Rosemount Analytical 3300HT, 3300HTVP Specifications

Page 1
Product Data Sheet
00813-0100-3300, Rev AA
September 2019
Rosemount™ 3300HT/3300HTVP, 3400HT/ 3400HTVP, and 3500P/3500VP
PERpH-X™ High Performance pH/ORP Sensors
A sensor design that is flexible to meet your demanding pH/ORP needs
Rosemount PERpH-X sensors are versatile sensors suited to meet a number of difficult application requirements. Thse high performance sensors are ideal for use in high temperature and high pressure processes and feature a wide array of options to solve your pH/ORP measurement needs.
Page 2

Overview

Extended application flexibility
September 2019
Selectable reference electrolytes allow sensors to be used in a large number of applications.
Electrolyte options include: — High temperature: most suitable for use in highly acidic,
basic, or oxidative solutions and high temperature applications.
— Bio-film resistant: inhibits the growth of bacteria and
algae.
— Poisoning resistant: most suitable for use in processes
containing sulfides, mercaptans, and cyanides.
— Oil resistant: aimed for use in applications where light
oils and greases can foul a sensor.
— Scaling resistant: targets applications where the
precipitation of calcium magnesium salts like gypsum or water hardness coat over the electrode.
— Metal resistant: most suitable for use in applications
where chloride in the reference electrode would react with the process.
Minimal maintenance and lower total cost of ownership
A long lasting rebuildable reference junction and electrolyte extends sensor life and allows sensors to be easily rebuilt where normally other sensors would have to be replaced.
SMART preamplifier allows automatic recognition of pH sensors by Rosemount™ transmitters: 1066, 1057, 1056, and 56.
pH calibration data is stored, which allows sensors to be calibrated in advance for "plug and play" installations in the field.
Versatile mounting options to meet your installation requirements
Ryton body sensors feature 1-in. MNPT front and rear facing process threads.
Titanium body sensors offered in 21-in. (533,4 mm) and 36-in. (914,4 mm) lengths suitable for use with retraction assemblies.
Contents
Overview........................................................................................................................................................................................... 2
Ordering information........................................................................................................................................................................ 3
Specifications.................................................................................................................................................................................... 8
Dimensional and installation drawings............................................................................................................................................ 10
Accessories......................................................................................................................................................................................25
2 Emerson.com/Rosemount
Page 3
September 2019
Variopol (VP8) cable connection option, for quick cable-to-sensor release, eliminates cable twisting.

