Rosemount™ 3300HT/3300HT VP,
3400HT/3400HTVP, and 3500/3500VP
PERpH-X™ High Performance pH/ORP Sensors
May 2017
LIQ-PDS-PERpH-X
A sensor design that is flexible to meet your demanding pH/ORP application needs
Rosemount 3500 PERpH-X sensors are versatile sensors suited to meet a number of difficult application requirements. These high
performance sensors are ideal for use in high temperature and high pressure processes and feature a wide array of options to solve
your pH/ORP measurement needs.
Page 2
Rosemount PERpH-X Sensors
Overview
May 2017
Expanded Application Flexibility
Selectable reference electrolytes allow sensors to be used in a
large number of applications.
Electrolyte options include:
• High Temperature – most suitable for use in highly acidic, basic or
oxidative solutions, and high temperature applications.
• Bio-Film Resistant – inhibits the growth of bacteria and algae
• Poisoning Resistant – most suitable for use in processes containing
sulfides, mercaptans, and cyanides
• Oil Resistant – aimed for use in applications where light oils and
greases can foul a sensor
• Scaling Resistant – targets applications where the precipitation of
calcium magnesium salts like gypsum or water hardness coat over
the electrode
• Metal Resistant – most suitable for use in applications where
chloride in the reference electrolyte would react with the process
Minimize Maintenance and Lower Total Cost of
Ownership
A long lasting rebuildable reference junction and electrolyte
extends sensor life and allows sensors to be easily rebuilt
where normally other sensors would have to be replaced.
SMART preamplifier allows automatic recognition of pH
sensors by Rosemount transmitters: 1066, 1057, 1056,
and 56.
pH calibration data is stored which allows sensors to be
calibrated in advance for “Plug and Play” installations in the
field.
Versatile Mounting Options to Meet Your
Installation Requirements
Ryton body sensors feature 1 in. MNPT front and rear facing
process threads.
Titanium body sensors offered in 21 in. and 36 in. lengths
suitable for use with retraction assemblies.
Variopol (VP8) cable connection option, for quick
Rosemount PERpH-X High Performance pH/ORP Sensors feature various sensor body
options. Sensors are available in a chemically resistant Ryton plastic body
(3500P/3500VP) or titanium tube housing (3300HT/3300HTVP/3400HT/3400HTVP).
Rosemount 3500P/3500VP sensors are constructed with front and rear facing 1 in.
MNPT threads for insertion or submersion type installations. Rosemount
3300HT/3300HTVP/3400HT/3400HTVP) sensors must be used with a process
connector allowing for various insertion depths. These titanium body sensors also have
extended sensor length options allowing installation through a ball valve assembly.
Sensors feature a Teflon reference junction and a Pt-100 RTD for temperature
compensation. Rosemount PERpH-X pH/ORP sensors are available with either an
integral cable connection or Variopol (VP8) connector. Variopol cables sold separately.
Accessories ).
(see
Table 1. Rosemount 3300HT pH/ORP Sensor ordering information
ModelSensor type
3300HTpH/ORP sensor
Rosemount PERpH-X Sensors
Measuring electrode
10pH - GPHT glass
12ORP
O-ring material
30EPDM
31Viton
32Kalrez
Typical model number: 3300HT-10-30
www.Emerson.com/RosemountLiquidAnalysis
3
Page 4
Rosemount PERpH-X Sensors
Table 2. Rosemount 3300HTVP pH/ORP Sensor ordering information
ModelSensor type
3300HTVPpH/ORP sensor
Measuring electrode
10pH - GPHT glass
12
O-ring material
30EPDM
31Viton
32Kalrez
Preamplifier option
_No preamplifier
70SMART preamplifier
Typical model number: 3300HTVP-10-30-70
ORP
(1)
May 2017
1.Only available if selected with option 10.
4
www.Emerson.com/RosemountLiquidAnalysis
Page 5
May 2017
Table 3. Rosemount 3400HT pH/ORP Sensor ordering information
ModelSensor type
3400HTpH/ORP sensor
Measuring electrode
10pH - GPHT glass
12
ORP
Sensor length
2121 in. Titanium tube
2536 in. Titanium tube
O-ring material
30EPDM
31Viton
32Kalrez
Cable length
619.5 in. cable without BNC
6215 ft. cable without BNC
(1)
(2)
Rosemount PERpH-X Sensors
Typical model number: 3400HT-10-21-30-62
1.For use with sensor head junction boxes.
2.For wiring directly to transmitter or junction box.
www.Emerson.com/RosemountLiquidAnalysis
5
Page 6
Rosemount PERpH-X Sensors
Table 4. Rosemount 3400HTVP pH/ORP Sensor ordering information
ModelSensor type
3400HTVPpH/ORP sensor
Measuring electrode
10pH - GPHT glass
12
Sensor length
2121 in. Titanium tube
2536 in. Titanium tube
O-ring material
30EPDM
31Viton
32Kalrez
Preamplifier option
ORP
May 2017
_No preamplifier
70SMART Preamplifier
(1)
Typical model number: 3400HTVP-10-21-30-70
1.Only available if selected with option 10.
