Rosemount Analytical 3900 Installation Guide

Page 1
Installation & Calibration Equipment
Sensor & Instrument Mounting, Accessories, Systems and More
Visualize > Analyze > Optimize
Page 2
INSTALLATION & CALIBRATION EQUIPMENT
Emerson Process Management offers a variety of Rosemount Analytical accessories to fit many requirements. This brochure includes mounting tees, ball valve kits, and other configurations used for mounting conductivity, pH, ORP and amperometric (dissolved oxygen, chlorine, and ozone) sensors and instruments. Please note that each mounting accessory is only valid for the stated sensor(s). Also included is information on instrument mounting, cables, junction boxes and various chemicals. The accessories in this brochure can be used in a wide variety of applications commonly found in the following industries:
TABLE OF CONTENTS
Mounting Recommendations for Sensors includes best practices for in-line and tank mounting............................4-5
> Chemical
> Food and Beverage
> Water and Wastewater Treatment
> Power
> Water Treatment
> Utilities
> Pulp and Paper
> Metals and Mining
> Hydrocarbon Processing
> Textile
> Pharmaceutical
Insertion Sensor Mounting including tees, low flow cells, handrail mounting, floating balls,
jet spray cleaner, mounting adapters, and process connectors.............................................................................6-18
Retraction Assemblies including ball valve kits, retraction assemblies
for higher pressures, and a rotary retraction unit ...............................................................................................19-27
Instrument Mounting with mounting brackets for wall, pipe, or panel mount....................................................28-30
Wiring Accessories includes best practices for cable lengths. Also includes junction boxes,
extension cables, VP connector cables and cable glands....................................................................................31-37
Calibration Accessories with pH buffer solutions and conductivity standard solutions Also included is a conductivity validation unit, a pH benchtop calibration unit,
a pH simulator, and VP plug simulator for pH and conductivity ..........................................................................38-40
Flow Accessories including rotameters and a constant head flow controller.......................................................41-42
Tags for Sensors and Instruments............................................................................................................................43
Sample Cooler.........................................................................................................................................................43
Special Parts and Assemblies including examples of customer requested panels with sensors,
flow cells, instruments, and sample cooling system mounted for easy installation.............................................44-45
Certificates for Sensors, Instruments and Chemicals including Certificate of Calibration,
Electronic Calibration, Loop Calibration, & Material Certificates. .............................................................................46
Index for quick and easy searching for parts by number or description....................................................................47
2
Page 3
CHOOSING THE RIGHT MOUNTING SOLUTION
VP CONNECTOR SHOWN; SENSORS ARE ALSO AVAILABLE WITH INTEGRAL CABLE
PROCESS END (FRONT END) THREADS, CLOSEST TO SENSING/MEASURING END OF THE SENSOR
Figure 1. The 3900 general purpose pH sensor is offered with 1 inch MNPT threads in two locations: one on the process end and one on the cable end. It also
has a 3/4” MNPT process end thread.
Understanding sensor process connections:
Rosemount Analytical sensors are offered with a wide variety of threaded process connections.
The sensor shown above is a 3900 pH sensor. All sensors use similar terminology for process connection threads. Note that not all sensors are offered with front and back end threads. Below are examples of sensors using other types of sensor-to-process connection threads.
The sensors shown in Figures 2 and 3 are much different than the sensor shown in Figure 1. Figure 2 is a 228 Toroidal sensor with 3/4" MNPT threads and is also available with 5/8" UNC threads. Figure 3 shows a non-threaded 3300HT pH sensor which is used with a
BACK END PROCESS THREADS, CLOSEST TO CABLE OR VP CONNECTOR
process connector (sold separately) to allow variable insertion depths for the sensor and easy sensor removal without twisting the cable.
Ordering Information
All sensors used with the accessories mentioned in this catalog must be purchased separately. Also, please note that additional accessory items must be purchased along with the stated mounting accessories in order for the complete mounting to be accomplished. Please consult your local representative for more information about sensors or consult the factory at 1-800-854-8257 (in the U.S.) or 1-949-757-8500 (outside of the U.S.). You can also visit us at RosemountAnalytical.com.
Figure 2. The 228 Toroidal Conductivity Sensor is offered with back end process threads only.
3
Process connector required for connection of sensor to process.
Figure 3. The 3300HT sensor is offered without process threads. A compression fit, process connector is used on the sensor body and gives the user flexibility for insertion. Retractable sensors (a longer version of what is shown above) are also offered without process threads. This is so the sensor can be mounted through a ball valve.
Page 4
MOUNTING RECOMMENDATIONS FOR SENSORS
Sensors mounted in tees and flow cells:
> To prevent leakage, wrap the sensor threads with pipe
tape before inserting into its mounting.
> Do not over tighten the sensor when installing. Hand
tighten the sensor, and then tighten with a wrench.
> The electrodes must be completely submerged in the
process liquid to measure properly
> Air bubbles accumulating on sensors will cause erro-
neous or erratic readings.
> To control air bubble formation, apply a small amount
of back pressure to the flow cell or pipe.
> All sensor cables should be run through a conduit to
protect the back end from moisture intrusion.
Figure 4. pH sensor recommended mounting orienta­tion. pH sensors should be mounted at least 10° from the
horizontal to ensure correct measurement. Sensors can also be mounted vertically.
Figure 5. Chlorine, dissolved oxygen, and ozone sensor recommended mounting orientation. Chlorine, dissolved
oxygen, ozone sensors should be mounted vertically or within 45° of the horizontal position. Also, these sensors have recommended flow rates that should be observed. Check instruction manual for more information on recom­mended flow rates.
Figure 6. Conductivity (contacting) sensors should have at least 1/4 in. (0.6 cm) clearance between electrodes and piping (as shown in both illustrations above). The recom-
mended sensor orientation is either horizontal or no more than 30° below horizontal. It is best that the inlet flow is facing the sensor to ensure full pipe and minimal air bubbles.
4
Page 5
MOUNTING RECOMMENDATIONS FOR SENSORS
Metal Pipe
Plastic Pipe
True Conductivity
Distance to Wall
Measured Conductivity
3/4 to 1 Toroidal Sensor diameter
Conductivity (toroidal) sensors should be kept at a far enough distance between sensor and pipe wall to ensure the pipe wall does not interfere with the reading. If clearance is too small, the sensor should be calibrated
Figure 7. The measured toroidal conductivity value can be affected when installed close to a pipe wall.
Sensors submerged in tanks, ponds, or basins
> To prevent leakage, wrap the sensor threads with pipe
tape before inserting into its mounting part.
> Do not over tighten the sensor when installing. Hand
tighten the sensor, and then tighten one or two (1–2) turns with a wrench.
while in the pipe. The sensor should be completely submerged in liquid. Mounting the sensor in a vertical pipe run with the flow from bottom to top is best. If the sensor must be mounted in a horizontal pipe run, orient the sensor in the 3 o’clock or 9 o’clock position.
Figure 8. Correct sensor orientation
> The electrodes must be completely submerged in the
process liquid to measure properly.
> All sensor cables should be run through a pipe conduit;
do not use the sensors cable to dangle the sensor into the tank, pond, or basin.
Do not install sensor by hanging it directly into water.
Incorrect Installation
Figure 9. Correct sensor installation is critical to to proper sensor performance.
Correct Installation
Always install sensor using pipe conduit.
5
Page 6
INSERTION SENSOR MOUNTING
he mounting parts contained in this section include in-
T line tees, low flow cells, a handrail mounting assembly, a floating ball system, a jet spray cleaner, mounting
dapters, and process connectors. The in-line tees are
a offered in various configurations to assist in easy sensor insertion and removal without twisting cables. The low
low cells are available for sidestream samples with 1/4"
f OD tubing. The handrail mounting assembly is available with all the parts needed to install a sensor into a pond
S
e
nsor
P
r
o
d
uct
N
um
b
MEASUREMENT
e
TYPE
pH and ORP
389, 396, 396P
, T
F396, 3300HT
Conductivity
r basin; the jet-spray cleaner can be used with this
o assembly for dirty applications. The floating ball system can be used as an alternative method of installing a
ensor into a pond or basin. Many mounting adapters
s are offered for tank and pipe insertion, as well as for submersion applications. The process connectors are
ffered specifically for use with metal body sensors for
o insertion, submersion, and retractable applications.
r
s*
, 3500, 3900, Hx338, 3800, RB546
140, 400, 228
Dissolved Oxygen
499ADO, 499AT
RDO
Ozone
4
99AOZ
Chlorine
499ACL, 498Cl
* Sensors with integral cable or VP connector can be used.
6
Page 7
IN-LINE TEES
These versatile in-line tees can be used with Rosemount Analytical pH, ORP, conductivity, dissolved oxygen, ozone, and chlorine sensors having 1-inch process connections. Compatible sensor models are listed in the table below. Tees can be installed directly in process piping or in slip streams. PN 23567-00 and PN 915240­03, -04, and -05 have union fittings that allow the sensor
In-Line Tees
2002011 23567- 00 915240- 03 915240- 04 915240-05
Size 1.5" 2.0"
Process Connection Ends 1.0" thread 1.5" socket .75" threads 1.0" threads 1.5" threads
Material CPVC, Schedule 80 PVC, Schedule 80
O- ring or Gasket Material Buna N
Maximum rating 150 psig (1136 kPa) 65 psig (549 kPa) 60 psig (515 kPa) at 120°F (49°C)
@150°F (65°C) @ 122°F (50°C)
Compatible Sensor Models*
pH and ORP 396P, 389, 3500, 3900 396P, 389, 3500,
Conductivity 150
Dissolved Oxygen 499ADO 499ADO
Ozone 499AOZ 499AOZ
Chlorine 499ACl 499ACl
*Note: All sensor models noted in this graph are listed with integral cables.
The sensor models with the “VP” connectors can also be used.
In-Line Tee Part Number
to be removed without the need to unscrew it, thus eliminating the possibility of the cable becoming twisted and damaged. Tees are offered in PVC and CPVC and can connect to ¾, 1, and 1 ½ inch process lines.
lways install tees so that the sensor is completely
A submerged in the process liquid, and allow enough headroom to remove the sensor.