Ordering information

Rosemount™ PERpH-X™ High Performance pH/ORP sensors feature various sensor body options.
Figure 1: Rosemount 3500P pH/ORP Sensor
Table 1: Rosemount 3300HT Ordering Information
Option Description
3300HT pH/ORP sensor - PERpH-X High Performance
Measuring electrode
10 GPHT hemi pH glass, 0-14 pH
12 ORP
O-ring material
30 EPDM O-rings
31 Viton
32 Kalrez
Sensors are available in a chemically resistant Ryton plastic body (Rosemount 3500P/ 3500VP) or titanium tube housing (Rosemount 3300HT/3300HTVP/3400HT/ 3400HTVP). Emerson constructs Rosemount 3500P/3500VP sensors with front and rear facing 1-in. MNPT threads for insertion or submersion type installations. Rosemount 3300HT/3300HTVP/3400HT/3400HTVP sensors must be used with a process connector allowing for various insertion depths. These titanium body sensors also have extended sensor length options allowing installation through a ball valve assembly. Sensors feature a Teflon reference junction and a Pt-100 resistance temperature device (RTD) for temperature compensation. Rosemount PERpH-X pH/ORP sensors are available with either an integral cable connection or Variopol (VP8) connector. Variopol cables sold separately (see Accessories).
®
®
Cable
02 Without preamplifier, 15-ft. (4,6 m) cable
07 Without preamplifier, 4-ft. (1,2 m) cable
08 Without preamplifier, 10-ft. (3 m) cable
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
Note
The Rosemount 3300HT pH/ORP sensor is housed in a titanium tube with a replaceable reference junction and refilliable reference electrolyte. The sensor includes a Pt-100 temperature compensator. The sensor is available with a standard 15-ft. (4,6 m) cable. You can mount the sensor using a process connector, ordered separately. You can also order junction box kits with preamplifiers separately if the transmitter does not have an integral preamplifier within 15-ft. (4,6 m) of the sensor.
Rosemount PERpH-X 3
Page 4
Table 2: Rosemount 3300HTVP Ordering Information
Option Description
3300HTVP pH/ORP sensor - PERpH-X High Performance with Variopol connector
Measuring electrode
10 GPHT hemi pH glass, 0-14 pH
12 ORP
O-ring material
30 EPDM
31 Viton
32 Kalrez
Preamplifier option
_ No selection
70 SMART preamplifier (for use with -10 only)
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
September 2019
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
Note
The Rosemount 3300HTVP Sensor is available with an integral Variopol (VP 8.0) connector. A mating Variopol connector cable is required for use with these sensors. SMART preamplifier (-70) is the standard preamplifier option.
Table 3: Rosemount 3400HT Ordering Information
Option Description
3400HT pH/ORP sensor - PERpH-X High Performance Retractable
Measuring electrode
10 GPHT hemi pH glass, 0-14 pH
12 ORP
Sensor length
21 21-in. (533,4 mm) titanium tube
25 36-in. (914,4 mm) titanium tube
O-ring material
30 EPDM
31 Viton
32 Kalrez
Options
61 9.5-in. (241,3 mm) cable without BNC (for preamplifier options)
4 Emerson.com/Rosemount
Page 5
September 2019
Table 3: Rosemount 3400HT Ordering Information (continued)
Option Description
62 15-ft. (4,6 m) cable without BNC
07 Without preamplifier, 4-ft. (1,2 m) cable
08 Without preamplifier, 10-ft. (3 m) cable
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
Note
The Rosemount 3400HT High Temperature pH/ORP Retractable Sensor is housed in a titanium tube for use with a ball valve (order separately). You can mount the sensor with a ball valve or a process connector, both ordered separately.
Table 4: Rosemount 3400HTVP Ordering Information
Option Description
3400HTVP pH/ORP sensor - PERpH-X Retractable/Variopol Connector
Measuring electrode
10 GPHT hemi pH glass, 0-14 pH
12 ORP
Sensor length
21 21-in. (533,4 mm) titanium tube
25 36-in. (914,4 mm) titanium tube
O-ring material
30 EPDM
31 Viton
32 Kalrez
Preamplifier option
_ No selection
70 SMART preamplifier (for use with -10 only)
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
Rosemount PERpH-X 5
Page 6
September 2019
Note
The standard Rosemount 3400HTVP is offered with an integral Variopol connector.
Table 5: Rosemount 3500P Ordering Information
Option Description
3500P pH/ORP sensor - PERpH-X Insertion/Submersion
Electrolyte selection
BF Bio-film resistant
HT High temperature
MR Metal resistant
OR Oil resistant
PR Poisoning resistant
SR Scaling resistant
Preamplifier/cable
01 SMART integral preamplifier, 25-ft. (7,6 m) for Rosemount 1056, 1057, 1066, 56,
5081, 6081, and Xmt. (Standard preamp if used with ORP)
02 Without integral preamplifier, 15-ft. (4,6 m) cable
03 SMART preamplifier, 33-ft. (10 m) cable (standard preamp if used with ORP)
04 SMART preamplifier, 50-ft. (15,2 m) cable (standard preamp if used with ORP)
05 SMART preamplifier, 66-ft. (20 m) cable (standard preamp if used with ORP)
06 SMART preamplifier, 100-ft. (30,5 m) cable (standard preamp if used with ORP
07 Without preamplifier, 4-ft. (1,2 m) cable
08 Without preamplifier, 10-ft. (3 m) cable
Measuring electrode type
10 GPHT hemi pH glass, 0-14 pH
12 ORP
Reference type
21 Double junction reference
O-ring material
30 EPDM
31 Viton
32 Kalrez
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
6 Emerson.com/Rosemount
Page 7
September 2019
Note
The Rosemount 3500P Sensor is a versatile sensor platform for measuring pH or ORP. The rugged Ryton body and rebuildable reference electrode with front and rear facing 1-in. MNPT threads allows use in either insertion or submersion applications.
Table 6: Rosemount 3500VP Ordering Information
Option Description
3500P pH/ORP sensor - PERpH-X Insertion/Submersion
Electrolyte selection
BF Bio-film resistant
HT High temperature
MR Metal resistant
OR Oil resistant
PR Poisoning resistant
SR Scaling resistant
Preamplifier/cable
01 SMART integral preamplifier, 25-ft. (7,6 m) for Rosemount 1056, 1057, 1066, 56,
5081, 6081, and Xmt. (Standard preamp if used with ORP)
02 Without integral preamplifier, 15-ft. (4,6 m) cable
Measuring electrode type
10 GPHT hemi pH glass, 0-14 pH
12 ORP
Reference type
21 Double junction reference
O-ring material
30 EPDM
31 Viton
32 Kalrez
Calibration and conformance certificates - optional level
CC Certificate of Calibration (no test data given)
LC Loop Calibration Certificate (sensor and transmitter calibrated together, with test
data)
EC Electronic Calibration Certificate (sensor calibrated against factory instrument, with
test data)
Q1 Certificate of Conformance
Note
The Rosemount 3500VP uses a mating Variopol cable (purchased separately). The sensor is offered with six different gel electrolytes to match the application.
Rosemount PERpH-X 7
Page 8

