Emerson designs, manufactures and tests its products to meet many national and international stan-
dards. Because these sensors are sophisticated technical products, you MUST properly install, use,
nd maintain them to ensure they continue to operate within their normal specifications. The
a
following instructions MUST be adhered to and integrated into your safety program when installing,
using, and maintaining Rosemount products. Failure to follow the proper instructions may cause
any one of the following situations to occur: loss of life; personal injury; property damage; damage
to this sensor; and warranty invalidation.
•Read all instructions prior to installing, operating, and servicing the product.
• If you do not understand any of the instructions, contact your Emerson representative for
clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance
of the product.
•Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to
the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update,
program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson. Unauthorized parts and procedures can affect the product's
performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY.
Third-party substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and
personal injury.
The information contained in this document is subject to change without notice.
DANGER
Hazardous Area InstallationN
This sensor is not Intrinsically Safe. or Explosion Proof. Installations near flammable liquids or in
hazardous area locations must be carefully evaluated by qualified on site safety personnel.
To secure and maintain an intrinsically safe installation, an appropriate transmitter/safety
barrier/sensor combi nation must be used. The installation system must be in accordance with the
governing approval agency (FM, CSA or BASEEFA/CENELEC) hazardous area classification requirements. Consult your transmitter instruc tion manual for details.
Proper installation, operation and servicing of this sensor in a Hazardous Area Instal lation is entirely
the responsibility of the user.
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating
conditions. Application compatibility is entirely the responsibility of the user.
WARNING
Retractable sensors must not be inserted nor retracted when process pressures are in excess of 64 psig
(542kPa) for option 21 or 35 psig (343 kPa) for option 25.
Page 4
CAUTION
Special Conditions for Safe Use
. All pH/ORP sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid
1
the danger due to a build up of an electrostatic charge.
2. All pH/ORP sensor models are intended to be in contact with the process fluid and may not meet the
500V r.m.s. a.c. test to earth.
This must be taken into consideration at installation.
About This Document
This manual contains instructions for installation and operation of the Rosemount 398R and
398RVP pH/ORP Sensors.
The following list provides concerning all revisions of this document.
Rev. LevelDateNotes
001/99This is the initial release of the product manual. The manual has
been reformatted to reflect the Emerson documentation style
and updated to reflect any changes in the product offering.
A07/02Revised multiple drawings.
B08/02Added drawing #40105549, rev. D.
C10/02Revised drawing on page 10.
D08/03Added Silcore information.
E06/04Updated ordering info and added/revised wiring drawings.
F10/05Delete obsolete options.
G01/11Miscellaneous revisions.
H03/17Updated the Ordering Information, Specifications, Product
Certifications, and Emerson logo and Address.
Page 5
Instruction ManualTable of Contents
LIQ-MAN-398R-398RVPMarch 2017
Contents
Section 1: Description and Specifications
1.1 Features and Applications....................................................................................1
EC Declaration of Conformity.............................................................................45
Intrisicallly Safe Sensor Installation Drawing - FM ................................47
Table of Contentsi
Page 6
Table of ContentsInstruction Manual
March 2017LIQ-MAN-398R-398RVP
iiTable of Contents
Page 7
Instruction ManualDescription and Specifications
LIQ-MAN-398R-398RVPMarch 2017
Section 1: Description and Specifications
1.1Features and Applications
Rosemount 398R and 398RVP sensors feature a chemical resistant construction of Tefzel, titanium,
and a TUpH reference junction which is ideal for measuring pH in harsh process liquids. These
sensors can be used to measure pH in sour water strippers, in pulp bleaching towards that use
chlorine dioxide, and in process streams containing a variety of organic solvents. These sensors are
designed for use with a 1-1/4 inch or 1-1/2 inch ball valve.
Figure 1-1: Cross Section Diagram of the TUpH Reference Technology
All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an
AccuGlass pH glass bulb. This sensor technology ensures superior performance while only requiring
minimal maintenance.
Hemi bulb: 250 psig (1825 kPa abs) at 212°F (100°C)
Flat bulb: 100 psig (790 kPa abs) at 212°F (100°C)
Wetted Materials
Titanium, Tefzel®, glass, platinum (ORP only), and choice of Viton® or Kalrez®
Reference
Permeable Tefzel with secondary helical pathway
Weight/Shipping Weight
SensorCode 21; 2.0 lb/3.0 lb (.9kg/1.40kg)
Code 25; 3.0 lb/4.0 lb (1.40 kg/1.80kg)
Process Connections
With ball valve1-1/2 in.
Without ball valve1 in.
Cable Length
Rosemount 398RStandard 15 ft. Integral Cable; Optional 9.5 inch Cable for wiring to Sensor Head Junction
Boxes
Rosemount 398RVPMust use VP interconnect cable (sold separately)
2Description and Specifications
Page 9
Instruction ManualDescription and Specifications
LIQ-MAN-398R-398RVPMarch 2017
1.3Product Certifications
Please see online certificates for further details.
