Profibus Communications. Add Appendix 1 – HART and Device and Appendix 2 - HART Status Bits.
CAUTION
Page 7
Instruction Manual Table of Contents
LIQ-MAN-56 April 2017
Contents
Section 1: Description and Specifications
.1 Features and Applications ................................................................................1
1
1.2 Enhanced Features ..........................................................................................2
1.3 Specifications-General ....................................................................................3
1.4 Contacting Conductivity .................................................................................6
1.5 Toroidal Conductivity ......................................................................................7
1.6 pH/ORP ...........................................................................................................8
1.7 Flow ................................................................................................................9
1.8 4-20 mA Current Input.....................................................................................9
1.9 Chlorine.........................................................................................................10
1.10 Dissolved Oxygen .........................................................................................12
1.11 Dissolved Ozone............................................................................................12
1.12 Turbidity........................................................................................................13
1.13 Ordering Information.....................................................................................14
Section 2: Installation
2.1 Unpacking and Inspection .............................................................................15
2.2 Installation ....................................................................................................15
Section: 3 Wiring
3.1 General .........................................................................................................21
3.2 Preparing Conduit Openings..........................................................................22
3.3 Preparing Sensor Cable..................................................................................22
3.4 Power, Output, Alarms and Sensor Connections ............................................22
Section 4: Display and Operation
4.1 User Interface ................................................................................................31
4.2 Instrument Keypad........................................................................................31
4.3 Main Display ..................................................................................................31
4.4 Menu System.................................................................................................32
4.5 USB Data Port ................................................................................................33
4.6 56 Data Logger and Event Logger Download Procedure .................................33
4.7 Software Upgrade .........................................................................................35
4.8 Configuration Transfer...................................................................................35
Section 5: Programming the Analyzer - Basics
5.1 General .........................................................................................................37
5.2 Changing the Startup Settings ......................................................................37
5.3 Programming Temperature ..........................................................................38
5.4 Configuring and Ranging the Current Outputs...............................................38
5.5 Setting a Security Code .................................................................................39
5.6 Security Access..............................................................................................39
5.7 Using Hold ....................................................................................................40
5.8 Resetting Factory Defaults – Reset Analyzer ..................................................40
5.9 Programming Alarm Relays............................................................................40
Table of Contents i
Page 8
Table of Contents Instruction Manual
April 2017 LIQ-MAN-56
Section 6: Programming - Measurements
6.1 Programming Measurements – Introduction ................................................43
6.2 pH .................................................................................................................43
6.3 ORP ..............................................................................................................44
6.4 Contacting Conductivity................................................................................45
6.5 Toroidal Conductivity ....................................................................................46
6.6 Chlorine.........................................................................................................47
6.7 Dissolved Oxygen ..........................................................................................50
6.8 Dissolved Ozone ...........................................................................................51
6.9 Turbidity .......................................................................................................51
6.10 Flow ..............................................................................................................52
6.11 Current Input ................................................................................................52
Section 7: PID Control
7.1 Introduction .................................................................................................55
7.2 PID Setup.......................................................................................................59
Section 8: Time Proportional Control
8.1 Introduction .................................................................................................63
8.2 TPC Setup......................................................................................................63
Section 9: Alarm Relay Functions
9.1 General..........................................................................................................67
9.2 High/Low Concentration Alarm......................................................................67
9.3 Delay Timer: ..................................................................................................68
9.4 Bleed and Feed ..............................................................................................70
9.5 Totalizer Based Relay Activation.....................................................................71
9.6 Interval Timer ................................................................................................72
9.7 Date and Time Activation ..............................................................................74
Section 10: Calibration
10.1 Calibration – Introduction .............................................................................75
10.2 pH Calibration ...............................................................................................75
10.3 ORP Calibration .............................................................................................76
10.4 Contacting Conductivity Calibration .............................................................77
10.5 Toroidal Conductivity Calibration ..................................................................79
10.6 Chlorine Calibration ......................................................................................80
10.7 Oxygen Calibration .......................................................................................82
10.8 Ozone Calibration .........................................................................................84
10.9 Calibrating Temperature................................................................................85
10.10 Turbidity .....................................................................................................85
10.11 Pulse Flow ...................................................................................................86
ii Table of Contents
Page 9
Instruction Manual Table of Contents
LIQ-MAN-56 April 2017
Section 11: HART®Communications
11.1 Introduction ..................................................................................................87
11.2 Physical Installation and Configuration ..........................................................88
11.3 Measurements Available via HART .................................................................89
11.4 Diagnostics Available via HART ......................................................................90
11.5 HART Hosts ...................................................................................................91
11.6 Wireless Communication using the 56...........................................................94
11.7 Field Device Specification (FDS) .....................................................................94
Section 12: Profibus Communications
12.1 General ..........................................................................................................95
12.2 Profibus Features ...........................................................................................95
12.3 Profibus Communications..............................................................................96
12.4 Data Transmission .......................................................................................100
12.5 Installation and Wiring.................................................................................111
Section 13: Maintenance
13.1 Overview .....................................................................................................113
13.2 Analyzer Maintenance..................................................................................113
13.3 USB Port ......................................................................................................113
Section 14: Return of Material
14.1 General........................................................................................................115
14.2 Warranty Repair ..........................................................................................115
14.3 Non-Warranty Repair ..................................................................................115
HART Appendix 1 ................................................................................................117
HART Appendix 2 ................................................................................................121
EC Declaration of Conformity .....................................................................125
Table of Contents iii
Page 10
Table of Contents Instruction Manual
April 2017 LIQ-MAN-56
iv
Page 11
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
Section 1: Description and Specifications
1.1 Features and Applications
This multi-parameter unit serves industrial, commercial and municipal applications with the widest range of liquid measurement inputs and digital communications available.
The 56 advanced dual-input analyzer supports continuous measurement of liquid analytical inputs from one or two sensors. The modular design allows signal input boards to be field replaced, making configuration changes easy. The high resolution full-color display gives unsurpassed visibility and functionality for liquid analytical instrumentation.
Dual Input Instrument: single or dual measurement of pH/ORP, Resistivity/ Conductivity, % Concentration, Total Dissolved Solids, Total Chlorine, Free Chlorine, Monochloramine, Dissolved Oxygen, Dissolved Ozone, Turbidity, Pulse Flow, Temperature, and 4-20 mA input from any device.
Full Color Display: The high resolution full-color display allows at-a-glance viewing of process readings – indoors or outdoors. Six additional process variables or diagnostic parameters are displayed for quick determination of process or sensor condition. The contrast of back-lit display can be adjusted and the main screen can be customized to meet user requirements.
Digital Communications: HART®version 5 and 7 digital communications are available on the
56. An optional Profibus®DP digital communications board is available for Profibus installations. 56 HART units communicate with the 475 HART hand-held communicator and HART hosts such as AMS Intelligent Device Manager. 56 Profibus units are fully compatible with Profibus DP networks and Class 1 or Class 2 masters. HART and Profibus DP configured units will support any single or dual measurement configurations of the 56.
Menus: Easily-managed window screens for easy navigation to local configuration and routine calibration. Quick Start and all menu screens are available in multiple locally displayed languages. Alpha-numeric keypad allows easy entries during configuration and calibration.
Quick Start Programming: Popular Quick Start screens appear the first time the unit is powered. The instrument auto-recognizes each measurement input type and prompts the user to configure each sensor loop in a few quick steps for immediate commissioning.
User Help Screens: A complete user guide and troubleshooting manual is embedded in the instrument’s memory and easily accessed via the INFO key on the local display. Detailed instructions and troubleshooting tips in multiple languages are intended to provide adequate guidance to resolve most problems on site.
Hazardous Area Approvals and Safety Approvals: None.
Enclosure: The instrument enclosure fits standard ½ DIN panel cutouts. The versatile
enclosure design supports panel-mount, pipe-mount, and surface/wall-mount installations. No Enclosure ratings – None.
Security Access Codes: Two levels of security access are available. Program one access code for routine maintenance and hold of current outputs; program another access code for all configuration menus and functions.
Description and Specifications 1
Page 12
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
Diagnostics: The analyzer continuously monitors itself and the sensor(s) for fault and warning conditions. A display banner flashes red to indicate a Fault condition and yellow for a
arning condition to visually alert field personnel. Details and troubleshooting information
W for any specific fault or warning can be readily accessed by pressing the INFO key.
Local Languages: Rosemount extends its worldwide reach by offering nine menu languages – English, French, German, Italian, Spanish, Portuguese, Chinese, Russian and Polish. Every unit includes user programming menus; calibration routines; faults and warnings; and user help screens in all nine languages.
Current Outputs: Every unit includes four 4-20 mA or 0-20 mA electrically isolated current outputs giving the ability to transmit the measurement value and the temperature for both sensors. Users have wide latitude to assign any measurement value or live diagnostic to any current output for reporting. Output dampening can be enabled with time constants from 0 to 999 seconds. HART digital communications transmitted via current output 1 is standard on all units (option code HT).
1.2 Enhanced Features
Process Trending Graphs: High-resolution color graphs of measurement data can be displayed
on-screen to pinpoint process disruptions or measurement problems and to estimate probe maintenance frequency. The analyzer gives the user the ability to zoom in to a specific narrow timeframe of process measurements for detailed on-screen evaluation.
Data Logger and Event Logger: Extensive onboard data storage captures measurement data from both channels every 30 seconds for 30 days for on-screen display or local upload to a USB 2.0 memory device. 300 significant analyzer events are recorded including start-up time, calibrations, hold outputs, configurations, alarms, power interruptions, faults, and more. All process data and events are time/date stamped.
USB 2.0 Data Transfer Port: A USB port is built-in to allow local data transfer of process data and events using a standard USB memory device. Cleanly formatted EXCEL data is useful for evaluation of process data on a computer and identification of critical alarm or fault events.
PID Control: Proportional, Integral and Derivative settings allow the analog current outputs to adjust a control device that has continuous adjustability by acting on process measurements or temperature. PID is typically used on modulating control devices such as automated control valves or variable volume pumps. Any current output can be programmed for PID functions.
Alarm Relay Capabilities: Four Single Pole Double Throw alarm relays are fully assignable and programmable to trigger alarms upon reaching measurement or diagnostics setpoints or fault conditions. Further relay settings include TPC, synchronized interval timers and four specialized timer functions described below. All relays are independently activated. Failsafe operation and programming of relay default state (normally open or normally closed) is software selectable.
Timer Functions: Basic TPC (Time Proportional Control) settings are available. Interval timers set relays by interval time, on-time and recovery time for discrete on/off control devices based on measurement inputs. In addition, four real-time clock relay functions are implemented including: bleed and feed, day and time interval timers, delay timer and a flow totalizer. These advanced timer features support a number of specialized applications that normally require dedicated timer control devices or DCS programming.
2 Description and Specifications
Page 13
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
Wireless Thum Adaptor Compatible: Enable wireless transmissions of process variables and
iagnostics from hard-to-reach locations where it is impractical to run wires for current
d outputs. When commissioned with the THUM Adaptor, 56 HART Emerson wireless networks using HART 7 wireless protocol.
Smart-Enabled pH: Rosemount SMART pH capability can eliminate field calibration of pH probes through automatic upload of calibration data and history – fully calibrating the pH loop. pH probe changes are literally plug and play using SMART pH sensors with VP cables connections.
Advanced Functions: Several specialty measurements are supported including: high reference impedance pH sensors, Ion Selective Electrode measurements, pH loop calibration by entering pH slope and reference offset, Isopotential point for pH, inferred pH determination using dual contacting conductivity inputs, differential conductivity, differential flow, totalized flow, current input from any 4-20 mA source, dual range calibration for chlorine sensors, programmable polarizing voltage for amperometric oxygen sensors and software selectable normally open or normally closed alarm relays – to name a few.
®
units can communicate on
1.3 Specifications - General
Case: Polycarbonate. Type 4X, IP66.
NOTE:
To ensure a water-tight seal, tighten all four front panel screws to 6 in-lbs of torque.
Dimensions: 6.2 x 6.2 x 5.2 in. (157 x 157 x 132 mm)
Conduit openings: Accepts (6) PG13.5 or 1/2 in. conduit fittings
Display: Large 3.75 x 2.2 in. (95.3 x 55.9mm) high resolution color LCD displays large
process variables and user-definable display of diagnostic parameters. Calibration, programming and information screens display clear, easy-to-read characters. The color display is back-lit and backlighting intensity is user adjustable. Measurement character height: (.5") 13mm. Main display can be customized to meet user requirements.
Ambient temperature and humidity: -10 to 60 °C, (14 to 140 °F) RH 5 to 95% (non­condensing). For Turbidity only: 0 to 55 °C (32 to 131 °F). RH 5 to 95% (non-condensing).
NOTE:
The analyzer is operable from -5 to 55 °C (-23 to 131 °F) with some degradation in display response or per­formance. Above 60 °C, the following components will progressively and automatically shut down: dis­play, USB communications port, current outputs, alarm relays, main circuit board.
WARNING
Always remove USB memory device at ambient temp above 60 °C. Do not access USB port if combustible atmosphere is present.
Storage temperature: -20 to 60 °C, (-4 to 140 °F)
Power: Code -02: 20 to 30 VDC. 20 W
Code –03: 85 to 264 VAC, 47.5 to 65.0 Hz, 20 W
Real time clock back-up: 24 hours.
Description and Specifications 3
Page 14
Section 1: Description and Specifications Instruction Manual
U
L
US
C
LISTED
April 2017 LIQ-MAN-56
Hazardous Location Approvals:
Options for CSA: 02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38, HT and DP
Class I, Division 2, Groups A, B, C, & D Class Il, Division 2, Groups E, F, & G
lass Ill
C T4 Tamb = 60 °C Enclosure Type 4X, IP66
See Non-Incendive Field Wiring drawing 1400668. The ‘C’ and ‘US’ indicators adjacent to the CSA Mark signify that the product has been evaluated to the applicable CSA and ANSI/UL Standards, for use in Canada and the U.S. respectively. Evaluated to CSA Standards: 22.2 Numbers: 1-10. 0.4-04, 25-1996, 94-M1991, 142-M1987, 213-M1987, 60529:05. ANSI/IEC 60529:04. ANSI/ISA:12.12.01:2007. UL No. 50:11th Ed. and No. 508:17th Ed.
Note: Single-input Turbidity configurations (models 56-02-27-38 or -HT, 56-03-27-38 or -HT) and dual­input Turbidity only configurations (56-02-27-37 or -HT, 56-03-27-37 -HT) are CSA approved class I Div. 2 for hazardous area installation.
Options for FM: -02, 03, 20, 21, 22, 23, 24, 25, 26, 27, 30, 31, 32, 33, 34, 35, 36, 37, 38, HT and DP.
Class I, Division 2, Groups A, B, C, & D Class Il & lll, Division 2, Groups E, F, & G T4 -10 °C ≤ Tamb ≤ 60 °C IP66
S
ee Non-Incendive Field Wiring drawing 1400667.
Evaluated to FM Standards: 3600:2011, 3611:2004, 3810:2005, ANSI/IEC: 60529:2004.
Note: Single-input Turbidity configurations (models 56-02-27-38 or -HT, 56-03-27-38 or -HT) and dual-input Turbidity only configurations (56-02-27-37 or -HT, 56-03-27-37 or -HT) are FM approved class I Div. 2 for hazardous area installation.
Ordinary Locations (only with - UL ordering option):
Options for UL: -02, 03, 20, 21, 22, 24, 25, 26, 27, 30, 31, 32, 34, 35, 36, 37, 38,
HT and DP.
Pollution Degree 2: Normally only non-conductive pollution occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Altitude: for use up to 2000 meter (6562 ft.)
RFI/EMI: EN61326-1:2006
LVD: EN-61010-1:2010
Input: One or two isolated sensor inputs. Measurement choices of pH/ORP, resistivity/conductivity/ TDS, %
concentration, ratio conductivity, total and free chlorine, monochloramine, dissolved oxygen, dissolved ozone, turbidity, pulse flow, temperature and raw 4-20 mA input. For contacting conductivity measurements, temperature element can be a Pt100 RTD or Pt1000 RTD. For other measurements (except ORP, flow and turbidity), use either a PT100 RTD, PT1000 RTD, or 22k NTC (D.O. only).
4 Description and Specifications
Page 15
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
Outputs: Four 4-20 mA or 0-20 mA isolated current outputs. Fully scalable. Max Load: 550 Ohms. Output 1 superimposes the HART®digital signal. Outputs can be programmed for
ID control. Output dampening can be enabled with time constants from 0 to 999 seconds.
P HART digital communications transmitted via current output 1 is standard on all units (option code HT).
Alarms: Four alarm relays for process measurement(s) or temperature. Any relay can be programmed for any measurement, timer, TPC or fault alarm operation, instead of a process alarm. When selected, a fault alarm will activate the relay when a sensor or analyzer fault occurs. Each relay can be configured independently. Alarm logic (high or low activation or USP*) and deadband are user-programmable.
*USP alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the limit. conductivity/resistivity measurement only)
Relays: Form C, SPDT, epoxy sealed
Maximum Relay Current
Power Input Resistive
28 VDC 5.0 A 5.0 A
115 VAC 5.0 A 5.0 A
230 VAC 5.0 A 5.0 A
Inductive load: 1/8 HP motor (max.), 115/240 VAC
Terminal Connections Rating:
Power connector ( 02 order code, 24 VDC power supply and 03 order code, 85-264 VAC power supply): 24-12 AWG wire size.
Signal board terminal blocks: 26-16 AWG wire size.
Current output connectors: 26-16 AWG wire size.
Alarm relay terminal blocks: 24-12 AWG wire size.
Weight/Shipping Weight: (rounded up to nearest lb or nearest 0.5 kg): 3 lbs/4 lbs
(1.5 kg/2.0 kg)
Description and Specifications 5
Page 16
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
1.4 Contacting Conductivity (Codes -20 and -30)
easures conductivity in the range 0 to 600,000 µS/cm (600 mS/cm). Measurement choices are
M conductivity, resistivity, total dissolved solids, salinity, and % concentration. Temperature compensation can be disabled, allowing the analyzer to display raw conductivity.
NOTE:
When two contacting conductivity sensors are used, The 56 can derive an inferred pH value. Inferred pH is calculated pH, not directly measured pH. Inferred pH is calculated from straight and cation con­ductivity. It is applicable only if the alkalizing agent is NaOH or NH It is strictly an application for power plants.
Performance Specifications - Analyzer
Measurement Range: see table below Solution temperature compensation: manual slope (X% / °C), high purity water (dilute
sodium chloride), and cation conductivity (dilute hydrochloric acid).
