Rosemount Analytical 5081FG Operating Manual

Page 1
Instruction Manual
IM-106-5081, Rev 2.0 October 2008
Model 5081FG Two-Wire In Situ Oxygen Analyzer (550° to 1600°C)
http://www.raihome.com
Page 2
Page 3
Effective October 31, 2008 Rev. 2.0
Page Summary
General Reformatted entire manual in accordance with Emerson Process Management style
guide. Revised all illustration formats. Cover Updated manual revision level and release date. 1-5 Revised Figure 1-2. 1-7 Revised Specifications. 1-9 Revised Table 1-1, Product Matrix. 2-11 Corrected reference to installation drawings in the warning regarding intrinsically safe
applications. 2-14 Revised Reference Air Package illustrations. 3-3 Revised Figure 3-3. 3-5 Revised Figure 3-5 to reflect update to Diagnostics Menu. 3-7 Added Operator Adjustable Parameters discussion and Table 3-1. 3-18, 3-19 Added Current Slope and Current Constant diagnostic descriptions. 4-2, 4-3 Revised Figures 4-1 and 4-2. 4-5, 4-6, 4-7 Revised Figure 4-3, sheets 1, 2, and 3. 5-8 Added Calibration Passes discussion. 6-1 through 6-5 Revised all procedures to reflect changes to Figure 6-1. 6-2 Revised Figure 6-1 to reflect components configuration changes. 7-1 Revised Table 7-1, Replacement Parts list. Updated part numbers. Appendix C Added new Appendix C containing installation drawings for intrinsically safe applications.
Page 4
Page 5
Instruction Manual
IM-106-5081, Rev 2.0 October 2008
Model 5081FG
Table of Contents
SECTION i Introduction
SECTION 1 Description and Specifications
SECTION 2 Installation
Preface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Technical Support Hotline: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
System Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Handling the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Inspect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Packing List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Locating Oxygen Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Installing Oxygen Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Locating Model 5081 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Installing Model 5081 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Oxygen Probe Signal Connections . . . . . . . . . . . . . . . . . . . . . . . .2-12
Model 5081 Transmitter 4-20 mA and Signal Connections . . . . . 2-13
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Reference Air Package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Instrument Air (Reference Air) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Calibration Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
SECTION 3 Startup and Operation
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Establishing Proper Calibration Gas Flow Rate . . . . . . . . . . . . . . . . .3-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Menu Tree. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Operator Adjustable Parameters . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Display Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Fault Val . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Upper Range Val . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
TOC-1
Page 6
Model 5081FG
Instruction Manual
IM-106-5081, Rev 2.0
October 2008
Cell T Hi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Reset Max Cell T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Set O2 Filter Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Trim 4mA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Trim 20mA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Set Hi Bottle O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Set Lo Bottle O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Set O2 Tracking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Set Code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Show Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
T/C mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
O2 CELL mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Cell Impedance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Current Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Current Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Previous Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Previous Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Max Cell T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
SW Ver (SOFt) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Unit Ser # (SEr) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
SW Build Number (bLdn) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
SW Build Date (bd). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Cal Check Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
In Manual? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
Accept High O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Accept Low O2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
SECTION 4 HART/AMS
SECTION 5 Troubleshooting
SECTION 6 Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Field Communicator Signal Line Connections. . . . . . . . . . . . . . . . . . . 4-2
Field Communicator PC Connections . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Off-Line and On-Line Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Field Communicator Start Cal Check Method . . . . . . . . . . . . . . . . . . . 4-8
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Probe Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Fault Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Identifying And Correcting Fault Indications . . . . . . . . . . . . . . . . . . . 5-4
Calibration Passes, But Still Reads Incorrectly . . . . . . . . . . . . . . . . . . 5-8
Probe Passes Calibration, O2 Still
Reads High. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Electronics
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Display Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Spare Board Stack Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Oxygen Probe Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
TOC-2
Page 7
Instruction Manual
IM-106-5081, Rev 2.0 October 2008
SECTION 7 Replacement Parts
Model 5081FG
APPENDIX A Safety Data
APPENDIX B Return of Material
APPENDIX C Schematics
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . . . . A-24
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-30
Returning Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Installation Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
TOC-3
Page 8
Model 5081FG
Instruction Manual
IM-106-5081, Rev 2.0
October 2008
TOC-4
Page 9
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
Oxygen Analyzers
READ THIS PAGE BEFORE PROCEEDING!

ESSENTIAL INSTRUCTIONS

Emerson Process Management designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson’s Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions
product.
• If you do not understand any of the instructions, contact your Emerson Process Management representative for clarification.
Follow all warnings, cautions, and instructions supplied with the product.
Inform and educate your personnel in the proper installation,
operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and
pressure sources.
• To ensure proper performance, use qualified personnel operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson Process Management. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers
are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
prior to installing, operating, and servicing the
marked on and
to install,
http://www.raihome.com
The information contained in this document is subject to change without notice.
If a Model 275/375 Universal HART® Communicator is used with this unit, the software within the Model 275/375 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-433-6076 or 1-888-433-6829.
Page 10
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
ii
Page 11
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section i Introduction

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page i-iv
Model 5081FG
http://www.raihome.com
Page 12
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

PREFACE The purpose of this manual is to provide information concerning the

components, functions, installation and maintenance of the Model 5081FG Two-Wire In Situ Oxygen Analyzer (550° to 1600°C).
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS The following definitions apply to WARNINGS, CAUTIONS, and NOTES

found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

SYMBOLS

Technical Support Hotline:

NOTE
Highlights an essential operating procedure, condition, or statement.
:
EARTH (GROUND) TERMINAL
:
PROTECTIVE CONDUCT OR TERMINAL
:
RISK OF ELECTRICAL SHOCK
:
WARNING: REFER TO INSTRUCTION MANUAL
NOTE TO USERS
The number in the lower right corner of each illustration in this publication is a manual illustration number. It is not a part number, and is not related to the illustration in any technical manner.
For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-6076 1-440-914-1261 In addition to the CSC, you may also contact Field Watch. Field Watch
coordinates Emerson Process Management’s field service throughout the U.S. and abroad.
Phone: 1-800-654-RSMT (1-800-654-7768) Emerson Process Management may also be reached via the Internet through:
e-mail: GAS.CSC@emerson.com World Wide Web: www.raihome.com
iv
Page 13
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Section 1 Description and Specifications

Component Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-7

COMPONENT CHECKLIST

A typical Rosemount Analytical Two-Wire In Situ Oxygen Analyzer should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your system in the table located on the first page of this manual.
Also, use the product matrix in Table 1-1 at the end of this section to compare your order number against your unit. The first part of the matrix defines the model. The last part defines the various options and features of the Model 5081FG Analyzer. Ensure the features and options specified by your order number are on or included with the unit.
http://www.raihome.com
Page 14
Model 5081FG
Figure 1-1. Typical System Package
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
1. Instruction Manual
2. Model 5081 Transmitter
3. Oxygen Probe
4. Adapter Plate with Mounting Hardware and Gasket (Optional)
5. Infrared Remote Control (IRC)
6. Reference Air Set (Optional)
7. Field Communicator Package (Optional)
8. Pipe Mounting Kit (Optional)
1-2
Page 15
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

SYSTEM OVERVIEW

Scope This Instruction Bulletin is designed to supply details needed to install, start

up, operate, and maintain the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer. The analyzer consists of an oxygen probe and Model 5081 transmitter. The signal conditioning electronics of the Model 5081 transmitter outputs a 4-20 mA signal representing an O control (IRC) allows access to setup, calibration, and diagnostics. This same information, plus additional details, can be accessed with the HART field communicator or Asset Management Solutions (AMS) software.
value. An infrared remote
2

System Description The Rosemount Analytical Two-Wire In Situ Oxygen Analyzer is designed to

measure the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The oxygen probe is permanently positioned within an exhaust duct or stack and performs its task without the use of a sampling system. The Model 5081 transmitter is mounted remotely and conditions the oxygen probe outputs.
The equipment measures oxygen percentage by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
EMF = KT log10(P1/P2) + C
Where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
2. P1 is the partial pressure of the oxygen in the reference air on the opposite side of the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry, instrument air (20.95% oxygen) as the reference air.
NOTE
The probe uses a Type B thermocouple to measure the cell temperature.
When the cell is at operating temperature and there are unequal oxygen concentrations across the cell, oxygen ions will travel from the high oxygen partial pressure side to the low oxygen partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade.
The output is proportional to the inverse logarithm of the oxygen concentration. Therefore, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer to provide exceptional sensitivity at low oxygen concentrations.
1-3
Page 16
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
The oxygen analyzer measures net oxygen concentration in the presence of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the portable apparatus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
October 2008

System Configuration The equipment discussed in this manual consists of two major components:

the oxygen probe and the Model 5081 tModel 5081 transmitterransmitter. Oxygen probes are available in three length options, providing in situ penetra-
tion appropriate to the size of the stack or duct. The options on length are 20 in. (508 mm), 26 in. (660 mm), or 38 in. (965 mm).
The Model 5081 transmitter is a two-wire transmitter providing an isolated out­put, 4-20 mA, that is proportional to the measured oxygen concentration. A customer-supplied 24 VDC power source is required to simultaneously pro­vide power to the electronics and a 4-20 mA signal loop. The transmitter accepts millivolt signals generated by the probe and produces the outputs to be used by other remotely connected devices. The output is an isolated 4-20 mA linearized current.

System Features 1. The cell output voltage and sensitivity increase as the oxygen

concentration decreases.
2. High process temperatures eliminate the need for external cell heating and increase cell accuracy.
3. HART communication is standard. To use the HART capability, you must have either:
(a) Field Communicator (b) Asset Management Solutions (AMS) software for the PC
4. Easy probe replacement due to the light-weight, compact probe design.
5. Remote location of the Model 5081 transmitter removes the electronics from high temperature or corrosive environments.
6. Power is supplied to the electronics through the 4-20 mA line for intrinsic safety (IS) purposes.
7. Infrared remote control (IRC) allows interfacing without exposing the electronics.
1-4
Page 17
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
8. An operator can operate and diagnostically troubleshoot the Two-Wire In Situ Oxygen Analyzer in one of two ways:
a. Infrared Remote Control. The IRC allows access to fault indication
menus on the Model 5081 transmitter LCD display. Calibration can be performed from the IRC keypad.
b. Optional HART Interface (Figure 1-2). The Two-Wire In Situ Oxygen
Analyzer's 4-20 mA output line transmits an analog signal proportional to the oxygen level. The HART output is superimposed on the 4-20 mA output line. This information can be accessed through the following:
Field Communicator - The field communicator requires Device Description (DD) software specific to the Two-Wire In Situ Oxygen Analyzer. The DD software will be supplied with many field communicators but can also be programmed into existing units at most Fisher-Rosemount Analytical service offices. See Section 4: HART/AMS, for additional HART information.
Personal Computer (PC) - The use of a personal computer requires AMS software available from Fisher-Rosemount.
9. Selected Distributed Control Systems - The use of distributed control systems requires input/output (I/O) hardware and AMS Security codes are provided to (by infrared remote control) prevent unintended changes to analyzers adjacent to the one being accessed.
10. A calibration check procedure is provided to determine if the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer is correctly measuring the net oxygen concentration in the industrial process.
Figure 1-2. Two-Wire In Situ Oxygen Analyzer HART Connections and AMS Application
1-5
Page 18
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Handling the Analyzer The probe was specially packaged to prevent breakage due to handling. Do

not remove the padding material from the probe until immediately before installation.
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to prevent possible equipment damage.
The oxygen probe is designed for industrial applications. Treat each component of the system with care to avoid physical damage. Some probe components are made from ceramics, which are susceptible to shock when mishandled.

System Considerations Prior to installing your Rosemount Analytical Two-Wire In Situ Oxygen

Analyzer, make sure you have all the components necessary to make the system installation. Ensure all the components are properly integrated to make the system functional.
After verifying that you have all the components, select mounting locations and determine how each component will be placed in terms of available line voltage, ambient temperatures, environmental considerations, convenience, and serviceability. Figure 1-1 shows a typical system package. A typical system installation is shown in Figure 1-3.
Figure 1-3. Typical System Installation
1-6
Page 19
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
A source of instrument air is optional at the oxygen probe for reference air use. Since the unit is equipped with an in place calibration feature, provisions can be made to permanently connect calibration gas bottles to the oxygen probe.
If the calibration gas bottles will be permanently connected, a check valve is required next to the calibration fittings on the integral electronics.
This check valve is to prevent breathing of the calibration gas line and subsequent flue gas condensation and corrosion. The check valve is in addition to the stop valve in the calibration gas kit.
NOTE:
The electronics is rated NEMA 4X (IP65) and is capable of operation at temperatures up to 65°C (149°F).
Retain the packaging in which the Rosemount Analytical Two-Wire In Situ Oxygen Analyzer arrived from the factory in case any components are to be shipped to another site. This packaging has been designed to protect the product.