IECEx
Sensors with no preamp (pH and ORP) – Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (398RVP pH only) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Per standards IEC60079-0 : 2011, IEC 60079-11 : 2011
ATEX
Sensors without preamp (pH and ORP) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (398RVP pH only ) – II 1 G Ex ia IIC T4 Ga (-20˚C ≤ Ta ≤ +60 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
FM
See online FM Certificate of Compliance for applicable sensor options:
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature
Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C
to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C
to +60 °C Hazardous (Classified) Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60°C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity;
NI/I/2/ABCD/T6 Ta = 60 °C; S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
See online CSA Certificate of Compliance for applicable sensor options:
Sensors with preamp – Intrinsically Safe:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups
ABCD; Ambient temperature rating -20 °C to +60 °C; Ex ia IIC; T6
Sensors without preamp – Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups
ABCD; Ex ia IIC; T6; Ambient temperature rating -20 °C to +60 °C: (Simple Apparatus)
Inspect the outside of the carton for any damage. If damage is detected, contact the carrier
immediately. Inspect the instrument and hardware. Make sure all items in the packing list are
present and in good condition. Notify the factory if any part is missing.
Note: If the sensor is to be stored, the protective boot should be filled with either KCl electrolyte
solution or pH 4.0 buffer solution and replaced on sensor tip until ready to use.
Note: Save the original packing cartons and materials as most carriers require proof of damage due
to mishandling, etc. Also, if it is necessary to return the instrument to the factory, you must pack
the instrument in the same manner as it was received. Refer to Section 6.0 for instructions.
WARNING
Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.
2.3Mechanical Installation
Both models may be installed through a weldalet or in a pipe tee or “Y”, as shown in Figure 2-2,
when used with a ball valve. Insert the end of the sensor to a depth sufficient to ensure that the
glass bulb is continuously wetted by the process fluid. Each model can also be inserted directly into
the process without the use of a ball valve for appli cations not requiring continuous operation
during sensor maintenance.
CAUTION
Allow sufficient room for safe retraction and insertion of the sensor. Personnel should have room for
stable footing while performing removal or insertion of the sensor.
The sensor must be mounted within 10-90 degrees of the horizontal with the tip pointed
downward. This ensures that the inside surface of the pH-sensitive glass bulb is completely wetted
and that there is a good electrical connection between the bulb and the internal silver/silver
chloride reference electrode. If the retraction version is to be installed without a ball valve follow the
installation procedure for insertion service (Section 2.3.2). Perform the following steps for sensor
installation through a ball valve:
2.3.1Installation Through Ball Valve
1.Carefully remove the liquid filled rubber boot which protects the glass electrode and keeps
the liquid junction wet during shipping and storage. Discard the liquid and boot. Make sure
the lubricated O-ring is in place in the groove inside the male connector on the sensor body.
CAUTION
Buffer solution, in the protective boot, may cause skin or eye irritation.
2.With the male connector on the sensor’s body, insert the sensor into the ball valve until it
gently touches the closed valve. The molded electrode guard will protect the glass bulb
from breakage. (extra caution should be taken when inserting the flat glass sensor into the
valve because it does not have an electrode guard).
8Installation
Page 15
Instruction ManualInstallation
LIQ-MAN-398R-398RVPMarch 2017
3.Thread the male connector body tightly into the ball valve assembly. DO NOT tighten the
hex nut on the male connector body; doing so would not allow the sensor to be inserted
through the ball valve.
4.Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that
the sensor cannot come free of the ball valve assembly. The built-in retraction stop will butt
against the shoulder of the male connector if properly installed.
CAUTION
The sensor must be captured by the valve assembly and the male connector so that it cannot be blown
free by process pressure if mishandled during insertion or retraction.
5.After confirming that the sensor assembly is properly secured by the valve assembly, the
valve may be opened and the sensor positioned into the process at the desired depth and
orientation.
6.While holding the sensor in position, tighten the hex nut of the male connector to firmly
secure the sensor in place. When the hex nut is tightened, the Teflon ferrule inside the
compression fitting clamps the sensor tube.
CAUTION
Over tightening the hex nut may damage the ferrule.
NOTICE
A stainless steel ferrule is available if the Teflon ferrule does not inadequately grip. When using the
metallic ferrule, care must be taken to avoid over tightening and damaging the sensor tube. If the male
connector leaks during insertion or retraction, replace the O-ring in the male connector.
Figure 2-1: Exploded View of Ball Valve Kit PN 23240-00 used with process connector
PN 23166-00 (or PN 23166-01)
Ball Valve Kit includes 1-1/2 in. x 1 in. reducer, 1-1/2 in. close nipple, and 1-1/2 in. ball valve]
Installation9
Page 16
InstallationInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 2-2: Exploded View of Ball Valve Kit PN 23765-00
2.3.2Installation Without Ball Valve
Rosemount 398R and 398RVP Sensors may be installed through a weldalet or pipe tee or “Y” when
used with a process connector (PN 23166-00 or 23166-01). The sensor should be installed within
80° of vertical, with the electrode facing down.
Figure 2-3: Typical Mounting Details-Retraction Version
Note: Sensor must be mounted at an angle between 10° and 90° above the horizontal.
Pipe tees and weldalets provided by customer.