3900 using adapter PN S10283-LQD
Figure 10. When using sensor Model 3900VP
with Tee PN 915240-03, -04, or -05, order
adapter PN S10283-LQD (top of photo)
Direct mount into tee threads
IN-LINE TEE (PN 2002011) PRESSURE/TEMPERATURE SPECIFICATIONS
psig (kPa) °F (°C)
150 (1136) 150 (65)
128 (984) 160 (71) 102 (805) 170 (77)
80 (653) 180 (82) 57 (494) 200 (93) 48 (432) 210 (99)
FIGURE 11. 11⁄2 in. tee with 1 in. threaded process connection (PN 2002011) shown with TUpH 396P mounted for angle or straight process flow.
Figure 12. 389 pH sensor in PN 2002011 Tee
Figure 13. Sensor tip in process stream flow path.
7
Page 8
IN-LINE TEES
Tees with union adapters offer quick and easy access and eliminate damage from cable twisting.
Figure 14. 1½ in. tee with 1½ in. socket process connection (PN 23567-00) shown with a Model 499A amperometric sensor. The tee assembly is offered with
a sensor mounting adapter that allows the user to remove the sensor without twisting the cable.
Figure 15. 499ACL sensor in PN 23567-00 Tee
Figure 18. PERpH-X 3500VP pH sensor in PN 915240-03 Tee
Figure 16. Sensor tip in process stream flow path.
Figure 19. Sensor tip in process stream flow path
Figure 17. 2 in. tee with ¾, 1, or 1½ in. threaded process connections (PN 915240-03, -04, -05 respectively) shown with a Model 499A amperometric sensor. The tee assembly is offered with a sensor mounting adapter that allows the
user to remove the sensor without twisting the cable.
8
Figure 20. 3900VP pH sensor in PN 915240-03 Tee with adapter PN S10293-LDQ
Figure 21. Sensor tip in process stream flow path
Page 9
LOW FLOW CELLS
Emerson offers a variety of low flow cells for sidestream applications where it is impractical to divert a large volume of sample, particularly if the sample must be sent to waste. Choose a transparent plastic or stainless steel body—both have ¼ inch FPT process connections. Plastic flow cells PN 24091-00, -01, and -02 are also supplied with ¼ inch MPT to ¼ inch OD tubing fittings
Low Flow Cells
Process Connection 1/4 inch 316 SST male tube connector
Wetted Materials Body and Nut - Polycarbonate/polyester blend; 316 SST
Maximum Temperature 158°F (70°C) Consult Factory
Maximum Pressure 90 psig (621 kPa) Consult Factory
Sensor Threaded Connection 1 inch 1 inch 3/4 inch 1 inch 3/4 inch
Special features None
Compatible Sensor Models*
pH and ORP Sensors
Conductivity Sensors ––400, 400 VP 400, 400 VP
Dissolved Oxygen 499ATrDO, 499ADO 499ADO 499A DO
Ozone 499AOZ ––––
Chlorine 499ACL 498CL, 499ACL –––
*Note: All sensor models noted in this graph are listed with integral cables. The sensor models with the “VP” designation can also be used.
PN 24091-00 PN 24091-01 PN 24091-02 S10240 (SQ 7716 ) S10290 (SQ 7637)
1/4" Fittings - 316 SST; O-ring - Silicone
Bubble shedding nozzle
396P, 389, 3500, 3900
RB-546
and have a union connection that allows easy removal of the sensor.
Valved rotameters are also available (see page 41) to adjust and measure sample flow in sidestream
nstallations. Accurate control of flow is especially
i important when measuring dissolved oxygen, chlorine, and ozone.
None Order as a special request only
396P, 389, 3500, 3900
Figure 22. PN 24091-00 with 3900VP pH sensor
Figure 23. Low Flow Cell (PN 24091-00, 24091-01 or 24091-02) dimensions
Figure 24. PN S10290 with 400VP sensor
9
Figure 25. Metal Low Flow Cell (PN S10240 and S10290) dimensions
Page 10
HANDRAIL MOUNTING
Use a handrail mounting assembly whenever a pH, ORP, or dissolved oxygen sensor must be installed in a tank, aeration basin, pond, or open channel. Two versions are available. PN 11275-01 consists of a 13-ft (4.0-m) long PVC pipe with a 45° bend about two feet (0.6 m) from the lower end where the sensor is connected. The angle allows bubbles to slide off the sensor. Clamps attach the pipe to a frame that can be bolted to a handrail.
Handrail Mounting Assemblies
Part Number 11275-01
Process Connection Mounting brackets for Various Sizes of Handrails
Wetted Materials PVC NA NA
Temperature Rating 140°F (60°C) max. NA NA
Compatible Sensors*
pH and ORP Sensors 396P, 389, 3500, 3900
Dissolved Oxygen Sensors 499ADO**
Optional Accessory
* Note: All sensor models noted in this graph are listed with integral cables. The sensor models with the “VP”
connector can also be used.
* If used with the Handrail Mounting Assembly in a tank, the correct measurement can only be made if there is a
*
continuous flow or movement past the membrane of the sensor. These sensors will not operate properly in standing liquid.
*** 1 1/2 inch pipe not supplied in these systems.
Jet Spray Cleaner (see pg 13)
HRMS-01
***
HMRS-02
***
The pipe insertion depth is adjustable. The HRMS consists of a 1½ inch pipe clamp held by a clevis and pin assembly, which can be attached to a railing. The user supplies the 1 ½ inch pipe. The clamp can be rotated from side to side and up and down and locked in place once the pipe is in the desired position. The sensor attaches to the pipe using either a threaded adapter (HRMS-02) or a union coupling (HRMS-01).
HINT: Using a sensor with a VP connector and cable will allow for easier access to the sensor when cleaning or calibration is necessary.
Figure 26. Photo on left shows sensor being dangled into the pond, an incorrect installation. Photo on
right shows correct sensor mounting – sensor is threaded into a pipe conduit while cable is run through the pipe of the Handrail Mounting Assembly.
10
Figure 27. HRMS-01 shown with user­supplied 1-1/2" pipe (top) and assem­bled (bottom).
Page 11
HANDRAIL MOUNTING ASSEMBLY
Figure 28. Handrail Mounting Assembly (PN 11275-01) shown with Model 499A Dissolved Oxygen Sensor.
11
Page 12
FLOATING BALL SYSTEM
The floating ball system when used with the HRMS provides a convenient way of keeping the sensor a fixed distance below the surface of a pond or basin independent of the water level. The system accommodates any sensor having 1-inch back facing threads. The sensor attaches to the bottom of the float through a union coupling, and the sensor is held a 45° angle allows air bubbles to side off the sensor. The float can be readily inflated using a hand pump.
Floating Ball System
Part No. Description
BB11-01 For Model 499ADO; includes sensor union adapter
BB11-03 For any sensor with a 1 inch rear-facing MNPT
connection; includes union adapter
Wetted Material: Flexible PVC, rigid PVC
Temperature Range: 32°F to 140°F (0°C to 60°C)
Size (Diameter x Length): 11 x 13 inches (28 x 33 cm)
Maximum Inflation Pressure: 2.1 psig (0.15 bar)
Maximum Circumference: 34-1/2 inches (88.0 cm)
Buoyancy: 27lbs (12 kg)
Minimum Sensor Insertion Depth Into Process:
BB11-01 with 499ADO sensor: 13 inches (33 cm) BB11-03 with standard 1 inch sensor: 17 inches (43 cm)
Pipe Boom Support Arm: 1-1/2 inch Sch 80 PVC pipe (by others)
Sensor Compatibility: 1 inch MNPT on sensor cable end
12
Figure 29. Assembled Floating Ball BB11-01 for the 499ADO dissolved oxygen sensor
Figure 30. Floating Ball and Sensor with 1 inch Rear Facing MNPT Connection
Page 13
JET SPRAY CLEANER
The jet spray cleaner is intended for use with pH, ORP, or oxygen sensors. It directs a spray of air, water, or other cleaning fluid at the end of the sensor flushing away suspended solids that have accumulated there. The cleaner can be used only if the sensor is installed in a tank or basin. It is not suitable for in-pipe installations.
Cleaning frequency and spray duration is controlled by
n interval timer in the analyzer. During cleaning the
a analyzer can be programmed to hold the last reading, thus avoiding spurious outputs and relay actions. Hold can be continued beyond the end of the cleaning
Timer: supplied By
user or uTilize
imer funcTion
T
inside rosemounT
AnAlyTicAl
nsTrumenTs
i
period, allowing the sensor time to stabilize before the analyzer returns to normal operation.
Cleaning intervals and frequency as well as the best cleaning agent can be determined only by experience. Several trials will likely be necessary to determine the best program.
Jet Spray Cleaner
Part Number 12707-00
Process Connection 2" NPT threads OR use sensor/conduit
Wetted Materials 316 SST, polypropylene, PVC
Maximum Pressure 50 psig (446 Kpa)
Compatible sensors*
pH and ORP 396P, 389, 3900, 3500
Chlorine, dissolved 499ACL, 499ADO, 499AOZ oxygen, and ozone
*Note: All sensor models noted in this graph are listed with integral
cables. The sensor models with the “VP” designation can also be used.
Figure 31. Example of Jet Spray Cleaner with timer and solenoid valve for interval timed cleaning
Figure 32. Jet Spray Cleaner shown with Model 3900 pH sensor using the front threads to connect to the cleaner and the back threads to connect to a watertight pipe conduit
Typical install uses these threads
13
Page 14
MOUNTING ADAPTERS – THREADED CONNECTIONS
Use these adapters to install sensors with back facing threads into process piping or sample tees. The sensor screws into a plastic adapter that is held in the process connector by a union nut. The rear of the adapter is threaded for a conduit connection, if required. The union connection allows the sensor to be removed without the danger of twisting and damaging the cable.
Mounting Adapters
ADAPTER 23242-02 23242-03 2001990
Materials of 316 SST, 316 SST, CPVC, PEEK, Construction PEEK, Viton PEEK, Viton Viton
Process Connection 1-1/2" MNPT 1-1/2" MNPT 2" MNPT
Sensor Adapter/ 1" x 3/4" FPT 3/4" x 5/8 - 3/4" x 3/4" FPT Union Thread Size 11 UNC
Temperature, 392F (200C) 392F (200C) 100F(38C) at Maximum
Pressure, 295 psig 295 psig 100 psig Maximum (2135 kPa abs) (2135 kPa abs) (791 kPa abs)
Compatible Sensor Models*
pH and ORP 389, 396P
3500, 3900 --
Conductivity 228-21 228-20 228-21
*Note: Most sensor models noted in this graph are listed with integral cables. The sensor models with the “VP” designation can also be used.