SPECIFICATIONS

Transmitter
Range 0 to 25% O
Net O
2
System Accuracy ±1.5% of reading or 0.05% O2, whichever is greater System Response in Flue Gas Initial – less than 3 seconds,
Probe Lengths 20 in. (508 mm)
Process Temperature Limits: 550° to 1400°C (1022° to 2552°F)
Ambient Temperature Limits: -40° to 149°C (-40° to 300°F) Mounting and Mounting Position Vertical or horizontal Materials of Construction (Process Wetted Parts):
Inner Probe Outer Protection Tube
Probe Junction Box Speed of Installation/Withdrawal 1 in. (25,4 mm) per minute Hazardous Area Certification Intrinsically safe per EN50 014 (1977), clause 1.3 Reference Air Requirement 100 m per minute (2,119 scfh) of clean, dry
Calibration Check Gas Fittings 1/4 in. tube fittings Cabling Two twisted pairs, shielded
T90 – less than 8 seconds
26 in. (660 mm) 38 in. (965 mm)
Operation to 1600°C (2912°F) with reduced cell life
Zirconia Alumina [1600°C (2912°F) limit] Inconel 600 [1000°C (1832°F) limit] Cast aluminum
instrument air; 1/4 in. tube fittings
Continued on Next Page
2
(1)
1-7
Page 20
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Electronics
Electronics Enclosure IP65 (NEMA 4X), weatherproof, and
Materials of Construction Low copper aluminum Ambient Temperature Limits -20° to 65°C (-4° to 149°F) Relative Humidity 95% with covers sealed Power Supply and Load Requirements See Figure 1-4 Inputs (from O2 Probe) Two wires - O2 signal
Output One 4-20 mA signal with superimposed digital HART
Hazardous Area Certification ATEX EEx ia IIC T4 or T5
Power Transient Protection IEC 801-4 Shipping Weight 10 lbs (4,5 kg)
Infrared Remote Control
Power Requirements Three AAA Batteries Hazardous Area Certification ATEX EEx ia IIC Class 1, Div 1, Group A, B, C, D
(1)
Thermocouple and O2 probe cell are both unpowered, developing a millivolt emf, and are considered a "simple apparatus" by certifying agencies.
(2)
Dependent on ambient temperature limits.
corrosion-resistant
Two wires - type B thermocouple
signal
NEC Class 1 Div 1 Group B,C,D
(2)
Figure 1-4. Power Supply and Load Requirements
1-8
Page 21
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Table 1-1. Product Matrix
5081FG High Temperature Oxygen Flue Gas Analyzer
High Temperature Analyzer - Instruction Manual
Code Sensing Probe Type
1 20 in. (508 mm) probe, 1/4 in. tube fittings 2 26 in. (660 mm) probe, 1/4 in. tube fittings 3 38 in. (965 mm) probe, 1/4 in. tube fittings
Code Probe Outer Tube Material - Maximum Operating Temperature
1 Alumina - 2912°F (1600°C) maximum - 1.25 NPT mounting 2 Inconel Alloy - 1832°F (1000°C) maximum - 1.25 NPT mounting
Code Mounting Adapter- Stack Side
0 No adapter plate required uses 1.25 NPT
("0" must also be chosen under "Mounting Adapter" below) 1 New flanged installation - Square weld plate with studs (matches "Mounting Adapter" below) 2 Model 450 mounting ("4" must also be chosen under "Mounting Adapter" below) 3 Competitor's Mount ("5" must also be chosen under "Mounting Adapter" below)
Code Mounting Adapter - Probe Side
0 No adapter plate 1 ANSI 2 in. 150 lb flange to 1.25 NPT adapter
(6 in. dia. flange, 4.75 in. BC with 4 x 0.75 in. dia. holes) 2 DIN to 1.25 NPT adapter (184 mm flange, 145 mm BC with 4 x 18 mm dia. holes) 3 JIS to 1.25 NPT adapter (155 mm flange, 130 mm BC with 4 x 13 mm dia. holes) 4 Model 450 to 1.25 NPT adapter 5 Competitor's mounting flange
Code Electronics & Housing - Intrinsically Safe, NEMA 4X, IP65
1 5081 Electronics (Hart-compatible) - ATEX EEx ia IIC T5 2 5081 Electronics (Hart-compatible) - CSA pending 3 5081 Electronics (Hart-compatible) - FM Class 1, Div. 1, Groups B,C,D
Code Housing Mounting
1 Surface or wall mounting 2 1/2 to 2 in. pipe mounting
Code Communications
1 No remote control 2 Infrared Remote Control (IRC)
(LCD display through cover window
Code Calibration Accessories
1 No hardware 2 Calibration and reference air flowmeters and
reference air pressure regulator
Code Armored Cable Length
00 No cable
11 20 ft (6 m) 12 40 ft (12 m) 13 60 ft (18 m) 14 80 ft (24 m) 15 100 ft (30 m) 16 150 ft (45 m) 17 200 ft (61 m) 18 300 ft (91 m) 19 400 ft (122 m) 20 500 ft (152 m)
5081FG 2 1 0 0 1 1 1 2 11 Example
Model 5081FG
1-9
Page 22
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
1-10
Page 23
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section 2 Installation

Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-11
Pneumatic Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-14
Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Appendix A. Failure to follow safety instructions could result in serious injury or death.
Model 5081FG
http://www.raihome.com
Page 24
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

PRE-INSTALLATION

Inspect Carefully inspect the shipping container for any evidence of damage. If the

container is damaged, notify the carrier immediately.

Packing List Confirm that all items shown on the packing list are present. Notify

Rosemount Analytical immediately if items are missing.
Before installing this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Appendix A. Failure to follow safety instructions could result in serious injury or death.

MECHANICAL INSTALLATION

Avoid installation locations near steam soot blowers.

Locating Oxygen Probe 1. The location of the oxygen probe in the stack or flue is important for

maximum accuracy in the oxygen analyzing process. The probe must be positioned so the gas it measures is representative of the process. Longer ducts may require several analyzers since the O to stratification. A point too near the wall of the duct, or the in-side radius of a bend, may not provide a representative sample because of the very low flow conditions. The sensing point should be selected so the process gas temperature falls within a range of 550° to 1400°C (1022° to 2552°F). Figure 2-1 provides mechanical installation references.
2. Check the flue or stack for holes and air leakage. The presence of this condition will substantially affect the accuracy of the oxygen reading. Therefore, either make the necessary repairs or install the probe upstream of any leakage.
3. Ensure the area is clear of internal and external obstructions that will interfere with installation and maintenance access to the probe. Allow adequate clearance for probe removal (Figure 2-1).
can vary due
2
2-2
Page 25
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-1. Probe Installation Details
Model 5081FG
2-3
Page 26
Model 5081FG

Installing Oxygen Probe

Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
The probe was specially packaged to prevent breakage due to handling. Do not remove the padding material from the probe until immediately before installation.
1. Ensure all components are available to install the probe.
NOTE
Leave the probe inner protective cover in place until installation. This is required to protect the ceramic cell during movement.
2. If using an optional adapter plate (Figure 2-2) or an optional mounting flange (Figure 2-3), weld or bolt the component onto the duct. The through hole in the stack or duct wall and refractory material must be 2 in. (50,8 mm) diameter, minimum.
3. If the optional adapter plates are not used, a 2 in. NPT, schedule 40, pipe nipple (Figure 2-4) should be welded to the stack or duct wall. When a 2 in. NPT to 1.25 NPT adapter is threaded to the welded pipe nipple, the adapter provides the pipe threads needed for the probe's process fitting.
4. Where high particulate or slag is in the flue gas stream, it may be desirable to inset the probe in the refractory as shown in Figure 2-5. Use pipe couplings and nipples to adjust the probe insertion depth.
5. Use high temperature material (alumina wool) to seal around the probe during insertion. This prevents hot gases from escaping or cold air from entering the stack or duct.
6. Initially insert the probe to a depth of 3 in. (76,2 mm) or 1/2 the depth of the stack or duct refractory, whichever is greater.
2-4
After initial insertion, do not insert the probe at a rate exceeding 1 in. per minute (25.4 mm per minute) or damage to the probe may result due to thermal shock.
7. After initial insertion, insert the probe at a rate of 1 in. (25,4 mm) per minute until the probe is fully inserted.
8. Install anti-seize compound on the pipe threads and screw the probe into the process flange or adapter.
9. If insulation was removed to access the duct work for probe mounting, make sure the insulation is replaced. See Figure 2-4.
If the ducts will be washed down during outage, MAKE SURE to power down the probes and remove them from the wash area.
Page 27
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-2. Optional Adapter Plate
Model 5081FG
Figure 2-3. Optional Probe Mounting Flange
2-5
Page 28
Model 5081FG
Figure 2-4. Horizontal Probe Installation
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
2-6
Page 29
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-5. Adjusting Probe Insertion Depth
Model 5081FG

Locating Model 5081 Transmitter

1. Ensure the Model 5081 transmitter is easily accessible for maintenance and service and for using the infrared remote control (if applicable).
Do not allow the temperature of the Model 5081 transmitter exceed 65°C (149°F) or damage to the unit may result.
2. The ambient temperature of the transmitter housing must not exceed 65°C (149°F). Locate the electronics in an area where temperature extremes, vibration, and electromagnetic and radio frequency interference are minimal.
3. Locate the Model 5081 transmitter within 150 ft (45,7 m) of the oxygen probe due to wiring and signal considerations.
2-7
Page 30
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

Installing Model 5081 Transmitter

Figure 2-6. Flat Surface Mounting Dimensions
1. Ensure all components are available to install the Model 5081 transmitter.
2. Choose a method or location to mount the transmitter. a. Flat Surface Mounting. The transmitter may be mounted on a flat
b. Pipe Mounting. An optional pipe mounting bracket is available for
surface using the threaded mounting holes located on the bottom of the transmitter housing. Refer to Figure 2-6 for installation references.
this type of installation. Refer to Figure 2-7 for installation references.
2-8
Page 31
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 2-7. Pipe Mounting Dimensions
Model 5081FG
2-9
Page 32
Model 5081FG
Figure 2-8. Display Positioning Assembly
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
3. For correct viewing orientation, the display may be changed 90 degrees, using the following procedure:
a. Refer to Figure 2-8. Loosen the cover lock screw until the cover lock
is disengaged from the knurled surface on the threaded circuit end
cap. b. Remove the circuit end cap. c. Remove the three screws retaining the display board in place. d. Lift and rotate the display board 90 degrees either way. e. Reposition the display board on the standoffs. Install and tighten all
three screws. f. Install the circuit end cap and tighten the cover lock screw to secure
the cover lock in place.
2-10
Page 33
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

ELECTRICAL INSTALLATION

All wiring must conform to local and national codes.
For intrinsically safe applications, refer to the installation drawings in Appendix C of this Instruction Manual.
Disconnect and lock out power before connecting the unit to the power supply.
Install all protective equipment covers and safety ground leads after installation. Failure to install covers and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, connection to the main electrical power supply must be made through a circuit breaker (min 10 A) which will disconnect all current-carrying conductors during a fault situation. This circuit breaker should also include a mechanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.

General The power supply and signal wiring should be shielded. Also, make sure the

signal wiring is grounded at the Model 5081 transmitter end only. Do not ground the signal loop at more than one point. Twisted pairs are recommended. Ground the transmitter housing to an earth ground to prevent unwanted electromagnetic interference (EMI) or radio frequency interference (RFI).
NOTE
For optimum EMI/RFI immunity, shield the 4-20 mA current loop cable and enclose in an earth grounded metal conduit.
NOTE
Never run signal or sensor wiring in the same conduit, or open tray, with power cables. Keep signal or sensor wiring at least 12 in. (0,3 m) away from other electrical equipment and 6.5 ft (2 m) from heavy electrical equipment.
2-11
Page 34
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Moisture accumulation in the transmitter housing can affect its performance and may void its warranty.
It is necessary to prevent moisture from entering the Model 5081 transmitter housing. The use of weather-tight cable glands is required. If conduit is used, plug and seal connections on the transmitter housing to prevent moisture accumulation in the terminal side of the housing.

Oxygen Probe Signal Connections

Figure 2-9. Oxygen Probe Terminal Block
1. Two signals represent the O2 value and the cell temperature. The probe provides these values to the Model 5081 transmitter for processing and
signal conditioning.
2. Wiring connections for the probe are shown in Figure 2-9.
2-12
Page 35
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Model 5081 Transmitter 4-20 mA and Signal Connections

Figure 2-10. Transmitter Terminal Block
1. A 4-20 mA signal represents the O2 value. Superimposed on the 4-20 mA signal is HART information that is accessible through a
HART field communicator or AMS software.
2. Two signals representing the O supplied to the Model 5081 transmitter from the oxygen probe.
3. Wiring connections for the Model 5081 transmitter are shown in Figure 2-10.
NOTE
The ground arrangement shown in Figure 2-10 limits the amount of noise introduced into the electronics.
4. Connect wire shields to terminal 1. Connect earth ground as shown.
value and the cell temperature are
2
2-13
Page 36
Model 5081FG

PNEUMATIC INSTALLATION

Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

General Reference air is required for O

required during a calibration check. Refer to Figure 2-11 for the gas connections on the oxygen probe.
calculation, and calibration check gas is
2

Reference Air Package After the oxygen probe is installed, connect the reference air set. Install the

reference air set according to Figure 2-11.

Instrument Air (Reference Air)

Figure 2-11. Air Set, Plant Air Connections
Instrument air is required for reference. Refer to the reference air schematic, Figure 2-12. Use 10 psig (68,95 kPa gage) minimum, 225 psig (1551,38 kPa gage) at 0.2 scfh (100 m hydrocarbons. Regulator outlet pressure should be set at 5 psi (35 kPa).
/min.); less than 40 parts-per-million total
Figure 2-12. Reference Air Schematic
2-14
Page 37
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Calibration Gas Two calibration check gas concentrations are used with the Two-Wire In Situ

Oxygen Analyzer: Low Gas - 0.4% O balance in nitrogen.
and High Gas - 8% O2, each with the
2
Figure 2-13. Gas Connections at Oxygen Probe
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero) be between 0.4% and 2.0% O than 40 parts per million. Failure to use proper gases will result in erroneous readings.
. Do not use gases with hydrocarbon concentrations of more
2
Do not use 100% nitrogen for the low (zero %) check gas. See Figure 2-13 for the probe connections. Set both calibration check gases at the same flow rate: 5 scfh (2,5
/min).
2-15
Page 38
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
2-16
Page 39
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section 3 Startup and Operation

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Establishing Proper Calibration Gas Flow Rate . . . . . . .page 3-3
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-4
Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-7
Diagnostics Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-16
Cal Check Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-21

GENERAL

Install all protective equipment covers and safety ground leads before equipment startup. Failure to install covers and ground leads could result in serious injury or death.
Model 5081FG
Verify Mechanical Installation
Ensure the Two-Wire In Situ Oxygen Analyzer is installed correctly. See Mechanical Installation in Section 2: Installation for mechanical installation information.
Verify Terminal Block Wiring
Ensure the wiring of both the oxygen probe terminal block and Model 5081 transmitter terminal block is correct. Refer to Electrical Installation in Section 2: Installation for electrical installation and wiring information.