Only Rosemount 398R should be used with a sensor head junction box.
10Calibration and Maintenance
Page 17
Instruction ManualInstallation
LIQ-MAN-398R-398RVPMarch 2017
igure 2-4: Dimensional Warning Label and Sensor Diagram
F
Note: For Rosemount 398RVP, the overall dimensional length increases by 1.9 inches (48 mm).
Calibration and Maintenance11
Page 18
InstallationInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 2-5: Dimensional Drawing for Ball Valve Kit PN 23765-00
(shown with Rosemount 398RVP sensor)
12Installation
Page 19
Instruction ManualInstallation
LIQ-MAN-398R-398RVPMarch 2017
Figure 2-6: Dimensional Drawing for Ball Valve Kit PN 23240-00
shown with Rosemount 398R).
(
Note: For Rosemount 398RVP, the overall dimensional length increases by 1.9 inches (48 mm).
Installation13
Page 20
InstallationInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 2-7: Dimensional Drawing of Rosemount 398R shown with Sensor Head Junction
Box, with and without 1-1/2 in. Ball Valve PN 23240-00
Note: For the installations shown, the 1 in. x 1 in. process connector (PN 23166-00 or 23166-01), sensor
head junction boxes (various part numbers; see page 3), and ball valve kit (PN 23240-00) must be
purchased separately.
Note: Sensor head junction boxes should be used with Rosemount 398R sensor
14Installation
Page 21
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
Section 3: Wiring the Rosemount 398R/398RVP
Sensor
3.1Wiring for Rosemount 398R/398RVP
For wiring diagrams not shown below, please refer to the Liquid Wiring Diagrams.
Figure 3-1: Wire Configurations
Figure 3-2: Wiring for Rosemount 398R-54-62 to 3081, 4081, and 5081 Transmitters.
Wiring the Rosemount 398R/398RVP Sensor
15
Page 22
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 3-3: Wiring for 398R-54-62 Through Remote Junction Box PN 23555-00
Note: Sensor Model 398R-54 or 398R-54-60 can also be wired as shown above, but customer must
prepare the BNC as shown in Figure 3-4. For preparing wires on end of extension cable, use Figure 3-5.
16Wiring the Rosemount 398R/398RVP Sensor
Page 23
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
Figure 3-4: BNC Preparation Instructions
Wiring the Rosemount 398R/398RVP Sensor17
Page 24
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 3-5: Extension Cable Preparation
18Wiring the Rosemount 398R/398RVP Sensor
Page 25
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
igure 3-6: Rosemount 398R-54-61 Wiring to Rosemount 1056/56 Transmitters
F
Figure 3-7: Rosemount 398R-54-61 Wiring to Rosemount 1057 Transmitter
Wiring the Rosemount 398R/398RVP Sensor19
Page 26
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 3-8: Rosemount 398R-54-61 Wiring to Rosemount 1066 Transmitter
Figure 3-9: Rosemount 398R-54-61 Wiring to Rosemount 5081
20Wiring the Rosemount 398R/398RVP Sensor
Page 27
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
Figure 3-10: Rosemount 398R-xx-xx-54-62 Wiring to Rosemount 1056/56/1057 Transmitters
Figure 3-11: Rosemount 398R-xx-xx-54-62 Wiring to Rosemount 1066 Transmitter
Wiring the Rosemount 398R/398RVP Sensor21
Page 28
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 3-12: Rosemount 398R-xx-xx-54-62 Wiring to Rosemount 5081 Transmitter
Figure 3-13: Rosemount 398RVP Wiring to Rosemount 1056/56 Transmitters
22Wiring the Rosemount 398R/398RVP Sensor
Page 29
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
Figure 3-14: Rosemount 398RVP Wiring to Rosemount 1057 Transmitter
Figure 3-15: Rosemount 398RVP Wiring to Rosemount 1066 Transmitter
Wiring the Rosemount 398R/398RVP Sensor23
Page 30
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Figure 3-16: Rosemount 398RVP Wiring to Rosemount 5081 Transmitter
Figure 3-17: Rosemount 398RVP-70 Wiring to Rosemount 1056/56/1057 Transmitters
24Wiring the Rosemount 398R/398RVP Sensor
Page 31
Instruction ManualWiring the Rosemount 398R/398RVP Sensor
LIQ-MAN-398R-398RVPMarch 2017
Figure 3-18: Rosemount 398RVP-70 Wiring to Rosemount 1066 Transmitter
Figure 3-19: Rosemount 398RVP-70 Wiring to Rosemount 5081 Transmitter
Wiring the Rosemount 398R/398RVP Sensor25
Page 32
Wiring the Rosemount 398R/398RVP SensorInstruction Manual
March 2017LIQ-MAN-398R-398RVP
26Wiring the Rosemount 398R/398RVP Sensor
Page 33
Instruction ManualCalibration
LIQ-MAN-398R-398RVPMarch 2017
Section 5: Calibration
5.1General Information
.New sensors must be calibrated before use.
1
2.Regular recalibration is also necessary and is determined by the user.
3.The use of a two-point buffer calibration is always recommended.
4.Refer to the transmitter instruction manual for more specific calibration procedures.
