Rosemount Analytical 410VP Specifications

Rosemount™ 410VP
Four-Electrode Conductivity Sensors
Product Data Sheet
June 2017
LIQ-PDS-410VP
The Rosemount 410VP four-electrode conductivity sensor is primarily intended for the measurement of electrolytic conductivity in the pharmaceutical and food and beverage industries. Typical applications include monitoring the concentration of CIP solutions and rinses, monitoring eluents in chromatographic separations, and detecting liquid interfaces.
Rosemount 410VP Four-Electrode Conductivity Sensor

Overview

Meet Various Sanitary Requirements
Sensor body and elastomers are compliant with 21CFR177; a
certificate of conformance is supplied with each sensor.
3-A Sanitary Standard 74-06 compliant.
Wetted surfaces (except electrodes) have a 16 micro inch
(0.4 micrometer) Ra finish.
Available in Tri-Clamp, Varivent N, and G 1-¼ fittings.
Animal-derived Ingredient (ADI) free optional.
High Performance and Robust Sensor Design
Four-electrode design provides exceptional linearity between
1 μS/cm and 1400 mS/cm.
June 2017
Unfilled PEEK sensor body.
Steam sterilization (SIP) tolerant up to 284 °F (140 °C).
Rapid response to temperature changes with a Pt-1000 RTD.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Engineering Specification . . . . . . . . . . . . . . . . . . . . . . . . . . 7
www.Emerson.com/RosemountLiquidAnalysis
June 2017
Rosemount 410VP Four-Electrode Conductivity Sensor

Ordering Information

The Rosemount 410VP four-electrode conductivity sensor is intended for measuring conductivity in a variety of pharmaceutical and food and beverage applications. All wetted plastics and elastomers are compliant with 21CFR177, and all wetted surfaces except the electrodes have a 16 micro-inch (0.4 micrometer) Ra surface finish. A certificate of conformance is provided with each sensor. Interconnecting VP8 cable must be ordered separately and is required for all first time installations.
Additional Information
Specifications: see “Specifications” on page 3 Dimensional drawings: see “Installation” on page 5 Accessories: see “Accessories” on page 7 Engineering Specifications: see “Engineering Specification” on page 7
Table 1. Rosemount 410VP Four-Electrode Conductivity Sensor ordering information
Model Sensor type
410VP Four-electrode conductivity sensor
Process Connection
20 1½ in. Tri-clamp
21 2 in. Tri-clamp
22 G 1¼
23 Varivent N
Material Specification
_ No selection
40 Free of Animal-Derived Ingredients (ADI-Free)
Special Option
99Q8 Supplier material traceability certificates
Typical Model Number: 410VP-20-40

Specifications

Table 2. Rosemount 410VP Four-Electrode Conductivity Sensor specifications
Wetted Materials
Electrodes 316L stainless steel
Sensor body unfilled PEEK; compliant with 21CFR177.2415
O-ring (option -22 only) EP; compliant with 21CFR177.2600
Conformance to 3-A Sanitary Standards
Sensors with option -20 and -21 meet 3-A sanitary standards for sensors and sensor fittings and connections used on milk and milk products equipment (74-06)
www.Emerson.com/RosemountLiquidAnalysis
3
Loading...
+ 5 hidden pages