POWER UP General

The Two-Wire In Situ Oxygen Analyzer displays the current oxygen reading on the LCD face of the Model 5081 transmitter. The O temperature, and 4-20 mA output current are displayed as shown in Figure 3-1. This and other information may also be accessed using HART/AMS.
Startup Display
When the probe is first inserted into the stack, some time is required until the minimum operating temperature [550°C (1022°F)] is reached. Some time is also required for the electronics to reach an operating state. Therefore, when the unit is first powered up, a faulted operation display as shown in Figure 3-2 may be displayed by the transmitter until the probe operating temperature is reached and the electronics are working properly (approximately 5 minutes).
concentration, cell
2
http://www.raihome.com
Page 40
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Operating Display
After the probe has reached operating temperatures, the Model 5081 transmitter display should look similar to Figure 3-1. The display will now track the O
concentration, cell temperature, and 4-20 mA output current.
2
Figure 3-1. Normal Operation Display
Figure 3-2. Faulted Operation Display
3-2
Page 41
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

ESTABLISHING PROPER CALIBRATION GAS FLOW RATE

Figure 3-3. Proper Calibration Check Gas Flow Rate
The calibration gas flow must be enough to ensure that no combustion flue gases mix with the calibration check gases and only clean, good calibration check gas surrounds the cell without expending excess gas (Figure 3-3). Monitor the O calibration check gas flow rate as follows:
NOTE
Only set the calibration check gas flow rate at startup. It is not necessary to perform this procedure for each calibration check.
1. Adjust the calibration check gas flow to 5 scfh (2,5 cell is surrounded by calibration check gas. Due to the cooling effect of the gas, the cell temperature will decrease slightly, causing the O concentration to drop. Once the electronics compensates for this effect, the O
2. Next, slowly reduce the calibration check gas flow until the O concentration changes, which indicates that the calibration check and flue gases are mixing. Increase the flow rate until this effect is eliminated.
2
concentration will stabilize.
2
concentration using an IRC or a field communicator. Set the
/min.) to ensure the
2
2
3-3
Page 42
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

OPERATION

Overview This section explains the operator controls and displays of the Two-Wire In

Situ Oxygen Analyzer. The use of the Infrared Remote Control (IRC) and the Model 5081 transmitter Liquid Crystal Display (LCD) are described in detail.
HART/AMS operation is not covered here. Refer to Section 4: HART/AMS.

Display The LCD on the circuit end of the Model 5081 transmitter displays O

concentration, cell temperature, and 4-20 mA output current during normal operation (Figure 3-4). The LCD will also display fault conditions when they occur. To interact with the transmitter, use the IRC and navigate through a series of menus displayed on the LCD.
2

Menu Tree The screens that can be displayed are shown in the menu tree of Figure 3-5.

These screens are displayed on the LCD and are accessed using the IRC keypad.
Figure 3-4. Normal Operation Display
3-4
Page 43
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 3-5. Transmitter Menu Tree
Model 5081FG
3-5
Page 44
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Navigation The IRC in Figure 3-6 is used to interact with the Model 5081 transmitter and

navigate through the screens on the LCD.
1. Hold the IRC within 6 ft (1,8 m) of the Model 5081 transmitter and within 15 degrees from the centerline of the transmitter LCD. The amount of ambient light may also affect IRC performance:
NOTE
The LCD may react slowly to IRC commands. Allow sufficient time between key presses to avoid undesired or repeated commands from accumulating in the command queue.
2. Use the keys on the IRC to navigate through the menu screens. Refer to Figure 3-6. General usage is as follows:
a. RESET. Returns to the PROCESS DISPLAY screen at the top of the
menu tree. Any non-entered number in the exited state will be
ignored, and the previous data will be used. b. HOLD. Not used. c. Left/Right Arrow. Moves left and right among editable digits on the
display. d. Up/Down Arrow. Increases or decreases the value of the currently
selected digit on the display. e. CAL. Accesses the CALCHECK MENU branch of the menu tree.
Only works from the PROCESS DISPLAY screen.
Figure 3-6. Infrared Remote Control (IRC)
3-6
Page 45
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
f. PROG. Accesses the PROGRAM MENU branch of the menu tree.
Only works from the PROCESS DISPLAY screen. g. DIAG. Accesses the DIAGNOSTICS MENU branch of the menu
tree. Only works from the PROCESS DISPLAY screen. h. ENTER. Initiates the editing process and causes the most
significant digit of the edited item to start flashing. Also processes the
entry so the previous value updates to the new value entered using
the arrow keys. Failure to press ENTER before exiting a screen will
cancel the input value and revert to the previous value. i. NEXT. Accesses the next user screen as shown in the menu tree.
Any non-entered number in the exited state will be ignored, and the
previous data will be used. j. EXIT. Exits from sub-branches of the menu tree where an exit option
is explicitly shown. Otherwise, returns to the PROCESS DISPLAY
screen at the top of the menu tree. Any non-entered number in the
exited state will be ignored, and the previous data will be used.

PROGRAM MENU The PROGRAM MENU branch of the menu tree allows you to program and

edit some process parameters, faults, outputs, and security codes. To access this branch of the menu tree, press the PROG key on the IRC when in the PROCESS DISPLAY screen (Normal or Faulted). If security is enabled, you must enter the analyzer code to gain further access to the screens in this branch. Each screen in this branch is accessed sequentially using the NEXT key. Refer to Figure 3-5 during the following menu and screen descriptions.
NOTE
To edit a screen value, press ENTER to access the data field. Use the left and right arrow keys to move among the digits in the data field. Note that the editable position will be flashing. To change the value of a digit, use the up and down arrow keys to increase or decrease the value. When finished editing, press ENTER to accept the value. To go to the next screen in the menu, press NEXT.

Operator Adjustable Parameters

Figure 3-1 lists the range and default value of operator-adjustable variables used by the Model 5081 transmitter. Each of these variables may be adjusted using the PROGRAM MENU screens.
Table 3-1. Adjusatble Parameters
Parameter Function Range Factory Default
CODE Use to enter the access code for this analyzer; select 555
FAULT VAL Use to designate a 4-20mA value that when displayed
UPPER RANGE VAL Use to set upper O
CELL T HI Use to set upper cell temperature for no-fault condition 650°C - 1600°C 1600°C SET O2 FILTER TIME Use to adjust analyzer response time to changing O SET HI BOTTLE O2 Use to define actual O2% of high calibration check gas 8% SET LO BOTTLE O2 Use to define actual O SET CODE Use to set security code for this analyzer 000 - 999 (excluding 000 and 555) 000 (no code)
to display designated analyzer code
will indicate an analyzer-faulted condition
% limit equivalent to 20mA output
(adjust the slope of the analyzer)
2
% of low calibration check gas 2%
2
000 - 999 (excluding 000 and 555) 000 (no code)
3.8mA - 24mA 3.6mA
2.0% - 25.0% 25%
% 0 - 300 seconds 0 seconds
2
3-7
Page 46
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Code Refer to Figure 3-7. After pressing the PROG key, this screen will display if

security is enabled (see Set Code). Use this screen to identify a specific analyzer in a process to prevent accessing an adjacent analyzer when using the IRC.
Press ENTER to begin editing. At this point, you can either specify the analyzer by its access code or view its code if it is unknown.
1. To gain further access to the screens in the PROGRAM MENU branch, enter the correct three-digit analyzer access code using the arrow keys and press ENTER. If security is disabled, this screen does not appear and the system displays the FAULT VAL screen.
2. If the analyzer access code is un-known, enter 555 and press ENTER to access the DISPLAY CODE screen. In that screen you will be able to view the analyzer access code.
Figure 3-7. Code
3-8
Page 47
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Display Code Refer to Figure 3-8. This screen is accessible from the CODE screen by

entering 555 and pressing ENTER. The DISPLAY CODE screen identifies the analyzer access code so you can return to the CODE screen and enter the code as described in Code. To return to the CODE screen, press NEXT.

Fault Val Refer to Figure 3-9. Use this screen to set the value that the 4-20 mA output

will drive to and display during a fault condition. Press ENTER to begin editing. Use the arrow keys to enter a fault value. The fault value can be between 3.8 and 24 mA. Then, press ENTER to accept the value. Pressing NEXT displays the UPPER RANGE VAL screen. Refer to Section 5: Troubleshooting, for the actual fault conditions.
Figure 3-8. Display Code
Figure 3-9. Fault Val
3-9
Page 48
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Upper Range Val Refer to Figure 3-10. Use this screen to set the value of the upper range limit.

This value is the maximum limit of the O used to scale the 4-20 mA output. Press ENTER to begin editing. Use the arrow keys to select and change the value. The upper range value can be between 0 and 25%. Then, press ENTER to accept the value. Pressing NEXT displays the CELL T HI screen.
concentration measurement and is
2

Cell T Hi Refer to Figure 3-11. Use this screen to set the value of the upper cell

temperature fault condition. This value is the maximum allowed cell temperature before a fault condition is indicated. Press ENTER to begin editing. Use the arrow keys to select and change the value. The value must be between 550° and 1600°C. Press ENTER to accept the value. Pressing NEXT displays the RESET MAX CELL T screen.
Figure 3-10. Upper Range Val
Figure 3-11. Cell T Hi
3-10
Page 49
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Reset Max Cell T Refer to Figure 3-12. The transmitter tracks the maximum cell temperature

obtained. Use this screen to reset the maximum cell temperature attained value to the current cell temperature. Press ENTER to begin editing. Use the arrow keys to select and change the value (Y/N). Then, press ENTER to accept the value. Pressing NEXT displays the SET O
FILTER TIME screen.
2

Set O2 Filter Time Refer to Figure 3-13. In some applications it is beneficial to dampen the raw

O
signal coming from the cell. Use this screen to enter the amount of time it
2
will take the O editing. Use the arrow keys to select and change the screen value to the O filter value (in seconds). Enter a value between 0 and 300 seconds and press ENTER to accept the value. Press NEXT to access the TRIM 4 mA? screen.
to reach 90% of the new reading. Press ENTER to begin
2
2
Figure 3-12. Reset Max Cell T
Figure 3-13. Set O2 Filter Time
3-11
Page 50
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Trim 4mA? Refer to Figure 3-14. Use this screen to trim the 4 mA value of the 4-20 mA

output.
NOTE
Before trimming the 4 mA value you must break the loop to add the ammeter. Power down the unit, connect the ammeter in series with Model 5081 transmitter terminals 15(-) and 16(+), power up the unit, and return to the TRIM 4 mA? screen.
Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the value displayed on the installed ammeter. Press ENTER to accept the value. After the value is entered, the unit calibrates itself to ensure it outputs 4 mA. Both the display and the ammeter will display 4 mA. Pressing EXIT returns to the initial TRIM 4 mA? screen, and pressing NEXT displays the TRIM 20 mA? screen.
Figure 3-14. Trim 4mA?
3-12
Page 51
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Trim 20mA? Refer to Figure 3-15. Use this screen to trim the 20 mA value of the 4-20 mA

output.
NOTE
Before trimming the 20 mA value you must break the loop to add the ammeter. Power down the unit, connect the ammeter in series with Model 5081 transmitter terminals 15(-) and 16(+), power up the unit, and return to the TRIM 20 mA? screen.
Press ENTER to begin editing. Use the arrow keys to select and change the screen value to the value displayed on the installed ammeter. Press ENTER to accept the value. After the value is entered, the unit calibrates itself to ensure it outputs 20 mA. Both the display and the ammeter will display 20 mA. Pressing EXIT returns to the initial TRIM 20 mA? screen, and pressing NEXT displays the SET HI BOTTLE O
screen.
2
Figure 3-15. Trim 20mA?
3-13
Page 52
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Set Hi Bottle O2 Refer to Figure 3-16. Use this screen to identify, within the electronics, the

percentage of O begin editing. Use the arrow keys to select and change the screen value to the O
percentage of the high calibration check gas. Press ENTER to accept
2
the value.
used as the high calibration check gas. Press ENTER to
2
Set Lo Bottle O
2
Figure 3-16. Set Hi Bottle O
Refer to Figure 3-17. Use this screen to identify, within the electronics, the percentage of O
used as the low calibration check gas. Press ENTER to
2
begin editing. Use the arrow keys to select and change the screen value to the O
percentage of the low calibration check gas. Press ENTER to accept
2
the value. Press NEXT to display the SET O
2
TRACKING screen.
2
Figure 3-17. Set Lo Bottle O
3-14
2
Page 53
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Set O2 Tracking Refer to Figure 3-18. Use this screen to permit the 4-20 mA line to track the

value during a calibration check. Press ENTER to begin editing. Use the
O
2
arrow keys to select Y or N. Entering Y (yes) will allow the 4-20 mA line to track the O O
value steady during the calibration check. Press ENTER to accept the
2
value during the calibration check. Entering N (no) will hold the
2
value. Press NEXT to display the SET CODE screen.