5.2Use of Calibration Buffers (Standards)
1.Good buffers lead to good calibrations: A pH measurement is only as good as the
calibration, and the calibration is only as good as the buffers used. A careful buffer
calibration is the first step in making an accurate pH measurement.
2.Use appropriate buffers: Calibrate with buffers having pH values that bracket the pH of the
process. For example, if the pH is between 8 and 9, calibrate with pH 7 and 10 buffers.
Commercial buffers for intermediate range pH are readily available.
3.Sensor and buffers must be at the same temperature: Allow time for the sensor and buffers
to reach the same temperature. If the sensor was just removed from a process having a
temperature more than 10°C different from the buffer, allow at least 20 minutes.
4.Buffers must be at process temperature: For best results, calibrate with buffers having the
same temperature as the process. If the buffer and process temperature differ by more
than about 15°C an error as great as 0.1pH may result.
5.Be careful using buffers at very high temperatures: Protect the solution from evaporation.
Evaporation changes the concentration of the buffer and its pH. Be sure the pH of the buffer
is defined at high temperatures. The pH of many buffers is undefined above 60°C. Finally,
no matter what the temperature is, allow the entire measurement cell, sensor and
solution, to reach constant temperature before calibrating.
6.The pH of a buffer changes with temperature: The pH of a buffer is a function of
temperature. The pH of alkaline buffers depends more strongly on temperature than the
pH of acidic or neutral buffers. Most process pH instruments, including those
manufactured by Rosemount, have an auto calibration feature. The instrument recognizes
the buffer being used and automatically corrects for the change in buffer pH with
temperature. If the instrument does not perform the correction, the user must enter the
appropriate value. Buffer manufacturers usually list the temperature dependence of the
buffer on the label.
7.Buffers have limited shelf lives: Do not use a buffer if the expiration date has passed. Store
buffers at controlled room temperature.
8.Do not reuse buffers: Do not return used buffer to the stock bottle. Discard it.
9.Protect buffers from excessive exposure to air: Atmospheric carbon dioxide lowers the pH
of alkaline buffers. Other trace gases commonly found in industrial environments, for
example, ammonia and hydrogen chloride, also affect the pH of buffers. Molds, from
airborne spores, grow readily in neutral and slightly acidic buffers. Mold growth can
substantially alter the pH of a buffer.
Calibration27
Page 34
CalibrationInstruction Manual
March 2017LIQ-MAN-398R-398RVP
5.3Two Point Buffer Calibration
1.Remove the protective vinyl cap from the sensor tip.
2.Rinse the sensor and immerse it in the first buffer. Ensure that the glass bulb and the
temperature element are completely submerged by keeping the sensor tip about 3 inches
below the liquid level. Swirl the sensor to dislodge trapped air bubbles. Do not allow the
weight of the sensor to rest on the glass bulb.
3.Once the reading is stable, enter the buffer value in the analyzer. If the analyzer does not
correct for changes in buffer pH with temperature, be sure to enter the temperaturecorrected value.
4.Remove the sensor from the first buffer. Rinse the sensor and place it in the second buffer.
Follow the same precautions given in step 2.
5.Once the reading is stable, enter the buffer value in the analyzer. If the analyzer does not
correct for changes in buffer pH with temperature, be sure to enter the temperaturecorrected value.
6.After calibration, note the sensor slope. Slope has units of mV per unit change in pH. An
ideal sensor has a slope of 59 mV/pH at 25°C. Slope decreases as the sensor ages. Once
the slope drops to between 47 and 49 mV/pH, the sensor should be replaced.
7.Remove the sensor from buffer 2 and return it to the process liquid.
5.4pH Standardization
Standardization means making the process instrument match the reading from a second pH meter.
The second pH reading is usually made on a grab sample.
1.Take the sample from a point as close as possible to the process sensor. To avoid starving
the process sensor, use a downstream sample point.
2.Wait until the process pH is constant or, at worst, slowly drifting before taking the grab
sample.
3.To ensure that measured pH is truly the pH of the process liquid, determine the pH of the
grab sample immediately. pH is a function of temperature. If the temperature of the
process differs from ambient, measure the pH of the grab sample before its temperature
changes. Some process liquids are poorly buffered. The pH of the sample may change
significantly upon exposure to air or to the sample container. To avoid deterioration of the
sample, measure the pH immediately.
4.Following the instructions in the instrument manual, adjust the process reading to the
value measured on the grab sample.
28Calibration
Page 35
Instruction ManualCalibration
LIQ-MAN-398R-398RVPMarch 2017
5.5ORP Standardization
There are relatively few ORP calibration standards available. The most popular one is a solution
containing 0.1 M iron (II) and 0.1 M iron (III) in 1 M sulfuric acid. The standard is available from
osemount as PN R508- 16OZ. The poten-tial of the solution measured against a silver-silver
R
chloride reference electrode is 475 ±20 mV at 25 °C.
1.Rinse the sensor with deionized water and place it in the ORP standard along with a reliable
thermometer. Submerge the sensor tip at least three inches below the surface of the liquid.
Swirl the sensor to dislodge trapped bubbles. Adjust the temperature of the standard
to 25 ±5 °C.