Specifications

Table 7: Percent Linearity over pH Range
pH range HT series
0 to 2 pH 94%
2 to 12 pH 99%
12 to 13 pH 97%
13 to 14 pH 92%
Table 8: Rosemount 3300HT/3300HTVP pH/ORP Sensor Specifications
Measurement range
pH 0 to 14
ORP -1500 to +1500 mV
Operating temperature
Without preamplifier 41 to 311 °F (5 to 155 °C)
September 2019
With preamplifier Up to 212 °F (100 °C)
Storage temperature
14 to 138 °F (-10 to 70 °C)
Maximum process pressure
400 psig (2859 kPa [abs])
CRN rating 200 psig at room temperature
Wetted materials
Titanium, Ryton, Teflon™, glass, and user-specified O-ring matieral (EPDM, Viton®, or Kalrez®)
Reference electrode
Double junction with replaceable process side electrolyte and Teflon junction
Temperature sensor
Pt-100 resistance temperature device (RTD)
Process connections
Must use 1-in. compression process connector (PN 23166-00 or 23166-01)
Cable length
4 to 15 ft. (1,2 to 4,6 m) integral cable (Rosemount 3300HT) or VP8 cable for Rosemount 3300HTVP (sold separately)
Weight/shipping weight
1 lb./2 lb. (0,5 kg/0,9 kg)
Table 9: Rosemount 3400HT/3400HTVP pH/ORP Sensor Specifications
Measurement range
pH 0 to 14
ORP -1500 to +1500 mV
8 Emerson.com/Rosemount
Page 9
September 2019
Table 9: Rosemount 3400HT/3400HTVP pH/ORP Sensor Specifications (continued)
Operating temperature
Without preamplifier 41 to 311 °F (5 to 155 °C)
With preamplifier Up to 212 °F (100 °C)
Storage temperature
14 to 138 °F (-10 to 70 °C)
Maximum process pressure
400 psig (2859 kPa [abs])
CRN rating 200 psig at room temperature
Maximum pressure at retraction or insertion
21-in. (533,4 mm) length 64 psig (542 kPa [abs])
36-in. (914,4 mm) length 35 psig (343 kPa [abs])
Wetted materials
Titanium, Ryton, Teflon, glass, and user-specified O-ring matieral (EPDM, Viton, or Kalrez)
Reference electrode
Double junction with replaceable process side electrolyte and Teflon junction
Temperature sensor
Pt-100 resistance temperature device (RTD)
Process connections
Must use 1-in. compression process connector (PN 23166-00 or 23166-01). Can be inserted through a ball valve.
Cable length
4 to 15 ft. (1,2 to 4,6 m) integral cable or options 9.5-in. (241,3 mm) for use with sensor head junction box (Rosemount 3400HT)
VP8 cable for Rosemount 3400HTVP (sold separately)
Weight/shipping weight
1 lb./2 lb. (0,5 kg/0,9 kg)
Table 10: Rosemount 3500P/3500VP pH/ORP Sensor Specifications
Measurement range
pH 0 to 14
ORP -1500 to +1500 mV
Temperature range
41 to 248 °F (5 to 120 °C)
Storage temperature
14 to 122 °F (-10 to 50 °C)
Maximum process pressure
100 psig (790 kPa [abs])
CRN rating 40 psig at room temperature
Rosemount PERpH-X 9
Page 10
September 2019
Table 10: Rosemount 3500P/3500VP pH/ORP Sensor Specifications (continued)
Wetted materials
Titanium, Ryton, Teflon, glass, and user-specified O-ring matieral (EPDM, Viton, or Kalrez)
Reference electrode
Double junction with replaceable process side electrolyte and Teflon junction
Temperature sensor
Pt-100 resistance temperature device (RTD)
Process connections
1-in. MNPT front and rear facing threads
Cable length
Rosemount 3500P 4 to 100 ft. (1,2 to 30,5 m) integral cable. Maximum of 15 ft.
(4,6 m) for sensors without a preamplifier.
Rosemount 3500VP VP8 cable (sold separately)
VP8 cable for Rosemount 3400HTVP (sold separately)
Weight/shipping weight
1 lb./2 lb. (0,5 kg/0,9 kg)

