Rosemount Analytical 3900 Specifications

Page 1
Rosemount™ 3900/3900VP
General Purpose pH/ORP Sensors
Product Data Sheet
June 2017
LIQ-PDS-3900
High performance for your process pH/ORP needs
Page 2
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Overview

A Robust Sensor Design
Extended sensor life and protection against poisoning ions
using a double junction reference.
Enhanced performance and increased life with cracking
resistant glass.
Maximum chemical resistance provided by a rugged
polyphenylene sulfide body.
Operates in sub-zero temperatures down to 14 °F (-10 °C).
Built-in solution ground for advanced diagnostics.
Versatile Installation Options
The sensor body features a one-piece construction with both
front facing and rear facing process threads.
June 2017
Contents
¾ in. and 1 in. MNPT process connections to meet a variety of
application installation requirements.
Variopol (VP8) cable connection option, for quick
cable-to-sensor release, eliminates cable twisting.
SMART Preamplifier
Automatic recognition of pH sensors by Rosemount
transmitters: 1066, 1057, 1056, and 56
pH calibration data is stored which allows sensors to be
calibrated in advance for “Plug and Play” installations in the field.
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Dimensional and Installation Drawings . . . . . . . . . . . . . . 6
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
www.Emerson.com/RosemountLiquidAnalysis
Page 3
June 2017
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Ordering Information

The Rosemount 3900 and 3900VP General Purpose pH/ORP Sensors feature a chemically resistant Ryton plastic body, along with a built-in solution ground for advanced diagnostics and a Pt-100 RTD for temperature compensation. These sensors are available with either an integral cable connection (3900) or Variopol (VP8) connector (3900VP). Variopol cables are sold separately (see
Table 1. Rosemount 3900 pH/ORP Sensor ordering information
Model Sensor Type
3900 pH/ORP sensor
Preamplifier Option
01 Preamplifier
02 No preamplifier
(1)
(2)
Accessories ).
Measuring Electrode
10 General Purpose Low Resistivity (GPLR) pH glass
12 Platinum ORP
Typical Model Number: 3900-01-10
1. Preamplifier is SMART with -10 option and a standard preamplifier if with -12 option. Comes with 32 ft. (10 m) of integral cable.
2. Comes standard with 15 ft. (4.7 m) of integral cable.
Table 2. Rosemount 3900VP pH/ORP Sensor with Variopol cable connection ordering information
Model Sensor Type
3900VP pH/ORP sensor
Preamplifier Option
01 Preamplifier
02 No preamplifier
(1)
Measuring Electrode
10 General Purpose Low Resistivity (GPLR) pH glass
12 Platinum ORP
Typical Model Number: 3900VP-01-10
1. Preamplifier is SMART with -10 option and a standard preamplifier if with -12 option.
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Specifications

Table 3. Percent linearity over pH range
June 2017
pH Range
Linearity
0-7 1-7 4-7 7-10 7-12 7-13 7-14
97% 98% 98% 99% 97% 96% 95%
Table 4. Rosemount 3900/3900VP pH/ORP Sensor specification
Measurement Range
pH 0 to 14
ORP -1500 to +1500 mV
Operating Temperature
-10 to 100 °C (14 to 212 °F), Automatic temperature compensation from -10 to 100 °C (14 to 212 °F)
Maximum Pressure
100 psig (790 kPa [abs]) at 100 °C (212 °F) CRN Rating: 60 psig up to 100 °C (212 °F)
Materials of Construction
Sensor Body Ryton - Polyphenylene Sulfide (PPS)
O-ring EPDM
pH Electrode Glass
ORP Electrode Glass, platinum
Solution Ground Stainless steel
Reference Junction PTFE (Teflon)
Conductivity
Responds to changes in pH at a minimum conductivity of 0.1 μS/cm when used with the low flow panel. The sample flow rate must be controlled to 2 GPH (7.6 L/hr)
Process Connection
Front Facing ¾ in. and 1 in. MNPT
Rear Facing 1 in. MNPT
Temperature Compensation
Pt-100
Cable
3900 32 ft (10m) cable with integral preamplifier; 15 ft (4.7m) cable without preamplifier
3900VP Use 24281-XX, 2.5 ft (.8m) to 100 ft (31m) (see
Accessories )
Weight/Shipping Weight
0.45 kg/0.9 kg (1 lb/2 lb)
4
www.Emerson.com/RosemountLiquidAnalysis
Page 5
June 2017
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Product Certifications

IECEx
3900/3900VP without preamp (pH and ORP) – Ex ia IIC T4 Ga (-20 °C Ta +60 °C)
3900/3900VP with SMART preamp (pH only) – Ex ia IIC T4 Ga (-20 °C Ta +60 °C)
Per standards IEC60079-0: 2011, IEC 60079-11: 2011
ATEX
3900/3900VP without preamp (pH and ORP) – II 1 G Ex ia IIC T4 Ga (-20 °C Ta +60 °C)
3900/3900VP with SMART preamp (pH only) – II 1 G Ex ia IIC T4 Ga (-20 °C Ta +60 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
FM
3900/3900VP with SMART preamp (pH only), with standard preamp (ORP only), and without preamp (pH and ORP):
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C Hazardous (Classified)
Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity; NI/I/2/ABCD/T6 Ta = 60 °C; S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2044, 3810:2005
CSA
3900/3900VP with SMART preamp (pH only) – Intrinsically Safe:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ambient temperature rating -20 °C to +60 °C; Ex ia IIC; T6
3900/3900VP without preamp (pH and ORP) – Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; Ambient temperature rating -20 °C to +60 °C: (Simple Apparatus)
Per standards C22.2 No. 142 – M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50, UL 508, UL 913, UL 60079-0: 2005, UL 60079-11: 2002
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Dimensional and Installation Drawings