Set Code Refer to Figure 3-19. Use this screen to set the security code for the Model

5081 transmitter. Press ENTER to begin editing. Use the arrow keys to select and change the value. Select any value between 000 and 999, excluding 000 and 555. Code 000 indicates that no code is set. Code 555 accesses the DISPLAY CODE screen. Press ENTER to accept the value. Pressing NEXT returns to the FAULT VAL screen at the beginning of the PROGRAM MENU.
Figure 3-18. Set O
Tracking
2
Figure 3-19. Set Code
3-15
Page 54
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

DIAGNOSTICS MENU The DIAGNOSTICS MENU branch of the menu tree allows you to examine

outputs, current faults, and unit information. None of the items in the DIAGNOSTICS MENU are editable. This branch of the menu tree may be accessed by pressing DIAG on the IRC when in the PROCESS DISPLAY screen (Normal or Faulted). Each screen in this branch is accessed sequentially by pressing NEXT. Refer to the menu in Figure 3-5 when reviewing the following menu and screens.

Show Fault Refer to Figure 3-20. After pressing DIAG, this screen displays. Pressing

ENTER accesses a screen displaying the current fault (if any). If more than one fault exists, and you are in the FAULT screen, press NEXT to go to the next fault. Information on the fault screens can be found in Section 5: Troubleshooting. Press EXIT to return from this fault sub-menu and press NEXT to access the T/C mV screen.
Figure 3-20. Show Fault
3-16
Page 55
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

T/C mV Refer to Figure 3-21. Use this screen to examine the cell thermocouple mV

output. Three decimal places are displayed. Pressing NEXT accesses the O CELL mV screen.
2

O2 CELL mV Refer to Figure 3-22. Use this screen to examine the O

Pressing NEXT accesses the CELL IMPEDANCE screen.
Figure 3-21. T/C mV
CELL mV output.
2
Figure 3-22. O2 CELL mV
3-17
Page 56
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Cell Impedance Refer to Figure 3-23. Use this screen to examine the O
status. GOOD indicates the cell is operating normally. WARN indicates the cell has degraded but is still operational. HI indicates that the cell has degraded but is still operational; however, failure will occur soon. Pressing NEXT accesses the CURRENT SLOPE screen.
NOTE
Temperature influences cell impedance. Wait until the cell is at operating temperature before checking cell impedance. If checked before the cell reaches 550°C (1022°F), this screen displays a fail indication.

cell impedance

2

Current Slope Refer to Figure 3-24. Use this screen to examine the slope calculated from

the most recent calibration check. The slope is the amount of cell voltage
Figure 3-23. Cell Impedance
generated for a given O over the life of the probe. Tracking the slope will indicate if the probe is degrading. Press NEXT to access the CURRENT CONSTANT screen.
value. For each calibration check, record the slope
2
Figure 3-24. Current Slope
3-18
Page 57
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Current Constant Refer to Figure 3-25. Use this screen to examine the cell zero constant

calculated from the most recent calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O access the PREVIOUS SLOPE screen.
on the reference and process sides of the cell. Press NEXT to
2

Previous Slope Refer to Figure 3-26. Use this screen to examine the slope value stored from

the second to last calibration check. The slope is the amount of cell voltage
Figure 3-25. Current Constant
generated for a given O over the life of the probe. Tracking the slope will indicate if the probe is degrading. Press NEXT to access the PREVIOUS CONSTANT screen.
value. For each calibration check, record the slope
2
Figure 3-26. Previous Slope
3-19
Page 58
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
October 2008

Previous Constant Refer to Figure 3-27. Use this screen to examine the cell zero constant stored

from the second to last calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O the reference and process sides of the cell. Press NEXT to access the MAX CELL T screen.
2
on

Max Cell T Refer to Figure 3-28. Use this screen to examine the maximum temperature

attained by the O Pressing NEXT accesses the SW VER screen.
cell. This value can be reset under the PROGRAM MENU.
2
Figure 3-27. Previous Constant
Figure 3-28. Max Cell T
3-20
Page 59
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

SW Ver (SOFt) Use this screen to see the software version number for the Model 5081

transmitter. Pressing NEXT accesses the UNIT SER # screen.

Unit Ser # (SEr) Use this screen to see the unit serial number for the Model 5081 transmitter.

Pressing NEXT accesses the SW BUILD NUMBER screen.

SW Build Number (bLdn) Use this screen to see the software build number for the Model 5081

transmitter. Pressing NEXT accesses the SW BUILD DATE screen.

SW Build Date (bd) Use this screen to see the software build date for the Model 5081 transmitter.

Pressing NEXT returns to the beginning of the DIAGNOSTICS MENU branch (the SHOW FAULT screen).

CAL CHECK MENU The CALCHECK MENU branch of the menu tree (Figure 3-5) allows you to

perform a calibration check of the analyzer. Before performing a calibration check, ensure the high calibration check gas and low calibration check gas O percentages are entered into the electronics via the PROGRAM MENU. To set these values, refer to Set Hi Bottle O2 and Set Lo Bottle O2.
Once these values are set, access the CAL-CHECK MENU branch by pressing CAL on the IRC when in the PROCESS DISPLAY screen (Normal or Faulted). Each screen in this branch identifies a process step in the calibration check procedure. The first screen in the sequence is the IN MANUAL? screen.
2

In Manual?

Figure 3-29. In Manual?
Failure to remove the analyzer from automatic control loops prior to performing this procedure may result in a dangerous operating condition.
Refer to Figure 3-29. If the O2 output value is used in any automatic process control loops the loop must be placed in manual to begin a calibration check.
Once the analyzer is removed from any automatic control loops press ENTER to edit the screen. Use the arrow keys to select Y (yes); press ENTER to start a calibration check and to display the ACCEPT HIGH O
screen.
2
3-21
Page 60
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Accept High O
Accept Low O
2
2
Figure 3-30. Accept High O
Refer to Figure 3-30. After pressing ENTER to begin the calibration check, the high calibration check gas starts to flow. After waiting approximately three minutes for the displayed O
value to settle, press NEXT to accept the high
2
calibration check gas reading and apply the low calibration check gas. The next screen to display is the ACCEPT LOW O
screen.
2
Refer to Figure 3-31. Once the low calibration check gas is applied, wait approximately three minutes for the displayed O
value to settle. Once the
2
value settles, press NEXT to accept the reading and to display the SLOPE screen.
2
Figure 3-31. Accept Low O
2
3-22
Page 61
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Slope Refer to Figure 3-32. Use this screen to examine the slope calculated from

current calibration check. The slope is the amount of cell voltage generated for a given O life of the probe. Tracking the slope will indicate if the probe is degrading. Press NEXT to access the CONSTANT screen.
value. After each calibration check, record the slope over the
2

Constant Refer to Figure 3-33. Use this screen to examine the cell zero constant

calculated from the current calibration check. The constant represents the voltage generated by the cell when no difference exists between the amount of O
on the reference and process sides of the cell. Note this value for
2
comparison against future calibration checks. Press RESET or EXIT to return to the PROCESS DISPLAY screen.
Figure 3-32. Slope
Figure 3-33. Constant
3-23
Page 62
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
3-24
Page 63
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Section 4 HART/AMS

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Field Communicator Signal Line Connections . . . . . . . . page 4-2
Field Communicator PC Connections . . . . . . . . . . . . . . . page 4-4
Off-Line and On-Line Operations . . . . . . . . . . . . . . . . . . . page 4-4
HART/AMS Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-4
Field Communicator Start Cal Check Method . . . . . . . . . page 4-8

OVERVIEW The HART field communicator is a handheld communications interface

device. It provides a common communications link to all microprocesor-based instruments that are HART compatible. The field communicator has a liquid crystal display (LCD) and keypad. A pocket-sized manual, included with the field communicator, details the specific functions of the keypad keys.
To interface with the Model 5081FG Analyzer, the field communicator requires a termination point along the 4-20 mA current loop and a minimum load resis­tance of 250 ohms between the field communicator and the power supply.
The field communicator accomplishes its task using a frequency shift keying (FSK) technique. With the use of FSK, high-frequency digital communication signals are superimposed on the analyzer's 4-20 mA current loop. The field communicator does not disturb the 4-20 mA signal, since no net energy is added to the loop.
The field communicator may be interfaced with a personal computer (PC), providing that special software has been installed. To connect the field com­municator to a PC, an interface adapter is required. Refer to the proper field communicator documentation regarding the PC interface option.
http://www.raihome.com
Page 64
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

FIELD COMMUNICATOR SIGNAL LINE CONNECTIONS

Figure 4-1. Signal Line Connections,
The field communicator can connect to the analyzer's analog output signal line at any wiring termination in the 4-20 mA current loop. There are two meth­ods of connecting the field communicator to the signal line. For applications in which the signal line has a load resistance of 250 ohms or more, refer to method 1. For applications in which the signal line load resistance is less than 250 ohms, refer to method 2.
Method 1, For Load Resistance
Refer to Figure 4-1 and the following steps to connect the field communicator to a signal line < 250 ohms or more of load resistance.
Explosions can result in death or serious injury. Do not make connections to the field communicator's serial port, 4-20 mV signal line, or NiCad recharger jack in an explosive atmosphere.
Using the supplied lead set, connect the field communicator in parallel with to the Model 5081FG Analyzer. Use any wiring termination points in the analog output 4-20 mA signal line.
250 Ohms
250 Ohms Load Resistance
4-2
Page 65
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Method 2, For Load Resistance < 250 ohms
Refer to Figure 4-2 and the following steps to connect the field communicator to a signal line with < 250 ohms load resistance.
Explosions can result in death or serious injury. Do not make connections to the field communicator's serial port, 4-20 mA signal line, or NiCad recharger jack in an explosive atmosphere.
1. At a convenient point break the analog output 4-20 mA signal line and install the optional 250 ohm load resistor.
2. Plug the load resistor into the rear panel of the field communicator.
Figure 4-2. Signal Line Connections, < 250 Ohms Load Resistance
Model 5081FG
4-3
Page 66
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

FIELD COMMUNICATOR PC CONNECTIONS

OFF-LINE AND ON-LINE OPERATIONS

There is an option to interface the field communicator with a personal com­puter. Refer to the applicable field communicator documentation regarding the PC interface option.
The field communicator can be operated both off-line and on-line. Off-line operations are those in which the communicator is not connected to
the Model 5081FG Analyzer. Off-line operations can include interfacing the field communicator with a PC (refer to applicable HART documentation regarding HART/PC applications.
In the on-line mode, the communicator is connected to the 4-20 mA analog output signal line. The communicator is connected in parallel to the Model 5081FG Analyzer or in parallel to the 250 ohm load resistor.
NOTE
If the field communicator is turned on while connected to the 4-20 mA analog output signal line, an undefined status indication appears while the communicator warms up. Wait until the warm-up period ends to continue.
The opening menu displayed on the LCD is different for on-line and off-line operations. When powering up a disconnected (off-line) communicator, the LCD will display the Main Menu. When powering up a connected (on-line) communicator, the LCD will display the On-line Menu. Refer to the field com­municator manual for detailed menu information.

HART/AMS MENU TREE This section consists of a menu tree for the field communicator. This menu is

specific to Two-Wire In Situ Oxygen Analyzer applications.
4-4
Page 67
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 4-3. HART/AMS Menu Tree (Sheet 1 of 3)
Model 5081FG
4-5
Page 68
Model 5081FG
Figure 4-3. HART/AMS Menu Tree (Sheet 2 of 3)
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
4-6
Page 69
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 4-3. HART/AMS Menu Tree (Sheet 3 of 3)
Model 5081FG
4-7
Page 70
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

FIELD COMMUNICATOR START CAL CHECK METHOD

To perform a calibration check on the Two-Wire In Situ Oxygen Analyzer with the field communicator, use the following procedure. If necessary, use the menu tree in Figure 4-3 (sheet 1 of 3) for reference.
NOTE
To select a menu item use either the up and down arrow keys to scroll to the menu item. Press the right arrow key or use the number keypad to select the menu item number. To return to a preceding menu, press the left arrow key.
NOTE
Pressing ABORT at any time during this process will purge the calibration check gases and end the calibration check procedure.
1. From the DEVICE SETUP SCREEN, select DIAG/SERVICE.
2. From the DIAG/SERVICE screen, select O
3. Before starting the calibration check procedure, first set up the high calibration check gas, low calibration check gas, and tracking using HART/AMS.
4. From the O2 CALCHECK screen, select menu item 1, START CALCHECK, to access the calibration check procedure.
CALCHECK.
2
Failure to remove the analyzer from automatic control loops prior to performing this procedure may result in a dangerous operating condition.
5. In the first screen, a "Loop should be removed from automatic control" warning appears. Remove the analyzer from any automatic control loops to avoid undesirable equipment performance and press OK.
6. The next screen prompts you to apply the high calibration check gas. This message will only display for approximately three seconds. Press OK.
7. At this point, calibration check gas will flow for approximately three min­utes until the gas reading is taken. Once the gas is measured, the mes­sage, "Hi gas reading taken," displays for three seconds.
8. Next, the screen prompts you to apply the low calibration check gas. Press OK.
9. The low calibration check gas will flow for approximately three minutes until the reading is taken. Once the gas is measured, the message, "Low gas reading taken," displays for three seconds.
10. Next, the screen prompts you to disconnect the calibration check gases. Press OK. Once the gases are disconnected the system will purge the gases for approximately three minutes.
11. When the "Loop may be returned to automatic control" note appears, return the analyzer to the automatic control loops previously removed and press OK.
4-8
Page 71
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Section 5 Troubleshooting

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Probe Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1
Fault Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3
Identifying And Correcting Fault Indications . . . . . . . . . . page 5-4
Calibration Passes, But Still Reads Incorrectly . . . . . . . . page 5-8

GENERAL This troubleshooting section describes how to identify and isolate faults that

may develop in the Two-Wire In Situ Oxygen Analyzer.