Dimensional and installation drawings

Figure 2: Proper Sensor Installation Orientation
Install sensor within 80 ° of vertical.
10 Emerson.com/Rosemount
Page 11
September 2019
Figure 3: Rosemount™ 3500VP Sensor Dimensional Drawing
Rosemount PERpH-X 11
Page 12
Figure 4: Rosemount 3300HT Sensor Dimensional Drawing: Standard Hemi Bulb Sensor
September 2019
A. Sensor body B. 1-in. swage fitting process connector assembly PN 23166-00/01 C. 1-in. NPT
12 Emerson.com/Rosemount
Page 13
September 2019
Figure 5: Rosemount 3300HTVP: Standard Hemi Bulb Sensor
A. Cable with Variopol receptacle (female) connector B. Variopol plug (male) connector C. Sensor body
D. Retraction stop collar
Rosemount PERpH-X 13
Page 14
Figure 6: Rosemount 3400HT Sensor Dimensional Drawing (with and without 1½-in. Ball Valve Assembly)
September 2019
A. Dimension (see Table 11). B. Dimension (see Table 11). C. Dimension (see Table 11).
D. Cable.
E. Cable bushing polypropylene. F. 1-in. x 1-in. swage fitting kit (PN 23166-00 or 23166-01) required to connect sensor directly to process or to ball valve.
G.
CAUTION
Residual pressure and process may remain trapped between ball valve and male connector.
64 psig (option 21) 35 psig (option 25)
H. O.D. titanium housing.
I. O.D. retraction stop collar. J. Electrode housing Ryton.
K. pH electrode.
L. A process connector (PN 23166-00 or -01 must be used to connect the sensor to a ball valve kit 23240-00. (Process connectors sold
separately).
M. Ball valve kit (PN 23240-00) used with extended length retractable sensor.
N. 1½-in. FPT ball valve PN 9340065. O. 1½-in. MPT close nipple.
14 Emerson.com/Rosemount
Page 15
September 2019
P. 1½-in. FPT reducing coupling.
Q. Ball valve kit PN 23240-00 (optional).
Table 11: Rosemount 3400HT Dimensions
Option A (in./mm)
(1)
B (in./mm) C (in./mm)
21 21.6/549 14.0/355 23.1/587
25 36.1/917 28.5/724 37.6/955
Add five inches of length to dimension A if mounting a sensor head junction box onto the sensor.
(1)
Rosemount PERpH-X 15
Page 16
Figure 7: Rosemount 3400HTVP Dimensional Drawing (with and without 1¼-in. Ball Valve Assembly)
September 2019
A. Dimension (see Table 11). B. Dimension (see Table 11). C. Knurl retainer.
D. Packing adapter.
E. Ball valve.
F. 1¼-in. MPT close nipple. G. Ball valve kit PN 23765-00. H. Housing titanium.
I. Retraction stop collar. J. Electrode housing.
K. pH electrode.
L. Ball valve kit PN 23765-00 contains a 1¼-in. full port ball valve, 1¼-in. close nipple, and a retraction kit PN 23796-00 with carbon
graphite packing for easy sensor insertion and removal.
Table 12: Rosemount 3400HTVP Dimensions
Option A (in./mm)
(1)
B (in./mm)
21 21.6/549 12.2/310
25 36.1/917 26.7/678
Add five inches of length to dimension A if mounting a sensor head junction box onto the sensor.
(1)
16 Emerson.com/Rosemount
Page 17
September 2019
Figure 8: Rosemount 3300HT/3300HTVP Straight Flow Installation
A. 1½-in. pipe tee with 1-in. threaded connections PN 2002011
Figure 9: Rosemount 3300HT/3300HTVP Angle Flow Installation
A. 1½-in. pipe tee with 1-in. threaded connections PN 2002011 B. Flow
Rosemount PERpH-X 17
Page 18
