Rosemount Analytical 54e, MCL Specifications

Page 1
Product Data Sheet
71-54e/rev.H January 2006
CONDUCTIVITY, pH/ORP, & AMPEROMETRIC HART
®
ANALYZERS/CONTROLLERS
Model 54e
• NEMA 4X (IP65) WEATHERPROOF CORROSION-RESISTANT ENCLOSURE.
• BACK-LIT DISPLAY with easy-to-use interface.
• TWO INDEPENDENT OUTPUTS.
• THREE ALARMS WITH PROGRAMMABLE LOGIC, plus one dedicated fault alarm.
• OPTIONAL TPC AND PID CONTROL CAPABILITY.
Model 54e pH/ORP Analyzer/Controller
COMPREHENSIVE pH GLASS DIAGNOSTICS1warn user of the need for calibra­tion, maintenance, or sensor replacement.
AUTOMATIC CALIBRATION with buffer recognition and stabilization check.
SELECTIVE ION CAPABILITY.
1
U.S. Patent No. 5,469,070.
Model 54eA Amperometric Analyzer/Controller
• MEASURES dissolved oxygen (ppm & ppb level), free chlorine, total chlorine, and ozone.
• ON-BOARD PRESSURE SENSOR for automatic air-calibration of oxygen sensors.
• SECOND INPUT FOR pH SENSOR ALLOWS AUTOMATIC pH CORRECTION for free chlorine measurement. No expensive reagents needed.
• AUTOMATIC BUFFER RECOGNITION for pH calibration.
• INTERVAL TIMER operates spray cleaner for oxygen sensor in aeration basins.
Model 54eC Conductivity/Resistivity Analyzer/Controller
MEASURES Conductivity, Resistivity, or Percent (%) Concentration.
USES EITHER CONTACTING OR INDUCTIVE SENSORS to meet most application requirements.
PERCENT CONCENTRATION CURVES FOR 0-12% NaOH, 0-15% HCl, and 0-25% or 96-99.7% H2SO4.
TEMPERATURE CORRECTION ALGORITHMS for high purity water, cation conduc­tivity, & linear temperature coefficient. Temperature correction can also be turned off.
Page 2
FEATURES AND APPLICATIONS—GENERAL
The Model 54e family of analyzers, with the appropriate sensors, monitor and control pH/ORP, conductivity/resis­tivity, dissolved oxygen (ppm and ppb level), free chlorine, total chlorine, or ozone in a variety of process liquids.
The analyzer/controller is housed in a rugged NEMA 4X (IP65) weatherproof, corrosion resistant enclosure of epoxy-painted aluminum. It is suitable for panel, pipe or wall mounting. All functions are accessible through the front panel membrane keyboard that features tactile feed­back. The large back-lit dot-matrix liquid crystal display continuously indicates the measured variable in large numerals, along with the temperature, output value and two programmable process parameters (such as alarms or diagnostics variables). Three levels of security are avail­able: 1) calibration, 2) output range and alarm setpoint adjustment, and 3) advanced configuration. A hinged front panel provides convenient wiring access. All operations and descriptive messages can be field selected for Eng­lish, French, German, Italian, or Spanish.
The Model 54e is a member of the Rosemount SMART FAMILY®of instruments which is designed to communi-
cate with the Model 275 HART
®
hand-held communicator and any other host, including AMS, that supports the 275 HART communication protocol.
The two independent, isolated outputs provide 4-20 or 0-20 mA signals for the process measurement and tem­perature. The controller option also allows PID control choices acting on process measurement or temperature. Manual control of the output may also be selected for routine maintenance and tests.
Three process alarms are a standard feature. All alarms have programmable high or low activation, independent setpoints, adjustable hysteresis (deadband), and time delay. One relay may be configured as a timer relay for chemical cleaning, which may be activated at a specific time interval. An overfeed timer feature is also selectable for any one process relay. The control option allows each alarm the additional choice of Time Proportional Control. During any relay activation, a corresponding Green LED is illuminated.
®
SMART FAMILY is a registered trademark of Rosemount Inc.
®
HART is a registered trademark of the HART Communication Foundation.
SPECIFICATIONS - GENERAL
Enclosure: Epoxy-painted (dark gray) cast aluminum,
NEMA4X (IP65). 144 x 144 x 132 mm (5.7 x 5.7 x 5.2 in.), DIN size panel cut-out.
Front Panel: Membrane keypad with tactile feedback.