100 psig (791 kPa abs)
OR
185F (85C) at
45 psig (412 kPa abs)
Figure 33. 3500 PERpH-X pH sensor with adapter PN 23242-02
Figure 35. The 228 Toroidal Conductivity Sensor with Adapter PN 23242-03
Figure 34. 228 Toroidal Conductivity sensor with adapter PN 23242-02 (or -03)
A “A” Dimension
Sensor* inches mm
396P 3.5 88.9
389 3.75 95.2 3900 4.5 114.3 3500 5.5 139.7
Figure 36. The 396P pH Sensor with Adapter PN 23242-02. Can also be used for connection
to a 228-21 toroidal conductivity sensor using the reversible PEEK adapter.
*The VP connector
versions can also be used.
Figure 37. Adapter PN 2001990 with 228-21 Toroidal Conductivity sensor
14
Page 15
MOUNTING ADAPTERS – SUBMERSION
Use this adapter to connect sensors with 1 inch back facing threads to standpipes for installation in tanks and basins.
Figure 38. Example of pipe mounting sensor with mounting adapter PN 9320057.
Figure 39. 499ADOVP Dissolved Oxygen Sensor with adapter PN 9320057
ADAPTER 9320057
Materials of PVC, Sch 80 Construction
Process Connection 1" FNPT
Sensor Adapter/ 1" FNPT Union Thread Size
Temperature, Ambient Maximum
Compatible Sensor Models*
pH and ORP 389, 396P,
3500, 3900
Conductivity -
Dissolved Oxygen 499ADO
Ozone 499AOZ
Chlorine 499ACL
*Note: Most sensor models noted in this graph are listed with integral cables. The sensor models with the “VP” designation can also be used.
15
Page 16
MOUNTING ADAPTERS – TRI-CLAMP®CONNECTIONS
Use these adapters to install threaded sensors into 1 ½ inch or 2 inch Tri-Clamp
®
process connections
Mounting Adapters
ADAPTER S10021 (SQ8940) SQP-10098-LQD* SQ 10904A & B
Materials of Construction 316 SST, 16 Ra finish 316 SST 304 SST
Process Connection 1.5" Tri-Clamp 2" Tri-Clamp
Sensor Adapter/Union Thread Size PG 13.5 1" 1" FNPT
Compatible Sensor Models**
pH and ORP 3800, Hx338 385+, 396, 396R, 398R, 389, 396P, 3500, 3900
TF396, 398, 3300HT, 3400HT
Conductivity –– –
Dissolved Oxygen Hx438 ––
* Formerly PN 23513-00 **Note: Most sensor models noted in this graph are listed with integral cables.
The sensor models with the “VP” connection can also be used.
2" (A) or 1.5" (B) Tri-Clamp
Figure 40. SQ10904 1.5" Tri-Clamp adapter (bottom) and 2" adapter (top)
Figure 41. 396P TUpH sensors shown with 1 inch threaded, 1.5 and 2 inch Tri-Clamp adapters
Figure 42. PN SQP-10098-LQD Compression fitting with Tri-Clamp Flange shown on TF396 sensor
16
Page 17
PROCESS CONNECTORS FOR 1 INCH SENSOR TUBES
Certain models of pH and ORP sensors lack threads and must be installed in the process piping using one of the fittings shown below. The sensor slides into the fitting, and a nut and ferrule tightened against the sensor tube holds the sensor in place. The insertion depth is adjustable. A collar on the tube prevents the sensor from
being ejected by the process pressure if the nut is not sufficiently tightened. The fittings are available with 1-in MPT and 2-in Tri-Clamp process connections. To prevent the sensor from turning and twisting the cable while a threaded fitting is being installed or removed, loosen the nut to free the sensor.
Process Connectors
ADAPTER PN 23166-00 PN 23166-01 PN 9510066 SQP-10098-LQD**
Materials of Construction 316 Stainless Steel Titanium Nylon 316 Stainless Steel
O-Ring Material EPDM EPDM Polyethylene Viton
Process Connection 1" MNPT 1" MNPT 1" MNPT 2" Tri-Clamp
Sensor Tube OD 1" (25.4 mm) 1" (25.4 mm) 1" (25.4 mm) 1" (25.4 mm)
Compatible pH and 385+, 396, 398, 396R, 385+, 396, 398, 396R, 396, 398, 3300HT 385+, 396, TF396, 398, ORP Sensors 398R, 3300HT, 3400HT 398R, 3300HT, 3400HT
*Note: Most sensor models noted in this graph are listed with integral cables. The sensor models with the “VP” designation can also be used. **Formerly PN 23513-00
396R, 398R, 3300HT, 3400HT
17
Page 18
PROCESS CONNECTORS FOR 1 INCH SENSOR TUBES
Figure 43. Metal process connector (PN 23166-00 [316 SST] or PN 23166-01 [titanium]) shown with TUpH Model 396 pH sensor. This process connector can be used for all 396, 398, and 3300HT sensor
installations. Also, it must be used for 396R, 398R, and 3400HT installations with or without a ball valve. See ball valve section for more information.
Figure 44. The 23166-00 (or -01) metal process connector, shown with the 3300HT PERpH-X sensor, allows various insertions depths, depending on where the user locates the compression fitting.
Also the threads can be switched to face the top end of the sensor (right) for connection to submersion pipes.
Figure 45. Nylon Process Connector PN 9150066 can be installed with threads facing up or down. NOTE: This
connector should be used for low pressure, non-aggressive chemical processes only.
Figure 46. 398 TUpH pH Sensor with SQP-10098­LQD Tri-Clamp flange adapter
18
Figure 47. PN SQP-10098-LQD Compression fitting with Tri-Clamp Flange shown here with TF396. Can
also be used with the 396, 398, and 3300HT sensors.
Page 19
RETRACTION ASSEMBLIES
Retraction assemblies allow a pH, ORP, or conductivity sensor (both contacting and toroidal) to be inserted or removed from a vessel or pipe without the need for depressurizing or draining it. Insertion depth is adjustable. Five retraction assemblies are available. Two
S
e
nsor
M
o
d
e
l
N
um
b
ME
A
S
UR
E
T
Y
PE
p
H
a
nd
O
R
Conductivit
MEN
P
y
T
385+, 396, 396P
228, 402
e
, 396R, 398, 398R, 3300HT
include the ball valve needed to close off the process line before the sensor is removed. Two require the ball valve to be purchased separately. The fifth assembly is a rotary retraction valve.
r
s
, 3400HT
See following pages for retractable mounting options.
19
Page 20
RETRACTION ASSEMBLIES – BALL VALVE INCLUDED
Two retraction assemblies are available. PN 23240-00 has a 1 ½ inch full port ball valve and PN 23765-00 has a 1 ¼ inch ball valve. Both assemblies can be used with the pH sensors listed in the table below. Model 402 and 402VP sensors can be used only with PN 23765-00.
PN 23765-00 includes a retraction chamber, 1 ¼ inch ball valve, and close nipple for attaching the retraction
Ball Valve Kits
1 1/2 in. Ball Valve Kit 1 1/4 in. Ball Valve Kit
Process Connections 1 1/2" FNPT 1 1/4" MNPT
Wetted Materials 316 SST 316 SST
Temperature Rating 32-212°F(0-100°C) 32-212°F(0-100°C)
Pressure Rating 100 psig (790 kPa) 200 psig (1481 kPa)
Maximum Retraction For 21 inch pH sensors: 64 psig (542 kPa) Pressure 64 psig (542 kPa)
For 36 inch pH sensors:
35 psig (343 kPa)
Compatible Sensor Models*
pH and ORP 396R, 398R, 385, 385+, 3400HT, RB-547
Conductivity 402
Other information Use PN 23166-00 or -01 If ball valve is already
sensor to ball valve.** alone is available (PN 23796-00)
Use 1-1/2" close nipple (PN 93101-02)
to attach ball valve to FTP
process connection.
assembly to an FPT process connection. PN 23240-00 includes only the ball valve and retraction chamber. A fitting (PN 23166-00 or 23166-01) is required to connect the sensor to the retraction chamber, and a 1 ½ inch close nipple (PN 9310102) is needed to attach the ball valve to a FPT process connection.
PN 23240-00 PN 23765-00
to connect available, the retraction assembly
* All sensor models noted in this graph are listed with integral cables. The sensor models with the
“VP” connector can also be used.
**Process connector is included with the 385 and 385+ sensors
Short insertion
Long insertion
Figure 48. Retractable sensors have the distinct advantage of having an adjustable insertion depth. Sensor tubes can be installed with a short (top photo) or a longer (bottom) insertion. See sensor model information for more about sensor tube lengths.
20
Page 21
RETRACTION ASSEMBLIES – BALL VALVE INCLUDED
pH sensors should be mounted at least 10 degrees above horizontal
Conductivity sensors should be mounted between 0 and -30 degrees from the horizontal
Figure 49. Examples of sensor mounting using a ball valve retraction kit connected to a standard pipe tee, pipe “Y”, or pipe weldolet.
To retract a sensor:
1) Loosen the retraction chamber or process connector nut.
2) Slide the sensor into the retraction chamber. The stop collar on the sensor prevents the sensor from sliding out of the retraction chamber.
0
0
-
5
6
7
3
2
PN
Kit
lve
Va
ll
a
B
¼
1
Figure 50. Observe the above mounting angles to obtain best measuring results.
3) Close the ball valve.
4) Remove the process connector or retraction chamber and sensor from the ball valve.
1.
” Ball Valve Kit PN 23240-00
½
1
Figure 51. Follow these four steps to easily remove the sensor from a pressurized system.
2
sensor out of process
e
Slid
.
retraction chamber
to
in
Close ball valve
3 .
4 .
t
u
n
n
Loose
21
Remove retraction chamber plus sensor
Sensor stop collar
Page 22
RETRACTION ASSEMBLIES – BALL VALVE INCLUDED
BASENSOR TUBE
mm/in mm/in
LENGTH
327.7/12.9 586.2/23.1 21 in.
660.0/27.4 954.5/37.58 36 in.
Figure 52. 1½ in. FNPT Ball Valve Kit (PN 23240-00) shown with process connector (sold separately) and TUpH Model 396R pH sensor. Ball valve kit includes a 1½ in. full port ball valve, a 1½ in. FTP to 1 in. FTP reducer (which serves as the retraction
chamber) and a 1½ in. close nipple that connects the reducer to the ball valve. A connector (PN 23166-00 or 23166-01) is required to connect sensor to the reducer.
ro
P PN
o
(s
q
re sens va
c
1
0
-
r
o
0
0
-
6
6
1
3
2
epar
s
ld
d
e
ir
u
r tub
o
it .
e k
lv
t
a
o
e to
tel co
is
)
y
ct
e
n
n
the bal
e
th
l
Ball valve kit PN 23240-00 includes
1
inch ball valve, a close nipple
2
a 1
o connect the reducer to the ball
t
1
lo
h c
inc
2
1
lu
inc
(not
ble fo
a
l
vai
a an M the 1
co
T
P
N
1
inch FNPT threads.