PROBE LIFE The zirconium oxide technology for measuring oxygen is very stable and

should provide accurate service for several years. Life of the probe is negatively impacted by:
• Continued operation at elevated temperatures above 1300°C (2372°F).
• Operation in processes that contain high levels of sulfur, SO acidic compounds.
, or other
2
Operating conditions with simultaneously high levels of SO O
are particularly damaging.
2
The health and accuracy of a given cell is closely related to the resistance, or impedance, of the cell. Figure 5-1 illustrates that the amount of output from a cell for a given O point where cell impedance increases to approximately 100 ohms.
Frequently conduct calibration checks to look for the following conditions:
• Continued degradation of cell slope.
• Sluggish response. (Note how long it takes the cell to respond to the application of calibration check gases.) See Figure 5-2.
The slope will be valid only for the process temperature at which the calibration check gases are flowed, so no adjustments to the electronics are made as a result of a calibration check.
Note that cells exposed to temperatures above 1300°C (2372°F) may lose the ability to measure accurately and respond quickly when returned to the lowest end of the operating temperature range [550°C (1022°F)].
value (represented as slope) will remain very stable to the
2
and low levels of
2
http://www.raihome.com
Page 72
Model 5081FG
Figure 5-1. Slope vs. Impedance
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Figure 5-2. Speed of Response
5-2
Page 73
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

FAULT INDICATIONS The fault conditions for the Two-Wire In Situ Oxygen Analyzer will be indi-

cated by the faulted operation display as shown in Figure 5-3. This screen displays when a fault that invalidates the O error is corrected, the screen will return to a normal operation display unless another error exists.
reading is present. When the
2
Figure 5-3. Faulted Operation Display
5-3
Page 74
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008

IDENTIFYING AND CORRECTING FAULT INDICATIONS

Figure 5-4. Transmitter Terminal Block
A fault in the operation of the Two-Wire In Situ Oxygen Analyzer is indicated by the faulted operation display. If no faults exist the display will indicate NONE. Information on the current fault is found under the DIAGNOSTICS MENU as detailed in Section 3: Startup and Operation.
The following paragraphs describe the faults, possible causes, and corrective actions. Refer to Figure 5-4 as needed for test points and wiring information.
NOTE
Allow adequate time for the oxygen probe to reach its operating temperature [approximately 500°C (932°F)] before investigating a fault. The SHOW FAULTS screen of the DIAGNOSTICS menu will indicate a fault until the unit reaches operating temperature.
NOTE
The probe uses a Type B thermocouple to measure the cell temperature. A Type B thermocouple output table may be useful for troubleshooting.
5-4
Page 75
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 5-5. Fault 1, Open Thermocouple Fault 1, Open Thermocouple
The thermocouple connection is open. The fault displays as shown in Figure 5-5.
1. Refer to Figure 5-4 and check the thermocouple wiring connections at terminals 8 and 10. Ensure the wires are properly connected.
2. Remove power. Disconnect the thermocouple wires (gray and red) from terminals 10 and 8. Measure the continuity across the gray and red thermocouple leads. The measurement should read approximately 1-2 ohms. Larger values indicate the thermocouple is open.
3. If the thermocouple is open, replace the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
Model 5081FG
Figure 5-6. Fault 2, Reversed Thermocouple Fault 2, Reversed Thermocouple Active
The thermocouple connections are reversed. The fault displays as shown in Figure 5-6.
1. Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2. Refer to Figure 5-4. Check the gray (to terminal 10) and red (to terminal 8) wires for the proper placement.
3. Using a multimeter, measure between terminals 8(-) and 10(+). If the reading is negative, the thermocouple wiring is reversed. Rewire as necessary.
4. If the wiring is correct and the probe is at operating temperature, then the transmitter electronics are bad. Replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
5-5
Page 76
Instruction Manual
IM-106-5081, Rev. 2.0
Model 5081FG
Figure 5-7. Fault 3, Shorted Thermocouple Fault 3, Shorted Thermocouple
The thermocouple connections are shorted. The fault displays as shown in Figure 5-7.
1. Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2. Refer to Figure 5-4. Using a multimeter, measure between terminals 8(-) and 10(+).
3. If the multimeter reading, in voltage mode, is between -0.5 and +0.5 mV, the thermocouple is shorted.
4. If the thermocouple is shorted, replace the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
5. If the thermocouple is not shorted, then replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
October 2008
Figure 5-8. Fault 4, High Probe Temperature Fault 4, High Probe Temperature
The probe's temperature has exceeded the maximum cell temperature setpoint. The fault displays as shown in Figure 5-8.
1. If the probe temperature exceeds the maximum cell temperature setpoint, the 4-20 mA signal output will become invalid and go to the default value.
2. Verify that the upper cell temperature setpoint is configured as desired under the PROGRAM MENU in Section 3: Startup and Operation.
5-6
Page 77
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 5-9. Fault 5, O2 Cell Open Fault 5, O2 Cell Open
Model 5081FG
The O
cell connection is open. The fault
2
displays as shown in Figure 5-9.
1. Allow adequate time for the oxygen probe to reach operating temperatures. Probe temperatures below approximately 500°C (932°F) may result in this fault.
2. Refer to Figure 5-4 and check the O
2
wiring connections at terminals 7 and 8. Ensure the wires are properly connected.
NOTE
Check the cell output voltage at the probe terminals -- not at the electronics.
3. Apply low calibration check gas (0.4% O Measure the cell output from the O
2
wires at the probe terminal block. The cell output should be 100 ±20 mV. If no voltage can be measured the cell is open.
4. If the O
cell is open, replace the oxygen
2
probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
cell
cell
).
2
Figure 5-10. Fault 6, Cell Impedance Too High Fault 6, Cell Impedance Too High
The O
cell impedance has exceeded 100 ohms.
2
The fault displays as shown in Figure 5-10.
1. This fault is usually indicated in conjunction with Fault 5, Cell Open. Correcting Fault 5 should correct Fault 6.
2. Fault 6 appears independently, the cell has degraded beyond specification.
3. If the O
cell has become too old, replace
2
the oxygen probe per Oxygen Probe Replacement in Section 6: Maintenance and Service.
5-7
Page 78
Instruction Manual
Model 5081FG
Figure 5-11. Fault 7, Reversed O2 Cell Fault 7, Reversed O2 Cell
IM-106-5081, Rev. 2.0
October 2008
The O displays as shown in Figure 5-11.
cell connections are reversed. The fault
2
1. Refer to Figure 5-4. Check the black (to terminal 7) and white (to terminal 8) wires for the proper placement. Rewire if necessary.
2. Apply the low calibration check gas (0.4% O
3. Using a multimeter measure between terminals 7(+) and 8(-). If the cell output reading is negative, the O reversed.
4. If the wiring is correct, check if the multimeter reading is the same as the reading shown on the O diagnostics screen (Section 3: Startup and Operation).
5. If the reading is different the transmitter electronics are faulty. Replace the faulty analog or CPU board per Electronics Replacement in Section 6: Maintenance and Service.
).
2
cell wiring is
2
CELL mV
2

CALIBRATION PASSES, BUT STILL READS INCORRECTLY

Probe Passes Calibration, O2 Still Reads High

There are fault conditions where no alarm indication is present and the probe passes calibration, but the O
External Reference Air Leak - There may be a leak that is permitting ambient air to mix with the process gases. Since many combustion processes are slightly negative in pressure, ambient air can be drawn into the cell area, biasing the O
1. Make sure that the calibration gas line is capped tightly between calibrations.
Bad Reference Side Cell Electrode - A bad reference side cell electrode can cause an elevated O "Calibration Recommended" alarm and increasing cell impedance readings. A high cell impedance can be calibrated out, but if the impedance continues to increase rapidly, the sensing cell must be replaced.
reading upward.
2
reading. This fault is usually indicated by a frequent
2
reading may still be incorrect:
2
5-8
Page 79
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Section 6 Maintenance and Service

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Electronics Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Oxygen Probe Replacement . . . . . . . . . . . . . . . . . . . . . . . page 6-3

OVERVIEW This section provides the procedures to maintain and service the Model

5081FG Two-Wire In Situ Oxygen Analyzer.
Install all protective equipment covers and safety ground leads after equipment repair or service. Failure to install covers and ground leads could result in serious injury or death.

ELECTRONICS REPLACEMENT

Display Board Replacement

Disconnect and lock out power before working on any electrical components.
Before replacing any electronic components verify that the power to the Model 5081 transmitter is removed. Refer to Section 7: Replacement Parts for replacement part numbers.
Use the following procedure to replace the display board (12, Figure 6-1).
1. Loosen screw (15) until cover lock (16) disengages from the knurled surface of circuit end cap (14). Remove circuit end cap.
2. Remove three screws (13) retaining the electronics in place.
3. Lift display board (12) to disconnect from CPU board (11). The display board is plugged into the CPU board; use care when removing the display board.
4. Align and plug the replacement display board (12) in the desired orientation. Carefully plug the display board into the mating connector on CPU board (11). Ensure the display board is fully seated.
5. Install and tighten all three screws (13).
6. Install circuit end cap (14). Tighten cover lock screw (15) until cover lock (16) engages knurled surface of circuit end cap (14).
http://www.raihome.com
Page 80
Model 5081FG
Figure 6-1. Two-Wire In Situ Oxygen Analyzer - Exploded View
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
6-2
Page 81
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG

Spare Board Stack Replacement

The spare board stack is composed of the analog board (10, Figure 6-1) and the CPU board (11). Use the following procedure to replace either of these boards.
1. Loosen cover lock screw (15) until cover lock (16) disengages from the knurled surface of circuit end cap (14). Remove circuit end cap.
2. Remove three screws (13).
3. Lift display board (12) to disconnect from CPU board (11). The display board is plugged into the CPU board; use care when removing the display board.
4. Remove terminal end cap (5).
5. Remove two screws (4) and lockwashers (3). Lift terminal block (6) until the analog board (10) is unplugged from the terminal board.
6. Lift failed board stack from housing (1) by the standoffs. Reinstall terminal block (6), lockwashers (3), and screws (4).
NOTE
Model 5081 analyzers shipped after August 2008 incorporate a new display board. If display board replacement is needed for an earlier analyzer unit, both the display and spare board stack must be replaced as matching parts. The earlier model analyzer uses a ribbon cable to connect the display board (12) to the CPU board (10).

OXYGEN PROBE REPLACEMENT

7. Install new spare board stack into housing (1). Carefully seat the analog board onto housing pins. Press firmly on the CPU board standoffs to ensure good contact.
8. Align and plug the replacement display board (12) into the mating connector on the CPU board (11). Ensure the display board is fully seated.
9. Install and tighten all three screws.
10. Install terminal end cap (5) and circuit end cap (14). Tighten cover lock screw (15) until cover lock (16) engages knurled surface of circuit end cap (14).
The oxygen probe is designed with ceramic materials to provide maximum life at elevated temperatures and is not rebuildable. The condition of the sensing cell can be determined periodically by two methods:
• Note the cell impedance at the electronics. When the impedance displays a warning indication (WARN), increase the frequency of impedance readings. A cell with a sustained high impedance indication (HI) indicates a probe that is beyond its useful life.
• Conduct a calibration check. Follow the prompts provided by the electronics through the process of flowing two calibration check gases of known values. Record the generated slope and constant values.
6-3
Page 82
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
Use heat resistant gloves and clothing when removing the probe. The probe can be as hot
°C (2912°F). This can cause severe burns.
as 1600
Do not install or remove probes from a process where pressures are more than a few inches of H personal injury.
Do not insert or withdraw a probe into or out of a hot process faster than 1 in. (25,4 mm) per minute or instrument damage from thermal shock may occur.
Also, ash, slag, or other materials can build up on the probe body in some applications. If this buildup is causing difficulty when withdrawing the probe, DO NOT FORCE. Rotate the probe back and forth to attempt to loosen the material on the probe body. Or, wait until the process cools down and access the buildup from inside the furnace.
O positive pressure. Hot gases may escape from the stack and cause severe
2
Probe replacement may be conducted online as long as the process in which the probe is mounted is operating at a negative, or slightly positive, pressure. Refer to Section 5: Troubleshooting for more information.
Refer to Table 7-1 for replacement probe part numbers. Before replacing the probe, verify that the reference air and calibration check gas lines are turned off and disconnected from the probe.
1. Remove the end cap of the probe to expose the terminal block.
2. Refer to Figure 6-2. Disconnect the four wires (two oxygen signal wires and two thermocouple wires) from the terminal block.
6-4
3. Disconnect the reference air and the calibration check gas lines.
4. Unscrew the probe from the stack and remove.
5. Using a replacement probe, refer to Mechanical Installation in Section 2: Installation for mechanical installation instructions.
6. Refer to Electrical Installation in Section 2: Installation for electrical installation instructions.
7. Refer to Pneumatic Installation in Section 2: Installation for reference air and calibration check gas installation instructions.
Page 83
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Figure 6-2. Oxygen Probe Terminal Block
Model 5081FG
6-5
Page 84
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
6-6
Page 85
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Section 7 Replacement Parts