Three green LEDs indicate alarm status. Red LED indicates fault condition.
Display: Three-line, back-lit, dot matrix LCD, 70 x 35 mm.
First line is measurement reading. Second line is tem­perature and current output. Third line is user-selec­table. Character heights: 1st line - 16 mm (0.6 in.), 2nd and 3rd lines - 7 mm (0.3 in.).
Hazardous Location Approvals:
Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
FM: Max. relay contact rating: 28 Vdc
resistive
150 mA - Groups A & B; 400 mA - Group C; 540 mA - Group D
CSA:
Max. relay contact rating: 28 Vdc; 110 Vac; 230 Vac; 6 amps resistive. Enclosure Type 4.
Power:
Code -01: 100 - 127 Vac, 50/60 Hz ± 6%, 8 W
200 - 253 Vac, 50/60 Hz ± 6%, 8 W
Code -02: 20 - 30 VDC, 6 W
RFI/EMI: EN-61326
LVD (Code -01 only): EN-61010-1
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Outputs can be assigned to the primary measurement, glass impedance (pH only), reference impedance (pH only), or temperature. Output dampening is user-selectable. Maximum load at 24 Vdc or 115/230 Vac is 500 ohms. Maximum load at 100/200 Vac is 500 ohms. Output 1 has superimposed HART signal (option -09 only). Outputs can be programmed for PID control (option
-20 only).
Output Accuracy: ± 0.05 mA
Alarms:
Relay 1 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 4 - Sensor/analyzer and process fault alarm
Each relay has a dedicated LED on the front panel.
Relay Contacts: Relays 1-3: Epoxy sealed form A
contacts, SPST, normally open
Relay 4: Epoxy sealed form C, SPDT
Resistive
Inductive 28 Vdc 5.0 Amps 3.0 Amps 115 Vac 5.0 Amps 3.0 Amps 230 Vac 5.0 Amps 1.5 Amps
Temperature Sensors Accepted: Pt 100 RTD,
Pt 1000 RTD, 22K NTC thermistor (54eA only).
Ambient Temperature: 0 to 50°C (32 to 122°F). Analyzer
can be operated between -20 and 60°C (-4 to 140°F) with some degradation in display quality.
Relative Humidity: 95% (maximum) non-condensing Weight/Shipping Weight: 4 lb/5 lb (1.8 kg/2.3 kg)
Page 2
Page 3
Page 3
PANEL MOUNTING DIMENSIONS
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 4
Page 4
PIPE AND WALL MOUNTING DIMENSIONS
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 5
Page 5
CODE MEASUREMENT SELECTION
pH/ORP pH/ORP
C Conductivity/Resistivity
A Amperometric (Chlorine, Dissolved Oxygen, and Ozone)
MODEL
54e MICROPROCESSOR ANALYZER
CODE REQUIRED SELECTION
01 115/230 VAC, 50/60 Hz Power 02 24 VDC Power
CODE OPTIONS
09 HART Communications Protocol 20 Controller Outputs - PID and TPC
ACCESSORIES
PART NO. DESCRIPTION
2002577 Wall and two inch pipe mounting kit
23545-00 Panel mounting kit
23554-00 Cable glands, kit (Qty 5 of PG 13.5)
9240048-00 Stainless steel tag (specify marking)
ORDERING INFORMATION
54e pH/ORP -01 -20 EXAMPLE
Page 6
Page 6
MODEL 54e pH/ORP ANALYZER
FEATURES AND APPLICATIONS
The Model 54e pH/ORP Microprocessor Analyzer, with the appropriate sensor, is designed to monitor and control pH or ORP in industrial and municipal processes using the most advanced technology available.
The continuous diagnostic capability of the analyzer elimi­nates troubleshooting guesswork by alerting the user to the following problems:
In the event of one of these conditions, the analyzer will display a descriptive message of the problem, illuminate a red LED on the front panel, and drive the outputs to prese­lected fault mode values. A dedicated fault relay may be used to operate a visible or audible alarm.
The glass impedance is continuously monitored and can be programmed to read out on the main display. Fault set­points for the glass impedance are programmable. The
glass impedance is temperature compensated for the greatest accuracy.
The automatic buffer recognition feature uses stored buffer values and their temperature curves for the most common buffer standards available worldwide. The analyzer will recognize the last two buffers used for the two point cali­bration, and perform a self-stabilization check on the sen­sor in each buffer.