2.Wait until temperature and ORP readings are stable.
3.Following the instructions in the instrument analyzer, store ORP value (475 mV) in memory.
4.Remove the sensor from the ORP standard, rinse it, and return it to the process fluid.
Calibration29
Page 36
CalibrationInstruction Manual
March 2017LIQ-MAN-398R-398RVP
30Calibration
Page 37
Instruction ManualMaintenance
LIQ-MAN-398R-398RVPMarch 2017
Section 6: Maintenance
6.1Maintenance
he frequency at which a sensor should be inspected, cleaned, and calibrated can be determined
T
only by experience. Generally, the greater the tendency of the process liquid to coat or foul the
sensor, the more frequently maintenance should be done. Rosemount 398R and 398RVP sensors
are fouling resistant and, they usually require maintenance less often than other pH (or ORP)
sensors. Sensors exposed to extreme pH values or to high temperature require more frequent
inspection than sensors installed in less severe environments. The best way to evaluate a sensor is
to check its performance in buffers. If the sensor cannot be calibrated or has low slope, it is dirty or
has failed. Refer to the troubleshooting guide in this manual for assistance.
6.2Sensor Removal
Please refer to the appropriate paragraph for instructions regarding removal of the sensor for
periodic maintenance.
6.2.1Retractable Version
WARNING
System pressure may cause the sensor to blow out with great force unless care is taken during removal.
Make sure the following steps are adhered to.
A.Rosemount 398R-21 and 398RVP-21 (21” tube)
1.Be certain system pressure at the sensor is below 64 psig before proceeding with
the retraction. It is also recommended that the personnel wear a face shield and
have a stable footing. Refer to Figure 6-1. Push in on the cable end or the top of the
junction box and slowly loosen the hex nut (B) of the process end male
connector (A).
Figure 6-1: Example of Sensor Tube Replacement
Maintenance31
Page 38
MaintenanceInstruction Manual
March 2017LIQ-MAN-398R-398RVP
B.Rosemount 398R-25 and 398RVP-25 (36” tube)
2.Be certain that pressure at the sensor is below 35 psig before proceeding with the
retraction. It is also recommended that the personnel wear a face shield and have
a stable footing. Refer to Figure 6-1. Push in on the cable end or the top of the
unction box and slowly loosen the hex nut (B) of the process end male
j
connector (A).
CAUTION
Do not remove nut at this time.
3.When the hex nut is loose enough, slowly ease the sensor back completely until
the retraction stop collar (mechanical safety stop) is reached.
CAUTION
Failure to withdraw the sensor completely may result in damage to the sensor when the valve is closed.
4.Close the ball valve slowly. If there is resistance, the valve may be hitting the sensor.
Double check that the sensor has been retracted to the retraction stop collar.
WARNING
Before removing the sensor from the ball valve, be absolutely certain that the ball valve is fully closed.
Leakage from the male connector threads may indicate that the male connector is still under pressure.
Leakage through a partially open valve could be hazardous, however with the ball valve closed, some
residual process fluid may leak from the connector's pipe threads.
5.The Male Connector Body (A) may now be completely unthreaded from the
reducing coupling and the sensor removed for servicing.
CAUTION
If the male connector leaks during insertion or retraction, replace the O-ring (PN 9550167) in the male
connector A.
6.3Cleaning Procedures - pH Sensors
To remove the sensor from the process piping, follow the instructions in Section 6.2.
CAUTION
Only persons thoroughly familiar with the procedure for diluting concentrated hydrochloric acid should
prepare the solution. Dilute the acid in a fume hood or in a well-ventilated area. Point the acid bottle
away from people when opening it. Wear appropriate safety equipment, including chemical goggles
and gloves. Do not let acid touch the skin or clothing. If acid solutions contact the skin or eyes, rinse
thoroughly with water. Seek medical assistance.
32Maintenance
Page 39
Instruction ManualMaintenance
LIQ-MAN-398R-398RVPMarch 2017
ProblemCleaning Suggestions
Loose scale or debrisUse a stream of water from a wash bottle to rinse away solids from the tip of the
sensor. If water does not work, gently wipe the glass bulb and liquid junction with
a soft cloth, tissue, cotton-tipped swab, or a soft bristle brush.
Oil and greaseWash the glass bulb with mild detergent solution and rinse thoroughly with
water.
Hard scale (carbonate and sulfate scales and
corrosion products)
When using acid or alkaline solvents, be careful to keep the solvent away from the liquid junction.
See Figure 6-2. If the cleaning solvent contacts the junction, hydrogen ions (acid solvent) or
hydroxide ions (alkaline solvent) will diffuse into the junction. Because hydrogen and hydroxide
ions have much greater mobility than other ions, they produce a large junction potential.
When the electrode goes back in service, the hydrogen or hydroxide ions slowly diffuse out of the
junction, causing the liquid junction potential and the pH reading to drift. It may take hours or days
for the reading to stabilize.
Always recalibrate the sensor after cleaning. If the sensor was cleaned with detergent or acid, soak
the sensor in pH 4 or pH 7 buffer for at least an hour before calibrating.