Table 7-1. Replacement Parts List
Figure and
Index Number
1-1, 2 6A00093G01 Model 5081 Transmitter ATEX 1-1, 2 6A00093G02 Model 5081 Transmitter CSA 1-1, 2 6A00093G03 Model 5081 Transmitter FM 1-1, 3 5R10092G01 20" Replacement Oxygen Probe, with Alumina Outer
1-1, 3 5R10092G02 26" Replacement Oxygen Probe, with Alumina Outer
1-1, 3 5R10092G03 38" Replacement Oxygen Probe, with Alumina Outer
1-1, 3 5R10092G09 20" Replacement Oxygen Probe, with Inconel 600 Outer
1-1, 3 5R10092G10 26" Replacement Oxygen Probe, with Inconel 600 Outer
1-1, 3 5R10092G11 38" Replacement Oxygen Probe, with Inconel 600 Outer
6-1, 6 1A99777H04 Terminal Block
6-1, 10, 11,
12 and 13
6-1, 12 1A99777H05 Display Board Assembly 6-1, 14 23593-01 Circuit End Cap (with Glass)
6-1, 5 1A99777H01 Terminal End Cap (witout Glass)
Part Number Description
Protection Tube
Protection Tube
Protection Tube
Protection Tube
Protection Tube
Protection Tube
1A99777H06 Spare Board Stack, HART Compatible, with Display
Board Assembly
Model 5081FG
NOTE
Model 5081 analyzers shipped after August 2008 incorporate a new display board. If display board replacement is needed for an earlier analyzer unit, both the display and spare board stack must be replaced. The earlier model analyzer uses a ribbon cable to connect the display board (12) to the CPU board (11).
http://www.raihome.com
Page 86
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
7-2
Page 87
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008

Appendix A Safety Data

Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page A-2
Safety Data Sheet for Ceramic Fiber Products . . . . . . . . . . . . . page A-24
Model 5081FG
http://www.raihome.com
Page 88
Model 5081FG