Three process alarms are a standard feature. All alarms have programmable high or low activation, independent setpoints, adjustable hysteresis (deadband), and time delay. One relay may be configured as a timer relay for chemical cleaning, which may be activated at a specific time interval. An overfeed timer feature is also selectable for any one process relay. The control option allows each alarm the additional choice of Time Proportional Control. During any relay activation, a corresponding Green LED is illuminated.
The Model 54e pH/ORP is compatible with either an exter­nal Pt 1000 or a Pt 100 RTD located in the sensor to com­pensate the pH reading for process temperature fluctuations. Manual temperature compensation is key­board selectable. Processes with pH changes due to tem­perature can be compensated using a programmable temperature coefficient or isopotential point.
• Cracked or broken pH elec­trode
• Aged glass - calibration rec­ommended
• Worn out or non- immersed electrode
• Poisoned reference (off-line only)
• Faulty temperature element
• Open wiring
• pH electrode slope error (off-line)
• Analyzer electronics failure.
EASY TO USE MENU-DRIVEN INTERFACE
AUTO CALIBRATION with buffer recognition and stabi-
lization check removes calibration guesswork.
DIAGNOSTIC MESSAGES are clearly spelled out on the main display and on all other screens.
4.01
pH
Autocal buffer 1
Exit Buf 2 Enter
7.10
pH
25.0°C 12.1 mA
Cracked glass failure
In the screen above, pressing Enter (Function key #4) initiates the automatic calibration procedure for buffer 1.
Page 7
Page 7
MODEL 54e pH/ORP ANALYZER
The Model 54e when configured as a pH analyzer,
requires a dual (glass and reference) impedance pream­plifier (patent pending). This preamp converts the high impedance pH glass electrode signal to a low impedance signal. The preamplifier may be located in one of three areas: 1) in the pH sensor for best performance, 2) in a remote junction box when process temperatures exceed 80°C (176°F) in submersion applications, or 3) in the ana­lyzer (included in circuitry) when the distance between the pH sensor and the analyzer is 4.5 meters (15 feet) or less.
The Model 54e pH measures over the full range of 0-14 pH. The current output may be calibrated to represent any 1 to 14 pH span.
A two-point calibration is made by immersing the sensor in two different buffer solutions and entering the pH values. When two buffers are used, the microprocessor automati­cally calculates the electrode slope. The electrode slope can be read on the display and manually adjusted. A one­point process standardization is easily accomplished by entering the pH value of a grab sample.
INSTRUMENT SPECIFICATIONS @ 25°C
Measurement Range: 0 to 14 pH Output Scale Expansion:
Zero suppression: up to 13 pH units Span: Any pH from 1 to 14
Accuracy: ± 0.01 pH Repeatability: ± 0.01 pH Stability: ± 0.01 pH/month, non-cumulative Temperature Coefficient: Input: ± 0.003 pH/°C
Output: ± 0.006 pH/°C
Temperature Compensation:
Pt 100 or Pt 1000 RTD,
Automatic or Manual; –15 to 120°C (5 to 248°F)
RECOMMENDED SENSORS:
Model 320B Flow Through pH Model 320HP High Purity pH *Model 328A Steam Sterilizable pH *Model 370 and 371 EuroSenz pH Model 381+ Insertion/Submersion/Flow Through pH *Model 389 Disposable pH *Model 396 Disposable pH *Model 396VP Disposable pH with VP 6.0 connector Model 396P Disposable pH Model 396PVP Disposable pH Model 396R Retractable pH Model 396RVP Retractable pH with VP 6.0 connector Model 397 Quik Disconnect pH Model 398 Insertion/Submersion pH Model 398VP Insertion/Submersion with VP 6.0 connector *Model 398R Retractable pH *Model 398RVP Retractable pH with VP 6.0 connector *Model 399 Disposable pH *Model Hx338 Steam Sterilizable pH *Model Hx348 Steam Sterilizable pH
The Model 54e when configured as an ORP analyzer, measures over a range of -1400 mV to +1400 mV in either the American convention (Oxidation Reduction Potential), or the European convention (Reduction Oxidation-Redox). Although temperature compensation is not used for ORP measurements, the process temperature is measured and displayed. Temperature measurement is made by an RTD located in the sensor assembly.
