Rosemount Analytical 54e, MCL Specifications

Product Data Sheet
71-54e/rev.H January 2006
CONDUCTIVITY, pH/ORP, & AMPEROMETRIC HART
®
ANALYZERS/CONTROLLERS
Model 54e
• NEMA 4X (IP65) WEATHERPROOF CORROSION-RESISTANT ENCLOSURE.
• BACK-LIT DISPLAY with easy-to-use interface.
• TWO INDEPENDENT OUTPUTS.
• THREE ALARMS WITH PROGRAMMABLE LOGIC, plus one dedicated fault alarm.
• OPTIONAL TPC AND PID CONTROL CAPABILITY.
Model 54e pH/ORP Analyzer/Controller
COMPREHENSIVE pH GLASS DIAGNOSTICS1warn user of the need for calibra­tion, maintenance, or sensor replacement.
AUTOMATIC CALIBRATION with buffer recognition and stabilization check.
SELECTIVE ION CAPABILITY.
1
U.S. Patent No. 5,469,070.
Model 54eA Amperometric Analyzer/Controller
• MEASURES dissolved oxygen (ppm & ppb level), free chlorine, total chlorine, and ozone.
• ON-BOARD PRESSURE SENSOR for automatic air-calibration of oxygen sensors.
• SECOND INPUT FOR pH SENSOR ALLOWS AUTOMATIC pH CORRECTION for free chlorine measurement. No expensive reagents needed.
• AUTOMATIC BUFFER RECOGNITION for pH calibration.
• INTERVAL TIMER operates spray cleaner for oxygen sensor in aeration basins.
Model 54eC Conductivity/Resistivity Analyzer/Controller
MEASURES Conductivity, Resistivity, or Percent (%) Concentration.
USES EITHER CONTACTING OR INDUCTIVE SENSORS to meet most application requirements.
PERCENT CONCENTRATION CURVES FOR 0-12% NaOH, 0-15% HCl, and 0-25% or 96-99.7% H2SO4.
TEMPERATURE CORRECTION ALGORITHMS for high purity water, cation conduc­tivity, & linear temperature coefficient. Temperature correction can also be turned off.
FEATURES AND APPLICATIONS—GENERAL
The Model 54e family of analyzers, with the appropriate sensors, monitor and control pH/ORP, conductivity/resis­tivity, dissolved oxygen (ppm and ppb level), free chlorine, total chlorine, or ozone in a variety of process liquids.
The analyzer/controller is housed in a rugged NEMA 4X (IP65) weatherproof, corrosion resistant enclosure of epoxy-painted aluminum. It is suitable for panel, pipe or wall mounting. All functions are accessible through the front panel membrane keyboard that features tactile feed­back. The large back-lit dot-matrix liquid crystal display continuously indicates the measured variable in large numerals, along with the temperature, output value and two programmable process parameters (such as alarms or diagnostics variables). Three levels of security are avail­able: 1) calibration, 2) output range and alarm setpoint adjustment, and 3) advanced configuration. A hinged front panel provides convenient wiring access. All operations and descriptive messages can be field selected for Eng­lish, French, German, Italian, or Spanish.
The Model 54e is a member of the Rosemount SMART FAMILY®of instruments which is designed to communi-
cate with the Model 275 HART
®
hand-held communicator and any other host, including AMS, that supports the 275 HART communication protocol.
The two independent, isolated outputs provide 4-20 or 0-20 mA signals for the process measurement and tem­perature. The controller option also allows PID control choices acting on process measurement or temperature. Manual control of the output may also be selected for routine maintenance and tests.
Three process alarms are a standard feature. All alarms have programmable high or low activation, independent setpoints, adjustable hysteresis (deadband), and time delay. One relay may be configured as a timer relay for chemical cleaning, which may be activated at a specific time interval. An overfeed timer feature is also selectable for any one process relay. The control option allows each alarm the additional choice of Time Proportional Control. During any relay activation, a corresponding Green LED is illuminated.
®
SMART FAMILY is a registered trademark of Rosemount Inc.
®
HART is a registered trademark of the HART Communication Foundation.
SPECIFICATIONS - GENERAL
Enclosure: Epoxy-painted (dark gray) cast aluminum,
NEMA4X (IP65). 144 x 144 x 132 mm (5.7 x 5.7 x 5.2 in.), DIN size panel cut-out.
Front Panel: Membrane keypad with tactile feedback.
Three green LEDs indicate alarm status. Red LED indicates fault condition.
Display: Three-line, back-lit, dot matrix LCD, 70 x 35 mm.
First line is measurement reading. Second line is tem­perature and current output. Third line is user-selec­table. Character heights: 1st line - 16 mm (0.6 in.), 2nd and 3rd lines - 7 mm (0.3 in.).
Hazardous Location Approvals:
Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
FM: Max. relay contact rating: 28 Vdc
resistive
150 mA - Groups A & B; 400 mA - Group C; 540 mA - Group D
CSA:
Max. relay contact rating: 28 Vdc; 110 Vac; 230 Vac; 6 amps resistive. Enclosure Type 4.
Power:
Code -01: 100 - 127 Vac, 50/60 Hz ± 6%, 8 W
200 - 253 Vac, 50/60 Hz ± 6%, 8 W
Code -02: 20 - 30 VDC, 6 W
RFI/EMI: EN-61326
LVD (Code -01 only): EN-61010-1
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Outputs can be assigned to the primary measurement, glass impedance (pH only), reference impedance (pH only), or temperature. Output dampening is user-selectable. Maximum load at 24 Vdc or 115/230 Vac is 500 ohms. Maximum load at 100/200 Vac is 500 ohms. Output 1 has superimposed HART signal (option -09 only). Outputs can be programmed for PID control (option
-20 only).
Output Accuracy: ± 0.05 mA
Alarms:
Relay 1 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 4 - Sensor/analyzer and process fault alarm
Each relay has a dedicated LED on the front panel.
Relay Contacts: Relays 1-3: Epoxy sealed form A
contacts, SPST, normally open
Relay 4: Epoxy sealed form C, SPDT
Resistive
Inductive 28 Vdc 5.0 Amps 3.0 Amps 115 Vac 5.0 Amps 3.0 Amps 230 Vac 5.0 Amps 1.5 Amps
Temperature Sensors Accepted: Pt 100 RTD,
Pt 1000 RTD, 22K NTC thermistor (54eA only).
Ambient Temperature: 0 to 50°C (32 to 122°F). Analyzer
can be operated between -20 and 60°C (-4 to 140°F) with some degradation in display quality.
Relative Humidity: 95% (maximum) non-condensing Weight/Shipping Weight: 4 lb/5 lb (1.8 kg/2.3 kg)
Page 2
Page 3
PANEL MOUNTING DIMENSIONS
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH
Page 4
PIPE AND WALL MOUNTING DIMENSIONS
WHEN INCH AND METRIC DIMS
ARE GIVEN
MILLIMETER
INCH