Figure 10: Rosemount 3300HT/3300HTVP Pipe "Y" Installation
A. 1½-in. x 1-in. reducing bushing B. 1½-in. pipe "Y"
Note
Valves and fittings by others. Mount the sensor at least 10 degrees from horizontal.
September 2019
18 Emerson.com/Rosemount
Page 19
September 2019
Figure 11: Typical Retraction Mounting for the Rosemount 3400HT Sensor (Shown with Sensor Head Junction Box)
A. Weldalet (1½-in. FPT) B. Electrode C. Electrode housing tip
D. Flow
E. Pipe tee
F. Pipe "Y"
Note
Sensor must be mounted at an angle between 10 degrees and 90 degrees above the horizontal. Pipe tees and weldalets provided by customer.
Rosemount PERpH-X 19
Page 20
September 2019
Figure 12: Rosemount 3500P/3500VP Sensor with Jet Spray Cleaner (PN 12707-00) for Submersion Installations
Note
This accessory is especially useful for keeping the sensor clean in dirty ponds or tanks. You can mount it using a handrail mounting assembly or similar submersion assembly.
20 Emerson.com/Rosemount
Page 21
September 2019
Figure 13: Jet Spray Cleaner
A. Cleaning solution by others.
Rosemount PERpH-X 21
Page 22
B. Solenoid valve or manual valve (supplied by others). C. Corrosion resistant tubing (supplied by others).
D. Polypropylene ¼-in. compression fitting.
E. ¼-in. 316 stainless steel.
F. ¼-in. polypropylene. G. Stainless set screw for adjustable spray nozzle height. H. 2-in. NPT threads.
I. Sensor. J. 1-in. PVC coupling for submersible applications (supplied by others).
K. 1-in. PVC or stainless conduit (supplied by others).
L. Cable.
M. Timer supplied by others or use timer feature in Rosemount instrument.
Figure 14: Rosemount 3500P Sensor Flow-through Installations
September 2019
A. 1½-in. schedule 80 CPVC tee with 1-in. FNPT connections (code 16) straight flow shown. B. Always mount sensor at least 10 degrees above horizontal. C. 1½-in. pipe "Y".
D. 1½-in. x 1-in. reducing bushing.
E. Flow.
F. 1½-in. schedule 80 CPVC with view or cleaning port. G. Angle flow shown.
22 Emerson.com/Rosemount
Page 23
September 2019
Figure 15: Submersion Installation: Handrail Mounting Accessory (PN 11275-01)
A. Sensor cable. B. 1½-in. PVC pipe schedule 80. C. 1½-in. pipe clamp, three places.
D. 45 degree sweep ell extension pipe.
E. Regularly check to make sure connections are water tight.
F. Unistrut 1-in. x 1-in. aluminum. G. 1½-in. pipe clamp, two places. H. Can be any convenient dimension.
I. Customer handrail, two places. J. Mounting channel aluminum, two places.
K. Locking pin with bead chain.
Note
Unless otherwise specified.
Rosemount PERpH-X 23
Page 24
Low flow cell specifications
Figure 16: Low Flow Cell (PN 24091-00)
September 2019
A. Inches B. Millimeters
24 Emerson.com/Rosemount
Page 25
September 2019
C. Outlet
D. Inlet
Inlet and outlet connections are stainless steel and take ¼-in. OD tubing. Flow cell is polycarbonate with ¼-in. FNPT fittings.
Wetted materials
Flow cell ratings
Sensor threaded connection
Body and nut: polyester/polycarbonate Fittings: 316 stainless steel Seals: silicone
Temperature: 32 to 158 °F (0 to 70 °C) Maximum pressure: 90 psig (721 kPa [abs]) Flow rate: 2 to 5 gallons per hour (7,6 to 18,9 liters per hour)
24091-00: 1-in. NPT adapter