INSTRUMENT SPECIFICATIONS @ 25°C
Measurement Range: –1400 to +1400 mV
Output Scale Expansion:
Zero suppression: up to ±1300 mV
Span: Any ORP range from 100 to 2800 mV
Accuracy: ± 1.0 mV
Repeatability: ± 1.0 mV
Stability: ± 1.0 mV/month, non-cumulative
Temperature Coefficient: Input: ± 0.2 mV/°C
Output: ± 0.4 mV/°C
Temperature Measurement: –15 to 120°C (5 to 248°F)
Pt 100 or Pt 1000 RTD
RECOMMENDED SENSORS:
Model 330 Flow Through ORP *Model 371 EuroSenz ORP Model 381+ Insertion/Submersion/Flow Through ORP *
Model 389 Disposable ORP Model 396P Disposable ORP Model 396R Retractable ORP Model 398 Insertion/Submersion ORP Model 398VP Insertion/Submersion with VP 6.0 connector Model 398R Retractable ORP Model 398RVP Retractable ORP with VP 6.0 connector
The Model 54e when ordered for ISE capability, is suitable for use with a number of ion-selective electrodes. Consult the factory for available measurements and ranges.
* Reference diagnostics not available with these sensors
Page 8
MODEL 54eC CONDUCTIVITY ANALYZER
Page 8
FEATURES AND APPLICATIONS
The Model 54e C Analyzer/Controller, with the appro­priate sensor, monitors and controls conductivity/ resistivity in a variety of industrial processes. The analyzer can use either a contacting electrode or an inductive (toroidal) conductivity sensor, selected via keyboard entry. Process measurement selections are conductivity, resistivity, raw conductivity, or percent (%) concentration.
The temperature correction algorithm is user-selec­table. Choose from high purity water (dilute sodium chloride), cation conductivity (dilute hydrochloric acid), or an adjustable linear temperature coefficient (0 to 5.00%/°C). The high purity water and cation conductivity corrections are good to 100°C. The lin-
ear temperature correction applies between -5 to 200°C. The user can set the linear temperature coef­ficient's reference temperature.
The percent concentration measurement includes the choice of four common solutions (0 to 12% NaOH, 0 to 15% HCl, and 0 to 25% or 96 to 99.7% H2SO4) which include specific nonlinear temperature correc­tions that depend on the concentration being meas­ured and the process temperature. A versatile data entry menu for entry of a custom solution with con­centration and conductivity values for up to five points lets the analyzer generate a best fit quadratic curve. Reference temperature and temperature slope may be adjusted to provide optimum results.
INSTRUMENT SPECIFICATIONS @ 25°C
Measurement Range: –15 to 200°C (5 to 392°F)
Contacting: 0-20,000 µS/cm Toroidal: 0-2 S/cm
Accuracy of Analyzer: (Analyzer connected to simulated sensor input)
Contacting Sensors: ±0.5% of reading, ±0.5 µS/cm Inductive Sensors: ±1% of reading, 200 µS/cm to 2 S/cm, ± 5 µS/cm
Repeatability: ±0.25% of reading Stability: ±0.25% of output range/month, noncumulative Ambient Temperature Coefficient: ±0.01% of reading/°C Temperature Compensation: -15 to 200°C (5 to 392°F) (automatic or manual) Temperature Correction: High purity water (dilute sodium chloride), cation conductivity (dilute hydrochloric
acid), linear temperature coefficient (0.0 to 5.00%/°C), or none. High purity water and cation conductivity temperature correction apply between 0 and 100°C. Linear temperature coefficient can be applied between
-5 and 200°C (23 to 392°F).
EASY TO USE MENU-DRIVEN INTERFACE
CUSTOM % CONCENTRATION CURVE is easy to
set up and update.
CONTROL SET-UP can be accomplished by entering four parameters from the control menu (code option -20 only).
Setpoint: 1000 µS/cm
Proportional: 100.0% Integral: 0 sec
Exit Edit
Number of points: 3
Ref temp: 25°C
Linear slope: 2.00%/°C
Exit Edit
Page 9
Page 9
MODEL 54eC CONDUCTIVITY ANALYZER
CONTACTING SENSORS
Conductivity Sensor 142, 400 142, 400 140, 141
Model Number 402, 403, 404 402, 403, 404 400, 402, 403
Cell Constant (/cm) 0.01 0.1 1.0
Recommended Conductivity 0-25 1-2000 10-10,000**
Range* (µµS/cm)
INDUCTIVE SENSORS
Conductivity Sensor
Model Number 226 228 225 222 (1in.) 222 (2 in.) 242
Nominal Cell Constant 1.0 3.0 3.0 6.0 4.0 *
Minimum Conductivity (µµS/cm) 50 200 200 500 500 100*
Maximum Conductivity (µµS/cm) 1,000,000 2,000,000 2,000,000 2,000,000 2,000,000 2,000,000*
* Model 242 values depend on sensor configuration and wiring.