2
pple PN
e ni
s
ded in ball val
io
at
all
inst
r
n to
io
nnect
9310102
) is
e kit
v
ns requirin
e to
at
m
g
r
o
ct
e
n
n
Co
s
es
valve. The reducer is also the retraction chamber.
Figure 53. Ball valve kit PN 23240-00 with other accessories (sold separately).
22
Page 23
RETRACTION ASSEMBLIES – BALL VALVE INCLUDED
Figure 54. 1¼ in. MNPT Ball Valve Kit (PN 23765-00 or 23765-01) shown with retraction kit and Endurance Model 402 conductivity sensor. Ball valve kit includes a 1¼ in. full port ball valve, a 1¼ in. nipple, and a retraction chamber.
Model A Dimensions B Dimensions C Dimensions
Minimum* Maximum
inches mm inches mm inches mm inches mm 402-11 3.21 82 5.34 136 18.13 460.5 1.19 30.2 402-12 2.34 59 4.47 114 17.33 440.2 1.19 30.2 402-13 2.36 60 4.49 114 17.33 440.2 1.19 30.2 402-14 3.89 99 4.91 125 18.83 478.3 1.19 30.2
396R-21 2.0 50.8 12.12 307.85 21 533.4 1.0 25.4 398R-21 2.0 50.8 12.12 307.85 21 533.4 1.0 25.4
3400HT-21 2.0 50.8 12.12 307.85 21 533.4 1.0 25.4
396R-25 2.0 50.8 26.62 676.15 36 914.4 1.0 25.4 398R-25 2.0 50.8 26.62 676.15 36 914.4 1.0 25.4
3400HT-25 2.0 50.8 26.62 676.15 36 914.4 1.0 25.4
*Minimum dimension must be met in order for sensor to measure correctly
Retraction
Chamber
1
" Full port ball valve
4
1
1
" nipple
4
1
Figure 55. The ball valve kit (PN 23765-00 or -01) has all parts needed to install the sensor. The retraction chamber (PN 23796-00) can be purchased separately.
23
Page 24
RETRACTION ASSEMBLIES (BALL VALVES NOT INCLUDED)
These retraction assemblies are primarily intended for use with Model 228 toroidal conductivity sensors.
owever, they can be modified for use with certain pH
H and contacting conductivity sensors. Consult the factory for information.
A mechanical (PN 23311-00) and manual (PN 23311-
01) retraction assembly are available. The mechanical
Shown with Model
28 Toroidal Sensor
2
retraction assembly, which uses a drive screw to insert and retract the sensor, can be operated against a
ressure as high as 295 psig (2036 kPa abs). Both
p retraction assemblies are fitted with 1/8" flush ports on the retraction chamber to allow the sensor to be
ashed clean of sample before it is removed. Both
w assemblies were designed to be used with a 1-1/2" full port ball valve (purchased separately – PN 9340065).
hown with Model
S 228 Toroidal Sensor
PART NUMBER 23311-00 PART NUMBER 23311-01 Mechanical Retraction Assembly Manual Retraction Assembly
Process Connection 1.5" MNPT 1.5" MNPT
Wetted Materials 316 SST, Teflon, EP 316 SST, Teflon, EP
Maximum Operating Temperature 392°F (200°C) 392°F (200°C)
Maximum Operating Pressure 295 psig (2036 kPa) 295 psig (2036 kPa)
Max. Insertion/Retraction Conditions 392°F (200°C) 266°F (130°C)
295 psig (2036 kPa) 35 psig (241 kPa)
Maximum Insertion Travel 8.5" (21.6 cm) 12" (30 cm)
Maximum Insertion Travel 6.25" (15.9 cm) 7.0" (17.8 cm) using PN 9340065 Ball Valve
Compatible Toroidal Model 228 Model 228 Conductivity Sensors
Compatible Contacting Consult factory for information Consult factory for information Conductivity Sensors
Compatible pH & ORP Sensors Consult factory for information Consult factory for information
24
Page 25
RETRACTION ASSEMBLIES (BALL VALVES NOT INCLUDED)
Figure 56. Mechanical Retraction/Insertion Assembly (PN 23311-00). To remove the sensor, use a socket
wrench to turn the drive screw until the sensor is inside the retraction chamber. Close the ball valve (not shown) and flush the retraction chamber. Loosen the hex union nut. Remove the stop collar and the front half of the guide clamp. Remove the sensor.
25
Figure 57. Manual Retraction/Insertion Assembly (PN 23311-01). To remove the sensor, loosen the
collet nut and pull the sensor into the retraction chamber. Close the ball valve (not shown) and flush the chamber. Loosen the hex union nut and remove the sensor.
Page 26
PASVE ROTARY RETRACTION VALVE
The PASVE is a mounting/service valve for pH sensors. It allows the cleaning and calibration of pH sensors without stopping the process. When required, this can be done automatically.
hen using the sensor in abrasive processes, it can be
W rotated to the measuring position only for the duration of the actual measurement.
The PASVE valve is available in a manually operated type or equipped with a pneumatic or electric actuator. It is offered in a version for welding directly onto the pipe or tank. It is also available with flanged connections. Contact factory for more details.
Figure 58. PASVE valve, shown in welded on tank, manual operation configuration
PASVE Specifications
Process Connections Code B: Welded on Tank at 15° angle
Code C: Welded on Tank Code P: Welded on Pipe Code F: Flange Code D: Flow through with flanged connection
Wetted Materials (Valve) AISI 316LSST or Titanium
Wetted Materials (Seals) PTFE (Teflon) or PTFE (With carbon graphite filling)
Maximum Operating 482°F (250°C) Temperature
Maximum Operating Pressure 570 psig (4000 kPa)
Cleaning Port Connections ¼ inch
Compatible pH Sensors 396, 396P, 389, 398, and 3300HT (with SQ9608)
385+, 396PVP, 3300HTVP (SQ 9608) and no others
26
Page 27
PASVE ROTARY RETRACTION VALVE
Figure 59. Operating PASVE Measuring Position Figure 60. Operating PASVE Servicing and Calibration
Position
Figure 61. Pasve valve with spring return, double action, and pneumatic (left) or electric (right) actuator
Flush ports to connect cleaning liquid
or air to spray directly onto the sensor.
Spray tube from cleaning port.
Sensing end of the pH sensor.
Figure 63. PASVE Example of Flush Water Set-Up
Figure 62. Close-up of PASVE valve with cleaning ports
27
Page 28
INSTRUMENT MOUNTING
The instrument mounting accessories found in this section include instrument mounting kits for pipe or wall mounting. These can be used with the following Rosemount Analytical Instruments:
TWO-WIRE INSTRUMENTS
5081 Transmitter
Features a rugged, weatherproof, corrosion-resistant enclosure.
> Meets NEMA 7B explosion proof
standards.
> Digital communication protocols:
®
HART
and FOUNDATION®fieldbus, allowing access to
AMS (Asset Management Systems)
1066
Tr
ansmit
Br
oad
r
ange
a
dv
anced
c
ap
a
b
i
lity,
>
U
ses
HA
F
O
UN
DA
T
comm
unic
>
U
se
r-
defina
WIRELESS INSTRUMENTATION
6081
Remote locations and installation costs are no longer barriers to getting information you require for critical applications.
> 6081-C measures conductivity
> 6081-P measures pH
> Self-organizing network for
high data reliability and network stability
> Industry leading wireless
security
of
comm
and
R
T
ve
IO
N
fie
ation
ble
te
mea
unic
uni
rs
ldbus
mea
que
io
pr
r
s
ur
ations
n
7
o
s
e
me
ea
se-
and
di
gital
tocols
ur
e
nts
me
F
OUR-WI
5
6
>
H c p
>
E
>
P t
> Pr
to a
,
of
-
use
.
.
nt
dia
gnos
tic
p
ar
amet
e
> Pr
De
> T
Wid
1056
>
Sin
rs
.
>
L allo g allo
>
E
1057
>
T
>
C Contacting Conductivity/ Resistivity, in any combination.
A
n
ig
h
o
lo
a
r
a
x
t
e
r
o
c
r
e
n
ocess
op
r
ivat
P
C (Du
g
a
rg
w
l
ance
wable
a
s
y to ins
hr
e
hoo
a
ly
r
eso
r
f
o
r
met
n
s
iv
e
ess
d
din
g
d
USB d
or
t
ional,
ive
t
t
h Mod
A
n
aly
le
or
e,
e
a
s
op
if
A
n
aly
e-in
se
f
R
E
ze
r
lu
imme
e
r
.
h
e
is
r
g
r
aphs
ata c
r
ive.
(PID
y Cy
zer
d
ual in
s
y
-to-r
e
r
t
he
par
tall an
zer
pu
t
r
om p
I
t
io
lp
u
ul
ators
pr
a
NS
T
R
UM
n
sc
r
e
e
n
d
i
a
t
e
r
e
sc
r
e
e
n
s
p
t
io
n
s
c
a
n
.
an b
e
do
Inte
g
r
a
l an
) c
ont
r
ol.
cle
C
ont
r
ol/Pulse
a
t
ion).
pu
t
ana
e
ad
displ
kno
w
ocess
is
amete
rs
.
d
w
ir
e.
naly
ze
r.
H/ORP/ISE or
c
f
w
at w
E
s
h
o
o
b
ly
NT
o
w
g
n
it
r
r
a
e
p
nload
d
ze
r.
a
y
a
it
hin
S
s
fa
u
lt
s
a
n
d
w
a
r
n
in
g
s
in
io
n
o
f
a
n
o
u
t
o
f
r
a
n
g
e
p
id
s
ta
r
t
u
p.
in
p
o
int
e
d
in
t
h
e
c
o
lo
r
e
d
28
Page 29
INSTRUMENT MOUNTING ACCESSORIES
R
os
e
m
o
u
n
t
A
na
l
y
ti
c
a
l
in
s
tru
m
e
n
ts
c
a
n
b
e
m
o
p
i
p
e,
in
a
p
a
n
e
l
,
or
on
a
w
a
ll
u
s
in
g
the
v
a
r
i
o
u
s
m
o
u
n
tin
g
a
c
c
e
s
s
or
i
e
s
.
