Rosemount Analytical 397 TUpH Specifications

Page 1
Rosemount™ 397
TUpH pH/ORP Sensor
Product Data Sheet
June 2017
LIQ-PDS-397
A high performance sensor with ease of maintenance in mind
Page 2
Rosemount 397 TUpH pH/ORP Sensor

Overview

June 2017
A TUpH Sensor Design
Excellent coating resistance and extended sensor life in
process solutions with high levels of suspended solids.
Highly chemical resistance with polypropylene and Tefzel
sensor body options.
Operates in process temperatures up to 100 °C and in
pressures up to 100 psig.
Helical reference pathway slows down the effects from
poisoning ions.
Ease of Maintenance and Installation
Quik-loc kit allows for quick and easy access to Rosemount
397 sensor without cable twisting.
Visual confirmation that the sensor is locked into position.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Dimensional and Installation Drawings . . . . . . . . . . . . . . 7
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
www.Emerson.com/RosemountLiquidAnalysis
Page 3
June 2017

Ordering Information

The Rosemount 397 sensor is housed in either a polypropylene or Tefzel body and is designed to be used with the Quik-Loc kit (sold separately). Sensors come standard with 15 ft. of integral cable and do not come with an integral preamplifier.
Table 1. Rosemount 397 pH/ORP Sensor ordering information
Model Sensor type
397 pH/ORP sensor
Preamplifier/Cable
02 Without Integral Preamplifier and 15 ft. Integral Cable
Measuring Electrode
10 pH - GPHT Hemi Glass Bulb
Rosemount 397 TUpH pH/ORP Sensor
12 ORP
Temperature Element
50 3k Balco
54 Pt-100
(1)
(2)
Temperature Element
_ No Selection - Polypropylene Body and Cable with BNC
62 Polypropylene Body and Cable without BNC
64 Tefzel Body Material
(4)
(3)
Typical Model Number: 397-02-10-54-64
1. For use with legacy transmitter model Rosemount 1181.
2. If selected with option -62, compatible with existing transmitter models Rosemount 1056, 1057, 56, 1066, and 5081. Without option -62, compatible with legacy transmitter models 1054, 2054, 2081, and XMT.
3. Only available with option -54.
4. Only available with options -02-10-54.
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount 397 TUpH pH/ORP Sensor

Specifications

Table 2. Rosemount 397 sensor specifications
Sensor type
TUpH Model 397 pH Sensor
pH range
ACCUGLASS™, GPHT: 0-14 pH
Percent Linearity Over pH Range
0-2 pH 94%
2-12 pH 99%
12-13 pH 97%
13-14 pH 92%
Temperature range
32° to 212 °F (0° to 100 °C)
June 2017
Pressure range
0-100 psig (100-790 kPa abs)
Minimum conductivity
75
μS/cm, nominal 100 μS/cm
Wetted materials
Glass, polypropylene, EP, Tefzel (optional)
Process connections
None, must use Quik-Loc Kit
Interconnecting cables
15 ft (4.6 m), for use with remote preamplifier
Weight/shipping weight
1 lb/2 lb (0.45 kg/0.9 kg)
4
www.Emerson.com/RosemountLiquidAnalysis
Page 5
June 2017
Rosemount 397 TUpH pH/ORP Sensor

Quik-Loc Features and Applications

The Quik-Loc Kit consists of an adapter and a coupler. The PEEK adapter enables the Rosemount 397 pH sensor to fit into a 1 in. MNPT coupler for quick disconnection without sensor cable twisting. The 316 stainless steel Dixon coupler is sealed with an EP gasket and features locking arms. The B100 mechanism keeps the coupler arms locked in place until removal of the sensor is desired. The Quik-Loc Kit is not recommended for use in processes with hazardous, cor- rosive, or strong oxidizing chemicals due to a risk of spray or bodily injury.
Figure 1. Cross Section Diagram of the TUpH Reference Technology
Table 3. Quik-Loc Kit Specifications
Process connection
1-in. MNPT
Wetted materials
316 SST, EP, PEEK
Weight/Shipping weight
1 lb / 2 lb (0.45 kg / 0.9 kg)
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount 397 TUpH pH/ORP Sensor
June 2017

Product Certifications

Please see online certificates for further details.
IECEx
Ex ia IIC T4 Ga (-20 °C Ta + 60 °C)
Per standards IEC60079-0: 2011, IEC 60079-11: 2011
ATEX
II 1 G Ex ia IIC T4 Ga (-20 °C Ta +60 °C)
Per standards EN 60079-0: 2012 + A11:2013, EN 60079-11:2012
FM
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C Hazardous (Classified) Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity; NI/I/2/ABCD/T6 Ta = 60 °C; S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; Ambient temperature rating -20 °C to +60 °C: (Simple Apparatus)
Per standards C22.2 No. 142 – M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50:11th Ed., UL 508:17th Ed., UL 913: 7th Ed., UL 60079-0: 2005, UL 60079-11: 2002
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
June 2017
Rosemount 397 TUpH pH/ORP Sensor

Dimensional and Installation Drawings

Figure 2. Rosemount 397 sensor and Quik-Loc Kit Components
Figure 3. Quik-Loc Mounting Dimension
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8
Rosemount 397 TUpH pH/ORP Sensor

Accessories

Table 4. Rosemount 397 Sensor accessories information
Part number Description
23555-00 Junction box with preamplifier
2002565 Mounting bracket kit for junction box
23646-01 Extension cable for junction box, prepped
9200273 Extension cable for junction box, unprepped
23757-00 Quik-Loc Kit; for use with 1 inch tees; 1.4 inch insertion depth
23757-01 Quik-Loc Kit; for use with 1-1/2 in. and 2 in. tees; 2 in. insertion depth
9160441 1 inch MNPT 316 Stainless Steel Coupler
23753-00 PEEK Adapter for insertion depth of 2 inches
23753-01 PEEK Adapter for insertion depth of 1.4 inches
9160447 1 inch E.P. Gasket for Quick-Loc Coupler
June 2017
8
www.Emerson.com/RosemountLiquidAnalysis
Page 9
June 2017

Notes:

Rosemount 397 TUpH pH/ORP Sensor
www.Emerson.com/RosemountLiquidAnalysis
9
Page 10
Rosemount 397 TUpH pH/ORP Sensor
LIQ-PDS-397
Global Headquarters
Emerson Automation Solutions
8200 Market Blvd Chanhassen, MN 55317
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001 Liquid.CSC@Emerson.com
North America Regional Office
Emerson Automation Solutions
8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RMT-NA.RCCRFQ@Emerson.com
Latin America Regional Office
Emerson Automation Solutions
1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Product Data Sheet
June 2017
Europe Regional Office
Emerson Automation Solutions GmbH
Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions Asia Pacific Pte Ltd
1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions
Emerson FZE P.O. Box 17033, Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Analyticexpert.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
Google.com/+RosemountMeasurement
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount and Rosemount logotype are trademarks of Emerson. All other marks are the property of their respective owners. © 2017 Emerson. All rights reserved.
Loading...