Salinity: uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65 Five percent concentration curves: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, 0-25% or 96-
99.7% H2SO4. The conductivity concentration algorithms for these solutions are fully temperature compensated. Four temperature compensation options: manual slope (X% / °C), high purity water (neutral salt), cation conductivity (dilute hydrochloric acid) and raw.
Input filter: time constant 1 - 999 sec, default 2 sec. Response time: 3 seconds to 95% of final reading
and the major contaminant is NaCl.
3
Recommended Sensors for Contacting Conductivity:
All Rosemount ENDURANCE 400 series conductivity sensors (Pt 1000 RTD) and 410VP 4­electrode high-range conductivity sensor.
PERFORMANCE SPECIFICATIONS
Recommended Range – Contacting Conductivity
Cell 0.01 S/cm 0.1 µS/cm 1.0 µS/cm 10 µS/cm 100 µS/cm 1000 µS/cm 10mS/cm 100mS/cm 1000mS/cm Constant
0.01
0.1
1.0
4-electrode
Temperature Specifications:
Temperature range 0-200 °C
Temperature Accuracy, Pt-1000, 0-50 °C
Temperature Accuracy,
Pt-1000, Temp. > 50 °C
0.01 µS/cm to 200 µS/cm
0.1 µS/cm to 2000 µS/cm
1 µS/cm to 20mS/cm
± 0.1 °C
± 0.5 °C
200 µS/cm to 6000 µS/cm
2000 µS/cm to 60 mS/cm
20 mS/cm to 600 mS/cm
2 µS/cm to 1400 mS/cm
Cell Constant Linearity
±0.6% of reading in recommended range
+2 to -10% of reading outside high recommended range
±5% of reading outside low recommended range
±4% of reading in recommended range
6 Description and Specifications
Page 17
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
1.5 Toroidal Conductivity (Codes -21 and -31)
easures conductivity in the range of 1 (one) µS/cm to 2,000,000 µS/cm (2 S/cm). Measurement
M choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration. Temperature compensation can be disabled, allowing the analyzer to display raw conductivity.
For more information concerning the use and operation of the toroidal conductivity sensors, refer to the product data sheets.
Performance Specifications- Analyzer
Measurement Range: see table below Repeatability: ±0.25% ±5 µS/cm after zero cal Salinity: uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65 Five percent concentration curves: 0-12% NaOH, 0-15% HCl, 0-20% NaCl, 0-25% or 96-
99.7% H2SO4. The conductivity concentration algorithms for these solutions are fully temperature compensated. For other solutions, the analyzer accepts as many as five data points and fits either a linear (two points) or a quadratic function (three or more points) to the data. Reference temperature and linear temperature slope august also be adjusted for optimum results. Three temperature compensation options: manual slope (X% / °C), neutral salt (dilute sodium chloride) and raw.
Input filter: time constant 1 - 999 sec, default 2 sec. Response time: 3 seconds to 95% of final reading
Recommended Sensors:
All Rosemount submersion/immersion and flow-through toroidal sensors.
Loop Performance (Following Calibration)
Temperature range
Temperature Accuracy,
Pt-100, -25 to 50 °C
Temperature Accuracy,
Pt-100,. 50 to 210 °C
-25 to 210 °C (-13 to 410 °F)
± 0.5 °C
± 1°C
PERFORMANCE SPECIFICATIONS Recommended Range - Toroidal Conductivity
Model 1 µS/cm 10 µS/cm 100 µS/cm 1000 µS/cm 10 mS/cm 100 mS/cm 1000 mS/cm 2000 mS/cm
226
225 & 228
242
5 µS/cm to 500 mS/cm
15 µS/cm to 1500 mS/cm
100 µS/cm to 2000 mS/cm
226: ±1% of reading ±5 µS/cm in recommended range
225 & 228: ±1% of reading ±10 µS/cm in recommended range 222, 242: ±4% of reading in recommended range
225, 226 & 228: ±5% of reading outside high recommended range
226: ±5 µS/cm outside low recommended range
225 & 228: ±15 µS/cm outside low recommended range
500 mS/cm to 2000 mS/cm
1500 mS/cm to 2000 mS/cm
222 (1in & 2in)
Description and Specifications 7
500 µS/cm to 2000 mS/cm
Page 18
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
1.6 pH/ORP (Codes -22 and -32)
or use with any standard pH or ORP sensors. Measurement choices are pH, ORP, Redox,
F Ammonia, Fluoride or custom ISE. The automatic buffer recognition feature uses stored buffer pH values and their temperature curves for the most common buffer standards available worldwide. The analyzer will recognize the pH value of the buffer being measured and perform a self stabilization check on the sensor before completing the calibration. Manual or automatic temperature compensation is menu selectable. Change in process pH due to temperature can be compensated using a programmable temperature coefficient. For more information concerning the use and operation of the pH or ORP sensors, refer to sensor product data sheets. The 56 can also derive an inferred pH value. Inferred pH can be derived and displayed when two contacting conductivity sensors are used.
Performance Specifications (pH input) - Analyzer
Measurement Range [pH]: 0 to 14 pH Accuracy: ±0.01 pH Diagnostics: glass impedance, reference impedance Temperature coefficient: ±0.002pH / °C Solution temperature correction: pure water, high pH (dilute base), Ammonia and custom Buffer recognition:NIST (including non-NIST pH 7.01 buffer), DIN 19267, Ingold, Merck, and
Fisher
Input filter: Time constant 1 - 999 sec, default 4 sec. Response time: 5 seconds to 95% of final reading
Recommended Sensors for pH:
Compatible with standard pH sensors with and without integral preamps. Supports Smart pH sensors from Rosemount (includes Smart integral preamps).
General purpose and high performance pH 396PVP, 3900VP and 3300HT sensors
Performance Specifications (ORP input) - Analyzer
Measurement Range [ORP]: -1500 to +1500 mV Accuracy: ± 1 mV Temperature coefficient: ±0.12mV / °C Input filter: Time constant 1 - 999 sec, default 4 sec. Response time: : 5 seconds to 95% of final reading
Recommended Sensors for ORP:
Compatible with standard ORP sensors with and without integral preamps.
NOTE:
Some older sensor preamps august not be compatible with the 56 (contact the factory for details).
8 Description and Specifications
Page 19
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
1.7 Flow (Code -23 and -33)
or use with most pulse signal flow sensors, the 56 user-selectable units of measurement
F include flow rates in GPM (gallons per minute), GPH (gallons per hour), cu ft/min (cubic feet per min), cu ft/hour (cubic feet per hour), LPM (liters per minute), LPH (liters per hour), or m3/hr (cubic meters per hour), and velocity in ft/sec or m/sec. When configured to measure flow, the unit also acts as a totalizer in the chosen unit (gallons, liters, or cubic meters). Dual flow instruments can be configured as a % recovery, flow difference, flow ratio, or total (combined) flow.
Performance Specifications - Analyzer
Frequency Range: 3 to 1000 Hz Flow Rate: 0 - 99,999 GPM, LPM, m3/hr, GPH, LPH, cu ft/min, cu ft/hr. Totalized Flow: 0 – 9,999,999,999,999 Gallons or m3, 0 – 999, 999,999,999 cu ft. Accuracy: 0.5% Input filter: Time constant 0-999 sec., default 5 sec.
1.8 4-20 mA Current Input (Codes -23 and -33)
For use with any transmitter or external device that transmits 4-20 mA or 0-20 mA current outputs. Typical uses are for temperature compensation of live measurements (except ORP, turbidity and flow) and for continuous pressure input for continuous measurement of % oxygen gas. External input of atmospheric pressure for oxygen measurement allows continuous partial pressure compensation while the 56 enclosure is completely sealed.
Externally sourced current input is also useful for calibration of new or existing sensors that require temperature measurement or atmospheric pressure inputs. In addition to live continuous compensation of live measurements, the current input board can also be used simply to display and trend the measured temperature or the calculated partial pressure from the external device. This feature leverages the large display variables on the 56 as a convenience for technicians. Temperature can be displayed in °C or °F. Partial pressure can be displayed in inches Hg, mm Hg, atm (atmospheres), kPa (kiloPascals), bar or mbar. The current input board serves as a power supply for loop-powered devices that do not actively power their 4­20 mA output signals.
Performance Specifications
Measurement Range *[mA]: 0-20 or 4-20 Accuracy: ±0.03 mA Input filter: Time constant 0-999 sec., default 5 sec.
*Current input not to exceed 22 mA
Description and Specifications 9
Page 20
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
1.9 Chlorine (Code -24 and -34)
Free and Total Chlorine
The 56 is compatible with the 499ACL-01 free chlorine sensor and the 499ACL-02 total chlorine sensor. The 499ACL-02 sensor must be used with the TCL total chlorine sample conditioning system. The 56 fully compensates free and total chlorine readings for changes in membrane permeability caused by temperature changes.
For free chlorine measurements, both automatic and manual pH corrections are available. For automatic pH correction, select code P and an appropriate pH sensor. For more information concerning the use and operation of the amperometric chlorine sensors and the TCL measurement system, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100 µA Automatic pH correction (requires Code P): 6.0 to 10.0 pH Temperature compensation: Automatic or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading
Recommended Sensors
Chlorine: 499ACL-01 Free Chlorine or 499ACL-02 Total Chlorine pH: The following pH sensor is recommended for automatic pH correction of free chlorine
readings: 3900
Monochloramine
The 56 is compatible with the 499A CL-03 Monochloramine sensor. The 56 fully compensates readings for changes in membrane permeability caused by temperature changes. Because monochloramine measurement is not affected by pH of the process, no pH sensor or correction is required. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0 nA – 100 µA Temperature compensation: Automatic or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading
Recommended Sensors
Rosemount 499ACL-03 Monochloramine sensor
10 Description and Specifications
Page 21
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
pH-Independent Free Chlorine
he 56 is compatible with the 498CL-01 pH-independent free chlorine sensor. The 498CL-01
T sensor is intended for the continuous determination of free chlorine (hypochlorous acid plus hypochlorite ion) in water. The primary application is measuring chlorine in drinking water. The sensor requires no acid pre-treatment, nor is an auxiliary pH sensor required for pH correction. The 56 fully compensates free chlorine readings for changes in membrane permeability caused by temperature. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0 nA – 100 µA pH independent Temperature compensation: Automatic (via RTD) or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading
Recommended Sensors
Rosemount 498CL-01 pH independent free chlorine sensor
Chlorine sensors with Variopol connection and cable connection 498CL-01
Description and Specifications 11
Page 22
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
1.10 Dissolved Oxygen (Codes -25 and -35)
he 56 is compatible with the 499ADO, 499ATrDO, Hx438, Gx438 and BX438 dissolved
T oxygen sensors and the 4000 percent oxygen gas sensor. The 56 analyzer displays dissolved oxygen in ppm, mg/L, ppb, µg/L, % saturation, % O compensates oxygen readings for changes in membrane permeability caused by temperature changes. An atmospheric pressure sensor is included on all dissolved oxygen signal boards to allow automatic atmospheric pressure determination during air calibration. Calibration can be corrected for process salinity if removing the sensor from the process liquid is impractical. The analyzer can be calibrated against a standard instrument. For more information on the use of amperometric oxygen sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor (when O2 <1.00 ppm); 0.1% Input Range: 0 nA – 100 µA Temperature Compensation: Automatic or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec Response time: 6 seconds to 95% of final reading
in gas, ppm O2in gas. The analyzer fully
2
Recommended Sensor
Rosemount amperometric membrane and steam-sterilizable sensors listed above
1.11 Dissolved Ozone (Code -26 and -36)
The 56 is compatible with the 499AOZ sensor. The 56 fully compensates ozone readings for changes in membrane permeability caused by temperature changes. For more information concerning the use and operation of the amperometric ozone sensors, refer to the product data sheets.
Performance Specifications - Analyzer
Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0 nA – 100 µA Temperature Compensation: Automatic or manual (0-35 °C) Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading
Recommended Sensor
Rosemount 499A OZ ozone sensor.
Dissolved Ozone 499AOZ
Dissolved Oxygen 499ADO sensor with Variopol connection
12 Description and Specifications
sensors with Polysulfone body Variopol connection and cable connection
Page 23
Instruction Manual Section 1: Description and Specifications
LIQ-MAN-56 April 2017
1.12 Turbidity (Codes -27 and -37)
he 56 instrument is available in single and dual turbidity configurations for the Clarity II
T turbidimeter. It is intended for the determination of turbidity in filtered drinking water. The other components of the Clarity II turbidimeter – sensor(s), debubbler/measuring chamber(s), and cable for each sensor must be ordered separately or as a complete system with the 56.
The 56 turbidity instrument accepts inputs from both USEPA 180.1 and ISO 7027­compliant sensors. Four fully programmable relays with timers are included.
Note: the 56 Turbidity must be used with Clarity II sensor, sensor cable and debubbler.
Performance Specifications - Analyzer
Units: Turbidity (NTU, FTU, or FNU); total suspended solids (mg/L, ppm, or no units) Display resolution-turbidity: 4 digits; decimal point moves from x.xxx to xxx.x Display resolution-TSS: 4 digits; decimal point moves from x.xxx to xxxx Calibration methods: User-prepared standard, commercially prepared standard, or grab
sample. For total suspended solids user must provide a linear calibration equation.
Inputs: Choice of single or dual input, EPA 180.1 or ISO 7027 sensors. Field wiring terminals: Removable terminal blocks for sensor connection. Accuracy after calibration at 20.0 NTU:
0-1 NTU ±2% of reading or 0.015 NTU, whichever is greater. 0-20 NTU: ±2% of reading.
Description and Specifications 13
Page 24
Section 1: Description and Specifications Instruction Manual
April 2017 LIQ-MAN-56
1.13 Ordering Information
he 56 Analyzer offers single or dual sensor input with an unrestricted choice of dual
T measurement combinations. Measurements capabilities include pH/ORP, Resistivity/ Conductivity, % Concentration, Total Chlorine, Free Chlorine, Monochloramine, Dissolved Oxygen, Dissolved Ozone, Turbidity, Pulse Flow, Temperature, and 4-20mA input.
The device includes two isolated inputs, nine local languages, four 4-20mA current outputs, removable connectors for power and current outputs, and four solid plugs for closure of openings. HART digital communications is included at no additional charge. Profibus digital communications is optional.
56 Advanced Dual-Input Analyzer
Level 1 POWER
02 24 VDC with four alarm relays 03 85-265 VAC switching, 50/60 Hz with four alarm relays
Level 2 MEASUREMENT 1
20 Contacting Conductivity 21 Toroidal Conductivity 22 pH/ORP 23 24 Chlorine 25 Dissolved Oxygen 26
27
Flow/Current Input
Ozone Turbidity
Level 3 MEASUREMENT 2
30 Contacting Conductivity 31 Toroidal Conductivity 32 pH/ORP/ISE 33 34 Chlorine 35 Dissolved Oxygen 36
37 38 None
Level 4 COMMUNICATIONS
HT HART®digital communication DP Profibus DP digital communication
Flow/Current Input
Ozone Turbidity
14 Ordering Information
Page 25
Instruction Manual Section 2: Installation
LIQ-MAN-56 April 2017
Section 2.0 – Installation
2.1 Unpacking and Inspection
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately.
2.2 Installation
2.2.1 General Information
1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in areas of extreme temperatures is not recommended unless a sunshield is used. Make sure to note the Ambient temperature specifications in section 1. The analyzer cannot be operated in ambient (shaded) conditions greater than 60 °C.
2. Install the analyzer in an area where vibration and electromagnetic and radio frequency interference are minimized or absent.
3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer.
4. The analyzer is suitable for panel, pipe, or surface mounting. See Figures 2-1 and 2-2.
5. Install cable gland fittings and plugs as needed to properly seal the analyzer on all six en­closure openings. The USB port cover must be fully installed on the front cover to ensure proper analyzer sealing.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol appears, consult the manual for appropriate action.
