Rosemount Analytical 5081FG Operating Manual

Instruction Manual
IM-106-5081, Rev 2.0 October 2008
Model 5081FG Two-Wire In Situ Oxygen Analyzer (550° to 1600°C)
http://www.raihome.com
Effective October 31, 2008 Rev. 2.0
Page Summary
General Reformatted entire manual in accordance with Emerson Process Management style
guide. Revised all illustration formats. Cover Updated manual revision level and release date. 1-5 Revised Figure 1-2. 1-7 Revised Specifications. 1-9 Revised Table 1-1, Product Matrix. 2-11 Corrected reference to installation drawings in the warning regarding intrinsically safe
applications. 2-14 Revised Reference Air Package illustrations. 3-3 Revised Figure 3-3. 3-5 Revised Figure 3-5 to reflect update to Diagnostics Menu. 3-7 Added Operator Adjustable Parameters discussion and Table 3-1. 3-18, 3-19 Added Current Slope and Current Constant diagnostic descriptions. 4-2, 4-3 Revised Figures 4-1 and 4-2. 4-5, 4-6, 4-7 Revised Figure 4-3, sheets 1, 2, and 3. 5-8 Added Calibration Passes discussion. 6-1 through 6-5 Revised all procedures to reflect changes to Figure 6-1. 6-2 Revised Figure 6-1 to reflect components configuration changes. 7-1 Revised Table 7-1, Replacement Parts list. Updated part numbers. Appendix C Added new Appendix C containing installation drawings for intrinsically safe applications.
Instruction Manual
IM-106-5081, Rev 2.0 October 2008
Model 5081FG
Table of Contents
SECTION i Introduction
SECTION 1 Description and Specifications
SECTION 2 Installation
Preface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Technical Support Hotline: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
System Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Handling the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Inspect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Packing List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Locating Oxygen Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installing Oxygen Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Locating Model 5081 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Installing Model 5081 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Oxygen Probe Signal Connections . . . . . . . . . . . . . . . . . . . . . . . .2-12
Model 5081 Transmitter 4-20 mA and Signal Connections . . . . . 2-13
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Reference Air Package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Instrument Air (Reference Air) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Calibration Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
SECTION 3 Startup and Operation
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Establishing Proper Calibration Gas Flow Rate . . . . . . . . . . . . . . . . .3-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Menu Tree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Operator Adjustable Parameters . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Display Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Fault Val . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Upper Range Val . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
TOC-1
Model 5081FG
Instruction Manual
IM-106-5081, Rev 2.0
October 2008
Cell T Hi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Reset Max Cell T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Set O2 Filter Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Trim 4mA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Trim 20mA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Set Hi Bottle O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Set Lo Bottle O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Set O2 Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Set Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Show Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
T/C mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
O2 CELL mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Cell Impedance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Current Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Current Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Previous Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Previous Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Max Cell T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
SW Ver (SOFt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Unit Ser # (SEr) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
SW Build Number (bLdn) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
SW Build Date (bd). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Cal Check Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
In Manual? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Accept High O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Accept Low O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
SECTION 4 HART/AMS
SECTION 5 Troubleshooting
SECTION 6 Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Field Communicator Signal Line Connections. . . . . . . . . . . . . . . . . . . 4-2
Field Communicator PC Connections . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Off-Line and On-Line Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Field Communicator Start Cal Check Method . . . . . . . . . . . . . . . . . . . 4-8
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Probe Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Fault Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Identifying And Correcting Fault Indications . . . . . . . . . . . . . . . . . . . 5-4
Calibration Passes, But Still Reads Incorrectly . . . . . . . . . . . . . . . . . . 5-8
Probe Passes Calibration, O2 Still
Reads High. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Electronics
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Display Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Spare Board Stack Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Oxygen Probe Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
TOC-2
Instruction Manual
IM-106-5081, Rev 2.0 October 2008
SECTION 7 Replacement Parts
Model 5081FG
APPENDIX A Safety Data
APPENDIX B Return of Material
APPENDIX C Schematics
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . . . . A-24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-30
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Installation Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
TOC-3
Model 5081FG
Instruction Manual
IM-106-5081, Rev 2.0
October 2008
TOC-4
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
Oxygen Analyzers
READ THIS PAGE BEFORE PROCEEDING!

