Rosemount Analytical 404 Specifications

Rosemount™ 404
Contacting Conductivity Sensors
Product Data Sheet
June 2017
LIQ-PDS-404
Reliable conductivity measurements for your process
Rosemount 404 Contacting Conductivity Sensors

Overview

June 2017
Minimize Start-up and Installation Time
A factory-measured cell constant ensures out-of-the-box
accuracy and no initial calibration requirements.
Available in cell constants of 0.01 and 0.1/cm.
A Robust Sensor Design
The sensors have concentric titanium electrodes separated by
a PEEK insulator.
An EPDM O-ring seals the internal parts of the sensor from the
process liquid.
Meet process compatibility requirements with a choice of
either a PVC or Stainless Steel body.
Maximum operating temperature up to 100 °C.
Maximum pressure rating up to 100 psig.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Engineering Specifications . . . . . . . . . . . . . . . . . . . . . . . . .7
www.Emerson.com/RosemountLiquidAnalysis
June 2017
Rosemount 404 Contacting Conductivity Sensors

Ordering Information

The Rosemount 404 Contacting Conductivity sensor features an integrated flow cell design. The flow through sensor design has a small holdup volume allowing for rapid response to sudden changes in process conductivity and temperature. The sensor must be used in a sidestream sample. Rosemount 404 sensors are available with either a PVC or stainless steel body. The stainless steel version can be disassembled for cleaning, whereas the PVC version cannot be taken apart.
Additional Information
Specifications: see “Specifications” on page 4 Dimensional drawings: see “Dimensional Drawings” on page 5 Accessories: see “Accessories” on page 6 Engineering Specifications: see “Engineering Specifications” on page 7
Table 1. Rosemount 404 Contacting Conductivity Sensor ordering information
Model Sensor type
404 Contacting Conductivity Sensor
Cell constant
11 0.01/cm
12 0.1/cm
Flow cell type
16 PVC
17 Stainless Steel
Temperature Compensation
_ Pt-1000
54 Pt-100
(1)
Options
_ No selection
50 Extended Integral Cable Length (50 ft; 15 m)
Typical Model Number: 404-12-17_-50
1. Recommended for use with Rosemount transmitters 1056, 56, 1057, 1066, and 5081.