Installation 15
Page 26
Section 2: Installation Instruction Manual
April 2017 LIQ-MAN-56
Fig. 2-1 Panel Mounting Installation dimensions
ILLIMETER
M
INCH
16 Installation
Page 27
Instruction Manual Section 2: Installation
LIQ-MAN-56 April 2017
Fig. 2-2 Pipe and Wall Mounting Installation dimensions
MILLIMETER
INCH
Shown with Mounting Kit PN 23820-00
Installation 17
Page 28
Section 2: Installation Instruction Manual
AREA
AREA
C
OR
OR
April 2017 LIQ-MAN-56
Fig. 2-3 FM Non-incendive field wiring installation for the 56-27-37 Analyzer
18 Installation
Page 29
Instruction Manual Section 2: Installation
OR
OR
U3
4
1
R56
R27
R
47
C
4
7
R28
C
4
3
C
4
4
Y1
C6
U16
+
U25
+
+
+
U23
4
1
1
U12
U2
+
+
C36
U4
U5
R
75
R73
R14
C13
C11
R16
R17
R67
C7
C12
R15
R66
R12
R13
C10
C24
C
3
R4
R5
R6
C4
C56
R68
R74
C53
J1
R52
C54
2
U7
U17
U18
R49
U8
U15
U13
U11
R35
U9
C25
U19
C30
C26
TB3
TB1
TB2
U14
C45
R70
U21
R71
C31
C32
C55
R42
R41
C42
R44
U10
R37
C35
C1
Z2
Z1
R2
R7
R8
R3
R9
R18
R53
C50
C46
C23
C22
C52
C21
C9
C2
C8
Z4
R10
Z3
R24
U22
R11
C15
C5
R22
R25
R20
R
21
Z5
R
29
Z6
Z7
R19
R23
R
40
R38C37
C28
C27
R58
C29
R59
R
34
C39
R36
R46
C41
R45
R63
R61
R60
R
62
C34
C33
R69
Q3
R72
C38
C48
R55
R54
R65
C51
C49
R57
AN SH
AMPEROMETRIC
RTD RET
RTD IN
RTD SH
ANODE
SNS
+5V
-4.5V
CA SH
CATH
ASSY 24203- REV
C22
R32
C47
2
1
J2
6
5
1
J4
4
R11
U9
U18
U15
U16
U13
U5
C49
C46
R77
R73
U3
U4
R3
R1
1
J3
2
U1
C
8
U2
R9
R4
R10
Z6
C16
R23
R15
U10
C6
R6
+
C19
+
C20
+
C28
C
24
+
C21
+
+
C55
R75
C41
R58
R57
C43
C
1
3
U6
R12
C52
R49
C37
C42
R67
C38
R70
R71
C54
R72
C56
R74
R68
C23
C35
R50
C25
C26
R28
R22
C10
R30
Z3
Z4
R35
C15
Z5
C12
C14
R46
U11
R41
C27 C36
R51
R47
J1
C30
R38
R63
C51 C29
R64
R44
U12
U8
R20
R31
R36
Y1
C32
C
3
3
C34
R8
R14
R21
R19
R16
R54
C44
R55
R56
C45
R2
R69
U7
C7
Q3
R29
Z1
C5
R17
Z
2
C3
C4
R5
R61
R65
R66
C18
R27
C1
R7
R13
U20
C48
C50
R62
R25
C2
U19
R76
R53
R45
C17
R60
R33
U22
U14
U26
D1
U21
C39
TB1
-5SHLD
RTD RTN
1
+5
SENSE
REF
GND
RTD IN
pH
SMART pH/ORP
ASSY 24312- REV
SHLD
R40
C10
1
Z4
R
6
4
3 7 R
6
6
R
C3
C2
Z
3
C
4
5
6
R
U
2
1
RTD RTN
CONTACTING
CONDUCTIVITY
ASSY 24355- REV
SHLDSHLD
RTD IN
SEN SEN
4CT4CT
SEN
SHLD
C
3
5
R40
R
5
2
C
3
8
C17
R
5
4
R
4
9
8
5
R
C44
C20
Q
3
R44
R41
C
1
3
2
4
R
R13
R
5
5
C12
5
3
R
9
2
C
C
24
R
3
7
R
6
0
C
3
4
R
36
R
7
5
R
7
0
C
2
5
7
4
C
C
40
R
2
0
U
1
4
C
42
C62
C
5
R
45
C
41
C54
C1
C
39
C45
R9
R6
R68
C49
R67
1
5
C
C
5
5
C57
R63
R
3
R
12
R
1
C
63
R
25
R18
R
30
Z1
C65
C64
R7 3
5
C
Z9
C23
C30
+
C52
U
1
7
R
4
8
U
16
1
R74
R19
C18
U19
U2
R69
R72
U1
C21
R
3
1
R
4
6
R14
R4
R
5
1
3
C
C28
+
R57
U
26
R53
7
2
C
+
0
5
C
+
C14
Z
8
R
2
R
6
2
R61
R8
U
1
0
R11
R76
R15
C16
R28
C56
R59
C22
01
R
C26
+
U
22
U
2
3
C
6
U11
TB1
2
3
C
C
3
6
C59
C
7
C48
Z2
U24
Y1
R38
U
2
5
U4
U15
R
7
1
R47
U
1
2
C43
U
2
8
9
2
U
C
58
C
46
J1
U
6
U
5
T
B2
1
6
C
8
C
9
C
11
C
15
C66
C
6
7
L1
Z
5
U
9
C60
+
U27
U
20
6
Z
Z7
C37
1
6
C
C41
R19
U19
J1
R38
C37
R40
R42
C28
C39
R34
R59
9
2
C
R58
C35
R37
R
7
4
R68
Q1
R64
R52
R21
C27
U10
R46
R36
R20
C5
R18
C6
R13
C14
R32
C7
C15
R26
R12
U3
C2
R24
R33
C17
C16
Z6
C3
R23
R22
R51
C13
C8
C24
R29
R50
Z5
C1
R30
R31
C10
Z2
Z1
C9
Z4
Z3
C12
C11
C21
U7
6
1
R
U12
C34
R62
R60
R61
R63
C33
U15
U13
R45
R41
C53
C43
C44
C42
Y1
U14
R44
R69
Q3
C54
C38
R72
C45
U21
C36
+
C55
C30
R70
R71
U11
C31
+
C26
R35
R11
U9
R25
R66
C4
R54
R55
C
48
R53
C46
0
5
C
U18
C49
C23
R57
U8
C52
U17
C32
U16
R65
C51
R73
R75
C22
R56
R49
R47
C47
J3
ASSY 24236- REV
TURBIDITY
1 3
+
+
+
+
C25
4
POWER SUPPLY
ALARM
WIRING (VAC)
(OPTIONAL)
ANALOG OUTPUT
(OPTIONAL)
SENSOR 1
ANY CSA APPROVED SENSOR
OR SIMPLE APPARATUS
UNCLASSIFIED AREA MODEL 56
CLASS 1 DIVISION 2, GROUPS ABCD 0-50 °C
CLASS II, III DIVISION 2 GROUPS EFG
9. THE EPA AND ISO CLARITY II TURBIDITY SENSORS ARE APPROVED FOR DIVISION 2
8.
THE 222, 225, 226 AND 228 TOROIDAL CONDUCTIVITY SENSORS ARE APPROVED BY CSA FOR USE WITH OPTIONS 21
AND 31.
1.3W. CONTACTING CONDUCTIVITY SENSORS AND pH, ORP, AMPEROMETRIC SENSORS WITHOUT
PREAMPS QUALIFY AS SIMPLE APPARATUS.
6
NON-INCENDIVE FIELD WIRING METHODS MAY BE USED FOR CONNECTING SENSORS TO THE
20/30,
21/31, 22/32,
24/34,
25/35,
OPTION BOARDS. SENSORS MUST BE CSA APPROVED AS NON-INCENDIVE FOR CLASS I,
Voc AND Isc LISTED IN TABLES 1A TO
1C AND THE Ci AND Li OF THE SENSOR AND INTERCONNECTED WIRING MUST BE
Ca AND La LISTED IN TABLES
1A TO 1C
OR BE CLASSIFIED AS SIMPLE APPARATUS
.
5
4 DURING INSTALLATION, LEAVE MAXIMUM AMOUNT OF JACKET INSULATION POSSIBLE ON N.I. FIELD WIRING WITHIN
INSTRUMENT ENCLOSURE. AFTER TERMINATION, WRAP N.I. FIELD WIRING WITHIN ENCLOSURE WITH MYLAR TAPE, TO
ENSURE ADEQUATE DOUBLE INSULATION REMAINS.
3
GROUND CONNECTION MAY BE MADE IN HAZARDOUS AREA.
2. SEAL REQUIRED AT EACH CONDUIT ENTRANCE.
1. INSTALLATION MUST CONFORM TO THE CEC.
SENSOR 2
ANY CSA APPROVED SENSOR OR
SIMPLE APPARATUS
UNCLASSIFIED AREA
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
METAL CONDUIT
SENSOR CABLE
IS SHIELDED
SENSOR CABLE
IS SHIELDED
METAL CONDUIT
3
SENSOR 1
ANY CSA APPROVED SENSOR OR
SIMPLE APPARATUS
SENSOR 2
ANY CSA APPROVED SENSOR OR
SIMPLE APPARATUS
WARNING
IF THE SENSOR TIP HAS EXPOSED ELECTRODES,
THEN IT MUST ONLY BE USED IN A NON-FLAMMABLE LIQUID PROCESS
TABLE 1A
ENTITY PARAMETERS FOR
OPTIONS 24/34, 25/35, 26/36 (AMPEROMETRIC SENSOR BOARD)
TABLE 1B
ENTITY PARAMETERS FOR
OPTION 22/32 (pH / ORP SENSOR BOARD)
TABLE 1C
ENTITY PARAMETERS FOR OPTION 20/30
(CONTACTING CONDUCTIVITY BOARD)
OPTION 21/31 (TOROIDAL CONDUCTIVITY SENSOR
BOARD) MAY ONLY BE USED WITH 200 SERIES SENSORS.
4 5 6
4 5 6
NON-INCENDIVE FIELD WIRING CONNECTIONS
FOR CLASS 1, DIVISION 2, GROUPS ABCD
OPTION 24/34, 25/35, 26/36 (CHLORINE,
DISSOLVED OXYGEN & OZONE SENSOR BOARD)
OPTION 22/32 (pH/ORP SENSOR BOARD)
OPTION 20/30 (CONTACTING
CONDUCTIVITY SENSOR BOARD)
MAY ONLY BE USED WITH CLARITY II SENSORS.
OUTPUT
PARAMETERS
AMPEROMETRIC
CONNECTORS
TB1, TB2, TB3
Voc, Vo
9.624 V
Isc, Io
250.4 mW
Ca
La
OUTPUT
PARAMETERS
pH TB1
CONNECTOR
Voc, Vo
9.624 V
Isc, Io
115 mA
Ca
La
6 mH
OUTPUT
PARAMETERS
CONDUCTIVITY
CONNECTORS
TB1, TB2
Voc, Vo
6.633 V
Isc, Io
30.45 mA
50.5 mW
Ca
La
85 mH
NOTES: UNLESS OTHERWISE SPECIFIED
SCALE: 1:1
WEIGHT:
SIZE
D
DWG NO
SHEET 1 OF 1
D
1400668
REV
THIS DOCUMENT IS CERTIFIED BY
CSA (REVISION C)
REVISIONS ARE NOT PERMITTED
WITHOUT CSA APPROVAL
LIQ-MAN-56 April 2017
Fig. 2-4 CSA Non-incendive field wiring installation
Installation 19
Page 30
Section 2: Installation Instruction Manual
April 2017 LIQ-MAN-56
20 Installation
Page 31
Instruction Manual Section 3: Wiring
LIQ-MAN-56 April 2017
Section 3.0 Wiring
3.1 General
The 56 is easy to wire. It includes removable connectors and slide-out signal input boards The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations.
3.1.1 Removable connectors and signal input boards
The 56 uses removable signal input boards and communication boards for ease of wiring and installation. Each of the signal input boards can be partially or completely removed from the enclosure for wiring. The 56 has three slots for placement of up to two signal input boards and one communication board
Slot 1-Left Slot 2 – Center Slot 3 – Right
Profi board Signal Board 1 Signal Board 2
3.1.2 Signal input boards
Slots 2 and 3 are for signal input measurement boards. Wire the sensor leads to the measurement board following the lead locations marked on the board. After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland. Tighten the cable gland nut to secure the cable and ensure a sealed enclosure.
NOTE:
For the purpose of replacing factory-installed signal input boards, Rosemount®Analytical Inc. is the sole supplier.
3.1.3 Digital communications
HART®digital communications is standard on 56. HART®versions 5 and 7 are available on the 56 and can be switched using the local keypad. A Profibus DP communication board is available as options for 56 communication with a host. HART communications supports Bell 202 digital communications over an analog 4-20 mA current output. Profibus DP is an open communications protocol which operates over a dedicated digital line to the host.
3.1.4 Alarm relays
Four alarm relays are supplied with the switching power supply (85 to 264 VAC, -03 order code) and the 24 VDC power supply (20-30 VDC, -02 order code). All relays can be used for process measurement(s) or temperature. Any relay can be configured as a fault alarm instead of a process alarm. Each relay can be configured independently and each can be programmed as an interval timer, typically used to activate pumps or control valves. As process alarms, alarm logic (high or low activation or USP*) and deadband are user­programmable. Customer-defined failsafe operation is supported as a programmable menu function to allow all relays to be energized or not-energized as a default condition upon powering the analyzer. The USP* alarm can be programmed to activate when the conductivity is within a user-selectable percentage of the limit. USP alarming is available only when a contacting conductivity measurement board is installed.
Wiring 21
Page 32
Section 3: Wiring Instruction Manual
April 2017 LIQ-MAN-56
3.2 Preparing Conduit Openings
here are six conduit openings in all configurations of 56 analyzer. (Note that four of the
T openings will be fitted with plugs upon shipment.)
Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with Type 4X or IP66 conduit plugs.
NOTE:
se watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the
U conduit before attaching the fitting to the analyzer.
3.3 Preparing Sensor Cable
The 56 is intended for use with all Rosemount sensors. Refer to the sensor installation instructions for details on preparing sensor cables.
3.4 Power, Output, and Sensor Connections
3.4.1 Power wiring
Two Power Supplies are offered for the 56:
a. 24 VDC (20 – 30V) Power Supply (-02 order code)
b. 85 – 265 VAC Switching Power Supply (-03 order code)
AC mains leads and 24 VDC leads are wired to the Power Supply board which is mounted vertically on the left side of the main enclosure cavity. Each lead location is clearly marked on the Power Supply board. Wire the power leads to the Power Supply board using the lead markings on the board.
The grounding plate is connected to the earth terminal of the -03 order code (85-265 VAC) power supply. The green colored screws on the grounding plate are intended for connection to some sensors to minimize radio frequency interference. The green screws are not intended to be used for safety purposes.
3.4.2 Current output wiring
All instruments are shipped with four 4-20 mA current outputs. Wiring locations for the outputs are on the Main board which is mounted on the hinged door of the instrument. Wire the output leads to the correct position on the Main board using the lead markings (+/positive, -/negative) on the board. Male mating connectors are provided with each unit.
3.4.3 Alarm relay wiring
Four alarm relays are supplied with the switching power supply (85 to 265 VAC, -03 order code) and the 24 VDC power supply (20-30 VDC, -02 order code). Wire the relay leads on each of the independent relays to the correct position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or Com/Common) on the board.
22 Wiring
Page 33
Instruction Manual Section 3: Wiring
LIQ-MAN-56 April 2017
3.4.4 Sensor wiring to signal boards
ire the correct sensor leads to the measurement board using the lead locations marked
W directly on the board. After wiring the sensor leads to the signal board, carefully slide the wired board fully into the enclosure slot and take up the excess sensor cable through the cable gland.
For best EMI/RFI protection use shielded output signal cable enclosed in an earth-grounded metal conduit. Connect the shield to earth ground. AC wiring should be 14 gauge or greater. Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as the disconnecting device for the analyzer.
Keep sensor and output signal wiring separate from power wiring. Do not run sensor and power wiring in the same conduit or close together in a cable tray.
WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable national or local codes.
CAUTION: This symbol identifies a risk of electrical shock.
CAUTION: This symbol identifies a potential hazard. When this symbol ap­pears, consult the manual for appropriate action.
Wiring 23
Page 34
Section 3: Wiring Instruction Manual
April 2017 LIQ-MAN-56
FIGURE 3-1 Power Wiring for 56 24 VDC Power Supply (-02 order code) PN 24365-00
FIGURE 3-2 Power Wiring for 56 85-264 VAC Power Supply (-03 order code) PN 24358-00
24 Wiring
Page 35
Instruction Manual Section 3: Wiring
LIQ-MAN-56 April 2017
FIGURE 3-3 Output Wiring for 56 Main PCB PN 24308-00
Wiring 25
Page 36
Section 3: Wiring Instruction Manual
April 2017 LIQ-MAN-56
Figure 3-4 56 Recommended Wire Entry Points
Figure 3-5 56 Recommended Wire Entry and THUM Adaptor Installation
26 Wiring
Page 37
Instruction Manual Section 3: Wiring
LIQ-MAN-56 April 2017
Figure 3-6 Contacting Conductivity signal board and Sensor Cable Leads
Figure 3-7 Toroidal Conductivity Signal board and Sensor Cable Leads
Wiring 27
Page 38
Section 3: Wiring Instruction Manual
April 2017 LIQ-MAN-56
Figure 3-8 pH/ORP/ISE Signal Board and Sensor Cable Leads
Figure 3-9 Amperometric signal (Chlorine, Oxygen, Ozone) board and Sensor cable leads
28 Wiring
Page 39
Instruction Manual Section 3: Wiring
LIQ-MAN-56 April 2017
Figure 3-10 Turbidity Signal Board with Plug-in Sensor Connection
Figure 3-11 Flow/Current Input Signal Board and Sensor Cable Leads
Wiring 29
Page 40
Section 3: Wiring Instruction Manual
April 2017 LIQ-MAN-56
30 Wiring
Page 41
Instruction Manual Section 4: Display and Operation
LIQ-MAN-56 April 2017
Section 4.0 Display and Operation
4.1 User Interface
The 56 has a large display which shows two live measurement readouts in large digits and up to six additional process variables or diagnostic parameters concurrently. The display is back-lit and the format can be customized to meet user requirements. The ENTER/MENU key allows access to Calibration, Hold (of current outputs), Programming, Display, Data and HART®functions. In addition, a dedicated INFO key is available to provide access to useful diagnostic and instrument information regarding installed sensor(s) and any problematic conditions. The display flashes a red banner to indicate a Fault condition and a yellow banner for a Warning condition. Help screens are displayed for fault and warning conditions to guide the user in troubleshooting. During calibration and programming, key presses guide the user step-by-step through procedures. An alpha-numeric keypad similar to a cell phone keypad is available to allow the user to enter data during programming and calibration or lengthy tags to describe process points, sensors, or instrumentation.
4.2 Instrument Keypad
There are three Function keys, four Navigation keys and an alpha-numeric keypad on the instrument keypad.
Function keys
The ENTER/MENU key is used to access menus for programming and calibrating the instrument as well as retrieving stored data. Eight top-level menu items appear when pressing the ENTER/MENU key from the main display of live readings:
Calibrate: calibrate attached sensors and analog outputs.
Program: Program outputs, relays, measurement, temperature, and security codes.
Hold: Suspend current outputs.
Display Setup: Program graphic trend display, brightness, main display format, tags,
language, and warnings.
Data storage and retrieval: Enable data and event storage, download data, and vie
events.
HART or Profibus: Program HART and Profibus communication parameters.
Time and Date: Set and view real-time clock settings.
Reset:Reset all instrument settings, calibration settings or current outputs to factory defaults.