ESSENTIAL INSTRUCTIONS

Emerson Process Management designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson’s Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions
product.
• If you do not understand any of the instructions, contact your Emerson Process Management representative for clarification.
Follow all warnings, cautions, and instructions supplied with the product.
Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
prior to installing, operating, and servicing the
marked on and
to install,
http://www.raihome.com
The information contained in this document is subject to change without notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software within the Model 275/375 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-433-6076 or 1-888-433-6829.
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
ii
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section i Introduction

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Model 5081FG
http://www.raihome.com
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

PREFACE The purpose of this manual is to provide information concerning the

components, functions, installation and maintenance of the Model 5081FG Two-Wire In Situ Oxygen Analyzer (550° to 1600°C).
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS The following definitions apply to WARNINGS, CAUTIONS, and NOTES

found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

SYMBOLS

Technical Support Hotline:

NOTE
Highlights an essential operating procedure, condition, or statement.
:
EARTH (GROUND) TERMINAL
:
PROTECTIVE CONDUCT OR TERMINAL
:
RISK OF ELECTRICAL SHOCK
:
WARNING: REFER TO INSTRUCTION MANUAL
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-6076 1-440-914-1261 In addition to the CSC, you may also contact Field Watch. Field Watch
coordinates Emerson Process Management’s field service throughout the U.S. and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768) Emerson Process Management may also be reached via the Internet through:
e-mail: GAS.CSC@emerson.com World Wide Web: www.raihome.com
iv
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Section 1 Description and Specifications

Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-7

COMPONENT CHECKLIST

A typical Rosemount Analytical Two-Wire In Situ Oxygen Analyzer should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your system in the table located on the first page of this manual.
Also, use the product matrix in Table 1-1 at the end of this section to compare your order number against your unit. The first part of the matrix defines the model. The last part defines the various options and features of the Model 5081FG Analyzer. Ensure the features and options specified by your order number are on or included with the unit.
http://www.raihome.com
Model 5081FG
Figure 1-1. Typical System Package
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
1. Instruction Manual
2. Model 5081 Transmitter
3. Oxygen Probe
4. Adapter Plate with Mounting Hardware and Gasket (Optional)
5. Infrared Remote Control (IRC)
6. Reference Air Set (Optional)
7. Field Communicator Package (Optional)
8. Pipe Mounting Kit (Optional)
1-2
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

SYSTEM OVERVIEW

Scope This Instruction Bulletin is designed to supply details needed to install, start

up, operate, and maintain the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer. The analyzer consists of an oxygen probe and Model 5081 transmitter. The signal conditioning electronics of the Model 5081 transmitter outputs a 4-20 mA signal representing an O control (IRC) allows access to setup, calibration, and diagnostics. This same information, plus additional details, can be accessed with the HART field communicator or Asset Management Solutions (AMS) software.
value. An infrared remote
2

System Description The Rosemount Analytical Two-Wire In Situ Oxygen Analyzer is designed to

measure the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The oxygen probe is permanently positioned within an exhaust duct or stack and performs its task without the use of a sampling system. The Model 5081 transmitter is mounted remotely and conditions the oxygen probe outputs.
The equipment measures oxygen percentage by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
EMF = KT log10(P1/P2) + C
Where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
2. P1 is the partial pressure of the oxygen in the reference air on the opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry, instrument air (20.95% oxygen) as the reference air.
NOTE
The probe uses a Type B thermocouple to measure the cell temperature.
When the cell is at operating temperature and there are unequal oxygen concentrations across the cell, oxygen ions will travel from the high oxygen partial pressure side to the low oxygen partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade.
The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer to provide exceptional sensitivity at low oxygen concentrations.
1-3
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
The oxygen analyzer measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
October 2008

System Configuration The equipment discussed in this manual consists of two major components:

the oxygen probe and the Model 5081 tModel 5081 transmitterransmitter. Oxygen probes are available in three length options, providing in situ penetra-
tion appropriate to the size of the stack or duct. The options on length are 20 in. (508 mm), 26 in. (660 mm), or 38 in. (965 mm).
The Model 5081 transmitter is a two-wire transmitter providing an isolated out­put, 4-20 mA, that is proportional to the measured oxygen concentration. A customer-supplied 24 VDC power source is required to simultaneously pro­vide power to the electronics and a 4-20 mA signal loop. The transmitter accepts millivolt signals generated by the probe and produces the outputs to be used by other remotely connected devices. The output is an isolated 4-20 mA linearized current.