* For sensor linearity equal to or better than 1% with ENDURANCE series. ** ENDURANCE sensors with cell constant of 1.0/cm may be used for conductivity up to
20,000 µS/cm with linearity equal to or better than 2%.
SENSOR CHOICE GUIDELINES
The Model 54eC is compatible with both contacting and inductive conductivity sensors. The best sensor for an application depends on many factors, among them are the conductivity to be measured, the compatibility of the sensor's wetted materials with the process chemicals and conditions, and the mounting arrangement. The tables below are provided as an aid for choosing an appropriate sensor.
RECOMMENDED SENSORS:
Model 140 Retractable Conductivity Model 141 Insertion Conductivity Model 142 Insertion Conductivity Model 222 Inductive Conductivity Model 225 Inductive Conductivity Model 226 Inductive Conductivity Model 228 Inductive Conductivity Model 242 Inductive Conductivity
Model 400 Screw-in Conductivity Model 401-14 Screw-in Conductivity Model 402 Retractable Conductivity Model 403 Conductivity with Sanitary Fitting Model 404 Flow Cell Conductivity Model 400VP Screw-in Conductivity Model 402VP Retractable Conductivity Model 403VP Conductivity with Sanitary Fitting
Page 10
MODEL 54eA AMPEROMETRIC ANALYZER
Page 10
The Model 54eA Analyzer/Controller with the appropri­ate sensor monitors and controls dissolved oxygen (ppm and ppb level), free chlorine, total chlorine, and ozone in a variety of process liquids. The analyzer is compatible with Rosemount Analytical series 499A amperometric sensors for oxygen, chlorine, and ozone; and with Hx438 and Gx448 steam sterilizable oxygen sensors.
For oxygen measurements, an on-board pressure sensor allows truly automatic air calibration. Simply expose the sensor to water-saturated air. Wait until readings are stable and press a button. The analyzer measures temperature and barometric pressure and automatically completes the calibration.
For free chlorine measurements, both automatic and manual pH correction are available. pH correction is necessary because amperometric chlorine sensors respond only to hypochlorous acid. To measure free chlorine (hypochlorous acid plus hypochlorite ion) most competing instruments require an acidified sam-
ple. Acid lowers the pH and converts hypochlorite ion to hypochlorous acid. The 54eA eliminates the need for messy and expensive reagents by measuring the sample pH and using it to correct the chlorine sensor signal. If the pH is relatively constant, a fixed pH cor­rection can be used. If the pH is greater than 7 and fluctuates more than about 0.2 units, continuous measurement of pH and automatic pH correction is necessary. Corrections are valid to pH 9.5.
The 54eA analyzer fully compensates oxygen, ozone, free chlorine, and total chlorine readings for changes in membrane permeability caused by temperature changes.
For pH measurements — pH is available with free chlorine only — the 54eA features automatic buffer recognition with stabilization check. Buffer pH and temperature data for commonly used buffers are stored in the analyzer. Glass impedance diagnostics warn the user of an aging or failed pH sensor.
EASY TO USE MENU-DRIVEN INTERFACE
CALIBRATING AN OZONE OR CHLORINE SENSOR IS SIMPLE
1.100 ppm
Stabilizing . . . Wait
Abort Cont
With the cursor on “Calibrate main sensor,” press Enter (key F4).
The analyzer starts monitoring sen­sor current. The Model 54eA ana­lyzer will not accept the calibration until readings are stable. Stability limits are user-selectable.
Once the reading is stable, take a grab sample of the process liquid and use a portable test kit to deter­mine the chlorine or ozone level. Press Edit (key F4). Use the arrow keys to change the reading to match the results of the grab sam­ple test.
Calibrate main sensor
Zero main sensor Adjust temperature
Exit Enter
CALIBRATING A DISSOLVED OXYGEN SENSOR TAKES ONLY A FEW KEY PRESSES
Press Enter (key F4) to start the calibration.
Calibrate in air
Calibrate in process Barometric pressure
Exit Enter
The display shows the current temperature and barometric pressure. Press Continue (key F3). The analyzer automatically com­pletes the calibration.