Calibrate
Program HART
Hold Time and Date
Display setup Reset
Software upgrade
Data storage and retrieval
Display and Operation 31
Page 42
Section 4: Display and Operation Instruction Manual
April 2017 LIQ-MAN-56
nd
a
g
n
i
m
m
a
r
g
o
r
p
e
l
b
a
n
e
r
o
s
n
o
ecti
l
e
s
r
e
nt
e
o
t
ed
s
u
o
s
l
a
s
i
ey
k
NU
E
M
/
R
E
NT
E
e
h
T
.
s
p
e
t
s
n
o
ati
r
b
i
l
a
c
.
l
ve
e
l
nu
e
m
s
u
o
i
v
e
r
p
e
th
o
t
s
n
r
tu
e
r
ey
k
T
EXI
e
h
T
d
n
g a
n
i
m
m
a
ogr
g pr
n
ri
s du
ion
t
a
n
a
pl
x
d e
n
s a
ion
t
c
ru
st
n
d i
e
il
a
s det
vide
o
pr
y
ke
O
F
N
I
e
Th
t
a
h
s t
g
n
i
rn
a
d w
n
s a
t
l
u
a
f
l
l
s for a
ip
g t
n
i
t
oo
sh
e
bl
u
o
r
s t
vide
o
so pr
l
a
t
I
s.
e
r
du
e
oc
g pr
n
i
t
a
br
i
l
ca
s.
s
e
oc
pr
n
i
ion
t
a
us oper
o
u
n
i
t
on
or c
ion
t
a
br
i
l
g ca
n
ri
du
r
u
occ
gust
u
a
ys
Ke
n
o
ti
viga
a
N
e
v
i
t
ui
t
n
i
an
n
i
e
t
opera
y
e
k
U
MEN
/
ER
T
EN
he
t
d
aroun
d
ge
arran
s
y
e
k
on
i
gat
vi
a
N
our
f
he
T
se
he
t
,
on
i
t
opera
u
en
m
g
n
uri
D
d.
boar
y
ke
er
t
pu
om
c
a
on
s
y
ke
on
i
t
a
g
i
v
a
n
he
t
o
t
lar
i
m
g en
p
l
e
Ni
On
On
On
e
r
ry
t
h
alph
e ke
n
e ke
e ke
e ke
si
use
a-nume
t
e
v
o
d to m
s use
key i
t
ef
l
e
h
t
,
i
r
a-numeric key
a-nu
ph
l
e a
r
ys a
cated “1
dedi
s a
i
y
cated “0
dedi
s a
i
y
cated “.” (deci
dedi
s a
i
y
e
h
c
igh
h
d to del
Ke
ad
p
eri
m
y
h
” ke
” ke
l
a
c
igh
p
s 1
y
y
te ete
ad
2
ma
d sc
en
ke
l
ys a
poi
en
e
r
ri
t
nt) ke
s du
e
s ou
sel
e ri
tl
c
n
n
i
y
to a
ion
t
c
g a
ed bel
t
e
v
i
ow.
g
n
ri
Du
.
tem
i
en
e
scr
t
en
c
dja
er a
h
t
o
n
.
ry
t
en
ter
c
a
r
a
h
c
eric
m
u
-n
a
ph
l
a
manner keys a
a
t
A
Th
c keypa
meri
u
n
pha-
l
a
The
ipl
t
l
ming a
gle
ave mu
nd ca
characte
to
br
i
l
nu pr
i
t
ogr
m
m
e
eric keys h
m
a
s to tog
4.3 Main Display
two pri
e or
ys on
pla
s
di
6
e 5
Th
ng ba
i
rn
nd wa
t a
l
s, fau
e
u
l
va
Process Measurements:
s a
e
bl
a
ri
s va
s
oce
Two pr
atu
per
s tem
s
oce
nd pr
a
a shows the Sensor 1 process reading. The Center display area shows the
Upper displ Senso assigned to d
For single inp the Cen
Secondary Values:
Up to six s
reen. All six s
sc parameter available. Possible secondary values include:
ay are
r 2 proc
er displ
t
econdar
ess read
fferen
i
ut co
ing. For dual conductivity, the Upper and Center display areas can be
t process variables as follows:
gurations, the Upper display area shows the live process variable and
fi
n
y area can be assigned to Temperature or blank.
a
y values are shown in six display quadrants at the bottom half of the
econdary value positions can be programmed by the user to any display
e ch
ation
e di
r
e i
r
d
rs t
n
s di
pe
o
a
step
ha
m
er, a
n
play
s
s
a
r
t
ry m
a
play
tes t
ra cter
s. C
are
a
l
ed if
he
s th
a
h
ail
v
a
su
a
e
elay f
r
rm
two sign
on
ed if
at r
a
ame
s
ca cter ab
em
r
e sign
as e
be en
n
sel
n e
o
le
t va
en
lags, a
bo
l
a
a
ect
l
bo
r
t
n
ed for
ter
ion
ach k
s, up to si
e
u
l
nd a
rds a
a
a
ies o
s a
ey.
digi
rd i
n a mob
tag en
ade by pr
e m
r
c
l
ta
sta
n
e i
r
sta
n
s i
x sec
om
l ed w
l
l
ile
e
ri
t
onda
mu
ed. On
l
p
s or
ic
n
th
i
ho
s
e
ry m
ation
on
ne.
du
ng t
si
e pr
The
ng
ri
h
a
e s ic
oce
e sen
nine
e key mu
en
em
r
su
on.
ri
s va
s
sor. Th
alp
t
bl
a
e
tipl
l
e
ha
-
e
4.4 Menu System
The 56 menu system is similar to a computer. Pressing the ENTER/MENU key at any time opens the top-level menu including Calibration, Hold, Programming, Display, Data and HART
32 Display and Operation
Page 43
Instruction Manual Section 4: Display and Operation
LIQ-MAN-56 April 2017
functions. To find a menu item, use the directional Navigation keys to highlight a menu item. Press ENTER/MENU and simply direct the cursor to the desired operation and follow
he screen prompts. Pressing the BACK screen control available on some menu screens will
t revert to the immediate previous menu screen. Pressing the EXIT key will return to the previous hierarchical menu level.
Fault and Warning banner:
If the analyzer detects a problem with itself or the sensor the word Fault banner (red) and/or Warning banner (yellow) will appear at the bottom of the main display. A fault requires immediate attention. A warning indicates a problematic condition or an impending failure. For detailed troubleshooting assistance, press INFO.
4.5 USB Data Port
The 56 menu system is similar to a computer. A USB 2.0 data port is accessible on the front panel of the 56 instrument. The USB data port can be used for download of measurement data and events using a USB memory device. It can also be used to download and upload complete analyzer configurations to copy all programmed settings to another 56 analyzer. NOTE: only 56 units which display the “Transfer Configurations” tab under the Data Storage and Retrieval menu are capable of downloading and uploading analyzer configurations.
The USB data port is easily accessed by inserting a coin in the vertical slot of the cover and rotating counterclockwise one quarter turn to remove the cover and Type 4 seal.
Caution: not all USB memory devices will physically fit into the 56 data port. After removing the USB cover and seal, make sure that the USB memory device can be easily and fully inserted into the USB data port without any mechanical conflict with the USB data port flange. The USB communications port is protected by a Type 4-rated seal and cover. Do not remove the cover during cleaning of the analyzer housing. Never remove the USB port cover when the instrument is operated in a hazardous rated area.
NOTE: the data logger and event logger are disabled by default setting upon initial startup from the factory. Always enable the data logger and event logger under the “Data Storage and Retrieval” menu to initiate internal recording of process data and event data.
4.6
56 Data Logger and Event Logger Download Procedure
4.6.1 Description
The 56 supports download of stored data logger and event logger data at the device. The download process is performed using a USB 2.0 flash drive memory device inserted into the USB data port on the front panel of the 56 analyzer. The data can be uploaded to a PC for viewing in preformatted EXCEL tables.
4.6.2 56 Data Logger Download Procedure
1. From the main menu, select Data storage and retrieval.
2. Select the Download tab.
3. Select “Download measurement” data.
Display and Operation 33
Page 44
Section 4: Display and Operation Instruction Manual
April 2017 LIQ-MAN-56
4. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to
ccess the data port.
a
. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB inser-
5
tion connector is keyed.
6. The Download screen will report Earliest data available and Latest data available that can be downloaded from the internal data logger file.
7. With a USB 2.0 flash drive properly inserted, choice Selected Range (default) or All Data.
8. If Selected Range is chosen, define the Start Date and End Date in the screen fields. Note that the Start Date and End Date reported will default to the current data that is recog­nized by the analyzer.
9. If All Data is selected, all stored date (up to 30 days) will be downloaded.
10. Select START and press ENTER.
11. A Data download information and instruction screen is displayed while data logger files are being downloaded. The download process will be completed in a few minutes.
12. When the data logger download is complete, a confirmation screen is display with the reported range of dates that were downloaded to the USB flash drive.
13. Carefully remove the USB 2.0 flash drive from the USB port.
14. Insert the NEMA cap into the 56 USB front display opening.
15. You august exit the Download completed screen by selecting BACK.
16. To view data logger files, insert the USB flash drive into a computer. On the designated drive associated with the USB flash drive, individual data logger files for each day can be opened as EXCEL formatted files in the root directory of the USB drive. Note that the data codes are assigned file names based on the analyzer’s recognized dates.
4.6.3 56 Event Logger Download Procedure
1. From the main menu, select Data storage and retrieval.
2. Select the Download tab.
3. Select “Download events”.
4. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to access the data port.
5. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB inser­tion connector is keyed.
6. A Download events information screen is briefly displayed while events files are being downloaded. The download process august require a few minutes to complete.
7. When the events download is complete, a confirmation screen appears.
8. Carefully remove the USB 2.0 flash drive from the USB port.
9. Insert the Type 4 cap into the 56 USB front display opening.
10. You august exit the Download completed screen by selecting BACK.
11. To view event logger files, insert the USB flash drive into a computer. On the designated drive associated with the USB device, the single event logger file can be opened as an EXCEL formatted file in the root directory of the USB drive. Note that the data codes are assigned file names based on the analyzer’s recognized dates.
34 Display and Operation
Page 45
Instruction Manual Section 4: Display and Operation
LIQ-MAN-56 April 2017
4.7 Software Upgrade
4.7.1 Description
All 56 advanced analyzers with serial number J12- or later (October 2012) allow software upgrades using the device’s USB data port. To download and install software upgrades, refer to Liquid Software Download.
4.7.2 Software upgrade checklist
Before upgrading, please confirm:
Serial Number date codes of the 56 analyzer is J12 (October 2012) or later.
Confirm that Safe Area requirements are met before installing software
USB 2.0 flash drive memory device is available
4.7.3 Software upgrade procedure
1. Download the 56 software upgrade file to a computer. Access software HERE. Save the file to the main root directory of a USB flash drive memory stick. Only one 56 software upgrade file should be present at the root directory.
2. Save all user settings before upgrading the software by copying the analyzer’s configura­tion to a flash drive. Access this procedure at the local 56 device menu location: MENU/Data storage and retrieval/Transfer Configuration.
3. Download the Data Logger measurement data and the Events before upgrading the soft­ware. Access these procedures at 56 menu location: MENU/Data storage and retrieval/Download/Download measurement data and at /Download/Download Events.
4. With the 56 analyzer powered up and the front enclosure panel completely closed, re­move the Type 4 cap from the front display by inserting a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap and seal to access the data port.
5. Carefully insert a USB 2.0 flash drive into the USB port on the 56. The USB insertion connector is keyed. Note that some USB devices will not fit into the USB port due to mechanical restrictions.
6. Press the ENTER/MENU key. Select Software upgrade in the main menu.
7. Select NEXT. Press ENTER/MENU to start software upgrade. The current software ver­sion and the new software version are reported on the screen.
8. Select NEXT. Press ENTER/MENU. “Software upgrade in process” and a progress bar will appear. The process august require up to 5 minutes to complete.
9. The Time and Date screen will appear. Enter local real time clock and date information. Press ENTER/MENU.
10. The main screen will appear reporting live process values. Programming of settings and calibration august be required.
4.8 Configuration Transfer
4.8.1 Description
All 56 unit with serial number J12- or later (October 2012) will support the function of Configuration Transfer from one 56 analyzer to another, or multiple 56 analyzers. The
Display and Operation 35
Page 46
Section 4: Display and Operation Instruction Manual
April 2017 LIQ-MAN-56
transfer is done by using a version 2.0 USB flash drive (USB memory stick) for downloading the existing configuration and uploading to another instrument.
4.8.2 Transfer Configuration Process User Notes
1. Configuration Transfer can only be performed between instruments of identical configu­ration. The analyzers must have the same signal boards installed.
2. Several text files downloaded to the flash drive can be saved to a PC and used later.
3. Only one set of configuration files will be stored on the flash drive. The files that are downloaded will over-write any existing files stored in the root directory of the flash drive.
4.8.3 Transfer Configuration Procedure - User Notes
1. Confirm all user settings before transferring configuration.
2. From the main menu, select Data storage and retrieval.
3. Select the Transfer Configuration tab
4. Select “Copy analyzer configuration to the flash drive”.
5. With the 56 analyzer powered, remove the NEMA cap from the front display by inserting a coin into the cap’s vertical slot and rotating counterclockwise. Remove the USB cap to access the data port.
6. Carefully insert a USB 2.0 flash drive into the USB port on the 56. Note that the USB inser­tion connector is keyed.
7. Select Copy data. Press ENTER.
8. An information screen appears warning users that any configuration files that exist in the root directly of the flash drive will be over-written upon configuration transfer. If No is se­lected (default setting), existing configuration files will not be overwritten. Select Yes to transfer the configuration file to the flash drive.
9. The configuration file will be transferred (downloaded) in about 20 seconds. A screen will report that the configuration file has been transfused.
10. Carefully remove the USB 2.0 flash drive from the USB port.
11. Insert the NEMA cap into the 56 USB front display opening.
12. Follow steps 2-3 above on the instrument that will receive the copied configuration.
13. On the 56 analyzer instrument that will receive the copied configuration, remove the Type 4 cap from the front display.
14. Carefully insert the USB 2.0 flash drive containing the configuration file into the USB port.
15. Select “Copy configuration from the flash drive to the analyzer”. Press ENTER.
16. You have two options for configuration transfer. Select one of the following and press ENTER:
a. Copy configuration data only to the analyzer.
b. Copy configuration and calibration data to the analyzer.
17. Select Copy data. Press ENTER.
18. The configuration file will be transferred (uploaded) in about 20 seconds. A screen will report that the configuration file has been transfused.
19. Carefully remove the USB 2.0 flash drive from the USB port.
20. Insert the NEMA cap into the 56 USB front display opening.
36 Display and Operation
Page 47
Instruction Manual Section 5: Programming Basics
LIQ-MAN-56 April 2017
Section 5.0. Programming the Analyzer - Basics
5.1 General
Typical programming steps include the following listed procedures. Each of these programming functions are easily and quickly accomplished using the intuitive menu systems.
Changing the measurement type, measurement units and temperature units.
Choose temperature units and manual or automatic temperature compensation mode
Configure and assign values to the current outputs
Set a security code for two levels of security access
Accessing menu functions using a security code
Enabling and disabling Hold mode for current outputs
Choosing the frequency of the AC power (needed for optimum noise rejection)
Resetting all factory defaults, calibration data only, or current output settings only
5.2 Changing Startup Settings
To change the measurement type, measurement units, or temperature units that were initially entered in Quick Start, choose the Reset function or access the Program menus for sensor 1 or sensor 2. The following choices for specific measurement type, measurement units are available for each sensor measurement board.
TABLE 5-1. Measurements and Measurement Units
Signal board Available measurements
pH/ORP (-22, -32)
Contacting conductivity
(-20, -30)
Toroidal conductivity
(-21, -31)
Chlorine
(-24, -34)
Oxygen
(-25, -35)
Ozone (-26, -36) Ozone Temperature (all)
pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
Conductivity, Resistivity, TDS, Salinity, NaOH (0-12%), HCl (0-15%), Low H2SO4, High H2SO4, NaCl (0-20%), Custom Curve
Conductivity, Resistivity, TDS, Salinity, NaOH (0-12%), HCl (0-15%), Low H2SO4, High H2SO4, NaCl (0-20%),
Custom Curve
Free Chlorine, pH Independ. Free Cl, Total Chlorine, Monochloramine
Oxygen (ppm), Trace Oxygen (ppb), Percent Oxygen in gas, Salinity
Temperature °C / °F
Measurements units:
pH, mV (ORP) %, ppm, mg/L, ppb,
S/cm, mS/cm, S/cm
µ
% (concentration)
S/cm, mS/cm, S/cm
µ
% (concentration)
ppm, mg/L
ppm, mg/L, ppb, µg/L % Sat, Partial Pressure, % Oxygen In Gas, ppm Oxygen In Gas
ppm, mg/L, ppb, µg/L
g/L, (ISE)
µ
Programming Basics 37
Page 48
Section 5: Programming Basics Instruction Manual
April 2017 LIQ-MAN-56
To change the measurement type, measurement units, or temperature
nits, access the Reset screens by
u pressing ENTER/MENU from the main screen.
To change the measurement type, measurement units, or temperature units, access the Program screens by pressing ENTER/MENU from the main screen.
5.3 Programming Temperature
Most liquid analytical measurements (except ORP) require temperature compensation. The 56 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature correction is off, the 56 uses the temperature entered by the user in all temperature correction calculations.
To select automatic or manual temp compensation, set the manual reference temperature, and to program temperature units as °C or °F, access the Temperature screens by pressing ENTER/MENU from the main screen.
5.4 Configuring and Ranging the Current Outputs
The 56 accepts inputs from two sensors and has four analog current outputs. Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs. This section provides a guide for configuring and ranging the outputs. ALWAYS CONFIGURE THE OUTPUTS FIRST.
To configure the outputs, access the Outputs screen by pressing ENTER/MENU from the main screen.
38 Programming Basics
Page 49
Instruction Manual Section 5: Programming Basics
LIQ-MAN-56 April 2017
5.5 Setting a Security Code
he security codes prevent accidental or unwanted changes to program settings, displays,
T and calibration. The 56 has two levels of security code to control access and use of the instrument to different types of users. The two levels of security are:
All: This is the Supervisory security
level. It allows access to all menu func­tions, including Programming, Cali­bration, Hold and Display.
Calibration/Hold: This is the operator
or technician level menu. It allows ac­cess to only calibration and Hold of the current outputs.
To set security codes, access the Security screen by pressing ENTER/MENU from the main screen. Upon entry of the proper code, the following security screen will appear.
5.6 Security Access
When entering the correct access code for the Calibration/Hold security level, the Calibration and Hold menus are accessible. This allows operators or technicians to perform routine maintenance. This security level does not allow access to the Program or Display menus. When entering the correct access code for All security level, the user has access to all menu functions, including Programming, Calibration, Hold and Display.
The 56 menus use a security code, access the Security screen by pressing ENTER/MENU from the main screen. If a security code is currently programmed, the follow security screen will appear. Enter the code.
1. If a security code has been
programmed, selecting the Calibrate, Hold, Program or Display top menu items causes the security access screen to appear.
2. Enter the three-digit security code for the appropriate security level.
3. If the entry is correct, the appropriate menu screen appears. If the entry is incorrect,
the Invalid Code screen appears. The Enter Security Code screen reappears after 2 seconds.
Programming Basics 39
Page 50
Section 5: Programming Basics Instruction Manual
April 2017 LIQ-MAN-56
5.7 Using Hold
he analyzer output is always proportional to measured value. To prevent improper operation
T of systems or pumps that are controlled directly by the current output, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the analyzer from hold once calibration is complete. During hold, all outputs remain at the last value. Once in
hold, all current outputs remain on Hold indefinitely.
To hold the outputs and alarm relays, access the Hold screen by pressing ENTER/MENU from the main screen.
5.8 Resetting Factory Default Settings
This section describes how to restore factory calibration and default values. The process also clears all fault messages and returns the display to the first Quick Start screen. The 56 offers three options for resetting factory defaults.
a. reset all settings to factory defaults
b. reset sensor calibration data only
c. reset analog output settings only
To reset to factory defaults, reset calibration data only or reset analog outputs only, access the Reset screen by pressing ENTER/MENU from the main screen.