System Features 1. The cell output voltage and sensitivity increase as the oxygen

concentration decreases.
2. High process temperatures eliminate the need for external cell heating and increase cell accuracy.
3. HART communication is standard. To use the HART capability, you must have either:
(a) Field Communicator (b) Asset Management Solutions (AMS) software for the PC
4. Easy probe replacement due to the light-weight, compact probe design.
5. Remote location of the Model 5081 transmitter removes the electronics from high temperature or corrosive environments.
6. Power is supplied to the electronics through the 4-20 mA line for intrinsic safety (IS) purposes.
7. Infrared remote control (IRC) allows interfacing without exposing the electronics.
1-4
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
8. An operator can operate and diagnostically troubleshoot the Two-Wire In Situ Oxygen Analyzer in one of two ways:
a. Infrared Remote Control. The IRC allows access to fault indication
menus on the Model 5081 transmitter LCD display. Calibration can be performed from the IRC keypad.
b. Optional HART Interface (Figure 1-2). The Two-Wire In Situ Oxygen
Analyzer's 4-20 mA output line transmits an analog signal proportional to the oxygen level. The HART output is superimposed on the 4-20 mA output line. This information can be accessed through the following:
Field Communicator - The field communicator requires Device Description (DD) software specific to the Two-Wire In Situ Oxygen Analyzer. The DD software will be supplied with many field communicators but can also be programmed into existing units at most Fisher-Rosemount Analytical service offices. See Section 4: HART/AMS, for additional HART information.
Personal Computer (PC) - The use of a personal computer requires AMS software available from Fisher-Rosemount.
9. Selected Distributed Control Systems - The use of distributed control systems requires input/output (I/O) hardware and AMS Security codes are provided to (by infrared remote control) prevent unintended changes to analyzers adjacent to the one being accessed.
10. A calibration check procedure is provided to determine if the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer is correctly measuring the net oxygen concentration in the industrial process.
Figure 1-2. Two-Wire In Situ Oxygen Analyzer HART Connections and AMS Application
1-5
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Handling the Analyzer The probe was specially packaged to prevent breakage due to handling. Do

not remove the padding material from the probe until immediately before installation.
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to prevent possible equipment damage.
The oxygen probe is designed for industrial applications. Treat each component of the system with care to avoid physical damage. Some probe components are made from ceramics, which are susceptible to shock when mishandled.

System Considerations Prior to installing your Rosemount Analytical Two-Wire In Situ Oxygen

Analyzer, make sure you have all the components necessary to make the system installation. Ensure all the components are properly integrated to make the system functional.
After verifying that you have all the components, select mounting locations and determine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considerations, convenience, and serviceability. Figure 1-1 shows a typical system package. A typical system installation is shown in Figure 1-3.
Figure 1-3. Typical System Installation
1-6
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
A source of instrument air is optional at the oxygen probe for reference air use. Since the unit is equipped with an in place calibration feature, provisions can be made to permanently connect calibration gas bottles to the oxygen probe.
If the calibration gas bottles will be permanently connected, a check valve is required next to the calibration fittings on the integral electronics.
This check valve is to prevent breathing of the calibration gas line and subsequent flue gas condensation and corrosion. The check valve is in addition to the stop valve in the calibration gas kit.
NOTE:
The electronics is rated NEMA 4X (IP65) and is capable of operation at temperatures up to 65°C (149°F).
Retain the packaging in which the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer arrived from the factory in case any components are to be shipped to another site. This packaging has been designed to protect the product.