Bar press : 760 mmHg
Temperature: 25.0°C Press F3 to air cal
Abort Cont
1.100 ppm
Calibrate : 1.100 ppm
Exit Edit
Page 11
Page 11
MODEL 54eA AMPEROMETRIC ANALYZER
SPECIFICATIONS — OXYGEN
Measurement Range: 0-99 ppm (mg/L), 0-200%
saturation
Resolution: 0.01 ppm, 0.1 ppb for 499A TrDO
sensor
Temperature correction for membrane perme-
ability: automatic between 0 and 50°C (can be
disabled)
Calibration: automatic air calibration or calibration
against a standard instrument
Pressure sensor range: 113 to 862 mmHg (151 to
1149 mbar)
RECOMMENDED SENSORS — OXYGEN:
Model 499A DO-54 for ppm level Model 499A TrDO-54 for ppb level Hx438 and Gx448 steam-sterilizable oxygen sensors
SPECIFICATIONS — FREE CHLORINE
Measurement Range: 0-20 ppm (mg/L) as Cl
2
Resolution: 0.001 ppm Temperature correction for membrane perme-
ability: automatic between 0 and 50°C (can be
disabled)
pH Correction: Automatic between pH 6.0 and 9.5.
Manual pH correction is also available.
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — FREE CHLORINE:
Model 499A CL-01-54
SPECIFICATIONS — pH
Application: pH measurement available with free
chlorine only.
Measurement Range: 0-14 pH Resolution: 0.01 pH Sensor Diagnostics: Glass impedance (for broken
or aging electrode) and reference offset. Refer­ence impedance (for fouled reference junction) is not available.
Repeatability: ±0.01 pH at 25°C Stability: ±0.01 pH/month, non-cumulative at 25°C
RECOMMENDED pH SENSORS:
Model 399-09-62, 399-14, and 399VP-09 See pH sensor product data sheet for complete order-
ing information.
SPECIFICATIONS — TOTAL CHLORINE
Measurement Range: 0-20 ppm (mg/L) as Cl
2
Resolution: 0.001 ppm Temperature correction for membrane perme-
ability: automatic between 5 and 35°C (can be
disabled)
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — TOTAL CHLORINE:
Model 499A CL-02-54 (must be used with SCS 921)
SPECIFICATIONS — OZONE
Measurement Range: 0-10 ppm (mg/L) Resolution: 0.001 ppm Temperature correction for membrane perme-
ability: automatic between 5 and 35°C (can be
disabled)
Calibration: against grab sample analyzed using
portable test kit.
RECOMMENDED SENSOR — OZONE:
Model 499A OZ-54
Page 12
ALSO AVAILABLE FROM ROSEMOUNT ANALYTICAL —
FOUNDATION®Fieldbus Transmitters
The Model 5081 has obtained FOUNDATION Fieldbus certification and is now authorized to bear the FOUNDATION Fieldbus registration checkmark. The Model 5081 transmitter successfully completed rigorous Fieldbus Foundation testing, including stack communication, physical layer conformance, and user layer interoperability testing.
Two digital communication protocols are available with Model 5081 transmitters: HART and F
OUNDATION
Fieldbus. Digital communications allows access to AMS (Asset Management Solutions). AMS can be used to set up and configure the transmitter, read process variables, and troubleshoot problems from a personal computer or host anywhere in the plant.
This transmitter features the latest in technology providing predictive pH sensor diagnostics and a large easy-to­read display. Ideal for harsh environments, the Model 5081 is housed in a rugged enclosure and is designed to meet NEMA 4X and NEMA 7B requirements.
Model 5081pH
pH/ORP Transmitter
• Large two-line display shows pH or ORP (mV) and temperature.
• Automatic two-point buffer calibration reduces errors.
• Solution temperature compensation allows pH adjustment for actual pH changes with temperature.
• Continuous diagnostics monitor sen­sor performance and warn the user of impending problems.
Model 5081A
Amperometric
Transmitter
• Measures dissolved oxygen (ppm and ppb level), free chlorine, total chlo­rine, and ozone.
• Second input for pH sensor allows automatic pH correction for free chlorine measurement. No expen­sive reagents needed.
• Automatic buffer recognition for pH calibration.
Model 5081C Conductivity
Transmitter
• Large two-line dis­play shows conduc­tivity and temperature.
• High loop accuracy (within ±1% of actual conductivity and resistivity measurement).
Credit Cards for U.S. Purchases Only.
The right people, the right answers, right now.
ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.