5.9 Programming Alarm Relays
The 56 24 VDC (02 order code) and the AC switching power supply (03 order code) provide four alarm relays for process measurement or temperature. Each alarm can be configured as a fault alarm instead of a process alarm. Also, each relay can be programmed independently and each can be programmed as an interval timer or one of four advanced timer functions. This section describes how to configure alarm relays, simulate relay activation, and synchronize timers for the four alarm relays.
This section provides details to program the following alarm features. To program the alarm relays, access the Program screen by pressing ENTER/MENU from the main screen and then select the Relay tab and the Configure relay control.
40 Programming Basics
Page 51
Instruction Manual Section 5: Programming Basics
LIQ-MAN-56 April 2017
The following relay functions can be programmed to any relay from the Configure Relay screen:
. assign a relay
1
. define a relay function
2
3. assign a Measurement
4. set relay logic
5. enter setpoints
6. set deadband
7. set normal state
8. set USP Safety level (contacting conductivity)
To program these relay functions, access the Configure Relay screen by pressing ENTER/MENU from the main Relay programming screen.
1. To assign a relay, highlight the desired Relay 1-4 and press ENTER/MENU.
2. To define a relay function, select from Setpoint, Interval Timer, TPC, Bleed and Feed, Water Meter, Delay timer, Date and Time, Fault or None and press ENTER/MENU.
3. To assign a measurement to a specific relay, select the desired measurement or temperature input and press ENTER/MENU.
4. To set relay logic to activate alarms at a High reading or a Low reading, select high or low and press ENTER/MENU.
5. To enter setpoints for relays, enter the desired value for the process measurement or temperature at which to activate an alarm event and press ENTER/MENU.
6. To set deadband as a measurement value, enter the change in the process value needed after the relay deactivates to return to normal (and thereby preventing repeated alarm activation) and press ENTER/MENU.
7. To set the Normal alarm condition, select Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
8. To set USP Safety, enter the percentage below the limit at which to activate the alarm and press ENTER/MENU. NOTE: USP Safety only appears if a contacting conductivity board is installed.
Programming Basics 41
Page 52
Section 5: Programming Basics Instruction Manual
April 2017 LIQ-MAN-56
This section provides details to simulate relay action. To simulate relays, access
he Program screen by pressing
t ENTER/MENU from the main screen and then select the Relay tab.
To simulate alarm relay conditions, access the Simulate Relay Action screen by pressing ENTER/MENU from the main Relay programming screen.
Alarm relays can be manually set for the purposes of checking devices such as valves or pumps. Under the Alarms Settings menu, this screen will appear to allow manual forced activation of the alarm relays. Select the desired alarm condition to simulate.
42 Programming Basics
Page 53
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
Section 6.0 Programming - Measurements
6.1 Programming Measurements – Introduction
The 56 automatically recognizes each installed measurement board upon first power-up and each time the analyzer is powered. Completion of Quick Start screens upon first power up enable measurements, but additional steps august be required to program the analyzer for the desired measurement application. This section covers the following programming and configuration functions:
1. Selecting measurement type or sensor type (all sections)
2. Identifying the preamp location (pH-see Sec. 6.2)
3. Enabling manual temperature correction and entering a reference temperature (all sections)
4. Enabling sample temperature correction and entering temperature correction slope (selected sections)
5. Defining measurement display resolution (pH and amperometric)
6. Defining measurement display units (all sections)
7. Adjusting the input filter to control display and output reading variability or noise (all sections)
8. Selecting a measurement range (conductivity – see Sections 6.4, 6.5)
9. Entering a cell constant for a contacting or toroidal sensor (see Sec’s 6.4, 6.5)
10. Entering a temperature element/RTD offset or temperature slope (conductivity­see Sec’s 6.4)
11. Creating an application-specific concentration curve (conductivity-see Sec’s 6.4, 6.5)
12. Enabling automatic pH correction for free chlorine measurement (Sec. 6.6.1)
To fully configure the analyzer for each installed measurement board, you august use the following:
1. Reset Analyzer function to reset factory defaults and configure the measurement board to the desired measurement. Follow the Reset Analyzer menu (Fig. 5-5) to reconfigure the analyzer to display new measurements or measurement units.
2. Program menus to adjust any of the programmable configuration items. Use the follow­ing configuration and programming guidelines for the applicable measurement.
6.2 pH Measurement Programming
The section describes how to configure the 56 analyzer for pH measurements. The following programming and configuration functions are covered.
1. Measurement type: pH Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
2. Preamp location: Analyzer Identify preamp location
3. Filter: 4 sec Override the default input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high reference impedance
5. Sensor wiring scheme: Normal or Reference to Ground
6. Resolution: 0.01pH Select 0.01pH or 0.1pH for pH display resolution
Programming Measurements 43
Page 54
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
7. Enabling pH sensor diagnostics
To configure the pH measurement board, access the Program screen by pressing ENTER/MENU from the main screen and
hen select the Measurement tab.
t
1. To Select a Measurement type, se­lect from: pH, ORP, Redox, Ammo­nia, Fluoride, and Custom ISE and press ENTER/MENU.
2. To program the Preamp location, se­lect Analyzer or Sensor/JBox, and press ENTER/MENU.
3. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
4. To program Reference Impedance. Select Low or High and press ENTER/MENU.
5. To choose the wiring scheme, Select Normal or Reference to Ground and press ENTER/MENU.
6. To program the display resolution, Select 0.01 pH or 0.1 pH and press ENTER/MENU.
7. To enable pH sensor diagnostics, select the “pH diagnostic setup” tab under Program­ming. Select Sensor 1 or Sensor 2. Select NEXT. Select On under sensor diagnostics to enable pH diagnostics.
NOTE: pH sensor diagnostics must be enabled to include diagnostic values such as Glass Impedance and Reference Impedance in EXCEL data log sheets after data download to USB. Enabling pH sensor diagnostics also allows assignment of Glass Impedance and Reference Impedance to the two-dimensional on-screen process graph accessible under “Display Setup/View Graph”.
6.3 ORP Measurement Programming
The section describes how to configure the 56 analyzer for ORP measurements.
The following programming and configuration functions are covered:
1. Measurement type: pH Select pH, ORP, Redox, Ammonia, Fluoride, Custom ISE
2. Preamp location: Analyzer Identify preamp location
3. Filter: 4 sec Override the default input filter, enter 0-999 seconds
4. Reference Z: Low Select low or high reference impedance
5. Sensor wiring scheme: Normal or Reference to Ground
To configure the ORP measurement board, access the Program screen by pressing ENTER/MENU from the main screen and then select the Measurement tab.
1. To Select a Measurement type, select ORP and press ENTER/MENU.
2. To program the Preamp location, select Analyzer or Sensor/JBox, and press ENTER/MENU.
3. To Override the default input
44 Programming Measurements
Page 55
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
filter, enter 0-999 seconds and press ENTER/MENU.
4. To program Reference Impedance. Select Low or High and press ENTER/MENU.
. To choose the wiring scheme, Select Normal or Reference to Ground and press
5
ENTER/MENU.
6.4 Contacting Conductivity Measurement Programming
The section describes how to configure the 56 analyzer for conductivity measurements using contacting conductivity sensors. The following programming and configuration functions are covered.
1. Measure: Conductivity, Select Conductivity, Resistivity, TDS. Salinity or % conc
2. Type: 2-Electrode Select 2-Electrode or 4-Electrode type sensors
3. Cell K: 1.00000/cm Enter the cell Constant for the sensor
4. Measurement units
5. Filter: 2 sec Override the default input filter, enter 0-999 seconds
6. Range: Auto Select measurement
Auto-range or specific range
7. Temp Comp: Slope Select Temp
Comp: Slope, Neutral Salt, Cation or Raw
8. Slope: 2.00% / °C Enter the linear
temperature coefficient
9. Ref Temp: 25.0 °C Enter the Refer-
ence temp
10. Cal Factor: default=0.95000/cm
Enter the Cal Factor for 4-Electrode sensors from the sensor tag
To configure the contacting conductivity measurement board, access the Program screen by pressing ENTER/MENU from the main screen and then select the Measurement tab.
1. To program a Measurement type, select Conductivity Select Conductivity, Resistivity,
TDS. Salinity or % conc. and press ENTER/MENU.
2. To program a sensor type, select 2-
Electrode Select 2-Electrode or 4­Electrode type sensors and press ENTER/MENU.
3. To program the Cell constant, enter
the exact cell constant value ex­pressed as 1.XXXXX/cm the for the sensor and press ENTER/MENU.
4. To program Measurement units, select uS/cm or mS/cm and press ENTER/MENU.
5. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
6. To program the measurement range, select a specific range appropriate for your process
and press ENTER/MENU.
7. To program the Temp Comp method, choose Slope, Neutral Salt, Cation or Raw
and press ENTER/MENU.
Programming Measurements 45
Page 56
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
8. To change the Temperature compensation Slope, enter the linear temperature
coefficient expressed as X.XX% / °C and press ENTER/MENU.
9. To program the Reference Temperature for Manual temperature compensation (not
rom probe RTD), enter the Reference temp expressed as XX.X °C and press
f ENTER/MENU.
10. To program the Cal Factor for 4-Electrode sensors, enter the value shown on the sensor
tag, expressed as X.XXXXX/cm and press ENTER/MENU.,
6.5 Toroidal Conductivity Measurement Programming
The section describes how to configure the 56 analyzer for conductivity measurements using inductive/toroidal sensors. The following programming and configuration functions are covered:
1. Measure: Conductivity Select Conductivity, Resistivity, TDS. Salinity or % conc
2. Sensor model: 228 Select sensor type
3. Measurement units
4. Cell K: 3.00000/cm Enter the cell Constant for the sensor
5. Temp Comp: Slope Select Temp Comp: Slope, Neutral Salt, Cation or Raw
6. Slope: 2.00% / °C Enter the linear temperature coefficient
7. Ref Temp: 25.0 °C Enter the Reference temp
8. Filter: 2 sec Override the default
input filter, enter 0-999 seconds
9. Range: Auto Select measurement
Auto-range or specific range
To configure the Contacting conductivity measurement board, access the Program screen by pressing ENTER/MENU from the main screen and then select the Measurement tab.
1. To program a Measurement type,
select Conductivity Select Conductivity, Resistivity, TDS. Salinity or % conc. and press ENTER/MENU.
2. To program the sensor model, select
228 or other toroidal model number and press ENTER/MENU.
3. To program Measurement units,
select uS/cm or mS/cm and press ENTER/MENU.
4. To program the Cell constant, enter the exact cell constant value expressed as
3.XXXXX/cm the for the sensor and press ENTER/MENU.
5. To program the Temp Comp method, choose Slope, Neutral Salt, Cation or Raw
and press ENTER/MENU.
6. To change the Temperature compensation Slope, enter the linear temperature
coefficient expressed as X.XX% / °C and press ENTER/MENU.
46 Programming Measurements
Page 57
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
7. To program the Reference Temperature for Manual temperature compensation
(not from probe RTD), enter the Reference temp expressed as XX.X °C and press
NTER/MENU.
E
. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
8
9. To program the measurement range, select a specific range appropriate for your process
and press ENTER/MENU.
6.6 Chlorine Measurement Programming
With a Chlorine measurement board installed, The 56 can measure any of four variants of Chlorine:
Free Chlorine
Total Chlorine
Monochloramine
pH-independent Free Chlorine
The section describes how to configure the 56 analyzer for Chlorine measurements.
6.6.1 Free chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Free Chlorine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display res-
olution 0.01 or 0.001
4. Free Cl Correct: Live Select Live/
Continuous pH correction or Manual
5. Manual pH: 7.00 pH For Manual pH
correction, enter the pH value
6. Filter: 5 sec Override the default input filter, enter 0-999 seconds
7. Dual Slope Calibration: Enable or Disable
1. To program the Measurement type, select Free Chlorine,
pH Ind Free Cl., Total Cl, or Mono­chloramine and press ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press ENTER/MENU.
3. To program the Measurement Resolu-
tion: Select 0.01 or 0.001 and press ENTER/MENU.
4. To program Free Cl Correction, se-
lect Live/Continuous pH correction or Manual and press ENTER/MENU
Programming Measurements 47
Page 58
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
5. To program for Manual pH correction, enter the pH value and press ENTER/MENU.
6. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
. To use Dual Slope Calibration, select Enable or Disable and press ENTER/MENU.
7
6.6.2 Total chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Total Chlorine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5 sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable or Disable
1. To program the Measurement type, select Free Chlorine, pH Ind Free Cl., Total Cl, or
Monochloramine and press ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press ENTER/MENU.
3. To program the Measurement Resolu-
tion: Select 0.01 or 0.001 and press ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press ENTER/MENU.
5. To use Dual Slope Calibration, select Enable or Disable and press ENTER/MENU.
6.6.3 Monochloramine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Monochloramine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5 sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable or
Disable
48 Programming Measurements
Page 59
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
1. To program the Measurement type, select Free Chlorine, pH Ind Free Cl., Total Cl, or
Monochloramine and press ENTER/MENU.
2. To program the Measurement Units: select ppm mg/L and press ENTER/MENU.
3. To program the Measurement Reso-
lution: Select 0.01 or 0.001 and press ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press ENTER/MENU.
5. To use Dual Slope Calibration, select
Enable or Disable and press ENTER/MENU.
6.6.4 pH-independent free chlorine measurement programming
This Chlorine sub-section describes how to configure the 56 analyzer for Free Chlorine measurements using the pH-independent free chlorine sensor, 498CL-01, manufactured by Rosemount. The following programming and configuration functions are covered:
1. Measure: Free Chlorine Select Free
Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine
2. Units: ppm Select units ppm
or mg/L
3. Resolution: 0.001 Select display
resolution 0.01 or 0.001
4. Filter: 5sec Override the default
input filter, enter 0-999 seconds
5. Dual Slope Calibration: Enable
or Disable
1. To program the Measurement type, select Free Chlorine,
pH Ind Free Cl., Total Cl, or Monochloramine and press ENTER/MENU.
2. To program the Measurement Units:
select ppm mg/L and press ENTER/MENU.
3. To program the Measurement
Resolution: Select 0.01 or 0.001 and
press ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press ENTER/MENU.
5. To use Dual Slope Calibration, select
Enable or Disable and press ENTER/MENU.
Programming Measurements 49
Page 60
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
6.7 Oxygen Measurement Programming
his section describes how to configure the 56 analyzer for dissolved and gaseous oxygen
T measurement using amperometric oxygen sensors. The following programming and configuration functions are covered:
1. Sensor type: Select Water/Waste,
Trace. BioRx, BioRx-Other, Brew, %O In Gas
2. Measure type: Select Concentration,
% Saturation, Partial Pressure, Oxygen in Gas
3. Units: ppm Select ppm, mg/L, ppb,
µg/L, % Sat, %O2-Gas, ppm Oxygen­Gas
4. Pressure Units: bar Select
pressure units: mm Hg, in Hg,. Atm, kPa, mbar, bar
5. Use Press: At Air Cal Select atmos-
pheric pressure source – internal or mA Input
6. Salinity: 00.0‰ Enter Salinity as ‰
7. Filter: 5 sec Override the default
input filter, enter 0-999 seconds
8. Partial Press: mmHg Select mm Hg,
in Hg. atm, kPa, mbar or bar for Par­tial pressure
2
1. To program Sensor type, Select
Water/Waste, Trace. BioRx, BioRx-Other, Brew, or %O2In Gas and press ENTER/MENU.
2. To program Measure type: Select Concentration, % Saturation, Partial Pressure, or
Oxygen in Gas and press ENTER/MENU.
3. To program measurement Units, Select ppm, mg/L, ppb, g/L, % Sat, %O2-Gas, or ppm
Oxygen-Gas and press ENTER/MENU.
4. To program Pressure Units: Select pressure units: mm Hg, in Hg,. Atm, kPa, mbar, or bar
and press ENTER/MENU.
5. To program which atmospheric Pressure source to use during Air Cal, Select internal or
mA Input and press ENTER/MENU.
6. To program Salinity, Enter the known Salinity value as ‰ and press ENTER/MENU.
7. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
8. To program Partial Press: Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial
pressure and press ENTER/MENU.
50 Programming Measurements
Page 61
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
6.8 Ozone Measurement Programming
his section describes how to configure the 56 analyzer for ozone measurement using
T amperometric ozone sensors. The following programming and configuration functions are covered:
1. Units: ppm Select ppm, mg/L, ppb, µg/L
2. Resolution: 0.001 Select display resolution 0.01 or 0.001
3. Filter: 5sec Override the default input filter, enter 0-999 seconds
6.9 Oxygen Measurement Programming
1. To program measurement Units, Select ppm, mg/L, ppb, µg/L, % Sat, %O2-Gas, or ppm
Oxygen-Gas and press ENTER/MENU.
2. To program the Measurement Resolution: Select 0.01 or 0.001 and press ENTER/MENU.
3. To Override the default input
filter, enter 0-999 seconds and press ENTER/MENU.
6.10 Turbidity Measurement Programming
This section describes how to configure the 56 analyzer for Turbidity measurements. The following programming and configuration functions are covered.
1. Measurement type: Turbidity Select
Turbidity or TSS calculation (estimated TSS)
2. Sensor type: Select EPA or ISO
3. Measurement units: NTU, FTU, FNU
4. Filter: 20 sec Override the default
input filter, enter 0-999 seconds
5. Bubble Rejection: On Intelligent
software algorithm to eliminate erroneous readings caused by bubble accumulation in the sample
1. To program the Measurement type,
Select Turbidity or TSS calculation (estimated TSS) and press ENTER/MENU.
2. To program the Sensor type:
Select EPA or ISO and press ENTER/MENU.
3. To program Measurement units:
NTU, FTU, FNU and press ENTER/MENU.
4. To Override the default input filter,
enter 0-999 seconds and press ENTER/MENU.
5. To program Bubble Rejection, select On or Off and press ENTER/MENU.
Programming Measurements 51
Page 62
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
6.11 Flow Measurement Programming
his section describes how to configure the 56 analyzer for flow measurement when used
T with a compatible pulse flow sensor. The following programming and configuration functions are covered:
To program pulse flow, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each programming routine.
1. Measurement type Pulse
Flow Select Pulse Flow or mA Current Input
2. Measurement units: GPH Select
GPM, GPH, cu ft/min, cu ft/hour, LPM, L/hour, m3/hr. or
3. Filter: 5sec Override the default
input filter, enter 0-999 seconds
1. To program Measurement type, Select Pulse Flow or mA Current Input and press
ENTER/MENU.