6
www.Emerson.com/RosemountLiquidAnalysis
Page 7
May 2017
Table 5. Rosemount 3500P pH/ORP Sensor ordering information
ModelSensor type
3500PpH/ORP sensor
Electrolyte selection
BFBio-film resistant
HT
MR
High temperature
Metal resistant
Rosemount PERpH-X Sensors
OR
PR
SR
Oil resistant
Poisoning resistant
Scaling resistant
Preamplifier/cable
01Preamplifier with 25 ft. cable
02No preamplifier with 15 ft. cable
(1)
Measuring electrode type
10pH – GPHT glass
12ORP
Reference type
21Double junction reference
O-ring material
30EPDM
31Viton
32Kalrez
Typical model number: 3500P-HT-01-10-21-30
1.Preamplifier is SMART if selected with option 10. Preamplifier is standard if selected with option 12.
www.Emerson.com/RosemountLiquidAnalysis
7
Page 8
Rosemount PERpH-X Sensors
Table 6. Rosemount 3500VP pH/ORP Sensor ordering information
ModelSensor type
3500VPpH/ORP sensor
Electrolyte selection
BFBio-film resistant
HT
High temperature
May 2017
MR
OR
PR
SR
Metal resistant
Oil resistant
Poisoning resistant
Scaling resistant
Preamplifier/cable
01Preamplifier
02No Preamplifier
(1)
Measuring electrode type
10pH–GPHT glass
12ORP
Reference type
21Double junction reference
O-ring material
30EPDM
31Viton
32Kalrez
Typical model number: 3500VP-HT-01-10-21-30
1.Preamplifier is SMART if selected with option 10. Preamplifier is standard if selected with option 12.
Titanium, ryton, teflon, glass, and user specified O-ring material (EPDM, Viton or Kalrez)
Reference electrode
Double junction with replaceable process side electrolyte and Teflon junction
Temperature sensor
Pt-100 RTD
Process connections
1 in. MNPT front and rear facing threads
Cable length
Rosemount 3500P option -0125 ft. Integral cable
Rosemount 3500P option -0215 ft. Integral cable
Rosemount 3500VPVP8 cable (sold separately)
Weight/shipping weight
1 lb/2 lb (0.5 kg/0.9 kg)
www.Emerson.com/RosemountLiquidAnalysis
11
Page 12
Rosemount PERpH-X Sensors
Product Certifications
Please see online certificates for further details.
IECEx
Sensors without preamp – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (pH only) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with standard preamp (ORP only) – Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +80 °C) or Ex ia IIC T5 Ga (-20 °C ≤ Ta ≤ +40 °C)
Per standards IEC60079-0: 2011, IEC 60079-11 : 2011
ATEX
Sensors without preamp – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with SMART preamp (pH only) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +60 °C)
Sensors with standard preamp (ORP only) – II 1 G Ex ia IIC T4 Ga (-20 °C ≤ Ta ≤ +80 °C) or – II 1 G Ex ia IIC T5 Ga
(-20 °C ≤ Ta ≤ + 40 °C)
Per standards EN 60079-0: 2012+A11:2013, EN 60079-11:2012
May 2017
FM
See online FM Certificate of Compliance for applicable sensor options:
Intrinsically Safe for use in Class I, II, and III, Division 1, Groups A, B, C, D, E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C
Intrinsically Safe for use in Class I, Zone 0, AEx ia IIC T6 Ta = -20 °C to +60 °C
Nonincendive for use in Class I, Division 2, Groups A, B, C, and D; Temperature Class T6 Ta = -20 °C to +60 °C
Suitable for use in Class II and III, Division 2, Groups E, F, and G; Temperature Class T6 Ta = -20 °C to +60 °C Hazardous (Classified)
Locations
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 °C - 1400332; Entity; I/0/AEx ia IIC/T6 Ta = 60 °C - 1400332; Entity; NI/I/2/ABCD/T6 Ta = 60 °C;
S/II,III/2/EFG/T6 Ta = 60 °C
Per standards 3600:1998, 3610:2010, 3611:2004, 3810:2005
CSA
See online CSA Certificate of Compliance for applicable sensor options:
Sensors with preamp – Intrinsically Safe:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ambient temperature
rating -20 °C to +60 °C; Ex ia IIC; T6
Sensors without preamp – Intrinsically Safe and Non-Incendive:
Class I, Division 1, Groups ABCD; Class II, Division 1, Groups EFG; Class III; Class I, Division 2, Groups ABCD; Ex ia IIC; T6; Ambient
temperature rating -20 °C to +60 °C: (Simple Apparatus)
The Emerson logo is a trademark and service mark of Emerson Electric Co.
Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.
All other marks are the property of their respective owners.
The contents of this publication are presented for information purposes only, and while
effort has been made to ensure their accuracy, they are not to be construed as warranties
or guarantees, express or implied, regarding the products or services described herein or
their use or applicability. All sales are governed by our terms and conditions, which are
available on request. We reserve the right to modify or improve the designs or
specifications of our products at any time without notice.
Analyticexpert.com
Facebook.com/Rosemount
Loading...