Rosemount Analytical 3800 Operating Manual

Page 1
Instruction Manual
IQ-MAN-3800
L
Rev. D
Rosemount
pH Sensors
3800
Page 2
asgkas
h
Page 3
Essential Instructions
Read this page before proceeding!
Emerson designs, manufactures and tests its products to meet many national and international stan-
dards. Because these sensors are sophisticated technical products, you MUST properly install, use,
nd maintain them to ensure they continue to operate within their normal specifications. The
a following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life; personal injury; property damage; damage to this sensor; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Third-party substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
CAUTION
Sensor/Process Application Compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions. Application compatibility is entirely the responsibility of the user.
WARNING
Do not exceed temperature and pressure limitations of 284 °F (140 °C) and 72 psig (600 kPa, 6 bar).
WARNING
Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level.
CAUTION
Internal electrolyte fill solution may cause skin or eye irritation.
Page 4
CAUTION
Special Conditions for Safe Use
1. All pH sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the
danger due to a build up of an electrostatic charge.
. All pH sensor models are intended to be in contact with the process fluid and may not meet the 500V
2 r.m.s. a.c. test to earth.
This must be taken into consideration at installation.
About This Document
This manual contains instructions for installation and operation of the Rosemount 3800 pH Sensors.
The following list provides concerning all revisions of this document.
Rev. Level Date Notes
D 04/17 Reformatted to reflect the latest Emerson documentation style
Wiring Diagrams, and EC Declaration of conformity.
Page 5
Instruction Manual Table of Contents
LIQ-MAN-3800 April 2017
Contents
Section 1: Specifications
1.1 Specifications ......................................................................................................1
1.2 Product Certifications ..........................................................................................2
Section 2: Installation
2.1 Unpacking and Inspection ...................................................................................3
2.2 Storage................................................................................................................3
2.3 Electrode Preparation ..........................................................................................3
Section 3: Wiring
3.1 Wiring for Rosemount 3800.................................................................................5
Section 4: Calibration
4.1 Two Point Buffer Calibration.................................................................................7
4.2 Recommended pH Sensor Standardization ..........................................................7
Section 5: Maintenance
5.1 Maintenance........................................................................................................9
Section 6: Accessories
6.1 Accessories........................................................................................................11
EC Declaration of Conformity.............................................................................13
Intrisicallly Safe Sensor Installation Drawing - FM ................................15
Table of Contents i
Page 6
Table of Contents Instruction Manual
April 2017 LIQ-MAN-3800
ii Table of Contents
Page 7
Instruction Manual Specifications
LIQ-MAN-3800 April 2017
Section 1: Specifications
1.1 Specifications
Table 1-1: Rosemount 3800 sensor specifications
Range
0 to 14 pH
Temperature range
up to 221 °F (steam up to 284 °F) [up to 105 °C (steam up to 140 °C)]
Temperature sensor
Pt 100
Wetted materials
Glass, ceramic and EPDM O-ring USP VI
Reference junction
Single ceramic junction
Dimensions
12 mm OD, shaft length -01= 120 mm, -02 = 225 mm , -03 = 325 mm
Electrolyte
Gelled polymer
Cable connector
Coaxial S8
Process connector
Pg 13.5 connector
Remarks
Suitable for all Rosemount pH-instruments. Quality certification includes list of wetted materials and calibration records
Specifications 1
Page 8
Specifications Instruction Manual
April 2017 LIQ-MAN-3800
1.2 Product Certifications
Please see online certificates for further details. IECEx
Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Per standards IEC60079-0 : 2011, IEC 60079-11 : 2011
ATEX
II 1 G Ex ia IIC T4 Ga (-20˚C ≤ Ta ≤ +60 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
FM
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C Hazardous (Classified) Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60°C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60°C - 1400332; Entity; NI/I/2/ABCD/T6 Ta = 60°C; S/II,III/2/EFG/T6 Ta = 60°C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; Ambient temperature rating -20°C to +60°C: (Simple Apparatus)
Per standards C22.2 No. 142 – M1987, C22.2 No 157 – M1992, CAN/CSA E60079-0:07, CAN/CSA E60079-11:02, UL 50:11th Ed., UL 508:17th Ed., UL 913: 7th Ed., UL 60079-0: 2005, UL 60079-11: 2002
2 Specifications
Page 9
Instruction Manual Installation
LIQ-MAN-3800 April 2017
Section 2: Installation
2.1 Unpacking and Inspection
Inspect the outside of the carton for any damage. If damage is detected, contact the carrier immediately. Inspect the instrument and hardware. Make sure all items in the packing list are present and in good condition. Notify the factory if any part is missing.