T
he
s
e
k
its
c
on
t
a
in
a
ll
p
n
eed
ed
for
a
s
u
c
c
e
s
s
f
u
l
in
s
tru
m
e
n
t
in
s
t
a
ll
a
ti
on
Mounting part 23820-00 23820-01
Bracket Carbon Steel, blue epoxy coating 316 SST
Bracket Dimensions 7.5" x 4.0" x 1.2"
U-Bolt SST
U-Bolt Dimensions for 2" pipe maximum
Screws and washers SST
Compatible Instruments 56, 1056, 1057, 1066, 5081, 6081
u
n
t
ed
v
e
rsa
ti
a
r
ts
.
PN 23820-00 and 23820-01 are Mounting Bracket Kits
on
a
for Pipe, Surface or Wall Mounting. Kit includes one
le
bracket and two u-bolts with washers and screws. Use for mounting Rosemount 56, 1056, 1057, 1066, 5081, and 6081 instruments.
Figure 64. These two graphics are examples of pipe or wall mounting a 6081 instrument using either of the mounting bracket kits. On the left is an example of the instrument mounting onto a
pipe. On the right, the instrument uses the kit to mount onto a wall. Other Rosemount Analytical instruments can be mounted similarly.
29
Page 30
INSTRUMENT MOUNTING ACCESSORIES
165
6.5
130
5.1
33.5
1.3
232
9.1
2 INCH PI PE SUPPLIED
BY CUSTOMER
SIDE VIEW
0.85”
22 mm
4.0”
102 mm
1.2”
30.5 mm
1.2”
30.5 mm
0.85”
22 mm
3.0”
76 mm
6.5”
1
65 mm
2
.8”
7
1 mm
3.0”
76 mm
Figure 65. 5081 pipe mounted with the PN 23820-01 SST Kit.
Figure 67. 1056 shown mounted on a pipe using either pipe mount kit, side view.
30
Figure 66. Dimensions of the mounting bracket found in kits PN 23820-00 and 23820-01
Figure 68. The 1056 is shown mounted on a wall/surface using either pipe mount kit, side view.
Page 31
WIRING ACCESSORIES
This section features junction boxes and cables compatible with Rosemount Analytical sensors and instruments.
Cable Extensions through Junction Boxes
Use a junction box when the analyzer must be installed
urther from the sensor than the standard cable length
f allows. To replace the sensor, disconnect the cable at the junction box. The cable between the junction box and the analyzer never needs to be disturbed.
> The distances between the sensor and instrument are
recommendations only.
> Long cable distances may degrade the signal; for best
performance, install long cable runs in a noise-free environment.
> For contacting conductivity sensors, long cable runs
will result in some loss of linearity.
> Consider an Emerson Wireless solution instead.
Mount the analyzer within standard cable length of the sensor and send the signal wirelessly to the control room. Emerson offers wireless pH and conduc­tivity instruments as well as the THUM, which converts any Rosemount Analytical HART instrument into a wireless device. All wireless instruments require
a Model 1420 Gateway to receive the signal.
Best Practice Tip
The best practice for extended distances are recom­mended below to assist with the easiest sensor replace­ment. Follow this guideline when installing sensors at a long distance from the instrument:
1. Install the junction box closest to the sensor.
2. Use an extension cable to run the signal from the junction box to the instrument. The use of factory prepared cable is strongly recommended. The cables designated as “prepped” are prepared with the cable jacket and all of the wires stripped and ready for a quick installation.
3. Distances between the sensor and instrument are cumulative and include the sensor cable plus the extension cable. For instance, if the maximum cable required is 200 ft (61.0 m) and the sensor cable is 15 ft (4.6 m), the extension cable maximum length is 185 ft (56.4m).
Chlorine, Dissolved Oxygen, and Ozone Sensors
Sensor to Analyzer Separation Distances
Sensor Cable Length
Measurement Sensor Standard Junction Box Prepped Unprepped
Dissolved Oxygen 499ADO 25 ft (7.6m) 200 ft (61.0m) 23550-00 23747-00 9200275
Chlorine 498CL or 499ACL 25 ft (7.6m) 200 ft (61.0m) 23550-00 23747-00 9200275
Ozone 499AOZ 25 ft (7.6m) 200 ft (61.0m) 23550-00 23747-00 9200275
Chlorine, Dissolved Oxygen or Ozone Sensor
Standard sensor cable or VP cable (50 ft (15m) max)
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
Recommended maximum distance
between sensor and instrument*
Junction box PN 23550-00
31
Extension cable
Extension Cable
Analyzer
Page 32
WIRING ACCESSORIES
Conductivity Sensors
Sensor to Analyzer Separation Distances
Sensor Cable Length
Recommended
maximum distance
etween sensor and
Measurement Sensor Standard Optional Junction Box Prepped Unprepped
b
instrument*
140, 141, 142 No Cable; use extension cable 200 ft (61.0m) Included 23747-00 9200275 Contacting 150 10 ft (3.1m) 200 ft (61.0m) 23550-00 23747-00 9200275 Conductivity 400, 401, 403, 404 10 ft (3.1m) 50 ft (15.2m) 200 ft (61.0m) 23550-00 23747-00 9200275
402 10 ft (3.1 m) 200 ft (61.0m) Included 23747-00 9200275
225, 226, 228 20 ft (6.1m) 200 ft (61.0m) 23550-00 23294-04 9200276 Toroidal 222 20 ft (6.1m) 100 ft (30.5m) 23550-00 23294-04 9200276 Conductivity 242, 245 Use extension 200 ft (61.0m) Included 23909-00
(flow through) cable
Extension Cable
Contacting Conductivity Sensors
Toroidal Conductivity Sensors
Standard sensor cable or VP cable (50 ft (15m) max)
J
unctio
P
N
n
23550-
b
o
x
00
Extension cable
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
Standard sensor cable or VP cable (50 ft (15m) max)
Junction box PN 23550-00
Extension cable
Maximum recommended distance between sensor and instrument is 100 ft (30.5m) to 200 ft
(61.0m) depending on the sensor model. See chart for more information.
Analyzer
Contacting or Toroidal Conductivity Sensors with sensor mounted junction box
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
Extension cable
32
Page 33
WIRING ACCESSORIES
pH and ORP Sensors
Sensor to Analyzer Separation Distances
Sensor Cable Length
Recommended
maximum distance
etween sensor and
Measurement Sensor Standard Optional Junction Box Prepped Unprepped
b
instrument*
No preamp 15 ft (4.6m) 200 ft (61.0m) 23555-00 23646-01 9200273
pH Sensors in the sensor (contains preamp)
Preamp in the sensor 25 ft (7.6m) 200 ft (61.0m) 23550-00 23646-01 9200273
Preamplifier in Sensor No Cable; 200 ft (61.0m) - 23646-01 9200273
Mounted Junction Box use extension cable
pH or ORP Sensor ­NON preamplified version
Extension Cable
pH or ORP Sensor with integral preamplifier
pH or ORP Sensor with integral preamplifier
Standard sensor cable or VP cable (50 ft (15m) max)
Junction box PN 23555-00
WITH preamplifier
Extension cable
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
Standard sensor cable or VP cable (50 ft (15m) max)
Junction box PN 23550-00
with extension board
Extension cable
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
Extension cable
Maximum recommended distance between sensor and instrument is 200 ft (61.0m)
33
Page 34
REMOTE JUNCTION BOXES
Use remote junction boxes to extend the sensor cable
for longer runs. Two captive screws allow easy access to
the board inside the box. There are two ports for cable;
the bottom port is for the sensor cable and the side port
is for the extension cable. These weatherproof junction
boxes are also ideal for indoor mounting.
> PN 23550-00 can be used to extend the cable of
lmost any Rosemount Analytical sensor. The terminal
a board allows point-to-point wiring between the sensor and extension cable.
> PN 23555-00 has an integral preamplifier (PN 23557-
00) and is used for pH and ORP sensors only. The preamplifier has FM, CSA and ATEX approval for use in hazardous locations. Any non-preamplified Rosemount Analytical pH or ORP sensor can be used with this junction box. The preamplified signal is compatible with Rosemount Analytical 56, 1056, 1057, 1066, 5081, and 6081 instruments.
> Both junction boxes have identical dimensions.
Figure 69. Junction Box PN 23550-00.
Figure 70. Junction Box PN 23550-00 is used for extending any Rosemount Analytical sensor’s cable by connecting the sensor cable to an internal extension board. Extension cable
sold separately.
Figure 71. Junction Box PN 23555-00 is used for pH and ORP sensors only. A preamplifier board is mounted internally.
Figure 72. Junction Box PN 23555-00 dimensions. Junction Box PN 23550-00 has identical dimensions.
34
Page 35
MOUNTING BRACKET FOR REMOTE JUNCTION BOXES
Remote mount junction boxes can be mounted on a wall or pipe using the Mounting Bracket PN 2002526.
Mounting Bracket Specifications
Mounting part 2002526
U-Bolt SST
U-Bolt Dimensions for 2" pipe maximum
Screws and washers SST
Compatible Junction Boxes 23550-00 and 23555-00
SENSOR HEAD JUNCTION BOX
Sensor Head Junction box PN 23709-00 is for use with all Rosemount Analytical retractable style pH and ORP sensors. It is mounted on the back end of the sensor using a compression fitting. The junction box contains an integral preamplifier (PN 23557-00) and has termi­nals to connect an extension cable (sold separately) directly to an instrument.
11.5 inches 292 mm
.4 inches
4
—————
12 mm
1
> Dimensions: 5"D x 5
3
8" H x 4"W
> Materials: Aluminum Alloy
> Preamplifier: FM, CSA, ATEX
> Compatible Sensors: 396R, 398R, 3400HT and RB547
35
Figure 72. Sesor Head Junction Box PN PN 23709-00.
Page 36
VP CONNECTOR CABLES, EXTENSION CABLES, AND CABLE GLANDS
Rosemount Analytical offers various extension and connector cables for the best EMI protection in the industry. VP connector cables are used with sensors
VP Connector Cables
pH and ORP Sensor Cables
The VP8 connector cable is compatible with the following sensors: 3300HTVP, 3400HTVP, 3500VP, 3900VP, 389VP, 396VP, 396PVP, 396RVP, 398VP, 398RVP, HX338, 3800VP.