SAFETY INSTRUCTIONS IMPORTANT

SAFETY INSTRUCTIONS FOR THE WIRING
AND INSTALLATION OF THIS APPARATUS
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non-EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazardous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment - and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Certain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the following standards: EN61010-1, IEC417, and ISO3864.
11. Where equipment or labels are marked "Do Not Open While Energized" or similar, there is a danger of ignition in areas where an explosive atmosphere is present. This equipment should only be opened when power is removed and adequate time as specified on the label or in the instruction manual has been allowed for the equipment to cool down ­and then only by trained service personnel.
circuits carrying conductors during a fault situation. The
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
A-2
Page 89
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
DŮLEŽITÉ
Bezpečnostní pokyny pro zapojení a instalaci zařízení Následující bezpečnostní pokyny se speciálně vztahují na všechny
členské státy EU. Pokyny by měly být přísně dodržovány, aby se zajistilo splnění Směrnice o nízkém napětí. Pokud nejsou pokyny nahrazeny místními či národními normami, měly by je dodržovat i nečlenské státy EU.
1. U všech zemnicích bodů, interních a externích, by mělo být vytvořeno odpovídající uzemnění.
2. Po instalaci nebo odstranění problémů musí být vyměněny všechny bezpečnostní kryty a uzemnění. Vždy musí být zajištěna integrita všech zemnicích svorek.
3. Sí˙ové kabely by měly odpovídat požadavkům normy IEC227 nebo IEC245.
4. Všechna zapojení by měla být vhodná pro použití při vnějších teplotách nad 75 °C.
5. Všechna použitá kabelová hrdla by měla mít takové vnitřní rozměry, aby zajistila odpovídající zakotvení kabelu.
6. Správnou činnost zařízení zajistíte, vytvoříte-li připojení k napájecímu zdroji pouze přes jistič, který v případě poruchy odpojí všechny s konduktory. Jistič může také obsahovat mechanický odpojovač. Pokud ho neobsahuje, musí být zajištěn a jasně označen jiný způsob odpojení zařízení od zdroje. Jističe nebo přepínače musí odpovídat uznávaným normám, např. IEC947. Všechna zapojení musí odpovídat místním normám.
7. Je-li zařízení nebo kryt označen symbolem na pravé straně, pravděpodobně se uvnitř nachází nebezpečné napětí. Tyto kryty by měly být sejmuty pouze po odpojení zařízení od zdroje - a to pouze kvalifikovaným zaměstnancem.
8. Je-li zařízení nebo kryt označen symbolem na pravé straně, povrch zařízení může být velmi horký. Tyto kryty by měly být sejmuty pouze kvalifikovaným zaměstnancem po odpojení zařízení od zdroje. Některé povrchy mohou být stále horké.
9. Je-li zařízení nebo kryt označen symbolem na pravé straně, přečtěte si nejprve instrukce v návodu k obsluze.
10. Všechny grafické symboly používané u výrobku pocházejí z následujících norem: EN61010-1, IEC417 a ISO3864.
11. Pokud je zařízení nebo štítky označeno varováním „Je-li za napětím, neotvírejte jej“ či podobným, může dojít ve výbušném prostředí ke vznícení. Zařízení lze otevřít pouze po jeho odpojení od zdroje a ponechání dostatečného času na vychladnutí, jak je uvedeno na štítku nebo v návodu k obsluze - a to pouze kvalifikovaným zaměstnancem.
obvody
řízení pod
A-3
Page 90
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
VIGTIGT
Sikkerhedsinstruktion for tilslutning og installering af dette udstyr. Følgende sikkerhedsinstruktioner gælder specifikt i alle
EU-medlemslande. Instruktionerne skal nøje følges for overholdelse af Lavsspændingsdirektivet og bør også følges i ikke EU-lande medmindre andet er specificeret af lokale eller nationale standarder.
1. Passende jordforbindelser skal tilsluttes alle jordklemmer, interne og eksterne, hvor disse forefindes.
2. Efter installation eller fejlfinding skal alle sikkerhedsdæksler og jordforbindelser reetableres.
3. Forsyningskabler skal opfylde krav specificeret i IEC227 eller IEC245.
4. Alle ledningstilslutninger skal være konstrueret til omgivelsestemperatur højere end 75°C.
5. Alle benyttede kabelforskruninger skal have en intern dimension, så passende kabelaflastning kan etableres.
6. For opnåelse af sikker drift og betjening skal der skabes beskyttelse mod indirekte berøring gennem afbryder (min. 10A), som vil afbryde alle kredsløb med elektriske ledere i fejlsitua-tion. Afbryderen skal indholde en mekanisk betjent kontakt. Hvis ikke skal anden form for afbryder mellem forsyning og udstyr benyttes og mærkes som sådan. Afbrydere eller kontakter skal overholde en kendt standard som IEC947.
7. Hvor udstyr eller dæksler er mærket med dette symbol, er farlige spændinger normalt forekom-mende bagved. Disse dæksler bør kun afmonteres, når forsyningsspændingen er frakoblet - og da kun af instrueret servicepersonale.
8. Hvor udstyr eller dæksler er mærket med dette symbol, forefindes meget varme overflader bagved. Disse dæksler bør kun afmonteres af instrueret servicepersonale, når forsyningsspænding er frakoblet. Visse overflader vil stadig være for varme at berøre i op til 45 minutter efter frakobling.
9. Hvor udstyr eller dæksler er mærket med dette symbol, se da i betjeningsmanual for instruktion.
10. Alle benyttede grafiske symboler i dette udstyr findes i én eller flere af følgende standarder:- EN61010-1, IEC417 & ISO3864.
11. Når udstyr eller etiketter er mærket "Må ikke åbnes, mens udstyret tilføres strøm” eller lignende, er der fare for antændelse i områder, hvor der er en eksplosiv atmosfære. Dette udstyr må kun åbnes, når strømkilden er fjernet, og der er gået tilstrækkelig tid til, at udstyret er kølet ned. Den nødvendige tid hertil er angivet på etiketten eller i brugervejledningen. Udstyret må kun åbnes af en faglært person.
A-4
Page 91
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
BELANGRIJK
Veiligheidsvoorschriften voor de aansluiting en installatie van dit toestel. De hierna volgende veiligheidsvoorschriften zijn vooral bedoeld voor de
EU lidstaten. Hier moet aan gehouden worden om de onderworpenheid aan de Laag Spannings Richtlijn (Low Voltage Directive) te verzekeren. Niet EU staten zouden deze richtlijnen moeten volgen tenzij zij reeds achterhaald zouden zijn door plaatselijke of nationale voorschriften.
1. Degelijke aardingsaansluitingen moeten gemaakt worden naar alle voorziene aardpunten, intern en extern.
2. Na installatie of controle moeten alle veiligheidsdeksels en -aardingen terug geplaatst worden. Ten alle tijde moet de betrouwbaarheid van de aarding behouden blijven.
3. Voedingskabels moeten onderworpen zijn aan de IEC227 of de IEC245 voorschriften.
4. Alle bekabeling moet geschikt zijn voor het gebruik in omgevingstemperaturen, hoger dan 75°C.
5. Alle wartels moeten zo gedimensioneerd zijn dat een degelijke kabel bevestiging verzekerd is.
6. Om de veilige werking van dit toestel te verzekeren, moet de voeding door een stroomonderbreker gevoerd worden (min 10A) welke alle draden van de voeding moet onderbreken. De stroomonderbreker mag een mechanische schakelaar bevatten. Zoniet moet een andere mogelijkheid bestaan om de voedingsspanning van het toestel te halen en ook duidelijk zo zijn aangegeven. Stroomonderbrekers of schakelaars moeten onderworpen zijn aan een erkende standaard zoals IEC947.
7. Waar toestellen of deksels aangegeven staan met het symbool is er meestal hoogspanning aanwezig. Deze deksels mogen enkel verwijderd worden nadat de voedingsspanning werd afgelegd en enkel door getraind onderhoudspersoneel.
8. Waar toestellen of deksels aangegeven staan met het symbool is er gevaar voor hete oppervlakken. Deze deksels mogen enkel verwijderd worden door getraind onderhoudspersoneel nadat de voedingsspanning verwijderd werd. Sommige oppper-vlakken kunnen 45 minuten later nog steeds heet aanvoelen.
9. Waar toestellen of deksels aangegeven staan met het symbool gelieve het handboek te raadplegen.
10. Alle grafische symbolen gebruikt in dit produkt, zijn afkomstig uit een of meer van devolgende standaards: EN61010-1, IEC417 en ISO3864.
11. Op plaatsen waar uitrusting of etiketten zijn voorzien van een melding als "Niet openen bij aanwezigheid van spanning" bestaat er brandgevaar in omgevingen waar een explosieve atmosfeer aanwezig is. Deze uitrusting mag uitsluitend worden geopend wanneer het niet meer onder spanning staat en de uitrusting gedurende de voorgeschreven tijd op het etiket of in de handleiding is afgekoeld - en dan uitsluitend door voldoende opgeleid onderhoudspersoneel.
A-5
Page 92
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
BELANGRIJK
Veiligheidsinstructies voor de bedrading en installatie van dit apparaat. Voor alle EU lidstaten zijn de volgende veiligheidsinstructies van
toepassing. Om aan de geldende richtlijnen voor laagspanning te voldoen dient men zich hieraan strikt te houden. Ook niet EU lidstaten dienen zich aan het volgende te houden, tenzij de lokale wetgeving anders voorschrijft.
1. Alle voorziene interne- en externe aardaansluitingen dienen op adequate wijze aangesloten te worden.
2. Na installatie, onderhouds- of reparatie werkzaamheden dienen alle beschermdeksels /kappen en aardingen om reden van veiligheid weer aangebracht te worden.
3. Voedingskabels dienen te voldoen aan de vereisten van de normen IEC 227 of IEC 245.
4. Alle bedrading dient geschikt te zijn voor gebruik bij een omgevings temperatuur boven 75°C.
5. Alle gebruikte kabelwartels dienen dusdanige inwendige afmetingen te hebben dat een adequate verankering van de kabel wordt verkregen.
6. Om een veilige werking van de apparatuur te waarborgen dient de voeding uitsluitend plaats te vinden via een meerpolige automatische zekering (min.10A) die alle indien een foutconditie optreedt. Deze automatische zekering mag ook voorzien zijn van een mechanisch bediende schakelaar. Bij het ontbreken van deze voorziening dient een andere als zodanig duidelijk aangegeven mogelijkheid aanwezig te zijn om de spanning van de apparatuur af te schakelen. Zekeringen en schakelaars dienen te voldoen aan een erkende standaard zoals IEC 947.
7. Waar de apparatuur of de beschermdeksels/kappen gemarkeerd zijn met het volgende symbool, kunnen zich hieronder spanning voerende delen bevinden die gevaar op kunnen leveren. Deze beschermdeksels/ kappen mogen uitsluitend verwijderd worden door getraind personeel als de spanning is afgeschakeld.
8. Waar de apparatuur of de beschermdeksels/kappen gemarkeerd zijn met het volgende symbool, kunnen zich hieronder hete oppervlakken of onderdelen bevinden. Bepaalde delen kunnen mogelijk na 45 min. nog te heet zijn om aan te raken.
9. Waar de apparatuur of de beschermdeksels/kappen gemarkeerd zijn met het volgende symbool, dient men de bedieningshandleiding te raadplegen.
10. Alle grafische symbolen gebruikt bij dit produkt zijn volgens een of meer van de volgende standaarden: EN 61010-1, IEC 417 & ISO 3864.
11. Op plaatsen waar uitrusting of etiketten zijn voorzien van een melding als "Niet openen bij aanwezigheid van spanning" bestaat er brandgevaar in omgevingen waar een explosieve atmosfeer aanwezig is. Deze uitrusting mag uitsluitend worden geopend wanneer het niet meer onder spanning staat en de uitrusting gedurende de voorgeschreven tijd op het etiket of in de handleiding is afgekoeld - en dan uitsluitend door voldoende opgeleid onderhoudspersoneel.
spanningvoerende geleiders verbreekt
A-6
Page 93
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
WICHTIG
Sicherheitshinweise für den Anschluß und die Installation dieser Geräte. Die folgenden Sicherheitshinweise sind in allen Mitgliederstaaten der
europäischen Gemeinschaft gültig. Sie müssen strickt eingehalten werden, um der Niederspannungsrichtlinie zu genügen. Nichtmitgliedsstaaten der europäischen Gemeinschaft sollten die national gültigen Normen und Richtlinien einhalten.
1. Alle intern und extern vorgesehenen Erdungen der Geräte müssen ausgeführt werden.
2. Nach Installation, Reparatur oder sonstigen Eingriffen in das Gerät müssen alle Sicherheitsabdeckungen und Erdungen wieder installiert werden. Die Funktion aller Erdverbindungen darf zu keinem Zeitpunkt gestört sein.
3. Die Netzspannungsversorgung muß den Anforderungen der IEC227 oder IEC245 genügen.
4. Alle Verdrahtungen sollten mindestens bis 75°C ihre Funktion dauerhaft erfüllen.
5. Alle Kabeldurchführungen und Kabelverschraubungen sollten in Ihrer Dimensionierung so gewählt werden, daß diese eine sichere Verkabelung des Gerätes ermöglichen.
6. Um eine sichere Funktion des Gerätes zu gewährleisten, muß die Spannungsversorgung über mindestens 10 A abgesichert sein. Im Fehlerfall muß dadurch gewährleistet sein, daß die Spannungsversorgung zum Gerät bzw. zu den Geräten unterbrochen wird. Ein mechanischer Schutzschalter kann in dieses System integriert werden. Falls eine derartige Vorrichtung nicht vorhanden ist, muß eine andere Möglichkeit zur Unterbrechung der Spannungszufuhr gewährleistet werden mit Hinweisen deutlich gekennzeichnet werden. Ein solcher Mechanismus zur Spannungsunterbrechung muß mit den Normen und Richtlinien für die allgemeine Installation von Elektrogeräten, wie zum Beispiel der IEC947, übereinstimmen.
7. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, die eine gefährliche (Netzspannung) Spannung führen. Die Abdeckungen dürfen nur entfernt werden, wenn die Versorgungsspannung unterbrochen wurde. Nur geschultes Personal darf an diesen Geräten Arbeiten ausführen.
8. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, in bzw. unter denen heiße Teile vorhanden sind. Die Abdeckungen dürfen nur entfernt werden, wenn die Versorgungsspannung unterbrochen wurde. Nur geschultes Personal darf an diesen Geräten Arbeiten ausführen. Bis 45 Minuten nach dem Unterbrechen der Netzzufuhr können derartig Teile noch über eine erhöhte Temperatur verfügen.
9. Mit dem Symbol sind Geräte oder Abdeckungen gekennzeichnet, bei denen vor dem Eingriff die entsprechenden Kapitel im Handbuch sorgfältig durchgelesen werden müssen.
10. Alle in diesem Gerät verwendeten graphischen Symbole entspringen einem oder mehreren der nachfolgend aufgeführten Standards: EN61010-1, IEC417 & ISO3864.
11. Wenn Geräte oder Etiketten mit dem Hinweis "Nicht unter Spannung öffnen" oder ähnlichen Hinweisen versehen sind, besteht in explosionsgefährdeten Umgebungen Entzündungsgefahr. Das Gerät darf nur geöffnet werden, wenn es nicht ans Stromnetz angeschlossen und entsprechend der Zeitangaben auf dem Etikett bzw. in der Betriebsanleitung ausreichend abgekühlt ist. Das Gerät darf nur von geschultem Service-Personal geöffnet werden.
A-7
Page 94
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
ΣΗΜΑΝΤΙΚΟ
Οδηγιεσ ασφαλειασ για την καλωδιωση και εγκατασταση τησ συσκευησ Οι ακόλουθες οδηγίες ασφαλείας εφαρµόζονται ειδικά για όλες τις χώρες
µέλη της Ευρωπαϊκής Κοινότητας. Θα πρέπει να ακολουθούνται αυστηρά ώστε να εξασφαλιστεί η συµβατότητα µε τις οδηγίες για τη Χαµηλή Τάση. Χώρες που δεν είναι µέλη της Ευρωπαϊκής να ακολουθούν τις οδηγίες, εκτός εάν αυτές αντικαθίστανται από τα Τοπικά ή Εθνικά πρότυπα.
1. Επαρκείς συνδέσεις γείωσης θα πρέπει να γίνονται σε όλα τα σηµ εία γείωσης, εσωτερικά και εξωτερικά, όπου υπάρχουν.
2. Μετά την εγκατάσταση ή την αντιµετώπιση σφαλµάτων, όλα τα καλύµµατα ασφαλείας και οι γειώσεις ασφαλείας πρέπει να επανεγκαθίστανται. Η καλή κατάσταση όλων των ακροδεκτών γείωσης πρέπει να συντηρείται διαρκώς.
3. Τα καλώδια τροφοδοσίας πρέπει να πληρούν τις απαιτήσεις των IEC227 ή IEC245.
4. Όλες οι καλωδιώσεις θα πρέπει να είναι κατάλληλες για χρήση σε θερµοκρασία χώρου υψηλότερη από 75°C.
5. Όλοι οι στυπιοθλίπτες θα διαστάσεων, ώστε να παρέχουν επαρκή στερέωση των καλωδίων.
6. Για τη διασφάλιση ασφαλούς λειτουργίας αυτής της συσκευής, η σύνδεση τροφοδοσίας θα πρέπει να γίνεται µόνο µέσω ασφαλειοδιακόπτη, ο οποίος θα αποσυνδέει όλους κυκλωµάτων, στη διάρκεια κατάστασης σφάλµατος. Ο ασφαλειοδιακόπτης µπορεί επίσης να περιλαµβάνει µηχανικό διακόπτη αποµόνωσης. Εάν δεν περιλαµβάνει, τότε άλλα µέσα αποσύνδεσης της συσκευής από την τροφοδοσία πρέπει να παροχηθούν και να σηµανθούν σαφώς ως τέτοια. Οι ασφαλειοδιακόπτες ή διακόπτες αναγνωρισµένα πρότυπα όπως το IEC947. Όλες οι καλωδιώσεις πρέπει να συµµορφώνονται µε τα τοπικά πρότυπα.
7. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το σύµβολο που εικονίζεται δεξιά, επικίνδυνες τάσεις ενυπάρχουν κάτω από αυτά. Αυτά τα καλύµµατα θα πρέπει να αφαιρούνται µόνο όταν έχει αφαιρεθεί η τροφοδοσία από τη συσκευή - και στην περίπτωση αυτή,
µόνο από ειδικευµένο τεχνικό προσωπικό.
8. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το
σύµβολο που εικονίζεται δεξιά, υπάρχει κίνδυνος από καυτές επιφάνειες κάτω από αυτά. Τέ το ια καλύµµατα πρέπει να αφαιρούνται µόνο από ειδικευµένο τεχνικό προσωπικό, όταν έχει αφαιρεθεί η τροφοδοσία από τη συσκευή. Κάποιες επιφάνειες µπορούν να παραµένουν ζεστές στην αφή.
9. Όπου συσκευές ή καλύµµατα είναι σηµασµένα µε το σύµβολο που εικονίζεται δεξιά, ανατρέξτε στις οδηγίες χρήσης της
10. Όλα τα γραφικά σύµβολα που χρησιµοποιούνται σε αυτό το προϊόν είναι από ένα ή περισσότερα από τα εξής πρότυπα: EN61010-1, IEC417 και ISO3864.
