A high performance sensor with ease of maintenance in mind
Rosemount 397 pH/ORP sensors feature a TUpH reference design and are designed to be used with the Quik-Loc kit allowing for easy
installation and maintenance. These low maintenance sensors are ideal for measuring pH or ORP in aqueous solutions with high levels
of suspended solids.
Page 2
Rosemount 397 TUpH pH/ORP Sensor
Overview
June 2017
A TUpH Sensor Design
Excellent coating resistance and extended sensor life in
process solutions with high levels of suspended solids.
Highly chemical resistance with polypropylene and Tefzel
sensor body options.
Operates in process temperatures up to 100 °C and in
pressures up to 100 psig.
Helical reference pathway slows down the effects from
poisoning ions.
Ease of Maintenance and Installation
Quik-loc kit allows for quick and easy access to Rosemount
397 sensor without cable twisting.
Visual confirmation that the sensor is locked into position.
The Rosemount 397 sensor is housed in either a polypropylene or Tefzel body and is
designed to be used with the Quik-Loc kit (sold separately). Sensors come standard
with 15 ft. of integral cable and do not come with an integral preamplifier.
Table 1. Rosemount 397 pH/ORP Sensor ordering information
ModelSensor type
397pH/ORP sensor
Preamplifier/Cable
02Without Integral Preamplifier and 15 ft. Integral Cable
Measuring Electrode
10pH - GPHT Hemi Glass Bulb
Rosemount 397 TUpH pH/ORP Sensor
12ORP
Temperature Element
503k Balco
54Pt-100
(1)
(2)
Temperature Element
_No Selection - Polypropylene Body and Cable with BNC
62Polypropylene Body and Cable without BNC
64Tefzel Body Material
(4)
(3)
Typical Model Number: 397-02-10-54-64
1.For use with legacy transmitter model Rosemount 1181.
2.If selected with option -62, compatible with existing transmitter models Rosemount 1056, 1057, 56, 1066, and 5081. Without option -62, compatible with legacy
transmitter models 1054, 2054, 2081, and XMT.
3.Only available with option -54.
4.Only available with options -02-10-54.
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount 397 TUpH pH/ORP Sensor
Specifications
Table 2. Rosemount 397 sensor specifications
Sensor type
TUpH Model 397 pH Sensor
pH range
ACCUGLASS™, GPHT: 0-14 pH
Percent Linearity Over pH Range
0-2 pH94%
2-12 pH99%
12-13 pH97%
13-14 pH92%
Temperature range
32° to 212 °F (0° to 100 °C)
June 2017
Pressure range
0-100 psig (100-790 kPa abs)
Minimum conductivity
75
μS/cm, nominal 100 μS/cm
Wetted materials
Glass, polypropylene, EP, Tefzel (optional)
Process connections
None, must use Quik-Loc Kit
Interconnecting cables
15 ft (4.6 m), for use with remote preamplifier
Weight/shipping weight
1 lb/2 lb (0.45 kg/0.9 kg)
4
www.Emerson.com/RosemountLiquidAnalysis
Page 5
June 2017
Rosemount 397 TUpH pH/ORP Sensor
Quik-Loc Features and Applications
The Quik-Loc Kit consists of an adapter and a coupler. The PEEK adapter enables the Rosemount 397 pH sensor to fit into a 1 in. MNPT
coupler for quick disconnection without sensor cable twisting. The 316 stainless steel Dixon coupler is sealed with an EP gasket and
features locking arms. The B100 mechanism keeps the coupler arms locked in place until removal of the sensor is desired. The
Quik-Loc Kit is not recommended for use in processes with hazardous, cor- rosive, or strong oxidizing chemicals due to a risk of spray
or bodily injury.
Figure 1. Cross Section Diagram of the TUpH Reference Technology
Table 3. Quik-Loc Kit Specifications
Process connection
1-in. MNPT
Wetted materials
316 SST, EP, PEEK
Weight/Shipping weight
1 lb / 2 lb (0.45 kg / 0.9 kg)
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount 397 TUpH pH/ORP Sensor
June 2017
Product Certifications
Please see online certificates for further details.
IECEx
Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ + 60 °C)
Per standards IEC60079-0: 2011, IEC 60079-11: 2011
ATEX
– II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Per standards EN 60079-0: 2012 + A11:2013, EN 60079-11:2012
FM
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C Hazardous (Classified)
Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity; NI/I/2/ABCD/T6 Ta = 60 °C;
S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; Ambient
temperature rating -20 °C to +60 °C: (Simple Apparatus)
Per standards C22.2 No. 142 – M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50:11th Ed., UL
508:17th Ed., UL 913: 7th Ed., UL 60079-0: 2005, UL 60079-11: 2002
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
June 2017
Rosemount 397 TUpH pH/ORP Sensor
Dimensional and Installation Drawings
Figure 2. Rosemount 397 sensor and Quik-Loc Kit Components
Figure 3. Quik-Loc Mounting Dimension
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8
Rosemount 397 TUpH pH/ORP Sensor
Accessories
Table 4. Rosemount 397 Sensor accessories information
Part numberDescription
23555-00Junction box with preamplifier
2002565Mounting bracket kit for junction box
23646-01Extension cable for junction box, prepped
9200273Extension cable for junction box, unprepped
23757-00Quik-Loc Kit; for use with 1 inch tees; 1.4 inch insertion depth
23757-01Quik-Loc Kit; for use with 1-1/2 in. and 2 in. tees; 2 in. insertion depth
91604411 inch MNPT 316 Stainless Steel Coupler
23753-00PEEK Adapter for insertion depth of 2 inches
23753-01PEEK Adapter for insertion depth of 1.4 inches