Part Number VP8 Cable for pH & ORP sensors
Feet Meters
24281-02 2.5 0.8
24281-05 4 1.2
24281-06 10 3.0
24281-00 15 5
24281-07 20 6.1
24281-01 25 7.6
24281-08 30 9.1
24281-03 50 15.2
24281-04 100 30.5
having VP quick disconnect fittings. Extension cables are used with sensor mounted or remote junction boxes. Follow this simple guide to select the appropriate cable.
The VP8 Cable PN 24281-XX is a 6 conductor cable with 1 coax and 3 shields.
Jacket Material: PVC
Jacket Maximum Diameter: 0.31 inches
Temperature Range: -40C to 105C; sunlight resistant
Figure 73. VP8 connector cables (bottom) can be used with all pH and ORP sensors with a VP designation. 3500VP (top)
and 396PVP shown.
Conductivity, DO, OZ, and Chlorine Sensor Cables
VP6 cable PN 23747-XX is compatible with the following sensors: 400VP, 402 VP, 403VP, 499ACL-VP, 499ADO-VP, 400ATrDO-VP, and 499AOZ-VP. For 410VP (four-electrode contacting sensor) use PN 24287-00 or -01.
VP6 Cable
for conductivity,
dissolved oxygen, ozone
& chlorine sensors
Part Number Feet Meters
23747-02 10 3.0
23747-03 50 15.2
23747-04 4 1.2
23747-05 2.25 0.7
23747-06 2.5 0.8
23747-07 15 4.6
23747-08 20 6.1
23747-09 25 7.6
23747-10 30 9.1
23747-11 100 30.5
24287-00 10 3.0
24287-01 20 15.2
The VP6 Cable PN 23747-XX is a 3 conductor cable with 2 coax and 3 shields.
Jacket Material: PVC
Jacket Maximum Diameter: 0.31 inches
Temperature Range: -40C to 105C; sunlight resistant
The VP6 cable PN 24287-XX is a cable with 2 twisted pairs and 1 triad.
Jacket Material: PVC
Jacket Maximum Diameter: 0.31 inches
Temperature Range: 0C to 105C; sunlight resistant
Figure 74. VP6 connector cables can be used with conduc­tivity and amperometric sensors with a VP designation. 499ADO-VP and 400VP (bottom) shown.
36
Page 37
EXTENSION CABLES
Rosemount Analytical offers two types of cables; prepped and unprepped. The unprepped cables are raw cables; user must cut the cable jacket and strip all of the wires before use. The cables designated as prepped are
prepared with the cable jacket and all of the wires stripped and ready for quick installation. The use of factory­prepped cables is strongly recommended. All extension cables can be ordered at user specified length.
pH and ORP Sensor Extension Cables
3300HT/HTVP
3400HT/HTVP
3500/VP
3900/VP
385+
389/VP
396/VP / 398/VP
396P/VP
396R/VP 398R/VP
398R/VP
Cable Part No. Cable Description
9200273 Gray Ext Cable, 11-cond, shielded, unprepped XXX XXXXXXX
23646-01 Extension Cable, 11-cond, shielded, XXX X XX XXX
prepped (raw cable PN 9200273)
Conductivity
Amperometric*
Conductivity and Amperometric Sensor Extension Cables
Cable Part Extension Cable Description Number
9200275 Shielded 9-cond, unprepped XX X
9200276 Unprepped XXX
9200334 For 410VP sensor X
23294-00 Unshielded, Prepped XX
23294-04 Shielded for improved EMI/RFI protection. XX
Prepped
23294-05 Shielded with additional shield wire for -03 option. XX
Prepped
23747-00 Prepped XX X X
23909-00 Prepped XXX
24289-00 For 410VP sensor X
9200275 is the raw cable for prepped cable 23747-00 9200276 is the raw cable for prepped cables 23909-00, 23294-00, 23294-04, 23294-05 * Amperometric sensors measure dissolved oxygen, ozone, and/or chlorine
140, 141, 142
150
222
225, 226, 228
242
245
400 & 400VP, 401,
402 & 402VP, 403
& 403VP, 404
410VP
499ACl, 499ADO
499AOZ,
499ATRDO
Cable Glands
Use cable glands to seal the cable where it enters the analyzer or junction box. Cable glands also prevent the weight of the cable from pulling on and dislodging wiring connections in the analyzer. Cable gland kit PN 23554-00 provides 5 cable glands that fit all Rosemount Analytical instruments and junction boxes.
37
Figure 75. Cable gland found in kit PN 23554-00
Page 38
CALIBRATION ACCESSORIES
Portable Conductivity Validation Unit
he conductivity validation unit (CVU) consists of a
T factory-calibrated Model 1056 conductivity analyzer and Model 404 flow-through conductivity sensor
oused in a rugged carrying case. The CVU is ideal for
h applications where calibrating a conductivity sensor against a standard solution is impractical. This includes
ensors that cannot easily be removed from the process
s piping and sensors having 0.01/cm cell constants for which stable low conductivity calibration standards are not available. The CVU is widely used for calibrating conductivity sensors in water for injection applications.
Using the CVU is easy. Connect the Model 404 sensor in series with the process sensor using the tubing provided with the CVU. Allow the sample to flow past both sensors. Once readings are stable, adjust the process analyzer reading to match the CVU reading.
The calibration system is completely enclosed. The solu­tion never contacts atmospheric carbon dioxide, which can contaminate low conductivity standards, leading to erroneous calibration. Thus, the CVU is ideal for cali­brating 0.01/cm cell constant sensors.
Figure 76. The 1056 conductivity instrument and 404 conductivity sensor are hard mounted into a protective case. Each unit includes all cabling and tubing needed for
connection to a water system.
The CVU is available in two versions:
> CVU-01: One 404-11-17 Stainless Steel Sensor with a
0.01 cell constant pre-wired to the 1056-01-20-38-AN analyzer.
> CVU-02: Two sensors, one 404-11-17 Stainless Steel
Sensor with a 0.01 cell constant and one 404-12-17 Stainless Steel Sensor with a 0.1 cell constant pre­wired to the model 1056-01-20-30-AN analyzer.
See Product Data Sheet PDS 71-CONDVALII for more information.
38
Page 39
CALIBRATION STANDARDS
pH Buffer Solutions
Emerson offers pH 4.01, pH 6.86, and pH 9.18 NIST buffer solutions that meet most pH calibration require­ments. Accuracy is ±0.02 pH at 25°C (77°F). A chart listing the pH of the buffer at other temperatures is provided on the label.
Other buffer solutions, pH 4.00, pH 7.00, and pH 10.00, are available. These solutions are color-coded for easy recognition. They are available as single bottles or in packages of six.
All Rosemount Analytical pH instruments have an auto-
matic buffer calibration features that recognizes the pH of the buffer and uses the temperature-adjusted pH value during calibration.
A 475 mV ORP (redox) standard solution is also available.
Part Number Description Size Color Code
9210012 Rosemount Analytical pH 4.01 buffer solution 16 oz (473 ml) Pink
9210013 Rosemount Analytical pH 6.86 buffer solution 16 oz (473 ml) Clear
9210014 Rosemount Analytical pH 9.18 buffer solution 16 oz (473 ml) Clear
905-3506 pH 4 buffer solution 16 oz (473 ml) Red
905-582517 6 pack of pH 4 buffer solution (PN 905- 3506) Box of 6 16 oz (473 ml) bottles Red
905- 3501 pH 7 buffer solution 16 oz (473 ml) Green
905- 582521 6 pack of pH 7 buffer solution (PN 905- 3501) Box of 6 16 oz (473 ml) bottles Green
905- 3505 pH 10 buffer solution 16 oz (473 ml) Blue
905- 582525 6 pack of pH 10 buffer solution (PN 905- 3505) Box of 6 16 oz (473 ml) bottles Blue
R508- 8oz ORP Standard, 475mV, 8oz 8 oz (237 ml) NA
Conductivity Standards
Emerson offers a variety of conductivity standards in both 16 oz (473 mL) and 1 gal (3.8 L) sizes. To determine the best standard to use, consult the analyzer product data sheet and choose a standard whose conductivity is in the recommended range for the cell constant.
Part Number Standard Solution Value Size Color Code
05010781899 (SS- 6) 200 μS/cm 16 oz (473 ml) Red
05000705464 (SS-1) 1409 μS/cm 16 oz (473 ml) Blue
05010782468 (SS-5) 1000 μS/cm 16 oz (473 ml) Yellow
05010782147 (SS-7) 5000 μS/cm 16 oz (473 ml) Violet
05010797875 (SS- 6A) 200 μS/cm 1 gallon (3.8 L) Red
05000709672 (SS-1A) 1409 μS/cm 1 gallon (3.8 L) Blue
05010783002 (SS-5A) 1000 μS/cm 1 gallon (3.8 L) Yellow
05010782026 (SS-7A) 5000 μS/cm 1 gallon (3.8 L) Violet
39
Page 40
CALIBRATION ACCESSORIES
Bench Top Calibrator
he bench top calibrator lets you take full advan-
T tage of the power of Emerson’s SMART pH sensors. Instead of carrying buffer solutions,
eakers, a waste container, and a wash bottle
b
into the plant to calibrate the sensor, just discon-
nect the SMART sensor and bring it to the
hop or laboratory. Connect the sensor to
s
the bench top calibrator and complete
the calibration. The results are auto-
matically stored in the sensor. When
the sensor is reconnected to the field analyzer, it transfers the stored calibra­tion data to the field instrument.
The calibrator has a pre-wired VP cable
and pre-wired spring-loaded push connec-
ors, so it can be used with both VP and integral cable
t sensors. Although the calibrator is most convenient for use with SMART sensors, it can also be used with non-
MART pH sensors. In the latter instance, the user notes
S the calibration data and enters it manually in the field instrument.
The bench top calibrator (PN S10119) consists of a dual channel Model 1056 pH analyzer—one channel connected to the VP cable, the other to the push connectors—mounted on a stand with openings in front to accommodate three 250 mL beakers for buffers and rinse water. Beakers are not provided.
pH sensors and buffers solutions are sold separately. See product data sheet PDS 71-BTC for more information.
Handheld pH Simulator
The handheld pH simulator PN S10104 (SQ7428) supplies switch-selectable pH 4, 7, and 10 inputs to an analyzer. Two other switches allow the user to simulate a cracked glass membrane and a coated reference junc­tion. The simulator is compatible with any Rosemount Analytical pH instrument. The unit is battery operated.