Figure 1. Proper sensor installation orientation
Figure 2. Dimensional Drawing for the Rosemount 3900/3900VP Sensor
June 2017
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
June 2017
Rosemount 3900/3900VP General Purpose pH/ORP Sensors
Table 5. Horizontal pipe tee (PN 2002011) pressure/temperature ratings
psig (kPa [abs]) °F (°C)
150 (1136) 150 (65)
128 (984) 160 (71)
102 (805) 170 (77)
80 (653) 180 (82)
57 (494) 200 (93)
48 (432) 210 (99)
Figure 3. Typical flow through insertion installation using PN 2002011 pipe tee
Table 6. Low flow cell specifications PN 24091-00/24091-02
Wetted Materials
Body and nut Polyester/Polycarbonate
Fittings 316 SST
Seals Silicone
Flow Cell Ratings
Temperature 32 to 158 °F (0 to 70 °C)
Maximum pressure 90 psig (721 kPa [abs])
Flow Rate 2 to 5 GPH (7.6 to 18.9 LPH)
Sensor Threaded Connection
24091-00 1 in. NPT adapter
24091-02 ¾ in. NPT adapter
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8
Rosemount 3900/3900VP General Purpose pH/ORP Sensors
Figure 4. Low flow cell PN 24091-00/24091-02
June 2017
Figure 5. Flow-through tee (2 in.) (PN 915240-XX)
8
www.Emerson.com/RosemountLiquidAnalysis
Page 9
June 2017
The following flow-through tees are available for the Rosemount 3900/3900VP:
915240-03
915240-04
915240-05
For a new installation, use the appropriate tee above and add the S10283-LQD adapter.
When replacing an existing Rosemount 399 with a Rosemount 3900 sensor, add the S10283-LQD adapter to convert the tee to accept the Rosemount 3900 sensor.
Rosemount 3900/3900VP General Purpose pH/ORP Sensors
Figure 6. Jet spray cleaner PN 12707-00
The jet spray cleaner eliminates routine, manual sensor maintenance by cleaning the sensor with water or compressed air. Flow through the cleaner can be controlled by a solenoid valve.
Note: The jet spray cleaner can be used with handrail mounting assembly (PN 11275-01, not shown) or can be mounted through conduit as shown above.
www.Emerson.com/RosemountLiquidAnalysis
9
Page 10
Rosemount 3900/3900VP General Purpose pH/ORP Sensors
Figure 7. Low flow panel PN SQP10077-LQD
June 2017
Table 7. Low flow panel specifications
Inlet flow
Inlet pressure
Temperature
(*) The minimum inlet pressure is required to open a check valve, which prevents the flow cell from draining if sample flow is lost. Removing the check valve lo wers the inlet
pressure requirement to a few feet of water head.
3-80 gph (11.4-304 L/hr)
3 – 65 psig (122 – 549 kPa abs)*
32 - 122 °F (0 - 50 °C)
10
www.Emerson.com/RosemountLiquidAnalysis
Page 11
June 2017
Rosemount 3900/3900VP General Purpose pH/ORP Sensors

Accessories

Table 8. Rosemount 3900/3900VP pH/ORP Sensor accessories information
Part number Description
24281-00 15 ft. cable with mating VP8 connector
24281-01 25 ft. cable with mating VP8 connector
24281-02 2.5 ft. cable with mating VP8 connector
24281-03 50 ft. cable with mating VP8 connector
24281-04 100 ft. cable with mating VP8 connector
24281-05 4 ft. cable with mating VP8 connector
24281-06 10 ft. cable with mating VP8 connector
24281-07 20 ft. cable with mating VP8 connector
24281-08 30 ft. cable with mating VP8 connector
12707-00 Jet spray cleaner
23242-02
24091-00 Low flow cell, 1 in. NPT adapter
24091-02 Low flow cell, ¾ in. NPT adapter
23555-00 Weatherproof junction box, with preamplifier
2002011 CPVC In-line Tee, 1.5 in. size, 1.0 in. threaded process connection
11275-01 Handrail mounting assembly
9200273 11-Conductor extension cable, shielded and unprepped (for use with remote junction box)
915240-03 PVC flow through Tee, ¾ in. NPT process connection
915240-04 PVC flow through Tee, 1 in. NPT process connection
915240-05 PVC flow through Tee, 1-½ in. NPT process connection
9210012 Buffer solution, 4.01 pH, 16 oz
9210013 Buffer solution, 6.86 pH, 16 oz
9210014 Buffer solution, 9.18 pH, 16 oz
SQP10077-LQD Low flow panel
Insertion mounting adapter, 1.5 in. MNPT process connection, 1 in. x ¾ in. FPT sensor adapter/union thread size
www.Emerson.com/RosemountLiquidAnalysis
11
Page 12
Rosemount 3900/3900VP General Purpose pH/ORP Sensors
LIQ-PDS-3900
Global Headquarters
Emerson Automation Solutions
8200 Market Blvd Chanhassen, MN 55317
+1 800 999 9307 or +1 952 906 8888
+1 952 949 7001 Liquid.CSC@Emerson.com
North America Regional Office
Emerson Automation Solutions
8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RMT-NA.RCCRFQ@Emerson.com
Latin America Regional Office
Emerson Automation Solutions
1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Product Data Sheet
June 2017
Europe Regional Office
Emerson Automation Solutions GmbH
Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions Asia Pacific Pte Ltd
1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions
Emerson FZE P.O. Box 17033, Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Analyticexpert.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
Google.com/+RosemountMeasurement
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount and Rosemount logotype are trademarks of Emerson. All other marks are the property of their respective owners. © 2017 Emerson. All rights reserved.