Figure 6-2: Tip of Rosemount 398R pH Sensor
If wiping the sensor tip with a tissue or cotton swab does not remove the scale,
soak the the glass bulb ONLY in a solution of 5% hydrochloric acid. To prepare the
acid solution, add 15 mL of concentrated hydrochloric acid to 85 mL of water with
continuous stirring. See CAUTION below. Keep the acid away from the liquid junction. Rinse the sensor thoroughly with deionized water. Some scales (for
example, calcium sulfate) cannot be removed easily with acid. Soaking the glass
bulb in a 2% solution of disodium EDTA for 20 minutes may be helpful.
The figure shows the tip of the Rosemount 398R pH sensor. The bottom of the liquid junction is
about even with the top of the wings that form the slotted tip. Keep acidic and alkaline solvents
away from the liquid junction. If acids or bases get into the junction, subsequent pH readings may
drift for several hours.
Maintenance33
Page 40
MaintenanceInstruction Manual
March 2017LIQ-MAN-398R-398RVP
6.4Cleaning Procedures - ORP Sensors
Clean platinum ORP electrodes by using a tissue to rub the metal surface with a paste of baking
soda (sodium bicarbonate). A clean platinum electrode is bright and shiny.
6.5Checking the Reference Electrode
Some processes contain substances, for example, sulfides, that poison the reference electrode.
Poisoning alters the electrode potential. For example, sulfide poisoning converts the reference
electrode from a silver/silver chloride electrode into a silver/silver sulfide electrode, causing a shift
in potential of several hundred millivolts.
A good way to check for poisoning is to compare the voltage of the reference electrode with a
silver/silver chloride electrode that is known to be good. The reference electrode from a new sensor
is the best choice. To check the suspect electrode, place both sensors in a beaker containing buffer
or a solution of potassium chloride. Connect the reference leads to a voltmeter and measure the
potential difference. If the suspect electrode is good, the difference should be no more than
about 20 mV.
Refer to Figure 6-3. A poisoned reference electrode usually requires replacement. A laboratory
silver/silver chloride reference electrode can be used in place of the second sensor. All Rosemount
pH sensors have a silver/silver chloride reference, and most sensors use gelled saturated potassium
chloride for the fill. The potentials of a good sensor reference electrode and a saturated silver/ silver
chloride laboratory electrode will agree within about 20 mV.
Figure 6-3: Checking the Potential of the Reference Electrode
34Maintenance
Page 41
Instruction ManualMaintenance
LIQ-MAN-398R-398RVPMarch 2017
6.6Rejuvenating Reference Electrodes
Occasionally, a poisoned or plugged reference electrode can be reconditioned. Although the
electrode seldom recovers completely, the procedure might extend the life of the sensor by a few
eeks.
w
a.Clean the sensor as thoroughly as possible.
b.Soak the sensor for several hours in a hot (NOT BOILING) 3% potassium chloride solution.
Prepare the solution by dissolving 3 g of potassium chloride in 100 mL of water.
c.Soak the sensor in pH 4 buffer at room temperature overnight.
d.Calibrate the sensor in buffers and retest it in the process liquid.
Table 6-1: Resistance as a Function of Temperature for Selected RTDs.
Temperature °CResistance (Ohms) ±1%
3KPT-100
02670100.0
102802103.8
202934107.7
253000109.6
303066111.5
403198115.4
503330119.2
603462123.1
703594126.9
803726130.8
903858134.6
1003990138.5
6.7Temperature Element
Rosemount 398R/398RVP pH sensors produce a voltage which the instrument converts to pH using
a temperature-dependent factor. A Pt 100 or a Balco 3K RTD built into the sensor measures
temperature. To permit cor-rection for changes in lead resistance with temperature, a three-wire
configuration (Figure 6-4) is used. To check the RTD, disconnect the leads and measure the
resistances shown. The measured resistance should agree with the value in the table to within about
±1%. If the measured resistance is appreciably dif-ferent (between 1 and 5%), the discrepancy can
be calibrated out. See the instrument instruction manual.
Figure 6-4: Three-wire RTD circuit
Maintenance35
Page 42
MaintenanceInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Consult Table 6.1 (above) for temperature-resistance data. Lead resistance is about 0.05 ohm/ft at
5°C. Therefore, 15 ft of cable increases the resistance by about 1.5 ohm. The resistance between
2
the RTD return and RTD sense leads should be less than 2 ohms. If a connection is open or shorted
and should not be, replace the sensor. If the measured resistance is greatly in error, replace the
sensor.
6.8Sensor Tube Replacement When Used With A
Sensor Head Junction Box
Replace ment of the retraction versions sensor tube assembly involves the removal and installation
of two sets of male connectors: One at the process end of the sensor, and the other at the junction
box end (See Figure 6-1 and Figure 6-5). Refer to Section 6.2 for proper removal of the sensor from
process.
1.Remove sensor from process before proceeding. The junction box with attached male
connector must be recovered from the old sensor for reuse. Unscrew the junction box cover
and set aside. Disconnect electrical connections from printed circuit board inside junction
box. Disconnect BNC connector to preamp. Unscrew hex nut (F) from male connector body
(E). Separate junction box from used sensor. Set aside.