11. Όπου συσκευή ή ετικέτα είναι σηµασµένη µε την ένδειξη "Μην ανοίγετε ενώ βρίσκεται σε λειτουργία" ή άλλη παρόµοια, υπάρχει κίνδυνος ανάφλεξης σε περιοχές ανοίγεται µόνο όταν είναι εκτός ρεύµατος και αφού παρέλθει ο κατάλληλος χρόνος που αναγράφεται στην ετικέτα ή στο εγχειρίδιο οδηγιών ώστε να ψυχθεί και µόνο από εκπαιδευµένο προσωπικό συντήρησης.
συσκευής.
µε εκρηκτική ατµόσφαιρα. Ο παρών εξοπλισµός πρέπει να
πρέπει να είναι τέτοιων εσωτερικών
τους ηλεκτροφόρους αγωγούς των
Κοινότητας θα πρέπει επίσης
πρέπει να συµµορφώνονται µε
θα
A-8
Page 95
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
OLULINE TEAVE
Juhtmestiku ja seadme paigaldamisega seotud ohutusjuhised Alljärgnevad ohutusjuhised rakenduvad eriti kõigi Euroopa Liidu
liikmesriikide suhtes. Antud juhiseid tuleb täpselt järgida, et kindlustada vastavus madalpinge direktiiviga. Euroopa Liitu mittekuuluvad riigid peavad samuti alljärgnevaid juhiseid järgima, va juhul, kui on olemas vastavad kohalikud riiklikud standardid.
1. Ettenähtud maanduspunktide, nii sisemiste kui väliste jaoks tuleb tagada nõuetekohased maaühendused.
2. Pärast paigaldamist või rikketuvastust tuleb kõik turvaümbrised ja turvamaandused uuesti oma kohale seada. Kõigis olukordades tuleb säilitada kõigi maandusklemmide terviklikkus.
3. Toitejuhtmed peavad vastama IEC227 või IEC245 nõuetele.
4. Kogu juhtmestik peab sobima kasutamiseks üle 75°C õhutemperatuuri juures.
5. Kõik juhtmetihendid peavad sisemõõtmete poolest tagama nõuetekohased kaabliühendused.
6. Seadme ohutu töötamise tagamiseks peab ühendus toiteallikaga toimuma vaid läbi automaatkorgi, mis veaolukorras lülitab välja kõik voolukandjad. Automaatkorgil võib olla ka mehhaaniliselt reguleeritav lahklüliti. Vastasel juhul peab seadme toiteallikast lahtiühendamiseks olema teine ja selgelt osutatud moodus. Automaatkorgid või -lülitid peavad vastama tunnustatud standarditele nagu nt IEC947. Kogu juhtmestik peab vastama kohalikele standarditele.
7. Seadmel või ümbristel asuv paremale osutav sümbol tähistab selle all leiduvat ohtlikku pinget. Selliste sümbolitega ümbriseid võib eemaldada vaid juhul, kui seade on toiteallikast lahti ühendatud ning ka siis ainult vastavate oskustega spetsialisti poolt.
8. Seadmele või ümbristele märgitud paremale osutava sümboli all valitseb kuumadest pindadest tulenev oht. Nimetatud sümbolitega ümbriseid võib eemaldada vaid vastavate oskustega spetsialist, kui seade on toiteallikast lahti ühendatud. Teatud pinnad võivad puudutamise jaoks liiga kuumad olla.
9. Seadmel või ümbristel leiduva paremale osutava sümboli korral vt juhiste jaoks Toimimisjuhendit.
10. Kõik selle toote juures kasutatavad graafilised sümbolid lähtuvad ühest või enamast järgmistest standarditest: EN61010-1, IEC417 ja ISO3864.
11. Kui seadmele või siltidele on kirjutatud "Ärge avage voolutarbimine korral" vms, valitseb plahvatusohtlikus keskkonnas süttimise oht. Seadet võib avada ainult siis, kui toide on lahti ühendatud ning seadmel on võimaldatud sildil või kasutusjuhendis osutatud aja jooksul maha jahtuda -- ning ka sellisel juhul ainult vastavate oskustega spetsialisti poolt.
A-9
Page 96
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
TÄRKEÄÄ
Turvallisuusohje, jota on noudatettava tämän laitteen asentamisessa ja kaapeloinnissa.
Seuraavat ohjeet pätevät erityisesti EU:n jäsenvaltioissa. Niitä täytyy ehdottomasti noudattaa jotta täytettäisiin EU:n matalajännitedirektiivin (Low Voltage Directive) yhteensopivuus. Myös EU:hun kuulumattomien valtioiden tulee nou-dattaa tätä ohjetta, elleivät kansalliset standardit estä sitä.
1. Riittävät maadoituskytkennät on tehtävä kaikkiin maadoituspisteisiin, sisäisiin ja ulkoisiin.
2. Asennuksen ja vianetsinnän jälkeen on kaikki suojat ja suojamaat asennettava takaisin pai-koilleen. Maadoitusliittimen kunnollinen toiminta täytyy aina ylläpitää.
3. Jännitesyöttöjohtimien täytyy täyttää IEC227 ja IEC245 vaatimukset.
4. Kaikkien johdotuksien tulee toimia >75°C lämpötiloissa.
5. Kaikkien läpivientiholkkien sisähalkaisijan täytyy olla sellainen että kaapeli lukkiutuu kun-nolla kiinni.
6. Turvallisen toiminnan varmistamiseksi täytyy jännitesyöttö varustaa turvakytkimellä (min 10A), joka kytkee irti kaikki jännitesyöttöjohtimet vikatilanteessa. Suojaan täytyy myös sisältyä mekaaninen erotuskytkin. Jos ei, niin jännitesyöttö on pystyttävä katkaisemaan muilla keinoilla ja merkittävä siten että se tunnistetaan sellaiseksi. Turvakytkimien tai kat-kaisimien täytyy täyttää IEC947 standardin vaatimukset näkyvyydestä.
7. Mikäli laite tai kosketussuoja on merkitty tällä merkillä on merkinnän takana tai alla hengenvaarallisen suuruinen jännite. Suojaa ei saa poistaa jänniteen ollessa kytkettynä laitteeseen ja poistamisen saa suorittaa vain alan asian-tuntija.
8. Mikäli laite tai kosketussuoja on merkitty tällä merkillä on merkinnän takana tai alla kuuma pinta. Suojan saa poistaa vain alan asiantuntija kun jännite-syöttö on katkaistu. Tällainen pinta voi säilyä kosketuskuumana jopa 45 mi-nuuttia.
9. Mikäli laite tai kosketussuoja on merkitty tällä merkillä katso lisäohjeita käyt-töohjekirjasta.
10. Kaikki tässä tuotteessa käytetyt graafiset symbolit ovat yhdestä tai useammasta seuraavis-ta standardeista: EN61010-1, IEC417 & ISO3864.
11. Jos laitteessa tai tarrassa on merkintä "Älä avaa, kun virta on kytketty" tai vastaava, räjähdysvaarallisissa tiloissa on syttymisen vaara. Nämä laitteet voidaan avata vain silloin, kun virta ei ole kytkettynä ja laitteen on annettu jäähtyä tarrassa tai oppaassa määritetyn ajan. Tällöinkin laitteet saa avata vain koulutettu huoltohenkilökunta.
A-10
Page 97
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
IMPORTANT
Consignes de sécurité concernant le raccordement et l'installation de cet appareil.
Les consignes de sécurité ci-dessous s'adressent particulièrement à tous les états membres de la communauté européenne. Elles doivent être strictement appliquées afin de satisfaire aux directives concernant la basse tension. Les états non membres de la communauté européenne doivent également appliquer ces consignes sauf si elles sont en contradiction avec les standards locaux ou nationaux.
1. Un raccordement adéquat à la terre doit être effectuée à chaque borne de mise à la terre, interne et externe.
2. Après installation ou dépannage, tous les capots de protection et toutes les prises de terre doivent être remis en place, toutes les prises de terre doivent être respectées en permanence.
3. Les câbles d'alimentation électrique doivent être conformes aux normes IEC227 ou IEC245.
4. Tous les raccordements doivent pouvoir supporter une température ambiante supérieure à 75°C.
5. Tous les presse-étoupes utilisés doivent avoir un diamètre interne en rapport avec les câbles afin d'assurer un serrage correct sur ces derniers.
6. Afin de garantir la sécurité du fonctionnement de cet appareil, le raccordement à l'alimentation électrique doit être réalisé exclusivement au travers d'un disjoncteur (minimum 10A.) isolant tous en cas d'anomalie. Ce disjoncteur doit également pouvoir être actionné manuellement, de façon mécanique. Dans le cas contraire, un autre système doit être mis en place afin de pouvoir isoler l'appareil et doit être signalisé comme tel. Disjoncteurs et interrupteurs doivent être conformes à une norme reconnue telle IEC947.
7. Lorsque les équipements ou les capots affichent le symbole suivant, cela signifie que des tensions dangereuses sont présentes. Ces capots ne doivent être démontés que lorsque l'alimentation est coupée, et uniquement par un personnel compétent.
8. Lorsque les équipements ou les capots affichent le symbole suivant, cela signifie que des surfaces dangereusement chaudes sont présentes. Ces capots ne doivent être démontés que lorsque l'alimentation est coupée, et uniquement par un personnel compétent. Certaines surfaces peuvent rester chaudes jusqu'à 45 mn.
9. Lorsque les équipements ou les capots affichent le symbole suivant, se reporter au manuel d'instructions.
10. Tous les symboles graphiques utilisés dans ce produit sont conformes à un ou plusieurs des standards suivants: EN61010-1, IEC417 & ISO3864.
11. Les équipements comportant une étiquette avec la mention " Ne pas ouvrir sous tension " ou toute autre mention similaire peuvent créer un risque d'incendie dans les environnements explosifs. Ces équipements ne doivent être ouverts que lorsqu'ils sont hors tension et que la durée de refroidissement requise indiquée sur l'étiquette ou dans le manuel d'instructions s'est écoulée. En outre ils ne doivent être ouverts que par un personnel qualifié.
les conducteurs
A-11
Page 98
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
FONTOS
Biztonsági előírások a készülék vezetékeléséhez és üzembeállításához A következő biztonsági előírások kifejezetten vonatkoznak az összes
EU-tagállamra. Ezeket szigorúan be kell tartani a Kisfeszültségű irányelvnek való megfelelés biztosításához. A nem EU-tagállamok szintén tartsák be a következőket, kivéve ha a helyi és nemzeti szabványok azt másként nem írják elő.
1. A megfelelő földelést biztosítani kell az összes rendelkezésre álló földelési ponton, legyen az belső vagy külső.
2. Az üzembeállítás vagy hibaelhárítás után az összes biztonsági burkolatot és biztonsági földvezetéket ki kell cserélni. A földelőkapcsok sértetlenségét mindig biztosítani kell.
3. A tápvezetékeknek eleget kell tenniük az IEC227 vagy IEC245 szabványokban megfogalmazott követelményeknek.
4. Az összes vezetéknek alkalmasnak kell lennie a 75 °C-nál magasabb környezeti hőmérséklet melletti használatra.
5. Az összes használt kábelvezető tömszelencének olyan belső méretűnek kell lennie, hogy biztosítsák a kábelek megfelelő lekötését.
6. A berendezés biztonságos működésének biztosításához az elektromos hálózathoz való csatlakozást csak megszakítón keresztül szabad megvalósítani, amely az összes hibahelyzet esetén. A megszakító magában foglalhat egy mechanikusan működtethető áramtalanító kapcsolót is. Ellenkező esetben biztosítani kell a berendezés elektromos hálózatról történő lekapcsolásának más módját, és ezt világosan jelezni kell. A megszakítóknak vagy kapcsolóknak meg kell felelniük egy elismert szabványnak, például az IEC947 szabványnak. Az összes vezetéknek meg kell felelnie az összes helyi szabványnak.
7. Ha a berendezés vagy a burkolata a jobb oldalon látható szimbólummal jelzett, alatta valószínűleg veszélyes feszültség van jelen. Az ilyen burkolat csak a berendezés áramtalanítása után távolítható el - és csak képzett szervizszakember végezheti el.
8. Ha a berendezés vagy a burkolata a jobb oldalon látható szimbólummal jelzett, fenn áll a veszélye, hogy alatta forró felületek találhatóak. Az ilyen burkolatot csak képzett szervizszakember távolíthatja el a berendezés áramtalanítása után. Bizonyos felületek érintésre forróak maradhatnak.
9. Ha a berendezés vagy a burkolata a jobb oldalon látható szimbólummal jelzett, tekintse meg az Üzemeltetési útmutató arra vonatkozó utasításait.
10. A terméken használt grafikus szimbólumok a következő szabványok legalább egyikéből származnak: EN61010-1, IEC417 és ISO3864.
11. Ha a berendezésen vagy a címkéken a „Ne nyissa ki bekapcsolt állapotban” vagy hasonló felhívás szerepel, robbanásveszélyes környezetben fennáll a gyulladás veszélye. Ez a berendezés csak áramtalanítás után nyitható ki, a címkén vagy a kezelési útmutatóban szereplő, a berendezés lehűlését biztosító megfelelő idői ráhagyás után ­és csak képzett szervizszakember végezheti el.
áramot szállító vezetéket bontja
A-12
Page 99
Instruction Manual
IM-106-5081, Rev. 2.0 October 2008
Model 5081FG
IMPORTANTE
Norme di sicurezza per il cablaggio e l'installazione dello strumento. Le seguenti norme di sicurezza si applicano specificatamente agli stati
membri dell'Unione Europea, la cui stretta osservanza è richiesta per garantire conformità alla Direttiva del Basso Voltaggio. Esse si applicano anche agli stati non appartenenti all'Unione Europea, salvo quanto disposto dalle vigenti normative locali o nazionali.
1. Collegamenti di terra idonei devono essere eseguiti per tutti i punti di messa a terra interni ed esterni, dove previsti.
2. Dopo l'installazione o la localizzazione dei guasti, assicurarsi che tutti i coperchi di protezione siano stati collocati e le messa a terra siano collegate. L'integrità di ciscun morsetto di terra deve essere costantemente garantita.
3. I cavi di alimentazione della rete devono essere secondo disposizioni IEC227 o IEC245.
4. L'intero impianto elettrico deve essere adatto per uso in ambiente con temperature superiore a 75°C.
5. Le dimensioni di tutti i connettori dei cavi utilizzati devono essere tali da consentire un adeguato ancoraggio al cavo.
6. Per garantire un sicuro funzionamento dello strumento il collegamento alla rete di alimentazione principale dovrà essere eseguita tramite interruttore automatico (min.10A), in grado di disattivare tutti i conduttori di circuito in caso di guasto. Tale interruttore dovrà inoltre prevedere un sezionatore manuale o altro dispositivo di interruzione dell'alimentazione, chiaramente identificabile. Gli interruttori dovranno essere conformi agli standard riconosciuti, quali IEC947.
7. Il simbolo riportato sullo strumento o sui coperchi di protezione indica probabile presenza di elevati voltaggi. Tali coperchi di protezione devono essere rimossi esclusivamente da personale qualificato, dopo aver tolto alimentazione allo strumento.
8. Il simbolo riportato sullo strumento o sui coperchi di protezione indica rischio di contatto con superfici ad alta temperatura. Tali coperchi di protezione devono essere rimossi esclusivamente da personale qualificato, dopo aver tolto alimentazione allo strumento. Alcune superfici possono mantenere temperature elevate per oltre 45 minuti.
9. Se lo strumento o il coperchio di protezione riportano il simbolo, fare riferimento alle istruzioni del manuale Operatore.
10. Tutti i simboli grafici utilizzati in questo prodotto sono previsti da uno o più dei seguenti standard: EN61010-1, IEC417 e ISO3864.
11. L'indicazione "Non aprire sotto tensione" o simili sull'apparecchiatura o sulle etichette segnala il pericolo di accensione nelle aree in cui è presente un'atmosfera esplosiva. L'apparecchiatura può essere aperta solo quando l'alimentazione è scollegata ed è trascorso il tempo indicato sull'etichetta o nel manuale delle istruzioni per consentirne il raffreddamento. L'operazione può essere effettuata esclusivamente da personale dell'assistenza qualificato.
A-13
Page 100
Model 5081FG
Instruction Manual
IM-106-5081, Rev. 2.0
October 2008
SVARBU
šio prietaiso laidų prijungimo ir instaliacijos saugos instrukcijos
Toliau išvardinti saugumo reikalavimai taikomi konkrečiai visoms ES šalims narėms. Jų turi būti griežtai paisoma, kad būtų užtikrintai laikomasi Žemos įtampos direktyvos. Ne ES narės taip pat turi laikytis toliau pateikiamų reikalavimų nebent juos pakeičia vietiniai ar Nacionaliniai standartai.
1. Turi būti atliktas tinkamas įžeminimas visuose įžeminimo taškuose, vidiniuose ir išoriniuose, kur numatyta.
2. Visos apsauginės dangos ir įžemikliai po instaliacijos ar remonto turi būti pakeisti. Visų įžeminimo terminalų vientisumo priežiūra turi būti atliekama nuolat.
3. Matinimo tinklo laidai turi atitikti IEC227 ar IEC245 reikalavimus.
4. Visi laidai turi būti tinkami naudojimui aplinkos temperatūtoje, aukštesnėje nei 75°C.
5. Visi naudojamų kabelių riebokšliai turi būti tokių vidinių matmenų, kad būtų galimas tinkamas kabelio pritvirtinimas.
6. Saugaus šio prietaiso veikimo užtikrinimui, prijungimas prie maitinimo tinklo turi būti atliekamas tik per automatinį pertraukiklį, kuris atjungs visas
grandines nešančius konduktorius linijos gedimo metu. Automatinis pertraukiklis taip pat gali turėti mechaniškai veikiantį įzoliavimo jungiklį. Jeigu ne, tuomet turi būti nurodytos kitos įrenginio atjungimo priemonės, ir aiškai pažymėtos, kad jos tokios yra. Automatiniai perjungikliai ar jungikliai turi atitikti pripažintus standartus, tokius kaip IEC947. Visi laidai turi atitikti visus vietinius standartus.
7. Kur įrenginys ar dangos yra pažymėti simboliu dešinėje, žemiau turi būti pavojinga įtampa. Šios dangos turi būti nuimamos tik tada, kai srovė yra pašalinta iš įrenginio ­ir tik tuomet tai turi atlikti apmokytas personalas.
A-14
8. Ten kur įrenginys ar dangos yra pažymėti simboliu dešinėje, ten yra pavojus nuo karštų paviršių apačioje. Šios dangos gali būti nuimamos tik apmokyto personalo, kai srovė yra pašalinta iš įrenginio. Tam tikri paviršiai gali išlikti karšti liečiant.
9. Ten kur įrenginys ar dangos yra pažymėti simboliu dėšinėje, žr. nurodymus Valdymo instrukcijose.
10. Visi grafiniai simboliai naudojami šiam produktui yra iš vieno ar daugiau toliau išvardintų standartų: EN61010-1, IEC417, ir ISO3864.
11. Ten, kur įrenginys ar etiketės yra pažymėti "Neatidaryti esant srovės tiekimui" ar panašiai, yra užsidegimo pavojus tose vietose, kur yra sprogstamoji atmosfera. Šis įrenginys gali būti atidarytas tuomet, kai yra pašalinta srovė, ir praė valdymo instrukcijoje, pakankamas įrenginio ataušimui - ir tai tik apmokyto personalo.
jęs atitinkamas laikas, nurodytas etiketėje ar
Loading...