Consult the factory for more information.
pH and Conductivity Simulator Module with VP plug
Simulator modules, when connected to a VP cable, simulate either a pH or contacting conductivity sensor. They are used for checking the proper working of the cable and analyzer.
PN SQ8705 simulates pH 7.00, 25°C, 100 MΩ glass impedance, and 5 kΩ reference impedance. Use it with any VP pH cable having part number 24281-XX.
PN 23979-00 contains two modules for checking VP conductivity cables and conductivity analyzers. One module puts a 20 kΩ resistor across the electrode
terminals; the other puts a 100 kΩ resistor across the terminals. The expected conductivity or resistivity depends on the cell constant stored in the analyzer. Both modules simu­late 25°C. Use the module with any VP conductivity cable having part number 23747-XX
Consult the factory for more infor­mation.
40
Page 41
FLOW ACCESSORIES
Rosemount Analytical offers a variety of rotameters with integral valves to help measure and control flow in sidestream samples. A constant head flow controller is also available.
Rotameters – Sensor Recommendations
Sensor Model Flow Recommended
Rate* Valve Rotameter
All pH Sensor Models 2-5 g/hr (7.6 - 19 L/hr) PN 9390004 or PN 196-898754
499A CL 8-15 g/hr (30-57 L/hr) PN 196-898754
499A DO 2-5 g/hr (7.6-19 L/hr) PN 9390004
499A OZ 2-5 g/hr (7.6-19 L/hr) PN 9390004
When used with low flow cells (see page 9)
*
Rotameter Specifications
PN 9390004 PN 196-898754
Range of Flow Required 0.4 - 5.0 g/hr (1.5 - 19.0 L/hr) 2.0 - 20.0 g/hr (7.6 - 76.0 L/hr)
Wetted Parts Acrylic, 316 SST, Viton polycarbonate, 316 SST, brass, Buna N
Process Connection 1/4 inch FNPT (316 SST) 1/8 inch FNPT (brass)
Maximum Pressure 100 psig (790 kPa) 100 psig (790 kPa)
Maximum Temperature 130°F (54°C) 130°F (54°C)
41
Page 42
CONSTANT HEAD FLOW CONTROLLER
The constant head flow controller SQP-10077-LQD (SQ10957) uses gravity to control flow in sidestream samples, eliminating the need for pressure regulators, valves, and rotameters. The overflow sampler consists of two concentric tubes with the annular space closed at the bottom. The sample enters through a fitting at the bottom of the outside tube and fills the space between the tubes, eventually overflowing the inside tube and running to drain. A second fitting on the outside tube connects to a low flow cell, which holds the sensor. The outlet from the low flow cell joins the drain from the overflow sampler. The flow of sample to the sensor is controlled by the difference between the height of the overflow tube and the outlet from the low flow cell as well as by an orifice in the flow cell inlet. The flow controller provides a steady flow of 2 gph (126 mL/min)
Flow Controller Specifications
Inlet flow 3 – 80 gph (0.2 – 30 L/hr)
Inlet pressure 3 – 65 psig (122 – 549 kPa abs)*
Temperature 32 - 122°F (0 - 50°C)
*The minimum inlet pressure is required to open a check valve, which prevents the flow cell from draining if sample flow is lost. Removing the check valve lowers the inlet pressure requirement to a few feet of water head.
as long as the inlet flow is enough to cause the sample to overflow the center tube of the flow controller.
The flow controller is part of the FCL, FCLi, and MCL systems for measuring free chlorine and monochlo-
amine. The system shown can also be used with pH and
r ORP sensors. It is not recommended for ENDURANCE conductivity sensors because the orientation of the flow
ell does not allow the sensor to be completely
c submerged in the sample. The flow controller may not be suitable for ozone or oxygen sensors. Consult the factory for these applications.
The flow controller is mounted on a plastic back plate for ease of installation.
Figure 78. Example of a panel using a specially design constant head flow controller designed to control the flow to three flow cells. Panels with choice of one, two, or three flow cells and choice of one or two instruments (shown with the 56 are offered under SQ 11102. Consult factory for more information
Figure 77. Example of Overflow Sampler with Low flow cell PN 24091-00 (see page 9).
42
Page 43
TAGS FOR SENSORS AND INSTRUMENTS
It is common for most facilities to tag their sensors and instruments. It is especially critical in large facilities so that all instruments and sensors are correctly located. Rosemount Analytical makes it simple to tag all of your
Figure 79. PN 2001492 SST Tag
Used for sensors, commonly installed on sensor with wire Dimensions: 3.125"W x 1.31"H
SAMPLE COOLERS
new equipment before it gets installed into your facility. Tags are also useful for large projects and OEM users. Up to three lines of information can be added with 13 charac­ters per line including spaces.
Figure 80. PN 9240048-00 SST Tag
Used for most instruments Dimensions: 3.125"W x 1.31"H
Figure 81. PN 9241178-00 SST Tag
Used for 5081 and 6081 instruments only Dimensions: 2.56"W x 1.28"H
Rosemount Analytical offers a pre-packaged sample cooler suitable for cooling high temperature and pres­sure liquid samples, such as boiler and boiler feedwater. The cooler is not recommended for steam. The cooler consists of a sediment trap to remove suspended solids, a coil-in-shell heat exchanger, a needle valve for control-
Figure 82. Sample Cooler example
ling flow and pressure, a pressure relief valve set to 75 psig (618 kPa abs), and a bimetal thermometer for local display of temperature. The entire system is mounted on a strut channel frame. The system is available in two versions: the sample cooler alone (Model BSA), or the sample cooler with a Model 141-06 conductivity sensor (Model 99-14).
Uncooled sample
Temperature: 600 F (315C) max Pressure: 1500 psig (10,440 Kpa abs)
Cooled sample
Operating Temperature Range: 32 F to 203 F (0C to 95C) Preferred: 70F to 90 F (21 C to 32 C)
Cooling Water
Temperature: 122F (50C) maximum Pressure: 200 psig (1480 Kpa abs) maximum
Contact factory for more information.
43
Page 44
SPECIAL PARTS AND ASSEMBLIES
Need a small modification to a standard product? Want a small sampling system but don’t have the facilities or personnel to design and build it yourself? Whether your requirement is as simple as increasing the insertion length of a retractable sensor or building a special sample panel, our Special Request Department can help.
Figure 82 shows an example of a sample cooling system
uilt to customer specifications. It includes a sample
b cooler, throttle valve, bimetal thermometer, stainless steel flow cell, and pressure relief valve. The entire sample system is mounted on a stainless steel back plate along with a Model 5081 transmitter.
Figure 83 shows a stainless steel enclosure with air purge. A 5081 transmitter and flow cell is visible inside the enclosure.
Figure 84 is an example of a customer specified special system with an air purge system located on the outside of the Hoffman enclosure. The 5081 instrument, 499DO sensor, and rotameter are mounted in the enclosure and was build under the special request part number SQ10645E1 / SQP 10081-LQD. Similar systems can be specified to mount multiple sensors and instruments, such as that found in Figure 85.
Examples of other special requests include sensor material substitutions, modifications of a mounting parts, and cable length adjustments. Some of special requests are found throughout this catalog and are designated with a part number starting with SQP or SQ. Other types of special requests can be accommodated upon request. Consult the factory for more information about your special needs.
Figure 82. Customer specified system including a sample cooling system with a 5081 instrument and a conductivity sensor was created as PN SQ 8413B.
Figure 84. Stainless steel Hoffman enclosure with 5081 instrument mounted internally. The process inlet is
controlled with a flow meter with regulator valve. The liquid flows into a plastic low flow cell and pass the 499ADO dissolved oxygen sensor. Systems can be special ordered to have instruments and sensors mounted in a similar manner. Special order part number for system shown is SQ10645E1 / SQP 10081-LQD. Contact factory for details.
44
Figure 83. 5081 and sensor mounted inside customer specified Hoffman enclosure.
Page 45
SPECIAL PARTS AND ASSEMBLIES
Figure 85. Stainless steel Hoffman enclosure with 3 5081 instruments mounted internally. A corrosion protection air purge system is mounted
on the outside of the enclosure. A flow meter with regulator valve controls the flow to the 400VP conductivity, 499ADO dissolved oxygen, and 389 pH sensors. All three sensors are mounted in low flow cells. Customized systems for wall or panel mount can be special ordered similarly, having instruments and sensors mounted and ready to connect to 1/4 inch tubing. This system was designed per customer specifications – special order part number is SQ10645F1 / SQP 10082-LDQ. Contact factory for details.
Figure 86. Clarity II turbidity sampling system with Z purge heated enclosure for hazardous areas. One of many examples that is offered
through our special request system. This system was designed per customer specifications – special order part number is SQ 10341 / SQP 10030-LQD. Contact factory for details.
Figure 87. Multiple 1056 instruments with HART communication, shown mounted on a panel for simple Plug-N-Plumb installation. Four sensors (conductivity, pH,
dissolved oxygen, and turbidity) are mounted with a flow regulator. The system shown was designed on a stand-alone panel for table-top mounting but can be purchased as a panel mount version.
45
Page 46
CERTIFICATES FOR SENSORS, INSTRUMENTS, AND CHEMICALS
Level 1 Code 99EC 0506-24 9002 ,3 .guA :etaD .veR
Rosemount Analytical Inc.
2
400 Barranca Parkway
Irvine, CA 92606-5018 USA
T
el 1(800) 854-8257
F
ax 1(949) 863-0171
Fax 1 (949) 757-8520 (QA Office)
www.emersonprocess.com
Certificate of Calibration
TO:
DATE:
R
CM or SO #
P
O #:
Line Item
Model Number/ Customer
Part Number
Range K= Serial Number Tag Number
1 5081-C-HT-20-60-99CC 0-20mS 1.0 A10-1220470-017 40-AT-402
6 3200HP-01-99CC Refer. Manual N/A A10-1218201-008 40-AT-401
7 5081-P-HT-21-69-99CC 0-14pH±1400mV N/A A10-1226016-001 N/A
8 400-11-60-99CC 0-50µS 0.01/cm J09-698212 N/A
This certifies tha t the furnished materials have been manufactured and tested in accordance with all a
pplicable product specifications under compliance of our Quality System. Rosemount Analytical is certified
t
o ISO 9001:2008 by Det Norske Veritas Certification # 09242-2006-AQ-HOU-ANAB
a
nd complies with Standard CSA Z299.3. Our calibration system is maintained in accordance with all requirements of ANSI/NCSL Z540-1-1994, American National Standard for Calibration System Requirements, and is traceable to the National Institute of Standards and Technology (NIST).