2.Pry off split ferrule from sensor and set aside for reuse. Remove hex nut (F) and set aside for
reuse. Check that the internal O-ring is in place in the male connector body (E) attached to
the junction box.
3.Remove hex nut (B) from male connector body (A) at process end of sensor and set aside.
Slide the Teflon ferrule and the male connector off sensor in the direction of junction box
and set.
Note: If stainless steel ferrule was used, male connector body (A) will have to be dis carded with
the sensor tube.
4.Discard used O-ring from male connector body (A). Coat new O-ring with a thin film of the
O-ring lubricant provided. Position it in the machined O-ring groove in place of the
discarded O-ring.
CAUTION
Make sure lubricant does not contact any part of the sensor tip particularly the glass bulb.
5.Cover the 1" MNPT pipe threads of the male connector body (A) with Teflon tape (not
provided) to protect them from galling during reinstallation.
6.Pass the wires from the new sensor through the process end male connector (A). Make
sure that the beveled edge of the ferrule faces the process end of the sensor. Snug the hex
nut (B) to keep it in place. Do not tighten down fully on the hex nut at this time.
7.Pass the wires from the new sensor through the hex nut (F), the split ferrule (from the old
sensor), male connector body (E), O-ring, and through the junction box from the “neck”
opening and out to the printed circuit board in the junction box. Butt the ferrule’s beveled
edge and the sensor tube against the junction male connector (E).
8.Screw the hex nut (F) by hand until the tube is “locked” into the male connector body.
36Maintenance
Page 43
Instruction ManualMaintenance
LIQ-MAN-398R-398RVPMarch 2017
9.Make sure that the male connector body (E) is sufficiently tightened.
10.The sensor will “click” into place by pulling the sensor tube away from the junction box,
ut will not move from side to side or pull clear of the male connector.
b
11.If the sensor tube is correctly attached to the junction box, wrench tighten hex nut (F) on
male connector body (E). See Figure 6-1.
12.Do not put the sensor tube in a vise or use a pipe wrench to tighten the hardware as these
will damage the sensor. If sensor tube is not correctly attached to the junction box, loosen
hex nut (F) and repeat.
13.Connect the sensor wires to the terminals on the printed circuit board in the junction box
in the manner recommended on the junction box cover,and reattach the BNC connector
to the preamp.
14.Screw on the cover of the junction box aside. Discard sensor tube.
15.Insert the sensor in the process fitting. Stop it against the closed ball valve. Slide the
process-end male connector down the sensor tube to mate with the process fitting.
Tighten the male connector into the process fitting.
16.Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that
the sensor cannot come free from the valve assembly and male connector. The built-in
retraction stop collar at the end of the sensor will butt against the shoulder of the male
connector.
17.Open ball valve and position the sensor at the desired insertion depth and orientation.
Using a crescent or open end wrench, tighten the hex nut (B) to secure the sensor in place.
See Figure 6-5.
Note: A stainless steel ferrule is available if the Teflon ferrule does not adequately grip. Be careful
and avoid over tightening. This can damage the sensor tube.
Figure 6-5: Male Connector Tightening Diagram
CAUTION
If the male connector leaks during insertion or retraction, replace the O-ring (PN 9550167) in the male
connector A.
If the sensor is to be stored, the rubber boot should be filled with 7pH buffer solution and replaced
on sensor tip until ready to use.
Maintenance37
Page 44
MaintenanceInstruction Manual
March 2017LIQ-MAN-398R-398RVP
38Maintenance
Page 45
Instruction ManualTroubleshooting
LIQ-MAN-398R-398RVPMarch 2017
Section 7: Troubleshooting
7.1Troubleshooting
Table 7-1, below, lists common problems, causes and remedies typically encountered in process
measurement. For more specific troubleshooting information, please refer to the appropriate
transmitter manual.
Table 7-1: Troubleshooting
Note: For any repair or warranty inquiries please contact our Customer Care group.
Troubleshooting39
Page 46
TroubleshootingInstruction Manual
March 2017LIQ-MAN-398R-398RVP
Table 7-1: Troubleshooting (continued)
40Troubleshooting
Page 47
Instruction ManualTroubleshooting
LIQ-MAN-398R-398RVPMarch 2017
Table 7-1: Troubleshooting (continued)
Troubleshooting41
Page 48
42Heading title
Page 49
Instruction ManualAccessories
LIQ-MAN-398R-398RVPMarch 2017
Section 8: Accessories
8.1Accessories
Table 8-1: Accessories for Rosemount 398R/398RVP
Part NumberDescription
23557-00Preamplifier for remote junction box (PN 23555-00),
23550-00Remote Junction box with extension board
9550099O-ring, Viton, for process connector
9550220O-ring, Kalrez, for process connector
9210012Buffer solution, 4.01 pH, 16 oz
9210013Buffer solution, 6.86 pH, 16 oz
9210014Buffer solution, 9.18 pH, 16 oz
Accessories43
Page 50
AccessoriesInstruction Manual
March 2017LIQ-MAN-398R-398RVP
44Accessories
Page 51
Instruction ManualEC Declaration of Conformity
LIQ-MAN-398R-398RVPMarch 2017
EC Declaration of Conformity
Note: Please see website for most recent Declaration.