2.2 Storage
1. It is recommended that electrodes be stored in their original shipping containers until needed.
2. Do not store at temperatures below 23 °F (-5 °C).
3. Electrodes should be stored with a protective cap containing KCl solution (PN 9210342).
4. For overnight storage, immerse the sensor in tap water or 4 pH buffer solution.
5. A pH glass electrode does have a limited shelf life of one year.
2.3 Electrode Preparation
1. Remove electrode from shipping container.
2. Remove the protective boot covering the electrode bulb.
3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there.
Note: Do not allow lubricant to coat electrode bulb or reference junction. If it does, wipe it clean
before installation.
Installation 3
Page 10
Installation Instruction Manual
April 2017 LIQ-MAN-3800
Figure 2-1: Rosemount 3800 dimensional drawing
Figure 2-2: Sensor orientation
4 Installation
Page 11
Instruction Manual Wiring
LIQ-MAN-3800 April 2017
Section 3: Wiring
3.1 Wiring for Rosemount 3800
For other wiring diagrams not shown below, please refer to the Liquid Transmitter Wiring Diagrams.
Figure 3-1: Rosemount 3800 sensor wiring to Rosemount 1056/56/1057 transmitters
Wiring 5
Page 12
Wiring Instruction Manual
April 2017 LIQ-MAN-3800
Figure 3-2: Rosemount 3800 sensor wiring to Rosemount 1066 transmitter
Figure 3-3: Rosemount 3800 sensor wiring to Rosemount 5081 transmitter
6 Wiring
Page 13
Instruction Manual Calibration
LIQ-MAN-3800 April 2017
Section 4: Calibration
4.1 Two Point Buffer Calibration
elect two stable buffer solutions, preferably pH 4.0 and 7.0 (pH buffers other than pH 4.0 and
S pH 7.0 can be used as long as the pH values are at least two pH units apart).
Note: A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approximately
±59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by transmitter.
2. Once the first buffer has been acknowledged by the transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over time the sensor will age, both in the process and in storage, and will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mV/pH.
4.2 Recommended pH Sensor Standardization
For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
Calibration 7
Page 14
Calibration Instruction Manual
April 2017 LIQ-MAN-3800
8 Calibration
Page 15
Instruction Manual Maintenance
LIQ-MAN-3800 April 2017
Section 5: Maintenance
5.1 Maintenance
Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60 minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% @ 25 °C (77 °F), then it will be approximately 25% @ 80 °C (176 °F), and approximately 5% @ 120 °C (248 °F).
Maintenance 9
Page 16
Maintenance Instruction Manual
April 2017 LIQ-MAN-3800
10 Maintenance
Page 17
Instruction Manual Accessories
LIQ-MAN-3800 April 2017
Section 6: Accessories
6.1 Accessories
Table 6-1: Accessories for Rosemount 3800
Part Number Description
200339
9
160478
9
24167-03 ¾ in. NPT 1 in. x Pg 13.5 insertion adapter
9999SQ10057 1 in. NPT x Pg 13.5 insertion adapter
24166-00 EDPM O-Ring
8 cable, free rotating connector, single pole 10m long
S
1 ¼ in. x Pg 13.5 insertion adapter
G
Accessories 11
Page 18
Maintenance Instruction Manual
April 2017 LIQ-MAN-3800
12 Accessories
Page 19
Instruction Manual EC Declaration of Conformity
LIQ-MAN-3800 April 2017
EC Declaration of Conformity
Note: Please see website for most recent Declaration.