SPECIFICATIONS

Transmitter
Range 0 to 25% O
Net O
2
System Accuracy ±1.5% of reading or 0.05% O2, whichever is greater System Response in Flue Gas Initial – less than 3 seconds,
Probe Lengths 20 in. (508 mm)
Process Temperature Limits: 550° to 1400°C (1022° to 2552°F)
Ambient Temperature Limits: -40° to 149°C (-40° to 300°F) Mounting and Mounting Position Vertical or horizontal Materials of Construction (Process Wetted Parts):
Inner Probe Outer Protection Tube
Probe Junction Box Speed of Installation/Withdrawal 1 in. (25,4 mm) per minute Hazardous Area Certification Intrinsically safe per EN50 014 (1977), clause 1.3 Reference Air Requirement 100 m per minute (2,119 scfh) of clean, dry
Calibration Check Gas Fittings 1/4 in. tube fittings Cabling Two twisted pairs, shielded
T90 – less than 8 seconds
26 in. (660 mm) 38 in. (965 mm)
Operation to 1600°C (2912°F) with reduced cell life
Zirconia Alumina [1600°C (2912°F) limit] Inconel 600 [1000°C (1832°F) limit] Cast aluminum
instrument air; 1/4 in. tube fittings
Continued on Next Page
2
(1)
1-7
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Electronics
Electronics Enclosure IP65 (NEMA 4X), weatherproof, and
Materials of Construction Low copper aluminum Ambient Temperature Limits -20° to 65°C (-4° to 149°F) Relative Humidity 95% with covers sealed Power Supply and Load Requirements See Figure 1-4 Inputs (from O2 Probe) Two wires - O2 signal
Output One 4-20 mA signal with superimposed digital HART
Hazardous Area Certification ATEX EEx ia IIC T4 or T5
Power Transient Protection IEC 801-4 Shipping Weight 10 lbs (4,5 kg)
Infrared Remote Control
Power Requirements Three AAA Batteries Hazardous Area Certification ATEX EEx ia IIC Class 1, Div 1, Group A, B, C, D
(1)
Thermocouple and O2 probe cell are both unpowered, developing a millivolt emf, and are considered a "simple apparatus" by certifying agencies.
(2)
Dependent on ambient temperature limits.
corrosion-resistant
Two wires - type B thermocouple
signal
NEC Class 1 Div 1 Group B,C,D
(2)
Figure 1-4. Power Supply and Load Requirements
1-8
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Table 1-1. Product Matrix
5081FG High Temperature Oxygen Flue Gas Analyzer
High Temperature Analyzer - Instruction Manual
Code Sensing Probe Type
1 20 in. (508 mm) probe, 1/4 in. tube fittings 2 26 in. (660 mm) probe, 1/4 in. tube fittings 3 38 in. (965 mm) probe, 1/4 in. tube fittings
Code Probe Outer Tube Material - Maximum Operating Temperature
1 Alumina - 2912°F (1600°C) maximum - 1.25 NPT mounting 2 Inconel Alloy - 1832°F (1000°C) maximum - 1.25 NPT mounting
Code Mounting Adapter- Stack Side
0 No adapter plate required uses 1.25 NPT
("0" must also be chosen under "Mounting Adapter" below) 1 New flanged installation - Square weld plate with studs (matches "Mounting Adapter" below) 2 Model 450 mounting ("4" must also be chosen under "Mounting Adapter" below) 3 Competitor's Mount ("5" must also be chosen under "Mounting Adapter" below)
Code Mounting Adapter - Probe Side
0 No adapter plate 1 ANSI 2 in. 150 lb flange to 1.25 NPT adapter
(6 in. dia. flange, 4.75 in. BC with 4 x 0.75 in. dia. holes) 2 DIN to 1.25 NPT adapter (184 mm flange, 145 mm BC with 4 x 18 mm dia. holes) 3 JIS to 1.25 NPT adapter (155 mm flange, 130 mm BC with 4 x 13 mm dia. holes) 4 Model 450 to 1.25 NPT adapter 5 Competitor's mounting flange
Code Electronics & Housing - Intrinsically Safe, NEMA 4X, IP65
1 5081 Electronics (Hart-compatible) - ATEX EEx ia IIC T5 2 5081 Electronics (Hart-compatible) - CSA pending 3 5081 Electronics (Hart-compatible) - FM Class 1, Div. 1, Groups B,C,D
Code Housing Mounting
1 Surface or wall mounting 2 1/2 to 2 in. pipe mounting
Code Communications
1 No remote control 2 Infrared Remote Control (IRC)
(LCD display through cover window
Code Calibration Accessories
1 No hardware 2 Calibration and reference air flowmeters and
reference air pressure regulator
Code Armored Cable Length
00 No cable
11 20 ft (6 m) 12 40 ft (12 m) 13 60 ft (18 m) 14 80 ft (24 m) 15 100 ft (30 m) 16 150 ft (45 m) 17 200 ft (61 m) 18 300 ft (91 m) 19 400 ft (122 m) 20 500 ft (152 m)
5081FG 2 1 0 0 1 1 1 2 11 Example
Model 5081FG
1-9
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
1-10
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section 2 Installation

Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-11
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-14
Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Appendix A. Failure to follow safety instructions could result in serious injury or death.
Model 5081FG
http://www.raihome.com
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

PRE-INSTALLATION

Inspect Carefully inspect the shipping container for any evidence of damage. If the

container is damaged, notify the carrier immediately.