EC Declaration of Conformity45
Page 52
EC Declaration of ConformityInstruction Manual
March 2017LIQ-MAN-398R-398RVP
46EC Declaration of Conformity
Page 53
ANY FM APPROVED
ASSOCIATED APPARATUS
HAVING ENTITY PARAMETERS
NO
N-
H
AZARDOUS
(UNCLASSIFIED)
AREA
CLASS I, II, III, DIVISION 1, GROUPS A-G
T6 Ta = 60°C
SENSOR
ENTITY PARAMETERS
Ui = 13.1U, Ii = 358mA
Pi = 698 mW
Ci = 0.967 µF, Li = 0.1mH
1. NO REVISION TO THIS DRAWING IS PERMITTED WI THOUT
FM APPROVAL.
2. U
max
> U
t
; I
max
> I
t
; (C
i
OF ALL LOOPS + C CABLE) < C
a
;
(L
i
OF ALL LOOPS + L CABLE) < L
a
, P
max
OR P
i
> P
0.
3. SINGLE MULTI-CHANNEL
IS BARRIER OR APPARATUS MUST BE FM APPROVED,
4. SINGLE MULTI-CHANNEL
IS BARRIER OR APPARATUS MANUFACTURE'S
CONTROL DRAWI NGS MUST BE FOLLOWED WHEN INSTALLING THE SYSTEM.
IS BARRIER OR EQUIPMENT MAY BE INSTALLED WITHIN THE HAZARDOUS
(CLASSIFIED) LOCATION FOR WHICH IT IS APPROVED
.
5. INSTALLATION MUST BE IN ACCORDANCE WITH ARTICL
E 500 OF THE NEC
(ANSI/NFPA 70) AND ANSI/ISA RP 12.6.
WARNING: SUBSTITUTION OF COMPONENTS MAY IMPA
IR INTRINSIC SAFETY.
6. pH & AM PEROMETRIC SENSORS WITHOUT PREAMPS ARE SIMPLE APPARATUS.
ZON E
ZONE 0
7. CONTROL EQUIPMENT CONNECTED TO THE ASSOCIATED APPARATUS MUST
NOT USE OR GENERATE MORE THAN 250V.
RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND
MUST BE LESS THAN OR EQUAL TO 1 OHM.
ANY
FM APPROVED TRANSMI TTER FOR
DIVISION 1 WITH INTRINSICALLY SAFE
OUTPUT PARAMETERS. THIS
FM
APPROVED DEVICE MUST BE INSTALLED
PER ITS INSTALLATION DRAWING.
FM
APPROVED EQUIPMENT (MAY BE
MULTIPLE DEVICES, NUMBER IS LIMITED
BY REQUIREMENTS TO MEET ALL OTHER
IS REQUIREMENTS FOR THE NETWORK)
WITH EQUIVALENT HAZARDOUS AREA
APPROVALS.
H
AZA RDOUS
(CL ASSI
F
IED)
AREA
INTRISICALLY SA FE
9. pH/ORP SENSOR MODELS THAT MAY CONTAI N THE PREAMPLIFIER:
3900/VP
3500/VP
3300HT/VP
3400HT/VP
396/VP
396R/VP
396P/VP
398R/VP
399/VP
389/VP
385/385+
10. WARNING: TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE
ATMOSPHERES, DISCONNECT POWER BEFORE SERVIC
ING.
11. THE ENTITY CONCEPT ALLOWS INTECONNECTION OF INTRINSICALLY SAFE
APPARATUS WITH ASSOCIATED APPARATUS WHEN THE FOLLOWING IS TRUE:
U
i
> U
o;
I
i
> I
o;
P
i
> P
o;
C
o
> C
i
+ C CABLE; L
o
> L
i
+ L CABLE.
12. COPY REVISIONS TO 1400332 TO pH/ORP SHIPPING MANUALS.
13 Ci INCLUDES THE CAPACITANCE OF 500 FEET OF SENSOR CABLE.
13
DWG NO.
1400332
8.
Instruction ManualFM Installation
LIQ-MAN-398R-398RVPMarch 2017
Intrisicallly Safe Sensor Installation Drawing - FM
FM Installation47
Page 54
LIQ-MAN-398R-398RVP
Rev. H
March 2017
www.Emerson.com/RosemountLiquidAnalysis
Youtube.com/user/Rosemount
Twitter.com/Rosemount_News
Emerson
8200 Market Blvd.
Chanhassen, MN 55317,
USA
Tel +1 800 999 9307
Fax +1 952 949 7001
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of
one of the Emerson family of companies. All other marks are the property of their respective
owners.
The contents of this publication are presented for information purposes only, and while effort has
been made to ensure their accuracy, they are not to be construed as warranties or guarantees,
express or implied, regarding the products or services described herein or their use or applicability.
All sales are governed by our terms and conditions, which are available on request. We reserve the
right to modify or improve the designs or specifications of our products at any time without notice.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.