2. To program Measurement units: Select GPM, GPH, cu ft/min, cu ft/hour, LPM, L/hour, or
m3/hr. and press ENTER/MENU.
3. To Override the default input filter, enter 0-999 seconds and press ENTER/MENU.
6.12 Current Input Programming
This section describes how to configure the 56 analyzer for current input measurement when wired to an external device that transmits 4-20 mA or 0-20 mA analog current output. The following programming and configuration functions are covered.
1. Measurement type mA input Override the default (Flow) and select mA current input
2. mA Input Temperature Select Temperature, Pressure, Flow or Other
3. Measurement units: °C. Select measurement units based on selected input device type
4. Input Range: 4-20 mA Select 4-20
mA or 0-20 mA
5. Low Value: 0.000oC Enter
the low measurement value to as­sign to 4 mA
6. High Value: 100.0oC Enter
the high measurement value to as­sign to 20mA
7. Filter: 05 sec Override the
default input filter, enter 0-999 seconds
52 Programming Measurements
Page 63
Instruction Manual Section 6: Programming Measurements
LIQ-MAN-56 April 2017
1. To override the default the
Measurement type (Flow)
elect mA current input and press
s
0/4
ENTER/MENU.
2. To program the mA Input type,
Select Temperature, Pressure, Flow or other and press ENTER/MENU.
3. To program measurement units,
Select measurement units based on selected input device type and press ENTER/MENU.
4. To program the Input Range: Select 4-20 mA or 0-20 mA and press ENTER/MENU.
5. To program the Low input Value, enter the low measurement value to assign to 4mA
and press ENTER/MENU.
6. To program the High input Value, enter the high measurement value to assign to 20mA
and press ENTER/MENU.
7. To override the default input filter, enter 0-999 seconds and press ENTER/MENU.
Programming Measurements 53
Page 64
Section 6: Programming Measurements Instruction Manual
April 2017 LIQ-MAN-56
54
Page 65
Instruction Manual Section 7: PID Control
LIQ-MAN-56 April 2017
Section 7.0 PID Control
7.1 Introduction
7.1.1 Measurement and Set Point (Feedback Control)
The 56 controller is given two items of information: measurement and set point. The controller reacts to the difference in value of these two signals and produces an analog output signal to eliminate that difference. As long as the difference exists, the controller will try to eliminate it with the output signal. When measurement and set point are equal, the condition of the controller is static and its output is unchanged. Any deviation of measurement from set point will cause the controller to react by changing its output signal.
7.1.2 Proportional Mode
The simplest control is proportional control. In this control function, the error from set point, divided by the control range, is multiplied by the Gain constant to produce the output.
The control range is the percent of the analog output span (the difference between the 4 (or
0) mA and 20 mA settings) through which the measured variable must move to change the output from minimum to maximum.
The smaller the Gain, the less the controller reacts to changes in the measured variable. The larger the Gain, the more the controller reacts to changes in the measured variable.
The proportional control output is given by the expression below. As can be seen, the overall gain is determined by the control range chosen (URV and LRV) and the Gain:
Proportional Output (%) = Gain * (PV – SP) * 100 / (URV, upper range value and LRV, lower range value).
7.1.2.1 Direct Acting Control Action
Direct acting control action increases the control output as the measured variable increases above the setpoint. The LRV is usually set to the setpoint value, so that the control output is 0% at the setpoint, and the URV is greater than the setpoint so that the 100% control output is at a higher measurement value. The Gain parameter can then be adjusted to produce the desired gain.
Fig. 7-1 Direct Acting Control
PID Control 55
Page 66
Section 7: PID Control Instruction Manual
April 2017 LIQ-MAN-56
xample of direct acting control: Lower the pH of a solution at 10 pH by adding acid to
E
control it at 8 pH with the Gain parameter assumed to be 1.0. The higher the measured pH, the more acid is required to lower the pH toward the setpoint, but as the pH approaches the setpoint less acid is required:
Fig. 7-2 Example of Direct Acting Control
7.1.2.2 Reverse Acting Control Action
Reverse acting control action, decreases the control output as the measured variable increases toward the setpoint. The LRV is usually set to the setpoint value, so that the control output is 0% at the setpoint, and the URV is less than the setpoint value so that the 100% control output is at a lower measurement value. The Gain can then be adjusted to produce the desired gain.
Fig. 7-3 Reverse Acting Control
Example of reverse acting control: Add base to a solution at 8.0 pH, to control the pH to
10.0 pH with an assumed Gain parameter of 1.0. The lower the measured pH, the more base is required to raise the pH toward the setpoint, but as the pH approaches the setpoint less base is required:
56 PID Control
Page 67
Instruction Manual Section 7: PID Control
LIQ-MAN-56 April 2017
ig. 7-4 Example of Reverse Acting Control
F
7.1.3 Proportional Bias
Most processes require that the measured variable be held at the set point. The proportional mode alone will not automatically do this, if an output greater than 0% is needed to keep the PV at setpoint. At setpoint, the control output is 0%, and if a non-zero control output is needed to keep the PV at the setpoint, proportional alone will only stabilize the measured variable at some offset (deviation) from the desired setpoint.
Bias is used to provide a constant control output at the setpoint to maintain PV at the setpoint. The effect of Bias is expressed as follows:
Proportional Output (%) = [Gain * (PV – SP) * 100 / (URV – LRV)] + BIAS
Fig. 7-5 Direct Acting Control with Bias
PID Control 57
Page 68
Section 7: PID Control Instruction Manual
April 2017 LIQ-MAN-56
Fig. 7-6 Reverse Acting Control with Bias
7.1.4 Proportional Plus Integral (Reset)
For the automatic elimination of deviation, Integral mode, also referred to as Reset, is used. The proportional function is modified by the addition of automatic reset, rather than a constant Bias value. With the reset mode, the controller continues to change its output until the deviation between measurement and set point is eliminated.
The action of the reset mode depends on the overall gain. The rate at which it changes the controller output is based on the overall gain band size and the reset time (I). The reset time is the time required for the reset mode to repeat the proportional action once. It is expressed as seconds per repeat, adjustable from 0-3,000 seconds.
The reset mode repeats the proportional action as long as an offset from the set point exists. Reset action is cumulative. The longer the offset exists, the more the output signal is increased. If the PV overshoots the setpoint, the reset action will decrease. When the measurement reaches the setpoint and the proportional control action becomes zero, there will be an accumulated integral control action to keep the process at the setpoint.
The controller configured with reset continues to change until there is no offset. If the offset persists, the reset action eventually drives the controller output to its 100% limit - a condition known as "reset windup".
Once the controller is "wound up", the deviation must be eliminated or redirected before the controller can unwind and resume control of the measured variable. The integral time can be cleared and the "windup" condition quickly eliminated by manually overriding the 56 analog output using the manual mode to reduce the control output and then setting the reset time to 0 seconds to make integral control action 0%. The reset time can then be changed to avoid reset windup.
The proportional plus integral control output is given below. Note that the larger the reset time (I), the slower the integral response will be:
% Output = [Gain × 100 / (URV – LRV)] × [(PV – SP) + 1/I Σ (PVt– SP) Δt]
58 PID Control
Page 69
Instruction Manual Section 7: PID Control
LIQ-MAN-56 April 2017
7.1.5 Derivative Mode (Rate)
erivative mode provides a 3rd control mode, which responds to the rate of change of the
D Proportional control output, multiplied by the Derivative parameter D which has units of seconds. The contribution of the derivative response is given below:
% Output = [Gain × 100 / (URV – LRV)] × D × [(PV
The purpose of derivative action is to provide a quick control response to changes in the measured parameter. In general, it is not often used in concentration control, and in fact, it has been estimated that 90 to 95% of all control applications use only Proportional plus Integral control. Any noise in the measurement causes problems with derivative action. Temperature measurements tend to be less noisy than other measurements, and derivative action is most often used for temperature control.
– SP) (PV
t
7.1.6 Process Characterization and Tuning
Control loops are tuned by the choice of the control range and the selection of the control parameters. How these parameters are chosen should depend on how the process responds to controller output. The process response is characterized by certain behaviors, which are due to such factors as mixing and reaction time, response time of the process to control output changes, and the characteristics of the final operator, i.e. control values, pumps, heaters, etc. With these characteristics known, initial control settings can be developed.
A good reference to PID control is provided by the book, “Control Loop Foundation—Batch and Continuous Processes”, by Terrence Blevins and Mark Nixon, International Society of Automation, Research Triangle Park, NC, © 2011.
A guide to tuning control is provided by the book, “Good Tuning: A Pocket Guide”, by Gregory K. McMillan, International Society of Automation, Research Triangle Park, NC, ©
2005.
– SP)] / Δt
t-1
7.2 PID Setup
7.2.1 PID Control
The 56 current ¬outputs (one or all four) can be programmed for PID control. PID control can be applied to any of the measurements provided by the sensor boards, such as pH, conductivity, and concentrations. In addition, PID control can be applied to temperature and any measurement input to the 56 using the flow/4-20 mA board.
The output signal of PID control is used with a final control element, which can vary is output from 0 to 100% in response to the control signal. Final control elements can include control valves, pumps or heaters.
7.2.2 Selecting PID Control
Select PID control, the analog output to be used, and the measurement and range from the main analog output setup window:
PID Control 59
Page 70
Section 7: PID Control Instruction Manual
April 2017 LIQ-MAN-56
Basic Definitions
Output – Select the analog output (1 through 4) to be configured for PID control.
Analog/PID/Simulate – Choose PID
Assign – Select the Measurement to be controlled. Note: This measurement can also
be a 4-20 mA signal input brought in by the flow/ 4-20 mA board.
Range – Select either 0-20 mA or 4-20 mA range, depending on the signal range used
by the final control element, e.g. a pump or valve.
Select Next to go to the PID Setup parameters.
7.2.3 PID Setup Parameters
The PID control setup window contains the PID control tuning parameters.
Also note that the upper portion of the screen shows the measurement chosen for control (PV), and the control output in mA and % Output. This makes it possible to observe the primary variable (PV) and the control output, in terms of percent and milliamps, while tuning PID control.
PID Control Parameters: Basic Definitions
Setpoint – Select the desired setpoint.
URV – The value of PV (in the above example, 14.00 pH) at which the control will be 20
mA or 100% output.
LRV -- The value of PV (in the above example 0.00 pH) at which the control will be 0 or 4
mA or 0% output.
60 PID Control
Page 71
Instruction Manual Section 7: PID Control
LIQ-MAN-56 April 2017
Note: If you want the control output to increase as PV (in this case pH) increases, URV should be greater than LRV. This is direct acting control action. Examples of direct acting control are
he addition of acid to decrease pH and adding water to a solution to decrease the
t concentration.
If you want the control output to decrease as PV increases, i.e. reverse acting control action, the URV should be less than LRV. Examples of reverse acting control are adding caustic to increase pH and adding a concentrated solution to water to make a solution of lower concentration.
Bias (range: 0 to 100%; default 0%) – Bias is a fixed control output which allows the con-
trol output to be greater than zero when the measurement (PV) is at setpoint. It is used in proportional only control to prevent cycling resulting from the control output going to zero at the setpoint.
Gain (range: 0.0 to 1000.0; default 0.0) – In proportional (P) only control, the output is
directly proportional to the difference between the process variable (PV) and the set­point divided by the output span (URV – LRV). Gain is a factor which multiplies the pro­portional output to meet the requirements of the process being controlled. Using Gain values less than 1 reduce the proportional output while Gain values greater than 1 in­crease the proportional output.
Integral (Reset, I) (range: 0 to 3,000 seconds; default 0 seconds) – Integral repeats the
proportional action in a time period given by the reset time (I). The reset time is given in seconds per repeat and is adjustable from 0 to 3,000 seconds. Integral control acts as an automatic bias which increases or decreases the overall control output in re­sponse to the error (PV – SP) to keep the PV at the setpoint.
Derivative (Rate, D) (range: 0 to 3,000 seconds; default 0 seconds) –Derivative action,
gives an immediate control output in response to changes in the proportional output with time (derivative). The amount of increase or decrease depends on the rate of change of the error. The rate constant (D) allows the user to adjust the amount deriva­tive control contributes to the control signal. Smaller values reduce the effect of deriva­tive control.
Mode: Mode has two settings, Auto (Automatic) and Man (Manual). In the Auto mode
the control output is controlled automatically by the PID algorithm. In manual mode, the control output can be set to a constant value; this is useful during transmitter cali­bration or servicing.
Value (Manual) (range: 0 100%) – When the Manual mode is chosen; this control ap-
pears on the screen and allows you to write the constant control output value in the Man­ual mode.
Using the Next button lead to the final PID control window:
7.2.4 Transport Time
Transport Time makes it possible to apply PID control action to a process flowing in a pipe for a short period of time (run time), and then hold the control output fixed to allow the treated sample time to mix and travel to the pH or other analytical sensor (transport time). If properly tuned the PV should reach the setpoint after successive time periods.
It is best used when raw sample pH (or concentration) remains relatively steady for long periods of time, as is the case for samples flowing from a large body of water. It should not be used where process upsets are possible because the delay in applying control will make recovery from the upset slow, and can result in overshoot after the incoming sample has returned to a normal range.
PID Control 61
Page 72
Section 7: PID Control Instruction Manual
April 2017 LIQ-MAN-56
Transport Time Parameters: Basic Definitions
Transport Time (On/Off) – Turns the Transport Time feature on or off.
Transport Time (range 1 to 600 seconds) – When Transport Time is turned On; a
control appears at the right of it, which allows the value of the Transport Time to be enter. This will be the time period that PID control output is held constant, while the treated sample travels to the sensor.
Run Time (range 0 to 60 seconds) – This is the time period that PID control action is
automatic. It always must be a shorter time than Transport Time.
Fault: Basic Definitions
1. Fault – When a measurement fault occurs (either sensor or transmitter) the control
output can be setup to continue providing a live control output or the output can be set to a fixed value.
If the live reading is used during a fault condition, the control output could be based on an erroneous measurement, which might cause problems. Using a fixed value for control output during a fault condition can ensure that the control output goes to an acceptable value.
2. Fault Current – If a fixed value on fault is chosen, this parameter selects the output. The control output on fault can be set to a value to prevent a major upset or an unsafe condition.
Note: If a fixed fault current output is chosen and PID control Mode is set to Manual, the Manual output value will override the Fault Current value.
62 Time Proportional Control
Page 73
Instruction Manual Section 8: Time Proportional Control
LIQ-MAN-56 April 2017
Section 8.0 Time Proportional Control
8.1 Introduction
8.1.1 Time Proportional Control
Time Proportional Control is more commonly known as Duty Cycle or Pulse Width Modulation. It applies PID control to the activation of a relay rather than using an analog output.
The TPC output is defined as the percent of time that a relay is on (% On Time), during a user selected time period (Time Period). As the control output increases the on time increases:
Fig. 8-1 TPC Periods and On Time
The proportional, integral, and derivative are defined the same as analog PID control, but use % On Time instead of % Output:
Proportional On Time (%) = [Gain × (PV – SP) x 100 / (URV – LRV)] + BIAS
Proportional and Integral Control:
% On Time = [Gain × 100 / (URV – LRV)] + [(PV – SP) + 1/I Σ (PVt– SP) Δt]
Derivative Mode:
% On Time = [Gain × 100 / (URV – LRV)] × D × [(PVt– SP) (PV
As with analog PID control, TPC can be direct acting (URV > LRV) or reverse acting (LRV > URV). For more detail see Section X.1, PID Control Introduction.
– SP)] / Δt
t-1
Time Proportional Control 63
Page 74
Section 8: Time Proportional Control Instruction Manual
April 2017 LIQ-MAN-56
8.2 TPC Setup
8.2.1 Selecting TPC
Select TPC control, the relay to be used, and the measurement to be controlled from the main relay setup window:
Basic Definitions
Relay – Select the relay (1 through 4) to be use for TPC control.
Type – Choose TPC
Assign – Select the Measurement to be controlled. Note: This measurement can also be a 4-20 mA signal input brought in by the flow/ 4-20 mA board.
• Select Next to go to the PID Setup parameters
8.2.2 TPC Setup Parameters
The TPC setup window contains the PID control tuning parameters.
Also note that the upper portion of the screen shows the relay number and the value of the measurement assigned to it, the % On Time for the relay, and the current state of the relay, i.e. On or Off. This makes it possible to observe the primary variable (PV), the % On Time, and the relay state, while tuning time proportional control.
64 PID Control
Page 75
Instruction Manual Section 8: Time Proportional Control
LIQ-MAN-56 April 2017
PID Control Parameters: Basic Definitions
Setpoint – Select the desired setpoint.
URV – The value of PV (in the above example, 14.00 pH) at which the control will be
00% On Time.
1
LRV – The value of PV (in the above example 0.00 pH) at which the control will be 0% On
Time.
Note: If you want the control output to increase as PV (in this case pH) increases, URV should be greater than LRV. This is direct acting control action. Examples of direct acting control are the addition of acid to decrease pH and adding water to a solution to decrease the concentration.
If you want the control output to decrease as PV increases, i.e. reverse acting control action, the URV should be less than LRV. Examples of reverse acting control are adding caustic to increase pH and adding a concentrated solution to water to make a solution of lower concentration.
Time Period (range: 10 to 3,000 seconds; default 30) -- The time period for each cycle of
TPC.
Bias (range: 0 to 100%; default 0%) – Bias is a fixed control output which allows the % on
time to be greater than zero when the measurement (PV) is at setpoint. It is used in pro­portional only control to prevent cycling resulting from the % on time going to zero at the setpoint.
Relay default (Close, Open, None) – Select the relay action during a fault condition.
Gain (range: 0.0 to 1000.0; default 0.0) – In proportional (P) only control, the output is
directly proportional to the difference between the process variable (PV) and the setpoint divided by the output span (URV – LRV). Gain is a factor which multiplies the proportional output to meet the requirements of the process being controlled. Using Gain values less than 1 reduce the proportional output while Gain values greater than 1 increase the proportional output.
Integral (Reset, I) (range: 0 to 3,000 seconds; default 0 seconds) – Integral repeats the
proportional action in a time period given by the reset time (I). The reset time is given in seconds per repeat and is adjustable from 0 to 3,000 seconds. Integral control acts as an automatic bias which increases or decreases the overall control output in response to the error (PV – SP) to keep the PV at the setpoint.
Derivative (Rate, D) (range: 0 to 3,000 seconds; default 0 seconds) –Derivative action,
gives an immediate control output in response to changes in the proportional output with time (derivative). The amount of increase or decrease depends on the rate of change of the error. The rate constant (D) allows the user to adjust the amount derivative control contributes to the control signal. Smaller values reduce the effect of derivative control.