Packing List Confirm that all items shown on the packing list are present. Notify

Rosemount Analytical immediately if items are missing.
Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Appendix A. Failure to follow safety instructions could result in serious injury or death.

MECHANICAL INSTALLATION

Avoid installation locations near steam soot blowers.

Locating Oxygen Probe 1. The location of the oxygen probe in the stack or flue is important for

maximum accuracy in the oxygen analyzing process. The probe must be positioned so the gas it measures is representative of the process. Longer ducts may require several analyzers since the O to stratification. A point too near the wall of the duct, or the in-side radius of a bend, may not provide a representative sample because of the very low flow conditions. The sensing point should be selected so the process gas temperature falls within a range of 550° to 1400°C (1022° to 2552°F). Figure 2-1 provides mechanical installation references.
2. Check the flue or stack for holes and air leakage. The presence of this condition will substantially affect the accuracy of the oxygen reading. Therefore, either make the necessary repairs or install the probe upstream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance access to the probe. Allow adequate clearance for probe removal (Figure 2-1).
can vary due
2
2-2
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-1. Probe Installation Details
Model 5081FG
2-3
Model 5081FG

Installing Oxygen Probe

Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
The probe was specially packaged to prevent breakage due to handling. Do not remove the padding material from the probe until immediately before installation.
1. Ensure all components are available to install the probe.
NOTE
Leave the probe inner protective cover in place until installation. This is required to protect the ceramic cell during movement.
2. If using an optional adapter plate (Figure 2-2) or an optional mounting flange (Figure 2-3), weld or bolt the component onto the duct. The through hole in the stack or duct wall and refractory material must be 2 in. (50,8 mm) diameter, minimum.
3. If the optional adapter plates are not used, a 2 in. NPT, schedule 40, pipe nipple (Figure 2-4) should be welded to the stack or duct wall. When a 2 in. NPT to 1.25 NPT adapter is threaded to the welded pipe nipple, the adapter provides the pipe threads needed for the probe's process fitting.
4. Where high particulate or slag is in the flue gas stream, it may be desirable to inset the probe in the refractory as shown in Figure 2-5. Use pipe couplings and nipples to adjust the probe insertion depth.
5. Use high temperature material (alumina wool) to seal around the probe during insertion. This prevents hot gases from escaping or cold air from entering the stack or duct.
6. Initially insert the probe to a depth of 3 in. (76,2 mm) or 1/2 the depth of the stack or duct refractory, whichever is greater.
2-4
After initial insertion, do not insert the probe at a rate exceeding 1 in. per minute (25.4 mm per minute) or damage to the probe may result due to thermal shock.
7. After initial insertion, insert the probe at a rate of 1 in. (25,4 mm) per minute until the probe is fully inserted.
8. Install anti-seize compound on the pipe threads and screw the probe into the process flange or adapter.
9. If insulation was removed to access the duct work for probe mounting, make sure the insulation is replaced. See Figure 2-4.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-2. Optional Adapter Plate
Model 5081FG
Figure 2-3. Optional Probe Mounting Flange
2-5
Model 5081FG
Figure 2-4. Horizontal Probe Installation
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
2-6
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-5. Adjusting Probe Insertion Depth
Model 5081FG

Locating Model 5081 Transmitter

1. Ensure the Model 5081 transmitter is easily accessible for maintenance and service and for using the infrared remote control (if applicable).
Do not allow the temperature of the Model 5081 transmitter exceed 65°C (149°F) or damage to the unit may result.
2. The ambient temperature of the transmitter housing must not exceed 65°C (149°F). Locate the electronics in an area where temperature extremes, vibration, and electromagnetic and radio frequency interference are minimal.
3. Locate the Model 5081 transmitter within 150 ft (45,7 m) of the oxygen probe due to wiring and signal considerations.
2-7
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

Installing Model 5081 Transmitter

Figure 2-6. Flat Surface Mounting Dimensions
1. Ensure all components are available to install the Model 5081 transmitter.
2. Choose a method or location to mount the transmitter. a. Flat Surface Mounting. The transmitter may be mounted on a flat
b. Pipe Mounting. An optional pipe mounting bracket is available for
surface using the threaded mounting holes located on the bottom of the transmitter housing. Refer to Figure 2-6 for installation references.
this type of installation. Refer to Figure 2-7 for installation references.
2-8
Loading...
+ 100 hidden pages