EC Declaration of Conformity 13
Page 20
EC Declaration of Conformity Instruction Manual
April 2017 LIQ-MAN-3800
14 EC Declaration of Conformity
Page 21
ANY FM APPROVED
ASSOCIATED APPARATUS
HAVING ENTITY PARAMETERS
NO
N-
H
AZARDOUS
(UNCLASSIFIED)
AREA
CLASS I, II, III, DIVISION 1, GROUPS A-G
T6 Ta = 60°C
SENSOR
ENTITY PARAMETERS
Ui = 13.1U, Ii = 358mA
Pi = 698 mW
Ci = 0.967 µF, Li = 0.1mH
1. NO REVISION TO THIS DRAWING IS PERMITTED WI THOUT
FM APPROVAL.
2. U
max
> U
t
; I
max
> I
t
; (C
i
OF ALL LOOPS + C CABLE) < C
a
;
(L
i
OF ALL LOOPS + L CABLE) < L
a
, P
max
OR P
i
> P
0.
3. SINGLE MULTI-CHANNEL
IS BARRIER OR APPARATUS MUST BE FM APPROVED,
4. SINGLE MULTI-CHANNEL
IS BARRIER OR APPARATUS MANUFACTURE'S
CONTROL DRAWI NGS MUST BE FOLLOWED WHEN INSTALLING THE SYSTEM.
IS BARRIER OR EQUIPMENT MAY BE INSTALLED WITHIN THE HAZARDOUS
(CLASSIFIED) LOCATION FOR WHICH IT IS APPROVED
.
5. INSTALLATION MUST BE IN ACCORDANCE WITH ARTICL
E 500 OF THE NEC
(ANSI/NFPA 70) AND ANSI/ISA RP 12.6.
WARNING: SUBSTITUTION OF COMPONENTS MAY IMPA
IR INTRINSIC SAFETY.
6. pH & AM PEROMETRIC SENSORS WITHOUT PREAMPS ARE SIMPLE APPARATUS.
ZON E
ZONE 0
7. CONTROL EQUIPMENT CONNECTED TO THE ASSOCIATED APPARATUS MUST
NOT USE OR GENERATE MORE THAN 250V.
RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND
MUST BE LESS THAN OR EQUAL TO 1 OHM.
ANY
FM APPROVED TRANSMI TTER FOR
DIVISION 1 WITH INTRINSICALLY SAFE
OUTPUT PARAMETERS. THIS
FM
APPROVED DEVICE MUST BE INSTALLED
PER ITS INSTALLATION DRAWING.
FM
APPROVED EQUIPMENT (MAY BE
MULTIPLE DEVICES, NUMBER IS LIMITED
BY REQUIREMENTS TO MEET ALL OTHER
IS REQUIREMENTS FOR THE NETWORK)
WITH EQUIVALENT HAZARDOUS AREA
APPROVALS.
H
AZA RDOUS
(CL ASSI
F
IED)
AREA
INTRISICALLY SA FE
9. pH/ORP SENSOR MODELS THAT MAY CONTAI N THE PREAMPLIFIER:
3900/VP
3500/VP
3300HT/VP
3400HT/VP
396/VP
396R/VP
396P/VP
398R/VP
399/VP
389/VP
385/385+
10. WARNING: TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE
ATMOSPHERES, DISCONNECT POWER BEFORE SERVIC
ING.
11. THE ENTITY CONCEPT ALLOWS INTECONNECTION OF INTRINSICALLY SAFE
APPARATUS WITH ASSOCIATED APPARATUS WHEN THE FOLLOWING IS TRUE:
U
i
> U
o;
I
i
> I
o;
P
i
> P
o;
C
o
> C
i
+ C CABLE; L
o
> L
i
+ L CABLE.
12. COPY REVISIONS TO 1400332 TO pH/ORP SHIPPING MANUALS.
13 Ci INCLUDES THE CAPACITANCE OF 500 FEET OF SENSOR CABLE.
13
DWG NO.
1400332
8.
Instruction Manual FM Installation
LIQ-MAN-3800 April 2017
Intrisicallly Safe Sensor Installation Drawing - FM
FM Installation 15
Page 22
LIQ-MAN-3800
Rev. D
April 2017
www.Emerson.com/RosemountLiquidAnalysis
Youtube.com/user/Rosemount
Twitter.com/Rosemount_News
Emerson
8200 Market Blvd. Chanhassen, MN 55317, USA Tel +1 800 999 9307 Fax +1 952 949 7001
Liquid.CSC@Emerson.com
Analyticexpert.com
facebook.com/Rosemount
©2017 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.