Mode: Mode has two settings, Auto (Automatic) and Man (Manual). In the Auto mode
the control output is controlled automatically by the PID algorithm. In manual mode, the control output can be set to a constant value; this is useful during transmitter calibration or servicing.
% On Time (Manual) (range: 0 100%) – When the Manual mode is chosen; this control
appears on the screen and allows you to write the constant On Time value in the Manual mode.
PID Control 65
Page 76
Section 8: Time Proportional Control Instruction Manual
April 2017 LIQ-MAN-56
66
Page 77
Instruction Manual Section 9: Alarm Relay Functions
high alarm
tnioptes
high alarm logic
1.5
1.3
deadband (0.2)
time
PV
alarm activates
alarm deactivates
LIQ-MAN-56 April 2017
Section 9.0 Alarm Relay Functions
9.1 General
An alarm is a relay that closes a set of contact points (a switch) inside the analyzer. In doing so, the relay closes an electrical circuit and turns on a device wired to the contacts. The 56 Advanced Analyzer has four alarm relays and seven relay control functions. The relays are turned on and off by the analyzer based on the control points, setpoints or control parameters that you program into the analyzer through the keypad. See Section 9.2 through 9.7 to program the alarm relay functions. Each relay functions section includes a description, a figure detailing its operation, a step-by-step setup procedure, and a table or default and programmable limit settings.
The 56 has the following relay control functions:
Table 9-1. Alarm Relay Functions
Relay Control Functions Common applications Section
High/Low Concentration Alarm measurement setpoint control Sec. 9.2
Delay Timer chemical mixing and neutralization Sec. 9.3
Bleed and Feed blowdown and chemical addition Sec. 9.4
Totalizer Relay Activation chemical dosing in reactors Sec. 9.5
Interval Timer periodic probe cleaning Sec. 9.6
Date and Time Activation seawater-cooled condensers Sec. 9.7
9.2 High/Low Concentration Alarm
9.2.1 Description
High/Low concentration alarms are setpoint alarms with adjustable deadband. These operate as simple on/off alarms used for applications requiring discrete on/off control of pumps and valves. Typical applications include demineralizer bed regeneration and blowdown in boilers and cooling towers. Any active device variable in the 56 analyzer can be programmed as a high/low concentration alarm including the primary or secondary variables, temperature, raw values and diagnostics.
A schematic of the high/low concentration (setpoint) alarm operation is shown below.
Figure 9-1. High/Low Concentration Alarm operation
Alarm Relay Functions 67
Page 78
Section 9: Alarm Relay Functions Instruction Manual
April 2017 LIQ-MAN-56
9.2.2 Setup
ccess high/low concentration (setpoint) alarms by pressing ENTER/MENU from the main
A screen and then Program/Relays/Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Setpoint as the relay type and press ENTER/MENU.
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to the designated relay and press ENTER/MENU.
4. Logic: Set High for a high setpoint or Low for a low setpoint and press ENTER/MENU.
5. Setpoint: Enter the desired setpoints value. Press ENTER/MENU.
6. Deadband: To set deadband as a measurement value, enter the change in the process value needed after the relay deactivates to return to normal (and thereby preventing re­peated alarm activation). Press ENTER/MENU.
7. Select NEXT. Press ENTER/MENU to advance to the next setup screen.
8. Normal state: Set the normal alarm condition as Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to nor­mally open or normally closed upon power up.
Table 9-2. Defaults and programmable limits
Relay Function Limits and Selections Default
Setpoint NA NA
Logic Low/High High
Setpoint * *
Deadband * 0.000
On time 0 to 999.9 min 0 min
Delay time 0 to 999.9 min 0 min
Normal state Close/Open Open
* See Appendix 1 – HART and Device Variables
9.3 Delay Timer
9.3.1 Description
Delay Timer is a concentration control scheme which delays live measurement after chemical addition using (one or all four of) of the 56 alarm relays. This ensures sufficient mixing time in a vessel or recirculation loop before live sensor measurement, preventing unmixed readings that might cause overshooting. Relay On time and Delay times are field­programmable. Typical applications that would utilize the Delay Timer are: concentration control in vessels, pH adjustments for neutralization and endpoint control for oxidation­reduction reactions.
68 Alarm Relay Functions
Page 79
Instruction Manual Section 9: Alarm Relay Functions
high alarm
tnioptes
deadband
time
P
V
delay time
on time
measurement
LIQ-MAN-56 April 2017
A schematic of the Delay Timer operation is shown:
Figure 9-2. Delay Timer Alarm operation
9.3.2 Setup
Access Delay Timer by pressing ENTER/MENU from the main screen and then Program/Relays/ Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Delay Timer as the relay type and press ENTER/MENU.
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to the designated relay and press ENTER/MENU.
4. Logic: Set High for high reading setpoint logic or Low relay logic for low reading setpoint logic and press ENTER/MENU.
5. Setpoint: Enter the desired setpoints value. This will activate an alarm event when the process measurement reaches the entered setpoint value. Press ENTER/MENU. See Table 9-2 for entry limits.
6. Deadband: To set deadband as a measurement value, enter the change in the process value needed after the relay deactivates to return to normal (and thereby preventing re­peated alarm activation). Press ENTER/MENU.
7. Select NEXT Press ENTER/MENU to advance to the next setup screen.
8. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The as­signed measurement value will be on hold during this time.
9. Delay time: Enter the time in minutes (X.X min) to take the assigned measurement off hold after the relay is re-energized to begin reporting live values.
Alarm Relay Functions 69
10. Normal state: Set the normal alarm condition as Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to nor­mally open or normally closed upon power up.
Page 80
Section 9: Alarm Relay Functions Instruction Manual
time
bleed on
on
off
AL1
(bleed)
AL2
(feed)
feed on
delay
feed time = X% of bleed time
April 2017 LIQ-MAN-56
able 9-3. Defaults and programmable limits
T
Relay Function Limits and Selections Default
Delay Timer NA NA
Logic Low/High High
Setpoint * *
Deadband * 0.000
Normal state Close/Open Open
See Appendix 1 – HART and Device Variables
*
9.4 Bleed and Feed
9.4.1 Description:
A bleed and feed timer is typically used to replace chemicals lost during blowdown. It involves two or more relays. The bleed relay is a normal setpoint alarm relay. Once the bleed relay deactivates, one or more feed relays activate for a percentage of the time the bleed relay was on.
Bleed and Feed supports continuous monitoring of blow-down water conductivity to determine the point of excessive conductivity. At a programmable maximum concentration value, dumping (bleeding) of the excessively dirty blow-down water is triggered. Subsequently, pumping (feeding) of additional make-up water chemicals is enabled to account for lost blow-down water. Through level control, make-up water is added in proportion to the volume of blowdown material lost through dumping and evaporation.
A schematic of the Bleed and Feed timer operation is shown:
Figure 9-3. Bleed and feed timer alarm operation
9.4.2 Setup
Access Bleed and Feed Timers by pressing ENTER/MENU from the main screen and then Program/Relays/Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign relay 1 for Bleed and Feed and press ENTER/MENU.
2. Type: Select Bleed and Feed as the relay type and press ENTER/MENU.
70 Alarm Relay Functions
3. Assign: Assign S1 (sensor 1), S2 (sensor 2 if available) or other available parameters to the designated relay and press ENTER/MENU.
4. Logic: Set High for high setpoint or Low for low setpoint and press ENTER/MENU.
Page 81
Instruction Manual Section 9: Alarm Relay Functions
LIQ-MAN-56 April 2017
5. Setpoint: Enter the desired setpoints value. This will activate an alarm event when the process measurement reaches the entered setpoint value. Press ENTER/MENU. See Table
-3 for entry limits.
9
6. Select NEXT Press ENTER/MENU to advance to the next setup screen.
7. Deadband: To set deadband as a measurement value, enter the change in the process value needed after the relay deactivates to return to normal (and thereby preventing re­peated alarm activation). Press ENTER/MENU.
8. Normal state: Set the normal alarm condition as Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to nor­mally open or normally closed upon power up.
9. Select NEXT Press ENTER/MENU to advance to the next setup screen: Configure feed relay.
10. Feed relay: Assign relay 2, 3, or 4 as a feed relay and press ENTER/MENU.
11. Linked to bleed relay: No entry required. The relay originally programmed as the Bleed relay is displayed.
12. Delay time: Enter the time in minutes (X.X min) after the Bleed time is activated before triggering this feed relay.
13. Feed time equals: Enter the percent of time that the Bleed timer is on (X.X%) to activate this feed relay (for pumping make-up water chemicals).
Relay Function Limits and Selections Default
Bleed and Feed NA NA
Logic Low/High High
Setpoint * *
Deadband * 0.000
Normal state Close/Open Open
Feed relay 1, 2, 3, 4 not assigned
Linked to bleed relay 1, 2, 3, 4 1
Delay time 0 - 999.9 min 1.0 min
Feed time equals 0 - 999.9% of bleed time 10.0% of bleed time
Normal state Close/Open Open
* See Appendix 1 – HART and Device Variables
9.5 Totalizer Based Relay Activation
9.5.1 Description:
A totalizer based timer feeds chemicals for a preset period every time a programmed volume of liquid has been added to or removed from a vessel. The relay energizes when the volume has been reached and remains energized for a fixed time. The process repeats once the volume has been reached again.
Totalizer Based Relay Activation triggers a relay at user-defined intervals based on accumulated totalized flow. The scheme uses pulse inputs from a flow meter or 4-20mA current input(s) from a flow transmitter to calculate total flow (as volume).
Alarm Relay Functions 71
Page 82
Section 9: Alarm Relay Functions Instruction Manual
time
relay on
time
X volume
accumulated
X volume
accumulated
April 2017 LIQ-MAN-56
A typical application for totalized flow relay activation is controlling chemical dosing in reactors.
schematic of the Totalizer timer operation is shown:
A
Figure 9-4. Totalizer alarm operation
9.5.2 Setup
Access Totalizer Timers by pressing ENTER/MENU from the main screen and then Program/Relays/ Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Totalizer timer as the relay type and press ENTER/MENU.
3. Assign: Assign Pulse flow S1 (sensor 1) or S2 (sensor 2) as the measurement input and press ENTER/MENU.
4. Active relay after: Enter accumulated volume (XX.XXXX) and units of measurement (gal, thousand gal, million Gal, trillion Gal).
5. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The assigned measurement value will be on hold during this time.
6. Normal state: Set the normal alarm condition as Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
Table 9-5. Defaults and programmable limits
Relay Function Limits and Selections Default
Totalizer Based Timer NA NA
Activate relay after 0 to 99.9990 10
units Gal, Liters, cuft,m3 accumulated E3gal (x1000 gal)
On time 0 to 999.9 min 0 min
Normal state Close/Open Open
accumulated
9.6 Interval Timer
9.6.1 Description:
72 Alarm Relay Functions
The interval timer august be used for periodic sensor cleaning or periodic process adjustment. The cycle begins at the Interval time when the switch is turned on. When the
Page 83
Instruction Manual Section 9: Alarm Relay Functions
time
interval time
on time
hold
recovery time
LIQ-MAN-56 April 2017
Interval time has expired, the analyzer activates hold mode on the assigned measurement and the relay is energized for the On time period.
A schematic of the Interval timer operation is shown:
Figure 9-5. interval timer Alarm operation
9.6.2 Setup
Access Interval timer by pressing ENTER/MENU from the main screen and then Program/Relays/ Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Interval timer as the relay type and press ENTER/MENU.
3. Interval time: Enter the time in hours (XX.X hours) between complete interval cycles.
4. On time: Enter the time in minutes (X.X min) for the relay to remain energized. The assigned measurement value will be on hold during this time.
5. Recovery time: Enter the time in minutes (XX min) before the process is restored and live measurements can resume.
6. *Hold while active: Select which sensors outputs should be on hold (S1, S2 or both) during the interval timer activation time. Press ENTER/MENU.
7. Select NEXT Press ENTER/MENU to advance to the next setup screen.
8. Normal state: Set the normal alarm condition as Open or Closed and press ENTER/MENU. Program the normal state to define the desired alarm default state to normally open or normally closed upon power up.
*56 units with software ver. 2.1X and greater allow override of interval timer to ensure that all relays and outputs are held if desired.
Table 9-6. Defaults and programmable limits
Relay Function Limits and Selections Default
Interval timer NA NA
Alarm Relay Functions 73
Interval time 0 to 999.9 hr 24.0 hr
On time 0 to 999.9 sec 10 sec
Recovery time 0 to 999 sec 60 sec
Hold while active 0 to 999 sec 0 sec
Normal state Sensor 1, Sensor 1, both Sensor 1
Page 84
Section 9: Alarm Relay Functions Instruction Manual
April 2017 LIQ-MAN-56
9.7 Date and Time Activation
9.7.1 Description:
This relay feature allows programming of 1 to 4 relays to activate on an assigned day of the week and time of day or night for an assigned interval. They function like sprinkler timers. The programmable timeframe cycle is two weeks.
An example application for Date and Time Activation is daily chlorine dosing in seawater­cooled condensers.
The Date and Time relay setup screen is shown:
Figure 9-6. Date and Time alarm operation
9.7.2 Setup
Access Date and Time timer by pressing ENTER/MENU from the main screen and then Program/Relays/Configure Relay. From the main relay programming screen, program this feature as follows:
1. Relay: Assign a relay by highlighting the desired relay 1-4 and press ENTER/MENU.
2. Type: Select Date and time timer as the relay type and press ENTER/MENU.
3. Select NEXT Press ENTER/MENU to advance to the next setup screen.
4. The Week 1 calendar will appear. Program the start relay activate time by entering day(s) of week, hour(s) of day and minutes for each hour. Enter the duration of time in minutes (XX min) for relay activation. Up to four relays can be simultaneously energized for any programmed times
5. Select NEXT Press ENTER/MENU to advance to the next setup screen.
6. The Week 2 calendar will appear. Repeat the programming entries for week 2 in the same manner as week 1. Up to four relays can be simultaneously energized for any programmed times.
CAUTION
Date and Time timer operation depends on accurate setup of the internal real time clock. Continu­ous powered operation of the 56 analyzer is recommended to preserve programmed Date and Time timer clock settings.
74 Alarm Relay Functions
Page 85
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
Section 10.0 Calibration
10.1 Calibration – Introduction
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a commercial buffer). The auto-recognition feature of the analyzer will enable the appropriate calibration screens to allow calibration for any single sensor configuration or dual sensor configuration of the analyzer. Completion of Quick Start upon first power up enables live measurements but does not ensure accurate readings in the lab or in process. Calibration should be performed with each attached sensor to ensure accurate, repeatable readings. This section covers the following programming and configuration functions:
1. Auto buffer cal for pH (pH Cal - Sec.7.2)
2. Manual buffer cal for pH (pH Cal - Sec.7.2)
3. Set calibration stabilization criteria for pH (pH Cal - Sec.7.2)
4. Standardization calibration (1-point) for pH, ORP and Redox (pH Cal - Sec.7.2 and 7.3)
5. Entering the cell constant of a conductivity sensor (Conductivity Cal - Sec. 7.4 and 7.5)
6. Calibrating the sensor in a conductivity standard Conductivity Cal - Sec. 7.4 and 7.5)
7. Calibrating the analyzer to a laboratory instrument (Contacting Conductivity Cal - Sec.7.4)
8. Zeroing an chlorine, oxygen or ozone sensor (Amperometric Cal - Sec’s 7.6, 7.7, 7.8)
9. Calibrating an oxygen sensor in air (Oxygen Cal - Sec’s 7.6)
10. Calibrating the sensor to a sample of known concentration (Amperometric Cal - Sec’s
7.6, 7.7, 7.8)
11. Enter a manual reference temperature for temperature compensation of the process measurement
10.2 pH Calibration
New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors. The analyzer recognizes the buffers and uses temperature-corrected pH values in the calibration. Once the 56 successfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25 °C.
To calibrate the pH loop with a connected pH sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Auto Calibration - pH 2 point buffer calibration with auto buffer recognition
2. Manual Calibration - pH 2 point buffer calibration with manual buffer value entry
3. Standardization - pH 1 point buffer
calibration with manual buffer value entry
Calibration 75
Page 86
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
4. Entering A Known Slope Value - pH
lope calibration with manual entry
S of known slope value
1. To Auto Calibrate the pH loop using 2 point buffer calibration with auto buffer recognition, select Auto Buffer and follow the step-by-step proce­dures displayed on-screen.
2. To Manual Calibrate the pH loop using 2 point buffer calibration with manual buffer value entry, select Manual Buffer and follow the step-by-step procedures displayed on-screen.
3. To Standardization Calibrate the pH loop using 1 point buffer calibration with manual buffer value entry, select Standardize and follow the step-by-step procedures displayed on­screen.
4. To Calibrate the pH loop using with manual entry of a Known Slope Value and Reference off­set value, select Slope/Offset and follow the step-by-step procedures displayed on-screen.
10.3 ORP Calibration
For process control, it is often important to make the measured ORP agree with the ORP of a standard solution. During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.
To calibrate the ORP loop with a connected ORP sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routine is covered:
1. To Standardization the ORP loop using 1 point buffer calibration with manual buffer value entry, follow the step-by-step procedures displayed on-screen.
76 Calibration
Page 87
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
10.4 Contacting Conductivity Calibration
ew conductivity sensors rarely need calibration. The cell constant printed on the label is
N sufficiently accurate for most applications.
CALIBRATING AN IN-SERVICE CONDUCTIVITY SENSOR
After a conductivity sensor has been in service for a period of time, recalibration august be necessary. There are three ways to calibrate a sensor.
a. Use a standard instrument and sensor to measure the conductivity of the process stream. It is
not necessary to remove the sensor from the process piping. The temperature correction used by the standard instrument august not exactly match the temperature correction used by the 56. To avoid errors, turn off temperature correction in both the analyzer and the standard instrument.
b. Place the sensor in a solution of known conductivity and make the analyzer reading
match the conductivity of the standard solution. Use this method if the sensor can be easily removed from the process piping and a standard is available. Be careful using standard solutions having conductivity less than 100 µS/cm. Low conductivity standards are highly susceptible to atmospheric contamination. Avoid calibrating sensors with 0.01/cm cell constants against conductivity standards having conductivity greater than 100 µS/cm. The resistance of these solutions august be too low for an accurate measurement. Calibrate sensors with 0.01/cm cell constant using method c.
c. To calibrate a 0.01/cm sensor, check it against a standard instrument and 0.01/cm
sensor while both sensors are measuring water having a conductivity between 5 and 10 µS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide, saturate the sample with air before making the measurements. To ensure adequate flow past the sensor during calibration, take the sample downstream from the sensor. For best results, use a flow-through standard cell. If the process temperature is much different from ambient, keep connecting lines short and insulate the flow cell.