All products, including packaging, are free of Mercury and are certified to contain no hazardous materials or wood products. Records of inspection, testing and other required supporting documentation are on file and available for your examination upon request.
QA INSPECTOR
eltiT evitatneserpeR ynapmoC AQ -UEIT .K
Level 3 Code 99LC 42-6048
Rev. Date: Aug. 3, 2009
R
osemount Analytical Inc.
2400 Barranca Pkwy. Irvine, CA 92606-5018 Tel 1(800) 854-8257 Fax 1(949) 863-0171 Fax 1 (949) 757-8520 (QA Office)
www. emersonprocess.com
Certificate of Loop Calibration
To: Date:
SO# or RMA# P.O.#
T
his certifies that the furnished materials have been manufactured and tested in accordance with all
a
pplicable product specifications under compliance of our Quality System. Rosemount Analytical is certified
to ISO 9001:2008 by Det Norske Veritas Certification # 09242-2006-AQ-HOU-ANAB
and complies with Standard CSA Z299.3. Our calibration system is maintained in accordance with all requirements of A
NSI/NCSL Z540-1-1994, American National Standard for Calibration System Requirements, and is
t
raceable to the National Institute of Standards and Technology (NIST).
A
ll products, including packaging, are free of Mercury and are certified to contain no hazardous materials or wood products. Records of inspection, testing and other required supporting documentation are on file and available for your examination upon request.
Units Under Test (UUT)
Calibrated Range: 0-100mS Verified Range:
N/A
L
ine
Item
I
nstrument/Sensor
Model Number
C
ustomer
Part Number
S
erial
Number
T
ag
Number
C
ell
Constant
1
1056-02-21-38-AN-99LC N/A J09-694219 N/A N/A
2
225-07-56-99CB-99LC-99Q8 N/A J09-694079 N/A 3.3000
S
tandards (STD):
S
tandards
Asset No. /
NIST Report No.
Lot
Number
Calibration
Date
Calibration
Due/Expiration
Date
KCL ( 650001 ) N/A G30422 N/A N/A
01-70-20 90-70-80 A/N 2082 02-CR
THERMOMETER 1808/48380-2 N/A 02-27-09 02-27-10
01-90-10 90-90-90 A/N 7401 53Y05H
Test Results:
STD
Value
UUT
Reading Tolerance Deviation
In/Out of Tol.
STD Temp.
Reading
UUT
Tolerance
UUT Temp
Reading
111.624mS 111.700mS ±1% .07 % IN 21.5 °C ± 1°C 21.5 °C ±1% % °C ± 1°C °C
After C
alibration
±1% % °C ± 1°C °C
pH Sensor Only
SLOPE =
mV/pH @ 25 °C OFFSET= mV
Note: UUT and RTD must be within tolerance specified of the Standard Value
Sr. Test Technician
eltiT evitatneserpeR ynapmoC - serolF .J
CERTIFICATES and CALIBRATIONS: All certificates state our compliance with our quality system which is in accordance with ISO-9001, ANSI/NCSL Z540-1-1994, and traceability to NIST.
EVEL 1: Analyzer or Sensor Certificate of Calibration
L
(NO actual test data provided): Code 99CC. Calibration certificate at standard factory ranges only. Identifies the
odel, Serial and/or Tag numbers. Does not include
M actual test data. Can be provided after the fact. Certificates states compliance to our quality system, as noted above.
LEVEL 2: Electronic Calibration Test Data Report, Code 99EC or Electronic Calibration Tag, Code 99ET: Calibration certificate for specific, customer specified ranges. Identifies the Model, Serial and/or Tag numbers, applicable test standards, and NIST report numbers. This certificate includes the actual test data recorded during the electronic calibration. It also states that instrumentation on the specified order is compliant with our quality system.
LEVEL 3: Loop Calibration (with Test Data Report ) Code 99LC with one or two sensors. A custom calibration with validation point within Cell Constant range is also available. A loop calibration is where the analyzer is cali­brated with its mating sensor(s). The loop is calibrated against our NIST traceable certified reference instru­ments, at our standard or customer specified range. The calibration certificate identifies the Model, Serial and/or Tag numbers, test standards, and NIST report numbers. In addition, this certificate includes the actual test data recorded during the calibration, therfore this certificate can not be provided after shipment has been made. It also states that equipment on the specified order is compliant with our quality system.
OTHER CERTIFICATES
1. Hazardous Area Certificates: (FM, CSA, Baseefa) Specify requirements with placement of order.
2. NAFTA Certificates of Origin with or without Chamber of Commerce stamp.
3. Material Certificate: for various Sensors and calibra­tion chemicals.
46
Page 47
INDEX
Part Number Description Page
2001492 Tag 43
2001990 Adapter 14
002011 Tee 7
2
2002526 Junction box 35
9200245 Extension Cable 32
9200273 Extension Cable 33, 37
9200275 Extension Cable 31, 37
9200276 Extension Cable 37
9200334 Extension Cable 37
9210012 pH Buffer 39
9210013 pH Buffer 39
9210014 pH Buffer 39
9310102 Nipple 22
9320057 Adapter 15
9390004 Rotameter 41
9510066 Process Connector 17
05000705464 Conductivity Standard 39
05000709672 Conductivity Standard 39
05010781899 Conductivity Standard 39
05010782026 Conductivity Standard 39
05010782147 Conductivity Standard 39
05010782468 Conductivity Standard 39
05010783002 Conductivity Standard 39
05010797875 Conductivity Standard 39
11275-01 Handrail Mounting Assy 10
12707-00 Jet Spray Cleaner 13
196-898754 Rotameter 41
23166-00 Process Connector 17
23166-01 Process Connector 17
23240-00 Ball Valve Kit 20
23242-02 Adapter 14
23242-03 Adapter 14
23294-04 Extension Cable 32, 37
23294-05 Extension Cable 37
23311-00 Retraction Assembly 24
23311-01 Retraction Assembly 24
23550-00 Junction box 31-33, 34
23554-00 Cable Gland Kit 37
23555-00 Junction box 33, 34
23567-00 Tee 7
23646-01 Extension Cable 33, 37
23709-00 Junction box 35
23747-00 Extension Cable 37
23747-XX VP6 Cable 31, 32, 36
23765-00 Ball Valve Kit 20
23796-00 Retraction Kit 20
Part Number Description Page
23820-00 Instrument Mounting Kit 29
23820-01 Instrument Mounting Kit 29
3909-00
2
23979-00 Conductivity Simulator 40
24035-00 Rotameter 41
24091-00 Low Flow Cell 9
24091-01 Low Flow Cell 9
24091-02 Low Flow Cell 9
24281-XX VP8 Cable 36
24287-XX VP6 Cable 36
24289-00 Extension Cable 37
905-3501 pH Buffer 39
905-3505 pH Buffer 39
905-3506 pH Buffer 39
905-582517 pH Buffer 39
905-582521 pH Buffer 39
905-582525 pH Buffer 39
915240-03 Tee 7
915240-04 Tee 7
915240-05 Tee 7
9240048-00 Tag 43
9241178-00 Tag 43
99-14 Sample Cooler 43
BB11 Floating Ball System 12
Calibration Certificate Certificates 46
CVU Conductivity Validation Unit 38
Electronic Calibration Cert. Certificates 46
HRMS-01 Handrail Mounting System 10
HRMS-02 Handrail Mounting System 10
Loop Calibration Certificate Certificates 46
Material Certificate Certificates 46
Pasve Retraction Assembly 26
R508-8oz ORP Standard 39
S10021 (SQ 8940) Triclamp 16
S10104 (SQ 7428) pH Simulator 40
S10119 Bench Top Calibrator 40
S10240 (SQ 7716) Low Flow Cell 9
S10283-LQD Tee Adapter 7
S10290 (SQ 7637) Low Flow Cell 9
SQ 10904 Triclamp 16
SQ 8705 pH Simulator 40
SQP 10081-LQD Special Assembly 44
SQP-10077-LQD (SQ 10957) Overflow Sampler 42
SQP-10098-LQD (formerly PN 23513-00)
xtension Cable
E
Triclamp 16, 17
2, 37
3
47
Page 48
COMPLETE PROCESS ANALYTICAL SOLUTIONS
GAS CHROMATOGRAPH CENTER OF EXCELLENCE
Emerson Process Management Rosemount Analytical 10241 West Little York, Suite 200 Houston TX 77040 USA T 1.713.396.8700 T 866.422.3683 F 1.713.466.8175 gc.csc@emerson.com
LATIN AMERICA
Emerson Process Management Rosemount Analytical 10241 West Little York, Suite #200 Houston, TX 77040 USA T 713.467.6000 F 713.827.3328
GAS AND COMBUSTION CENTER OF EXCELLENCE
Emerson Process Management Rosemount Analytical 6565 P Davis Industrial Parkway Solon, OH 44139 USA T 1.440.914.1261 T 800.433.6076 F 1.440.914.1262 gas.csc@emerson.com
EUROPE
Emerson Process Management AG Blegistrasse 21 CH-6341 Baar-Walterswil Switzerland T 41.41.768.6111 T 41.41.761.8740
RosemountAnalytical.com
Online ordering available.
ANALYTICAL HEADQUARTERS
Emerson Process Management Rosemount Analytical 2400 Barranca Parkway Irvine, CA 92606 USA T 1.949.757.8500 T 800.854.8257 F 1.949.474.7250 liquid.csc@emerson.com
MIDDLE EAST AND AFRICA
Emerson Process Management EPM Building P. O. Box 17033 Jebe Ali Free Zone Dubai, United Arab Emirates T 971.4.8835235 F 971.4.8835312
FLAME AND GAS DETECTION
Emerson Process Management Net Safety Monitoring 2721 Hopewell Place N.E. Calgary, AB T1Y 7J7 Canada T 1.403.219.0688 T 866.347.3427 F 1.403.219.0694 safety.csc@emerson.com
ASIA- PACIFIC
Emerson Process Management Asia Pacific Private Ltd. 1 Pandan Crescent Singapore 0512 Republic of Singapore Phone: +65.777.8211 Fax: +65.777.0947
Emerson Process Management, Rosemount Analytical, PlantWeb, TUpH and ENDURANCE, are marks of Emerson Process Management group of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for informational purposes only, and while every effort has been made to ensure its accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modif y or improve the designs or specifications of such products at any time without notice.
© Rosemount Analytical Inc. 2012. All rights reserved. Printed in the U.S.A.
91-6020/rev. C
Loading...