Rosemount Analytical 396R, 396P, 396PVP, 396RVP Specifications

Product Data Sheet
Rosemount™ 396/396VP/396P/396PVP/ 396R/396RVP
TUpH pH/ORP Sensors
June 2017
LIQ-PDS-396
Rosemount TUpH pH/ORP sensors are robust sensors designed to meet a number of difficult application requirements. Sensors feature a variety of options suited to meet a number of installation requirements and are ideal for use in dirty, abrasive, coating, and high solid applications.
Rosemount TUpH 396, 396VP, 396P, 396PVP, 396R, 396RVP Sensors
Rosemount 396/396VP
Rosemount 396P/396PVP
Rosemount 396R/396RVP

Overview

Robust Sensor Design
TUpH reference technology minimizes sensor maintenance
and ensures a steady pH/ORP signal even when sensor is coated.
Maximize sensor life with poisoning resistant helical reference
pathway design.
Optional flat pH glass electrode for abrasive and coating
applications.
Maximum process pressure 150 psig.
Maximum process temperature 100 °C.
Versatile Mounting Options to Meet Your Installation Requirements
June 2017
Rosemount 396P/396PVP sensors feature 1 in. MNPT front
and rear facing process threads.
Rosemount 396R/396RVP offered in 21 in. and 36 in. lengths
suitable for use with retraction assemblies – easily remove sensors for maintenance and calibration without having to shutdown your process.
Variopol(VP8) cable connection option, for quick
cable-to-sensor release, eliminates cable twisting.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Dimensional and Installation Drawings . . . . . . . . . . . . . 10
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
www.Emerson.com/RosemountLiquidAnalysis
June 2017
Rosemount TUpH 396, 396VP, 396P, 396PVP, 396R, 396RVP Sensors

Ordering Information

Rosemount 396 and 396VPpH sensors feature a stainless steel body for use with optional 1 inch threaded process connectors suitable for insertion, submersion, or flow through installation. Sensors are available with either a standard pH hemi glass bulb or optional pH flat glass bulb. Rosemount 396 sensors feature a standard 15 ft. integral cable and Rosemount 396VP sensors feature a Variopol cable connection for use with VP8 quick disconnect interconnecting cables (sold separately; see Accessories Table 9).
Table 1. Rosemount 396 pH/ORP Sensor ordering information
Model Sensor type
396 pH Sensor
Temperature element
50 3k Balco
54 Pt-100
(1)
Cable options
_ 15ft. cable with BNC
62 15 ft. cable without BNC
(2)
Glass options
_ No selection –pH –GPHT hemi glass bulb
71 pH–GPHT flat glass bulb
Typical model number: 396-54-62
1. For use with legacy transmitter models 1181pH, 1050, and 1003.
2. For use with existing transmitter models Rosemount 56, 1056, 1057, 1066, and 5081. Also compatible with legacy transmitter models 6081, 81, 1181, 2081, 3081, 4081, 54, 1055, and XMT.
Table 2. Rosemount 396VP pH/ORP Sensor ordering information
Model Sensor type
396VP pH Sensor
Temperature element
50 3k Balco
54 Pt-100
55 Pt-100
(1)
(2)
(3)
Glass Options
_ No selection –pH –GPHT hemi glass bulb
71 pH –GPHT flat glass bulb
www.Emerson.com/RosemountLiquidAnalysis
3
Rosemount TUpH 396, 396VP, 396P, 396PVP, 396R, 396RVP Sensors
June 2017
Table 2. Rosemount 396VP pH/ORP Sensor ordering information
Preamplifier Option
_ No preamplifier
70 SMART preamplifier
(4)
Typical Model Number: 396-54-62
1. For use with legacy transmitter models 1181pH, 1050, and 1003.
2. For use with existing transmitter models Rosemount 56, 1056, 1057, 1066, and 5081. Also compatible with legacy transmitter models 6081, 81, 1181, 2081, 3081, 4081, 54, 1055, and XMT.
3. For use with existing transmitter models Rosemount 56, 1056, 1057, 1066, and 5081. Only available with option -70.
4. Only available with option -55.
Rosemount 396P and 396PVP pH/ORP sensors feature a molded reinforced polypropylene body with forward and rear facing 1 in. MPT threads suitable for insertion, submersion, and flow through type installations. The Rosemount 396P sensor has an integral cable connection and the Rosemount 396PVP sensor has a Variopol cable connection for use with VP8 interconnecting cables (sold separately; see Accessories Table 9).
Table 3. Rosemount 396P pH/ORP Sensor ordering information
Model Sensor type
396P pH/ORP Sensor
Preamplifier/cable
01 25ft. Integral cable with preamplifier
02 15 ft. Integral cable without preamplifier
(1)
Measuring electrode
10 pH –GPLR hemi Glass Bulb
12 Platinum ORP
13 pH –GPLR Flat Glass Bulb
Temperature element/Transmitter compatibility
50 3K Balco, for use with model 1181
54 Pt-100, for use with model 1054,2054, and 2081
55 Pt-100, for use with model 56, 1056, 1057,1066, 5081
Sensor tip option
_ No selection – Standard shrouded tip
41 Slotted tip
(3)
(2)
Typical model number: 396P-01-10-55
1. Preamplifier is SMART for pH electrodes select ed with option -55.
2. Also compatible with legacy transmitter models 54, 1055, 6081, 81, 3081, 4081, and XMT.
3. Not available with option -13.
4
www.Emerson.com/RosemountLiquidAnalysis
June 2017
Rosemount TUpH 396, 396VP, 396P, 396PVP, 396R, 396RVP Sensors
Table 4. Rosemount 396PVP pH/ORP Sensor ordering information
Model Sensor type
396PVP pH/ORP Sensor
Measuring electrode
10 pH–GPLR hemi glass bulb
12 Platinum ORP
13 pH –GPLR flat glass bulb
Temperature element/Transmitter compatibility
50 3K Balco, for use with model 1181
54 Pt-100, for use with model 1054,2054, and 2081
55 Pt-100, for use with model 56, 1056, 1057,1066, 5081
(1)
Sensor tip option
_ No selection –Standard shrouded tip
41 Slotted tip
(2)
Preamplifier option
_ No selection -No preamplifier
70 SMART preamplifier
(3)
Typical model number: 396PVP-10-55-41-70
1. Also compatible with legacy transmitter models 54, 1055, 6081, 81, 3081, 4081, and XMT.
2. Not available with option -13.
3. Only available for pH electrodes selected with option -55.
www.Emerson.com/RosemountLiquidAnalysis
5
Loading...
+ 11 hidden pages