INSTALLATION
For sensor orientation, see Figure 2.
For wiring, see Figures 3 and 4.
TWO POINT BUFFER CALIBRATION
Select two stable buffer solutions, preferably pH 4.0 and 10.0
(pH buffers other than pH 4.0 and pH 10.0 can be used as
long as the pH values are at least two pH units apart).
NOTE
A pH 7 buffer solution reads a mV value of approx.
zero, and pH buffers read approximately ± 59.1 mV
for each pH unit above or below pH 7. Check the pH
buffer manufacturer specifications for millivolt values
at various temperatures since it may affect the actual
value of the buffer solution mV/pH value.
1. Immerse sensor in the first buffer solution. Allow sensor to
equilibrate to the buffer temperature (to avoid errors due
to temperature differences between the buffer solution
and sensor temperature) and wait for reading to stabilize.
Value of buffer can now be acknowledged by
analyzer/transmitter.
2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor
with distilled or deionized water.
3. Repeat steps 1 and 2 using the second buffer solution.
4. The theoretical slope value, according to the Nernst equation for calculating pH, is approximately 59.17 mV/pH. Over
time the sensor will age, both in the process and in storage,
and will result in reduced slope values. To ensure accurate
readings, it is recommended that the electrode be replaced
when the slope value falls below 47 to 49 mV/pH.
RECOMMENDED pH SENSOR
STANDARDIZATION
For maximum accuracy, the sensor can be standardized online or with a process grab sample after a buffer calibration
has been performed and the sensor has been conditioned to
the process. Standardization accounts for the sensor junction
potential and other interferences. Standardization will not
change the sensor’s slope but will simply adjust the analyzer’s reading to match that of the known process pH.
MAINTENANCE
Electrodes should respond rapidly. Sluggishness, offsets,
and erratic readings are indicators that the electrodes may
need cleaning or replacement.
1. To remove oil deposit, clean the electrode with a mild
non-abrasive detergent.
2. To remove scale deposits, soak electrodes for 30 to 60
minutes in a 5% hydrochloric acid solution.
3. Temperature effect on life expectancy: If glass electrode
life expectancy is 100% @ 25°C (77°F), then it will be
approximately 25% @ 80°C (176°F), and approximately
5% @ 120°C (248°F).
FIGURE 4. Model 372-02 (without preamp)FIGURE 3. Wiring Model 372-01 (with preamp)
Emerson Process Management
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2011
ORDERING INFORMATION
Code Cable (Required Selection)
01 25 ft cable with integral preamp for Models 54e, 1055, 3081, 4081, and 5081
02 15 ft cable for remote preamp (Models 54e, 1055, 3081, 4081, and 5081)
372 -01 EXAMPLE
MODEL
372 pH SENSOR
FIGURE 2. Sensor Orientation