Rosemount Analytical 3900 Operating Manual

Model 3900/3900VP
General Purpose pH/ORP Sensor
For additional information, please visit our website at
www.emersonprocess.com/raihome/liquid/.
CAUTION
The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water.
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
WARNING
The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the
responsi-
bility of the user.
CAUTION
SENSOR/PROCESS
APPLICATION COMPATIBILITY
SPECIFICATIONS
Sensor Type: General purpose 3900 Measured Range: pH: A
CCUGLASS 0-14
ORP: -1500 to +1500 mV
Percent Linearity Over pH Range:
Operating Temperature: -10° to 100°C (14° to 212°F)
Automatic temperature compensation -10° to 100°C (14° to 212°F)
Maximum Pressure: 790 kPa [abs] (100 psig) at 100°C
(212°F)
Materials of Construction: stainless steel, glass, Teflon,
polyphenylene sulfide (PPS), and EPDM
Materials of ORP: Platinum Process Connections
Front facing:
3
4˝ and 1˝ MNPT
Rear facing: 1˝ MNPT
Integral Cable: 32 ft (10m) cable with integral SMART pre-
amp; 15 ft (4.7m) cable without preamp sensor
VP8 Cable: use 24281-XX, 2.5 ft (.8m) to 100 ft (31m) Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)
Instruction Sheet
PN 51A-3900 September 2010/rev.A
MODEL 3900
Insertion/Submersion
Sensor
Range Linearity
0 - 7 97%
1 - 7 98%
4 - 7 98%
7 - 10 99%
7 - 12 97%
7 - 13 96%
7 - 14 95%
2
STORAGE
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below -10°C (14°F).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode has a limited shelf life of one year.
ELECTRODE PREPARATION
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
INSTALLATION
For sensor orientation, see Figure 1.
For sensor dimensions, see Figure 2.
For mounting options, see Figures 3-6.
For wiring, see Figures 7-22.
For additional wiring information, please visit our website at www.emersonprocess.com/raihome/liquid/wiring.
TWO POINT BUFFER CALIBRATION
Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).
NOTE
A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approximately ± 59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor tempera­ture) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/ transmitter.
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.1 mV/pH. Over time the sensor will age, both in the process and in storage, which will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.
RECOMMENDED pH SENSOR STANDARDIZATION
For maximum accuracy, the sensor can be standard­ized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
MAINTENANCE FOR pH ELECTRODE
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 1 to 5 minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% at 25°C (77°F), then it will be approximately 25% at 80°C (176°F), and approximately 10% at 100°C (212°F).
MODEL 3900 INSTALLATION
CAUTION
Hydrochloric acid is toxic and highly corrosive. Avoid skin contact, wear protective gloves. Use only in a well ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately.
3
ORP CALIBRATION
1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone PN R508-8OZ (475mV). This can also be made quite simply by adding a few crystals of quinhydrone to either pH 4 or pH 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required.
2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize.
3. Standardize the instrument to the solution value shown in the table, right. The resulting potentials, measured with a clean platinum electrode and saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not sta­ble over long periods of time. Therefore, these standards should be made up fresh each time they are used.
4. Remove the sensor from the buffer, rinse, and install in the process.
MAINTENANCE FOR ORP ELECTRODE
Electrodes should respond rapidly. Sluggishness, off­sets, and erratic readings are indicators that the elec­trodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a
mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 1 to
5 minutes in a 5% hydrochloric acid solution.
3. ORP (metallic) electrodes should be polished with
moistened baking soda.
pH 4 Solution pH 7 Solution
Temp °C 20 25 30 20 25 30
mV Potential 268 264 260 94 87 80
ORP of Saturated Quinhydrone Solution (millivolts)
MODEL 3900 INSTALLATION
FIGURE 1. Sensor Orientation
FIGURE 2. Sensor Dimensions
Model 3900
Model 3900VP
TEMP COMP
SOLN GND
REFERENCE
JUNCTION
pH ELECTRODE
3/4 IN MNPT
.92
1 IN MNPT
2.33
WRENCH FLATS
1.30 IN ACROSS
6.91
8.63
1 IN MNPT
4
psig (kPa [abs]) °F (°C)
150 (1136) 150 (65)
128 (984) 160 (71)
102 (805) 170 (77)
80 (653) 180 (82)
57 (494) 200 (93)
48 (432) 210 (99)
Horizontal Pipe Tee (PN 2002011)
Pressure/Temperature Ratings
MODEL 3900 INSTALLATION
FIGURE 3: Typical Flow Through insertion installation using PN 2002011 Pipe Tee
(sensor must be installed at least 10° above the horizon)
MILLIMETER
INCH
Inlet and outlet connections are stainless steel and take 1/4-inch OD tubing. Flow cell is polycarbonate with 1/4-inch FNPT fittings.
WETTED MATERIALS:
BODY - ACRYLIC
NUT - CPVC
FITTINGS - 316 SST
SEALS - BUNA N
FLOW CELL RATINGS
TEMPERATURE: 32˚-194˚F (0-90˚C) MAX. PRESSURE: 90 PSIG (721 kPa [abs]) FLOW RATE: 2-5 GPH (7.6-18.9 LPH)
FIGURE 4: Low Flow Cell PN 24091-00
5
MODEL 3900 INSTALLATION
Jet Spray Cleaner shown with
Model 3900VP pH sensor
The Jet Spray Cleaner eliminates routine, manual sensor maintenance by cleaning the sensor with water or compressed air. Flow through the cleaner can be controlled by a solenoid valve.
NOTE: The Jet Spray Cleaner can be used with Handrail Mounting Assembly (PN 11275-01, not shown) or can be mounted through conduit as shown above.
FIGURE 5: Jet Spray Cleaner PN 12707-00
FIGURE 6. Submersion Installation
Using the Handrail Mounting
Assembly
(PN 11275-01)
All parts shown are supplied;
sensor is sold separately.
MILLIMETER
INCH
Loading...
+ 11 hidden pages