To calibrate the conductivity loop with a connected contacting conductivity sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zero the analyzer with the sensor attached
2. In Process Cal Standardize the sensor to a known conductivity
3. Cell K: 1.00000/cm Enter the cell Constant for the sensor
4. Meter Cal Calibrate the analyzer to a lab conductivity instrument
5. Cal Factor: 0.95000/cm Enter the Cal Factor for 4-Electrode sensors from the sensor tag
Calibration 77
Page 88
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
1. To Zero Calibrate the analyzer with the sensor attached, follow the step-
y-step procedures
b displayed on-screen.
2. To perform an In-Process Calibration of the conductivity loop by Standardizing the sensor to a known conductivity, follow the step-by-step procedures displayed on-screen.
3. To calibrate the conductivity loop by entering a Cell constant, Enter the cell Constant for the sensor and follow the step-by-step procedures displayed on-screen.
4. To Meter Cal Calibrate the analyzer to a lab conductivity instrument, follow the step-by­step procedures displayed on-screen.
5. To enter the Cal Factor to support calibration of a 4-Electrode sensors, enter the Cal Factor for the 4-Electrode sensors from the sensor tag and follow the step-by-step procedures displayed on-screen.
10.5 Toroidal Conductivity Calibration
Calibration is the process of adjusting or standardizing the analyzer to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a conductivity standard). This section contains procedures for the first time use and for routine calibration of the 56 analyzer.
To calibrate the conductivity loop with a connected contacting conductivity sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zero the analyzer with the sensor attached
2. In Process Cal Standardize the sensor to a known conductivity
3. Cell K: 1.00000/cm Enter the cell Constant for the sensor
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on­screen.
2. To perform an In-Process Calibration of the conductivity loop by Standardizing the sensor to a known conductivity, follow the step­by-step procedures displayed on­screen.
3. To calibrate the conductivity loop by entering a Cell constant, Enter the cell Constant for the sensor and fol­low the step-by-step procedures displayed on-screen.
78 Calibration
Page 89
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
10.6 Calibration —Chlorine
ith a Chlorine measurement board and the appropriate sensor, the 56 can measure any of
W four variants of Chlorine:
Free Chlorine
Total Chlorine
Monochloramine
pH-independent Free Chlorine
The section describes how to calibrate any compatible amperometric chlorine sensor. The follow­ing calibration routines are covered in the family of supported Chlorine sensors:
Air Cal
Zero Cal
In Process Cal
10.6.1 Calibration — Free Chlorine
A free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale standard). The zero calibration is necessary because chlorine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution with zero free chlorine
2. Grab Cal Standardizing to a sample of known free chlorine concentration
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on­screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-by-step procedures displayed on-screen.
Calibration 79
Page 90
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
10.6.2 Calibration — Total Chlorine
otal chlorine is the sum of free and combined chlorine. The continuous determination of
T total chlorine requires two steps. First, the sample flows into a conditioning system (TCL) where a pump continuously adds acetic acid and potassium iodide to the sample. The acid lowers the pH, which allows total chlorine in the sample to quantitatively oxidize the iodide in the reagent to iodine. In the second step, the treated sample flows to the sensor. The sensor is a membrane-covered amperometric sensor, whose output is proportional to the concentration of iodine. Because the concentration of iodine is proportional to the concentration of total chlorine, the analyzer can be calibrated to read total chlorine. Because the sensor really measures iodine, calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a solution containing a known amount of iodine (full-scale standard). The Zero calibration is necessary because the sensor, even when no iodine is present, generates a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a total chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because stable total chlorine standards do not exist, the sensor must be calibrated
against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this
purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solu­tion with zero total chlorine
2. Grab Cal Standardizing to a sample of known total chlorine con­centration
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on-screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-by-step procedures displayed on-screen.
80 Calibration
Page 91
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
10.6.3 Calibration — Monochloramine
monochloramine sensor generates a current directly proportional to the concentration
A of monochloramine in the sample. Calibrating the sensor requires exposing it to a solution containing no monochloramine (zero standard) and to a solution containing a known amount of monochloramine (full-scale standard). The Zero calibration is necessary because monochloramine sensors, even when no monochloramine is in the sample, generate a small current called the residual or zero current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a monochloramine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable monochloramine standards do not exist, the sensor must be calibrated
against a test run on a grab sample of the process liquid. Several manufacturers offer portable test
kits for this purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solu­tion with zero total chlorine
2. Grab Cal Standardizing to a sample of known monochloramine concen­tration
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on­screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-by-step pro­cedures displayed on-screen.
Calibration 81
Page 92
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
10.6.4 pH-Independent Free Chlorine Measurement
free chlorine sensor generates a current directly proportional to the concentration of free
A chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale standard). The zero calibration is necessary because chlorine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solu-
tion with zero total chlorine
2. Grab Cal Standardizing to a sample of known chlorine concentration
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on-screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-by-step procedures displayed on-screen.
10.7 Calibration — Oxygen
Oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero standard) and to a solution containing a known amount of oxygen (full-scale standard). The Zero Calibration is necessary because oxygen sensors, even when no oxygen is present in the sample, generate a small current called the residual current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used. The 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low residual current and does not normally require zeroing.The residual current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because the solubility of
82 Calibration
Page 93
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated
ater is difficult to prepare and use, so the universal practice is to use air for calibration. From the
w point of view of the oxygen sensor, air and air-saturated water are identical. The equivalence comes about because the sensor really measures the chemical potential of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same. Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical potential between oxygen in the sensor and oxygen in the sample.
An electrochemical reaction, which destroys any oxygen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sensor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate and the sensor current. When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water­saturated air for calibration. Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air. The analyzer monitors the sensor current. When the current is stable, the analyzer stores the current and measures the temperature using a temperature element inside the oxygen sensor. The user must enter the barometric pressure. From the temperature the analyzer calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxygen, the analyzer calculates the partial pressure of oxygen. Once the analyzer knows the partial pressure of oxygen, it uses the Bunsen coefficient to calculate the equilibrium solubility of atmospheric oxygen in water at the prevailing temperature. At 25 °C and 760 mmHg, the equilibrium solubility is 8.24 ppm. Often it is too difficult or messy to remove the sensor from the process liquid for calibration. In this case, the sensor can be calibrated against a measurement made with a portable laboratory instrument. The laboratory instrument typically uses a membrane-covered amperometric sensor that has been calibrated against water-saturated air.
To calibrate the oxygen sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in a medium with zero oxygen
2. Air Cal Calibrating the sensor in a water-saturated air sample
3. In Process Cal Standardizing to a sample of known oxygen
concentration
4. Sen@ 25 °C:2500 µA/ppm Entering a known slope value for sensor re­sponse.
Calibration 83
Page 94
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
1. To Zero Calibrate the analyzer with the sensor attached, follow the step-
y-step procedures displayed on-
b screen.
2. To Air Cal Calibrating the sensor in a water-saturated air sample, follow the step-by-step procedures dis­played on-screen.
3. To perform an In-Process Calibration by Standardizing the sensor, follow the step-by-step procedures displayed on-screen.
4. To calibrate the oxygen sensor by manually Entering a known slope value for sensor re­sponse, follow the step-by-step procedures displayed on-screen.
10.8 Calibration — Ozone
An ozone sensor generates a current directly proportional to the concentration of ozone in the sample. Calibrating the sensor requires exposing it to a solution containing no ozone (zero standard) and to a solution containing a known amount of ozone (full-scale standard). The Zero Calibration is necessary because ozone sensors, even when no ozone is in the sample, generate a small current called the residual or zero current. The analyzer compensates for the residual current by subtracting it from the measured current before converting the result to an ozone value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because stable ozone standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose.
To calibrate the ozone sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution with zero total chlorine
2. Grab Cal Standardizing to a sample of known ozone concentration
1. To Zero Calibrate the analyzer with the sensor attached, follow the step­by-step procedures displayed on-screen.
2. To perform a Grab Calibration by Standardizing the sensor, follow the step-by-step procedures displayed on-screen.
84 Calibration
Page 95
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
10.9 Calibrating Temperature
ost liquid analytical measurements require temperature compensation (except ORP). The
M 56 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature correction is off, the 56 uses the manual temperature entered by the user in all temperature correction calculations.
To calibrate temperature, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Temperature and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routine is covered:
1. Temperature with manual temperature entry
2. To Calibrate Temperature, follow the step-by-step procedures displayed on-screen.
10.10 Turbidity
This section describes how to calibrate the turbidity sensor against a user-prepared standard as a 2-point calibration with deionized water, against a 20 NTU user-prepared standard as a single point calibration, and against a grab sample using a reference turbidimeter.
To calibrate the turbidity sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routine is covered:
1. Slope Calibration Slope cal with pure water and a standard of known tur­bidity
2. Standardize Calibration Standardizing the sensor to a known turbidity
3. Grab Calibration Standardizing the sensor to a known turbidity based on a reference turbidimeter
Calibration 85
Page 96
Section 10: Calibration Instruction Manual
April 2017 LIQ-MAN-56
1. To calibrate the turbidity loop using Slope Calibration with pure water
nd a standard of known turbidity,
a follow the step-by-step procedures displayed on-screen.
2. To calibrate the turbidity loop using Standardize Calibration by Standard­izing the sensor to a known turbidity, follow the step-by-step procedures displayed on-screen.
3. To calibrate the turbidity loop using Grab Calibration by Standardizing the sensor to a known turbidity based on a reference turbidimeter, follow the step-by-step procedures displayed on-screen.
10.11 Pulse flow
A variety of pulse flow sensors can be wired to the Flow signal input board to measure flow volume, total volume and flow difference (if 2 Flow signal boards are installed). The 56 Flow signal board will support flow sensors that are self-driven (powered by the rotation of the impeller paddle-wheel).
To calibrate the pulse flow sensor, access the Calibration screen by pressing ENTER/MENU from the main screen, select S1 or S2 Measurement and press ENTER/MENU. Press INFO at any time to learn more about this procedure. A yellow screen will appear with detailed instructions and information.
The following calibration routine is covered:
1. K Factor A constant value represent­ing pulses/Gal of flow
2. Frequency/Velocity & Pipe Alternate cal method – requires man­ual entry of frequency (Hz) per veloc­ity and Pipe diameter used
3. In process Calibration based on known volume per unit of time
4. Totalizer Control User settings to stop, restart and reset total volume meter
1. To enter a K Factor constant value representing pulses/Gal of flow, follow the step-by-step procedures displayed on-screen.
2. To calibrate pulse flow Frequency/Velocity & Pipe as an alternate cal method, follow the step-by-step procedures displayed on-screen.
3. To In-process Calibration the pulse flow sensor based on known volume per unit of time, follow the step-by-step procedures displayed on-screen.
4. To stop, restart and reset Totalizer Control total volume meter, follow the step-by-step procedures displayed on-screen.
86 Calibration
Page 97
Instruction Manual Section 10: Calibration
LIQ-MAN-56 April 2017
Section 11.0 HART®Communications
11.1 Introduction
The 56 can communicate with a HART host using HART Revision 5 or HART Revision 7. The revision of HART used by the 56 can be selected using the keypad/display or a HART master such as AMS or the 475 Handheld Communicator. The default version of HART is Revision 5. Since some HART hosts cannot accommodate HART 7, the choice of HART Revision should be based on the capabilities of the host, and should be chosen as a first step in configuration.
When HART 5 is chosen, the Device Revision of the 56 is Device Revision 1; when HART 7 is chosen the Device Revision is Revision 2, or higher, for later revisions of the 56. The Device Revision of the DD (Device Description) and install files for AMS and DeltaV used should be the same as the Device Revision of the 56.
HART 5 Device Identification (56 Revision 1):
Manufacturer Name: Rosemount
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 86 (0x0056)
HART Protocol Revision: 5.1
Device Revision: 1
Capabilities: Supports all signal boards except the turbidity and flow/4-20mA input signal
boards.
HART 7 Device Identification (56 Revision 2):
Manufacturer Name: Rosemount
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 11862 (0x2E56)
HART Protocol Revision: 7.1
Device Revision: 2
Capabilities: Supports all signal boards.
HART 7 Device Identification (56 Revision 3):
Manufacturer Name: Rosemount
Model Name: 56
Manufacturer ID: 46 (0x2E)
Device Type Code: 11862 (0x2E56)
HART Protocol Revision: 7.1
Device Revision: 3
Capabilities: Supports all signal boards and the complete set of parameters for standardized
PID with PID transport time feature and TPC control.
HART Communications 87
Page 98
Section 11: HART Communications Instruction Manual
April 2017 LIQ-MAN-56
11.2 Physical Installation and Configuration
11.2.1 HART Wiring and Output Configuration
ART communications is superimposed on Analog Output 1 for all of the measurements and
H parameters of the 56. The 4-20 mA current of Analog Output 1 can be configured by the keypad display to be powered by the 56 (Output 1 power: Internal), or by an external 24 VDC power supply, or an I/O that provides power (Output 1 power: External).
11.2.2 HART Multidrop (Bus) Configuration
The HART Polling Address should be left at its default value of “0”, unless the 56 is used in a Multidrop configuration with up to 14 other transmitters. When the Polling Address is greater than “0”, the 4-20 mA output is held at 4 mA or below, and does not change in response to changes in the measurement in HART 5.
In HART 7, Loop Current Mode should be set to “On” to hold the current output to a minimum value. In both HART 5 and HART 7, Output Power should be set to “External” so that an external 24 VDC power supply can be used to power the multidrop bus.
11.2.3 HART Configuration
To access the HART Configuration screens, press the “HART” button in the Main Menu. The following controls are available:
HART Configuration Screen 1: Basic Definitions
Host HART mode – toggles between HART version 5 and HART version 7. If the HART
host being used can accommodate HART 7, HART 7 should be chosen due to its larger feature set. If the host can only use HART 5, then HART 5 must be chosen.
NOTE: If the 56 is connected to a HART host and the HART version is changed, the host will likely detect the transmitter as a new transmitter with a different device revision number.
Tag – The traditional 8 character HART tag number
Long tag – HART tag number of up to 32 characters (HART 7 only).
Polling address – Choose “0” unless Multidrop is being used. If Multidrop is being
used, each transmitter should have its own polling address of from 1 to 15.
Loop current mode – Set Output 1 current to a minimum value for multidrop applica-
tions (HART 7 only).
88 HART Communications
Page 99
Instruction Manual Section 11: HART Communications
LIQ-MAN-56 April 2017
Output 1 Power – Select “Internal” to power Output 1 with the 56. Select “External”
to power the current loop with an external power supply, e.g. a host I/O that provides
ower (source) to the transmitter (sink).
p
HART Configuration Screen 2 Basic Definitions
Burst command:
Off – Turns burst mode off
Cmd 1 – Bursts the Primary Value
Cmd 2 – Bursts Loop Current + % of range of the Primary Value
Cmd 3 – Bursts Dynamic Variables (PV, SV, TV, & QV) + Loop Current
Cmd 9 – Bursts up to 8 Device Variables with time stamp and status (HART 7 only)
Cmd 33 – Bursts 4 Device Variables
Cmd 48 – Bursts Additional Transmitter Status Bits (HART 7 only)
Cmd 93 – Bursts Trend Data (HART 7 only)
Find device cmd – Setting Find Device to “On”, enables the 56 to be indentified by the
host. The transmitter returns identity information including device type, revision level, and device ID.
Response preambles – Preambles synchronize the receiver to the incoming data. Re-
sponse preambles are the number of bytes of preambles (2 to 20) sent by the 56 at the start of a response packet.
11.3 Measurements Available via HART
A number of live measurements are made available by HART in addition to the main measurements such as pH or Conductivity. All of these measurements are called Device Variables, which can be mapped to the Dynamic Variables PV, SV, TV, and QV for regular reading by the typical HART host.
The 56 assigns the Dynamic Variables PV, SV, TV, and QV to Analog Outputs 1, 2, 3, and 4 respectively. Conversely, measurements assigned to Outputs 1 through 4, will automatically be assigned to PV through QV.
Each measurement board will have its own set of Device Variables, based on the secondary measurements used in making the main measurement. Appendix 1 shows the Device Variable for the each sensor boards used in the 56, and the Dynamic Variables, which they can be mapped to.
HART Communications 89
Page 100
Section 11: HART Communications Instruction Manual
April 2017 LIQ-MAN-56
11.4 Diagnostics Available via HART
11.4.1 Status Information -- Device Status Bits
Bit 0 Primary Variable out of Limits:
This bit is set when PV is out of its limits.
Bit 1 Non-primary Variable out of Limits:
This bit is set when non PV is out of its limits.
Bit 2 Loop Current Saturated:
This bit is set when the Analog Output 1 current is less than 1.0 mA or greater than
22.0 mA and the Device Status Bytes bit 3 is not set.
Bit 3 Loop Current Fixed:
This bit is set when Analog Output 1 does not follow the process. This bit is cleared when Analog Output 1 follows the process.
Bit 4 More Status Available:
The “more status available” bit will be set when the device status condition occurs (i.e. bit goes from 0 to 1) on at least one of the Additional Transmitter Status bits are set.
Bit 5 Cold Start:
This bit is set when a Master Reset is performed either by Command 42, or a power cycle. This bit is cleared after the first response or burst. In the case of a burst, the first burst always goes to the primary master.
Bit 6 Configuration Changed:
This bit is set when the last bytes of an EEPROM writing sequence is completed. Since EEPROM writes are delayed until after the response, the immediate acknowledgement will not have the bit set, however, after the EEPROM write completes, the bit will be set. This bit applies to all EEPROM writes whether or not they apply directly to the configuration of the instrument or not. This bit is cleared when Command 38 is executed.
Bit 7 Field Device Malfunction:
This bit is set when any of the following conditions is true, and cleared when all of the following conditions are false:
a) There is at least one main board fault.
b) There is at least one Sensor 1 fault.
c) There is at least one Sensor 2 fault.
11.4.2
90 HART Communications
Status Information – Extended Device Status Bits (HART 7 only)
Bit 0 Maintenance Required:
Any calibration error will set this bit. Any calibration for either Sensor 1 or Sensor 2 that fails will set this bit. This bit gets cleared if all device variables are calibrated successfully.
This bit is set to indicate that, while the device has not malfunctioned, the Field Device requires maintenance.
Bit 1 Device Variable Alert:
This bit will get set when at least one of device status byte of all the valid Device Variables does not equal to "good and not limited" (i.e. the higher 4 bits not equal to
Loading...