Accessories

Table 13: Connector Cable (Required for all First Time Installations)
Part number Description
24281-00 15-ft. (4,6 m) VP8 cable
24281-01 25-ft. (7,6 m) VP8 cable
24281-02 2.5-ft. (0,8 m) VP8 cable
24281-03 50-ft. (15,2 m) VP8 cable
24281-04 100-ft. (30,5 m) VP8 cable
24281-05 4-ft. (1,2 m) VP8 cable
24281-06 10-ft. (3 m) VP8 cable
24281-07 20-ft. (6,1 m) VP8 cable
24281-08 30-ft. (9,1 m) VP8 cable
Table 14: Ball Valve Assembly
Part number Description
23240-00 Ball valve assembly, 316 stainless steel, 1½-in. (process connector required)
23765-00 1¼-in. ball valve assembly
Table 15: Extension Cables (Requires a Remote Junction Box)
Part number Description
23646-01 Extension cable, 11-conduit, shielded, prepped, per foot
9200273 Extension cable, 11-conduit, shielded, unprepped, per foot
Table 16: Mounting Assemblies
Part number Description
11275-01 Handrail mounting assembly
12707-00 Jet spray cleaner
Rosemount PERpH-X 25
Page 26
Table 16: Mounting Assemblies (continued)
Part number Description
2002011 CPVC flow through tee, 1½-in. NPT process connection
24091-00 Cell, low flow, ¼-in. inlet and outlet
915240-03 Tee, flow-through, 2-in. PVC, ¾-in. NPT
915240-04 Tee, flow-through, 2-in. PVC, 1-in. NPT
915240-05 Tee, flow-through, 2-in. PVC, 1½-in. NPT
Table 17: O-rings for Process Connectors (Optional)
Part number Description
23594-01 O-ring, 2-214, EPDM, 4 each with tube
9550220 O-ring, 2-214, Kalrez®, for process connector
Table 18: Process Connectors (Required for all First Time Installations)
Part number Description
23166-00 Connector, 1-in. x 1-in. 316 stainless steel with O-ring groove
September 2019
23166-01 Connector, 1-in. x 1-in. titanium with O-ring groove
Table 19: Remote Junction Boxes
Part number Description
2002565 Mounting bracket kit
23555-00 Junction box, Rosemount 54/5081/1055/Xmt compatible preamplifier
Table 20: Sensor Head Junction Box
Part number Description
23709-00 Junction box, sensor head with preamplifier for Rosemount 54/3081
Table 21: Other Accessories
Part number Description
24231-00 High temperature reference kit
24231-01 Bio-film resistant reference kit
24231-02 Poisoning resistance reference kit
24231-03 Oil resistant reference kit
24231-04 Scaling resistant reference kit
24231-05 Metal resistant reference kit
24238-00 High temperature porous Teflon™ liquid junction
24238-01 Bio-film porous Teflon liquid junction
24238-02 Poisoning resistant porous Teflon liquid junction
24238-03 Oil resistant porous Teflon liquid junction
26 Emerson.com/Rosemount
Page 27
September 2019
Table 21: Other Accessories (continued)
Part number Description
24238-04 Scaling porous Teflon liquid junction
24238-05 Metal resistant porous Teflon liquid junction
24239-00 High temperature junction and Viton® O-ring kit
24240-00 High temperature junction and Kalrez O-ring kit
24250-00 Viton O-ring kit
24251-00 Kalrez O-ring kit
24270-00 EPDM O-ring kit
34116-00 Junction cap, Ryton
34017-00 Cap, high temperature guard pH/ORP molded PPS
9210012 Buffer solution, pH 4.01, 16 oz. (473,2 ml)
9210013 Buffer solution. pH 6.86, 16 oz. (473,2 ml)
9210014 Buffer solution, pH 9.18, 16 oz. (473,2 ml)
9210392 Reference fill gel high temperature, silica, 1 oz. (30 ml)
9210422 Metal resistant refill kit, 30 cc syringe (4-5 refills per syringe)
9210423 Oil resistant refill kit, 30 cc syringe (4-5 refills per syringe)
9210424 Scale resistant refill kit, 30 cc syringe (4-5 refills per syringe)
9210425 Poison resistant refill kit, 30 cc syringe (4-5 refills per syringe)
9210426 Bio-film refill kit, 30 cc syringe (4-5 refills per syringe)
R508-8OZ ORP standard, 475 mV, 8 oz. (236,6 ml)
Rosemount PERpH-X 27
Page 28
00813-0100-3300
Rev. AA
September 2019
GLOBAL HEADQUARTERS
Emerson Automation Solutions 6021 Innovation Blvd Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 F +1 952 949 7001 liquid.csc@emerson.com
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a P.O. Box 1046 CH-6340 Baar Switzerland
T + 41 (0) 41 768 6111 F + 41 (0) 41 768 6300 liquid.csc@emerson.com
Linkedin.com/company/Emerson-Automation-Solutions
Twitter.com/Rosemount_News
Facebook.com/Rosemount
Youtube.com/user/RosemountMeasurement
NORTH AMERICA
Emerson Automation Solutions 8200 Market Blvd Chanhassen, MN 55317
Toll Free +1 800 999 9307 F +1 952 949 7001 liquid.csc@emerson.com
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
T +971 4 811 8100 F +971 4 886 5465 liquid.csc@emerson.com
©
2019 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Singapore
T +65 777 8211 F +65 777 0947 liquid.csc@emerson.com
Loading...