Product Data Sheet
Rosemount™ 3300HT/3300HT VP,
3400HT/3400HTVP, and 3500/3500VP
PERpH-X™ High Performance pH/ORP Sensors
May 2017
LIQ-PDS-PERpH-X
A sensor design that is flexible to meet your demanding pH/ORP application needs
Rosemount 3500 PERpH-X sensors are versatile sensors suited to meet a number of difficult application requirements. These high
performance sensors are ideal for use in high temperature and high pressure processes and feature a wide array of options to solve
your pH/ORP measurement needs.
Rosemount PERpH-X Sensors
Overview
May 2017
Expanded Application Flexibility
Selectable reference electrolytes allow sensors to be used in a
large number of applications.
Electrolyte options include:
• High Temperature – most suitable for use in highly acidic, basic or
oxidative solutions, and high temperature applications.
• Bio-Film Resistant – inhibits the growth of bacteria and algae
• Poisoning Resistant – most suitable for use in processes containing
sulfides, mercaptans, and cyanides
• Oil Resistant – aimed for use in applications where light oils and
greases can foul a sensor
• Scaling Resistant – targets applications where the precipitation of
calcium magnesium salts like gypsum or water hardness coat over
the electrode
• Metal Resistant – most suitable for use in applications where
chloride in the reference electrolyte would react with the process
Minimize Maintenance and Lower Total Cost of
Ownership
A long lasting rebuildable reference junction and electrolyte
extends sensor life and allows sensors to be easily rebuilt
where normally other sensors would have to be replaced.
SMART preamplifier allows automatic recognition of pH
sensors by Rosemount transmitters: 1066, 1057, 1056,
and 56.
pH calibration data is stored which allows sensors to be
calibrated in advance for “Plug and Play” installations in the
field.
Versatile Mounting Options to Meet Your
Installation Requirements
Ryton body sensors feature 1 in. MNPT front and rear facing
process threads.
Titanium body sensors offered in 21 in. and 36 in. lengths
suitable for use with retraction assemblies.
Variopol (VP8) cable connection option, for quick
cable-to-sensor release, eliminates cable twisting.
Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2
Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Dimensional and Installation Drawings . . . . . . . . . . . . . 13
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
www.Emerson.com/RosemountLiquidAnalysis
May 2017
Rosemount 3500P pH/ORP Sensor
Ordering Information
Rosemount PERpH-X High Performance pH/ORP Sensors feature various sensor body
options. Sensors are available in a chemically resistant Ryton plastic body
(3500P/3500VP) or titanium tube housing (3300HT/3300HTVP/3400HT/3400HTVP).
Rosemount 3500P/3500VP sensors are constructed with front and rear facing 1 in.
MNPT threads for insertion or submersion type installations. Rosemount
3300HT/3300HTVP/3400HT/3400HTVP) sensors must be used with a process
connector allowing for various insertion depths. These titanium body sensors also have
extended sensor length options allowing installation through a ball valve assembly.
Sensors feature a Teflon reference junction and a Pt-100 RTD for temperature
compensation. Rosemount PERpH-X pH/ORP sensors are available with either an
integral cable connection or Variopol (VP8) connector. Variopol cables sold separately.
Accessories ).
(see
Table 1. Rosemount 3300HT pH/ORP Sensor ordering information
Model Sensor type
3300HT pH/ORP sensor
Rosemount PERpH-X Sensors
Measuring electrode
10 pH - GPHT glass
12 ORP
O-ring material
30 EPDM
31 Viton
32 Kalrez
Typical model number: 3300HT-10-30
www.Emerson.com/RosemountLiquidAnalysis
3
Rosemount PERpH-X Sensors
Table 2. Rosemount 3300HTVP pH/ORP Sensor ordering information
Model Sensor type
3300HTVP pH/ORP sensor
Measuring electrode
10 pH - GPHT glass
12
O-ring material
30 EPDM
31 Viton
32 Kalrez
Preamplifier option
_ No preamplifier
70 SMART preamplifier
Typical model number: 3300HTVP-10-30-70
ORP
(1)
May 2017
1. Only available if selected with option 10.
4
www.Emerson.com/RosemountLiquidAnalysis
May 2017
Table 3. Rosemount 3400HT pH/ORP Sensor ordering information
Model Sensor type
3400HT pH/ORP sensor
Measuring electrode
10 pH - GPHT glass
12
ORP
Sensor length
21 21 in. Titanium tube
25 36 in. Titanium tube
O-ring material
30 EPDM
31 Viton
32 Kalrez
Cable length
61 9.5 in. cable without BNC
62 15 ft. cable without BNC
(1)
(2)
Rosemount PERpH-X Sensors
Typical model number: 3400HT-10-21-30-62
1. For use with sensor head junction boxes.
2. For wiring directly to transmitter or junction box.
www.Emerson.com/RosemountLiquidAnalysis
5
Rosemount PERpH-X Sensors
Table 4. Rosemount 3400HTVP pH/ORP Sensor ordering information
Model Sensor type
3400HTVP pH/ORP sensor
Measuring electrode
10 pH - GPHT glass
12
Sensor length
21 21 in. Titanium tube
25 36 in. Titanium tube
O-ring material
30 EPDM
31 Viton
32 Kalrez
Preamplifier option
ORP
May 2017
_ No preamplifier
70 SMART Preamplifier
(1)
Typical model number: 3400HTVP-10-21-30-70
1. Only available if selected with option 10.
6
www.Emerson.com/RosemountLiquidAnalysis
May 2017
Table 5. Rosemount 3500P pH/ORP Sensor ordering information
Model Sensor type
3500P pH/ORP sensor
Electrolyte selection
BF Bio-film resistant
HT
MR
High temperature
Metal resistant
Rosemount PERpH-X Sensors
OR
PR
SR
Oil resistant
Poisoning resistant
Scaling resistant
Preamplifier/cable
01 Preamplifier with 25 ft. cable
02 No preamplifier with 15 ft. cable
(1)
Measuring electrode type
10 pH – GPHT glass
12 ORP
Reference type
21 Double junction reference
O-ring material
30 EPDM
31 Viton
32 Kalrez
Typical model number: 3500P-HT-01-10-21-30
1. Preamplifier is SMART if selected with option 10. Preamplifier is standard if selected with option 12.
www.Emerson.com/RosemountLiquidAnalysis
7
Rosemount PERpH-X Sensors
Table 6. Rosemount 3500VP pH/ORP Sensor ordering information
Model Sensor type
3500VP pH/ORP sensor
Electrolyte selection
BF Bio-film resistant
HT
High temperature
May 2017
MR
OR
PR
SR
Metal resistant
Oil resistant
Poisoning resistant
Scaling resistant
Preamplifier/cable
01 Preamplifier
02 No Preamplifier
(1)
Measuring electrode type
10 pH–GPHT glass
12 ORP
Reference type
21 Double junction reference
O-ring material
30 EPDM
31 Viton
32 Kalrez
Typical model number: 3500VP-HT-01-10-21-30
1. Preamplifier is SMART if selected with option 10. Preamplifier is standard if selected with option 12.
8
www.Emerson.com/RosemountLiquidAnalysis